Rochester Institute of Technology
RIT Scholar Works
Theses
Thesis/Dissertation Collections
5-3-1965
An experimental apparatus for measuring the
relative illumination of finite conjugate lenses
Claude Fedele
Leo Neumann
Follow this and additional works at:
http://scholarworks.rit.edu/theses
This Thesis is brought to you for free and open access by the Thesis/Dissertation Collections at RIT Scholar Works. It has been accepted for inclusion
in Theses by an authorized administrator of RIT Scholar Works. For more information, please contact
ritscholarworks@rit.edu
.
Recommended Citation
SENIOR
RESEARCH
PROJECT
AN
EXPERIMENTAL APPARATUS
FOR
MEASURING
THE
RELATIVE
ILLUMINATION
OF
FINITE
CONJUGATE
LENSES
by
Claude
Fedele
Leo
Neumann
ROCHESTER
INSTITUTE
OF
TECHNOLOGY
ROCHESTER,
NEW
YORK
Submitted:
TABLE
OF
CONTENTS
ABSTRACT
i
INTRODUCTION
1
OBJECTIVES
3
DESCRIPTION
OF
APPARATUS
3
CALIBRATION
OF
EQUIPMENT
12
EVALUATION PROCEDURE
:
. . .13
RESULTS
15
CONCLUSIONS
18
DISCUSSION
OF
RESULTS
18
ACKNOWLEDGEMENT
20
ABSTRACT
The
apparatusdescribed
is
designed
to
measure relativeillumination
offinite
conjugatelenses
with comparableaccu -||T[|
illT**""T"""1
racy
andreliability
than
existing
methods and provides agraphical output with
less
time
expendedfor data
aquisition. AAn
opto-mechanicalscanning
systemis
utilizesfor
trans-mitting
light
pulsesfrom
known
image
field
positionsto
aphotomultiplier
tube.
The
light
pulses aretransmitted
to
the
phototube via an optical system
consisting
of anarray
oflucite
rods(light
pipes) positionedin
the
image
plane ofthe
lens
undertest.
The
rodsterminate
at a cylindricalhousing
which contains a
rotating
lens
prismassembly
which scansthe
end of each
lucite
rodin
sequence and relaysthe
illumination
values
to
the
photomultiplier.The
output current ofthe
tube
is
displayed
on a cathoderay
tube
oscilloscope and photographed,Data
obtained withthis
apparatus -was comparedto
data
obtained
using
photographic photometry.Results
indicate
that
the
apparatusis
very
repeatable andthat
data
aquisitiontime
INTRODUCTION;
A
number of photographic systems produce excellent qualitativeresults.
By
suitable calibration ofthe
various components ofthe
system,
the
information
gathering
capability
ofthe
systemcan
be
greatly
increased.
One
ofthe
calibrationsfrequently
encountered
in
the
field
of photographicinstrumentation
involves
the
measurement of relativeillumination
orthe
amount ofillumin
ation
"fall
off"in
the
image
plane of a photographic objective.This
decrease
in
illuminance
- atthe
off axis positionsmay
be
due
to
such causes asbarrel
vignetting,
absorption and cosinevariations.
Relative
illumination
is
defined
asthe
ratio ofthe
illuminance
atthe
focal
plane,
for
off-axisfield
positions,
to
the
illuminance
for
the
center ofthe
field.
This
assumesthat
the
luminance
ofthe
objectfield,
as observedfrom
the
lens,
is
the
samethroughout
the
field,
orthat
the
field
is
aLambert's
law
surface.Relative
illumination
is
specified asthe
per centof axial
illuminance
for
image
points at given angulardistances.
Two
methods of measurement exist.Extended
Source
Method
This
methodis
based
onfilling
the
lens
withlight
from
anextended uniform source of adequate
size,
placedin
-.the objectplane of
the
lens.
The
extended source shouldbe
uniformly
bright
over
the
useful areato
within 3$.A
photodetectoris
displaced
laterally
to
the
positioncorresponding
to
the
required angularpositions and
the
corresponding
percentage of axialillumination
are plotted of p|r/e,cent
illuminance
vsfield
angle ordistance
from
the
axis.Densitometric
Method
:In
this
method,
the
photoelectricdetector
is
replacedby
a photographic emulsion.Exposures
are made andthe
films
processed with control, strips which were exposed
in
a sensitometer.
The
negatives aredensitometered
atthe
positionscorresponding
to
the
required angular positions.The
illuminance
is
determined
from
a calibration curvederived
from
the
sensitometric
data.
A
curve of per centilluminance
versusdistance
from
axisis
plotted.Both
ofthese
methods aretime
consuming
andthe
accuracy
of results
doubtful
when equipmentis
adaptedto
obtaindata.
Recording
ofdata
andplotting
of curvesbecomes
tedious
when a great number oflenses
are evaluated.The
proposed apparatusOBJECTIVES;
To
design,
construct and evaluate a systemfor
measuring
the
relativeillumination
offinite
conjugatelenses.
This
apparatus will measure relative
illumination
with equal orbetter accuracy than existing
methods with a considerablereduction
in
data
aquisitiontime.
DESCRIPTION
OF APPARATUS;
cy^Pr.,ri>&A
The
apparatusis
ampri-.&e4 offive
maincomponents;
1.
Light
Source
2.
Optical
Bench
3.
Scanning
System
4.
Multiplier Photometer
5.
Readout
Equipment
Light
Source;
The
requirementfor
a maximum ofJ%
variationin
luminance
acrossthe
objectformat
was achievedby
using
three
high
intensity
projectionlamps
in
a plane parallelto
adiffusing
screen comprised offlashed
opal and groundglass.The
sourceis
enclosedto
preventstray
light
from
reaching
the
scanning
system,
and air cooledby
two
fans.
This
arrang
ement resulted
in
a sourceuniformly
bright
to
within2%
overa six
inch
diagonal.
(See
Figure
1)
Optical
Bench;
The
opticalbench
consists of alens
holder
andlight
source
holder
which slidealong
two
parallel sixfoot
long
steel ways.The
scanning
system andimage
plane are mountedpermanently
on one end ofthe
bench,
sothat
changesin
magnification are
facilitated
by
moving
the
light
source andlens
relativeto
the
scanning
system.(Refer
to
Fig.
2)
Scanning
System;
The
scanning
system consists of anarray
oY nineteendiameter
lucite
rods mounted apartin
the
image
plane of
the
lens
undertest.
The
image
diagonal
is
6.25"long.
The
light
pipes serveto
transmit
the
light
from
the
image
plane
to
the
inner
wall of a cylindrical aluminumhousing.
The
aluminumhousing
contains arotating
lens-prism
assembly
which
transmits
the
light
from
the
end of eachlight
pipein
sequence,
to
a photomultipliertube.
The
scanneris
driven
by
a
synchronous,
6
RPM
motor,
resulting
in
a scan rate across ..the
image
plane of4
seconds,
with one complete scanevery
10
seconds.
It
is
mountedin
aball
bearing
racefor
rigidity
and smooth operation.(Refer
to
Fig.
3)
Lens
#1
(
6mm,,
f;1.2
)
forms
animage
ofthe
light
pipe exitface
atthe
entrance pupil oflens#2.
Lens
#2
(
50mm,
f:2
)
produces an enlarged
image
ofthe
exit pupil oflens#l
atthe
photocathode of
the
photomultipliertube,
ensuring
that
the
entire surface ofthe
photocathodeis
uniformly
illuminated.
A
folded
optical pathis
usedto
keep
the
apparatus as compactas possible*
(in
principle)
X
o
z
<
O
K-Q.
o
g
o v.H
ED
^S^^
OPTICAL
BENCH
Fig.
2a-Overall
view of systemA
groundglassis
located
in
the
image
plane,
just
below
the
light
pipearray
to
facilitate
focussing
and measurement ofmagnification.
A
reticle ofknown
dimensions
mounted atthe
light
sourceis
usedin
this
operation.(Refer
to
Fig.
3
)
Photomultiplier Photometer;
A
photomultiplier photometer was constructedfor
maximumsensitivity
andfor
the
ability
to
substitute phototubes ofdifferent
spectral sensitivities.The
direct
currentsupply
voltages
for
the
photomultiplier are providedfty
afull
waverectified power supply.
(For
circuit,
referto
Fig.
4)
Voltages
for
eachdynode
andfor
the
anode are providedby
equally
spacedtaps
on a voltagedivider
network acrossthe
rectified power supply.The
photomultipliertube
(RCA
931
-A)features
a combination ofhigh
photosensitivity,
high
secondary
emission and small
D.C.
dark
current.The
spectral response coversthe
rangefrom
approximately
3000
to
6200
angstroms,
with a peak at
4000
angstroms.The
output current ofthe
931A
is
alinear
function
ofthe
exciting
illumination
undernormal
operating
conditions.(Refer
to
Fig.
4a)
Readout
Equipment;
The
output ofthe
photomultiplier wasdisplayed
on acathode
ray
tube
oscilloscope.A
chart recordermay
alsobe
used,
if
a|
RPM
motoris
substitutedfor
the
6
RPM
motor.The
outputis
displayed
as voltage vshorizontal
sweep
time,
8
Image
Plane
Drive
Motor
Front
Surface
Mirror
SCANNING
APPARATUS
10
CO -p tH o > <H O CO ri w S Si o co O-P O -P o a o O CM CM ri u (D B fn O CO C CtiVO Jh I Eh P C s &, o o EnVoltage
Divider-10-20000/v,
1
watt1-esistors
PHOTOMETER
AND
READOUT
EQUIPMENT
11
Fig.
4a-Photomultiplier
Tube
Housing
12
Calibration
ofEquipment;
,n
6tu.,&
Considerable
variation was encounteredin
the
-^l4mAgmeexirt-i4g
"from
the
light
pipesdue
to
such causes as unequal(rod)
length,
losses
due
to
surfaceimperfections,
variationin
polishat
the
light
pipeends,
andlosses
atthe
radii wherethe
pipesenter
the
cylindricalhousing.
This
variation was minimizedby
utilizing
sections of continuous neutraldensity
wedges mountedover
the
image
plane end ofthe
light
rod array.With
the
light
source mounted close
to
the
image
plane,
the
continuous wedgeswere moved across each of
the
light
pipes,
one at atime,
whileobserving
the
output signal onthe
oscilloscope.Density
valuesranged
from
.10to
.56.Using
this
method,
the
illumination
exiting
from
each ofthe
pipes was matchedto
k%.
The
linearity
ofthe
phototube output was checkedby
adjusting
the
output ofthe
tube
sothat
a value of100^
wasobserved on
the
oscilloscope graticule andthen
introducing
known
neutraldensity
valuesinto
the
optical path.The
resulting
transmission
values were comparedto
calculated values andindicated
that
there
was no measureabledeparture
from
linear
ity.
,o-^xX^
The
combined s^aoiiity-ofthe
photomultiplier,
powersupply,
oscilloscope,
andlight
source showed adrift
(in
per cent:,illuminance
value)
of not morethan
li%
over a period often
minutes when all components were operatedfrom
a constant13
EVALUATION
PROCEDURE;
Method
ofData
Collection;
Electronic
Method
(scanning
apparatus);The
lens
to
be
tested
was placedin
the
lens
holder,
set at
the
required magnification(1;1)
andfocussed.
The
light
source,
photometer,
and oscilloscope wereturned
onfor
a
warm-up
period often
minutes.The
scannerdrive
motor wasturned
on andthe
output signal ofthe
photomultiplierdisplayed
on
the
oscilloscope.By
varying
eitherthe
input* voltageto
the
photomultiplier or
the
scale attenuators onthe
oscilloscope,
the
maximumdeflection
was adjustedto
100^.
(The
outputcurrent which
is
being
displayed
is
directly
proportionalto
the
illumination
level.)
The
oscilloscopetrace
was photographedon
Polaroid
material and processed,according
to
the
manufacturer'sinstructions.
(Refer
to
sampledata,
Fig.
5)
Densitometric
Method;
The
lens
was placedin
a view camera and set atthe
required magnification
(1;1)
andfocus
setting.The
uniformsource was photographed on
Panatomic-X
sheetfilm.
The
exposures chosen resulted
in
densities
that
fell
onthe
straightline
portion ofthe
characteristic curve.The
film
was storedfor
a period oftwelve
hours
priorto
processing,
allowing
the
latent
image
to
stabilize.The
six replicatesfor
eachlens
and a control
strip
were processedtogether
to
minimizeprocessing
variability.
The
film
wasdeveloped
for
k^
minutesin
Kodak
DK-50
developer
at68
F
using
A.S.A.
agitation,
dried
and14
SAMPLE
DATA
for
ELECTRONIC
METHOD
Fig.
5a-Oscilloscope
Trace-No
lens
in
systemFig.
5b-Relative
Illumination
for
6"f;4.5
lens
15
illuminance
atthe
sensitometerstep
wedge.(Refer
to
appendixI)
Densities
atthe
requiredimage
field
positions were convertedto
illuminance
values andtabulated.
Replicated
data
from
both
methods was averaged and plotted.(Refer
to
Fig.
6a, 6b)
The
resulting
curves were comparedstatistically,
using
the
densitometric
method as a standard.Variance
was calculatedto
determine
the
repeatability ofthe
two
methods.(Refer
to
Appendix
II,
III)^RESULTS
;1)
The
relativeillumination
asdetermined
by
the
densitomet
ric method and
by
the
electronic methodis
asfollows;
3"
f;1.9
Lens
/Distance
L
in
ii
Densitometric
Electronic
100
100
_2_92
92
80
79
ll
59
39
34
2i
11
>e
from
axisinches
Relative
Illumination
6"
f;4.5
Lens
0
2i1
ii
2
2*
3
Densitometric
100
99
98
96
93
89
8o
Electronic
100
99
98
97
94
91
86
ibi
stancefrom
axisv~
in
inches
K
Relative
Illumination
2)
The
variance of each ofthe
two
methodsis
asfollows;
3"
f;1.9
Lens
Densitometric; s =
169
Electronic s =
0.60
6"
f;4.5
Lens
\
Densitometric;
s^=58.4
Electronic
; s2=O.35
3)
The
data
aquisition and evaluationtime
for
both
methodsare as
follows;
Figure
6a
Relative
Illumination
of 6"f;4.5
lens
16
Figure
6b
Relative
Illumination
for
3"f;1.9
lens
17
JI
r~~*
' 5"
OFF
AXIS
DISTANCE IN
INCHES
A' Mil
18
CONCLUSIONS;
1.)
Data
aquiredusing
the
scanning
apparatusis
comparable
to
data
obtainedby
the
densitometric
method.2.)
The
apparatus exhibits ahigher
degree
of repeatability
than
the
densitometric
method.3.)
Data
aquisitiontime
is
reducedby
afactor
ofsix
to
ten
times.
DISCUSSION
OF
RESULTS;
Inspection
ofthe
relativeillumination
curves revealsthat
in
onecase,
(6"f;4.5 lens)
the
apparatus producesdata
that
are somewhathigher
than
the
densitometric
methodindicates,
while
in
the
case ofthe
3"f;
1.9
-lens,the
reverseis
true.
Statistical
tests,
however,
failed
to
detect
any
significantdifference
between
the
two
methods.A
possible explanationfor
this
is
that
most ofthe
electronicdata
falls
well withinthe
two
sigmalimits
marked onthe
densitometrically
derived
curves.
(Refer
to
Fig.
6a,
6b)
The
variability
ofthe
densitometric
data
is
very
high
when compared
to
the
electronicdata.
This
can,
in
part,
be
explained
by
the
nature of photographic photometry.When
measuring
relativeillumination,
only
a small portion ofthe
D-log
E
curve
is
utilized,
andrelatively
smalldensity
differences
must
be
detected.
There
are a great number of stepsin
the
photometric
process,
eachstep
introducing
a certain amount ofvariability.
Since
a smalldensity
difference
resultsin
arather
large
changein
transmission,
one would expectthat
19
The
scanning
apparatusis
simpleto
operate and servesto
minimizethe
number of steps involvedin
the
aquisition ofdata.
The
chancesfor
error(especially
human
error) are minimized
and considerable savingsin
time
are realized.1.
Department
ofDefense,
USA,
"Military
Standard
PhotographicLenses,
MIL-STD-150A"(1950)
2.
Radio
Corporation
ofAmerica,
"Data
Sheet-931-A
Multiplier
Phototube','
(1958)
3.
Jerome
Li,
"Introduction
to
Statistical
Inference','The
Science
Press
Inc.,
Pennsylvania,
(1957)
4.
"Symbols,
Definitions
andTables",
Industrial
Statistics
20
ACKNOWLEDGEMENTS
The
authors areindebted
to
professorsB.
Carroll,
A.
Rickmers,
W.
Shoemaker,
andH.
Todd
for
their
cooperationand guidance
in
the
execution ofthis
project.Furthermore,
we wishto
thank
Elgeet
Optical
Co.
for
the
loan
ofthe
oscilloscope andlight
source.The
help,
adviceand encouragement received
from
Robert
Volk
andKen
Hood
ofthe
R.I.T.
Mechanical
Department,
Richard
Neumann,
Lenore
Fedele,
Alfonso
Nazzaro,
Robert
Nobilini
andJames
Savage
wasAPPENDIX
APPENDIX
II
STATISTICAL
ANALYSIS
;X
Test
for
Goodness
ofFit
- 3"ftl.
9
lens
Hls
xe^xp
V
Xe=*p
Xe
*P
1
100
100
2
93
91
3
80
79
4
62
63
5
32
39
6
8
12
Xe
=Averages
of ElectronicData
X
=Averages
ofDensitomet-^
ric
Data
Using
the
generalformula
(Xe
Xp)
we obtain;
^V
ioo
91
80
63
39
12
X
=2.39
^C
=11.07
(book
value)y~S,os
On
the
basis
ofthe
above evidence, we state with95%
confidencethat
there
is
nodifference
between
the
two
methods.A
similartest
onthe
6"
f;4.5
data
yieldedthe
following
results;X-
=0.645
4.
"YL
=11.07
(book
value)Calculation of Standard Deviation
General
formula\
n-l-\2
?(*-*).
Calculated Values
for
3"
f
:19Densitometric Method; Electronic Method ;
s =
13
s =
0.775
Calculated
Values
for
6"
f;4.5
Densitometric Method;
Electronic Method :
s =
7.6
APPENDIX
III
F Test;
General
Formula;
F
=The
following
values were calculated:3"
ftl.
9
lens
2
p
=(13)
=281
Cal (.775)2
6"
f;4.5
lens
F
=(7>*>)2
=
168
alC
U590)2
Pbook
=^
5,5,
.052
la.
4
On
the
basis
ofthe
above evidence, we can state with95$
confidence,
that
there
is
a significantdifference
between
the
variancesof
the
two
methods.The
variance ofthe
photographic methodis