JOURNAL OFCLINICAL MICROBIOLOGY,June 1982,p.1111-1119
0095-1137/82/061111-09$02.00/0 Vol.15,No.6
Accuracy and Precision of the Autobac System for Rapid
Identification
of
Gram-Negative
Bacilli: a
Collaborative
Evaluation
A. L.
BARRY,"*
T. L.GAVAN,2
P. B.SMITH,3
J.
M.MATSEN,4
J. A.MORELLO,5
AND B. H.SIELAFF6t
UniversityofCalifornia-DavisMedical Center, Sacramento, California 958171;TheCleveland Clinic Foundation, Cleveland, Ohio
441062;
Centersfor Disease Control, Atlanta, Georgia303333; University ofUtahSchoolof Medicine, Salt Lake City, Utah 841324;Universityof Chicago Medical Center, Chicago, Illinois606375;andPfizer Inc., Groton, Connecticut063406
Received 23 November 1981/Accepted 28 January 1982
Gram-negative
bacilli were
identified within 3 to 6 h by determining
susceptibil-ity to 18 different antibacterial agents in the Autobac I system and by applying a
two-stage quadratic discriminant analysis to the susceptibility
patterns.
The
Autobac system was compared with standard reference methods for identifying
glucose nonfermenters
and glucose
fermenters. Intralaboratory and
interlabora-tory
precision
of the Autobac system was comparable to that of the reference
methods. Sensitivity (accuracy) and specificity of the two systems were also
comparable, although there were some differences with certain species. Autobac
responses were considered to be equivocal (needing additional tests) if the relative
probability
of an accurate
identification was <0.70. Only 5% of 2,889 strains
produced such equivocal results; a similar number of strains gave low probability
levels with the
reference methods. When the two systems disagreed, an
indepen-dent
reference
laboratory
arbitrated,
confirming 49% of the Autobac responses
and
36%
of the reference
identifications.
With
equivocal
responses excluded, the
overall
accuracy of the
Autobac
system was
95.3%
compared
with
95.9%
for the
reference method. The respective accuracy estimates would be 93.8% and 93.1%
if
all
first-choice identifications
were
evaluated.
In
general, previous
attempts to automate or
mechanize
identification
of
microorganisms
have been
accomplished by
simply adapting
traditional methods
for
use
in a mechanized test
system. In
1973, Friedman and
MacLowry
(5)
reported
acomputer-assisted
system
for
identifi-cation of bacteria based
on ananalysis
of the
pattern
of
susceptibility
to commonantimicrobi-al agents.
Sielaff
etal.
(8) mechanized this
approach
by
using
susceptibility
testresults
ob-tained after 3 to 5 h with the Autobac I system
(Pfizer
Diagnostics, Groton, Conn. [now
the
property
of
General
Diagnostics,
Morris
Plains,
NJ
07950]).
By
applying
aquadratic
discriminant
function
technique for
dataanalysis,
they
ob-tained
a97% correlation with conventional
iden-tification
procedures.
Matsen et al.(personal
communication)
have
examined
thepossibility
of
utilizing
awide
variety
of
antibacterial agents,
other
than the commonchemotherapeutic
agents, for
thepurpose of bacterial
identifica-tion. After
extensive
screening
of
potentially
useful
agents, 18 different antibacterial agents
t Present address: Minnesota Mining and Manufacturing Co.,St.Paul,MN 55144.
were
selected because of their
discriminatory
capabilities. Sielaff
etal.
(9) describe
a systemfor
rapid (3-
to6-h) identification
ofgram-nega-tive
bacilli, based on
analyzing the patterns of
susceptibility to
18different
antibacterial agents.
This
system used the Autobac
Isystem fitted
with a programmed computer which will
per-form
atwo-stage
quadratic
discriminant
analysis
of the
susceptibility
patterns.
The
present
report documents the
accuracy
and
precision
of the
proposed
Autobac system
for
rapid identification of
gram-negative
bacilli.
For this
evaluation,
standard
reference methods
were used to
confirm the identification of each
isolate. The
accuracy of
thereference methods
selected for this
collaborative
study
was alsoestimated
by
furthertesting
selectedstrains
atthe Centers for Disease
Control,
Atlanta,
Ga.
The
precision of each identification system
wasevaluated
by
testing
92isolates in
triplicate
in
each of five
independent
laboratories.
MATERIALSANDMETHODS
Thestudyconsisted oftwophases. The firstphase wasdesignedtodocumentprecision (reproducibility) of the reference methodsand of the Autobacsystems. 1111
on February 7, 2020 by guest
http://jcm.asm.org/
1112 ET AL.
The second phase consisted oftesting fresh clinical isolates and stock culturestodocument thespecificity andsensitivity of the Autobac system, in comparison withreference methods.
Bacterial isolates.For the firstphaseof thisstudy,92 stock cultureswereselectedatthe Centers forDisease Control. The identification of each isolatewas recon-firmed, and each strainwassuspended in rabbit blood, dispensed into 15 different Durhamtubes, and frozen at -70°C. The cell suspensions were identified only with randomly selected code numbers. Each of the fiveparticipating laboratories received 276 coded cell suspensions representing three tubes of each stock culture. Personnel performing these tests were not awareof the initialidentificationsorof which suspen-sions wereduplications. The frozensuspensions sent toeach testing laboratory were thawed and subcul-turedjust before being identified with the reference methods and with the Autobac system. Subsequently, bothinterlaboratoryandintralaboratory reproducibili-tyof bothidentification systemswasevaluated. This culture collection included 3Escherichiacoli,2 Kleb-siella pneumoniae, 3 Klebsiella ozaenae, 3 Proteus mirabilis, 3 Proteus vulgaris,4Morganella morganii, 3Providenciastuartii, 2 Providencia rettgeri,4 Salmo-nellaenteritidis, 2 Arizonahinshawii, 3Shigella son-nei, 3Serratiamarcescens,4Enterobacter aerogenes, 3Enterobactercloacae,3Enterobacteragglomerans, 3 Hafnia alvei, 3 Citrobacterfreundii, 3 Citrobacter diversus, 3Edwardsiellatarda, 3 Yersinia enterocoli-tica, 3 Yersiniapseudotuberculosis, 3Acinetobacter calcoaceticus subsp. anitratus, 3 Aeromonas hydro-phila, 2 Moraxellalacunata,2Flavobacterium menin-gosepticum, 2Alcaligenes odorans, 3 Pseudomonas aeruginosa, 3 Pseudomonasmaltophilia,3 Pseudomo-nas cepacia, 2 Pseudomonas putidalfluorescens group,3Pseudomonasputrefaciens,and 3 Pseudomo-nasstutzeristrains. Thelaboratories failed to report all 1,380pairs ofidentifications: 1,253 reference identifi-cations and 1,240 Autobac identifiidentifi-cations were avail-able foranalysis. The unreportedidentifications were randomly distributed, not likely to introduce signifi-cantbias to the data.
The secondphase of this study involved tests with recent clinical isolates and a few stock cultures, in-cludedasrepresentatives of the less common species. Thefour clinicallaboratoriesparticipating in this study tested 2,889isolates with standard reference methods and withthe Autobac system. For each isolate, both testsystems wereinitiated at the same time, and the resultsofall testswere stored in a central computer facility. Subcultures of 653 strains were submitted to the Centers for Disease Control forarbitration: 327 strains were identified asbelonging to the same spe-cies with theAutobac and reference test systems and 326strains haddiscrepant identifications with the two independent systems. The former 327 strains were submitted forarbitration because the initial program gave discrepant results, but they were found to be in agreement when the modified program that is de-scribed in this report was applied to the data. The first experimental program is not described because it provedtobeinaccurate.
Ten quality control strains were tested by both identification systems at approximately biweekly in-tervals during the second phase of this study. The control strains included E. coli, K.pneumoniae, P.
vulgaris,
M.morganii,
S.marcescens, C.diversus,P. aeruginosa, P. putrefaciens, M. lacunata, and A. calcoaceticussubsp.
antitratus. These strains wereselectedto
provide
atleastonepositiveresponseandone
negative
response toeach of the individualtests included in bothidentification systems.Autobacidentificationsystem. Thetestsystem evalu-ated in this study has been described in detail
by
Sielaffetal. (9). Briefly,eachisolatewassubcultured toblood agar and MacConkey agarplates. The next day, the following information was recorded: (i) growthonMacConkey agar, (ii)lactosefermentation
on
MacConkey
agar,(iii)presence ofprecipitatedbile aroundcoloniesonMacConkeyagar,(iv)spot oxidase test(6) results, (v)
spot indoletest(10) results,and(vi) swarming growthonblood agar. These six pieces of informationwereentered into thecomputerized identi-fication system before the Autobac cuvettes wereread. To inoculate a cuvette, one or more isolated coloniesweresuspendedinphosphate-bufferedsaline, and the suspension was then adjusted to a standard turbidity (ca. 107 colony-forming units per ml),
by
usingtheAutobacphotometer. For each 13-chamber cuvette, 2 ml ofinoculumsuspensionwasaddedto18 ml ofEugonbroth(BBL Microbiology Systems, Cock-eysville, Md.).In the present study,two13-chamber cuvettes were inoculated, and 24 different elution disksweretested.Afterourdatawereanalyzed, the 18 elution disks described by Sielaff et al. (9) were
selected, and the identifications presented here are basedonlyontheresults oftestswiththose 18disks. A19-chambercuvettewillsoonbe available fortesting the 18elutiondisks, providingonecontrolchamber.
Onceinoculated,thecuvettes wereallowedto incu-bate at 35°C in an Autobac incubator-shaker. The cuvettes were all read after 3 h of incubation. If sufficientgrowthwas notinitiated in the control cham-ber,thecuvettes werereincubated and readat
hourly
intervals for thenext3h. Ifsufficient growth hadnot been obtained after 6 h ofincubation, the test was aborted, and the strain was retested the next day. Repeated failure togrow within 6 h was a very rare occurrenceand when this didhappen, the strainwas removed from our study. As soon as satisfactory growthwasobtained in the controlchamber,turbidity in each test chamber wasmeasured, and the results were automatically entered into the computeralong with thepreliminarytestdata notedearlier. The pro-gramthen providedafirst- andsecond-choice identifi-cation along with the relative probability (R.P.) that expresses theconfidence with which one may accept each reported identification. A lowR.P. indicates that additional tests are needed toconfirm the identifica-tion; it mightrepresent a species that is not in the data base.
Referencemethods.Conventional tubed media were usedtoobtainareferenceidentification. These media (ScottLaboratories, Fiskeville, R.I.) were essentially identical to those used at the Centers for Disease Control for identification ofEnterobacteriaceae (3) or byG. L.Gilardi (personalcommunications) for identi-fication of theglucosenonfermenters. The media used toidentify glucosefermenters andnonfermenters are described more completely below. Most tests were read after 24 h and, ifnegative, they were read again after 48 h ofincubation.
o-Nitrophenyl-P-D-galacto-pyranoside tests were readafter 20min and 1, 2, 4, and J.CLIN. MICROBIOL.on February 7, 2020 by guest
http://jcm.asm.org/
AUTOBAC ID SYSTEM 1113 48 h or until a positive reaction was seen.
Voges-Proskauer, phenylalanine deaminase, and indole tests were performed after 48 h of incubation only, since externalreagents had to be added. The results of all tests were entered into the computer, which was programmed with percentage figures obtained from the appropriate sources (primarily from the Centers for DiseaseControl,EntericBacteriology Section, and from G. L. Gilardi). The latter computer program was developed to identify only the genera or species in-cluded in the Autobac data base, using rather tradi-tional methods. When the two systems disagreed, many more discriminatory tests were performed, as needed, for final identification of each isolate. This arbitration work was performed exclusively at the Centers for Disease Control by P. B. Smith, D. L. Rhoden, and A. 0.Esaias.Arbitration was performed with 653 isolates for which discrepant identifications were initially obtained by Autobac and reference methods. The Autobac program wasrevised at the end ofthis study, and when the revised program was used, 327of the 653 Autobac identifications were in agree-mentwith the initial referenceidentifications. Only the revised program is evaluated in this report.
Statistical analysis. Forthe purpose of this report, the term "accuracy"indicates agreement with a refer-ence identification (reference methods orarbitration tests when available). "Precision" is used inter-changeablywith"reproducibility"toindicate repeat-ability ofa particular test response. Interlaboratory and intralaboratory reproducibilities are each ex-pressed as "reproducibility index," rather than per-centageof responses in agreement with modesorwith theexpected responses. Thisallowed comparisonof allpossible pairsof data and expresses theproportion of datapairsthatwerein agreement(4,7).Some data did notpermitthe selection ofamodalresponsetobe used asanindex of agreement; thussomedata could not be analyzed in the traditional manner. Further-more,thereproducibilityindex couldeasily accommo-datemissing data,i.e., ifonlytwoof three responses were reported, one pair of data was available for analysis. "Sensitivity" ofan identification system is defined as the percentage of strains within a given speciesthatwereaccurately identified.
"Specificity,"
on the otherhand, examines the number of timesa
given species identification wasreported bythe test systemand expresses the percentage of times those reportswerein agreement with the reference identifi-cation(accurate).
C)C)Cx
. _C.
-0
C)
n 5
C) .
--o. °.n
o_.F _
,-C U)C
C
0X
to) -I
0X-C)oC
C) 0.
- U) RESULTSPrecision of referencetests.Tenqualitycontrol strainswereeach tested 29 and 41 differenttimes against the appropriate battery of reference tests,thus providinganopportunity toestimate theprecisionof each of the individualtest
proce-dures. False-positiveorfalse-negativereactions
wererecordedwith about 2to3% ofmost tests (Tables 1 and2). Overall precision variedfrom 93.5% forM.morganiito99.2%for A. calcoace-ticus. Additional estimates of precision were
obtained from replicate tests that were per-formed with the larger collection of isolates
_,
II
w+ A¢,+ " w
0%
-.
00 s.t-4
00 -+ + + ( + + + +-+ +- + I l- +
I + + I -l
+ + + w+ +I.-&4+ 1-- t-i+ + + + I -+
+ + ) +t
+ ,_"+
'IO A I-10% 4 t-'O w
o- + w -"
+ WJI
I + +..
+- + +
+
" I +-I
o-I +I.+ Iw-+Wv+o,
+w I I+ + +
w$ 00
.. .1 PO
-en~ >A ao w xo
VOL. 15, 1982
,0n
C)^ C2 ze z r' C) c c <i _ c ° o), c (1cti
r C) . 0 C) 0C)
C_ co C) 00 CD-U) -9 C) _.C)
_.0z
00O
C) 0C)
3-9 _. C) 3 -9a'
0-_. c D A r) D B B D 3 rv i2 CY. w 2 D. 3 D ?I 10 A 2 D 10 a. (c l< --I, o O I=FC),.
00 o x1a"-'
0.
C)
-CD, 2?. a' C) C) I CD,A' o
10 w ~-+ +son February 7, 2020 by guest
http://jcm.asm.org/
TABLE 2. Precisionof reference methods used for identification ofglucose-nonfermenting gram-negative bacilli'
Most commonresultband no. of errors recorded with: Overall Control strain
(no.of timestested) Mo- Oxi- Argi- H2S 10% 1% NO3 Growth precision
tility dase nine (KIA) Lactose Glucose togas at 42°C (%)
P.aeruginosa(40) + + + - - + + +
1 2 98.8
P.putrefaciens (33) + + - + - - +
4 1 3 1 3 94.5
A.calcoaceticus (32) - - - - + + - +
1 1 99.2
M.lacunata(14)d - + - - -
-2 98.2
Precision%C 95.8 100 99.2 100 100 96.6 96.6 95.0 97.9
aReplicatetestswithfour control strains tested in four
laboratories;
seeTable 3 fordescription
oftests.bPositive(+)ornegative (-), "errors"areeither
false-positive
orfalse-negative
results.cPrecisionexpressedaspercentageoftestswith
expected
results.dThecontrol strain ofM.
lacunata
astestedrepeatedly
inonly
twolaboratories.included in the first
phase of this
study.
Table
3describes the
individual
testsperformed and lists
the
precision estimated
on the basis of 15repli-cate tests
with each of 92 strains. In that series
of tests, false-positive or false-negative results
with
one or morereference
testsresulted in
amisidentification
(13.8%)
of the
1,380
identifica-tions
recorded
orin
an"unable
toidentify"
response
(3%);
83.2% of
the1,380
identifications wereconsidered
correct. The relativeprobabili-ty
of
an accurateresult
(R.P. value)
waslow
(<0.70) with 28% of incorrect
responses butwith only
3% of the
correctidentifications.
When low R.P.
values
areobtained,
erroneousreference
testsmaybe
suspected, and additional
testing is needed before
afinal
identification
canbe made with confidence.
Intralaboratory and
interlaboratory
reproduc-ibility. The
precision of Autobac identifications
was
established and compared with
thatof
thereference methods by
examining
results of the
first phase
of this study. Both interlaboratory
and
intralaboratory precisions
were expressed asreproducibility
indexes rather than
percent-agesof
responses in agreement with the modal response (as done in Tables 1 through 3). With this approach, all possible pairs of responses arecompared
(4,7).
For example, when three re-sponses arebeing
compared, there will be threepairs, i.e., first
and second, first and third, and second andthird. If
the three identifications were E. coli, E. coli, and E. cloacae, therepro-ducibility index
would be 0.33 because only oneof
thethree pairs (33.3%)
was in agreement. In moststudies
to date, such reports would have beenconsidered 66.7%
reproducible (two of three correct when compared with the mode).The
reproducibility
ratio is statistically morevalid
because
all
randomly occurring situations areconsidered.
Itpermits inclusion
of all data,evenwhen some data are
missing
or when theresults
aresuch that
amodal
response cannotbe
established
for
estimating precision. However,
the
reproducibility
index
provides figures
which
appear
much
lower than
percent agreementfig-ures
that
arecustomarily
quoted. Because the
Autobac and
reference results
wereanalyzed in
the
same way,the reader
is urged
to comparereproducibility
indexes
recorded for the
twomethods and
not tobe
concerned with their
absolute
magnitudes. Each reproducibility index
was
calculated by
dividing the number of pairs in
agreement
by the total number of pairs being
compared (Table 4).
Intralaboratory precision ranged from
0.67 to0.84
for the Autobac
systemcompared
with
0.58and
0.88
for the reference method.
When datafrom all laboratories
werecombined,
intralabor-atory
reproducibility
indexes
werenearly
com-parable. The
chi-square
testdemonstrated
nosignificant differences between the
reproducibil-ities
of the
twomethods
(X2
=1.62,
P >0.20).
Interlaboratory reproducibility indexes
were0.82
for both methods. Combined
ratios,
calcu-lated
by comparing
allpossible pairs
of respons-es, were 0.74and
0.76for the Autobac and
reference
methods,
respectively.
Because over8,800
pairs of
responses werecompared in
thisanalysis,
thedifference between
the twoindexes
was
statistically significant
(P < 0.05). Howev-er, thedifference is probably
toosmall
tobe of
clinical
importance.
Accuracy of identifications. When the
identifi-cation obtained
with the Autobac systemdis-agreed with
that obtained with the referencemethod, independent
arbitration was used todetermine
whether either of the methods was correct.Since
entirely different
approaches are used inthe
two systems to achieve anidentifica-tion, both
systems are not likely to be in error at 1114 BARRY ET AL.on February 7, 2020 by guest
http://jcm.asm.org/
AUTOBAC ID SYSTEM 1115 TABLE 3. Description and estimated precision of conventional tests utilized to obtain reference
identifications
Test Precision
(%)M
Test Precision(%)OGlucosenonfermenters
Ornithine
decarboxylase, MoellerH2S, Kligler
iron agar(KIA)slants 98.1 basalmedium .97.5Oxidasespottest, 1%
tetramethyl-p-
DNase,toluidine blue medium 97.2 phenylenediamine ... 97.5ONPG'
test forP-D-galactosidase
Arginine dihydrolase, Moeller basal
activity
.96.4broth
.96.2
Sucrosefermentation, peptone brothNitrate reduction togas, indole withAndrade indicator .96.1 nitrate broth .95.6
Xylose fermentation,
peptone broth10%o Lactose oxidation, purple agar withAndrade indicator.96.1 slants ... 95.4 Adonitol fermentation, peptone
Motility, hanging droppreparation. 93.7 broth withAndrade indicator .... 95.9
Glucose, HughandLeifson
OFb
Arabinose fermentation, peptonebasalmedium... 88.5 broth withAndrade indicator .... 95.4
Growthat42°C, tryptic soy agar Lactosefermentation, peptone broth slants... ... 74.9 withAndrade indicator... 94.6
Flagellum stains, selected strains Voges-Proskauer (VP) test, MR-VPe
only
...-c broth and Barrittreagent... 94.3Phenylalanine deaminase (PD),
Glucose fermenters
phenylalanine
agarslants .94.2Oxidase spot test, 1%tetramethyl-p- Salicinfermentation, peptone broth
phenylenediamine
... 99.4 with Andradeindicator... 92.5Lysinedecarboxylase, Moellerbasal Motility test medium with
triphenyl-medium... 98.3 tetrazolium chloride... 89.8
Indoleproduction, peptonebroth Malonateutilization, malonate agar
and Kovacreagent... 98.1 slants... 86.9
H2S,triple sugariron(TSI) agar Ureaseactivity, Christensen urea
slants... 97.7 agarslants... 86.3
a Precision of each conventional testexpressedas thepercentageof correct reactions noted whentriplicate tests were performed with 92 strains (66 glucose fermenters and 26 nonfermenters) in each offive separate laboratories.Fifteentest results were obtained with eachstrain,and the mostcommonresponsewasacceptedas thecorrect result forestimating precision.
bOF, Oxidation-fermentation.
c, Insufficient datato estimateprecision.
dONPG,
o-Nitrophenyl-p-D-galactopyranoside
test.'MR-VP, Methylred-VogesProskauer.
the
sametime,
andif they are, they are unlikely toprovide the
same erroneousidentification.
For that
reason, weassumed that the reference
method
wascorrectwhen it
confirmed the
Auto-bac
response,thus
allowing
us toevaluate the
accuracy
of the reference method without
arbi-tration of
anexcessive
number
of strains.
The data inTable
5confirm the
validity
of
thisassumption, i.e.,
when
both
systemsagreed,
323
of
327identifications
werefound
tobe
accurate.When the
two testsdisagreed,
arbitration
con-firmed 49o of the Autobac
responsesand
36% of the reference identifications.With both
testsystems,the computerprintout
provided
an R.P.value which indicates the
confidence that
maybe
attached to thefirst-choice
identification. The
response maybecon-sidered
equivocal if
alow
R.P. valueis
listed;
i.e.,
additional
tests areneeded before
the finalidentification
can bemade with confidence.
When
estimating
the accuracyof
atest system,equivocal
responsesshould
be excluded sincethey
canbe
neither
accurate norinaccurate.
Withother commercial systems, about 5 to 10% of the strains tested have been found to be equivocal (1, 2). The practical utility of a test system is diminished if more than 10% of re-sponsesindicateaneedfor additional confirma-tory tests.
The data in Table 6 were accumulated to determine how the accuracy of both systems was affected by excluding strains with various R.P.values. With bothtestsystems, elimination ofidentificationswithlowR.P. values excluded more erroneousresponses than correct identifi-cations, thus increasing the overall accuracy. Both systemswere
approximately
97%accurate ifallresponseswith R.P. <0.95 wereexcluded, but 11 and19% of the strains would be consid-eredequivocal. Withthe Autobac system, only 8.5% ofthe strains had R.P. values of <0.80, and 96.2% of theremaining
identifications were considered accurate. However, a significant number(about
25%)
of Enterobacterspp. would beexcludedbecause the R.P.was <0.80. Fewer Enterobacter spp. would be excluded if only VOL. 15,1982on February 7, 2020 by guest
http://jcm.asm.org/
TABLE 4.
Intralaboratory
andinterlaboratory reproducibility'
ofAutobac identificationcompared
with that ofidentificationsobtained with reference methodsPairs inagreement/totalno.ofpairs ReproducibilityindeXb Comparison
Autobac Reference Autobac Reference
Intralaboratory
Lab1 211/257 220/255 0.82 0.86
Lab 2 211/251 213/257 0.84 0.83
Lab 3 162/211 125/214 0.77 0.58
Lab 4 211/258 219/263 0.82 0.83
Lab 5 177/263 232/264 0.67 0.88
All laboratories 972/1240 1,009/1,253 0.78 0.81
Interlaboratoryc
675/826 699/857 0.82 0.82Combination'
6,516/8,809 6,733/8,823 0.74 0.76aFive laboratories tested 92 strains, each in
triplicate,
yielding
1,380 responses for each method; fewer responseswereavailableforanalysisbecause somelaboratoriesfailedtoreport alltest results.bNumber ofpairs inagreement divided
by
number ofpossiblepairs.Comparing
the mostfrequent
identification reported by each of five laboratories; when onelaboratory
reportedthree differentresponses,one wasrandomly selected forcomparison.
dOverall
combination ofintralaboratory
andinterlaboratory precision;
eachidentificationwascomparedwith all otheridentifications. For eachstrain, 105possible
pairsofresponseswerecompared.R.P. values of <0.70 were considered to be
equivocal. Thiswould excludeonly5%of allof
the responses and would result in an overall
accuracy of 95.3%. Consequently, we
recom-mend that first-choice Autobac responses be
accepted only if the R.P. value is .-0.70. When wedidthat,theoverallaccuracyof theAutobac
system wascomparabletothat of the reference
tests.
Sensitivityandspecificity.Thesensitivityofan
identification system may be defined as the
percentage of strains within a given species or
group ofspecies that was accurately identified
(agreedwith the referenceidentification).Onthe
otherhand, specificitydesignatestheconfidence
that may begiventoaparticular species
identifi-cationreportedbythe systemunderevaluation.
For example, 94 strains of C. freundii were
tested with the reference tests, but only 55%
were accuratelyidentified (sensitivity),whereas
94% of the C.freundii identifications obtained
TABLE 5. Resultofarbitrationtests comparedwith
first-choice identificationsinitially obtained with reference and Autobacsystemsin theparticipating
laboratories
Initial Autobac andreference
Arbitrationtests identification:
confirmed
Disagreed Agreed
Referencetests 118a _b
Autobactests
159-Neither 49 4
Both -323
Total 326 327
aDatapresentedas number of strains in each
cate-gory.
b-,Nonepossible.
with
the referencetestswereaccurate(specific-ity).
Further,
the referencemethods
were97%
sensitive
butonly
88%
specific
for A. calcoaceti-cus: allbut 3 of the 101 A.calcoaceticus
strains werecorrectly identified,
but otherspecies
were alsobeing
misidentified
asA.calcoaceticus.
Thecalculated
sensitivity
and
specificity
for
each ofthe
microbial
species
included
in the secondphase
of thisstudy
arepresented
in Table 7. Allfirst-choice identifications
obtained wth the ref-erence method may be contrasted with those obtained with theAutobac
system.
Inaddition,
the
percentages of strains
within eachspecies
that would be
excluded because
the R.P. valuesTABLE 6. Estimatedaccuracyof reference and Autobacidentifications: effect of
excluding equivocal
identifications,
basedonthe R.P. that each response wasaccurate%Excludedateach %Accurateaafter R.P. R.P. level excludingtests excluded
Referenceb Autobac Referenceb Autobac
Nonec -_c - 93.1 93.8
<0.60 4.1 3.0 95.4 95.2
<0.70 5.2 5.0 95.9 95.3
<0.75 6.3 6.5 96.2 95.6
<0.80 7.1 8.5 96.4 96.2
<0.85 7.8 11.1 96.6 96.5
<0.90 9.2 15.0 97.0 96.9
<0.95 11.2 19.3 97.4 97.4
aAccuracywasjudgedasagreementwith the refer-ence identifications or arbitration tests (when
avail-able).
"Eighteen
strains that could notbeidentifiedwith the initial referencetestswereexcludedwhencalculat-ing
accuracy.cEvaluation of allfirst-choice
identification,
regard-less ofthe R.P. valuegiven;-,none.
J. CLIN. MICROBIOL.
on February 7, 2020 by guest
http://jcm.asm.org/
AUTOBAC ID SYSTEM 1117
TABLE 7. Sensitivity andspecificity of the reference methods and of Autobac identifications (before and after excluding equivocal results with R.P. below 0.70)
% Sensitivity (accuracy) of: %Specificityc of:
No.of Autobac A
Referenceidentification strains responses Reference Autobac responses Reference Autobac responses tested excluded testb All Excluding testb All Excluding R.P. <0.70 methods tests R.P. < 0.70 methods tests R.P. < 0.70
A. calcoaceticus 101 0.0 97.0 97.0 97.0 88.2 97.0 98.0
Aeromonas sp. 38 0.0 97.4 100.0 100.0 92.5 100.0 100.0
Alcaligenes sp. 17 17.6 64.7 88.2 92.9 78.6 83.3 86.7
C.diversus 56 3.6 98.2 96.4 98.1 91.7 96.4 98.1
C.freundii 94 21.1 55.3 80.0 91.9 94.5 72.4 75.6
E. tarda 7 0.0 85.7 100.0 100.0 100.0 58.3 70.0
E. cloacae 229 14.8 89.4 83.0 83.9 96.7 90.9 94.7
E. aerogenes 173 6.4 98.8 91.3 93.8 96.6 95.2 96.2
E. agglomerans 33 15.2 87.5 75.8 75.0 43.8 64.1 70.0
E. coli 414 4.3 97.6 95.4 97.5 97.6 97.8 99.2
Flavobacterium sp. 13 0.0 38.5 100.0 100.0 83.3 86.7 86.7
H.alvei 39 5.1 87.2 92.3 94.6 87.2 85.7 89.7
K.pneumoniae 312 4.2 97.8 96.5 97.6 98.0 94.1 96.7
Klebsiella (other species) 9 22.2 77.8 77.8 100.0 63.6 70.0 70.0
Moraxella sp. 4 0.0 100.0 100.0 100.0 25.0 66.7 66.7
M.morganii 100 3.0 96.9 97.0 97.9 90.5 94.2 94.9
P. mirabilis 304 1.6 98.3 99.3 100.0 100.0 98.7 99.3
P. vulgaris 64 1.6 96.8 89.1 88.9 90.9 100.0 100.0
ProvidencialP. rettgeri 91 4.4 63.3 96.7 98.8 100.0 89.8 94.4
P. cepacia 9 0.0 77.8 77.8 77.8 100.0 70.0 77.8
P. maltophilia 57 3.5 71.9 94.7 98.2 95.3 93.1 93.1
P.putidalfluorescens 30 6.7 96.6 93.3 92.9 84.8 96.6 96.3
P. stutzeri 6 0.0 100.0 83.3 83.3 66.7 100.0 100.0
P. aeruginosa 308 1.0 98.7 97.4 98.7 99.3 98.7 99.0
Pseudomonas(otherspecies) 10 0.0 50.0 66.7 60.0 41.7 50.0 75.0
SalmonellalArizona 94 4.3 95.7 92.6 96.6 92.8 93.5 94.5
Serratia sp. 185 1.6 98.9 97.3 97.8 95.8 100.0 100.0
Shigella sp. 71 4.2 97.2 93.0 94.1 72.6 95.7 95.5
Y.enterocolitica 18 16.7 94.1 88.9 93.3 72.7 69.6 77.8
Y.pseudotuberculosis 3 33.3 33.3 33.3 50.0 33.3 100.0 100.0
Total 2,889 5.0 93.1 93.8 95.3
aAutobac responses were evaluated with and without excluding equivocal results, R.P. < 0.70. The
percentage of strains within eachspecieswithequivocalresponses islisted.
bExcluding 18 strains that couldnotbeidentified by the initial reference tests, includingallotherfirst-choice
identifications.
cSpecificityindicates theconfidence thatcanbeplaceduponagivenspecies identification, i.e.,thenumberof correctresults dividedbythe numberof times eachspecies wasreported.
for Autobac
responses were<0.70
are noted inTable
7.By
excluding
such
equivocal
identifica-tions,
thesensitivity
and
specificity
of
theAuto-bac
systemweresomewhat
improved.
In
addition
to the clinical isolatesreported
here,
tests wereperformed
with 10isolates
belonging
tospecies
thatare notincluded in the Autobac program(6
Achromobacterxylosoxi-dans,
3Pasteurella
multocida,
and 1Bordetella
bronchiseptica
isolate).
TheAchromobactersp.isolates
were all misidentified as Pseudomonas sp., as was the Bordetella sp. The Pasteurella sp. isolatesweremisidentifiedas Moraxella sp., E.agglomerans,
and Edwardsiella sp. All of thosemisidentificationswerereported
with R.P. valuesof
.0.70
and thus would not be consid-eredequivocal
identifications.
DISCUSSION
The Autobac
system for
rapid
identification of
gram-negative
bacilli
represents
aunique
ap-proach
tobacterial taxonomy based
onthepat-terns
of
susceptibility
tovarious antibacterial
agents.
Continuing
efforts to findnon-chemo-therapeutic
agents
that can be used in such asystem
should
improve
thereliability
of
thisapproach.
Resistant variantsmight
appear insome
environments,
leading
toatypical
suscepti-bility
patterns
whchmight
leadtomisidentifica-tions. Because there
should
be noselective
pressure for variants that become
resistant
to thenon-chemotherapeutic
agents,
efforts
should be made toreplace
thetherapeutic
agents
thatare
currently
included in thesystem.
VOL. 15, 1982
on February 7, 2020 by guest
http://jcm.asm.org/
The taxonomic groups that were used to de-velop the Autobac
identification
system were limited to 30species
orgroupsof species that are likely to be found in clinicalspecimens.
No attempt was made toidentify species
of Aetro-monas, Alcaligenes, Fl/atobacteriuln,
Morax-ella, Serri-atia, orSlhigella. In most clinicalsitua-tions, identifications
tothegeneric
level will besufficient. Serological
confirmation
of Sailinioniel-la spp. andShigella
spp. would be anecessary supplement to this system.Sailmonella
sp. andAr-izona
sp. are notdistinguished,
but many hold that Arizona sp. should beplaced
in the genus Sallmonella. P.rettugeri
is notdistinguished
from P.stiuartii
(ureapositive
and ureanegative).
Failure to make that distinction
currently
pre-sents noimportant
clinicalproblem.
K.pnieia-mnoniae
and K. oxytoca are notseparated,
but that could beaccomplished by
simply referring
to the spot indole test
(10).
K.o.xvtoca being
indole
positive.
Furthermore, the Autobac com-puterprogramdoes
notyetidentify
three recent-ly recognized species (Enterobacter sakaz-akii, E.gergo'ii,
andCitrobacteramnalonaticis).
Ourlimited
experience
withtesting
a few species that are notincluded
in the current program underscores thepossibility
that othermisidenti-fications
can occur withuncommonly
encoun-tered species.In spite of these
relatively minor limitations.
the Autobac systemcanrapidly
identify
the vast majority of gram-negative bacilli found in clini-cal material. Thesensitivity.
specificity.
andprecision
of the Autobac systemare comparable tothecorresponding
characteristics of the refer-encetestsystem. However, theAutobac system is much morerapid (3to 6hversus 48h).
and themechanization
and computer-assistedinterpre-tation minimize
thetechnologist
time required to obtain reliable results.In most studies of this nature, a new test system is normally compared directly with a
reference
system, and anydiscrepancies
are assumed to represent errors on the part of the new system. The present study provided a unique opportunity to evaluate the reference method, as well as the Autobac system. When the twodisagreed,
the Autobac system was in error about half of the time, and the reference methods were in error a little more than half of the time (Table 5). Although everyeffort
was madetostandardize and to control the reference methods,false-positive
and false-negative reac-tions were obtained with about 2 to3%
oftheindividual
tests. Occasionally, the erroneous testresults were important enough to result in amisidentification
and, subsequently, a disagree-ment with the Autobac system. For the same reason,false-positive
or false-negative results with the Autobac system could result inerrone-ousidentifications which
disagree
with the refer-ence methods. For that reason, it is easy to understandwhy
correlation between twoinde-pendent
identification systemsrarely
exceeds 90 to95%, depending
upon theprecision of thetwo systems. An even greater number ofdiscrepan-cies
might
be expected if conventional tests rather than standard reference methods had been used for evaluating the Autobac system. For example, conventional tests may represent standard tubed media read afterovernightincu-bation,
accepting some loss of precision and accuracy for the sake of convenience. Such procedures are not appropriate reference meth-ods forevaluating new identification systems.By excluding
testswithalowprobability ofan accurate response, a significant proportion of misidentifications were eliminated. The overall accuracy of both systems was as great as 97% when all responses with R.P. levels of <0.95 were excluded. But that would have required supplementary tests with 11 or 19% of the iso-latesincluded inthe secondphase of thisstudy.
We concluded that Autobac identifications with R.P. values of .0.70 could be accepted and those of <0.70 should be confirmed with supple-mentarytests. Only 5% of the isolates included in this study would require such confirmation, and the overall accuracy of both the Autobac and the reference methods was 95 to 96%. Sensitivity and specificity of the two methods varied somewhat with different species, but, in general, the Autobac system was as sensitive and specific asthe reference method.
In summary, the Autobac system
provides
a unique approach to the rapid identification of gram-negative bacilli. The overall results indi-catethat the system is just assensitive, specific, and precise as the standard reference methods. Themechanized system, with computer-assistedinterpretation,
requires aminimum
amount of technologist time and provides reliable results within 3 to6h.ACKNOWLEDGMENTS
We express oursinceregratitude to the following microbiol-ogists whoperformed thestudies described in this report: R. Aaron. C. William Bacon. Robert E. Badal. Ann 0. Esaials, B. B. Gardner. Michael H. Graves. C. Knapp, Dwane L.
Rhoden,and M.Telenson. We alsoacknowledgethe invalu-able assistance provided by microbiologists in the Enteric Bacteriology Section of the Centers for Disease Control and by G. L.Gilardi,whoprovided certainunpublisheddata that wereneeded forestablishingthedata base used for
interpreti-tionof thestandardreference tests.
LITERATURE CITED
1. Barry,A.L., and R. E. Badal. 1979.Rapid identification ofEnterobocterioaoea with the Micro-ID system versus API 20E and conventional media. J. Clin. Microbiol. 10:293-298.
2. Barry, A. L., R. E. Badal, and L. J. Effinger. 1979.
on February 7, 2020 by guest
http://jcm.asm.org/
AUTOBAC ID SYSTEM 1119
fication of Enterobocteriaceae in frozen microdilution
traysprepared by Micro-Media Systems. J.Clin. Micro-biol. 10:492-496.
3. Ewing, W. H.,andB. R. Davis. 1975.Media andtestsfor
differentiationofEnterobocterioceae. U.S. Department of
Health, Education and Welfare. National Communicable
DiseaseCenter. Atlanta. Ga.
4. Fleiss, J. L. 1971. Measuring nominal scale agreement
among manyraters. Psychol. Bull. 76:378-382. 5. Friedman, R., and J. MacLowry. 1973.Computer
identifi-cation of bacteriaonthe basis of their antibiotic
suscepti-bilitypatterns. AppI. Microbiol. 26:314-317.
6. Kovacs,N. 1956.IdentificationofPseilonionaos
ae(lruinl-osaby the oxidase reaction. Nature(London)178:703.
7. Light, R.J. 1971. Measures ofresponse agreement for qualitative data: some generalizations and alternatives.
Psychol. Bull.76:365-377.
8. Sielaff, B. H., E. A. Johnson, and J. M. Matsen. 1976.
Computer-assistedbacterial identification utilizing
antimi-crobial susceptibility profiles generatedby Autobac 1. J. Clin. Microbiol. 3:105-109.
9. Sielaff, B. H., J. M. Matsen, and J. E. McKie. Novel approachtobacterial identification thatusesthe Autobac
system.J. Clin. Microbiol. 15:1103-1110.
10. Vracko, R., and J. C. Sherris. 1963. Indole-spot test in
bacteriology. Am.J.Clin.Pathol. 39:429-432.
VOL. 15 1982