Relationship between canine transthoracic
impedance and defibrillation threshold.
Evidence for current-based defibrillation.
B B Lerman, … , C W Clark, O C Deale
J Clin Invest.
1987;
80(3)
:797-803.
https://doi.org/10.1172/JCI113136
.
The electrical parameter used to define defibrillation strength is energy. Peak current,
however, may more accurately reflect the field quantities (i.e., electric field strength and
current density) that mediate defibrillation and therefore should be a better clinical descriptor
of threshold than energy. Though transthoracic impedance is a major determinant of
energy-based threshold and is sensitive to operator-dependent changes in impedance
(electrode-subject interface), an ideal threshold descriptor should be invariant with respect to these
changes in impedance. We therefore compared the relative invariance of energy- and
current-based thresholds when transthoracic impedance was altered by one of two
methods: (a) change in electrode size (protocol A) or (b) change in electrode force (protocol
B). In protocol A, impedance was altered in each dog by a mean of 95%. Energy thresholds
determined at both low and high impedance were 44 +/- 21 J (mean +/- SD) and 105 +/- 35
J, respectively, P less than 0.0001. In contrast, peak current (A) thresholds were
independent of transthoracic impedance, 22 +/- 5 A (low impedance) vs. 24 +/- 6 A (high
impedance), P = NS. Energy and current thresholds showed a similar relationship for
animals tested in protocol B. Therefore, current-based thresholds, in contrast to energy
thresholds are independent of operator-dependent variables of transthoracic impedance
and are invariant for a given animal. These results suggest […]
Research Article
Relationship between Canine
Transthoracic Impedance
and Defibrillation Threshold
Evidence for
Current-based Defibrillation
Bruce B. Lerman, Henry R.Halperin, Joshua E. Tsitlik,KennethBrin, ChristopherW.Clark,and0.CarltonDeale Cardiology Division, Department ofMedicine, The University ofVirginiaMedical Center, Charlottesville, Virginia22908
Abstract
Theelectrical parameter used to define defibrillation strength is energy. Peak current, however, may more accurately reflect the
field
quantities (i.e.,electricfield
strength and currentden-sity)
thatmediatedefibrillation
and therefore should be a bet-terclinical
descriptor of threshold than energy. Thoughtrans-thoracic
impedance is a major determinant of energy-based threshold andis
sensitive
to operator-dependent changes in impedance (electrosubject interface), an ideal thresholdde-scriptor should be invariant with
respect to these changes in impedance. We therefore compared the relative invariance of energy- andcurrent-based
thresholds when transthoracicim-pedance
was alteredby
oneof
two methods: (a) change inelectrode size
(protocol A)
or(b)
change in electrode force (protocol B). In protocol A, impedance was altered in each dog by a meanof 95%.
Energy thresholds determined at both low andhigh impedance
were44±21 J(mean±SD) and 105±35J,
respectively, P < 0.0001. In contrast, peak current (A)
thresh-olds were
independent of
transthoracic impedance, 22±5 A (low impedance) vs. 24±6 A (high impedance), P= NS. En-ergyand
currentthresholds
showed asimilar relationship
foranimals tested
inprotocol
B.Therefore,
current-basedthresh-olds,
in contrast to energy thresholds areindependent of
opera-tor-dependent variables of transthoracic impedance
and are invariantfor a given animal. These results suggest thatrede-fining defibrillation
threshold in terms of peak current rather than energyprovides
asuperior method
ofdefibrillation.
Introduction
Considerable
controversyexists regarding the electrical
param-eterthat
best describes
defibrillation threshold.
Thestandard
clinical
practice
is
todeliver
ashock
calibrated in units
of
energy.
However, field
quantities
such asmyocardial electric
field strength and
currentdensity actually describe the
electri-cal
parametersof defibrillation (1).
Peak current may moreprecisely reflect these field quantities than
energy(for
adamped sine
wavepulse).
Identification of
theappropriate
threshold
descriptor
has significantclinical
importance
since
asuprathreshold shock
cancauseMB-creatine kinase
release(2),
frank
myocardial necrosis
(3, 4), and give rise toshock-in-AddressreprintrequeststoDr.Lerman, Division ofCardiology, Uni-versity of Virginia Medical Center, Box 158, Charlottesville, VA
22908.
Receivedfor publication
23July
1985and inrevisedform
15April
1987.
duced
arrhythmias
that mayimmediately
refibrillate the heart (5, 6).Clinical studies have yielded
confficting
dataregarding
theoptimal
energy doserequired
todefibrillate
apatient (7-10).
Despite
significant subject-to-subject variability in
energythresholds
(unrelated
tobody
weight (1 1-14), the
consensusrecommendation is
todeliver
afixed
energydose
toall
patients
(15).
This energy-based
methodofdefibrillation
mayhowever,result in
delivering excessive
current topatients with low
trans-thoracic
impedance
andinsufficient
current topatients with
high impedance.
To date, no study has shown that peak current (A)
is
abetter
descriptor
of threshold than energy.Rather,
experimen-tal studies have
only shown
thatpeak
currentnormalized
tobody
weight (1 A/kg) (16) is
abetter thresholddescriptor
than energy(17).
These studies have had minimal clinicalimpact
since
delivering
1A/kg
wouldresult
inmarkedly excessive
current
doses
compared with
thatdelivered by
presentenergy-based
practice.
Transthoracic impedance is
amajor determinant of
en-ergy-based defibrillation threshold (14, 18,
19).
Transthoracic
impedance is
not,however,
constantfor
agiven subject,
butrather
is sensitive
toand
dependent
onthe externalthoracic
interface between the electrodes and
subject.
Wehypothesized
that these
operator-dependent variables of
impedance
(e.g.,
electrode
force, size,
andelectrode-electrolyte interface)
pri-marily affect series rather than parallel changes
inthe
elec-trode-subject interface.
Therefore,
current-based thresholds
for
agiven subject
should beinvariant
andindependent
of
changes in transthoracic
impedance
incontrast toenergy- andvoltage-based thresholds.
We thus examined therelationshipbetween variations in transthoracic
impedance
and
defibrilla-tion threshold
asdefined
by
delivered energy,peak voltage
andpeak
current.Methods
General
Mongrel dogsweighing 16to24kgwereanesthetized with
pentobar-bital,30mg/kg,and ventilated withaHarvard
respirator.
Electrocar-diographicsurface lead II and arterialpressure
(Statham
P23Dbtrans-ducer, Statham Instruments, Inc., Oxnard, CA) were
continuously
monitoredon anelectrostatic recorder(model
ES-l000,
Gould, Inc.,Houston,TX). Electrodes coveredwith electrode paste(ReduxPaste, Hewlett-Packard, Inc., Palo Alto, CA) were
firmly
placed overtheshavenrightand left lateral chest wallsatthetransverselevel of the
heart.Damped-sinusoidalDCshocksweredeliveredbyeitheran
un-modified, commercially available defibrillator
(Lifepak 6,
Physio-ControlCorp.,Redmond, WA)inwhich theenergy
setting
isequiva-lentto theenergydischargedintoa50 Qload
(protocol A)
orby
a modified device inprotocolB.Voltage deliveredacrossthe thoraxwasmeasuredwitha 1,000:1
voltage divider inparallel with the defibrillatoroutput, and the deliv-eredcurrent wasmeasured with a0.1-0 aresistor in series withthe J.Clin. Invest.
©TheAmerican Society forClinical Investigation, Inc.
0021-9738/87/09/0797/07 $2.00
defibrillator output. Voltage andcurrentwaveforms weredisplayedon
atriggered-sweep storage oscilloscope (model 5113, Tektronix,
Bea-verton, OR) with a frequency response from DC to IMHz.
Time course
of
voltage and current
during discharge
Toderive asimplifiedrelationship between peak current, energy and transthoracic load, the timecoursesof deliveredvoltage and current waveforms were examined across a rangeof six energy settings (30to
400 J) in 13dogs aswellas between alldogsat 400J. The stored waveforms werephotographed withaPolaroidcameraand then traced onto thedigitizing tablet ofamicrocomputer (Hewlett-Packard 9815) for analysis.
Relationship
between transthoracic
impedance
and
energy-,
voltage-,
and
current-based
defibrillation
thresholds
Todetermine therelationshipbetweentransthoracicimpedanceand energy-,voltage-andcurrent-based thresholdstwoprotocolswere de-signed.
ProtocolA (alterationof
impedance by
electrodesize). Defibrilla-tion thresholdsweredetermined in 11 mongrel dogsat twodifferent impedances(high andlow)usingastandarddefibrillator. Transtho-racicimpedancewasalteredbyameanof 95% (range, 50to 188%) in eachanimalbyusingdifferent sized oval electrodes(19cm2-high im-pedance or 76cm2-low impedance). Ventricular fibrillation was in-duced byintroducing alternatingcurrentthroughapercutaneousqua-dripolarcatheter (No. 6F, United States Catheter and Instrument
Corp., Billerica, MA) positioned either in therightorleft ventricle. After 30 s of ventricularfibrillation,ovalelectrodes covered with elec-trode pastewerefirmlyheldbythe operatoroverthe rightandleft lateral chest wall at the approximate transverse level of the heart. Energywasdischargedat aninitial setting of 30 Jduringpeak
inspira-tion. Incremental energysettingswereselected(50, 100,200J)until defibrillationwasachieved. Peakcurrentand energywererecorded for each shock and thetransthoracicimpedancecalculated.
To confirm reproducibility of threshold measurements (at each impedance level) in each dog, ventricular fibrillation wasreinitiated
duringthreeadditional trialsfollowinga5-minequilibration period duringwhich blood pressure, arterial blood gases, and heartrate
re-turnedtobaseline. Resultswereaccepted foranalysisif thresholds for
atleast three offour trialswereidenticalateach of thetwoimpedance
levels. Defibrillation thresholdsweredeterminedinitiallyatlow im-pedance and thenrepeatedathigh impedance.
Protocol B(alteration of impedance byelectrodeforce). Transtho-racicimpedance was altered in this protocol by changing electrode forcethroughaprecisionforce-control system. Thresholdswere deter-minedathighelectrodeforce, 130 Newtons (N) (low impedance) and
atlowelectrodeforce, 20N(highimpedance).Adefibrillator modified
specificallyfor this protocol(Fig. 1)wascalibrated in units of current anddischargedthroughaprecisioncontrolsystem that regulated elec-trodeforce. Electrode size (60 cm2, circular stainless steel) was held
constant.
Thecurrent-baseddefibrillator (constant-load current divider cir-cuit)was amodified commercially availabledefibrillatorthat was cali-brated to deliver 3,000 V at 400 J across a 50 ( load. The dog's transthoracic impedance or resistance RT was determined
prospec-tivelywitha30-kHzsinewave(18,20)transmitted through the
exter-nalelectrodes. The variable resistorsRpand Rs (Fig. 1) were then adjustedtodeliver the desired current to RT and simultaneously maintain aconstant 50 0load to the defibrillator. Thus, a slightly overdamped pulse was delivered (that was within 2% of the selected current)during each defibrillation trial, regardless of the transthoracic
impedance.
The system tocontrol electrode force was adjusted by changing precision weights on a pair of levers. The tension of the weights was changedto acompressiveforce on the electrodes by a statically-bal-ancedpairof levers, each having one horizontal and one diagonal arm
(Fig. 2).
Allforces duetotheweight
of the levers and electrodeswerel+S X
Defibrillator 50ohms
I,0 Rp
RT
Figure1.Modified current-based defibrillator (constant-loadcurrent
divider circuit). Seetextfordetails. Rp, variable parallel resistor;Rs,
variable seriesresistor; RT,transthoracic resistance
(subject).
cancelled by adjusting counter-weights until the horizontal beamswere
statically balanced withnoapplied weight. Thus, thenetforceatthe electrodeswasequal in magnitudeto that supplied by the precision weights. The endotracheal tubewasclampedatpeak-inspiriation be-forebalancing the system and delivering the shock.
Defibrillation thresholds weredetermined atboth high and low electrode force. After induction of ventricular fibrillation, thresholds weredetermined, starting with 12Aandincrementingin steps of 2 A untildefibrillation occurred. Thresholdsweredetermined in three
con-secutive trials with low electrode force (high transthoracic impedance)
at5-min intervals. Then three consecutive thresholdsweredetermined withhigh electrode force (low transthoracic impedance) at 5-min in-tervals. Preliminary experiments performed in four dogstodetermine the stability of current-based thresholds with respectto number of shocks and time showed an approximate exponential decline of 30% in threshold overthe first six shocks (5-min intervals), whichwas fol-lowedbyastableplateau phase. For this reason, thresholdsduringthe experimental protocolweredetermined after six shocksweredelivered insinus rhythm at 5-min intervals. Voltage, current, and energy thresholds and transthoracic impedance determined during the
pla-teauphasewerethenaveraged for all defibrillation trials in each elec-trode configuration.
Statistical
analysis
Statistical comparisons of energy, voltage and currentdefibrillation thresholdsat twodifferent transthoracicimpedancesin each animal
wereperformed using Student's pairedt test.Thesignificanceof linear
relationshipsbetweenvariableswasexaminedusingleast squares
lin-earregression.Dataareexpressedasmean±SD.Differenceswere
con-sideredsignificantforP<0.05.
Results
Relationship
between transthoracic impedance and
defibrillation threshold
Protocol A: alteration ofimpedance due to electrode size.
Defi-brillation thresholds determined in each animal at a control impedance, 44±8 Q (low impedance, standard electrodes) and
at animpedance increased by 95%, 86±17 0 (small electrodes) (range, 50to 188%)showed marked variability with respect to energyandvoltagethresholds. For example, at low impedance
thresholdenergy was44±21J compared with 105±35 J at high impedance, P<0.0001. Threshold voltage was 937±237 V at
lowimpedance compared with 1,980±291 V at high imped-ance,P <0.0001 (Fig. 3). In contrast, peak current thresholds remainedinvariant with respect to impedance, 22±5 A (low impedance) vs. 24±6 A
(high impedance),
P= NS. Further-more, currentshowed a smallercoefficient
of variation(SD/
mean) than energy, 24 vs. 41%. Thus, each animal demon-stratedathresholdcurrentthatwas
independent of
transtho-racicimpedance,
whereas energy and voltage requirementsweredirectlyrelated to impedance for a given animal.
I
I-
Vertical Adjustment- Chain Drive
VerticalAdjustment
Support HorizontalAdjustment Ii Support
71-- -a Right Angle Drive =___________ I-) Fulcrum
-.- - -(Journal Bearing)
_ -- -~ ElectrodeAngle - Adjustment
I
Set A= .0=90
IFI=IW /--i'-Torque Decoupler e / of } }g, _ (Journal Bearing)
/ |2 -~ Electrode
F -a...- Counterbalance
____ E Weights/ .
---S--__- ____ DogFigure 2. Apparatusforcontrolling electrode force. See text for discussion.
electrode
force (20 N) increased the transthoracic impedance
in each animal
by a meanof 32%,
from 71±9
Q athigh
elec-trode force
(130 N)
to94±10
Q atlow electrode
force,
P<
0.0001.
Energythresholds
increased in
eachdog with
lowelectrode
force,
from 70±34
J(130 N)
to111 ±47J(20 N),
P<
0.001
asdid voltage thresholds,
1,452
±363
V(130
N)
to2,117+489
V(20
N),
P=0.0001
(Fig. 4).
In contrast,currentthresholds remained invariant with
respecttochange
intrans-thoracic
impedance,
21±6 A(low
impedance) and
23±5
A(high impedance),
P=NS.
Inaddition, the coefficient
ofvaria-tion
wasless
for
currentthanfor
energy-based thresholds,
25
vs
45%.
Aninitial shock of 30
Awould
havedefibrillated
90%of the animals in
protocols
Aand B.Current
thresholdswerenot related to
body
weight
ineither protocol
AorB.0LOWTTI
200 NHIGH TTI
Energy P.000
(j)
100II
I
I
1 I
I
I
a Voltage2
(kV)
1.
5
Current 2
(A)
p0.000,
1O.0
?5
CP's
|
I
|
|
Relationship
between
peak
current, energy, and
transthoracic
impedance
To
determine
whether thetime
courses of voltage and currentwaveforms
were dependent on the magnitude of thedis-charged pulse, superimposition of
thenormalized
voltage and currentwaveforms
wasperformed
ineach
dog atthesix
energy levels(Figs.
5 and 6).Similarly,
thedata were examined forinter-animal variability
bynormalizing
the 400-Jshocks acrossall dogs
(Fig. 7).
The curves in thesefigures
show that thetime
courses
ofdelivered
voltage and current weresimilar
across the rangeof
shocksin each
dog as well as between all dogs at a constant energylevel.
Since
the curves all haveasimilar
qualitative
shape
and thetransthoracic
impedance
is
predominantly
resistive (21, 22),
a24
Energy (J)
IC
3
Voltage 2
(kV)
E]LOW TTI
)tpoo.oo0
* HIGH TTIFl
Fl
I
l
I
fL P. O poo
xv p. 0.0001
.01111111
0 50T Current
(A) 2
1 2 3 4 5 6 7 8 9 10 11
Dog
Figure 3. Energy-,voltage-andcurrent-based thresholdsas a func-tion of transthoracicimpedance. Impedancewaschangedin each an-imal bya meanof 95%by alteringelectrode size.TTI,transthoracic
impedance.Seetextfor discussion.
p. NS
Mti
FMIrili.
1 2 3 4 5 6 7 8 9
Dog
Figure4.Energy-,voltage-andcurrent-based thresholds.
Impedance
waschangedineach animalbyalteringelectrode force. Abbrevia-tionsaspreviously designated.Seetextfor discussion.
fftrUMIFIIFFIrrUIrdMITTnTFrniiiiiiiiijit .
1'C'
2 3 4
io
I
1 2 3 4 5 TIME(m)
6
/f
a 9
a.t.
.l10 I1
LP
L
-.
12 13
Figure5. Eachpanel displaysthevoltagewaveformsforeachdogat6energylevels, (30-400 J).Each waveformwasnormalized toapeakvalue of100. The timecourseofvoltage decayisindependentof themagnitudeofdischargedenergy.
relationship
between peak delivered current anddischarged
energy canbe derived in thefollowing
manner:Bydefinition,
W= i2(t)Rdt (1)
DOG 1
1001
I5
NORMALIZED 50 CURRENT%)_
251
1 2 3 4 5
TYME() 6
10
where W is deliveredenergy,
i(t)
is the value ofcurrent attimetand R is transthoracic resistance. By
dividing
current by itspeak
value,
Ip,
one canobtainafunction
for normalized cur-rentiN(t):3 4
7
1 1
5
t.X\
A,
9
13 12
Figure6.Normalizedcurrentwaveformsfor each dog during6differentenergypulses(30-400 J). The timecourseof deliveredcurrentis inde-pendentofthemagnitude ofenergyandthe timecoursesofvoltage andcurrentareidenticaltoeach other.
ic
7
NORMALIZED 5 VOLTAGE (%)
2
A
10075
NORMALIZED VOLTAGE(%) 50
25
0
s/v .
t..
I
Is
f
.
The data derived from discharging six energy levels across 13 dogs were used to evaluate the validity of Eq. 5 according to ageneral linear model (Statistical Analysis System, SAS Insti-tute, Cary, NC). This procedure permitted the assessment of the overallslope and intercept of the
relationship
and the like-lihood that any individual animal had a different slope from theothers
(homogeneity
ofslopes).
Differences inslope
be-tween animals were notsignificant (P
=NS),
and thus an overallslope
andintercept
werecalculated from all animals in acombined
model. The data werestrongly linearly
related (r =0.999, P < 0.0001) with theregression equation:I,=22
fW/R-0.9
(6)
The
regression line
wasthen force-fitted
through
zerowith the resultantequation (Fig. 8) (r
=0.999,
P < 0.0001):B
1a7
NORMALIZED CURRENT (%) 5
2 101
lo
5j^i
V
Figure7.(A) Normalized voltage war dogsduringa400-J shock. Each
wavy
of 100. The time course of voltage de
Ip
=
22
VW/R
(7)Predictive accuracy
ofderived current equation
To
prospectively validate
thepredictive
accuracyof
Eq.7,six
N.4shocks
at energylevels from 30to400 Jweredischarged
trans-I'M
thoracically
inanadditional 10
dogs.
The energysetting
onthedefibrillator
was used as anapproximation for delivered en-ergysince
the two are nearlyequivalent for
the rangeof trans-thoracic impedance observed in this study.. The estimates for peak current derived from Eq. 7 are presented together with the measured values forpeak
current inFig.
9. For the six energy levels across the10
dogs, 92%of the estimates derived
from Eq. 7 were within 5% of the measured values, and 100%
5 of
the
estimates
were within8% of the measured values.
Therewas also no
significant difference (ANOVA)
between mea-TIME (ms) sured peak current, predicted peak current by Eq. 7, and thatveforms superimposed for 13 predicted by second order source-free RLC circuit equations eform was normalized to a peak (23), which completely describe the defibrillator circuit and cay was nearly identical for all load(Fig. 9).
dogs.(B) Normalized current waveforms superimposed for 400 J shocks. Thetime course of current decay was nearly identical for all dogs and thereforewasindependent of the transthoracic load.
LN(t)
=i(t)/Ip
(2)Substituting
Eq.
(2) into Eq. (1),
=
Ip2
iN2(t)Rdt
=
IP2R
f
iN2(t)dt
(3)
The
shapes of
thenormalized
currentwaveforms
across eachdog
andbetween
alldogs
werenearly identical
andsu-perimposable
for the
rangeof
energies
examined (30-400 J)
(Figs. 6 and 7).
Under theseconditions,
f
iN(t)dt
is
constant asis
fiN2(t)dt.
Settingthis latterintegralto
K,
wemayderive W=Ip2RK1
Discussion
The
major findings of this study
arethat current-basedthresh-olds, in
contrast toenergy- andvoltage-based
thresholds,
areinvariant for
agiven subject
and areindependent of
operator-dependent variables of transthoracic
impedance.
Impedance
was altered in each
animal
byeither
varying electrode size
orforce. These methods of
altering
impedance
werechosen sincechanges
in theelectrode-subject
interfaceareoperator-depen-1 00
I
(A)
I= 22 -/R
8 0- r=0.999
60
40
2 0
0
(4)
and
solving
for
Ip,
wehaveanequation
relating peak
currentto
delivered
energy and transthoracicresistance:
Ip
=K2 VW/R whereK2=l/VKi.
0 2 3 4
tW/R
(5)
Figure 8. Linearregressionfor peak delivered current(I)as a func-tion ofthesquarerootof thequotientof delivered energy(FW)andtransthoracic load (R).
'-A
10I
8(
6C
30J
* 20J
,
50J
o300J
A
1O
O
4001
0'
oA
A~~~~A
20
40
60
80
100
Predicted Peak Current
I21
1-
/
IA~~~~~~~~~~~~~~~~~~~
0-y
X
A
n-u
0n
20
40
60
80
100Predicted
Peak
Current
#Figure9.Comparisonbetweenmeasuredandpredicted peakcurrent across 10 animals atenergylevels 30 to 400 J.(A)Predictedpeakcurrent
determinedby Eq.(7)*.Dottedlinerepresentstheline ofidentity. (B).Predictedpeakcurrent basedonsecondordersource-free RLC
equa-tions(23)**. Capacitance,36 MF;inductance,28mH;internal resistance ofdefibrillator, 12.8U. dent and can alter
significantly
between successive shocks(24, 25).
An ideal descriptor of threshold should be invariant with
respect to operator-dependent changes in impedance.
How-ever,these
changes
inimpedance
areamajor determinant of defibrillation successinthe energy-based approach sincehighimpedance patients require more energy to defibrillate than low impedance patients (14, 18, 19). Experimentally, imped-ance-based energyadjustments have been showntoimprove defibrillation rates when impedance is increased by altering the
electrode-subject
interface (19). Our dataindicate,
how-ever, that this approach is indirect and suboptimal since (a) this method dictates that both energy andcurrent doses are
dependentontransthoracic impedance, and (b) incontrast to
energy
thresholds,
currentthresholdsareinvariant andinsen-sitivetochanges in transthoracic impedance.
Anadditional limitation of theenergy-baseddefibrillation
methodis the
difference
betweenindicated(orselected)energyonadefibrillator and theactualdeliveredenergy.Foragiven indicatedenergy,actualdeliveredenergywill show significant variability acrossallsubjects duetoitsdependence on
trans-thoracic
impedance.
Thisconceptcanbe understoodbyexam-ining
the relationship between threediscrete
but commonlymisunderstood variables: W., stored energy (capacitor); Wd,
actualdelivered energy;and Wi, indicatedorselectedenergy on adefibrillator. For example,
(RT+RO)
(8)
where RT - transthoracic resistance, and
Ro
= internalresis-tanceof the defibrillator. Fora50-f load, deliveredenergy is
equivalenttoindicatedenergy
(Wi).
Thatis,We _t o A ) (9)
W, canbe expressed intermsof storedenergy Wsby
rearrang-ingEq. 9.
Ws w
I( 50+
)(10)
Substituting
Efq.
10 intoEq.
8givestherelationship between indicated and deliveredenergy(1 1)
Therefore, accordingtoEq. 11iashock dischargedata200-J
setting
will delivermore energytoa100-X
subjectthan to a50-0 subject.
Though
the concept of defibrillation defined in terms ofcurrenthas beenpreviouslyintroduced(16),the dosewas
con-
100-80
z
u
60
X4
240
4,
20
0
B
4(
RT
Wd-Wi
i+
sidered weight dependent and was therefore normalized to
body weight (1 A/kg).
Minimalinterest in thisapproach
devel-oped
since it would deliver 70 A to the average person and morethan 100 A tolarge subjects, a dose far in excess of that deliveredby present energy-based practice (26).Our data do not address the threshold-impedance relation-ship due to changes in resistance ofintrathoracic structures. These changes, though
potentially important,
develop more slowly and are less frequently encountered clinically during successive shocks than areelectrode-subject alterations
in im-pedance. One would predict that thedevelopment of pleural effusion, pleural fibrosis, pulmonary edema or pericardial ef-fusion, etc., couldpotentially affect
changes inresistances
bothin series
and parallel to thetransthoracic
electrodes. Under some of theseconditions, therefore,
voltage rather than cur-rent could prove to be abetter threshold descriptor.The
electrical
strengthfor
defibrillation pulses hascustom-arily been calibrated
inunits
of
energy. Itis likely
that energy-based defibrillators wereoriginally
developed because oftech-nological considerations,
one of which was the ability to mea-sure theenergy stored in acapacitor. The simple relationship derived inthisstudybetween
transthoracic resistance(impe-dance),
energy andpeak current (Eq. 7)provides
aclinically
useful
relationship
thatpermits
the operator to select the en-ergysetting
necessary to deliver a selectedpeak
currentusing
astandard,
commercially available defibrillator (provided
that transthoracicimpedance
isprospectively determined).
Auto-mated current-baseddefibrillators
can be developedfor
clini-cal use
with
microprocessor-based technology.
Though our data show
variability
in current-based thresh-olds across alldogs,
it isconsiderably
less than that for energy orvoltage.
Our results suggest thata fixed-dose of 30 Awilldefibrillate
most animals.Extrapolation
ofpublished
clinical data,though
notanalyzed
orinterpreted
inthis
respect, sug-gest that an initial shock between 20 and 30Awould be anappropriate
firstapproximation
for human defibrillation (11,18).
Insummary, our
findings
demonstrate that current-basedthresholds,
in contrast toenergy andvoltage thresholds,
areindependent of operator-dependent variables of transthoracic
impedance
and areinvariant
for
agiven subject.
Theseresults
suggest
that
redefining
defibrillation threshold intermsofpeak
current
rather
than energyprovides
asuperior
method ofdefi-brillation.
Acknowledgments
Dr. Lerman isa-recipient ofa National Institutes ofHealth New
InvestigatorAward(HL-35860).
References
1.Chen,P-S.,P.D.Wolf,F. J.Claydon,E.G.Dixon,H. J. Vidail-let Jr.,N. D.Danieley, T.C.Pilkington,and R. E. Ideker. 1986.The
potential gradientfieldcreatedbyepicardialdefibrillation electrodesin
dogs. Circulation. 74:626-636.
2.Ehsani,A., G.A.Ewy,and B. E.Sobel. 1976.Effects of electrical
countershock on serum creatine phosphokinase isoenzyme activity.
Am.J.Cardiol. 37:12-18.
3.Dahl,E.F.,G. A.Ewy,E. D.Warner,and E. D. Thomas. 1974.
Myocardial necrosis from directcurrentcountershock:effectof paddle
electrode size and time interval between discharges. Circulation. 50:956-961.
4.Warner,E.D., C.Dahl, and G.A.Ewy. 1975.Mycardialinjury
from transthoracic defibrillator countershock. Arch Pathol. 99:55-61. 5. Jones, J. L., E. Lepeschkin,R. E. Jones, and S. Rush. 1978. Response of culturedmyocardialcellstocountershock-type electric field stimulation.Am.J.Physiol.235:H214-H222.
6. Jones, J. L., and R. E. Jones. 1980. Postshockarrhythmias-a possiblecauseofunsuccessful defibrillation. Crit. Care Med. 8:167-197.
7.Lown, B.,R.S.Crampton,R. A.DeSilva, andJ.Gascho. 1978. Theenergyfor ventriculardefibrillation-too littleor toomuch?N.
Engl.J.Med. 298:1252-1253.
8.Adgey,A. A.J.,J. N.Patton,N. P.S.Campbell, and S.W.Webb. 1979.Ventricular defibrillation: appropriateenergylevels.Circulation. 60:219-222.
9.Tacker,W.A.,Jr., and G.A.Ewy. 1979.Emergency defibrilla-tion doses: recommendadefibrilla-tions and radefibrilla-tionale. Circuladefibrilla-tion. 60:223-225. 10. Weaver,W.D.,L. A.Cobb, M. D. Copass, andA.P.Hallstrom. 1982. Ventriculardefibrillation-a comparative trialusing 175-Jand 320-J shocks.N.Engl. J. Med. 307:1101-1106.
11.Patton, J. S., and J. F.Pantridge. 1979. Current required for ventricular defibrillation.Br.Med. J. 1:513-514.
12. Kerber,R. E., and W. Sarnat. 1979. Factorsinfluencingthe
successof ventricular defibrillation inman.Circulation. 60:226-230. 13.Gascho,J.A.,R.S.Crampton,J.N.Sipes,M. L.Cherwek,F. P.
Hunter, and W. M. O'Brien. 1979. Energy levelsandpatientweightin ventricular defibrillation. JAMA (J.Am. Med. Assoc.). 252:1380-1384.
14.Kerber, R. E., S. R.Jenser,J. A. Gascho,J.Grayzel,and R.
Hoyt. 1983. Determinants of defibrillation.Prospectiveanalysisof 183
patients.Am.J.Cardiol. 52:739-745.
15.1986. Standards and Guidelines forCardiopulmonary Resusci-tation(CPR)andEmergencyCardiac Care(ECC).JAMA(J.Am.Med. Assoc.). 255:2942-2943.
16.Geddes,L.A., W.A.Tacker,J. P.Rosborough,A.G. Moore, andP. S.Cabler. 1974.Electrical dosefor ventricular defibrillation of large and small animals using precordial electrodes. J. Clin. Invest.
53:310-319.
17.Armayor, M.R., G.Savino,M.E.Valentinuzzi,0.E.Clavin,
J.E.Monzon, andM. T.Arrendondo. 1979. Ventriculardefibrillation thresholds withcapacitor discharge.Med. Biol.Eng.Comput.
17:435-442.
18.Kerber,R.E., C.Kouba, J.Martins,K.Kelly,R.Low,R.Hoyt,
D. Ferguson, L. Bailey, P.Benentt, and F. Charbonnier. 1984.
Ad-vancepredictionof transthoracicimpedancein humandefibrillation andcardioversion:Importanceofimpedanceindeterminingthe suc-cessoflow-energyshocks. Circulation. 70:303-308.
19.Kerber,R.E., D.McPherson,F.Charbonnier,R.Kieso,andP.
Hite. 1985. Automatedimpedance-basedenergyadjustmentfor defi-brillation.Experimentalstudies. Circulation. 71:136-140.
20.Geddes,L. A.,W. A. Tacker,W. Schoenlein, M.
Minton,
S.Grubbs,and P.Wilcox. 1976.The
prediction
of theimpedance
of thethoraxtodefibrillatingcurrent.Med.Instrum. 10:159-162.
21. Ewy,G.A.,M. D.Ewy,A.J.Nuttall,andA.W. Nuttall.1972. Canine transthoracic resistance.J.
Appl. Physiol.
39:91-94.22.Machin,J.W. 1978. Thoracicimpedanceofhumansubjects. Med. Biol.Eng.Comput. 16:169-178.
23.Trantham,J.L.,J. J.Gallagher,L. D.
German,
A.Broughton,
T.Guarnieri,and J.Kasell. 1983. Effects ofenergydelivery
viaaHisbundle catheterduring closed chest ablation of the atrioventricular
conductionsystem. J.Clin.Invest.72:1562-1574.
24.Connell,P. N.,G.A.Ewy,D. F.Dahl,and M. D.Ewy. 1973. Transthoracicimpedancetodefibrillatordischarge.Effect ofelectrode
sizeandelectrode-chestwallinterface.J.Electrocardiol. 6:313-317.
25. Kerber, R. E., J. Grayzel, R. Hoyt, M. Marcus, and B. S.
Kennedy. 1981.Transthoracic resistancein humandefibrillation.
In-fluenceofbodyweight,chestsize,serialshocks,paddlesize and
paddle
contactpressure.Circulation. 63:676-682.
26.