Contents

Contents

Contents ... 1 Contents ... 1 1

1 Brief Introduction ... 3Brief Introduction ... 3 1.1 Characteristics 1.1 Characteristics ... 3 ... 3 1.2 Composition 1.2 Composition ... 3 ... 3 2 2 Specifications Specifications ... ... 55 2.1 Specification 2.1 Specification ... 5 ... 5 2.2 Technical specification 2.2 Technical specification ... 5 ... 5

2.3 Working condition & Environment

2.3 Working condition & Environment... ... 55 3

3 Laser Gauge ... 6Laser Gauge ... 6 3.1 Measuring principle

3.1 Measuring principle ... 6 ... 6

3.2 Layout and function

3.2 Layout and function... ... 88

3.3 Application

3.3 Application ... 8 ... 8 4

4 BX-1A Operation Panel ... 13BX-1A Operation Panel ... 13 4.1 Parameter code

4.1 Parameter code ... 14 ... 14

4.2 Operation flow chart of BX-1A operation panel

4.2 Operation flow chart of BX-1A operation panel ... 19 ... 19

4.3 The alarm

4.3 The alarm... ... 2121

4.4 Calibration

4.4 Calibration ... 21 ... 21 5

5 YK-1A Remote YK-1A Remote Controller Controller ... 23... 23 5.1 Parameter code

5.1 Parameter code ... 24 ... 24

5.2 Operation flow chart of remote controller 

5.2 Operation flow chart of remote controller  ... 29 ... 29

5.3 Output interface

5.3 Output interface ... 29 ... 29 6

6 YK-1B Remote YK-1B Remote Controller Controller ... 31... 31

6.1 Control panel

6.1 Control panel ... 32 ... 32

6.2 Terminals

(2)
(3)

7

7 Remote Remote Display Display ... ... 3333 8

8 Installation Installation and and Operation Operation ... ... 3535 8.1 Installation

8.1 Installation ... 35 ... 35

8.2

8.2 Basic Basic operationoperation ... 37 ... 37

8.3

8.3 Feedback Feedback controlcontrol ... 37 ... 37 9

9 Communication ... 42Communication ... 42 10

10 Overall Dimensions ... 45Overall Dimensions ... 45 11

11 Maintenance Maintenance ... ... 5353 12

12 Malfunction and Malfunction and Solutions Solutions ... 53... 53 13

13 Accessories ... 54Accessories ... 54 14

14 After-saAfter-sale le service ... 54service ... 54 Attachment ... 55 Attachment ... 55

(4)

1

1

Brief Introduction

Brief Introduction

Opmac 

Opmac  series of laser scanning online diameter measurement instruments are specially designed series of laser scanning online diameter measurement instruments are specially designed for wires and cables, which not only measure and display diameter and deviation online, but also for wires and cables, which not only measure and display diameter and deviation online, but also automatically control diameter of wire and print parameters according to user’s demand.

automatically control diameter of wire and print parameters according to user’s demand. Opmac Opmac 

series instruments have two categories: the A series measure object in one axis, and the B series series instruments have two categories: the A series measure object in one axis, and the B series measure object in two vertical axes (X axis and Y axis). The two categories have same measure object in two vertical axes (X axis and Y axis). The two categories have same communication interfaces and operation functions.

communication interfaces and operation functions.

 Suitable for online measurement of any round wires and other objects.Suitable for online measurement of any round wires and other objects.

  Non-contact measurement Non-contact measurement,,high-speed inspection, high precision and stable performance.high-speed inspection, high precision and stable performance.

 Adopt FAdopt Fθθ optics optics,,Fewer measurement Fewer measurement error.error.

 Inside PID regulation module.Inside PID regulation module.

 All-dimensional revolving operating panel.All-dimensional revolving operating panel.

 Remote displays of various sizes are optional.Remote displays of various sizes are optional.

 Standard RS232/485 communication interfaces, supporting CAN and Modbus with extraStandard RS232/485 communication interfaces, supporting CAN and Modbus with extra  protocol adapter.

 protocol adapter.

 Wires quality management software can record the real time curve of changing externalWires quality management software can record the real time curve of changing external diameters.

diameters.

 Reliable operation in the most foul environments.Reliable operation in the most foul environments.

One basic

One basic Opmac Opmac  instrument is composed of laser gauge, operating panel and remote controller instrument is composed of laser gauge, operating panel and remote controller and is

and is able to able to be connected be connected with difwith different sizes of ferent sizes of remote display remote display as indicated as indicated in the in the Figure 1.Figure 1.

( aa)) one axis measurementone axis measurement Figure 1 Figure 1

1.1 Characteristics

1.1 Characteristics

1.2 Composition

1.2 Composition

(5)

( bb)) two axes measurementtwo axes measurement Figure 1

Figure 1

A high-precision laser scanning system and embedded digital processing electronics are inlayed in A high-precision laser scanning system and embedded digital processing electronics are inlayed in the measurement instrument. The signal of diameter processed by scanning system is converted to the measurement instrument. The signal of diameter processed by scanning system is converted to digital signal firstly and then comes out real diameter and related data through a series of operation. digital signal firstly and then comes out real diameter and related data through a series of operation. The communication interface of the measurement instrument contains one serial I/O and three The communication interface of the measurement instrument contains one serial I/O and three synchronized data channels. It could transfers measured data to peripheral device (e.g. remote synchronized data channels. It could transfers measured data to peripheral device (e.g. remote controller, PC and PLC, etc.) through the serial I/O and to remotely controls and displays through controller, PC and PLC, etc.) through the serial I/O and to remotely controls and displays through data channels.

data channels.

Operation panel is installed on top of the measurement instrument, which displays data and Operation panel is installed on top of the measurement instrument, which displays data and  parameters and adjusts optic angle easily.

 parameters and adjusts optic angle easily.

Remote controller is mainly used for automatic control, which obtains measured data from the Remote controller is mainly used for automatic control, which obtains measured data from the measurement instrument. PID signal comes out through the comparison between measured diameter measurement instrument. PID signal comes out through the comparison between measured diameter with pre-set diameter and controls traction speed thereby controls wire diameter. In addition, remote with pre-set diameter and controls traction speed thereby controls wire diameter. In addition, remote controller has a full-function operation panel and standard Serial I/O which communicates with PC controller has a full-function operation panel and standard Serial I/O which communicates with PC and PLC and supports diameter monitor software.

and PLC and supports diameter monitor software.

Opmac 

Opmac   series instruments are also applied for on-line measurement of optical fiber, glass tube,  series instruments are also applied for on-line measurement of optical fiber, glass tube,  plastic tube and other spherical wire.

(6)

2

2

Specifications

Specifications

M o d e l M o d e l M eM ea s u r i n g a s u r i n g r a n g er a n g e  A A cc cc uu rr aacc yy 25AL3/25BL3 0.1 25AL3/25BL3 0.1~~25mm25mm ±±1µm1µm±±0.008%0.008%××XXZZ 40AL3/40BL3 1 40AL3/40BL3 1~~40mm40mm ±±1.5µm1.5µm±±0.008%0.008%××XXZZ 50AL3/50BL3 1 50AL3/50BL3 1~~50mm50mm ±±1.5µm1.5µm±±0.008%0.008%××XXZZ 70AL2/70BL2 1 70AL2/70BL2 1~~70mm70mm ±±10µm10µm±±0.008%0.008%××XXZZ 90AL2 90AL2 one axis

one axis 11~~40mm40mm one axisone axis ±10µm±10µm±±0.008%0.008%××XXZZ edge edge measurement measurement 40 40~~ 100mm 100mm edge edge measurement measurement ±10µm±10µm±±0.008%0.008%××XXZZ 100AL3(2) 100AL3(2) one axis

one axis 11~~45mm45mm one axisone axis ±3(10)µm±3(10)µm±±0.008%0.008%××XXZZ edge edge measurement measurement 45 45~~ 120mm 120mm edge edge measurement measurement ±6(10)µm±6(10)µm±±0.008%0.008%××XXZZ 5AL3 0.03~5mm 5AL3 0.03~5mm ±1µm±0.008%×X±1µm±0.008%×XZZ 5AL4 0.03~5mm 5AL4 0.03~5mm ±0.4µm±0.008%×X±0.4µm±0.008%×XZZ Table 1 Table 1 X

XZZ:Standard diameter of the measured objective:Standard diameter of the measured objective

Power srequirements Power srequirements:: 176V ~ 176V ~ 264V 264V AC 50HzAC 50Hz Operating condition: 5 ~ 45 Operating condition: 5 ~ 45℃℃ ≤≤85%RH85%RH Consumed Power  Consumed Power :≤:≤12W12W

25

25

A

A

L

L

3

3

Display precision: Display precision: 2 2 for for 0.01mm0.01mm 3 3 for for 0.001mm0.001mm 4 4 for for 0.0001mm0.0001mm

Scanning mode: laser scanning Scanning mode: laser scanning Characteristic code:

Characteristic code:

 A for one axis measurement  A for one axis measurement B for two axes measurement B for two axes measurement Model Model

2.1 Specification

2.1 Specification

2.2 Technical specification

2.2 Technical specification

2.3

(7)

3

Laser Gauge

The laser beam irradiates to the rotating prism and after reflection and through lens 1, it becomes  parallel beam. The parallel beam converges to the receiving component through lens 2 [See Figure

2(a,b)]. When placing EUT on focal plane (FP) of lens 1, part of lights are shadowed, thus becomes electrical signal on receiving component and then comes out diameter and a serial of related signals after processing by measurement instrument. The signal processing is indicated in Figure 3.

The 90AL and 100AL laser gauge is different from other one-axis gauge, for it has two parallel axes, which not only can be used to measure 2 small objects respectively, but also can be used to measure one large object [See Figure 2(c-f)].

(a) one axis measurement Figure 2

(b) two axes measurement

Figure 2 Lens1 Prism Laser  Lens2 Receiving component Measuring objective Lens2 in X axis Lens2 in Y axis Prism Measuring objective Receiving component Laser  Lens1 in X axis Lens1 in Y axis

3.1 Measuring principle

(8)

(c) one axis measurement for 90/100AL using upper axis

Figure 2

(d) parallel measurement for 90/100AL Figure 2

(e) one axis measurement for 90/100AL using lower axis

Figure 2

(f) edge measurement for 90/100AL Figure 2

(9)

(a) one axis measurement Figure 3

(b) two axes measurement Figure 3

The functional layout of Laser diameter measurement instrument is indicated in Figure 4. Take the

Opmac  25AL and 25BL for examples.

Laser gauge equals to a high-precision diameter sensor which not only measures online in real time  but also boosts strong data communication function.

3.3.1 For some occasion requiring only diameter measure, the laser gauge with the operating panel is competent for diameter measurement, display and over-deviation alarm. [See Figure 5(a)]

3.3.2 For some occasion not only requiring diameter measure but also automatic control, the solution includes remote controller and laser gauge, which communicating with the former via serial I/O. [See Figure 5(b)]

Diameter Signal Count pulse    C  o   u   n    t  e  r    D  a    t  a    b  u    f    f  e  r    P  r   o   c   e   s   s    i  n  g   c    h    i  p

Synchronized data channels Operating panel

Serial I/O

Synchronized data channels Serial I/O    P  r   o   c   e   s   s    i  n  g   c    h    i  p    D  a    t  a    b  u    f    f  e  r    C  o   u   n    t  e  r Count pulse Operating panel Diameter Signal of X axis Diameter Signal of Y axis

3.2 Layout and function

(10)

1 . L a s e r g a u g e 2 . G u i d e r o l l e r h o l d e r 3 . G u i d e r o l l e r 4 . S y n c h r o n i z ed d at a c h a n n e l 1 5 . S e r i a l I / O 6 . S y n c h r o n i z ed d at a c h a n n e l 2 7 . S y n c h r o n i z ed d at a c h a n n e l 3 8 . P o w e r s w i t c h 9 . G r o u n d i n g b o l t 1 0 . P o w e r s o c k e t Figure 4(a)

(11)

1 . L a s e r g a u g e 2 . G u i d e r o l l e r h o l d e r 3 . G u i d e r o l l e r 4 . S y n c h r o n i z ed d at a c h a n n e l 1 5 . S e r i a l I / O 6 . S y n c h r o n i z ed d at a c h a n n e l 2 7 . S y n c h r o n i z ed d at a c h a n n e l 3 8 . P o w e r s w i t c h 9 . G r o u n d i n g b o l t 1 0 . P o w e r s o c k e t Figure 4(b)

3.3.3 For some occasion requiring management of several laser gauges simultaneously, the measured data from different instruments can be sent to a PC via serial I/O and processed there. [See Figure 5(c) ]

3.3.4 Synchronized data channel is applied to be connected with remote displayer to display data in long distance with transmission speed up to 1MB/s and distance of 300 m. The connecting method is indicated by Figure 6.

(12)

Figure 5(a)

Figure 5(b) connected with remote controller 

(13)

Figure 6

3.3.5 Serial I/O

The serial I/O is a standard serial communication port with RS485 protocol. The connecting cable function is indicated by Figure 7. See Chapter 8 for detail.

P i n P i n C o l o r F u n c t i o n 2 O r a n g e A 3 O r a n g e - w h i t e B 5 B l u e & B l u e - w h i t e D G N D O t h e r s N o n e N o n e Figure 7 3.3.6 Grounding

(14)

4

BX-1A Operation Panel

Operation panel is installed on top of the measurement instrument and connected with it through synchronized data channel 1 as indicated in Figure 8. Operation panel is applied for data display,  parameter management, over-deviation indication and alarm. The panel’s functional layout is

indicated by Figure 9.

One axis Instrument Two axes Instrument

Figure 8

1. Main display screen: displaying diameter value and parameters

2. Sub display screen: displaying deviation value, offset, and parameter code

3. Functional key: menu/option/number 3 4. Functional key: turn down / number 2 5. Functional key: turn up /number 1

6. Upper deviation light: when measured diameter exceeds admissible deviation upper limit, orange light on.

7. None-deviation light: when measured diameter is in admissible deviation, green light on.

8. Lower deviation light: when measured diameter exceeds admissible deviation lower limit, red light on.

9. Over-deviation light: when measured diameter exceeds deviation, light flickers. 10. Over-deviation alarm: when measured diameter exceeds deviation, alarm on.

(15)

Although the parameters of the two categories are different, yet the operating method of them is the same. In the next section, we will take one axis instrument for example.

P a r a m e t e r c o d e D e s c r i p t i o n M o d e l S e t t i n g r a n g e F a c t o r y d e f a u l t R e f e r e n c e d i a m e t e r R e f e r e n c e d i a m e t e r o f a v e r a g e 0 ~ 5 . 2 m m 0 . 5 m m 2 0 ~ 2 7 m m 5 m m 4 0 0 ~ 4 2 m m 1 0 m m 5 0 0 ~ 5 2 m m 1 0 m m 7 0 0 ~ 7 2 m m 5 0 m m 9 0 ~ 1 0 0 m m 5 0 m m 1 0 0 ~ 1 2 0 m m 5 0 m m U p p e r d e v i a t i o n1  l i m i t U p p e r d e v i a t i o n l i m i t o f a v e r a g e L 3 0 ~ 2 m m 0 . 1 m m L 2 0 ~ 2 0 m m 1 m m L o w e r d e v i a t i o n l i m i t L o w e r d e v i a t i o n l i m i t o f a v e r a g e L 2 0 ~ 2 0 m m 1 m m R e f e r e n c e d i a m e t e r o f X a x i s 0 ~ 5 . 2 m m 0 . 5 m m 2 0 ~ 2 7 m m 5 m m 4 0 0 ~ 4 2 m m 1 0 m m 5 0 0 ~ 5 2 m m 1 0 m m 7 0 0 ~ 7 2 m m 5 0 m m 9 0 ~ 1 0 0 m m 5 0 m m 1 0 0 ~ 1 2 0 m m 5 0 m m U p p e r d e v i a t i o n1  l i m i t o f X a x i s L 3 0 ~ 2 m m 0 . 1 m m L 2 0 ~ 2 0 m m 1 m m L o w e r d e v i a t i o n l i m i t o f X a x i s L 3 0 ~ 2 m m 0 . 1 m m L 2 0 ~ 2 0 m m 1 m m R e f e r e n c e d i a m e t e r o f Y a x i s 0 ~ 5 . 2 m m 0 . 5 m m 2 0 ~ 2 7 m m 5 m m 4 0 0 ~ 4 2 m m 1 0 m m 5 0 0 ~ 5 2 m m 1 0 m m 7 0 0 ~ 7 2 m m 5 0 m m 9 0 ~ 1 0 0 m m 5 0 m m 1 0 0 ~ 1 2 0 m m 5 0 m m U p p e r d e v i a t i o n1  l i m i t o f Y a x i s L 3 0 ~ 2 m m 0 . 1 m m

4.1 Parameter code

L 3 0 ~ 2 m m 0 . 1 m m L 2 0 ~ 2 0 m m 1 m m

(16)

L o w e r d e v i a t i o n l i m i t o f Y a x i s L 3 0 ~ 2 m m 0 . 1 m m L 2 0 ~ 2 0 m m 1 m m M a i n d i s p l a y o p t i o n : 0 : d i s p l a y t h e a v e r a g e v a l u e o f X a n d Y 1 : d i s p l a y t h e a v e r a g e d e v i a t i o n v a l u e o f X a n d Y 2 : d i s p l a y e l l i p t i c i t y (f),f=short axis/long axis

3 : d i s p l a y t h e d i a m e t e r v a l u e o f a x i s X 4 : d i s p l a y t h e d i a m e t e r v a l u e o f a x i s Y 5 : d i s p l a y t h e d e v i a t i o n v a l u e o f a x i s X 6 : d i s p l a y t h e d e v i a t i o n v a l u e o f a x i s Y 7 : d i s p l a y t h e p o s i t i o n2  v a l u e o f a x i s X 8 : d i s p l a y t h e p o s i t i o n v a l u e o f a x i s Y 2 5 B / 4 0 B 5 0 B / 7 0 B 0 ~ 8 0

Main display option:

0: automatically display measurement value.3

1: display value of upper axis if in one axis measure mode using upper axis; in other situation it will display

2: display value of lower axis if in one axis measure mode using lower axis; in other situation it will display

3,4,5: correspond to the deviation value of option 0,1,2, respectively.

6,7,8: correspond to the laser axis position value of 0,1,2, respectively.

9 0 / 1 0 0 A 0 ~ 8 0

Sub display option:

0: display the deviation value. 1: displaying the position value.

2 5 A / 4 0 A

5 0 A / 7 0 A 0 , 1 0

Sub display option:

0 : d i s p l a y t h e a v e r a g e v a l u e o f X a n d Y

1 : d i s p l a y t h e a v e r a g e d e v i a t i o n v a l u e o f X a n d Y 2 : d i s p l a y e l l i p t i c i t y (f),f=short axis/long axis

3 : d i s p l a y t h e d i a m e t e r v a l u e o f a x i s X 4 : d i s p l a y t h e d i a m e t e r v a l u e o f a x i s Y 5 : d i s p l a y t h e d e v i a t i o n v a l u e o f a x i s X 6 : d i s p l a y t h e d e v i a t i o n v a l u e o f a x i s Y 2 5 B / 4 0 B 5 0 B / 7 0 B 0 ~ 8 1

(17)

7 : d i s p l a y t h e p o s i t i o n2  v a l u e o f a x i s X 8 : d i s p l a y t h e p o s i t i o n v a l u e o f a x i s Y Sub display option:

0: automatically display measurement value.3

1: display value of upper axis if in one axis measure mode using upper axis; in other situation it will display

2: display value of lower axis if in one axis measure mode using lower axis; in other situation it will display

3,4,5: correspond to the deviation value of option 0,1,2, respectively.

6,7,8: correspond to the laser axis position value of 0,1,2, respectively. 9 0 / 1 0 0 A 0 ~ 8 3 Alarm control: 0: alarm disabled, 1: alarm enabled. 25A/40A 50A/70A 0 、 1 0 Alarm control: 0: alarm disabled,

1: alarm enabled when the average value is out of tolerance, 2: alarm enabled when the value from X(up) or Y(low) axis is out of tolerance,

3: alarm enabled when the value from X(up) axis is out of tolerance,

4: alarm enabled when the value from Y(low) axis is out of tolerance. 25B/40B 50B/70B 90A 100A 0 — 4 0 A d v a n c e d s e t t i n g : i n p u t p a s s w o r d t o e n t e r s u b m e n u A l l 5 d i g i t s 1 2 3 1 2 A v e r a g e t i m e s : s a m p l i n g s p e e d f o r d i s p l a y v a l u e A l l 1 ~ 1 0 0 0 1 0

The type of the measured objective: 0: non-transparent objective; 1: transparent objective

2 5 B / 4 0 B

5 0 B / 7 0 B 0 , 1 0

The type of the measured objective: 0: non-transparent objective;

1: the edge measurement of transparent objectives

9 0 / 1 0 0 A 0 , 1 0

(18)

B a u d r a t e f o r s e r i a l c o m m u n i c a t i o n A l l 1 9 2 0 0 9 6 0 0 4 8 0 0 2 4 0 0 1 2 0 0 9 6 0 0 C h e c k s u m m e t h o d : 0 f o r C R C c h e c k s u m ; 1 f o r e x c l u s i v e - o r c h e c k s u m ; 2 f o r M o d b u s R T U 2 a x e s 0 , 1 0 M o d i f y p a s s w o r d : c h a n g e p a s s w o r d o f a d v a n c e d s e t t i n g . A l l Z e r o p o i n t 1 : s e t b e f o r e s o l d , u s e r s h o u l d n ’ t s e t i t . 2 5 A / 2 5 B 4 0 A / 4 0 B 5 0 A / 5 0 B 7 0 A / 7 0 B Z e r o p o i n t 2 : s e t b e f o r e s o l d , u s e r s h o u l d n ’ t s e t i t . Z e r o p o i n t 3 : s e t b e f o r e s o l d , u s e r s h o u l d n ’ t s e t i t . Z e r o p o i n t 4 : s e t b e f o r e s o l d , u s e r s h o u l d n ’ t s e t i t . Z e r o p o i n t 5 : s e t b e f o r e s o l d , u s e r s h o u l d n ’ t s e t i t . I n i t i a l s t a t u s a f t e r p o w e r o n A l l W a i t i n g o r E U T b e y o n d m e a s u r e a r e a A l l F a u l t A l l U p p e r d e v i a t i o n v a l u e e x c e e d s 9 . 9 9 9 A l l L o w e r d e v i a t i o n v a l u e e x c e e d s 9 . 9 9 9 A l l v e r s i o n n u m b e r ( s e c o n d a r y d i s p l a y s c r e e n d i s p l a y t h e i n s t r u m e n t m o d e l ) A l l T h e s i z e o f t h e u p - a x i s m e a s u r e m e n t a r e a ; m e a n w h i l e , p r e s s " u p " a n d " d o w n " a n d t o s a v e 9 0 / 1 0 0 A z e r o p o i n t 1 o f u p - a x i s 9 0 / 1 0 0 A z e r o p o i n t 2 o f u p - a x i s 9 0 / 1 0 0 A z e r o p o i n t 3 o f u p - a x i s 9 0 / 1 0 0 A z e r o p o i n t 4 o f u p - a x i s 9 0 / 1 0 0 A z e r o p o i n t 5 o f u p - a x i s 9 0 / 1 0 0 A T h e s i z e o f t h e l o w - a x i s m e a s u r e m e n t a r e a ; m e a n w h i l e , p r e s s " u p " a n d " d o w n " a n d t o s a v e 9 0 / 1 0 0 A z e r o p o i n t 1 o f l o w - a x i s 9 0 / 1 0 0 A

(19)

z e r o p o i n t 2 o f l o w - a x i s 9 0 / 1 0 0 A z e r o p o i n t 3 o f l o w - a x i s 9 0 / 1 0 0 A z e r o p o i n t 4 o f l o w - a x i s 9 0 / 1 0 0 A z e r o p o i n t 5 o f l o w - a x i s 9 0 / 1 0 0 A z e r o p o i n t 1 f o r m i d - u p m e a s u r e m e n t , f o r u s e r a d j u s t m e n t 1 0 0 A L 3 z e r o p o i n t 2 f o r m i d - u p m e a s u r e m e n t , f o r u s e r a d j u s t m e n t 1 0 0 A L 3 z e r o p o i n t 1 f o r m i d - l o w m e a s u r e m e n t , f o r u s e r a d j u s t m e n t 1 0 0 A L 3 z e r o p o i n t 2 f o r m i d - l o w m e a s u r e m e n t , f o r u s e r a d j u s t m e n t 1 0 0 A L 3 T h e m i d d l e p r o o f r e a d p o i n t 1 9 0 / 1 0 0 A L 2 T h e m i d d l e p r o o f r e a d p o i n t 2 9 0 / 1 0 0 A L 2 T h e m i d d l e p r o o f r e a d p o i n t 3 9 0 / 1 0 0 A L 2 T h e m i d d l e p r o o f r e a d p o i n t 4 9 0 / 1 0 0 A L 2  Note:

1. The deviation value is the difference of the actual measurement diameter and the reference diameter. 2. The position value is the difference of the center of the measured objective and the center of the beam. 3. The last decimal fraction is not displayed if the value exceeds 99.999. The displayed value depends on

measure mode. If it is edge measurement, it will display edge measurement value; if it is one axis measurement, it will display one axis measurement value; if it is two axes measurement, it will display the average value of the two axes measurement.

(20)

Main menu, take one axis instrument for example:

(21)
(22)

4.2.1  Under any circumstance, press for 3 seconds, the displayer returns to measuring status and saves parameter simultaneously. Without any operation, it returns to measuring status after one minute without saving of parameter.

: The grey figure means flickering.

The status of the alarm is controlled by main menu item "SWH-B".

The alarm will beep and flash when there is over deviation if the alarm is on. During the alarm, if the measured objective is removed or any key in the operation panel is pressed, the alarm will be automatically shut down for the time being. Until there is a measured objective or no over deviation,the alarm is automatically recovered.

The instrument of model Opmac25□  and Opmac40□  can’t shut down alarm automatically if whose firmware version is older than V1.05 or can’t be displayed.

When the instrument measure value and the fact value have large discrepancy,user can adjust personally.Take the Opmac 100AL3 for example, explain the emendation method.

Preparation:take 2~5 master bar,clean up the surface of the master bar,emendation

 point

dot

□ ”

correspond with five master bar respectively. Arrange orderly from

small to large.

one axis emendation:enter into the second class menu,put the menu interface

switch to

“ □

dot1

,put the most small master bar into the measure area,adjust

the show value equal to the standard ctick value, at the same time press “up key” and “down key”to affirm.according to this method, put the menu interface switch

to

“ □

dot2

”、“

¯  dot3

”、“

¯  dot4

”、

¯  dot5

respect ively, put the other master

 bars to adjust. If the mas ter bars are less than five, then the after points are

unwanted to adjust. Emendation of Opmac 25

□ 、

Opmac 40

according to this

method is OK. Opmac 100AL3’s emendation also accordding to this method,

adjust the nether axes emendation point

 _dot

□ ”

.After the completion of

emendation,prss “menu” key for a certain time,until the display measure interface save the parameter.

edge measurement emendation:don’t place any measured good in the measure

area, put the menu interface switch to

¯dot0

a nd

 _dot0

, at the same time

 press “up key” and “down key”to affirm t he measure area.put the menu interface

switch to

-dot1

”、

-dot2

”、

-dot1

”、

-dot2

in turn.Put the stardand stick into

the measure area(see figure 10) ,emendate the value of master bar according to the one axes emendation method, and save.

4.3 The alarm

(23)
(24)

5

YK-1A Remote Controller

Remote controller collects diameter value and parameters from measurement instrument, and it is mainly applied for feedback control and communication with peripheral device. The functional layout is indicated by Figure 11.

Figure 11

1. Main display screen: displaying diameter and parameter.

2. Sub display screen: displaying deviation, offset and parameter code. 3. Automatic control light: light on in automatic control mode.

4. Power supply light: light on while powering on.

5. Automatic control key: press this key to enter automatic control mode, repress it to exit.

6. Power supply key: press for 0.5 second to power on and repress it for 0.5 second to power off.

7. Bus & control: serial I/O for the gauge, analog output, over-deviation alarm output. 8. RS232/485: serial I/O for peripheral equipments.

9. Power supply socket 10. Installation rack

11. Output setting: adjust voltage range of analog output when in deviation / PID control mode. The adjustable range is ±0.2V~2.3V, and the default value is ±2V.

(25)

P a r a m e t e r c o d e D e s c r i p t i o n M o d e l S e t t i n g r a n g e F a c t o r y d e f a u l t R e f e r e n c e d i a m e t e r

R e f e r e n c e d i a m e t e r o f a v e r a g e See the table 4.1

U p p e r d e v i a t i o n l i m i t

U p p e r d e v i a t i o n l i m i t o f a v e r a g e See the table 4.1

L o w e r d e v i a t i o n l i m i t

L o w e r d e v i a t i o n l i m i t o f a v e r a g e See the table 4.1

R e f e r e n c e d i a m e t e r o f X a x i s See the table 4.1

U p p e r d e v i a t i o n l i m i t o f X a x i s See the table 4.1

L o w e r d e v i a t i o n l i m i t o f X a x i s See the table 4.1

R e f e r e n c e d i a m e t e r o f Y a x i s See the table 4.1

U p p e r d e v i a t i o n l i m i t o f Y a x i s See the table 4.1

L o w e r d e v i a t i o n l i m i t o f Y a x i s See the table 4.1

D e v i a t i o n o u t p u t r a t i o , e n a b l e d w h e n p a r a m e t e r “ c o n _ s ” i s 1 . T h e d e v i a t i o n a n d t h e a n a l o g o u t p u t i s a l i n e a r r e l a t i o n s h i p , a n d t h e p a r a m e t e r ’ s v a l u e i s c o r r e s p o n d i n g t o t h e m a x a n a l o g o u t p u t v a l u e . T h e s e t t i n g r a n g e i s t h e s a m e a s t h e u p p e r ( l o w ) d e v i a t i o n

5.1 Parameter code

(26)

Main display option:

0: display XY average value

1: display XY average deviation value 2: display ellipse(|X-Y|)

3: display the diameter value of axis X 4: display the diameter value of axis Y 5: display the deviation value of axis X 6: display the deviation value of axis Y 7: display the position value of axis X 8: display the position value of axis Y

2 5 B / 4 0 B 5 0 B / 7 0 B

0 ~ 8 0

Main display option:

0: automatically display measurement value.When the value exceeds 99.999. The last decimal fraction is not displayed.If it is edge measurement, it will display edge measurement value; if it is one axis measurement, it will display one axis measurement value; if it is two axes measurement, it will display the average value of the two axes measurement. 1: if it is up measurement, it will display up measurement value; in other situation it will

display

2: if it is down measurement, it will display down measurement value; in other situation it will

display

3,4,5: correspond to the deviation value of 0,1,2

6,7,8: correspond to the laser axis position value of 0,1,2

9 0 / 1 0 0 A 0 ~ 8 0

Sub display option:

0: display the deviation value(the difference of the actual measurement diameter and the reference diameter)

1: display the position value(the measured objective is located in the vertical position of the beam)

2 5 A / 4 0 A 0 , 1 0

Sub display option:

0: display XY average value

1: display XY average deviation value 2: display ellipse(|X-Y|)

2 5 B / 4 0 B 5 0 B / 7 0 B

(27)

3: display the diameter value of axis X 4: display the diameter value of axis Y 5: display the deviation value of axis X 6: display the deviation value of axis Y 7: display the position value of axis X 8: display the position value of axis Y S u b d i s p l a y o p t i o n :

0: automatically display measurement value.When the value exceeds 99.999. The last decimal fraction is not displayed.If it is edge measurement, it will display edge measurement value; if it is one axis measurement, it will display one axis measurement value; if it is two axes measurement, it will display the average value of the two axes measurement. 1: if it is up measurement, it will display up measurement value; in other situation it will

display

2: if it is down measurement, it will display down measurement value; in other situation it will

display

3,4,5: correspond to the deviation value of 0,1,2

6,7,8: correspond to the laser axis position value of 0,1,2

(28)

Alarm control: 0: alarm disabled, 1: alarm enabled.

2 5 A / 4 0 A 0 、 1 0

Alarm output relay control: 0: alarm disabled,

1: alarm enabled when the average value is out of tolerance,

2: alarm enabled when the value from X(up) or Y(low) axis is out of tolerance,

3: alarm enabled when the value from X(up) axis is out of tolerance, 4: alarm enabled when the value from Y(low) axis is out of tolerance. 2 5 B / 4 0 B 5 0 B / 7 0 B 9 0 A 1 0 0 A 0 — 4 0 Ad va nc ed s et ti ng : i np ut p as sw or d t o e nt er s ub m en u A ll 5 d ig it s 1 2 3 1 2 A v e r a g e t i m e : s a m p l i n g s p e e d f o r d i a m e t e r d i s p l a y A l l 1 ~ 1 0 0 0 1 0 P I D c o n t r o l m o d e p a r a m e t e r P A l l 1 ~ 2 5 5 2 7 P I D c o n t r o l m o d e p a r a m e t e r I A l l 0 ~ 2 5 5 1 0 P I D c o n t r o l m o d e p a r a m e t e r D A l l 0 ~ 2 5 5 0 P I D c o n t r o l m o d e p a r a m e t e r A A l l 0 ~ 2 5 5 2 5 0 Voltage polarity:

0: positive (higher voltage for larger deviation value); 1: negative (lower voltage for larger deviation value).

A l l 0 , 1 0 C o n t r o l m o d e : 0 f o r P I D c o n t r o l , 1 f o r d e v i a t i o n v a l u e c o n t r o l A l l 0 , 1 0 T h e g a u g e ’ s a d d r e s s i n s e r i a l c o m m u n i c a t i o n A l l 0 ~ 1 2 7 1 B a u d r a t e f o r s e r i a l c o m m u n i c a t i o n w i t h t h e g a u g e A l l 1 9 2 0 0 9 6 0 0 4 8 0 0 2 4 0 0 1 2 0 0 9 6 0 0 S e l f - a d d r e s s f o r s e r i a l c o m m u n i c a t i o n w i t h p e r i p h e r a l e q u i p m e n t s A l l 0 ~ 1 2 7 1 B a u d r a t e f o r s e r i a l c o m m u n i c a t i o n w i t h p e r i p h e r a l e q u i p m e n t s A l l 1 9 2 0 0 9 6 0 0 4 8 0 0 2 4 0 0 1 2 0 0 9 6 0 0 F r a m e c h e c k m e t h o d : C R C c h e c k s u m o n l y . A l l 0 , 1 0 P a s s w o r d c h a n g e : c h a n g e p a s s w o r d t o e n t e r a d v a n c e d s e t t i n g A l l v e r s i o n n u m b e r ( s e c o n d a r y d i s p l a y s c r e e n d i s p l a y t h e i n s t r u m e n t m o d e l ) A l l

(29)

I n i t i a l s t a t u s a f t e r p o w e r o n A l l

W a i t i n g o r E U T b e y o n d m e a s u r e a r e a A l l

F a u l t A l l

U p p e r d e v i a t i o n v a l u e e x c e e d s 9 . 9 9 9 A l l

L o w e r d e v i a t i o n v a l u e e x c e e d s 9 . 9 9 9 A l l

Although the parameters of the two categories are different, yet the operating method of them is the same. In the next section, we will take one axis instrument for example.

(30)

The menu setting method of the remote controller is the same with the method of laser measurement instrument.

5.3.1 Bus

A standard 15-pin D-type socket contains serial I/O, analog output and over-deviation alarm switch. See Figure 12 for pins’ function.

P i n P i n c o l o r F u n c t i o n 2 O ra n g e A 3 O r a n g e - w h i t e B 5 Bl u e & B l u e - w h i t e D G N D 1 1 Bro w n D e vi a t io n / PI D 1 2 R e d D i a m e t e r 1 3 W h i t e G N D 1 4 Ye l l o w O v e r - d e v i a t i o n a l a rm o u t p u t 1 5 G re e n O v e r - d e v i a t i o n a l a rm o u t p u t O t h e r s N o n e N o n e Figure 12

 Serial I/O: for communicating with the gauge, using RS485, see Chapter 8 for detail.

 Analog output: two independent 12 bits D/A output (photoelectric isolation), 100

Ω

output impedance.

1. Diameter output: 0~10V, which can be calculated by the following formula:

Output Voltage =  Dim 10V 

000 .

25 × , here  Dim  is the measured diameter, 25.000 is the upper diameter range, and 10V is the maximum output voltage.

2. Deviation/PID output: the adjustable range is determined by “Output setting”, the default range is -2V~+2V. The output voltage can be calculated by the following formula:

5.2 Operation flow chart of remote controller

5.3 Output interface

﹠  control

(31)

Output Voltage = V max V   D  D dev ref  m × −

, here  Dm  is the measured diameter,  Dref   is the

reference value, V dev is the deviation limit (use upper limit if  Dm ≥ Dref , or lower limit if

 Dm < Dref ), and V max is the maximum output voltage, the default value of which is 2V.

 Over-deviation alarm output: the output contact points of the relay are normally opened, relay’s capacity: AC120V 1A/DC24V 1A.

5.3.2 RS232/RS485

The serial interface communicates with peripheral device using RS232/RS485 (customer could select while purchasing). Please see Chapter 8 for detail. The function of connecting cable is indicated by Figure 13. P i n P i n c o l o r F u n c t i o n 2 O ra n g e A(T X) 3 O ra n g e -wh i t e B(R X) 5 Bl u e & B l u e - w h i t e D G N D o t h e r s N o n e N o n e Figure 13

(32)

6

YK-1B Remote Controller

The YK-1B remote controller has same functions with YK-1A.

The functional layout is indicated by Figure 14.

Figure 14(a)

Figure 14(b)

1. Main display screen: displaying diameter value and parameters

2. Sub display screen: displaying deviation value, offset, and parameter code 3. Power indicator: the indicator is on when powered on

4. Auto control indicator: the indicator is on when the PID function is enabled

5. Auto control key: press this key to enable the PID function, press again to disable it 6. Functional key: turn up

7. Functional key: turn down 8. Menu key:

(33)

图 15

图 16

 模拟量输出:Analog output of diameter

 上位机通讯:RS485/232 with peripheral devices(computers, for example)

 测径仪通讯:RS485/232 with the gauge.

 接地:Ground

 220VAC输入:AC220V power supply

 上超差触点输出:Upper deviation output

 下超差触点输出:Lower deviation output

 模拟量公共端:The common terminal of analog

6 1 Control panel

(34)

7 Remote Display

Remote displayer is applied for displaying in long distance, which is connected with laser gauge by D-type socket (15pins/25pins).The transmission distance can amount to 200 meters.

Remote displayer (YX-□A)is connected by electric current loop, but user needs to provide connecting wire.YX-0.5B is the dual Remote displayer, which displaying content is in accordance with the date of laser gauge.

Connecting: the D-type socket(15pins)’s 2-pin 、10-pin or D-type socket(25pins)’s 12-pin 、24-pin connect with the D-type socket(15pins)’s 2-pin 、10-pin of the laser gauge at random, no need to distinguish voltage  polarity.

(35)

Figure 17

1. Display screen

2. Upper deviation light 3. None deviation light 4. Lower deviation light

5. Synchronized data channel 6. Power socket

7. Swinging ring

2

(36)

8 Installation and Operation

The installation of laser gauge has two forms---laying and standing[See Figure 18].When installing bracket, first lock the fastness staff on the ground with an expanding bolt  ; put in the adjustment staff; lock the splint to the adjustment pole with locking screws; lock the measurement instrument on the splint; adjust the height;lock the handle knob.

When loosing the handle knob,be careful that the adjustment pole will spring.

Figure 18

The measured objective should be in the guide roller and vertical to beam. Adjust the holder to making guide roller hold wire slightly, which can reduce objective

Splint locking screw screw Adjustment pole Handle knob Fixing pole Expanding screw

Diameter measurement instrument

Standing

Laying

Bracket Installation

(37)

dithering when moving. See Figure 19(25AL for example).

Measured wire is not in the center of guide roller

Incline of the wire results in the measured di ameter value larger than the real one

Figure 19(a)

For objectives of different diameters, the installation height of guide roller should  be 0.1-20mm,20-25mm or 0.1-10mm,10-20mm,20-25mm on purpose of placing the

objective near the center of beam to measure exact value.

Opmac 25AL Opmac 25BL

Figure 19(b) Figure 19(c) 20~25mm 0.1~20mm 20~25mm 10~20mm 0.1~10mm

(38)

When powering on, the measurement instrument is in initialization status, showing on screen and enters measuring status after 3 seconds, showing on main display screen. Placing wire in measuring area, the value on main screen is the diameter of wire and value on sub display screen is offset.

When powering on traction machine, the wire moves forward and diameter changes accordingly. The changing speed is depending on the average time [See Section 5.1]. The higher the value is ,the slower the diameter changes. Usually it is set to 10.

Set proper reference diameter, upper over-deviation limit and lower over-deviation limit. [See Section 4.2] when measured diameter is larger than reference diameter plus high over-deviation limit, “ ” light on; when measured diameter is less than reference diameter subtracted by low over-deviation limit, “ ” light on; when reference diameter subtracted by low over-deviation limit is less than or equals to measured diameter is larger than or equals to reference diameter plus high over-deviation limit, “ ” light on.

Once measured diameter exceeds upper and lower limits, over-deviation alarm contact closes. When sound and light alarm enables, operation panel gives an alarm and at same time the over-deviation light on top flickers.

The laser gauge and remote controller compose of feedback system, which adjusts speed of traction machine or extruder according to deviation of measured value and pre-set wire dimension. Here is an example of the control of wire/cable production line and please follow this one for others applications.

8 . 3. 1 In s t a l l a t i o n p o s i t i o n

The measurement instrument [See Figure 20 for installation position] can be installed in front of cooling flume [See Figure 20(b)], that is good position for rapid feedback control but the measured diameter is heated value of wire with a certain difference from cooled value(real value). In addition, due to the soft scarfskin of high temperature near extrusion port, guide roller cannot be installed. If the measurement instrument is installed behind cooling flume and dryer [See Figure 20(c)], the measured value is real value after cooling but the delayed feedback on dimension change effects control precision. Remote controller usually is installed on the electric cabinet and connected with measurement instrument through cable [See Figure 20(a)].

 Alm+

 Alm+

Norm

8.2 Basic operation

(39)

(a)

Measurer installed in front of flume

( b) Measurer installed behind flume

(c) Figure 20

8.3.2 C o n n e c t i o n

First of all, power off measurement instrument, remote controller and traction machine and open traction machine controller (could be DC motor controller or inverter), and then break the connection between the 10V power supply to the power pin of the regulation resistance and after

Creel stand   Spark

inspector    Laser 

measurement Cooling tank Extruder 

Feeding reel   Length

measurer  Creel stand   Spark inspector  Laser measurement Extruder 

Feeding reel Cooling tank   Length measurer  Distributing cabi

(40)

this, connect the white cable of “bus & control” interface on remote controller to that power pin, and the brown one to the 10V power supply. [See Figure 21]

Figure 21

 Note: when connected with the regulation resistance of the inverter, analog output cable should be as short as possible and the shielding braid must be connected with shielding cables of the inverter. 8.3.3 Testing and manufacture

8.3.3.1  Power on measurement instrument and remote controller and check parameters: average time=50, voltage polarity=0, control mode=0, P=15、I=14、D=0.

8.3.3.2  Place a wire on the measurement instrument and set reference diameter, 1mm less than

measured value. Power on traction machine controller and set traction speed to 0. Press to

enter control mode. If within 20 seconds, the traction machine is not working, which means a wrong connection (see Section 8.3.2 for details), just exchange positive pin and negative pin of the analog output cable or set polarity to 1.

Because the feedback control can’t work with wrong connection, so user should follow the instructions in Section 8.3.3.1 and Section 8.3.3.2 to ensure right connection.

8.3.3.3  While starting to manufacture, first of all, power on measurement instrument and remote controller, not enter automatic control status. Set reference diameter according to technical demand. If the measurement instrument is installed behind flume, the reference diameter can be set as real diameter. If the measurement instrument is installed in front of flume (displaying heated value), the reference value can be set a little bit higher.

8.3.3.4 Manually adjust extruding and traction speed to make the measured data close to reference diameter and after a while, when measurement data is stable (there is inertia of extruding and traction speed), press “ ” for automatic control. Under normal working circumstance, traction speed is about 1200r/m, should not be too high otherwise there is no adjusting scale of the remote controller. If the traction speed is too low while working, just manually set it at about 1200r/m. After completing manufacture, press “ ” to stop automatic control and then  power off extruder and traction machine.

8.3.3.5  The P and I parameters of remote controller decide the stability of control. The defaulted  parameters could work well under most situations. When the control is too slow or non-stable with

(41)

a. When there is deviation between measured diameter and reference (positive or negative) lasting for a long time [see Figure 22(a)], it suggests that the system has “steady-state error”. Increase  parameter I to remove “steady-state error”.

b. When measured diameter is changing constantly up and down near reference for a long time [see Figure 22 (b)], it suggests that system has “vibration”. Reduce parameter I to control “vibration”.

c. When measured diameter rapidly reaches reference and vibrates in control mode [see Figure 22(c)], it suggests that system is “overshoot”. Reduce parameter P to eliminate it.

d. When measured diameter costs long time to reach reference or could not reach in control mode [see Figure 22(d)], it suggests that system is “undershoot”. Increase parameter P to eliminate it In a word, parameter P decides adjusting speed, parameter I decides control precision. The ideal control curve is indicated by Figure 22(e).

(42)

The parameter must be saved after adjustment in case of loss when powering on next time. (a) (b) (c) (d) (e) Figure 22

Note: “ ” denotes reference diameter “ ” denotes measured diameter

Vibrating

Overshoot

Undershoot

Ideal control curve

Steady-state error (high)

Reference Value

Measured Value Measured Value

Reference Value Steady-state error (low)

(43)

9

Communication

The remote controller or peripheral device uses RS485 to communicate with the gauge, and  peripheral devices use RS232/RS485 to communicate with remote controller. The protocols for both serial communications are the same, except some parameters are special for remote controller. Table 2 shows parameters for transmission and indicates those only for remote controller. See section 4.1 and 5.1 for detail of parameters’ range.

When remote controller or peripheral device communicates with the gauge, the former is sender and the latter is receiver. When peripheral device communicates with remote controller, the former is sender and the latter is receiver. That means remote controller is a sender and a receiver in the same time when it is connected with peripheral device and the gauge.

Data format: 1 star bit, 8 data bits and 1 stop bit, no parity checkout. Transmission format of the sender when reading parameters:

g a u g e a d d re ss p a ra me t e r

Transmission format of acknowledge of the receiver:

g a u g e a d d re ss p a ra me t e r 2 -b yt e d a t a ( 90/100AL are 3

bytes,f r o m h i g h t o l o w ) C R C c h e c k s u m

Transmission format of the sender when writing parameters:

g a u g e a d d re ss p a ra me t e r 2 -b yt e d a t a( 90/100AL are 3

bytes,f r o m h i g h t o l o w ) C R C c h e c k s u m

There is no acknowledge from the receiver when the sender writing parameters.

CRC check sum is the CRC sum (1-byte) of all bytes before it. CRC checking uses 8 digits CCITT cyclic redundancy check (CRC-CCITT) and its generator polynomial is: G(x)=X8+X5+X4+1. The negative data is denoted in complement form.

 Example 1: reading diameter.

Suppose the address of the receiver is 1 (01H), and current diameter is 6.234(185AH), then the sender sends:

01H 41H

And the acknowledge of the receiver is:

01H 41H 18H 5AH CRC-check-sum

 Example 2: reading offset

Suppose the address of the gauge is 1(01H), and current offset is -5 then the sender sends:

01H 44H

And the acknowledge of the receiver is:

(44)

Suppose the address of laser diameter measurement instrument is 1(01H), and the sender is going to change reference diameter to 60.00(1770H) then it sends:

01H 66H 17H 70H CRC-check-sum

And there is no acknowledge from the receiver.

 Example 4

reading the diameter value of 100A

Suppose the address of laser diameter measurement instrument is 1(01H),The value of current diameter is 106.350(19f6eH),well then:

Input:

01H 41H

The laser diameter measurement instrument output:

01H 41H 01H 9FH 6EH CRC-check-sum

For reading and writing other parameters, please follow the examples.

Read p a r a m e t e r W r i t e p a r a m e t e r O n e a x i s Tw o a x i s Ty p e  A S C I I H e x A S C I I H e x A 4 1 H N U L L N U L L D i a m e t e r V a l u e A v e r a g e d i a m e t e r v a l u e R e a d o n l y B 4 2 H N U L L N U L L N U L L D i a m e t e r v a l u e o f X a x i s C 4 3 H N U L L N U L L N U L L D i a m e t e r v a l u e o f Y a x i s D 4 4 H N U L L N U L L O f f s e t O f f s e t o f X a x i s E 4 5 H N U L L N U L L N U L L O f f s e t o f Y a x i s F 4 6 H f 6 6 H R e f e r e n c e d i a m e t e r R e a d & w r i t e G 4 7 H g 6 7 H U p p e r d e v i a t i o n p r e s e t H 4 8 H h 6 8 H L o w e r d e v i a t i o n p r e s e t I 4 9 H i 6 9 H S u b d i s p l a y o p t i o n J 4 a H j 6 a H A l a r m o p t i o n[ 1 ] K 4 b H k 6 b H A v e r a g e t i m e L 4 c H l 6 c H  P p a r a m e t e r M 4 d H m 6 d H  I p a ra m e t e r N 4 e H n 6 e H  D p a r a m e t e r O 4 f H o 6 f H α p a r a m e t e r P 5 0 H p 7 0 H V o l t a g e p o l a r i t y Q 5 1 H q 7 1 H C o n t r o l m o d e R 5 2 H r 7 2 H  A u t o m a t i c C o n t ro l  [ 2 ]  Table 2 parameters

(45)

 Note[1]: It is the alarm option when communicating with the gauge and the over-deviation alarm output option when communicating with remote controller.

 Note[2]: The A u t o m a t i c C o n t r o l parameter is the same as the automatic control key in the panel of remote

controller, 1 for starting auto control, 0 for exit. Additional parameters for Opmac 5AL:

Rea d p a r a m e t e r   W r i t e p a r a m e t e r   T 54H T 74H A d d r e s s f o r s e r i a l c o m m u n i c a t i o n . R a n g l e : 0 ~ 1 2 7 . D e f a u l t v a l u e : 1 . R e a d & w r i t e U 55H U 75H B a u d r a t e f o r s e r i a l c o m m u n i c a t i o n . D e f a u l t v a l u e : 1 9 2 0 0 . R a n g l e : 0 f o r 9 6 0 0 1 f o r 1 9 2 0 0 2 f o r 3 8 4 0 0 3 f o r 5 7 6 0 0 4 f o r 1 1 5 2 0 0 V 56H v 76H C o m m u n i c a t i o n m o d e . D e f a u l t v a l u e : 0 . R a n g l e : 0 for CRC checksum 1 for XOR checksum 2 for MODBUS RTU mode

MODBUS

Opmac  series instruments with Version 3.02 or above can communicate with peripheral devices via Modbus(RTU mode). Users can select this protocol in the parameter menu, see section 4.1 for detail. Read parmeters in table 2 are the register address in Modbus.

 Example 5: reading diameter value via Modbus

Suppose the address of laser diameter measurement instrument is 1(01H),The value of current diameter is 6.234(185aH),well then:

Input:

01H 03H 00H 41H 00H 01H D4H 1EH

The laser diameter measurement instrument output:

01H 03H 02H 18H 5aH 32H 7FH

.

escription Type

(46)

10

Overall Dimensions

The outline and overall dimensions of instrument are indicated by Figure 23.

Figure 23(a)Opmac 25AL

(47)

Figure 23(c)Opmac 40AL

+

(48)

Figure 23( e) Opmac 100AL

(49)

Figure 23( g) Opmac 50AL

(50)

Figure 23( i) Opm ac 70B L

Figure 23( j) YK- 1A Re mote Contr oller

(51)

Figure 23( l ) Y X - 0 . 5 A remote displayer  Figure 23( m ) Y X - 0 . 5 B remote displayer 

(52)

Figure 23(o)Bracket       7       6       0    ~       9       6       0       1       5       0 270 240       1       2       0 4-φ13

(53)
(54)

11

Maintenance

11.1 Please keep the protection lens clean, which are at light-in area and light-out area of the measurement instrument while working. Use lens paper to clean dust .

11.2  Dry the wire from flume while using the production line in case of water on wire surface defect the measurement precision.

11.3 Reference bar should be cleaned with petrol or alcohol and rubbed with lens paper or absorbent cotton before use and smear oil after use to prevent rust.

11.4 suggests malfunction, maybe the scanning motor stopped, the laser generator fault

or measuring area is blocked.

12

Malfunction and Solutions

12.1 No scanning light at light-in area, showing on screen.

 Scanner is not working or no output of laser---internal malfunction

 Dislocation of protection lens cover in light-out area blocks scanning beam, causing no scanning signal. Just loosen the fixing snail of lens and adjust its position.

12.2 Only part of the scanning light is on light-out area, while showing on screen.

 If there is any object on protection cover or lens which shields the scanning beam, just clean it with dehydrated alcohol and rub with lens paper.

 If the protection cover in light-in and light-out area is dislocated blocking scanning light,  just re-locate the cover.

12.3 While EUT in measuring area but showing on screen or flickering with measured

value alternately.

 If there is any dust near light-in and light-out area which reduces light, just clean lens.

 If scanner’s speed is low or laser beam is reduced, it is inner malfunction. 12.4  No display after powering on and no reaction pressing any key.

 Damage of display panel or inner malfunction.

 Dislocation of measurement instrument when powering on shortly after power off. Just  power it off and again on it after 10 seconds.

12.5 Serious measuring error 

 If there is dust on protection lens effecting precision, just clean lens.

 If EUT is dislocated from measuring center or dithering , just adjust bracket and place the object in center of measuring area of the gauge.

12.6 Communication fault

 Wrong connection, please see Figure 13.

(55)

 Damage of inner communication module, please contact us for maintenance. 12.7 Analog output fault

 Wrong connection, please see Figure 12.

 Short circuit of peripheral interface, please check peripheral connection.

 Damage of measurement instrument’s interface.

12.8 The instrument display“Err-1”.

 The laser diameter measurement instrument of software after V1.07 version and the Remote controller of software after V1.06 version have this function.

 This phenomenon is caused by the within data of the instrument are littered by accident. At this time, enter into the second menu, check the parameters and the measuring data, under the condition of no problem, save the data, “Err-1” display disappear.

12.9 The instrument display “Err-2”

 The laser diameter measurement instrument of software after V1.07 version and the Remote controller of software after V1.06 version have this function.

 This phenomenon is caused by that when read the exterior backup parameters, the exterior  backup parameters come forth error. Please contact with our company’s technician.

 Note: If there is any inner malfunction of the mea surement instrument, please contact us for maintenance.

13

Accessories

Power cable. 2 pcs

Master bar. 1 pcs

Cleaning ball. 1 pcs

Lens paper. 1 copy

Control cable. 1 pcs

Serial communication cable. 1 pcs

Conform to delivery list.

14

After-sale service

Within one year, if the machine has fault, we guarantee for maintenance and change if the measurement instrument is properly stored and used.

References

Related documents

This article examines the White Paper’s treatment of China’s rise, and of the US-China relationship, the newly introduced construct known as the ‘Indo-Pacific

Developments in technology will continue to influence all facets of life, including criminal law and the way law enforcement will attempt to protect social

If we now recall from the rational expectation versions of the Philips curve literature that inflation rates are basically a function of monetary policy and the exogenously

Functional-association outcome RIDLs + transcript length + exonic conservation , in which functional-association outcome indicates the traits defined above; RIDLs indicates the

In Kenya, East African Breweries Limited, a subsidiary of Diageo, is the second largest listed company with a market capitalisation of US$2.4 billion, or 10.5% of the Nairobi

The Franco-Prussian War and the subsequent unification of Germany, occurring within a single year (1870/71), offer a great deal of academically unexplored facets, including

The purpose of this conference was to respond to some of these problems by bringing together mental health clinicians, substance abuse/ dependence counselors, VR

Recalling what we have discussed above concerning the centralized reverse supply chain model, for time-insensitive products, the centralized approach would be adopted