AMICO Information

Management System (AIMS)

ALERT- 2

(2)

Table of Contents

INTRODUCTION ...3

Theory of Operation ... 3

Network Diagram ... 4

INSTALLATION ... 5

Medical Gas Alarms ... 5

Network Interface Module ... 5

Connecting AREA / Master ... 6

Connecting NIM Modules ... 6

Host P.C. ... 7

Detecting and Verifying Alarms ... 7

Selecting Scan Speed Rate ... 8

Profile Colour & Manifold FSD Features ... 9

Alarm Order ... 10

Labelling Properties ... 11

Log File ... 13

APPENDIX A - Wiring Diagram Area Display to AIMS... 14

APPENDIX B - Wiring Diagram Master Module to AIMS ... 15

APPENDIX C - Wiring Diagram Manifold ... 16

APPENDIX D - S1 Dip Switch Settings (Set Node ID Address) ...17-19 APPENDIX E - PC Installation ... 20

Installation of the Device Driver ... 21

AIMS Universal Board APPENDIX F - AIMS Specifications for Universal Board ... 22

APPENDIX G - S1 Dip Switch Settings (Universal Board Setting) ...23

APPENDIX H - Changing Pressure Units on AIMS Universal ... 24

APPENDIX I - Data/Inputs from Sources of Supply ... 25

(3)

INTRODUCTION

The Amico Information Management System (AIMS) provides hospital personnel with real time information on a P.C. of the status of the Medical Gas device. The system consists of three parts:

1. Medical Gas devices (ie: Area Alarm, Master Alarm and Manifold). 2. Polling Network.

3. Host P.C. with PC Interface Card and AIMS software.

Theory of Operation

The system is intended to provide real time information to hospital personnel on the status of the medical gas pipeline.

AIMS is a computerized network system which will poll every medical device in a hospital and display each one on the screen of a Host P.C. The polling is automatic and independent of the scanning rate.

The network topology is shown in Fig.1, on page 4. The Host P.C. screen running AIMS will display every medical device as a “button”, which will be “green”, if the alarm is normal and “red” if the medical device has gone into an alarm condition. In the event all medical devices are “normal” with no fault conditions, all the “buttons” will be green indicating a normal condition. In this mode, every medical device will be displayed on the bottom part of the screen in turn, as it is scanned depending on the scan speed selected. When all the alarms have been scanned and displayed, the cycle will be repeated again. If the buttons are “white” in color it indicates an existing medical device with no power supplied or the communication link is broken.

In the event an alarm occurs in the hospital, the scanning of all medical devices stops, and the medical device, which is in alarm condition is displayed continuously. In the event other medical devices on the network go into an alarm condition, the scanning of all medical devices which are in the alarm condition (red) begins and continues until all the faults have been cleared. Once all the medical devices have cleared and all the buttons are “green”, the normal scanning and

(4)

This figure shows a typical installation of an AIMS System. Medical Gas Devices are connected to the polling network which terminates at a Host P.C. The P.C. can be installed anywhere in the hospital provided that the maximum length of the cable run on the polling network is 5,000 ft and #18 to #22 gauge. The P.C. can be accessed remotely during unattended operations using standard software not supplied by Amico (i.e. PC Anywhere).

NETWORK DIAGRAM

(5)

INSTALLATION

Medical Gas Alarms

The AIMS System will operate with the Amico ALERT-2 microprocessor based Medical Gas Alarm. For installation of the Medical Alarm please refer to Amico’s “Operation & Maintenance Manual” for the ALERT-2.

Network Interface Module

Each Medical Gas Alarm (Master or Area) requires one Network Interface Module (NIM) to be installed within the local alarm box to allow that alarm to be placed on the Network. Please ensure the Medical Gas Alarm is turned OFF during the installation of the Network Interface Module.

This module has exactly the same dimensions as any Alarm Module and hence it can be mounted on the frame of the Medical Gas Alarm. If the frame is fully populated with Alarm Modules, the NIM card can be mounted on the 4 standoffs in the alarm box as shown in Fig. 2.

Once the NIM card is in place, the ribbon cable must be inserted into the connector in order to power the module. This might require the entire ribbon cable connecting all the modules, to be replaced with a longer one, if the existing ribbon cable will not reach.

Fig. 2

(6)

Connecting Area NIM:

Each Area display module must be connected to the Area NIM card so that the information for that gas can be transmitted to the Host P.C. This is accomplished by connecting one wire from the “SLAVE” (-) terminal strip on the display module to the “LOCAL” terminal strip on the NIM module. The first display module (near the annunciator module) should be connected to position 1 on the “LOCAL” terminal strip of the NIM module, the second to position 2 and so on until all the display modules have been connected to the NIM module. Use #22 to #24 gauge hook-up wire (not supplied). Please refer to APPENDIX - A (page 14).

For proper communication between the display module(Area Board) and the NIM, dip switch SW 10 on S3 of all display modules has to be set to the “ON” position.

Connecting Master NIM:

Each Master module must be connected to the Master NIM card so that the information for that module can be transmitted to the host P.C. This is accomplished by connecting a cable from the 24 pin-female socket connector on the master module to the 24 pin-male socket connector on the Master NIM. The first Master module (near the annunciator module) should be connected to “Master 1” on the NIM module. The second Master module should be connected to “Master 2” on the NIM module and so on until all Master modules have been connected to the Master NIM module. Please refer to APPENDIX -B (page 15).

Each NIM (Area and Master) module must have a unique address as set on the dip switches # S1. The sequence on the network of these addresses are not important. The eight position NIM dip switch settings (S1) are to be set from 1 to the maximum number of alarms on the network and the setup screen on the AIMS Software should then be matched with the appropriate number. Refer to APPENDIX - D (page 17) for dip switch settings. It is very important that each NIM module has an unique address. Having two or more NIM modules with the same address will give unpredictable results.

Connecting NIM Modules:

To connect the NIM module to the polling network, the network wire coming into the Alarm must be terminated on the “NETWORK” terminal strip. The incoming wire should be connected to “OUT+” and “OUT-”. The outgoing wire should be connected to “IN+” and “IN-”. The “Network” connection “In” or “Out” is irrelevant, as long as the + and - are matched. Use #18 to #22 gauge twisted pair cable. The four position dip switches (S2*) should be set as shown. Set all switches to “off” except the last NIM Module on the Network, which should be set to “on, on, off, off”. The Alarm is now connected to the network.

21 21

6

6

)RUDOO1,0 H[FHSW/DVW )RU/DVW 1,0RQO\ 21 6

(7)

Host P.C.

The NBS-42 card must be connected with a 9-socket male connector, provided with the AIMS system. The cable supplied has a length of 5 feet, stripped one inch, exposing the black (-) and red (+) wires. Connect this wire to the first alarm in the AIMS Network terminal marked “OUT+” and “OUT-”.

Connect the Network cable to the back of the P.C. If the P.C. is not supplied by Amico, see APPENDIX - E (page 18) for instructions on how to install the Network Interface into the customer supplied P.C. The AIMS software runs as an independent application on the P.C. running under the Windows 98, 2000 and XP operating systems. To install AIMS on the Host P.C. Insert the AIMS CD disk provided by Amico into the CD drive follow the procedures on the screen:

Detecting and Verifying Alarms

1 Select the “Options” command at the top of the screen.

2 Click “Setup” and a dialog window will appear. 3 Move mouse pointer to the “Primary ISA Card based”

symbol and right click.

4. Ensure that ALL active medical devices are powered ON. Click on “Enumerate”.

The AIMS system will automatically poll the network and locate the active alarms. Properly connected medical devices will show up as “green” buttons on the screen.

(8)

DISPLAY SPEED

Selecting Scan Speed

The AIMS system will display alarms in three different scanning modes as required by the hospital staff: SLOW , MEDIUM and FAST. The rate of speed that the alarms are polled can be selected either by the hospital staff or the factory default can be used.

To change the scanning speed rate, click on “scan/stopped” and select the “Scanning” command at the top of the screen. Select your choice of “Slow Scan”, “Medium Scan” or “Fast Scan”.

Factory default settings: Slow = 5000

Medium = 3000 High = 500

Selecting Scan Rate

You can change the scanning rate by selecting a value from 1 to 9999 (1 = fast, 9999 = slow) in the “Setup Dialog” window. To change the scanning speed rate, select: “Options” | “Setup” | “Primary AIMS Server” symbol.

The server properties window will show the “Slow Scan Rate”, “Medium Scan Rate” and “Fast Scan Rate” values. Select the desired Scan Rate and double click the value. Enter your desired speed rate from 1 to 9999.

(9)

Profile Features

The AIMS profile colour scheme of Medical Gases is preset by the factory as USA mode (NFPA colour coding). To change the colour scheme to ISO mode (Canadian colour coding), select:

“Options” | “Setup” | “Amico Server” symbol. Click on the symbol to access the Organization Properties window. The window will show the colour scheme as the last icon on this window. Double click to change from one mode to the other. When selection has been determined, click on the “Close” button.

Manifold FSD Feature

The Full Scale Deflection on the manifold allows the AIMS manifold image to match the represention on the screen, for example the default setting for a manifold tank is 3000 psi so when a tank is 1500 psi the manifold tank picture on the screen is half full. The scale can be change to match the gas cyclinders by changing the FSD.

“Options” | “Setup” | “ and click on the manifold iconon the right side of the property window The left and right bank will be set to 3000. Simply double click on the number and change it to the required setting. When the selection has been made click on the “Close” button.

(10)

Alarm Order

The Alarm order feature allows the hospital personnel to move the medical device into any position (ID address) on the AIMS layout screen. 1. To change the alarm order, select:

“Options” | “Setup”, then move the mouse to the “Primary ISA Card based” symbol. 2. Right-click on the ISA symbol. Select

“Properties”. The “Network Properties: window will open.

3. Select “Alarm Order” tab.

4. Click on the ID Address of the alarm you wish to move.

5. Select the “Move Up” and/or “Move Down” buttons in order to move the device to the desired position.

(11)

Labeling Properties

The Labeling Properties Window allows the user to label the actual Green Text Button Display and the White Window Box Display. 1. To label the Text Button and the Window

Display select.

“Option” | “Setup”, then move the mouse pointer to the Alarm Symbol.

2. Right - click on the Alarm symbol. Select “Properties”. The Alarm Properties Window will open.

3. Click on the “Button Text” space and type

in the text that will be used in identifying that Alarm Unit.

4. Click on the “Message Text” space and type in the text that will be used in identifying that Alarm.

(12)

Labeling Master Module

Begin by clicking “Scanning” | “Stop” to stop scanning and then “Options” | “Setup...” | , then move the mouse pointer to the Master Alarm Symbol and click on that small icon. To begin labeling the Master Module, click with the mouse button on the button under the heading called “Display Text [1,1], double click and a small box will appear on the left of that window. Enter your desired text (i.e.: Vacuum Low Line pressure. After the line is complete, simply click on the next “Display Text [1.2]

In order to separate the text into two lines, you must place the following symbol”^” (Shift 6) after the first line (i.e.: OXYGEN^High Line Pressure

To add three lines to one legend, simply add another symbol “^” in the text line (i.e.: OXYGEN^High Line^Pressure). NOTE: By adding more lines to the text, the font size decreases.

OXYGEN High Line Pressure

OXYGEN

High Line Pressure

(13)

Log File:

A log file represents the record of events for alarms on the AIMS network. To enable the log file:

Click on “Scan” | “Stopped” | “Option” | “Setup...”. Move mouse pointer to the “Primary ISA Card based” symbol and right click | Click on “Properties” and then “Logfile”

Ensure that “Enable File Logging” is selected. Assign a file name to system “LOG File Name” such as:

AIMS.LOG

. Click “OK” when completed.

To view the log file:

Close AIMS program. Click on “Close” and then “File” and then “Exit”. This will unlock the log file from the AIMS program. Open the C:\AIMS\AIMS.LOG file with notepad or import the file into a spreadsheet (i.e.: MICROSOFT EXCEL). Once in a spreadsheet file you can create Tables and Graphs then print them out as a record.

(14)

APPENDIX - A

8 21 6 6 8 8 8 8 8 8 ,1,1287287 /2&$/ 1(7:25. 521 6 6 . . 8

8 8 8

8 3 $/ 8 8 5 6 21 6 6 3 $/ 8 8 8 8 8 8 83 '2:1 6/$9( 6(1625 /2:$/$50

121&&20 +,*+$/$50121&&20

$UHD

'LVSOD\

0RGXOH

1,0

0RGXOH

7R0RGXOH 7R0RGXOH 7R0RGXOH 7R0RGXOH 7R0RGXOH WR *DXJH7ZLVWHG 3DLU:LULQJ

:LULQJRI1,00RGXOH

WR$UHD0RGXOH

6HWWR2II2II2II2II )RU/DVW1,0RQO\ VHWWR2Q2Q2II2II 621 3RVLWLRQ 66HW,QGLYLGXDO 1RGH,'$GGUHVV WR *DXJH+RRN 8S&DEOH

(15)

APPENDIX - B

21 ,1,1287287 1(7:25. 6 0DVWHU 1,0 0RGXOH 0$67(5 0RGXOH WR *DXJH7ZLVWHG 3DLU:LULQJ

:LULQJRI0DVWHU1,00RGXOH

WR0DVWHU0RGXOH

)RUDOO1,0 VHWWR2II2II2II2II )RU/DVW1,0RQO\ VHWWR2Q2Q2II2II 5LEERQ&DEOHIRU³0DVWHU´ 7\SLFDOIRUHDFK0DVWHU0RGXOH 1RWH0DWFKWKHUHG'RWVRQWKH FLUFXLWERDUGVWRWKHUHGOLQHRQ WKHULEERQFDEOH 8 8 8 8 8 8 8 8 8 8 0$67(5 0$67(5 0$67(5 0$67(5 0$67(5 0$67(5 8 8 8 &+ &+ &+ &+ &+ &+ &+ &+ &+ &+ 8 21 6 66HW,QGLYLGXDO 1RGH,'$GGUHVV - - -- -

(16)

-APPENDIX - C

Wiring of Manifold

+

12V +5 +5V COM

V

+

PWR IN LINE PRE LIMIT SW

LEFT RIGHT

LEFT HP RIGHT HP

+5 V +I N +I N NC NC

-GN D NO GN D GN D CO M CO M Power Supply Left Limit Switch Line Transducer Master Module Right Limit Switch Right HP Transducer

NO Contact (held closed by Relay) To Master Alarm

DIGITAL MANIFOLD WIRING SCHEMATIC

-I

N -IN

Left HP Transducer VERSION 3.0 CO M NO +I N -IN +O UT -O U T A.I.M.S. CONNECTION Optional Remote Buzzer

Ensure Dip Switch 7 is off if no heaters are installed

S2 For all NIM

set to: Off, Off, Off, Off For Last NIM only set to: On, On, Off, Off S1

Set for individual node ID address

AIMS connections #18 to #22 gauge twisted pair wiring

(17)

APPENDIX - D

S1 Dip Switch Settings

(Set Network Address)

8 7 6 5 4 3 2 1 8 7 6 5 4 3 2 1

D o N o t U se o n o n o n o n o n o n o n o n Ala r m # 4 2 o n o n o ff o n o ff o n o ff o n

Ala r m # 1 o n o n o n o n o n o n o n o ff Ala r m # 4 3 o n o n o ff o n o ff o n o ff o ff

Ala r m # 2 o n o n o n o n o n o n o ff o n Ala r m # 4 4 o n o n o ff o n o ff o ff o n o n

Ala r m # 3 o n o n o n o n o n o n o ff o ff Ala r m # 4 5 o n o n o ff o n o ff o ff o n o ff

Ala r m # 4 o n o n o n o n o n o ff o n o n Ala r m # 4 6 o n o n o ff o n o ff o ff o ff o n

Ala r m # 5 o n o n o n o n o n o ff o n o ff Ala r m # 4 7 o n o n o ff o n o ff o ff o ff o ff

Ala r m # 6 o n o n o n o n o n o ff o ff o n Ala r m # 4 8 o n o n o ff o ff o n o n o n o n

Ala r m # 7 o n o n o n o n o n o ff o ff o ff Ala r m # 4 9 o n o n o ff o ff o n o n o n o ff

Ala r m # 8 o n o n o n o n o ff o n o n o n Ala r m # 5 0 o n o n o ff o ff o n o n o ff o n

Ala r m # 9 o n o n o n o n o ff o n o n o ff Ala r m # 5 1 o n o n o ff o ff o n o n o ff o ff

Ala r m # 1 0 o n o n o n o n o ff o n o ff o n Ala r m # 5 2 o n o n o ff o ff o n o ff o n o n

Ala r m # 1 1 o n o n o n o n o ff o n o ff o ff Ala r m # 5 3 o n o n o ff o ff o n o ff o n o ff

Ala r m # 1 2 o n o n o n o n o ff o ff o n o n Ala r m # 5 4 o n o n o ff o ff o n o ff o ff o n

Ala r m # 1 3 o n o n o n o n o ff o ff o n o ff Ala r m # 5 5 o n o n o ff o ff o n o ff o ff o ff

Ala r m # 1 4 o n o n o n o n o ff o ff o ff o n Ala r m # 5 6 o n o n o ff o ff o ff o n o n o n

Ala r m # 1 5 o n o n o n o n o ff o ff o ff o ff Ala r m # 5 7 o n o n o ff o ff o ff o n o n o ff

Ala r m # 1 6 o n o n o n o ff o n o n o n o n Ala r m # 5 8 o n o n o ff o ff o ff o n o ff o n

Ala r m # 1 7 o n o n o n o ff o n o n o n o ff Ala r m # 5 9 o n o n o ff o ff o ff o n o ff o ff

Ala r m # 1 8 o n o n o n o ff o n o n o ff o n Ala r m # 6 0 o n o n o ff o ff o ff o ff o n o n

Ala r m # 1 9 o n o n o n o ff o n o n o ff o ff Ala r m # 6 1 o n o n o ff o ff o ff o ff o n o ff

Ala r m # 2 0 o n o n o n o ff o n o ff o n o n Ala r m # 6 2 o n o n o ff o ff o ff o ff o ff o n

Ala r m # 2 1 o n o n o n o ff o n o ff o n o ff Ala r m # 6 3 o n o n o ff o ff o ff o ff o ff o ff

Ala r m # 2 2 o n o n o n o ff o n o ff o ff o n Ala r m # 6 4 o n o ff o n o n o n o n o n o n

Ala r m # 2 3 o n o n o n o ff o n o ff o ff o ff Ala r m # 6 5 o n o ff o n o n o n o n o n o ff

Ala r m # 2 4 o n o n o n o ff o ff o n o n o n Ala r m # 6 6 o n o ff o n o n o n o n o ff o n

Ala r m # 2 5 o n o n o n o ff o ff o n o n o ff Ala r m # 6 7 o n o ff o n o n o n o n o ff o ff

Ala r m # 2 6 o n o n o n o ff o ff o n o ff o n Ala r m # 6 8 o n o ff o n o n o n o ff o n o n

Ala r m # 2 7 o n o n o n o ff o ff o n o ff o ff Ala r m # 6 9 o n o ff o n o n o n o ff o n o ff

Ala r m # 2 8 o n o n o n o ff o ff o ff o n o n Ala r m # 7 0 o n o ff o n o n o n o ff o ff o n

Ala r m # 2 9 o n o n o n o ff o ff o ff o n o ff Ala r m # 7 1 o n o ff o n o n o n o ff o ff o ff

Ala r m # 3 0 o n o n o n o ff o ff o ff o ff o n Ala r m # 7 2 o n o ff o n o n o ff o n o n o n

Ala r m # 3 1 o n o n o n o ff o ff o ff o ff o ff Ala r m # 7 3 o n o ff o n o n o ff o n o n o ff

Ala r m # 3 2 o n o n o ff o n o n o n o n o n Ala r m # 7 4 o n o ff o n o n o ff o n o ff o n

Ala r m # 3 3 o n o n o ff o n o n o n o n o ff Ala r m # 7 5 o n o ff o n o n o ff o n o ff o ff

Ala r m # 3 4 o n o n o ff o n o n o n o ff o n Ala r m # 7 6 o n o ff o n o n o ff o ff o n o n

Ala r m # 3 5 o n o n o ff o n o n o n o ff o ff Ala r m # 7 7 o n o ff o n o n o ff o ff o n o ff

Ala r m # 3 6 o n o n o ff o n o n o ff o n o n Ala r m # 7 8 o n o ff o n o n o ff o ff o ff o n

D ip S w itc h S e tting s (S 1 ) D ip S w itc h S e tting s (S 1 )

Dip Switch Settings (S1) (S2 on AIMS Universal) Dip Switch Settings (S1)

(18)

APPENDIX - D (continued)

S1 Dip Switch Settings

(Set Network Address)

8 7 6 5 4 3 2 1 8 7 6 5 4 3 2 1

Ala r m # 8 4 o n o ff o n o ff o n o ff o n o n Ala r m # 1 2 6 o n o ff o ff o ff o ff o ff o ff o n

Ala r m # 8 5 o n o ff o n o ff o n o ff o n o ff Ala r m # 1 2 7 o n o ff o ff o ff o ff o ff o ff o ff

Ala r m # 8 6 o n o ff o n o ff o n o ff o ff o n Ala r m # 1 2 8 o ff o n o n o n o n o n o n o n

Ala r m # 8 7 o n o ff o n o ff o n o ff o ff o ff Ala r m # 1 2 9 o ff o n o n o n o n o n o n o ff

Ala r m # 8 8 o n o ff o n o ff o ff o n o n o n Ala r m # 1 3 0 o ff o n o n o n o n o n o ff o n

Ala r m # 8 9 o n o ff o n o ff o ff o n o n o ff Ala r m # 1 3 1 o ff o n o n o n o n o n o ff o ff

Ala r m # 9 0 o n o ff o n o ff o ff o n o ff o n Ala r m # 1 3 2 o ff o n o n o n o n o ff o n o n

Ala r m # 9 1 o n o ff o n o ff o ff o n o ff o ff Ala r m # 1 3 3 o ff o n o n o n o n o ff o n o ff

Ala r m # 9 2 o n o ff o n o ff o ff o ff o n o n Ala r m # 1 3 4 o ff o n o n o n o n o ff o ff o n

Ala r m # 9 3 o n o ff o n o ff o ff o ff o n o ff Ala r m # 1 3 5 o ff o n o n o n o n o ff o ff o ff

Ala r m # 9 4 o n o ff o n o ff o ff o ff o ff o n Ala r m # 1 3 6 o ff o n o n o n o ff o n o n o n

Ala r m # 9 5 o n o ff o n o ff o ff o ff o ff o ff Ala r m # 1 3 7 o ff o n o n o n o ff o n o n o ff

Ala r m # 9 6 o n o ff o ff o n o n o n o n o n Ala r m # 1 3 8 o ff o n o n o n o ff o n o ff o n

Ala r m # 9 7 o n o ff o ff o n o n o n o n o ff Ala r m # 1 3 9 o ff o n o n o n o ff o n o ff o ff

Ala r m # 9 8 o n o ff o ff o n o n o n o ff o n Ala r m # 1 4 0 o ff o n o n o n o ff o ff o n o n

Ala r m # 9 9 o n o ff o ff o n o n o n o ff o ff Ala r m # 1 4 1 o ff o n o n o n o ff o ff o n o ff

Ala r m # 1 0 0 o n o ff o ff o n o n o ff o n o n Ala r m # 1 4 2 o ff o n o n o n o ff o ff o ff o n

Ala r m # 1 0 1 o n o ff o ff o n o n o ff o n o ff Ala r m # 1 4 3 o ff o n o n o n o ff o ff o ff o ff

Ala r m # 1 0 2 o n o ff o ff o n o n o ff o ff o n Ala r m # 1 4 4 o ff o n o n o ff o n o n o n o n

Ala r m # 1 0 3 o n o ff o ff o n o n o ff o ff o ff Ala r m # 1 4 5 o ff o n o n o ff o n o n o n o ff

Ala r m # 1 0 4 o n o ff o ff o n o ff o n o n o n Ala r m # 1 4 6 o ff o n o n o ff o n o n o ff o n

Ala r m # 1 0 5 o n o ff o ff o n o ff o n o n o ff Ala r m # 1 4 7 o ff o n o n o ff o n o n o ff o ff

Ala r m # 1 0 6 o n o ff o ff o n o ff o n o ff o n Ala r m # 1 4 8 o ff o n o n o ff o n o ff o n o n

Ala r m # 1 0 7 o n o ff o ff o n o ff o n o ff o ff Ala r m # 1 4 9 o ff o n o n o ff o n o ff o n o ff

Ala r m # 1 0 8 o n o ff o ff o n o ff o ff o n o n Ala r m # 1 5 0 o ff o n o n o ff o n o ff o ff o n

Ala r m # 1 0 9 o n o ff o ff o n o ff o ff o n o ff Ala r m # 1 5 1 o ff o n o n o ff o n o ff o ff o ff

Ala r m # 1 1 0 o n o ff o ff o n o ff o ff o ff o n Ala r m # 1 5 2 o ff o n o n o ff o ff o n o n o n

Ala r m # 1 1 1 o n o ff o ff o n o ff o ff o ff o ff Ala r m # 1 5 3 o ff o n o n o ff o ff o n o n o ff

Ala r m # 1 1 2 o n o ff o ff o ff o n o n o n o n Ala r m # 1 5 4 o ff o n o n o ff o ff o n o ff o n

Ala r m # 1 1 3 o n o ff o ff o ff o n o n o n o ff Ala r m # 1 5 5 o ff o n o n o ff o ff o n o ff o ff

Ala r m # 1 1 4 o n o ff o ff o ff o n o n o ff o n Ala r m # 1 5 6 o ff o n o n o ff o ff o ff o n o n

Ala r m # 1 1 5 o n o ff o ff o ff o n o n o ff o ff Ala r m # 1 5 7 o ff o n o n o ff o ff o ff o n o ff

Ala r m # 1 1 6 o n o ff o ff o ff o n o ff o n o n Ala r m # 1 5 8 o ff o n o n o ff o ff o ff o ff o n

Ala r m # 1 1 7 o n o ff o ff o ff o n o ff o n o ff Ala r m # 1 5 9 o ff o n o n o ff o ff o ff o ff o ff

Ala r m # 1 1 8 o n o ff o ff o ff o n o ff o ff o n Ala r m # 1 6 0 o ff o n o ff o n o n o n o n o n

Ala r m # 1 1 9 o n o ff o ff o ff o n o ff o ff o ff Ala r m # 1 6 1 o ff o n o ff o n o n o n o n o ff

Ala r m # 1 2 0 o n o ff o ff o ff o ff o n o n o n Ala r m # 1 6 2 o ff o n o ff o n o n o n o ff o n

Ala r m # 1 2 1 o n o ff o ff o ff o ff o n o n o ff Ala r m # 1 6 3 o ff o n o ff o n o n o n o ff o ff

Ala r m # 1 2 2 o n o ff o ff o ff o ff o n o ff o n Ala r m # 1 6 4 o ff o n o ff o n o n o ff o n o n

Ala r m # 1 2 3 o n o ff o ff o ff o ff o n o ff o ff Ala r m # 1 6 5 o ff o n o ff o n o n o ff o n o ff

Ala r m # 1 2 4 o n o ff o ff o ff o ff o ff o n o n Ala r m # 1 6 6 o ff o n o ff o n o n o ff o ff o n

Ala r m # 1 2 5 o n o ff o ff o ff o ff o ff o n o ff Ala r m # 1 6 7 o ff o n o ff o n o n o ff o ff o ff

D ip S w itc h S e tting s (S 1 ) D ip S w itc h S e tting s (S 1 )

Dip Switch Settings (S1) (S2 on AIMS Universal)

Dip Switch Settings (S1) (S2 on AIMS Universal)

(19)

APPENDIX - D (continued)

S1 Dip Switch Settings

(Set Network Address)

8 7 6 5 4 3 2 1 8 7 6 5 4 3 2 1

Alarm #168 o ff o n o ff o n o ff o n o n o n Alarm #209 o ff o ff o n o ff o n o n o n o ff

Alarm #169 o ff o n o ff o n o ff o n o n o ff Alarm #210 o ff o ff o n o ff o n o n o ff o n

Alarm #170 o ff o n o ff o n o ff o n o ff o n Alarm #211 o ff o ff o n o ff o n o n o ff o ff

Alarm #171 o ff o n o ff o n o ff o n o ff o ff Alarm #212 o ff o ff o n o ff o n o ff o n o n

Alarm #172 o ff o n o ff o n o ff o ff o n o n Alarm #213 o ff o ff o n o ff o n o ff o n o ff

Alarm #173 o ff o n o ff o n o ff o ff o n o ff Alarm #214 o ff o ff o n o ff o n o ff o ff o n

Alarm #174 o ff o n o ff o n o ff o ff o ff o n Alarm #215 o ff o ff o n o ff o n o ff o ff o ff

Alarm #175 o ff o n o ff o n o ff o ff o ff o ff Alarm #216 o ff o ff o n o ff o ff o n o n o n

Alarm #176 o ff o n o ff o ff o n o n o n o n Alarm #217 o ff o ff o n o ff o ff o n o n o ff

Alarm #177 o ff o n o ff o ff o n o n o n o ff Alarm #218 o ff o ff o n o ff o ff o n o ff o n

Alarm #178 o ff o n o ff o ff o n o n o ff o n Alarm #219 o ff o ff o n o ff o ff o n o ff o ff

Alarm #179 o ff o n o ff o ff o n o n o ff o ff Alarm #220 o ff o ff o n o ff o ff o ff o n o n

Alarm #180 o ff o n o ff o ff o n o ff o n o n Alarm #221 o ff o ff o n o ff o ff o ff o n o ff

Alarm #181 o ff o n o ff o ff o n o ff o n o ff Alarm #222 o ff o ff o n o ff o ff o ff o ff o n

Alarm #182 o ff o n o ff o ff o n o ff o ff o n Alarm #223 o ff o ff o n o ff o ff o ff o ff o ff

Alarm #183 o ff o n o ff o ff o n o ff o ff o ff Alarm #224 o ff o ff o ff o n o n o n o n o n

Alarm #184 o ff o n o ff o ff o ff o n o n o n Alarm #225 o ff o ff o ff o n o n o n o n o ff

Alarm #185 o ff o n o ff o ff o ff o n o n o ff Alarm #226 o ff o ff o ff o n o n o n o ff o n

Alarm #186 o ff o n o ff o ff o ff o n o ff o n Alarm #227 o ff o ff o ff o n o n o n o ff o ff

Alarm #187 o ff o n o ff o ff o ff o n o ff o ff Alarm #228 o ff o ff o ff o n o n o ff o n o n

Alarm #188 o ff o n o ff o ff o ff o ff o n o n Alarm #229 o ff o ff o ff o n o n o ff o n o ff

Alarm #189 o ff o n o ff o ff o ff o ff o n o ff Alarm #230 o ff o ff o ff o n o n o ff o ff o n

Alarm #190 o ff o n o ff o ff o ff o ff o ff o n Alarm #231 o ff o ff o ff o n o n o ff o ff o ff

Alarm #191 o ff o n o ff o ff o ff o ff o ff o ff Alarm #232 o ff o ff o ff o n o ff o n o n o n

Alarm #192 o ff o ff o n o n o n o n o n o n Alarm #233 o ff o ff o ff o n o ff o n o n o ff

Alarm #193 o ff o ff o n o n o n o n o n o ff Alarm #234 o ff o ff o ff o n o ff o n o ff o n

Alarm #194 o ff o ff o n o n o n o n o ff o n Alarm #235 o ff o ff o ff o n o ff o n o ff o ff

Alarm #195 o ff o ff o n o n o n o n o ff o ff Alarm #236 o ff o ff o ff o n o ff o ff o n o n

Alarm #196 o ff o ff o n o n o n o ff o n o n Alarm #237 o ff o ff o ff o n o ff o ff o n o ff

Alarm #197 o ff o ff o n o n o n o ff o n o ff Alarm #238 o ff o ff o ff o n o ff o ff o ff o n

Alarm #198 o ff o ff o n o n o n o ff o ff o n Alarm #239 o ff o ff o ff o n o ff o ff o ff o ff

Alarm #199 o ff o ff o n o n o n o ff o ff o ff Alarm #240 o ff o ff o ff o ff o n o n o n o n

Alarm #200 o ff o ff o n o n o ff o n o n o n Alarm #241 o ff o ff o ff o ff o n o n o n o ff

Alarm #201 o ff o ff o n o n o ff o n o n o ff Alarm #242 o ff o ff o ff o ff o n o n o ff o n

Alarm #202 o ff o ff o n o n o ff o n o ff o n Alarm #243 o ff o ff o ff o ff o n o n o ff o ff

Alarm #203 o ff o ff o n o n o ff o n o ff o ff Alarm #244 o ff o ff o ff o ff o n o ff o n o n

(20)

APPENDIX - E

PC Installation

Hardware Platform

The following is the recommended hardware for optimum program operation. Though the program will function adequately on lower performance machines, the initialization of the Alarms and the Redraw time becomes annoying to the average user.

- Pentium 500 MHZ or better - 128 Mb of RAM

- 40 Gb Hard drive or better

- 17” SVGA monitor or better (19” recommended) - Windows 98, NT, 2000 or XP

Installation of Host PC

1. Turn the power off on the computer.

2. Remove all cords from the back of the computer tower. ( Please make a note of thecables you remove, after installation of the NBS-42 card you will have to re-connect them.)

3. Remove the computer tower cover, to gain access to the motherboard.

4. Check the computer mother board for an open PCI computer slot and insert the card.

5. Once the NBS-42 card has been installed, fasten the screw on the metal side of the card. This will ground the card and keep it from moving.

(21)

6. Re-attach the computer cover and cables.

7. Attach the cable from the Polling Network into the D9-female connector located on rear of card.

“OUT+” (Red)

“OUT-” (Black)

Installation of the Device Driver

MAKE SURE YOU HAVE CORRECTLY INSTALLED YOUR NBS-42 CARD BEFORE CONTINUING.

Power on the computer and insert the A.I.M.S. software and Driver CD.

1. The computer will have two windows being displayed on your computer. The computer will detect the new hardware automatically. The second window will be the computer asking for a unique name for the NBSX-1 card, change it to NBSX-1 then click ok.

2. The computer will then ask you “would you like the wizard to search for the drivers” | click next.

3. The computer will ask you “what would you like windows to do” | click where it reads a display list of the drivers in a specific location. Then | click next.

4. Check the box for | CD and then click | Next.

5. After installing the card, the computer will request that you restart the computer.

6. Once the computer has been re-started, the CD will automatically run the software installation setup program. Simply follow the instructions on the screen.

7. Click on the Aims icon that will be placed on your desk top or locate it by |Start| then |Program|, and then |Aims|.

8. An error message may pop up click ok, and then ok again. Click on |option|, then |setup|.

(22)

The Amico Information Management System (AIMS) shall be Amico Alert-2 series.

The AIMS Universal Board shall be a microprocessor based medical device. It shall display the topology and a clone image of the device when connected to a centrally located P.C. The AIMS Universal Board works in conjuction with the Aims System.

The AIMS Universal Board will support a Vacuum Pump, providing that the system has a dry contact which is dedicated to each starter relay motor. Each motor must also have a dedicated high temperature dry contact signal. An additional Amico Vacuum sensor would need to be installed in the supply line. See chart for connection details (APPENDIX - I, page 25) and configuration setup (APPENDIX - G, page 23).

The AIMS Universal Board will support a Air Compressor, providing that the Air Compressor has a dry contact which is dedicated to each starter relay motor. Each motor must also have a dedicated high temperature dry contact signal. Dew point and CO2 high can either be a dry contact or a 4-20mA signal. An additional Amico High Medical Air sensor would need to be installed in the supply line. See chart for connection details (APPENDIX - I, page 25) and configuration setup (APPENDIX - G, page 23).

The AIMS Universal Board will support a Bulk Tank, providing the following conditions are met: Main Bulk Tank Liquid Level has a (4-20 mA) signal, Reserve Liquid Level Tank has a (4-20 mA) signal, A Dry contact signal from the Main Tank when in use, A Dry contact signal from Reserve Tank when in use and an Optional (4-20 mA) for flow from a Flow Meter (see APPENDIX - I, page 25, and APPENDIX - G, page 23).

The AIMS system consists of one Network Interface Card in each device, one computer with Windows with a Gateway module and the AIMS software. The PC shall be accessible to the internet either directly or via the hospitals LAN network. The above items are not supplied with the AIMS Universal Board.

The system shall accommodate up to 249 medical devices and connect each device in series with a two wire, twisted pair cable (up to 5,000 ft.).

APPENDIX - F

A.I.M.S. Universal Specifications

3803/,7(9

*1'

WR *DXJH7ZLVWHG 3DLU:LULQJ

&RQILJXUDWLRQ 1HWZRUN$GGUHVV

S4

S1

(23)

APPENDIX - G

Pump & Compressor / Bulk reservoir PC board settings:

S3 Dip Sw itch setting

ID# De s cription 1 2 3 4 5 6 7 8

0 Do Not Us e on on on on on on on on

1 of f of f of f on on on on off

2 Air Com p. Duple x Tank M ount of f of f of f on on on off on

3 Air Com p. Duple x Stack M ount of f of f of f on on on off off

4 Air Com p. Triple x Stack M ount of f of f of f on on off on on

5 Air Com p. Quadruple x Stack M ount of f of f of f on on off on off

6 of f of f of f on on off off on 7 of f of f of f on on off off off

8 of f of f of f on off on on on 9 of f of f of f on off on on off

10 of f of f of f on off on off on 11 of f of f of f on off on off off

12 V acuum Pum p Duple x Tank M ounte d of f of f of f on off off on on

13 V acuum Pum p Duple x Stack M ounte d of f of f of f on off off on off

14 V acuum Pum p Triple x Stack M ounte d of f of f of f on off off off on

15 V acuum Pum p Quadruple x Stack M ounte d of f of f of f on off off off off

16 of f of f of f off on on on on 17 of f of f of f off on on on off

18 of f of f of f off on on off on 19 of f of f of f off on on off off

20 Bulk Tank of f of f of f off on off on on

Air Compressor - Dew Point Monitor & CO2 Monitor

Factory Default - Disable Disable

For Dew Point Monitor: set switch #1 to off position on S1 (CONFIGURATION)

For CO2 Monitor: set switch #2 to off position on S1 (CONFIGURATION)

Enable

For Dew Point Monitor: set switch #1 to on position on S1 (CONFIGURATION)

For CO2 Monitor: set switch #2 to on position on S1 (CONFIGURATION)

Bulk Tank - Main Tank/Reserve Tank/Flow

Factory Default - Disable Disable

For Reserve Tank set switch #1 to off position on S1 (CONFIGURATION)

For Main Tank set switch #2 to off position on S1 (CONFIGURATION)

For Flow set switch #3 to off position on S1 (CONFIGURATION)

Enable 21 6 S1 Configuration Configuration S1

(24)

APPENDIX - H

Changing Pressure Units on AIMS Universal

Factory set mode:

AIR COMPRESSOR - PSI VACUUM PUMP - InHg

To change the pressure unit of measure :

1. Select “Options - Setup”.

2. From the Setup menu, click on the unit (air compressor or

vacuum pump)

3. From the “Property” menu, double-click on the desired unit of

measure (PSI, kPa, BAR, or InHg).

AIR COMPRESSOR

VACUUM PUMP

OXYGEN BULK TANK

(25)

Input # Vacuum Pump Air Compressor Bulk Tank IN1 Vacuum Pump #1 Air Compressor #1 Main Tank In Use IN2 Vacuum Pump #2 Air Compressor #2 Reserve Tank In Use IN3 Vacuum Pump #3 Air Compressor #3

IN4 Vacuum Pump #4 Air Compressor #4 IN5 Reserve In Use Reserve In Use IN6 Temp. Malfunction #1 Temp. Malfunction #1 IN7 Temp. Malfunction #2 Temp. Malfunction #2 IN8 Temp. Malfunction #3 Temp. Malfunction #3 IN9 Temp. Malfunction #4 Temp. Malfunction #4 IN10 Change Filter Change Filter

IN11 Dew Point High

IN12 CO High

CH1 Dew Point Monitor (4-20 mA) Main Tank Level (4-20 mA) CH2 CO2 Monitor (4-20 mA) Reserve Tank Level (4-20 mA)

CH3 Flow (4-20 mA)

Sensor Vacuum Sensor HP Air Sensor Not Used

INPUT # VACUUM PUMP AIR COMPRESSOR BULK TANK

Please note that IN1-IN12 are dry contact signals in and CH1-CH3 are 4-20mA signals.

Connection required from other manufacture devices to AIMS universal board, a universal board is required for each gas service monitored.

APPENDIX - I

Data / Inputs from Sources of Supply

Model Number

Description

A2P-AIMS-AREA ... AIMS Circuit Board Assembly (AREA) A2P-AIMS-MAST ... AIMS Circuit Board Assembly (MASTER) A2P-AIMS-UNIV ... AIMS Circuit Board Assembly (UNIVERSAL)

(For Air Compressors, Vacuum Pumps & Bulk Tanks) A2P-AIMS-NETW ... AIMS Board for PC, License @ Software

A2P-AIMS-SOFTW ... AIMS License and Software ONLY A2P-AIMS-PC17 ... Personal Computer for AIMS - 17” Monitor A2P-AIMS-PC19 ... Personal Computer for AIMS - 19” Monitor A2P-AIMS-REPEAT ... AIMS Repeater 485RPTR

A2X-AIMS-CAB485 ... AIMS Cable RS-485

APPENDIX - J

(26)

References

Related documents

Though sacred music was still of great importance, secular music became more prominent in the renaissance

frauds and other irregularities; selection and application of appropriate accounting policies; making judgments and estimates that are reasonable and prudent; and the

The current version of TWITTIR O ` - UD comprises 1,424 tweets, annotated at multiple levels: a prag- matic level that attempts to model irony (see Sec- tion 2.1) and a syntactic

Full year away 15% fees Home/EU students Semester or..

The Vickers hardness measurement in the TZ sample showed that the hardness was lower compared with the TN and TY samples, due to the volume expansion in the lattice of the

DESIGN OF REINFORCED CONCRETE RECTANGULAR COLUMN TABLE OF CONTENTS PAGE CONTENTS 2 A... Longitudinal bars

Whilst “new” EU countries typically have lower ratings than “old” EU countries, after controlling for financial variables, all countries are found to have significantly