• No results found

CyAn ADP with Summit Software User Guide

N/A
N/A
Protected

Academic year: 2021

Share "CyAn ADP with Summit Software User Guide"

Copied!
147
0
0

Loading.... (view fulltext now)

Full text

(1)

CyAn™ ADP with Summit™ Software

User Guide

Document Number 0000050 Revision F July 2006

(2)

Copyright © 2001 - 2006 Dako. All rights reserved.

This document may not be copied in whole or in part or reproduced in any other media without the express written permission of Dako. Please note that under copyright law, copying includes translation into another language.

(3)

The CyAn ADP Analyzer is the next step in Advanced Digital Processing (ADP) with enhanced detection for flow cytometry. The CyAn ADP High-Performance Flow Cytometer is a research tool engineered for precision analysis of cells, bacteria, and other similarly sized particles. With CyAn ADP, Dako sets a new industry standard with a combination of features never before available on a bench-top analyzer. CyAn ADP gives users three excitation lines with independent, alignment-free focusing optics, simultaneous nine color and two scatter parameters, analysis rates of 70,000 events per second, a full 9 × 9 interlaser compensation matrix, and high sensitivity. The result is stable, user-friendly, and flexible technology. The instrument is optimized for cell cycle, kinetics, fluorescent protein work, and multi-color immunophenotyping. Rare-event analysis, such as MHC Dextramer studies, and no-lyse whole blood applications are easily performed on the CyAn ADP. The instrument also provides simplified compensation before, during, and after acquisition with unequaled sensitivity in all fluorescent parameters.

(4)

Contacting Dako

Dako should be contacted immediately for assistance in the event of any instrument malfunction. For further information please contact your local Dako office.

Australia Tel. 2 9316 4633 Fax 2 9316 4773 Germany Tel. 040 69 69 470 Fax 040 69 52 741 Switzerland Tel. 041 760 11 66 Fax 041 760 11 77 Austria Tel. 0800 0800 7153 Fax 0800 0800 7154 Italy Tel. 02 58 078 1 Fax 02 58 078 292 United Kingdom Tel. (0)1 353 66 99 11 Fax (0)1 353 66 89 89 Technical Support Tel. (0) 1 353 66 99 65 Belgium Tel. 016 38 72 20 Fax 016 38 72 21 Japan Tel. 075 211 3655 Fax 075 211 1755 Canada Tel. 905 858 8510 Fax 905 858 8801 The Netherlands Tel. 020 42 11 100 Fax 020 42 11 101 Czech Republic Tel. 420 541 423 710 Fax 420 541 423 711 Norway Tel. 23 14 05 40 Fax 23 14 05 42 Poland Tel. 058-661 1879 Fax 058-661 3390

United States of America Carpinteria, California Tel. 805 566 6655 Fax 805 566 6688 Technical Support Tel. 800 424 0021 Customer Service Tel. 800 235 5763 United States of America Fort Collins, Colorado Flow Instrumentation Tel. 800 822 9902 Fax 970 226 0107 Denmark Head Office Tel. 44 85 95 00 Fax 44 85 95 95 Sales Tel. 44 85 97 56 Fax 44 85 84 29 Spain Tel. 93 499 05 06 Fax 93 499 02 08 France Tel. 1 30 50 00 50 Fax 1 30 50 00 11 Sweden Tel. 08 556 20 600 Fax 08 556 20 619

(5)

User Resources

For the latest information on Dako products and services, please visit the Dako Web site at

http://www.dako.com, and the Dako Flow Cytometry Discussion Group at

http://www.dakousa.com/index/support/discussion_groups.htm.

Scope

This guide provides a detailed discussion of the architecture and operating procedures for the CyAn™ ADP High-Performance Flow Cytometer. Detailed operating instructions for Summit® software can be found in the Summit software online help system. The information contained in this document can be applied to all CyAn ADP products.

Disclaimers

This document is not a substitute for the detailed operator training provided by Dako or for other advanced instruction in general cytometric techniques. It is essential that the operator have a working knowledge of Microsoft® Windows® XP or current Operating System prior to using this guide. Dako also recommends the use of all of the networking services provided in the individual operator’s laboratory. Although daily maintenance and routine instrument adjustments are discussed here, your local Dako Technical Service Group should be contacted immediately for assistance in the event of any instrument malfunction.

Trademarks

Microsoft® and Windows® XP are registered trademarks of Microsoft Corporation. Macintosh™ is a registered trademark of Apple, Inc. MoFlo® is a registered trademark of Dako. CyAn™,

Summit™, SpectrAlign™, and SpectraComp™ are trademarks of Dako. All other trade names and trademarks are the property of their respective holders.

(6)

Table of Contents

Section 1 Safety... 9

General Hazard Avoidance ... 10

Biohazard Avoidance ... 11

Laser Safety ... 12

Section 2 Installation... 15

CyAn ADP Installation Requirements ... 15

General Laboratory Information... 15

CyAn ADP Power Configuration and Consumption ... 17

CyAn ADP Instrument Configurations... 17

Section 3 System Overview... 19

CyAn ADP Features ... 19

CyAn ADP Subsystems ... 20

Fluidics ... 21

Optics ... 22

Electronics ... 25

Peripheral Devices... 26

Software... 26

Summit Software Features ... 27

Section 4 Startup and Shutdown Procedures ... 29

Required Reagents for Startup and Shutdown ... 29

CyAn ADP Startup... 29

Daily Decontamination Procedure... 32

Shutdown Procedure... 33

Section 5 Sheath Management System... 35

Sheath Management System Indicators ... 35

Changing Out Sheath and Waste Containers... 41

Replacing Cleaner Fluid... 42

Section 6 Summit Software Basics ... 43

Acquire and Analyze up to 10 Million Events... 43

Create a New Database... 44

Open an Existing Database ... 46

Create a Histogram ... 47

Create a Dot Plot... 48

Open FCS (Listmode) Files... 49

Set a Gate from a Single Region ... 51

Set a Serial Gate ... 52

Define Gate Logic... 55

Shortcut Keys ... 57

Section 7 CyAn ADP Maintenance... 59

Daily Preventive Maintenance... 59

Weekly Preventive Maintenance... 59

Reusable Sheath Containers... 59

Head Unit and SMS Cart ... 60

Fluidic Lines ... 60

Monthly Preventive Maintenance ... 60

System Decontamination... 60

Laser Interlock Maintenance Procedure ... 61

Biannual Maintenance Procedure ... 61

Section 8 Troubleshooting ... 63

Section 9 Six-Color Compensation Experiment... 65

Background ... 65

Setting up for sample acquisition ... 66

(7)

Manual Compensation ... 71

Setting Spillover Percentage for Compensation ... 72

Running the Second Sample ... 73

Running the Remaining Samples... 75

Running the “All Stains” Sample ... 83

Putting Compensation into Practice ... 84

CD3+ T-Lymphocytes ... 84

CD 19+ B-Lymphocytes ... 84

CD4+/CD3+ T-Lymphocytes (T-Helper cells) ... 85

CD8+/CD3+ T-Lymphocytes (T-Cytotoxic/Suppressor cells)... 85

Natural Killer Lymphocytes ... 86

Section 10 Auto Compensation ... 87

Auto Compensation for Single Positive Controls ... 87

Section 11 VisiComp ... 91

Using the VisiComp Feature after Compensation... 91

Section 12 WorkList Builder... 93

Overview ... 93

Accessing the WorkList Builder... 94

Creating a New Panel ... 95

Setting up Tests and Defining Reagents for a Panel ... 97

Creating Reagents by Defining Antibody/Fluorochrome Combinations... 98

Deleting Reagents from the WorkList Builder Database... 100

Setting up Tests (Tubes or Samples) within a Panel ... 101

Setting up a Panel ... 105

Creating Worklists from Defined Panels and Tubes ... 107

Setting up Worklists for Multiple Experiments or Specimens... 108

Customizing the WorkList Builder Viewing Area... 110

Edit Columns... 110

Importing a Specimen List from a Text File ... 111

Saving and Reusing Worklists ... 112

Associating Protocols, Gate/Event Limits, or Other Information to Worklist Tests ... 113

Appendix A Approved Cleaners and Disinfectants ... 115

Cleaners ... 115

Disinfectants for Use in Sample Line*... 115

Disinfectants for Use in Sheath Line** ... 115

Disinfectants for Use in the Waste Tank ... 115

Appendix B CyAn ADP Consumables and Parts ... 117

Appendix C Technical & Instrument Specifications ... 119

Appendix D Symbol Definitions ... 123

(8)
(9)

Section 1

Safety

Electrical Safety

The Dako instrument product line conforms to international regulations

encompassing the accessibility of high voltages by the user. In the United States, each flow cytometer is manufactured to comply with applicable regulations from the American National Standards Institute (ANSI), the Occupational Safety and Health Administration (OSHA), and various state regulatory organizations. Internationally, each CyAn™ ADP High-Performance Flow Cytometer is manufactured to comply with the International Electrotechnical Commission’s Standard for Safety

Requirements for Electrical Equipment for Measurement, Control and Laboratory Use (IEC 61010-1) and subsequent amendments.

Biohazard Safety

The CyAn ADP instrument may be used for the analysis of pathogenic or other harmful agents. These operations can constitute a biohazard for the operator as well as bystanders in the surrounding laboratory space. Dako does not certify any instrumentation for use with any hazardous organism or agent. Dako strongly recommends explicit guidance be obtained from the institutional or company biosafety officer prior to operating this instrument with any potentially hazardous organism or agent. See the Biohazard Avoidance section for more information.

WARNING

If any biohazardous organism or agent is used in this instrument, the user must inform Dako in writing prior to any field service visit or the return of any part to Dako, or its vendors, for service. Safety of the CyAn ADP user and of all Dako employees is of primary concern. Proper decontamination procedures must be followed for returned parts.

WARNING

Wear Personal Protective Equipment (PPE) in accordance with your laboratory safety procedures when operating or maintaining this instrument. It is recommended that gloves, a laboratory coat, and safety glasses are worn at all times when using the CyAn ADP. Use of this instrument in a manner other than that specified in this manual may cause impairment of equipment, compromise safety, and/or result in injury.

(10)

General Hazard Avoidance

Safety of the CyAn ADP operator is of primary concern. It is important that all operators familiarize themselves with this safety section to avoid all potential hazards associated with the use of the CyAn ADP. Any questions or concerns regarding safety of the instrument should be directed to Dako Colorado immediately.

1. PINCHING: The lid of the CyAn ADP is on hinges that should keep the lid in the UP position when open. It is recommended by Dako that the lid NOT be left open for extended periods of time OR be opened and closed repeatedly to avoid excessive wear on the hinges which may, over time, cause the lid to fall unexpectedly, potentially falling on an operator’s hands or fingers, causing injury. Please use caution when the lid is open and do not leave unattended. If the lid of the CyAn ADP becomes loose and will not stay up on its own, please contact Customer Care so it can be repaired.

2. TRIPPING: The Sheath Management System (SMS) is attached to the CyAn ADP head unit by an 8 foot (2.4 meters)long umbilical cord. It is suggested that the SMS cart be placed underneath a laboratory bench to keep it out of the way of foot traffic through the laboratory. If this is not possible, it is recommended that the SMS be placed where it will minimize the likelihood of a tripping hazard. It is suggested that the umbilical cord be run behind the laboratory bench, if possible, to keep the cord from dangling and potentially creating a tripping hazard as well. If the cord must be run down in front of the laboratory bench to the SMS, please use caution when operating or walking past the CyAn ADP to avoid tripping on the umbilical cord. It is suggested that hazard tape be placed on these items to make them highly visible to minimize the possible tripping hazard in cases where laboratory set up is not ideal. If a person trips on either the umbilical cord or SMS cart of the CyAn ADP, please contact Customer Care.

3. CHEMICAL: PPE should be worn at all times when using any part of the CyAn ADP to ensure protection from all potential hazards. Use caution when handling disinfectant products used in the waste tank. It is important that the appropriate type and amount of disinfectant is used in the waste tank to ensure effective inactivation of biologics. Please see Appendix A for a list of approved cleaners and disinfectants. DO NOT use any cleaner or disinfectant that is NOT on this list, as this may cause damage to the system and void your warranty. Please contact Customer Care with any questions regarding chemical usage on your CyAn ADP. 4. ERGONOMICS – Computer Use: It is suggested by Dako that the safety officer or

designated individual evaluate the operator(s) workstation soon after the installation of the CyAn ADP to ensure proper ergonomic guidelines are followed to minimize any discomfort that may arise from the use of CyAn ADP peripheral equipment. If any operator experiences any physical discomfort when using the CyAn ADP, please contact Dako Customer Care. A regular keyboard and mouse are included with the computer. If desired, an ergonomic keyboard or trackball mouse may be changed out (at the institution’s expense) on the computer supplied with the CyAn. If problems arise after installation of alternative peripheral equipment, please contact Customer Care immediately.

5. ERGONOMICS - Lifting: The sheath and waste tank containers used on the SMS cart have a maximum capacity of 20L. When full, either container will weigh approximately 19kg (42 pounds). Use caution when lifting full containers, whether it be removing or replacing it on the cart, to avoid back strain or injury. Proper lifting techniques must always be used when handling heavy items to avoid back injury. If an operator is unable to comfortably lift a container, it is recommended to seek assistance with moving the container to the desired position. It is also suggested that the waste tank be emptied when half full so that it will be

(11)

easier to handle. If any operator experiences any physical discomfort when handling the sheath or waste containers on the SMS cart, please contact Dako Customer Care. 6. SKIN ABRASION: The sample probe is constructed of stainless steel and the end of the

probe is slightly squared off. Accidental direct contact with the end of the probe and the skin could cause an abrasion or cut. Use caution when loading and unloading samples on the probe. PPE should be worn at all times when using any part of the CyAn ADP. Samples of any kind should never be used on the CyAn without gloves being worn. If any injury related to the sample probe occurs, contact Dako Customer Care immediately.

Biohazard Avoidance

Safety of the CyAn ADP operator is of primary concern. It is important that all operators familiarize themselves with this safety section to avoid all potential biohazards associated with the use of the CyAn ADP. Any questions or concerns regarding biohazard safety and the CyAn ADP should be directed to Dako Customer Care immediately.

1. WASTE TANK: The reusable waste tank is located on the right side of the SMS cart and is labeled with a large black and orange biohazard symbol on the front. Emptying waste fluid generated from the CyAn ADP could be a daily occurrence for some operators. It is expected that all operators have undergone proper blood borne pathogens training provided by the institution of use, or similar authority, prior to using the CyAn. It is pertinent that the appropriate type and quantity of disinfectant is placed in the waste tank to ensure effective inactivation of the biologics in use when the tank is full. Please see Appendix A for a list of approved cleaners and disinfectants for use with the CyAn ADP. Please use caution when emptying the waste tank to avoid splashing of the waste fluid. PPE should be worn at all times when using any part of the CyAn ADP, including emptying the waste tank. Please contact Dako Customer Care if there are any questions or concerns about handling waste fluid generated by the CyAn ADP.

2. SAMPLE WASTE: Exposure to sample waste fluid from clogs in the waste line may occur; however, this is rare. If a clog occurs, exposure to biohazards is likely. PPE should be worn at all times when using the CyAn ADP to protect the operator from unexpected hazards. If exposure to biohazardous material occurs due to a fault of the CyAn ADP, Dako Customer Care should be contacted immediately.

3. SAMPLES: Sample handling is a normal part of the use of the CyAn ADP. It is expected that all operators have undergone proper bloodborne pathogens training provided by the

institution of use, or similar authority, prior to using the CyAn ADP. PPE should be worn at all times when using any part of the CyAn ADP. Samples of any kind should not be handled without wearing gloves. Caution should be used when handling any biohazardous samples. Any reagents used on the instrument, or in the preparation of samples for use on the CyAn ADP, should be handled by professional users only.

4. SAMPLE – O-Ring: There is an o-ring located in the top interior of the sample injection area. The purpose of this o-ring is to create a seal when a test tube is placed on the sample probe, which results in pressurization of the tube. Over time this o-ring may lose its elasticity or become damaged from repeated contact with polypropylene test tubes. Deterioration of this o-ring may cause test tubes to be loose when placed on the sample probe, leading to the potential for test tubes filled with biohazardous samples falling off, causing exposure to the operator or other laboratory personnel. This occurrence is not likely, but if an operator experiences this condition, Dako Customer Care should be contacted immediately so the o-ring can be replaced.

(12)

5. SAMPLE – Waste Manifold: Waste fluid is continuously generated when running the CyAn ADP and is a normal function of this instrument. This waste fluid is routed through a manifold located within the SMS cart. If this part malfunctions, a backup of fluid may occur, resulting in waste fluid dripping through a small hole in the top corner of the sample injection area. Because this waste fluid may contain biohazardous material, caution should be used if this fault is identified; however, this occurrence is highly unlikely. PPE should be worn at all times when using any part of the CyAn ADP to ensure protection of the operator from all potential hazards. If an operator identifies such a leak, Customer Care should be contacted

immediately.

6. AEROSOLS: The CyAn ADP is a flow cytometry analyzer and has no sorting capabilities. No aerosols are generated under normal operating conditions. In the event of extreme over pressurization of the system, there is a remote chance that sample and/or waste fluid could exit orifices at a rate which could potentially create aerosols; however, this has never been experienced. If at any time an operator identifies a potential overpressurization problem, Customer Care should be contacted immediately.

7. SAMPLE PROBE: The sample probe is constructed of stainless steel and the end of the probe is slightly squared off. Accidental direct contact with the end of the probe and the skin could cause an abrasion or cut. Because biohazardous samples are run through the sample probe, there is a chance that if the skin is broken due to direct contact with the end of the sample probe, an operator could be directly exposed to biohazardous material. Use caution when loading and unloading samples on the probe. PPE should be worn at all times when using any part of the CyAn ADP. Samples of any kind should never be used on the CyAn ADP without gloves being worn. If any injury related to the sample probe occurs, Dako Customer Care should be contacted immediately.

Laser Safety

WARNING

Wear Personal Protective Equipment in accordance with your laboratory safety procedures when operating or maintaining this instrument. Use of this instrument in a manner other than that specified in this manual may cause impairment of equipment, compromise safety, and/or result in injury.

CAUTION

Use of controls or adjustments or performance of procedures other than those specified in this manual may result in hazardous radiation exposure. This radiation will be in the visible regions of the electromagnetic spectrum. Do not attempt to defeat the interlocks or open covers or panels retained with screws.

CLASS 1 LASER PRODUCT

IEC/EN 60825 -1/A2:2001

The CyAn ADP instrument conforms to

international regulations encompassing laser safety. The CyAn ADP is a Class 1 laser device.

(13)

This designation indicates no hazardous laser energy is accessible to the user during normal operation or during a failure mode.

Note Because the CyAn ADP has been designed to operate as a Class 1 Laser Device, the instrument must be operated with all light containment tubes in place and all protective light seals intact. Most laser components of the CyAn ADP flow cytometer are Class IIIB to Class IV. As such, these lasers have the potential to cause injury. There is one small viewing window inside the CyAn ADP, but its purpose is to be used only by trained Dako personnel during service or repair. A safety interlock prevents operators from viewing laser light through this viewing window under normal operating conditions.

Certification of regulatory compliance of these lasers often requires significant involvement from the institutional or company laser-safety officer. Dako does not provide regulatory certification for the individual lab, but can offer assistance in laser safety matters.

All access plates and other user accessible points of potential laser exposure are clearly designated on the CyAn ADP instrument with the labels illustrated in figures 1.1, 1.2, and 1.3 below. Interlocks are designed to prevent accidental irradiation of the operator. The user should not defeat these interlocks.

Any questions or concerns regarding laser safety of the CyAn ADP should be directed to Dako Customer Care.

(14)
(15)

Section 2

Installation

CyAn ADP Installation Requirements

IMPORTANT

Your Dako representative is responsible for uncrating, installing, and initial set up of the CyAn ADP.

General Laboratory Information

Avoid placing the CyAn ADP directly below heating and air conditioning vents or fans in order to prevent the possible adverse effects of temperature fluctuation, vibration, and dust.

Table 2.1: CyAn ADP Installation Requirements General Requirements

Power Requirements 100V – 230V, 600VA(W), 50/60 Hz

Lab Bench Stable bench/table top to support the CyAn

ADP, one monitor, keyboard, and a mouse pad

Service Access 45.7 cm (18 in) minimum around instrument

components

Phone Location near CyAn ADP for contacting

technical support

Installation Category II

Pollution Degree 2

Internet Access Internet Service Provider or LAN connection

for downloading software updates Dimensions (not including Auxiliary

Components)

Height

front cover closed - 39.1 cm (15.4 in) front cover open – 72.1 cm (28.4 in) Width - 33.3 cm (13.1 in)

Depth - 49.8 cm (19.6 in)

with clearance for cables – 62.5 cm (24.6 in) Weight - 36.3 kg (80 lbs)

(16)

Auxiliary Components

Sheath Management System (with casters): Houses sheath container, waste container, cleaner fluid container, air compressor, and vacuum

Height – 61.3 cm (24.2 in) Width - 73.3 cm (28.9 in) Depth – 61.9 cm (24.4 in) Weight – 22.6 kg (50 lbs)

Summit Workstation Height – 42.9 cm (16.9 in)

Width – 19.1 cm (7.5 in) Depth – 45.7 cm (18.0 in) Weight – 10.5 kg (23 lbs) Optional Uninterruptible Power Supply Height – 20.3 cm (7.9 in)

Width – 14.7 cm (5.7 in) Depth – 44.5 cm (17.5 in) Weight – 20 kg (44 lbs)

Operating Environment

Ambient Temperature 15 to 30°C (59 to 86°F)

For optimum performance maintain at +/- 2°C

(17)

CyAn ADP Power Configuration and Consumption

Avoid placing the CyAn ADP directly below heating and air conditioning vents or fans in order to prevent the possible adverse effects of temperature fluctuation, vibration, and dust.

CyAn ADP Instrument Configurations

Instrument Options Laser Configuration Options

Code No. Lasers Parameters Colors 488 nm 635 nm 405 nm

CY 200* 2 9 7 5 2 -

CY 201* 3 11 9 5 2 2

(18)
(19)

Section 3

System Overview

The CyAn ADP is a state-of-the-art flow cytometer that utilizes one or more laser excitation sources to analyze biological cells, beads, or other microscopic particles as they are transported through an interrogation point in single file. Information can be gathered from large numbers of particles in a relatively short time so that populations can be studied or differentiated from other populations using simple to complex statistical methods. Physical and biochemical characteristics that can be measured using this instrument include but are not limited to forward scatter (size-related), side scatter (morphology-(size-related), fluorescence from tagged (stained) cells/particles, and auto fluorescence from non-stained cells/particles.

CyAn ADP Features

The CyAn ADP High-Performance Flow Cytometer is a research tool engineered for precision analysis of cells, bacteria, and other similarly sized particles. With CyAn ADP, Dako sets a new industry standard with a combination of features never before available on a bench-top analyzer. CyAn ADP gives users three excitation lines with independent, alignment-free focusing optics, simultaneous nine color and two scatter parameters, analysis rates of 70,000 events per second, a full 9 × 9 interlaser compensation matrix, and high sensitivity. The result is stable, user-friendly, and flexible technology. The instrument is optimized for cell cycle, kinetics, fluorescent protein work, and multi-color immunophenotyping. Rare-event analysis, such as MHC Dextramer studies, and no-lyse whole blood applications are easily performed on the CyAn ADP. The instrument also provides simplified compensation before, during, and after acquisition with unequaled sensitivity in all fluorescent parameters.

The main features of CyAn ADP include:

Walk-Up Operation

CyAn ADP offers walk-up operation that puts even a novice user at ease. With no optical alignment necessary, the user simply places a 5 mL sample tube in the tube holder and chooses the desired protocol in Summit software. The software is used to control instrument settings.

Bench-Top Configuration

The CyAn ADP was designed with today’s laboratories in mind. The CyAn ADP measures 30 cm wide x 45 cm deep x 40 cm high.

Summit Software

Summit software, Dako’s premier cytometry software product, offers full control of all CyAn ADP functions, coupled with robust user- and application-specific data acquisition, analysis, storage, reduction, and retrieval. Summit software controls all instrument parameters in an intuitive MS Windows®-based drag-and-drop environment using simple protocol-based menus. Summit

(20)

software also provides high-content data acquisition, database capabilities, and open architecture for data exchange. Additional features include off-line analysis and easy page layout.

Safety

The CyAn ADP High-Performance Cell/Bead Analyzer has been designed to incorporate the highest level of laboratory and operator safety.

CyAn ADP Subsystems

CyAn ADP has four key subsystems that form a powerful research tool. The four subsystems are:

ƒ Fluidics

ƒ Optics

ƒ Electronics

ƒ Software

(21)

A functional block diagram for the CyAn ADP High-Performance Flow Cytometer is shown in figure 3.1. The fluidics system pressurizes the system and transports particles to the interrogation point. Lasers are used as excitation sources and their beams, along with the ensuing scatter and fluorescence, are directed within the optical system. Electronics are used to power and control the instrument functions. The software provides the user interface for the above fluidics, optical, and electronic hardware, and the functions to acquire, analyze, and store data associated with the particles.

Fluidics

The Sheath Management System controls the transfer of sheath fluid, waste, and cleaner fluid throughout the CyAn ADP system. Sheath fluid is housed in either a replaceable 20 liter cubitainer or 20 liter plastic carboy. Waste is contained in a 20 liter plastic carboy. Cleaner fluid is contained in a 5 liter cubitainer. The Sheath Management System provides improved sheath pressure stability and prevents bubbles from entering the system when the sheath container is changed. The built-in cleaner fluid cubitainer allows the user to easily clean and rinse the head-unit sheath path using Dako Clean and Rinse solution and sheath or DI water as rinsing fluid. The 20 liter sheath and waste containers provide approximately 24 hours of system run time.

Figure 3.2 outlines the key fluidics components of CyAn ADP. In general, flow of air and fluids follows a path left to right in the diagram. An air compressor provides pressure for propulsion of the sample, cleaner, and sheath fluids. Air regulators condition and stabilize the pressure source prior to sheath and cleaner reservoirs and sample vessel. A number of electrically controlled valves control the flow state of the system and provide a means for cleaning and debubbling, in addition to routine flow conditions for sample analysis. Sample is forced through tubing, is introduced to the flow cell, and is fluidically focused by sheath fluid. This hydrodynamic focusing effect causes individual particles to be introduced in single file to each of the sequential laser beams. While a pinch valve is used to allow the flow of sample, the rate of sample flow is controlled by adjusting the over-pressure (differential pressure) of an electrically controlled air regulator relative to sheath pressure. Waste is drawn by a vacuum pump to a holding container for disposal.

(22)

Figure 3.2: CyAn ADP Fluidics Block Diagram

Optics

Three excitation sources are included on the CyAn ADP optical bench. Each path has its own independent, unique steering and focusing elements to provide optimal excitation of particles at the interrogation point. Each laser beam is focused to the quartz flow cell, where particles are transported past the laser beams.

(23)

The focused beams are separated vertically to ensure minimal optical crosstalk between detection paths. Further, the signals that are detected as particles that traverse the beams are electronically gated to reduce unwanted signal interference between parameters. A high numerical aperture custom achromatic objective is used to collect scatter and fluorescence and to generate an image of the interrogation point for each of three apertures (spatial filters). The apertures improve signal to noise by preventing unwanted scatter around the excitation region from entering the detector block. Light that transmits through the spatial filters is collimated and spectrally filtered on its way to the appropriate photomultiplier tubes (PMT) for detection.

(24)

Figure 3.4 illustrates the location, position, and orientation of the optical filters and PMTs within the detection block assembly of the CyAn ADP. Dichroic filters are located at 45 degrees to the direction of light propagation, while emission filters are located at 90 degrees to the light path. Filter sticks are interchangeable, thus allowing custom configurations to be implemented. Please contact Dako to order dichroic filter sets.

Laser 1

Figure 3.4: CyAn ADP Detector Block and Optical Filter Layout

Emission Filters APC APC-Cy7 CAS B CFP DAPI Pacific Blue CAS Y PE-Cy7 FL4 PE-Cy5 PerCP 7AAD PE-TxRed PI PE FITC GFP 730DLP 640DLP 595DLP 545DLP R95% 730DLP mirror 485DLP mirror mirror Emission Filters Dichroic Filters

Fluorescence & Scatter from Interrogation Point

750LP 680/30 613/20 575/25 530/40 488/10 450/50 750LP 665/20 530/40 488 nm Laser SSC FL1 FL2 FL3 FL5 Photomultiplier Tubes SSC 650DLP

* Filter configurations are easily modified to accommodate different applications. FL9 FL8 FL6 FL7 635 nm Laser 405 nm Laser

(25)

Electronics

CyAn ADP has an array of electronic components. The PC/workstation is used for instrument control, status, and data acquisition functions and communicates with the CyAn ADP instrument through a high-speed serial link. Housed within the instrument is a state-of-the-art electronic chassis that provides a communication backplane/bus with a number of available card slots to allow modular connection of key electronic control and sensing components. These components include trigger and signal processing, and multi-function input/output. Peripheral to the electronic chassis, but within the instrument are a number of devices that can be grouped as fluidics control and sense (pumps, regulators, valves, and sensors), laser/shutter control and sense, PMT voltage control, and PMT signal. The multifunction I/O card is used for controlling these peripheral devices and sensing instrument status. The PMT and photodiode detectors convert light emitted from particles excited at the interrogation point, into electrical signals. These signals are input to the trigger and signal processing cards. Amplification, analog to digital conversion, and sampling techniques provide quantitative measurements including peak, area/integral, log, and pulse width for a given triggered particle event.

(26)

Peripheral Devices

The CyAn ADP instrument includes the following peripheral devices: Summit Workstation with a printer, and the Sheath Management System (SMS). The SMS houses a sheath container, waste container, cleaner cubitainer, reusable plastic carboy, level indicators, in-line sheath filter, and compressor/vacuum pump.

Software

CyAn ADP uses Summit software for instrument control, data acquisition, and subsequent data analysis. Summit software is the Windows® user interface for the entire Dako flow cytometry product line.

Summit software offers users complete control of the CyAn ADP instrument at varying levels of complexity, depending on their needs. Summit software, on CyAn ADP with all parameters enabled, has the capability to acquire, save, and analyze up to 10 million events (see page 43). The instrument control panel (figure 3.5) provides access to laser control, event rate settings, system maintenance functions such as clean and rinse, and Sheath Management System functions. The sample parameters panel (figure 3.6) has software controls for adjusting parameter settings such as threshold, PMT voltage, and gain.

User documentation for Summit software is available from the Help menu.

Figure 3.5: Instrument Control Panel

(27)

Summit Software Features

Summit software offers compensation, flexibility, intelligence, organization, and intuitiveness.

Figure 3.7: Summit Software Desktop Display

Compensation

Summit software performs full inter-laser compensation of fluorescence parameters. Data saved as a sample file in Summit software will retain all compensation values as well as the uncompensated data for each parameter.

Flexibility

Summit software offers flexibility throughout the program to allow the data to be gathered and manipulated in ways that best suit the application and user. For example, virtually any logical expression can be defined using any number of regions in “AND” and “NOT” combinations for generating statistics, including non-rectangular regions with an almost limitless number of vertices. In addition, any and all parameters available with the CyAn ADP High-Performance Flow Cytometer can be selectively stored, including the ability to disable non-contiguous groups of parameters. Eliminating unnecessary parameters increases the display performance during acquisition and reduces data storage needs.

(28)

Intelligence

While performing batch analysis, Summit software's intelligent parameter matching feature dynamically determines the correct parameter to display for a sample based on antibody, stain, name, parameter type, channel, or a combination of these.

Organization

Summit software’s sample viewer tracks samples in user-definable folders. Tests and controls with different parameters can be grouped, and samples can be dragged and dropped between folders to help organize information. Statistics for each gate or region are automatically available under each histogram.

(29)

Section 4

Startup and Shutdown Procedures

Required Reagents for Startup and Shutdown

Dako Decontamination Solution, catalog # S2324 Dako Clean and Rinse Solution, catalog # S2323

CyAn ADP Startup

Note The Quick Start Guide for the CyAn ADP (Document Number. 0000023) is available for a quick reference for Startup and Shutdown procedures.

All CyAn ADP users:

1. Log on to the Summit Workstation.

2. Open Summit software by double-clicking the Summit icon on the Windows® desktop. 3. Select an existing database or create a new one, and then click OK.

(30)

4. Click the Instrument tab to view the CyAn Control Panel.

5. Check the status of the sheath container and the waste container by viewing the Sheath and

Waste levels in the CyAn Control Panel. Make sure the sheath container is at least half full and the waste container is at least half empty. If necessary, fill the sheath container and empty the waste container as described in Changing Out Sheath and Waste Containers on page 41.

WARNING

Remove the waste container and the sheath container from the cart before emptying or filling. Dispose of the contents of the waste container in accordance with local, state, and federal regulations.

WARNING

Do not drop the waste or sheath container on the Sheath Management System. Doing so may result in improper calibration of the load cells.

6. Click Startup on the Cyan Control Panel.

(31)

8. Open the required laser shutters by clicking the Closed buttons. These buttons should now read Opened.

9. Allow a minimum of 30 minutes for the lasers to stabilize.

10. It is recommended to perform a Clean and Rinse cycle while the instrument is warming up. Before running the Clean and Rinse cycle, make sure you have enough sheath fluid and cleaner fluid to complete the cycle.

ƒ On the CyAn Control Panel, click the Sys Clean button

ƒ A message box will warn you that the Clean and Rinse process will take 10-15 minutes. Click OK to continue. The clean cycle takes seven minutes to complete and the rinse cycle takes seven minutes to complete.

11. Repeat the clean cycle using a tube of DI water.

12. Open your QC file protocol generated by your laboratory, place a tube with SpectrAlign beads (106 concentration) on the sample probe, and acquire (F2) at a low event rate (~100 eps). Verify that the data is within your daily QC specification. Dako recommends the following CV specifications for instrument calibration.

Parameter Target CV FITC < 3.0 PE < 3.0 PE-Texas Red < 3.5 PE-Cy5 < 4.5 PE-Cy7 < 6.5 Pacific Blue < 6.0 Cyan Yellow < 6.0 APC < 6.0 APC-Cy7 < 6.0

13. If QC data is within specification, you are ready to run samples. If not, click Debubble on the CyAn Control Panel. Rerun the SpectrAlign beads. If necessary, repeat Debubble and then

(32)

IMPORTANT

ƒ After the lasers have stabilized, do not power them down; leave them powered for the duration of CyAn ADP operations for the day. If the lasers are powered down, you must allow another 30 minute stabilization period before running samples again. To conserve sheath fluid, you can shut down the SMS by clicking the Fluidics off button in the CyAn Control Panel.

ƒ If you have problems with the CyAn ADP or maintenance questions, please contact your local Dako Technical Service Group.

Daily Decontamination Procedure

It is recommended that after running samples, a decontamination of the sample line be performed on a daily basis.

1. Install a tube containing 3 mL of a 1:5 dilution of Dako Decontamination Solution (S2324) and run for 2 minutes.

2. Remove the tube of Decontamination Solution, replace with a tube of deionized water, and rinse for 2 minutes.

3. Click the Shutdown button on the CyAn Control Panel and perform the steps as indicated in the software.

Note: Dako Decontamination Solution contains sodium hypochlorite, which can stain clothing as well as affect the finish of the sample lever arm. Be careful when handling this fluid in order to avoid damage to these items.

(33)

Shutdown Procedure

Dako recommends that Sample InjectionClean and DI Water Clean are enabled for shutdown. Below are the instructions for how to set these criteria.

1. On the Edit menu, click Preferences.

2. Expand the Instrument list item by clicking the + sign. 3. Click CyAn.

4. In the Command Options list, select Shutdown, check the Sample InjectionClean and DI Water Clean check boxes, and then click Save and Close.

5. Click Shutdown on the CyAn Control Panel. This will automatically shut down the lasers and close the laser shutters.When prompted, place a tube full of Clean and Rinse solution on the sample probe and close the lever.

Note: If shutting down the instrument for the day, follow the instructions above, but use Decontamination solution instead of Clean and Rinse solution.

6. When prompted, remove the tube, replace with a tube of DI water, and move the sample lever in.

(34)

7. Check the status of the sheath container and the waste container by viewing the Sheath and

Waste levels in the CyAn Control Panel. Make sure the sheath container is at least half full and the waste container is at least half empty. If necessary, fill the sheath container and empty the waste container as described in Changing Out Sheath and Waste Containers on page 41.

8. Fill the test tube with DI water, place it on the sample probe and leave the lever out. 9. Close Summit software.

Note: Dako recommends waiting 30 seconds after closing Summit software before logging off ofthe Summit Workstation in order to allow all data sources to close.

10. Log off of the Summit Workstationand turn off the computer. 11. Turn off the computer monitor.

(35)

Section 5

Sheath Management System

Sheath Management System Indicators

The Sheath Management System provides 20 - 24 hours of sample run time. As the level of sheath fluid decreases and the level of waste fluid increases, both Summit software and Light Emitting Diodes (LED) indicators on the Sheath Management System front panel will provide fluid level status information and alerts to the user.

SMS indicators can be viewed in the Sheath Management System section of the Instrument

tab in the Control Panel. There are indicator lights in the SMS area. When all of the indicator lights are green, all components of the SMS are functioning properly.

Figure 5.1 – SMS section of Control Panel

If an error condition occurs or status changes for a subsystem, an indicator light will change from green to amber or red. When you place your cursor over the indicator light, a message appears that provides a description of the warning and instructions for how to fix the problem.

(36)

The following figures illustrate some of the possible error conditions in the SMS.

Figure 5.2 – Low sheath fluid warning

Figure 5.3 – Empty sheath fluid message

(37)

The LED indicators on the front panel of the Sheath Management System also change when fluid levels change in the SMS. Three fluid levels are monitored on the SMS panel: SHEATH LEVEL,

CLEANER LEVEL, and WASTE LEVEL.

When all fluids are at suitable levels for proper operation of the CyAn ADP, the SHEATH LEVEL,

CLEANER LEVEL, and WASTE LEVEL LEDs appear green beside the OK label:

When the sheath level or cleaner fluid level is low, thegreen OK LED changes to a flashing amber or a solid amber LED beside the LOW label:

When the waste fluid level is high, the green OK LED changes to a flashing amber LED beside the HIGH label:

(38)

When the sheath or cleaner level changes from low to empty, the flashing amber LED changes to a red LED beside the EMPTY label:

When the waste fluid level changes from high to full, the flashing amber LED changes to a red LED beside the FULL label:

(39)

Table 5.1 lists the indicators in the CyAn Control Panel user interface and the message text for each error or status condition.

Table 5.1: CyAn ADP Control Panel Indicators CyAn Control

Panel Message Text

Message 102: Your CyAn cover is open. The laser shutters have been closed and fluidic system is shutdown. Please close your cover and restart.

Message 301: Cleaner fluid is low. Please replace the cleaner solution and press the Startup button on the CyAn Control Panel.

Message 302: Cleaner fluid is empty. Please replace the cleaner solution and press the Startup button on the CyAn Control Panel.

Message 304: Cleaner quick connect is not completely engaged. Please check your connection.

Message 305: Internal reservoir overfilled. This will not prevent operation of the instrument but may require future service.

Message 306: Clean subsystem is halted. Please check that you have sufficient cleaner fluid.

Message 307: Cleaner subsystem switch error. This will not prevent operation of the instrument but may require future service.

Message 430: Less than 30 min of sheath fluid is remaining.

Message 410: Less than 10 minutes of sheath fluid is remaining. Please replenish your sheath fluid.

Message 401: Internal sheath reservoir level is low. CyAn will stop soon. Please replenish your sheath fluid.

(40)

CyAn Control

Panel Message Text

Message 402: Out of sheath fluid. Replenish sheath and press the Startup button on the CyAn Control Panel.

Message 404: Sheath quick connect is not completely engaged. Please check your connection.

Message 405: Internal reservoir overfilled. This will not prevent operation of the instrument but may require future service.

Message 406: Sheath subsystem is halted. Please check that you have sufficient sheath fluid.

Message 407: Sheath subsystem switch error. Service maybe required.

Message 705: Waste subsystem is halted. Please check waste tank level.

Message 730: Less than 30 min until the waste container is full.

Message 710: Less than 10 min until the waste container is full. Please empty the waste container.

Message 700: Waste container is full. Please empty the waste container.

Message 905: Low sheath pressure inside CyAn. Check connection between the CyAn and SMS.

Message 906: Low vacuum inside CyAn. Check connection between CyAn and SMS.

(41)

CyAn Control

Panel Message Text

Message 998: Waste subsystem halted due to loss of vacuum. If during operation, please check vacuum pump, waste quick connect or waste tank level. Check the waste filter on the waste container to see if it has clogged. If so, replace the waste filter.

<Message in top portion of Control Panel>

Message 999: SMS Power fault. Reset SMS to continue. If this problem reoccurs, please call Dako for technical support.

Changing Out Sheath and Waste Containers

Use the shutdown fluidics procedure to fill and empty sheath and waste containers when running samples.

1. Click Fluidics Off on the CyAn Control Panel.

WARNING

Remove the waste container and the sheath container from the cart before emptying or filling. Dispose of the contents of the waste container in accordance with local, state, and federal regulations.

Do not drop the waste or sheath containers on the Sheath Management System. Doing so may result in improper calibration of the load cells. 2. At the waste container, release the quick-connects.

3. Remove the waste container from the cart, remove the lid, and empty. Dispose of the contents of the waste container in accordance with local, state, and federal regulations. 4. Place an appropriate amount and type of disinfectant in the bottom of the waste tank to

ensure effective killing action when the tank is full. All biologicals introduced into the waste tank must be in contact with the disinfectant for a minimum of 10 minutes before being discarded. See Appendix A on page Error! Bookmark not defined. for a list of Approved Cleaners and Disinfectant products.

As an example: If 200mL of regular household bleach (5.25% active chlorine) is placed in the container, this will provide approximately 500ppm available chlorine when the tank’s full capacity of 20L has been reached. If the samples you are running will not be effectively killed by this concentration of sodium hypochlorite solution, an increased amount of bleach should be used to achieve effective disinfection concentration. Please keep in mind that diluted sodium hypochlorite solutions begin losing their effectiveness after 24 hours.

5. Replace the lid and tighten, place the container back in its position on the cart, and reconnect the waste container using the quick-connects.

(42)

6. Restore the sheath fluid by either replacing the entire sheath cubitainer or refilling the plastic carboy, depending on which type of sheath container is used with your Sheath Management System:

ƒ If you are using the disposable sheath cubitainer, release the quick-connects. Dispose of the entire cubitainer in accordance with local, state, and federal regulations (for example, recycle the outside cardboard box and place the plastic bladder in an appropriate

receptacle). Place a new sheath cubitainer back in its position on the cart. Locate the quick connect inside the box. Spray the quick connect with 70% ethanol. Reconnect the cubitainer to the SMS cart.

ƒ If you are using the plastic carboy, release the quick-connects. Remove the carboy from the cart, loosen the lid, and fill with particulate-free deionized water (dH2O) or a suitable sheath fluid. Seal the lid, and place the container back in its position on the cart. Spray the quick connect with 70% ethanol. Reconnect the sheath container to the SMS cart using the quick-connects.

7. Click Startup on the Cyan Control Panel.

Replacing Cleaner Fluid

1. Release the cleaner cubitainer quick-connect.

2. Unscrew the cap from the spent cleaner cubitainer. Retain this cap for use on the new cubitainer. Do not place this cap and its attached tubing on the floor or other surface where it will be in contact with contaminants. Place on clean paper towel if needed during change out of the cleaner cubitainers.

3. Dispose of the entire cleaner cubitainer in accordance with local, state, and federal

regulations. (for example, recycle the outside cardboard box and place the plastic bladder in an appropriate receptacle).

4. Remove the cardboard punch-out on a new cleaner cubitainer. Holding onto the ring around the cap, pull up on the lid so that the lid extends up to the cardboard.

5. Remove the cap and replace it with the cap from the previous cubitainer.

6. Place the new cleaner cubitainer back in its position on the cart. Spray the quick connect with 70% ethanol. Reconnect the cubitainer to the SMS cart.

(43)

Section 6

Summit Software Basics

Acquire and Analyze up to 10 Million Events

Summit software, on CyAn ADP with all parameters enabled, has the capability to acquire, save, and analyze up to 10 million events. The following graphics depict 10 million events acquired at 25,000 events per second, during an eight color experiment, with 37 parameters enabled on the CyAn ADP.

(44)

The following procedures describe some of the most commonly-used functions of Summit software. For complete Summit software documentation, see the Summit online help by clicking Table of Contents on the Summit Help menu.

Create a New Database

1. Double-click the Summit software shortcut on the Windows® desktop. The Select database

dialog box appears.

2. Select CYAN for the database from the drop-down list. This allows you to create a database for CyAn.

(45)

4. Specify the folder to create the new database in by clicking the Save in drop-down list and navigating to the desired folder location. To create a new folder, click

the Create New Folder button, type a folder name, and then press ENTER.

5. Specify a name for the database in the File name field, and then click Save.

The new database is created and saved to the specified folder. Immediately following this procedure, Summit software will open the newly-created database when launched.

(46)

Open an Existing Database

1. Double-click the Summit shortcut on the Windows® desktop. The Select database dialog box appears.

2. Click the Browse button. The Open Database dialog box appears.

3. Find the database you want to open by using the Look in drop-down list to browse all local drives and folders.

(47)

Summit software now opens using the selected database.

Create a Histogram

1. Select the Histogram tab in the Control Panel.

2. In the Create Histograms panel, double-click a parameter for the X Axis in the X-parameter

(48)

A histogram is created with the selected parameter.

Create a Dot Plot

1. Select a parameter for the X Axis in the X-parameter column. 2. Double-click a parameter for the Y Axis in the Y-parameter column.

(49)

A dot plot is created with the selected X and Y Axis parameters.

Open FCS (Listmode) Files

When a data file is opened in Summit software, the entire data set is not fully copied into the current database. Instead, a link or “shortcut” to the original file is created. A full listing of all opened files associated with the database can be viewed in the Database samples dialog box. 1. On the File menu, click Load Sample. The Open dialog box appears.

(50)

2. Locate the FCS file to open by browsing to a folder in the Look in list.

3. Select the FCS file you want to open. To see more detailed information about the file, click

More Info.

4. To select multiple files, press the CTRL key while selecting the files.

5. Click Open.

(51)

Set a Gate from a Single Region

1. Open a dot plot and create a region (R1) for gating.
(52)

3. Double-click in a target dot plot or histogram to apply the gate.

Set a Serial Gate

1. Open a dot plot gated on region (R1).

(53)

3. Right-click in region (R2) and then click Set Gates.

The Combine region and gate? dialog box appears.

4. Click Yes to activate serial gating. Click No to gate using only the specified region.

5. Check the Make my choice the default and don’t ask me again check box to disable this dialog box in the future. If checked, this option is set as the default in Summit software. The default can be changed in the Preferences dialog box.

(54)

6. Click in a histogram or dot plot to apply the gate.

If you clicked Yes in step 4, the serial (or sequential) gate logic is applied to the histogram or dot plot. The graphic below shows a gated histogram involving regions 1 and 2.

Note When a region in a dot plot or histogram is used for gating, the serial gate option works by automatically appending any gate regions applied to that plot to the new gate. For example, a dot plot is gated based on two regions (R1 & R2). If an R4 region is created in that plot and is used as a gate on another histogram or dot plot, the serial gate option defines the newly applied gate as (R1 & R2 & R4). Electing not to serial gate would apply a gate involving region (R4) only to the target histogram or dot plot.

(55)

Define Gate Logic

1. Select the Gating tab in Control Panel.

2. Click the upper-left icon in the Gate Logic Builder panel and then click New to create a new gate.

3. Double-click a text field in the Expression column. The Edit gate expression dialog box appears.

(56)

4. Select one or more regions for the gate and then click OK. Gates can be defined to include those events that fall inside or outside specific regions. To clear all selected regions, click

Clear.

Note The number of regions available correlates to the number of regions created in the current workspace.

The newly defined gate expression is displayed in the Expression column of the Gate Logic Builder column and the current gate scheme is displayed in the Gate Scheme panel.

(57)

Shortcut Keys

The following list describes available keyboard shortcuts in Summit software. These shortcuts are provided to help you quickly perform various operations in the software.

Shortcut Key Function

F1 Opens the Summit software online help system.

F2 Starts/Stops acquiring events.

F3 Saves acquisition data to a drive (C:, D:, etc.), the network, disk, or CD. F7 Opens the Create New Histograms dialog box for the current workspace. F8 Opens a window that displays all loaded sample files in the current database. Left arrow key Sets the displayed sample in the current workspace or protocol as the first file in the Database samples window. If more than one file group is present in the sample window (that is, organized in multiple folders), the last sample in the currently selected group is displayed. Also repositions a region to the left when selected.

Right arrow key Sets the displayed sample in the current workspace or protocol as the last file in the Database samples window. If more than one file group is present in the sample window (that is, organized in multiple folders), the last sample in the currently selected group is displayed. Also repositions a region to the right when selected.

Down arrow key Sets the displayed sample in the current workspace or protocol as the next file in the Database samples window. If more than one file group is present in the sample window (that is, organized in multiple folders), the last sample in the currently selected group is displayed. Also repositions a region downward when selected.

Up arrow key Sets the displayed sample in the current workspace or protocol as the previous file in the Database samples window. If more than one file group is present in the sample window (that is, organized in multiple folders), the last sample in the currently selected group is displayed. Also repositions a region upward when selected.

SHIFT+ARROW Expands a region in the arrow direction (left, right, up, down) when selected. CTRL+ARROW Contracts a region in the arrow direction (left, right, up, down) when selected

CTRL+O Opens a dialog box to open one or more FCS listmode files.

CTRL+S Saves modified items on the Summit desktop that have changed (or auto-save is invoked at periodic intervals and when Summit software is closed).

CTRL+P Prints the current view of the Summit software desktop.

CTRL+C Copies the selected region, which can be pasted into a histogram. CTRL+V Pastes the copied region into a histogram.

(58)

Shortcut Key Function

CTRL+Z Clears the event buffer of all acquired events.

ALT+F4 Exits Summit software.

+ Expands one node when viewing a folder list in a window (use + on the numeric keypad).

- Collapses one node when viewing a folder list in a window (use – on the numeric keypad).

* Expands an entire folder tree contained in a window (use * on the numeric keypad).

(59)

Section 7

CyAn ADP Maintenance

Regular maintenance of the CyAn ADP instrument is recommended as described in this section. System integrity can be effectively maintained by assuring cleanliness of the fluidics. In addition to performing preventive maintenance procedures, we also recommend that you establish and perform other laboratory procedures for routine operations such as backing up your data and experimental protocols.

Daily Preventive Maintenance

Follow the cleaning steps outlined in Section 4: Startup and Shutdown Procedures on page 32 on a daily basis.

Weekly Preventive Maintenance

Reusable Sheath Containers

The following steps should be performed on a weekly basis.

If using the reusable sheath carboy, it should be cleaned and disinfected once a week. The container should be handled as aseptically as possible during cleaning. A location where a sink, hot water, counter space, and DI water are available is required.

1. Put on gloves, a lab coat, and safety glasses.

2. Remove cap of container and set aside on clean paper towel.

3. Add a small amount of a phosphate-free laboratory grade soap and enough hot water to create a lather. Replace cap. Agitate or shake vigorously for 30 seconds.

4. Empty out soap and rinse with hot water until soap is removed. 5. Rinse with DI Water.

6. With 70% Ethanol, spray the cap and threads on the container and the exterior of container and wipe clean.

7. The carboy can now be filled with sheath fluid.

Note: If using the disposable sheath cubitainers, no weekly maintenance of the sheath container is needed; however, please follow the instructions starting on page 41 when switching out disposable sheath cubitainers.

(60)

Head Unit and SMS Cart

The exterior surfaces of the CyAn ADP should be cleaned on a weekly basis. Any general-purpose surface cleaner/disinfectant can be used. All surfaces should be wiped down to remove dust, fingerprints, and any other soiled marks. The sample injection area should be thoroughly cleaned and disinfected as well. It is highly recommended that the computer keyboard and especially the mouse are cleaned and disinfected on a weekly basis also.

Fluidic Lines

SMS Clean & Rinse Cycle

Note: Before running the clean & rinse cycle, make sure you have enough sheath fluid and cleaner fluid to complete the cycle.

ƒ On the CyAn Control Panel, click the Clean/ Rinse button.

A message box will warn you that the Clean and Rinse process will take 10-15 minutes. Click OK

to continue. The clean cycle takes seven minutes to complete, followed by a rinse cycle that takes seven minutes to complete.

Monthly Preventive Maintenance

System Decontamination

Performing this decontamination procedure is a necessary part of system maintenance and will ensure the instrument’s longevity. To prevent the buildup of debris and potential contamination of the fluidics system on the CyAn ADP, the system must be decontaminated on a monthly basis. A disinfectant should be run through the system to eliminate any sample buildup or microorganisms that may have entered the system through normal everyday use. Only Dako-approved

disinfectants can be used.

The following procedure must be followed once per month. A reusable sheath carboy is required. 1. Ensure that the instrument is ON and the software is open with at least one histogram

created.

2. Put on gloves, safety glasses, and a laboratory coat. Shut down the fluidics on the CyAn ADP.

3. In a clean reusable sheath carboy make a minimum of two liters of a 1:5 dilution of Dako Decontamination Solution (equivalent to 2000 ppm active chlorine).

4. Spray the quick connect with 70% Ethanol and hook up to sheath port on the SMS cart.* 5. Start up the fluidics. This will move the Decontamination Solution in the sheath tank through

the system.

6. Place a test tube with 4mL of 1:5 diluted Decontamination Solution on the sample probe. 7. Press F2 to acquire. Sample flow rate should be on LOW (corresponding to ~ 10-20 µL/min). 8. Let the Decontamination Solution run from both the sheath tank and the sample for 60

(61)

9. After 60 minutes, click Debubble on the CyAn Control Panel. After this function is complete, remove the tube from the sample probe and Backflush once.

10. Shut down the fluidics on the CyAn ADP.

11. Replace sheath with DI water. Hook up a disposable box of sterile DI water or thoroughly rinse the reusable carboy with DI water and then fill with 5 liters of fresh DI water. 12. Spray quick connect with 70% Ethanol and hook up to sheath port on SMS cart.* 13. Start up fluidics and place a tube with 4mL DI water on the sample probe. 14. Click F2 to acquire. Sample flow rate should be on LOW.

15. Let rinse for a minimum of 90 minutes. Note: DI water may need to be refilled on the sample probe during the rinse time. Refill as necessary.

16. After 90 minutes have elapsed, click Debubble on the CyAn Control Panel. After this function is complete, remove the tube from the sample probe and Backflush once.

17. Perform an SMS Clean & Rinse Cycle. See Weekly Preventative Maintenance – Fluidic Lines above.

The system is now fully decontaminated and ready to continue cellular analysis.

*Note: A sheath container with approximately 1L of fluid in it may alert the system that the tank is low and that it should be changed out soon. The decontamination cycle will be able to run for the specified time without the instrument fluidics stopping. Please call Customer Care if any problems arise with performing this procedure.

Laser Interlock Maintenance Procedure

1. Follow the instrument startup procedure. Make sure the lasers are on and the laser shutters are open.

2. Visually inspect the housing on the instrument to verify that panels or covers are fitted and tight so that all laser energy is contained in the interior.

3. Facing the instrument, lift the lid approximately one inch (2.54 cm). The LED light(s) on the right front of the cover should go out, verifying that the laser shutters are closed.

4. Close the lid.

5. If the LED light(s) did not go out when the lid was opened, contact your local Dako Support Representative.

Biannual Maintenance Procedure

A Dako Field Service Representative should perform an maintenance check on the CyAn ADP every 6 months. To schedule a 6-month or annual maintenance service, contact your local Dako Support Representative.

(62)
(63)

Section 8

Troubleshooting

Please contact your local Dako Field Service Representative immediately for assistance with any instrument malfunction or service need.

WARNING

Do not attempt any maintenance on the CyAn ADP laser components. Laser maintenance should only be performed by specially trained, certified Dako Field Service Representatives.

The following table is a guide for troubleshooting CyAn ADP problems.

Table 6.1: CyAn ADP Troubleshooting Guide

Problem Action

CyAn ADP has no power Make sure that the power plug is firmly attached to the wall.

Make sure that the power switch on the back of the CyAn ADP is on. There are no

operator-replaceable fuses in the CyAn ADP. Acquiring data, but no display. Make sure that the lasers are on.

Make sure that the laser shutters are open on the CyAn Control Panel.

Median values spike or fall suddenly and recover on all or most fluorescence channels across multiple laser paths.

Debubble (Click Debubble on the Control Panel.)

Indicator lights in the Sheath Management System portion of the Control Panel change from green to amber or red.

Hover the cursor over the indicator light to get more information about the error and instructions on how to correct the problem. Information on the problem can also be found in Section 5.

Data integrity – data looks questionable

during acquisition. Remove sample. Perform backflush and debubble. Check sample preparation. Re-run sample. If problem remains, check alignment with beads. Perform another backflush and debubble. If problem persists, contact Customer Care.

(64)
(65)

Section 9

Six-Color Compensation Experiment

Background

Compensation is the process of resolving the actual intensities from each antibody conjugate in a multicolored sample. Using single color staining or single color fluorochromes as a compensation control makes it possible to define the relative amount of light that ends up in another detector. For example, compensation can be used to determine how much light from FITC ends up in the FL2 detector and, conversely, how much light from PE ends up in the FL1 detector (see figure below). This percentage is defined as the spi

References

Related documents