High Sensitivity ICP-MS:
Overcome the problem of complex samples
Meike Hamester, Andrew Toms and René Chemnitzer Bruker Daltonics
Berlin, Germany - Milton, Canada
Historical Product Roadmap
2003 2005
Introduced world’s most sensitive ICP-MS (Varian)
Introduced model 810 and 820 using CRI technology (Varian)
2010 Bruker acquires Varian ICP-MS
2011
Bruker launches aurora M90, world’s most sensitive ICP-MS
Bruker 810
No CRI interface Highest sensitivity No selectivity
Bruker 820
CRI interface High sensitivity High selectivity
Bruker aurora M90
CRI interface for Highest Sensitivity and Selectivity
Non CRI interface for utmost sensitivity
aurora M90:
combining highest sensitivity with selectivity
translates to results in advantage
high
sensitivity/high signal to noise ratio
low detection limits low detection limits low detection limits
short integration times short analysis times higher sample throughput
higher dilution factors less matrix effects less cone depositions less cross contamination reduced washout time
less drifts
less QC failures less re-calibration less sample volume
higher sample throughput
higher sensitivity in
interference mode low detection limits for spectrally interferenced analytes
accurate results at lowest concentration levels
high sensitivity for laser
ablation low laser energy higher spatial resolution
lower detection limits in solid samples
lower detection limits for
LC / GC coupling lower detection limits for elemental species (As, Sn, Br…)
simpler procedures, less sample volume
Benefits and side benefits of high sensitivity
5
The Bruker aurora M90 ICP-MS
Multi Patented Technology by Bruker Daltonics
Sensitivity:
Reflecting Optics concept
F
Parabolic mirror 3D control
D E
energy spread
90 degree Reflection concept for ions9Be ~ 0.5eV
115In ~ 2eV
232Th ~ 5eV
All ions are focused into quad entrance
Sensitivity:
Ion Mirror for ICP-MS
unwanted particles
electrostatic field
MS quad
• Four electrode segments to provide 100% multidimensional control of the ion beam by
mean of parabolic field shaping
• Hollow structure allows
photons, neutrals, & particles to pass straight through to pump
Sensitivity / Signal to noise ratio:
Curved Fringing Rods to minimise noise
• Metastables can travel through the Quadrupole
• Reaching the detector area they may cause ion-electron emission
• Those pulses can not be distinguished from
analyte ions
• The background would sacrifice BECs & DLs values
• The Curved Fringes minimise continues background
Key for
low background/
high signal to noise ratio
1 2 3 4
5 6 7
Ions Neutrals Photons To detector for measurement
Removed by vacuum system
Ion Mirror
Sensitivity and Selectivity:
CRI Interface and 90 Degree ‘Reflecting’ Ion Optics
Kinetic Energy
Discrimination
gas
Exponential reduction Original
plasma jet
New, clean Plasma expansion Interference
gas
CRI cone
10
Uses He and (H2) gasses Collisional / Reactive
Heto remove 16O35Cl+ on 51V+ H2 to remove 40Ar+ on 40Ca+
Selectivity:
Collision-Reaction Interface – CRI (II)
Benefits
• One-button interference management
• Auto-optimization
• Fast switching between gases
• Low maintenance
• No extra consumables
Skimmer Cone
Typical aurora M90 performance:
Normal Sensitivity and High Sensitivity modes
Normal Mode High Sensitivity
Mode (1) High Sensitivity Mode (2)
9Be (cps/ppb) 30660 102483 289763
59Co (cps/ppb) 387949 987589 1906846
115In (cps/ppb) 426560 1032937 2073298
238U (cps/ppb) 528568 936320 1723943
Background
@5amu 0.1 cps 1 cps 2 cps
CeO+/Ce+ 0.9 % 0.7 % 3.3 %
Ba++/Ba+ 0.9 % 1.9 % 1.4%
• Nebulizer: concentric glass, 400 L/min
• Spraychamber: scott-type, glas
•
Focus point lifted up•
Ion beam diffused•
3 mm focus•
Sensitivity attenuatedComputer Modeling -Normal Mode
Ion beam focus Y dimension control
12
Normal
•
Ion beam focus onentrance of Quadrupole
•
Ion beam focus: 1mmComputer Modeling - High Sensitivity
Ion beam focus Y dimension control
13
High
Sensitivity: Uranium
(high sensitivity mode: Ion beam in focus)
1.700.000 cps/ppb
238U
9
Be in Urine
• Dilution 1:10
• MDL: 0.3 ppt
Sensitivity: Beryllium in Urine
Gadolinium
They are used for:
• conventional diagnostic radiology
• angiography
• computed tomography (CT)
• magnetic-resonance-tomography (MRT)
Gadolinium compounds are used as contrast agents in medical checkups as they intensify differences in density between different types of tissues
Water soluble unspecific complexes of Gadoliniumkomplexe with 0,5 Mol/l Gd
linear complexes Gd-DTPA; -BMA
Magnevist, Omniscan
makrocyclic complexes Gd-DOTA; Gd-HP-DO3A Dotarem, Prohance
Gadolinium in densely populated areas of Berlin
17
Sensitivity: Detection of Gadolinium (
158Gd)
508.000 cps/ppb Gd (158Gd)
Selectivity:
Typical Interference: 40Ar35Cl interference on 75As
19
Reducing CaO (and ArO) interference on Fe
Normal Sensitivity CRI no gas
Normal Sensitivity CRI He
High Sensitivity CRI no gas
High Sensitivity CRI He
cps/ppb 89219 22354 731393 124151
0 100000 200000 300000 400000 500000 600000 700000 800000
Sensitivity [cps/ppb]
Sensitivity: Different CRI Modes
59
Co
Sensitivity and Selectivity:
BEC: Different modes of operation (all CRI Interface)
59
Co
NS NS & Ca NS & Ca &
He HS HS & Ca HS & Ca & He
BEC [ppt] 6.00 60.9 5.10 5.8 82 6.2
0.00 10.00 20.00 30.00 40.00 50.00 60.00 70.00 80.00 90.00
BEC [ppt]
100 ppm Ca
Normal Sensitivity Mode
• Ion beam ‘de-focused’
• Sensitivities > 200.000 cps / ppb
• CRI II interface utilizing He and H2 for interference reduction
High Sensitivity Mode
• Ion beam in focus
• Sensitivities > 1.000.000 cps / ppb
• CRI II interface for interference reduction
• Sensitivities > 100.000 cps / ppb in He mode
Bruker aurora M90 ICP-MS:
combining highest sensitivity with selectivity
Many other features…
Plasma generation, all digital detector, automated aerosol dilution…….
Bruker aurora M90 ICP-MS
…unique features and fascinating applications
•
Ground-breaking Plasma exitation – the ‚Turner interlaced coils‘•
Intelligent detector design – full digital DDEM detector•
Full range of applicationsBruker RF Plasma System -”Twin Coil” design
• Two interlaced RF coils
Driven in opposing directions
• Results in balanced RF field for plasma generation
• Very robust plasma
Efficient energy transfer to plasma
• No secondary discharge
Low kinetic energy distribution of ions
• No mechanical shielding required
For cool plasma operation
High sensitivity
Low oxide ratios
• Ideal for laser ablation
Decomposition of laser aerosol
Detector Technology in the Bruker M90
•
Detectors in mass spectrometers convert ions into electrical pulses•
The pulses are measured or counted by the electronics, resulting in the signal observed by the operator26
Quadrupole
Signal Output e-
e- e-
+ e-
How a Detector Works
Discrete Dynode Electron Multiplier (DDEM)
•
Impact of ion results in multiple ejections of electrons from dynode surface (conversion)•
Multiple electron ejections continue at each dynode (amplification)•
Result - ion impact leads to the formation of a large pulse of electrons (detection)Quadrupole
Signal Output e-
e- e-
+ e-
Conventional ICP-MS Detectors – the Problem
• Traditional ICP-MS detectors will saturate once they reached a few million counts/sec
Quadrupole
Signal Output e-
e- e-
+ e-
• As ICP-MS instruments became more and more sensitive over time and were able to see lower concentrations of analytes, this meant the highest concentration the
instrument could read without over-ranging also dropped.
Higher concentration samples had to be diluted
Conventional ICP-MS Detectors -the “Solution”?
• To overcome this limitation, the dual-mode or “pulse- analog” detector was developed
Quadrupole
Signal Output e-
e- e-
+ e-
• Instead of counting individual pulses, it ‘averages’
high rates of pulses into a steady current
• This current is proportional to the incoming ion flux from the quadrupole
Conventional ICP-MS Detectors -the problems with the “Solution”
•
The slope of the response in analog mode is often different than in pulse-counting mode•
Difficult to make a single calibration curve over wide range•
The ‘cross-calibration’ can drift, and requires frequent verification or re-calibration – weekly or even daily•
Detector lifetime can begreatly shortened in analog mode
Schematic of Detector Operation
•
The detector converts ions into electrical pulses.Adjustment of voltage applied to control dynode provides attenuation of final output signal
•
Three operating modes; None (Off), Medium and High31
‘V’ Control 4th Dynode
Signal Output
Quadrupole
Gain Controlled
Ion to e-
Conversion
Amplification e-
e- e-
+ e-
32
•
Unique to Bruker•
Long lifetime due to gain control protection•
> 109 dynamic range•
Fast operation•
Simple attenuation correction set up in software•
Attenuation of electrons not ions, avoids mass biaseffects
•
Excellent long term stability and precisionBruker All Digital Detector
Easy – Fast - Stable
Bruker ICP-MS Detector
– Short and Long Term Stability
• Detector attenuation factors were re-measured/re- calibrated 33X over 21 months
• Even though detector operating voltage changes over time, attenuation factors remain stable
• Typically, detector needs recalibration every 2-3 months
Isotope Norm-Med Atten Med-High Atten
Pb208 1.94% 2.34%
Th232 1.87% 2.56%
U238 1.76% 2.4%
Bruker ICP-MS Detector -Accuracy
•
Measurements made in all 3 modes•
Margin ofexperimental error
encompasses
‘true’ value in all cases
Shaulis, B., T. J. Lapen, and A. Toms (2010),
Signal linearity of an extended range pulse counting detector: Applications to accurate and precise U‐Pb dating of zircon by laser ablation quadrupole ICP‐MS,
Geochem. Geophys. Geosyst., 11, Q0AA11,doi:10.1029/2010GC003198.
35
NIST 1515 Apple Leaves Standard Reference Material
ICPMS wt.% stdev. Reference wt.% stdev.
Ca 1.494 0.016 1.526 0.015
Mg 0.269 0.002 0.271 0.008
K 1.62 0.01 1.61 0.02
ICPMS µg/g stdev. Reference µg/g stdev.
Al 285 2 286 9
AS 0.039 0.0007 0.038 0.0077
Ba 48 0.4 49 2
B 27.6 2.6 27 2
Cu 5.54 0.02 5.64 0.24
Pb 0.48 0.023 0.47 0.024
Mn 55 0.2 54 3
Hg 0.041 0.002 0.044 0.004
Mo 0.104 0.007 0.094 0.013
Ni 0.93 0.07 0.91 0.12
Se 0.05 0.011 0.05 0.009
Na 27.8 7.9 24.4 1.2
Sr 23 2 25 2
V 0.28 0.02 0.26 0.03
Zn 13.0 0.8 12.5 0.3
NIST 8436 Durum Wheat Flour
A Joint Material of Agriculture Canada and NIST
ICPMS wt.% stdev. Reference wt.% stdev.
K 0.311 0.023 0.318 0.014
Mg 0.097 0.0003 0.107 0.008
P 0.247 0.01 0.290 0.022
ICPMS mg/kg stdev. Reference mg/kg stdev.
Ca 261 35 278 26
Fe 39.1 1.2 41.5 4
Zn 22.4 1.8 22.2 1.7
Mn 16.3 1.0 16 1
Na 16.8 0.4 16.0 6.1
Al <15 BW 11.7 4.7
Cu 4.6 0.2 4.3 0.6
Ba 1.94 0.06 2.11 0.47
Rb 1.8 0.04 2 0.4
Se 1.17 0.02 1.23 0.09
Sr 1.16 0.07 1.19 0.09
Mo 0.6 0.03 0.7 0.12
Ni 0.17 0.02 0.17 0.08
Cd 0.09 0.01 0.11 0.05
Pb 0.020 0.009 0.023 0.006
Cr <0.05 0.023 0.009
V 0.024 - 0.021 0.006
Co 0.010 0.002 0.008 0.004
Hg <0.002 0.0004 0.0002
Applications – Food/Agriculture
(CRI-He)35Cl16O,
34S16O1H
• Matrix and trace elements measured in one run using the automatic detector attenuation
• Elements interferred by molecular ions are measured with CRI in collision mode using He
36
Applications – Analysis of Nanoparticles
Analysis of Silver (Ag) Nanoparticles
•
The ‘number of signals’ is very dependent on the concentration.However, if the concentration is too high you will get signal overlap of multiple nanoparticles
•
Silver NP (PLANO GmbH, Wetzlar) where measured with an aurora M90 ICP-MS using a sample introduction with micro- concentric nebulizer and Twister Cyclonic Spray Chamber (GE, Australia).Important Parameters
•
The ‘Size’ of the particle determines the intensity of the signal•
The ‘Concentration’ defines the frequency of signals observed as number of peaks per minute37
Applications – Analysis of Nanoparticles
Analysis of 5nm Ag NPs using different dwell times
•
A criticle parameter in ICP-MS analysis of NP is the Dwell Time. Although a dwell time of <500 µs would be preferable, the sensitivity for the 5 nm particles is not sufficient to get a required signal to noise ratioApplications – Analysis of Nanoparticles
•
Single particle signals measured within 180 s were averaged.•
If the NP are assumed to be spherical in shape, the volume increases by third power with the diameter (Dd3).•
Therefore from 20 - 80nm (43) is a 64 fold increase in volume of the particle..!ICP-MS Signals of Silver NP with different diameter and concentration.
•
For Silver - good correlation between ‘Intensity and Particle Size’ can be demonstrated on the aurora M90 ICP-MSCd in Seawater (1/10 diluted)
40
Automated Aerosol Dilution
Minimizes sample preparation by extending matrix tolerance of ICP-MS
•
For samples containing 0.2 to 4% Total Dissolved Solids• Eg. seawater contains ~3.5% TDS
•
In line dilution via detuning of nebulizer efficiency& sheath gas dilution effect
•
Achieve 10-fold dilution without additional sample prep•
Suitable for seawater, metals and mining, high TDS effluentsAerosol Dilution-
One-Click Optimization - automatically
Auto-optimization progress
•
Automatic tuning routines optimize plasma and ion optic parameters for best sensitivity and low oxide and double charged formation43
Typical Analysis Settings
0.1-0.3 L/min 0.8-1.1 L/min
More efficient nebulization
Low gas dilution
Aerosol Dilution Settings
0.8-1.1 L/min 0.1-0.3 L/min
Less efficient nebulization
High gas dilution
Followed by automatic
optimization of ion optic settings…….
44
Auto-Optimization of Aerosol Dilution
Pre-Programmed Dilution Settings
Low Dilution
5-fold dilution
Samples with ~1-2% TDS Metals and mining, effluents
High Dilution
10-fold dilution
Samples with ~2-4% TDS
• Seawater
5-fold reduction in oxide interferences
• Rare Earths
Ability to measure As and Se directly at masses 75 and 78
Other instruments need to add oxygen to measure masses 91 and 94
% Recovery of 10ug/L multi-element spiked into undiluted seawater
0%
100%
% Recovery 105% 109% 99% 109% 92% 95% 110% 101%
56Fe 75As 78Se 52Cr 59Co 68Zn 114Cd Pb
Seawater Analysis using Aerosol Dilution
Summary
All-digital Detection System
• Only detector to offer 109 working range in pulse- counting (digital) mode
• No cross-calibration of digital-analog modes required
Unmatched Sensitivity
• High efficiency ion mirror
• Excellent detection limits
• Attractive to laser ablation
Simplified Interference Management
• Innovative Collision-Reaction Interface (CRI II)
• Removes interferences at plasma interface, not the ion beam
• Simplified setup and maintenance
Low Maintenance Design
• Hollow ion mirror design requires no cleaning
• No need for additional cleaning / replace of interference management technology
27.06.2012 46
Future article:
Atomic Perspectives Column:
Efficient Removal of Polyatomic
Spectral Interferences for the Multielement Analysis of Complex
Human Biological Samples by ICP-MS
Spectroscopy 27 (7), pp 20-27, July 2012