0095-1137/81/040742-08$02.00/0
Evaluation of Five
Gentamicin Assay Procedures For Clinical
Microbiology Laboratories
SALLY T. SELEPAK, FRANK G. WITEBSKY,* E. ARTHUR ROBERTSON, AND JAMES D. MAcLOWRY
ClinicalPathologyDepartment, National Institutesof Health,Bethesda, Maryland20205 Received 10October1980/Accepted 13January1981
Fivegentamicin assayprocedures (a bioassay,anenzymeimmunoassay,alatex agglutinationinhibitiontest,afluorescenceimmunoassay,and a radioimmunoas-say) were evaluated to determine which was optimal for our laboratory. The evaluationwasbased onrecoveryandprecisionstudies and results ofanalysesof patientsamples,aswell as technicalassayperformance factors. The latex agglu-tinationinhibition testappearsuseful for laboratoriesperforming onlyoccasional assays for gentamicin; however, the fact that some rheumatoid factor-positive sera,as well as some other sera for unknownreasons, maygive falselylow values isapotentialdrawback to thisprocedure.Because of itsaccuracy,precision, rapid turn-aroundtime,and relativesimplicityofperformance,weselected theenzyme immunoassay procedurefor routine use forgentamicin assaysin ourlaboratory.
Gentamicin continues to be a widely used antimicrobial agentforthetreatment ofa vari-ety of serious infections. It is important to be able to determine quickly and accurately the concentration ofgentamicin in serum, and
oc-casionallyin otherbothfluids,because of its low therapeutic index. A variety of techniques is available for gentamicin measurement. It was
thepurpose of thisstudytoevaluatefive differ-ent techniques, a bioassay, an enzyme
immu-noassay(EMIT),alatexagglutinationinhibition (card) test, afluorescence immunoassay (FIA),
and a radioimmunoassay (RIA), to determine whichwasoptimalforourlaboratory.
MATERIALS AND METHODS Recovery studies. Gentamicin sulfate (Schering Corp., Kenilworth, N.J.) concentrated solution was
prepared by drying powderedgentamicinsulfate for3 h at 60°C under partial vacuum and then weighing and dissolving the gentamicin in 0.1 M potassium phosphate buffer (pH8.0) to a finalconcentration of 1,000
Mg/ml.
This solution was then diluted in 18 separate humansera togivefinal concentrations of 0, 1, 2, 4, 8, 12, and16dg/ml.Fourcategoriesofsera were selected frompatientsnotreceivinggentamicin: nor-maltovisualinspection (nine patients); lipemic (three patients); icteric (three patients); hemolyzed (three patients). Portions (250 ,l) of each serum at each concentration werestored at-70°C until just before assay. These studies were not performed using the bioassay or RIA methods. For statistical handling of the data, when anassay gave a result of"lessthan"some lower limiton asampletowhichnogentamicin had been added, a concentration of0
Mg/ml
was as-signed to that sample for that determination. If anygentamicin had been added andaresult less than some
lowerlimit wasobtained,aconcentration of0.5kg/ml
wasassigned tothat sample for that determination. If a result of "greater than" some upper limit was ob-tained on one or both of duplicate determinations, neither of thevalueswasused in the statisticalanalysis of thedata.
Precision studies. Concentrated gentamicin sul-fate solution (described above) was added to pooled human serum from Clinical Center patients to give final concentrations of 3, 6, and 12 ig of gentamicin perml. Portions of250fland5ml ofserum ateach concentration were storedat -70°C until just before assay. To determine intrarun precision, 30 assays at each concentration were performed on the same day. Precision studies were not performed by the bioassay procedure.
Patient samples. A total of 110 serum samples wereselected fromClinical Center patients who were being treated with gentamicin and other antimicrobial agents.Samples were selected after assay by our rou-tine bioassay procedure to include a range of genta-micin concentrations from <1 to 10.5 fig/ml. After completion of the bioassay, the residual serum was
keptfrozen at -20°C in screw-capped glass vials for upto 7months beforebeing thawed and divided into
250-dl portions, which were refrozen and stored at -70°C. These were thawed just before assay by one of the non-bioassayprocedures. For statistical handling ofthe data, when an assay gave a result less than some lower limit (generally 1
tig/ml),
a gentamicin concen-tration of 0.5lig/ml
was assigned to that specimen. If aresult greaterthansomeupper limit was obtained, that value was not used in the statistical analysis of thedata.Bioassay procedure. Bioassay results were ob-tained by ourroutine procedure,performed as previ-ously described (2), except that the gentamicin sulfate was dissolved in 0.1 M potassium phosphate buffer 742
on February 7, 2020 by guest
http://jcm.asm.org/
(pH 8.0), standards were prepared on a weekly basis ratherthan daily,and specimens and standards were
storedinglass, rather than plastic, vials. The highest concentration measurable in ourroutineprocedureis 16,ug/ml.
EMIT. EMIT (Syva, Palo Alto, Calif.) was per-formed according to the directions of the reagents
manufacturer (Syva), except that calibrator curve
points were all established byduplicate rather than
single determinations. (The manufacturer recom-mends thatonly the0calibratorbe run induplicate.)
Theprinciple ofthis procedureinvolves competitive bindingbetweengentamicinin asampleand a known amountofenzyme-labeledgentamicin for an anti-gen-tamicinantibody. Theenzymeemployedisglucose
6-phosphate dehydrogenasederived from the bacterium Leuconostoc mesenteroides. Binding of the
enzyme-labeled gentamicin decreases the activity of the en-zyme. The enzymatic activity remaining results in
conversionof thecoenzyme nicotinamideadenine
di-nucleotidetoreducednicotinamide adenine dinucleo-tide; the resulting absorbance change is measured
spectrophotometrically. The instrumentation used was a model 1500automatic pipetter-diluter (CAVRO
Scientific Instruments, Los Altos, Calif.), a Gilford
Stasar III spectrophotometer (Gilford Instrument
Laboratories, Inc., Columbia, Md.) and a Syva CP-1000 timer-printer (manufactured for Syva by
Ox-bridge, Inc., Mountain View, Calif.). Values for the
absorbance changesof thecalibrator standardswere
plottedonspecial graphpaperprovidedwitheach kit
ofreagents;gentamicinconcentrations forthesamples
weredeterminedfrom theresultingcurves.The
high-estconcentration measurable bythisprocedureis 16
ig/ml.
Cardtest.The latex agglutination inhibition test
(Macro-Vue Card Test, Hynson, Westcott and
Dun-ning, Divisionof Becton Dickinson, Baltimore,Md.)
was done asspecified in the directionsofthe
manu-facturer, exceptthat all determinationsweredone in
duplicateand theresultswereaveraged.Theprinciple of thisprocedureinvolvesthe inhibitionof
agglutina-tionofgentamicin-sensitized latex particleswhichare
addedto amixtureofsample andaknownamountof anti-gentamicinantibody. Quantitationofgentamicin concentrationisachievedbymultiplyingthe recipro-cal ofthehighest dilution ofsampleshowing
aggluti-nationinhibitionbythe lowestconcentration of
gen-tamicin standard also showing agglutination inhibi-tion. The highest concentration measurable by this procedurevariesdependingupon theendpointof the
standard;it rangesfrom12.8to19.2,g/ml.
FIA.FIA(AmesDivision,MilesLaboratories,Inc., Elkhart,Ind.) wasdone asspecifiedinthe directions of the reagents manufacturer, except that up to 39
sampleswere run per curve; this wasspecified as a permissibleprocedure bythemanufacturer. The prin-ciple of this procedure involves competitive binding
betweengentamicininasampleandaknownamount of sisomicin labeled with a fluorogenic substrate (a derivativeof
umbelliferyl-,/-D-galactoside)
forananti-gentamicin-sisomicinantibody.Bindingof the labeled sisomicintoantibodypreventshydrolysisof the
fluo-rogenic substrate. /3-Galactosidase hydrolyzes
avail-able fluorogenic substrate to produce a fluorescent product,the amount of which is measured fluoromet-rically. Thefluorometer used was the Aminco Fluoro-Colorimeter (American Instrument Co., Savage, Md.). Fluorescence units of the standards were plotted againstgentamicinconcentrationon graph paper pro-vided by Ames; the gentamicinconcentrations of the samples were determined from the resulting curves. The highest concentration measurable by this proce-dure is 12
fig/ml.
RIA. RIAs were performed by Herner Analytics, Inc. (Rockville, Md.) using the Monitor Science ra-dioimmunoassay kit (Monitor Science Corp., Newport Beach, Calif.) as specified in the directions of the reagents manufacturer. The highest concentration
measurable by thisprocedureis 16
tig/ml.
RESULTS
Table 1 shows the results of the recovery studies on normal, hemolyzed, icteric, and li-pemic sera. The results from only seven normal serafor the FIA are shown in the concentration rangeofOto 8
lig/ml
because the sera from the other two patients were used up in runs from which no results could be obtained because of unsatisfactory standard curves. Table 2 shows the results oflinear regression analysis of the data presented in Table 1 exceptthat the anal-ysis has been done only for the concentration range of O to 8fig/ml.
The 12- and 16-ftg/mlconcentrations were not used in this analysis because a full set of data was not available from all the procedures at these concentrations. An ideal method for measuring gentamicin would give a slope (m) of 1.0 and a y-intercept (b) of zero for each of the different types of sera. In addition, the correlation coefficient (of measured concentrationscompared with concentrations of
gentamicin actually added) would be 1.00, and the standard error of the estimate (interpretable
as a standard deviation [5]) would be 0.00. All
threeprocedures evaluatedbyrecovery studies appearedto perform satisfactorilywith normal and hemolyzed sera. With icteric sera the FIA tendedtooverestimate theamountof
gentami-cin present. Withlipemic sera the EMIT proce-dure tended to underestimate the amount of gentamicin present. This can be seen more clearly from the results in Table1than from the linearregression analysisinTable 2, which does notinclude theresults obtainedat12and16
jug/
ml. The FIA appeared tooverestimate slightlythe amount of gentamicin present in lipemic
sera, but the number of measurements was
small.
Table 3 shows the results of the studies on
intra-runprecision.
Table4shows the correlation coefficients for the linear regression lines obtained by
plotting
on February 7, 2020 by guest
http://jcm.asm.org/
TABLE 1. Recoverystudies
Recovery bymethod
Gentamicin FIA EMIT Cardtest
Serum concn added
(,ugrnl No.of f o. No. of
Mean +SD' No.o Mean +SD
Nolueof
Mean±SD valuesVisually O 0.0 0.0 7 0.0±0.0 9 0.0±0.0 9C
normal 1 1.1 0.5 7 1.2±0.2 9 0.9±0.3 9C
2 1.9+0.4 7 2.1±0.3 9 1.5±0.8 9C
4 4.2±0.6 7 3.9±0.2 9 3.4±1.2 9C
8 8.3+0.8 7 7.8±1.4 9 8.0±0.9 9
12 11.0 4d 10.7 8e 10.9±1.6 9
16 -f 14.1 35 13.8 4
Hemolyzed O 0.0±0.0 3 0.0±0.0 3 0.0±0.0 3
1 1.4±0.3 3 1.4±0.1 3 0.7±0.4 3
2 2.1±0.4 3 2.3±0.1 3 2.2±0.6 3
4 4.4±0.5 3 4.3±0.3 3 3.6±0.0 3
8 8.8±0.4 3 8.6±0.8 3 7.9± 1.5 3
12 - 11.5 1' 13.2± 1.2 3
16 J J 14.4 2
Icteric 0 0.6±1.0 3 0.0±0.0 3 0.0±0.0 3
1 1.7±0.6 3 1.1±0.0 3 1.0±0.5 3
2 3.1 0.2 3 2.0±0.0 3 1.7±0.1 3
4 5.2 0.2 3 4.1±0.2 3 3.5±0.2 3
8 10.1±0.3 3 7.6±0.0 3 7.6±0.7 3
12 11.8±0.5 3 12.8±1.4 3
16 - 16.0 li 14.4 2'
Lipemic 0 0.4±0.7 3 0.0±0.0 3 0.0±0.0 3
1 1.7±0.6 3 1.0±0.4 3 0.9±0.4 3
2 2.9±0.8 3 1.8±0.3 3 2.0±0.3 3
4 4.9±0.8 3 3.6±0.2 3 3.7±0.5 3
8 9.0±1.5 3 7.1±0.8 3 7.4±0.3 3
12 11.2 lm 8.8±0.6 3 12.0±0.0 3
16 l 10.8±1.6 3 14.4 2
aMeanmeasuredgentamicin concentration (micrograms per milliliter) and standard
deviation
(SD).bNumber ofvalues used for determining each mean and standard
deviation.
cSeetextfor discussion ofproblemswithtwo sera.
dBy FIA threesamplesatthis concentration were measured as>12
,g/ml.
eBy EMITonesampleatthis concentrationwasmeasuredas >16
jug/ml.
fBy FIA allsamplesatthisconcentrationweremeasured as>12,ug/ml. gBy EMIT sixsamplesatthisconcentration were measured as >16jg/ml.
hBycardtestfivesamplesatthis concentration were measured as greater thanthe highest concentration measurablebyatleastoneof thetwocards used.
By EMITtwosamplesatthis concentration were measured as >16,ug/ml. By EMIT all three samplesatthis concentration were measured as >16,tg/ml.
kBy card test onesampleat thisconcentration was measured asgreater thanthe highest concentration measurable by one of the two cards used.
'By card test one sampleat this concentration was measured as greater than the highest concentration measurablebyboth of thecards used.
mByFIAtwosamples at thisconcentration were measured as >12
jg/ml.
gentamicin concentration measured by each method against that measured by each other method for ail thepatient samples tested.The best correlation was obtained between the EMITand the RIA.
Figures 1, 2, 3, and 4 show the results of
plottingthedata forpatientsamplesasobtained
by RIA, bioassay, FIA, and card test,
respec-tively, against the results obtained by EMIT.
The dataaredisplayed inthis fashion because weultimately selectedthe EMITfor routine use inourlaboratory.Note inFig.4that there were fourpatient samplesfor which thecard test gave avalue oflessthan 1, whereasthe EMIT gave
744 SELEPAK ET AL. J. CLIN. MICROBIOL.
on February 7, 2020 by guest
http://jcm.asm.org/
EVALUATION
TABLE 2. Linearregression analysisof datafromTableI
Parameters ofequation
Serum Method rh SEE'
y m x b
Visually FIA FIA 1.0 Concn added +0.0 0.99 0.50
normal EMIT EMIT 1.0 Concn added +0.1 0.97 0.65
Card Card 1.0 Concn added -0.3 0.96 0.81
Hemolyzed FIA FIA 1.1 Concn added +0.1 0.99 0.36
EMIT EMIT 1.0 Concn added +0.2 0.99 0.37
Card Card 1.0 Concn added -0.1 0.97 0.70
Icteric FIA FIA 1.2 Conen added +0.6 0.99 0.51
EMIT EMIT 0.9 Concn added +0.1 1.00 0.12
Card Card 0.9 Concn added -0.1 0.99 0.38
Lipemic FIA FIA 1.1 Concn added +0.6 0.97 0.83
EMIT EMIT 0.9 Concn added +0.1 0.99 0.38
Card Card 0.9 Concn added +0.0 0.99 0.31
aEquation of
regression
line inthe form y=mx+b.br,Correlation coefficient.
SEE,Standarderrorof the estimate.
TABLE 3. Intra-run precision
Recoveryby method
Gentamicin FIA EMIT Card
concn added
()4/mMean±
SD<a
No. of val- Mean±SD No.ofval- Mean±SD No. ofval--ea
SI)a uesh - ues ues3 3.1 0.3 30 3.3±0.1 30 3.0±0.6 30
6 7.7+0.6 30 6.4±0.1 30 5.8±0.7 30
12 - 11.1±0.6 30 10.8±1.8 28
aMean measured gentamicinconcentration and standard
deviation
(SD) (micrograms per milliliter).bNumber of values used fordetermining each mean and standard
deviation.
CNomean orstandard
deviation
wascalculated for FIA at this sample concentration because 20 of the 30 samplesweremeasuredas >12,tg/ml.dBy cardtest twosamplesatthis concentration weremeasuredasgreater thanthehighestconcentration
measurablebyoneof thetwocards used for eachsample.
TABLE 4. Correlationcoefficients forresults obtainedonpatient samples
Method Bioassay EMIT RIA FIA Card
Bioassay 1.00
EMIT 0.88 1.00
RIA 0.86 0.95 1.00
FIA 0.86 0.90 0.89 1.00
Card 0.82 0.85 0.84 0.85 1.00
values for these
samples
of2.5to8.1,ug/ml. Forthese
samples
the results of the other assayprocedures(except for thebioassayon one sam-ple) were in agreementwith the EMIT results. After this
discrepancy
wasnoted,
thesesamples
were checked for the presence of rheumatoid factor. Rheumatoid factorwaspresentintwoof thesamples,butnotinthesampleshowingthe greatest
discrepancy.
We retained thetwo rheu-matoidfactor-positive serainthe data base forE
`.10
co
o8
.2 6 o
EO
:>4 0
<: o Z~~~
ce 2 ç/ o
O
I
0 2 4 6 8 10 12
EMIT Gentamicin Conc. (,ug/mI)
FIG. 1. Linearregression lineforvaluesobtained by RIA versus EMITfor108clinicalspecimens (12 data pointssuperimposedonothers). See Table5for regression statistics.
on February 7, 2020 by guest
http://jcm.asm.org/
E12
10
0
c,
8
E6
c6
o2
m
0>
1~~~~~
co
o
Zoo
0
0 0 2
0
o cb
0
00
0 o
~0O
0 2 4 6 8 10 12
EMIT Gentamicin Conc. (ug/mI)
FIG. 2. Linear regressionlineforvaluesobtained by bioassay versusEMITfor99 clinicalspecimens (eight data points superimposedonothers).See Table
5for regression statistics.
12
E
S10r
c,' 8
*o 6
<r E
4, 0 Lw -~
2
Ï7~
o" ~
0 2 4 6 8 10 12
EMIT Gentamicin Conc. (i>g/ml)
FIG. 3. Linear regression lineforvalues obtained by FIA versus EMITfor 100clinical specimens (12
datapoints superimposedonothers).See Table 5for regression statistics.
the card test because we felt that only very
rarely would a clinical laboratory performing
gentamicinassaysknowthataparticular sample waspositiveforrheumatoidfactor.
Table 5 presents the results of the linear regressionanalysesofthe data inFig.1to4.The last column in this table shows the 95% confi-dence interval forasample determinedtobe6.0
kg/ml
by
the EMITprocedure,
whenassayed by
the otherprocedures (1). The numbersmay be
approximated bysubstituting the number6 for the EMIT value (x value) in the equations of the formy = mx + b andadding the quantity
twotimes the standarderrorof theestimate.
Table 6shows the reagent cost of measuring
a gentamicin serum concentration, either as a
single sample oras a single sample analyzed in abatch ofsixsamples,for theEMIT, FIA, and
cardtest.
14
12
a)
10
c,
r-8
E
2P 6
o
-4
0
2
o
o
o *9
ow 0°00
.0e-O
o oW
, o o
Xo
0 2 4 6 8 10 12 14
EMIT Gentamicin Conc. (,ug/mi)
FIG. 4. Linear regression lineforvaluesobtained by card test versus EMITfor109clinical specimens (16data pointssuperimposed on others). See Table 5
for regression statistics.
TABLE 5. Linearregressionanalysis ofdata inFig.
1, 2,3, and 4
Parameters ofequation" 95% confi-dence
in-Method rh SEE' terval for
y m x b predicted
value"
RIA RIA 1.1EMIT -0.1 0.95 0.82 4.8-8.1 value
Bioassay Bioassay 0.9EMIT +0.3 0.88 1.08 3.8-8.1 value
FIA FIA 1.1EMIT +0.3 0.90 1.12 4.9-9.4 value
Card Card 1.0EMIT -0.1 0.85 1.39 3.3-8.9 value
"Equationof
regression
line intheformy=mx +b. h r,Correlationcoefficient.'SEE,Standarderrorof theestimate.
" For asamplewithagentamicinconcentration of 6jug/ml asdeterminedbyEMIT.
TABLE 6. Costofmeasuringagentamicinserum
concentration'
Cost ofprocedure"
Method Single sample
Single sample inbatch of six samples
EMIT $20.46 $6.51
FIA $9.00 $2.75
Cardtest $4.00 $4.00
aIncludesonlycostof kits. Doesnotinclude cost of materials(e.g., testtubes) notprovided with kits, and does notinclude laborcost.
hCost based on
manufacturers'
recommendations
fornumber ofreplicatesofstandards, controls, and patient samples.1
on February 7, 2020 by guest
http://jcm.asm.org/
TABLE 7. Timesfor measuring gentamicin serum concentrationsofasingle sample and of a batch of
30samples
Time (min) forprocedure' Single sample Batchof30 samples Method
Elapsed Tech- Elasped
Technol-time nologisttime time ogisttirneEMIT 14 14 170b 170b
FIA 45 30 160 160
Card test 12 4 225 220
aTime required to complete indicated
procedure
plus a standard curve when applicable.
bThis timecanbereduced to2hwhen the CP5000
(Clinical Processor) is used.
Table 7 shows times required for the deter-mination of gentamicin serum concentrations either as asingle specimen or as abatch of 30 specimens.
Technical considerations. (i) Bioassay.
Advantages of the bioassay include the lack of need for sophisticated and costly instrumenta-tion, relative simplicity of performance, and broadapplicabilityto
virtually
anyantimicrobialagent.Thetechniques for
setting
up abioassayarefamiliar tomost
microbiologists.
Disadvan-tagesof the
bioassay
arethe timerequired from specimen receiptto result availability (a mini-mumof5h)andinterferenceby other antibiotics which eithercannotbeinactivatedorfor which resistantassayorganismsarenotavailable.(ii)
EMIT. Advantages of the EMITassay areextremely rapidturn-around time (about2min perspecimen after a standard curve has been
established), relative ease ofperformance, and
establishment ofastandardcurve
(usable
foratleast 24 h)
independently
ofspecimen assays.This last feature means that "stat" specimens
can easily be
handled,
and also meansthat,
should there beanyproblems
withtheestablish-mentofastandardcurve,therehas beenno
loss
of time, reagents, or
samples
from specimen processing. We also found thequality
control features of the EMIT to beespecially
useful.Specifically, precise
intervals aregiven
withinwhich certain calibrator
points
shouldfall,
and precise limits foracceptability
oftheresults ofthegentamicincontrolarestated.
Only
100pl
of serumisrequired
foraduplicate
determination.According
to thepackage
insert there is noknowninterferencewith theassay
by
antibiotics other thansisomicin ornetilmicin.A disadvan-tage of theprocedure
isthatoccasionally
there was some problemestablishing
anacceptable
standard curve; this
difficulty
wasusually
re-lated to the 16-,ug/ml calibrator and
generally
requiredonly re-assaying this calibrator one to
three times more to establish a satisfactory value. It wasourimpressionthat,after mostof the reagents in a given bottle (designated "A" or
"B" by the manufacturer)had been used up,it
wasoccasionallydifficult toestablishastandard
curve with these small volumes. This problem
mayhave been due to difficulties withaccurate
aspiration when onlysmall volumes ofreagent
remained in the bottles; we have since found
that the problem disappears when such smail volumes (with materialofthesame lotnumber)
arepooled.
(iii)
Cardtest.Advantages
ofthisprocedure
includeitslack of dependence on complex
equip-ment, lack of need for a standard curve, and rapidity.Accordingto information received from
the manufacturer, sisomicin does cross-react withthe assay; it is not known whether
netil-micinalsocross-reactswiththe assay. According tothe packageinsert there are no other known
cross-reacting antibiotics. Disadvantages we
foundwereasfollows. Sincethereis no
"batch-ing" with thistechnique, the time involved for running large numbers of specimens becomes
considerable. Also, as noted by the manufac-turer, anelevated rheumatoid factormaycause afalsely lowassayresult. Wefoundthis to be a
problem up to a gentamicinconcentration of 4
,ug/mIin onepatient sample, inwhich the level
was<1.0,g/mIbylatex agglutinationinhibition
butatleast4.0
,ug/ml
byallthe otherprocedures.In addition, as noted above, two other patient
samples gave inexplicably low results by this
procedure. In the recovery studies, in two sera
that were visually normal, an unusual type of
agglutinationoccurred in the lower dilutions of
all gentamicin concentrations. This agglutina-tion was presumed to be due to an interfering substance,since itdisappearedathigher serum
dilutions. The presence of this phenomenon
meant that in these two sera only the higher concentrations of gentamicin could be accu-ratelydetermined.Forthese two sera, gentami-cin concentrationsup to and including4,g/ml
wereall readasless than the lowest
concentra-tionmeasurable by each ofthe twocards used
ateach concentration. Therefore, forthe
sam-plestowhich 0, 1,2, and 4 ,ug ofgentamicin had
been added perml, valuesof 0, 0.5, 0.5, and 0.5 ,ug/ml, respectively, were assigned. Finally, there is sometimes difficulty in determiningthe
pre-ciseagglutinationinhibitionendpoint;this diffi-cultydecreases but does notdisappearwith fa-miliaritywith theprocedure. The
clarity
ofthe endpointswas also notedtovary somewhatbe-tweenlot numbers.
(iv) FIA. We were not able to discern any
on February 7, 2020 by guest
http://jcm.asm.org/
748 SELEPAK ET AL.
advantages to this procedure ascompared with the others evaluated. According to the package insert there is no known interference with the assay by antibiotics other than sisomicin or ne-tilmicin. Aprincipaldisadvantage ofthis proce-dure is that the standards on which the standard curve for agivenrun isbased are incorporated inthe same run asthesamples. Ifthe standard curveis notsatisfactory,the entire run,including
both standardsand samples, must be repeated
from the step in which standards and samples had been initially diluted in buffer. Duringthe course of the evaluationwe used two different fluorometers; with the firstinstrument, slightly
over one-third of the runshad to be repeated.
Performance was considerably better with the second instrument. Anotherdisadvantageis that manualmixing of dilutions isrequired, increas-ingthe risk ofspillage. Inaddition,quality con-trol is not well defined for this procedure. For
example, acceptableranges offluorescent units for the standardsare notspecified.
Also,
dupli-cate readings are supposed to be done, but no statement is made about maximum allowable
differences, if any, between
duplicate
readings. Finally,since the standard curveis constructed from point to point, rather thanusing
a "best-fit" line through ail thepoints, controls shouldprobablyberun at aconcentration in the inter-val between each two consecutive
standards,
rather thanusingonlytheone control,"a com-mercialcontrol,"recommended
by
the manufac-turer.DISCUSSION
We have previously evaluated a number of different RIAprocedures (2). Althoughmany of theseare accurateandprecise, they requirethe use of expensive and sophisticated equipment
and are not cost-effective for low-volume oper-ation, as astandard curvemustbedevelopedfor eachrun.
The particular FIA procedure we evaluated
suffers fromanumber of technicalproblems, as detailed above. We attempted to evaluate the FIAprocedureof BioRad(BioRadLaboratories, Richmond,Calif.) but hadtoabandonthatpart of the study because of numerous technical problems with the reagents and instrumenta-tion.
The card testappears useful forlaboratories performing only occasional assays for gentami-cin; however, the fact that some rheumatoid
factor-positive sera, as
weil
as some other sera for unknownreasons, will give falsely low values is a potential source ofdifficulty with this pro-cedure.After thisstudy had been completed, we were informed that others have foundthat theuseof serum which has been repeatedly frozen andthawed may produce results with the card test thatcorrelate poorly with the results of both EMIT and RIAprocedures, whereas the use of fresh serum or serum which has been frozen only onceproduces results that correlate better with these procedures (D. Bernstein, of Hynson, Westcott and Dunning, personal communica-tion). Itshould be noted that the patient samples we used in this study were all frozen twice; hence, it is possible that a better correlation of the card test results with both EMIT and RIA results might be obtained with fresh serum or serumfrozen and thawedonly once.
Thebioassay procedure appears less accurate than either RIA or EMIT andalso has an exces-sivelylong turn-around time.
The EMIT procedure has previously been compared with a bioassay (3) and an RIA (4) and was found to perform well in those studies. Because of its accuracy, precision, rapid turn-aroundtime, andrelativesimplicity of perform-ance as determined in ourstudy,weselected the EMIT procedure for routine use forgentamicin assays in ourlaboratory. Since the conclusion of thisstudy, Syva hasdevelopedacomputer pro-gram for use withthe assay; this program elim-inatesthe need for manual plotting of a standard curve and makesperformance of the assay pro-cedure still more simple. We should note that ourlaboratory has established the policy of di-luting any specimen and reassaying the dilution if theinitial determinationis -10 ig/ml.
Only small numbers of severely hemolyzed, icteric,andlipemic sera were tested, and all of these were sera to which gentamicin had been added in the laboratory. No patient sera that wereseverely hemolyzed, icteric, or lipemic were tested. Therefore, despite the fact that we de-tected noproblem with assaying hemolyzed sera, and no problem assaying icteric sera with the EMITprocedure, weare not prepared to state that one can accurately assay severely
hemo-lyzed oricteric sera by either the FIA or EMIT procedure. We do believe, however, that our data demonstrate that lipemic sera cannot be adequately assayed by the EMIT procedure.
With all oftheprocedures, it isimperativethat themanufacturer'sinstructions be followed pre-cisely.
There is considerable variation among the methods both inthe cost of reagents alone and intechnologist time, as noted in Tables 6 and 7. Actualreagent costs per test may differ greatly from these theoretical costs if standard curves or entire runs have to be repeated, as noted above. As withreagent costs per test, labor costs per testvary significantly with batch size for any
on February 7, 2020 by guest
http://jcm.asm.org/
procedure for which the number of standards and controls required is largelyindependent of the number ofclinical specimens inagivenrun.
It is important to note also that we have not
attemptedtoanalyze the effect ofcostsof
non-reagent materials, such as test tubes, nor the
effect of instrumentcosts onthe total costper test.
ACKNOWLEDGMENTS
We thank M. V.Ratnaparkhi and MarkH.Zweigforhelp with study design and data analysis,James E.Byrkitand Rita G. Minkerforhelpwith datahandling,andVirginiaM. Hol-combfor technical assistance.
LITERATURE CITED
1. Afifi, A. A.,andS. P. Azen.1972.Statisticalanalysis:a
computer oriented approach, p. 96. Academic Press,
New York.
2. Lantz,C.H., D. J. Lawrie,F. G. Witebsky, and J. D. MacLowry. 1980.Evaluation ofserumgentamicin
as-sayprocedures foraclinical microbiology laboratory. J.
Clin.Microbiol. 12:583-589.
3. O'Leary, T. D., R. M. Ratcliff,andT.D.Geary. 1980. Evaluation ofanenzymeimmunoassay forserum
gen-tamicin.Antimicrob.Agents Chemother.17:776-778.
4. Phaneuf, D.,E.Francke,and H. C.Neu.1980.Rapid, reproducible enzymeimmunoassay for gentamicin. J.
Clin. Microbiol.11:266-269.
5. Westgard, J. O., and M. R. Hunt.1973.Use and inter-pretation ofcommonstatisticaltestsin method-com-parison studies. Clin. Chem. 19:49-57.