Introduction into
Flow Cytometry
Stephanie Gurka, Andreas Hutloff, Timo Lischke,
• Use of fluorescence marked monoclonal antibodies
multi-parameter analysis (up to 18) for each individual cell
Flow Cytometry - FCM
FACS = Fluorescence Activated Cell Sorting
(also used for analytical cytometers)
• Analysis of the physical properties of single cells
or other biological particles
• Basic principle: a single cell passes through a flow cell
and is illuminated by a laser source
• Blood cells
• Tissue cells
• Algae
• Protozoa
• Chromosomes
• Yeast
Prerequisite:
single cell suspension
(Disaggregation: mechanical or enzymatic)
Instrument Overview
Sample
The Laser System
typically used monochromatic laser sources (nm)
• Gas laser systems which require complex air or water cooling
are more and more substituted with diode and solid-state lasers
Multi-Laser Systems
488 nm Laser
405 nm Laser
633 nm Laser
17 m s 55 m stime delay
Coherent lightsource (488 nm)
Optics - Forward Scatter Channel (FSC)
detect the amount of light scattered in the forward direction
(along the same axis that the laser light is traveling)
Forward Scatter Detector
particle passes through the focus
v ol ta ge time v ol ta ge v ol ta ge
10 9 8 7 6 5 4 3 2 1 0 C e ll c o u n t
is most influenced by the size of cells
FSC-Histogram
FSC-Threshold
• instrument is triggered when the signal exceeds a predefined threshold level
• Intensity of SSC is most influenced by the shape and optical homogeneity of cells
Optics - Side Scatter Channel (SSC)
• detect the amount of light scattered to the side
• the relative size
(FSC)
• the relative granularity
or complexity (SSC)
• the fluorescence intensity (FL1/2,
up to
FL X)
->
Which parameters can be measured?
Analysis of complex primary samples (heterogeneous cells), such as immune cells Detection of rare cell types
Using Fluorescence in Flow Cytometry
or cells
transfected with fluorescent proteins
• Nucleic acid fluorochromes
• Fluorochromes for membrane potential analysis or for ion flux (e.g. Ca2+) • Membrane label fluorochromes
What is Fluorescent Light ?
488 nm
FITC
530 nm
Incident Light Emitted Fluorescent Light
The fluorochrome absobs energy from the laser and releases the absorbed energy by:
a) Vibration and heat dissipation
b) Emission of photons of a longer wavelength =====> FLUORESCENCE
Stokes shift: energy difference between the wavelength of absorption and emission
Properties of Fluorochromes
Laser
SSC detector
photo multiplier tube (PMT)
FCS detector
photo diode
signal levels are high
Fluorescence detection
Fluorescence detector
(PMT3, PMT4 etc.)
FCS detector
photo diode
signal levels are high
SSC detector
photo multiplier tube (PMT)
• Specificity controlled by
the wavelength selectivity of optical filters and mirrors.
• Fluorescence emitted by each fluorochrome is usually
D = difference between positive and negative peak medians W = 2 x rSD (robust standard deviation)
Stain Index = D / W
Resolution Sensitivity:
W2
W1
D
Reagent Filter Stain Index
PE 585/40 356.3 Alexa 647 660/20 313.1 APC 660/20 279.2 PE-Cy7 780/60 278.5 PE-Cy5 695/40 222.1 PE-Alexa 610 610/20 80.4 Alexa 488 530/30 75.4 FITC 530/30 68.9 APC-Cy7 7801/60 42.2 Alexa 700 720/45 39.9 Pacific Blue 440/40 22.5 Clone RPA-T4
Fluorescence One Color Histogram
in conjunction with
fluorescence-based protein reporters (GFP)
Histogram:
Data Analyis
Light/Fluorescence Intensity ce ll co un t2D plot:
Intensities of 2 Light/Fluorescence Parameters plotted against each other
C
D
Gating - Statistics
Y % Y+X- % Y+X+ % A % B % CMFI =10
MFI =150
Quadrant Statistic
A laser beam of a single wavelength
is used to excite several fluorochromes
with different Stokes shifts and, thereby,
produce a variety of fluorescent colors.
Fluorescence dyes used for flow cytometry
http://www.bdbiosciences.com/spectra/
Fluorochrome excitation wavelength (nm) emission maximum (nm) LP Mirror (Canto2) BP Filter (Canto2) FITC, CFSE 488 525 502 530/30 PE 488 575 556 585/42 PI 488 620PerCP / PerCP-Cy5.5 / PE-Cy5.5 488 675 / 695 / 695 655 670LP
PE‑Cy7 488 767 735 780/60
AF647 / Cy5 / APC 633 665 / 667 / 660 685 660/20
A700 633 723 710 730/45
APC-Cy7 633 767 735 780/60
Pacific Blue / DAPI 405 451 / 460 - 450/50
Pacific Orange / DAPI 405 551 / 460 502 510/50
Fluorescence dyes used for flow cytometry
Fluorochrome
Company
FITC, PE, PerCP, APC, Cy5
Becton Dickinson
Alexa Fluor ___
Molecular Probes
(Invitrogen)
eFluor ___
eBioscience
V ___
BD Horizon
Pacific ___
___
Oregon ___
___
…
…
Fluorescence: points to consider
pH R e la tiv e in te n si ty Concentration fluorescein F lu o re se n ce in te nsi ty Time (sec) % in ita l f lu or e se n ce in te n si typH dependence
Quenching
Photobleaching
FITC
Photostability / Photobleaching
Brightness (high quantum yield)
Photostability (no bleaching)
pH insensitivity (stability of fluorescence emission)
Water solubility (little hydrophobic interactions)
Instrument compatibility (fit to excitation wavelenght)
Types of Fluorochromes
I)
Small dyes:
FITC, Cy5, AlexaFluor´s, eFluor´s,
accessory photosynthetic pigment of red (R-PE, PerCP) or bluegreen algae (APC). PE: 240-kDa protein with 34 phycoerythrobilin fluorochromes per molecule. APC: 105-kDa protein with 6 phycocyanobilin chromophores per molecule.
I)
g
I)
Large Protein dyes:
phycoerythrin, allophycocyanin, peridinin-chlorophyll-protein
FITCCoupling of fluorescent dyes to antibodies
I) Small dyes / haptens (FITC, Cy5, AFs, Dig, …)
II) Protein dyes (phycoerythrin, allophycocyanin)
Amine - thiol crosslinking
1) The bifunctional crosslinker Succinimidyl
trans-4-(maleimidylmethyl)cyclohexane-1-carboxylate (SMCC) reacts with amine groups of the fluorescent protein (R1)
thereby introducing a maleimide group
2) The mAb (R2) is partially reduced (with DTT) which yields free sulfhydryl groups
•
maximize signal:noise (pos/neg separation)
– This may occur at less than saturated staining
– This may or may not be the manufacturer’s recommended titer
•
Titer is affected by:
– Staining volume
– Number of cells
Antibody titration basics
1 10 100 10001 10 100 1000 10000 signal noise S:N ng antibody Intensity i n c re a si n g a m o u n t o f m A b
Basis of multicolor flow cytometry
A laser beam of a single wavelength is used to excite several fluorochromes with
different Stokes shifts and, thereby, produce a variety of fluorescent colors.
Two Color Experiment - 1 Laser
Filters collect 2 colors
positive population negative population
Fluorescence Compensation
mathematical subtraction
of the fluorescence due to one fluorochrome from the fluorescence due to another
Small errors
in compensation of a dim control can result
in large compensation errors with bright reagents
• Cells stained with a single fluorochrome-conjugated Ab (analyzed individually)
-> One control for each of the fluorochromes used in the experiment
+ Single control for every tandem conjugate
• Negative and positive populations are required (>10%)
• Use bright markers to setup proper compensation
Compensation controls
• manually (up to 4 FL) or automatic compensation (>4 FL)
• CompBeads
Specificity Controls
unstained / control: to detect "auto-fluorescence" or background staining
(monocytes/macrophages, cultured cells, or activated cells)
(to set up PMT-voltage for FSC, SSC and FL-channels)
secondary control: for indirect staining (Bio/SAv, Dig/anti-Dig) - secondary Ab alone
to control for non-specific binding of this polyclonal Ab
to dead or sticky cells.
specificity (experimental and gating) controls:
e.g.
Transfected cells: transfected / mock transfected / wt cell line,
Primary cells: WT / KO or activated / naive
Controls must undergo the same treatment (i.e., preparation, fixation)
as all the tubes in an experiment.
not necessary for (lineage) markers with clearly separated populations
Isotype Control:
Ab with the same Ig isotype as the test Ab,
specificity known to be irrelevant to the analyzed sample
-> whether observed fluorescence is NOT due to non-specific
(Fc receptors, dead cells) binding of the fluorescent Ab.
(one for each class of antibody used for staining,
with the same concentration and F/P ratio as Ab of interest)
FMO Control:
Fluorescence Minus One
leaving out the antibody of interest in the staining panel
-> fluorescence spillover of all other fluorochromes in channel of interest.
„Cold Block“:
Preincubation with an excess of unlabeled mAb
prior to addition of fluorophore labeled mAb (no wash between)
All events (cells) with fluorescence above the threshold set with the above controls
are considered positive for the marker of interest.
based on fluorescence height, fluorescence area and signal width.
Autofluorescence
• fluorescent signals generated by the cells themselves
(from pyridine and flavin nucleotides)• Present in all cells (viable and dead).
• Adds to fluorescence label of cells -> decreases fluorescence detection limit
• observed in all fluorescence channels,
General principle: Dye reacts with free amines.
Live cells (left) react with the fluorescent reactive dye only on their surface (weakly fluorescent cells).
Cells with compromised membranes (dead, right) react with the dye throughout their volume (brightly stained cells). In both cases, the excess reactive dye is washed away.
Dead cell exclusion
Dead cells, with compromised membrane integrity, tend to be sticky
-> bind all sorts of reagents unspecifically. -> exclude dead cells from analysis
• dye exclusion methods with DNA intercalating fluorochromes:
propidium iodide (PI), 7-amino-actinomycin D (7-AAD) or DAPI staining to positively identify dead cells
by their membrane permeability
Signal Separation: different fluorochromes
Isotype Control FITC PE PE-Cy7 13.68 26.84 75.29 a- hu CD4 conjugatesimportant for multicolor analysis: choice of which antibody to use with which fluorochrome (often many "correct" combinations possible)
consider: For any given mAb clone, the signal-to-noise ratio (positive/negative) can differ depending on the fluorochrome and instrument used
• Blocking of Fc receptors
with polyclonal Ig or specific mAb against Fc-Receptors (species specific!)
-> significantly reduces background staining, (usually not necessary with cell lines)
caution with indirect staining protocols and anti-rat-Ig
(use purified mouse-gamma globulin or mouse serum instead)
Specificity / Non-Specificity: Fc-Receptors
Ab bind to many cell types by their non-specific (Fc) ends.
Monocytes, BC and DC,
professionally bind many Ab through their Fc-receptors.
Sample preparation time, temperature, buffer (pH, salt concentration)
Lysis, digestion, fixation, permeabilisation, washing steps
instrument number and type of Lasers, Filters, Fluorescence Detectors
-> fluorochromes/ -combinations
antibody clone, affinity, monoclonal vs. polyclonal, Ig-Isotype, type of Fluorochrome, concentration, F/P ratio, (may differ from lot to lot)
Cell number and concentration: depending on the number of events to be analyzed
(due to cell loss during staining approx. 2 times more cells for staining than for analysis)
Cell concentration during staining:
Maximum density for staining is 5x107 cells/ml -> 50 μl staining volume for up to 2.5 Mio cells
20- 30 min at 4°C in the presence of NaN3 to be sure of minimizing capping / internalization/
staining procedure:
1) Choose brightest set of fluorochromes for particular instrument configuration.
1) Choose fluorochromes to minimize the potential for spectral overlap.
- high Compensation for adjacent channels, (FITC vs PE) - usually low Cross-beam compensation (blue vs red laser)
Exceptions: GFP and very bright FITC signals like CFSE (also excited by 405 nm detected in PacO channel);
PE-Cy5.5 / PerCP-Cy5.5 (excited by 633 nm detected in AF700 channel)
3) Reserve the brightest fluorochromes for “dim” antibodies, and vice versa.
- Highly expressed Antigens will be resolved with almost any fluorophore- Antigens expressed at lower density might require brighter flurophores to separate the positive cells adequately from the unlabeled cells
PacO < APC-Cy7 = PacB = FITC = AF700 = PerCP < PE-Cy7 < AF647 = PE = APC
1) Avoid spillover from bright cell populations
into detectors requiring high sensitivity for those populations.
- Strongly expressed Antigens impair the sensitivity/signal resolution of the adjacent channel - Preferentially, use this channel for Antigens which are not on the same cell as the Ag of interest
5) Take steps to avoid tandem dye degradation, and consider its impact upon results.
Set voltages: Decrease voltages for any detectors where events are off-scale Increase voltages for any detectors where low-end resolution is poor
Analytical Variables to consider
Data Acquisition, Analysis and Interpretation
Instrument setup and performance
• adjust and optimize PMT settings (optimal sensitivity)
Speed of analysis (high flow rate -> less intensity resolution)
• Run single-stained compensation controls for each experiment and set compensation • Run samples
• Run appropriate controls: Instrument setup controls (e.g., CompBeads) Gating controls (e.g., FMO)
Data Analysis / Interpretation appropriate number of acquired events to ensure reliable results gating strategy,
• Visually inspect compensation Create a template
containing dot plots of each color combination of the experiment,
then examine a fully stained sample for possible compensation problems • Check gating across all samples in the experiment.
Gates may need to be adjusted across donors and/or experimental runs.
-> Avoid classification errors and false conclusions
due to improper compensation and/or gating, or sample artifacts
Ask for interpreting the data, experiment and instrument setup
-> save time and labor
sensitivity and throughput rates enable detection of extremely rare populations and
events (frequencies < 10
-6),
■ Hematopoietic stem cells ■ Dendritic cells
■ Residual disease detection (tumor cell enumeration) ■ Antigen-specific T cells
■ Transient transfectants
Dump channels
use of an"dump channel"
significantly improves detection of rare cells or resolution of dim stains (e.g. CD11c). staining for an antigen not expressed by the cells of interest ("lineage negative„) -> exclusion of these cells for analysis
e.g. B220 for murine T cells,
CD3 + CD8 + Ly-6G/C + CD11b for B cells, CD3 + CD19 for dendritic cells.
-> also exclude cells binding antibodies unspecifically.
Preferentially,