• No results found

ISO_8466-1_1990_PDF_version_(en)

N/A
N/A
Protected

Academic year: 2021

Share "ISO_8466-1_1990_PDF_version_(en)"

Copied!
12
0
0

Loading.... (view fulltext now)

Full text

(1)

INTERNATIONAL

STANDARD

ISO

8466-1

First edition 1990-03-01

Water quality

- Calibration

and evaluation

of

analytical

methods

and estimation

of

Performance

characteristics

Part 1:

Statistical

Qualith de l’ea

evaluation of the linear calibration function

r- Etalonnage et Evaluation des mhthodes

d’analyse

et estimation

des caractkes de Performance

Partie 1: haha tion s ta tistique de Ia fonction lin&aire d’h talonnage

Reference number ISO 8466-1 : 1990 (EI Licensed to LAAM /MANUEL SOARES

ISO Store order #:913756/Downloaded:2008-05-21 Single user licence only, copying and networking prohibited

(2)

ISO 8466-1 : 1990 (EI

Contents

Page Foreword ... 1 Scope ... 2 Definitions.. ...

3

Symbols...~ .. 4 Performance ...

4.1 Choice of working

range

...

4.2 Calibration and characteristics of the method ... 4.3 Assessment ...

5

Example

...

5.1

Choice of working

range

...

5.2 Calibration and characteristics of the method ... 5.3 Evaluation ...

Annex A

Bibliography ... . . . Ill 1 1

2

3

3

4

5

6

6

7

7

8

0 ISO 1990

All rights reserved. No patt of this publication may be reproduced or utilized in any form or by any means, electronie or mechanical, including photocopying and microfilm, without Permission in writing from the publisher.

International Organization for Standardization Case postale 56 l CH-1211 Geneve 20 l Switzerland

Printed in Switzerland

ii

Licensed to LAAM /MANUEL SOARES

ISO Store order #:913756/Downloaded:2008-05-21 Single user licence only, copying and networking prohibited

(3)

ISO 8466-1 : 1990 (E)

Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national Standards bodies (ISO member bedies). The work of preparing International Standards is normally carried out through ISO technical committees. Esch member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International organizations, govern- mental and non-governmental, in liaison with ISO, also take patt in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.

Draft International Standards adopted by the technical committees are circulated to the member bodies for approval before their acceptance as International Standards by the ISO Council. They are approved in accordance with ISO procedures requiring at least 75 % approval by the member bodies voting.

International Standard ISO 8466-1 was prepared by Technical Committee ISO/TC 147, Water quaiity.

ISO 8466 consists of the following Parts, under the general title Water gua/ity - Calibration and evaluation of analytical methods and estimation of Performance charac teris tics :

- Part 7: Statisticaf evaluation of the linear calibration function - Part 2: Calibration strategy for non-linear calibration functions - Part 3: Method of Standard addition

- Part 4: Estimation basis method.

of Limit of detection and limit of determination of an analytical

Annex A of this part ISO 8466 is for information only.

Licensed to LAAM /MANUEL SOARES

ISO Store order #:913756/Downloaded:2008-05-21 Single user licence only, copying and networking prohibited

(4)

This page intentionally left blank

Licensed to LAAM /MANUEL SOARES

ISO Store order #:913756/Downloaded:2008-05-21 Single user licence only, copying and networking prohibited

(5)

INTERNATIONAL

STANDARD

ISO 8466-1 : 1990 (E)

Water

quality

- Calibration

and evaluation

of analytical

methods

and estimation

of Performance

characteristics

Part 1:

Statistical evaluation of the linear calibration function

1 Scope

This part of ISO 8466 describes the Steps to be taken in evaluating the statistical characteristics of the linear calibration function. lt is applicable to methods requiring a calibration. Fur- ther Parts of ttis lnternationa~ Standard will cover the deter- mina$.i!! ok IGr%t of detection and FEmit of determination, the effect of Ln$erferences and other performante chara,cteristics lt is intended especially for the evaluation of the pure analytical method and for the calculation of Performance characteristics of the calibration function.

In Order to derive comparable analytical results and as a basis for analytical quality control the calibration and evaluation of analytical methods have to be performed uniformly.

2 Definitions

For the purposes of this part of ISO 8466, the following defini- tions apply.

2.1

analytical

method:

An analytical method is composed of procedural, measuring, calibrating and evaluating instruc- tions (see figure 1).

Whereas the procedural and measuring instructions depend on the method, and are therefore the Object of standardization of the respective method, the calibrating and evaluating instruc- tions are valid for any analytical method requiring calibration.

2.2

calibrating

instruction

:

Describes the approach to determine the calibration function from information values, yi, obtained by measuring given Standard concentrations, xi. The slope of the calibration function,

b,

as a measure of sensitivity of the analytical method and the Standard deviation of the method, sXO, are figures of merit and characteristics which result from the calibration experiment.

The Standard deviation, sXO, allows the comparison of indepen- dent analytical methods.

For the user of the method, these characteristics present criteria for the internal laboratory quality control.

1 Originai Sample 1

_ , MeasUri!g ;X~~nstructi;

1

measuring instructions

1 Measur; value 1 ’

calibrating and evaluating instructions

r-YG&q

Figure 1 - The analytical

method

23

evaluating

instruction

:

A calculation guide for the computation of concentrations from the measured values by the use of the calibration function. Additionally, the confidence range permits an objective assessment of the imprecision of the analytical result[*J.

2.4

measured

values

: The concentration-dependent initial values (e.g. extinction) of a measuring System.

NOTE - Information value and measured volume are synonymous.

25

residual Standard deviation,

s : The residual Standard deviation describes the scatter of the information values about the calculated regression line. lt is a figure of merit, describing the precision of the calibration.

For the purpose of this Standard, the Standard deviation of the method means the Standard of deviation of the calibration pro- cedure.

1

Licensed to LAAM /MANUEL SOARES

ISO Store order #:913756/Downloaded:2008-05-21 Single user licence only, copying and networking prohibited

(6)

ISO 8466-1 : 1990 (EI

Yi

Information value of the Standard concentra- tion Xi calculated from the calibration function.

s2

i Variante of the information values for the analyses of Standard samples, having the con- centration Xi.

f

i

a

Degrees of freedom for the variance (f i = tli- 1).

Calculated blank (Ordinate bration straight line).

b

Sensitivity of the method (slope of bration line; coefficient of regression

x

Mean of the Standard concentrations resulting from the calibration experiment.

Y

Mean of the information values from the calibration experiment.

sY sY1

Residual Standard deviation. Residual Standard deviation regression calcula tion.

sY2 Residual Standard deviation

linear regression calculation. DS2 Differente of variances. Y n P x 2 tV;, l-a)

Information value of an analysed Sample. Number

Sample.

of replicates on the same analysed

values, Mean of information

replicates.

Concentration sf the analytical Sample, calculated from the i nformation value

Ym

Concentration of calculated from the

h

the analytical Sample, mean of the information values 7.

Tabled value of the t-distribution with

fl =

N

- 2 degrees of freedom and a confidence level of (1 - a) (f-factor of Student’s distribu- tion).

2.6

Standard

deviation

of the method

sXO: The ratio of the residual Standard deviation, s,,, to the sensitivity of the calibration function,

b.

lt is a figure of merit for the perfor- mance of the analytical method, and is valid within the working range (sec equation 13).

For the purpose of this Standard, the Standard deviation of the method means the Standard of deviation of the calibration pro-

cedure. calculation of the

2.7

coefficient

of Variation of the method,

VXO: The ratio of the Standard deviation of the method sXO to the appertaining mean, F, which is the centre of the working range.

intercept of the cali-

the cali- See also note to 2.5 and 2.6.

2.8

working

range

(of an analytical method): The interval, being experimentally established and statistically proved by the calibration of the method, between the lowest and highest quantity or mass concentration. The lowest possible limit of a

working range is the limit of detection of an analytical method. Yit resulting

2.9

homogeneity

of variances:

Homogeneity of variances of pooled data, such as those resulting from replicate analyses at different levels, is confirmed if these variances are not

significantly correlated to their appertaining concentrations. obtained by linear

2.10

sensitivity

of the analytical

method:

The slope of the calibration function of the complete analytical method, in- clusive of all procedural Steps, within the working range in question.

obtained by non-

2.11

measuring

Sample

(reaction Sample) : A Sample which tan be directly submitted to the measurement of the determi- nand. A measuring Sample is normally obtained by adding the required reagents to the analytical Sample. Obviously, measur- ing and analytical Sample are identical if no reagents have to be added to the analytical Sample.

resulting from

n

3 Symbols

Xi Concentration of the Ph Standard Sample.

i

Subscript of the concentration levels, where

i=

1,2, . . . . N.

N

Number of concentration levels (for this part of ISO 8466,

N

= IO).

Xl Concen tration of the Standard Sample at the lower level of the working range ( 1 st Standard Sample).

XI0 Concentration of the Standard Sample at the

upper level of the working range (10th Standard Sample).

FV;,f2,1

-a) Tabled value of the F-distribution (Fisher- Snedecor) with

fl

and

f2

degrees of freedom and a confidence level of (1 - a).

Yi,j jth information value for the concentration Xi.

%O Standard deviation of the method.

j

Subscript of the replicate j of level

i,

where j = 1,

2,

. . . .

ni.

V

XO Coefficient of Variation of the method.

VB (x, Confidence interval for the concentration .?.

% Number of replicates per level Xi.

Confidence interval of the mean Z?of the con- centration.

Mean of samples,

the information values Yi,j having the concentration Xi.

of Standard

2

Licensed to LAAM /MANUEL SOARES

ISO Store order #:913756/Downloaded:2008-05-21 Single user licence only, copying and networking prohibited

(7)

ISO 8466-1 :, 1990 IE)

4 Performance

4.1.2

Test for homogeneity

of the variances

4.1

Choice

of working

range

Esch calibration experiment is started with the choice of a preliminary working rangeL3].

The working range depends on

Both data sets of the concentrations x1 and xlo are used t~ calculate the variances sf and s$, as given in equation (1) :

10

t: (Y i j - El* t

S* i = j=l . . .

-1 (11

Ili

a) the practice-related objective of the calibration.

The working range shall cover, as far as possible, the application range for water, waste water, and sludge analy- sis. The most frequently expected Sample concentration should lie in the centre of the working range.

with the mean

10

c

Y- *

‘,J

.- j=l

Yj

= fori = 1 ori = 10 . . . (2) yli

b) feasibilities of technical realizability.

The measured values obtained must be linearly correlated to the concentrations. This requires that the measured values obtained near the lower limit of the working range tan be distinguished from the blanks of the method. The lower limit of the working range should therefore be equal to or greater than the limit of detection of the method. Dilution and concentrating Steps should be feasible without the risk of bias.

The variances are tested (F-test) for significant differentes at the limits of the working rangeL5t 6].

The test value PG is determined for the F-test from equation

(3).

PG = 2 forsyo > sf

1

.

. .

(3)

c) the variance of the information values must be indepen- dent of the concentration.

S*

PG = --$ for s: > sf0

10

The independence is verified by a statistical test on the

linearity16t *l. PG is compared with the tabled values of the F-distribution[? Decision :

4.1.1

Reparation

of the calibration

After establishing the preliminary working range, measured values of at least five (recommended N = 10) Standard samples are determined. The concentrations, xj, of these stan- dard samples shall be distributed equidistantly over the working range. In Order to check for the homogeneity of the variances, ten replicates of each of the lowest and the highest concentra- tions (x, and x,~) of the working range are determined. Ten in- formation values, )pi jl result from these series of measurements

a) If PG < Ffl. I f2. 099 the differente I I between the variances sf and SS is not significant.

b) If PG > q, ;

j-2;

0,99 the differente between the variances sf and s$ is significant.

If the differente between the variances is significant, the preliminary working range should be made smaller until the dif- (see table 1). ference between the variances is found to be random only.

Table 1 - Data sheet for the calibration

3

Licensed to LAAM /MANUEL SOARES

ISO Store order #:913756/Downloaded:2008-05-21 Single user licence only, copying and networking prohibited

(8)

lSO,8466-1 : 199O’E)

4.13

Test for linearity[*J ‘# *]

The easiest test for the linearity is the graphical representation of the calibration data with the calculated regression line. Any unlinearity is evident (see figure 2).

5 B c 0

06

I .- + E

05

1 8 z -

04

J

03

J \ , , , ., , ; . ,)c

12345-6789

Concentration

Figure 2 - Graphical

linearity

check

In the statistical linearity test the calibration data are used to calculate a linear calibration function as weil as a non-tin.ear calibration function, both with the residual Standard deviation

Syl Or 52-

The differente of the variances OS* is calculated from equation

(4) .

.

OS’ = (N -

2)

s;,

- (IV - 3) sy,

. . . (4)

Degrees of freedom : f = 1.

OS* and the variance of the non-linear calibration function sv2 are submitted to a F-test in Order to examine for significant dif- ferences.

The test value PG required for the F-test is calculated from equation (5) OS* PG =- . . .

(5)

S* Y* Decision :

a) If PG 4 F: The non-linear calibration function does not lead to a significantly better adjustment, e.g. the calibration function is linear.

b) If PG > F: The working range should be reduced as far as possible to receive a linear calibration function; otherwise the information values of analyzed samples must be evaluated using the non-linear calibration function.

4.2

Calibration

and characteristics

of the method

After the final working range is established, ten Standard samples are analyzed in accordance with all the Steps of the analytical method in Order to obtain ten (IV = 10) measured values Yj (see table 2).

The measurement against a blank is not allowed, since thereby valuable information on the magnitude of the blank will be lost. The comparison medium for zeroing the instrument is always, if possible, a pure solvent (e.g. pure water).

. Table 2 - Data sheet for simple linear regression

I i xj j x;

-vj

- i V2 Xj’uj l I 1 I

--y---l---

g:ppIIj

3 -- -t- ---- 4 --L--- , t ----t---l----l

.---j----~----~--q---j

5 6 ~-- .--~~ 7 --- --_.- ---+--- I I

--L---

1 t ---t-m

The ten data Sets, consisting of the va2ues of Xi and Yj, are sub- mitted to a linear regression analysis to obtain the coefficients a and b of the calibration function whsich describe the linear cor- relation between the concentration x as an independent variable, and the measured value y as a dependent variable. The calibration function as well as the characteristics of the method should result from data obtained from a working range x1 to xIo, as received from the measurement and not corrected for blanks. Generally, no blank value (concentration x = 0) is to be included in the calibration experiment and, consequently, in the least-squares fit of the regression.

The linear calibration function is given by equation (6)

Y =U+bX

. . .

(6)

The coefficients are obtained from equations (7) for sensitivity (slope of the calibration function) and (8) for the Ordinate intercept (calculated blank)

N c (Xj - XI ' (Yj - Y)

b

= p--_Ip--m_-

i=l N

. . .

(7)

Y (x; - XI*

M’ . I= 1

a

=

r--b%

. . . (8)

The coefficients provide an estimate of the true function, which is limited by the unavoidable procedural scatter. The precision of the estimate is quantified by the residual Standard deviation,

4

Licensed to LAAM /MANUEL SOARES

ISO Store order #:913756/Downloaded:2008-05-21 Single user licence only, copying and networking prohibited

(9)

ISO 8466-1 : 1990 (EI

sv, which is a measure of the scatter of the information values about the calibration line and is given by equation (9).

N

c

[Y

j - (a + bx,)l*

i=l (9)

N-2

4.3

Assessment

The concentration of an analyzed Sample is obtained a) from the measured valuey, to give ?

x

=-

Y

-a

. . . (IO)

b

or

b) from the mean of a series of replicates, 7, petformed on the same original Sample, to give ?

7-a

g=-

b

. . . (11)

As to the uncertainty of an analytical result, keep in mind that the analytical error is a combination of the uncertainty of the determination of the measured value, and the uncertainty of the estimation of the regression coeff icients[*!

a

/

/’

1

I

I

I

/

0

7

0

- 4

4

-

-

Fromsthe law of error propagation it follows that, for each value X, a confidence interval for the true value y exists whose limiting Points are on two hyperbolic paths bracketing the calibration line. Between these paths the true calibration func- tion tan be expected with a significance level of a (fl :

N

- 2, confidence level = 1 - a), determined by Student’s t-factor. The confidence intervals for analytical results, calculated from the calibration function, are given by the intersections with the respective hyperbolic paths in figure

3.

The estimation of the confidence intervals are given by equation (12) [71

7,,* =

% + VB (.%,

.a

Y -a

.2,,2 =

b

(jf

- Fl2

N

b*x

( Xi -

\ Z= 1 . / A NOTE - If i; = 1, F,,* = X1,2-

(12)

Equation (12) indicates that the confidence interval VB(?) brackets the true analytical value with a range governed by the statistical security of Student’s distribution. The magnitude of VB(g) is mainly determined by the number of replicates n^ and

I

I

I

I

I

vl3m

I

I

I

I

I

-

X

10

Working range

Figure 3 - Working

range x, to xlo, calibration

line with confidence

band and a Single analytical

result with its

appertaining

confidence

interval

5

Licensed to LAAM /MANUEL SOARES

ISO Store order #:913756/Downloaded:2008-05-21 Single user licence only, copying and networking prohibited

(10)

ISO 8466-1 : 1990 (E)

their results, the mean of the information values 7, as weil as the characteristics of the method, the residual Standard devi- ation s,,, and the sensitivity

b.

The quality of the analytical procedure increases therefore with increasing sensitivity and decreasing residual Standard deviation. The Standard deviation of the method sXO [see equation (1311 is the characteristic which allows the analyst to check the quality of his own work.

s

XO

=2

. . .

b

(13)

For the comparison of different standardized analytical methods, the coefficient of Variation of the method, expressed as a percentage, is given by equation (14)

v

xo =

s~xloo

. . . (14)

5 Example

The photometric determination of nitrite is used to demonstrate the calibration and the subsequent estimation of the statistical characteristics of the method and their influence on the final results of the evaluation.

5.1

Choice

of working

range

For the analysis of drinking water and surface water, a working range of 0,05 mg to 0,5 mg (NO;)/1 is appropriate.

5.1.1

Testing the hornogeneity

sf the variancesl)

According to the approach outlined in 4.1 .l, the variances ~2 of the information values obtained from the Standard concen: trations at the lower or upper limit of the working range respec-

’ tively, were determined (sec table 4).

The test value PG for the F-test is calculated from equation (3)

S2

PG=s=

13,54 x IO-6 = 2,9

1 4,67 x 10-6

Consulting the F-tablesL5] forfi = f2 = YE - 1 = 9 degrees of freedom for the variances ST and S$ gives

F(9,9; 0,991 = 5,35

The comparison of the calculated value PC with the tabled one indicates a random differente between the variances under examination. As the variances are homogeneous a simple regression analysis may be petformed.

Table 3 - Data sheet for the calibration

of NO;

i xi

mgll

Y*

l,l

Yi,2 Yi,3 Y- 1,4 Y- lt5 ’

Yfj

i, Y- 1,7 , i R,a , &,!3 j Yi, 10

' 1 0,05 0,140 0,143 0,143 0,146 0,144 0,145 0,344 0,146 0,145 0,148 2 0,lO 0,281 1 3 0,15 0,405 j I 4 0,20 0,535 / I 5 0,25 0,662 ) I 6 0,30 0,789 1 I 7 0,35 0,916 1 1 8 0,40 1,058 9 0,45 1,173 10 =N 0,50 1,303 1,302 1,300 1,304 1,300 1,296 1,295 1,301 1,296 1,306

Table 4 - Data sheet for the analysis sf variance.

Object:

nitrite

I

i xi Y. Yi,2 Yi,3 Y* Yi, 5 yi,6 Y* yi,8 Yi, 9 Yi, 10 s?

mgll

Ex% Ext.*) Ext.*) Ex;.:) Ext.*) Ext. *) EX:,’ ) Ext.“, Ext. *) Ext. *) mg& 12 1 0,05 0,140 0,143 0,143 0,146 0,144 0,145 0,144 0,146 0,145 0,148 4,67. IO-6 I 10 1 0,50 1 1,303 / 1,302 1 1,300 1 1,304 1 1,300 / 1,296 j 1,295 1 1,301 1 1,296 1 1,306 11356 . lW6 1

I

*c) Ext. : (extinction) I

1) Forthe sake oftransparency, alldimensions have intentionally been omitted in allequations, withoutambiguity. The dimensions are finallyadded to the result.

6

Licensed to LAAM /MANUEL SOARES

ISO Store order #:913756/Downloaded:2008-05-21 Single user licence only, copying and networking prohibited

(11)

5.1.2

Testing linearity

The Standard deviation of the method is calculated from equation ( 13)

A non-linear regression function may be derived [*l as

0,005 2 s =

sy

-

-

xo

b-

2,575 2 = 0,002 0 Y = 0,013 5 + 2,62 x - 0,081 8 ~2, giving a residual stan-

dard deviation of sY2 = 0,005 2 mg/l.

The coefficient of Variation of the method, expressed as

a

percentage, is given by equation (14) The residual Standard deviations of the linear and the non-linear

calibration function, sY, and su2, are compared:

Sxo

0,002o x 100

V xo = x x loo=

0,275

= 0,73

sYl = 0,005 2 mg/1 (see 5.2 for calculation procedure) sY* = 0,005 2 mg/l.

5.3

Evaluation

As both residual Standard deviations are equal, the differente of the variances DS* [see equation (411 does not need to be calculated. The non-linear calibration function does not lead to a significantly better adjustment, e.g. the calibration function is linear.

5.3.1

Single determination

The analysis of an unknown Sample, performed in the same way as the analysis of the Standards, gave an information value jk 0,641 (extinction). The analytical result was obtained using equation (12) with a confidence interva195 %, t(8; 0,951 = 2,31

5.2

Calibration

and characteristics

of the method

%,2

=

0,641

- 0,018

+

2,575 Since the prerequisites for the Performance of a simple linear

regression are fulfilled, the calibration function and the characteristics of the method tan be calculated using equations (71, (8) and (13). The results are shown in table 5.

0,002 0 2,31 (0,641 - 0,726 2)*

+ x - + - +

(2,575)* x 0,206 25 The slope,

b,

as a measure for the sensitivity, is calculated from

equation (7) = (0,242 + 0,005) mg/1 N c (Xi - 33 ’ (yi - jd i=l = N c (X i- 32 i=l

Thus the true value of the concentration tan be expected within the range 0,237 < x < 0,247 mg/l, with a confidence level of 0,95.

2,575

b =

5.3.2

Replicate

analysis

For three replicate determinations, the analytical method gave the information values 0,641; 0,631 and 0,633.

The Ordinate intercept, a, (calculated blank) is calculated from equation (8)

Calculation of the analytical result: a = v--

b X = 0,726 2 - 2

1

575 2 x 0

1

275

=

0

I

018

Ext .

%,2 =

0,635 - 0,018 + 2,575 The residual Standard deviation, sY, is calculated from

equation (9). (0,635 - 0,726 2)* + (2,575)* x 0,206 25

-

(Y i - yi)*

sy =

N-2

= 0,005 2 [Ext.l

= (0,240

+ 0,003)

mg/l

Thus the true value of the concentration tan be expected within the range 0,237 < x < 0,243 mg/I, with a confidence level of 0,95.

The equation for the straight line is given by equation (4) Y = 0,018 + 2,575 2 x

Table 5 - Data sheet for the regression

4 i xi mg0 d 1 V* 9 0,45 1,173 -- 10 0,50 1,303

1 1 1 $i 1 ,66,

I

6 0,30 0,789

r T--t--

i=l

c I 2f75

/

l

I I l 7,262 I 7 0,35 0,916 - 8 0,40 1,058 4 I 0,20 1 0,535 X = 0,275 mg /I 7 = 0,726 (Extinction)

7

Licensed to LAAM /MANUEL SOARES

ISO Store order #:913756/Downloaded:2008-05-21 Single user licence only, copying and networking prohibited

(12)

rSO8466-1 : 1990 (El

111

[Zl

[31

141

151

BI

[71

BI

191

no1

VONDERHEID, C., DAMMAN, V.

, DÜRR, w.,

FUNK, W. and KRUTZ, H., Statistical methods a nd Performance characteristics for the assessment and comparison of analytical procedures. An approach to standardization,

Vom Wasser 57, (1981),

pp. 59-74.

Annex

A

(informative)

Bibliography

MANDEL, J., The statistical analysis of experimental data,

lnterscience PubL,

J. Wiley & Sons, (1964), New York. GOTTSCHALK, G., Standardization of quantitative analytical procedures,

Z Ana!. Chem. 275, (1975),

pp. 1-10.

FRANKE, J.P., dE

ZEEW,

R.A. and HAKKERT, R., Evaluation and optimization of the Standard addition method for atomic absorp- tion spectrometry and anodic stripping voltammetry,

AnaL Chem. 50, (1978),

pp. 134-1380.

GRAF, U., HENNING, H.J. and STANGE, K., Formulae and tables of mathematical statistics, 2nd Edition,

Springer Verlag, (1966),

Berlin, Heidelberg, New York.

SACHS, L., Methods for

statistkaI evaluation, 3rd Edition,

Springer Vdag, (1971), Berlin, Heidelberg, New York.

BROWNLEE, K.A., Statistical theory and methodology in science and

enghering,

J. Wiley & Sens, (1965), New York. DOERFFEL, K., Statistics in Chemical analysis,

VEB- Verlag für die Grundstoffindustrie,

(1966), Leipzig.

WAGNER, R., Evaluation of BOD-dilution series - Presentation of a Computer program,

Vom Wasser 58, (1982),

pp. 231-255. Commissariat a I’energie atomique,

Statistique appliquee & l’exploitation

des mesures,

tome 2, Masson, (1978), pp. 345-379.

UDC 543.3 : 53.088

Descriptors: water, quality,

tests,

Chemical analysis, calibration, statistical analysis. Price based on 8 pages

8

Licensed to LAAM /MANUEL SOARES

ISO Store order #:913756/Downloaded:2008-05-21 Single user licence only, copying and networking prohibited

References

Related documents

• Follow up with your employer each reporting period to ensure your hours are reported on a regular basis?. • Discuss your progress with

4.1 The Select Committee is asked to consider the proposed development of the Customer Service Function, the recommended service delivery option and the investment required8. It

19% serve a county. Fourteen per cent of the centers provide service for adjoining states in addition to the states in which they are located; usually these adjoining states have

So, think of all people all over the world who buy sugared breakfast cereals that make health claims, for themselves, and their families including children.. Here is what

Descriptions of other animal prion diseases include Bovine spongiform encephalopathy (BSE) or mad cow disease in cattle, chronic wasting disease (CWD) in

National Conference on Technical Vocational Education, Training and Skills Development: A Roadmap for Empowerment (Dec. 2008): Ministry of Human Resource Development, Department

It was possible to improve the production process by applying the simulation technique, which shows that the general objective was achieved. Also, it was possible to

Since, financial development of India as of late is driven principally by administrations division and inside administrations segment by data innovation (IT) and