JOURNAL MICROBIOLOGY,Aug. 1982, 0095-1137/82/080271-05$02.00/0
Comparison
of
a
Fluorometric Method with Radial
Immunodiffusion Assays for
Determination of Complement
Components C3 and C4
MARGOTKOELLEANDWILLIAM R. BARTHOLOMEW*
Division ofClinical Microbiology andImmunology,Erie CountyLaboratory, Erie CountyMedical Center,
Buffalo,New York 14215, and Department of Microbiology, School of Medicine, andDepartmentofMedical
Technology, School ofHealth RelatedProfessions, State University ofNew YorkatBuffalo, Buffalo, New
York14214
Received 23 November1981/Accepted 29 April1982
Measurementsof patientserumcomplementcomponentsC3 and C4areuseful
indicators ofcomplement consumption in immune complex diseases. A fluoro-metric quantitative immunofluorescence system was evaluated in terms of
mea-suring these complementcomponents,and the resultswere compared with those
of radial immunodiffusionassays.Forcomparison of thetwosystems,232patient
sera wereevaluated for C3, and 202 specimensweretested for C4.Analysis of the
data by linearregression indicatedaproportional difference between the methods.
C3 andC4 concentrations measuredby the fluorometric method werelowerthan
those measured by radial immunodiffusion, especially concentrations exceeding thenormalranges.Indetecting lower concentrations (<120mg/dlfor C3 and <25
mg/dlforC4), thetwomethodsshowed betteragreement. Eachassaysystemwas
reproducible and could be usedtoevaluate changes thatoccurinconcentrations ofcomplementcomponents during therapeutic treatment. However, the ease in processingalarge volume ofspecimens and the short time neededtocomplete the
assay areadvantages that make the fluorometric methodmoresuitable thanradial
immunodiffusion foruseinalarge clinical laboratory.
Concentrations of
complement
componentsC3 and C4 in serum are used as indicators of
complement
consumption
in immune complex diseases. Concentrations of C3 and C4 havebeenofparticular valueasanaid in thediagnosis and management of
systemic
lupuserythmato-sis.Monitoring the concentration of theseserum
complement
componentshas beenusedto deter-mine the effectiveness oftherapeutic treatmentof this disease (1,
7).
The quantitation ofserumcomplement com-ponentsis
usually
performed by
radial immuno-diffusion (RID). RID assays are based on anantigen-antibody precipitation reaction. Antigen
is added to wellscut in anagarose matrix which
contains anoptimal concentration of
monospe-cific antiserum. In the method of Fahey and
McKelvey(6), theprecipitin ringsaremeasured within aspecifictimeperiod;complete diffusion
is notnecessary. Theprotein concentrations of
the specimens are determined by interpolation fromthe standard curve onwhichthediameters oftheprecipitinringsareplottedagainstthelog
of concentrations ofthe standards. One ofthe more recently developed test systems for the
quantitation of complement components and otherserum
proteins
isasolid-phase fluoromet-ricassay (3, 9).Thepurposeof this studywastoevaluatethe
quantitative immunofluorescence (QIF) system
forthe
quantitation
ofcomplement
componentsC3 and C4 and tocompare this
technique
with traditionalassays.TheQIFassayis basedonthe reaction ofserumcomplementcomponentC3orC4 with fluorescein
isothiocyanate
(FITC)-la-beled
monospecific
antibodies in a buffer solu-tion. After a 20-minincubation,
the unreacted antibodiesareadsorbedontosmalldisks coated withC3 orC4. The disks are 6mmin diameterand attached to the lower portion of plastic
probes, whichcan easilybeinsertedinto 75-by
100-mm test tubes.After adsorptionofthe
resid-ual labeledantibodies,theprobesarewashedin
buffer,and the amount of fluorescence
emanat-ing from the disk is determined by insertion of
theprobe intothe specifically designed
fluorom-eter. The amount offluorescence, measured in
fluorescentsignal units (FSU), is inversely
pro-portional to the concentration of the
comple-mentcomponent in the serum.
271
on February 7, 2020 by guest
http://jcm.asm.org/
The precision of the assay systems, as wellas
the total time for performing each assay, is discussed in this study.
MATERIALS AND METHODS
Serum samples. We evaluated the QIF system by
testing 232 patient serum samples for complement
component C3and 202samplesforC4, usingboth the
fluorometric and the RID methods. The sera were
randomly selected from hospital and clinic samples
submittedforcomplement assays. Many ofthe
speci-mens from clinic patients were from systemic lupus
erythmatosispatients who werebeingmonitored.
Normal rangesfor complement componentsC3and C4 wereestablished bytheQIFsystem,usingseraof
102 healthy adults. These sera were donated by
healthylaboratory personnel.
Bloodspecimenswereheldatroomtemperature for
1 hafterbeingdrawn toallowaclottoformand then
heldat4°C for1 h sothat theclotwould retract. At this
point, thespecimens wereimmediately centrifugedat
500 xg for 10min,divided intoportions,and stored at
-70°C until assayed.
Fluorometric analysis. (i) Instrumentation. The
FIAXfluorometer(InternationalDiagnostic
Technolo-gy,SantaClara,Calif.)wastheinstrumentusedforthe
QIF assays. Itisequippedwith atungstenlightsource.
Theexcitationwavelengthis 475nm, and the emission
is readat 540 nm. Aspecifically designed stage holds
theplasticprobes, which haveattached, cellulose-like
disksonwhichthereactions takeplace. A
microcom-puter interfaced with the fluorometer was used to
obtain a standard curve and calculate the values for
each specimen. A horizontal shaker was utilized to
agitate the tubes during the washing and various
incubation periods. It is designed to hold test tube
racks at a450 angleand operatesat200oscillationsper min at a distance of2.5 cm. All dilutions were
per-formed with a pipette diluter (model 1500; Cavro
Instruments, Los Altos, Calif.). This instrument can
beadjusted from5to50,ul for thedeliveryofreagents
orserumsamples,and thediluentmaybesetfrom0.25 to 1.0ml.
(ii) Reagents and method. Testkits forthe
determi-nations of C3 (3C/431A) and C4 wereobtained from
International Diagnostic Technology. The procedure
described by the manufacturer was followed without
modification.
To startthecomplementassayforC3,we mixed50 ,ul of a 1:101 dilution of standard or patient serum with 50 u1l ofFITC-labeled monospecific antiserum in 500
p.l ofphosphate-buffered saline (pH 7.4). Themixture
was incubated for 20 min. The tubes were placed on the horizontal shaker at the beginningof the
incuba-tion period. During the same time, thereaction
sur-faces of the disks were coated with the C3complement component by insertion of the plastic probes with attached disks into 500 p.l of the component reagent supplied with the kit. After 10 min, the probes were removed and placed into a second row of tubes
containing phosphate-buffered saline (pH 7.4) for a
wash. The disks were then ready to be used in the
assay and were placed into the incubated
serum-antiserum mixture. Residual, monospecific,
FITC-la-beled antibodies to C3 were absorbed by the C3-coated disks during a 20-min incubation. The disks
were then washed in phosphate-buffered saline (pH
7.7), and the amount offluorescence attributable to residual labeled antibody was determined with the
fluorometer;theintensitywasexpressedinFSU.The standardcurve wasbasedonthe FSU of four standard sera suppliedin thekit, FSU being inversely propor-tionaltothe concentrationsof C3in thestandardsera. The levels of C3 in patient serum samples were
determinedby interpolation from the standardcurve.
For the C4 assay, 20 p.l of a 1:13.5 dilution of standardorpatientserumwasmixed with 20p.l ofthe
FITC-labeled monospecific antiserum in 500 p.l of
phosphate-buffered saline (pH 7.4). The incubation
periods,aswellastheprocedureforcoatingthedisks withC4,werethesameasthose used for the C3 assay.
The determination ofFSU and standard curves were
similartothe determinationofC3.
RID. Test kits wereobtained from Meloy
Labora-tories, Inc., Springfield, Va. Theprocedure employed
wasthatgiven onthe insertincluded in each kit. The
incubation time for the C3determinationswas 18 hat
4°C, and that for the C4determinations was 22 h at 37°C. The antiserum used in the agarose for the measurement of serum C3byRID wasshown to react
with
P3IC
and ,1A. The diameters of the precipitinrings were measured with a vernier caliper and a
viewer-magnifier (National Instrument Laboratories,
Inc., Rockville, Md.). To obtainastandard curve,we
plottedtheconcentration ofeach serumstandard (log
[ordinate]) againstthediameterofeachprecipitinring
(abscissa) on semilog paper. For each new lot, we
establisheda standard curveby plotting threevalues
for each control serum. Each time the assay was
performed, high and low standards were included on
each plate and checked against the values of the
standard curve toensurethat the kitwasfunctioning
satisfactorily. Patientand standardsera wereassayed
in the same manner. After diffusion, the precipitin rings were measured, and the concentration of
com-plementcomponentin each patientserum was
deter-minedby interpolation from the standardcurve.
RESULTS
Complement concentrations found in sera of healthy individuals. The normal ranges for the complement components C3 and C4, as
mea-sured by thefluorometric system,werebased on the analysis of 102 sera. Thefrequency distribu-tion curves, plotted in increments of 0.5 stan-dard deviation, showed a bimodal distribution. The Kolmogorov-Smirnov test, used to deter-mine whether the observed data followed a
normaldistribution, indicated no significant de-viation for C3 (P>0.05) butdid show significant deviation for C4 (0.01 < P <0.05). On the basis of the mean ± two standard deviations, we calculated the normal range for C3 to be 72 to 170mg/dl (mean, 121 mg/dl), and thatfor C4 to be 12 to 46mg/dl (mean, 29mg/dl). The range for C4, calculated fromnonparametric data (exclud-ing 2.5% of the highestvalues and 2.5% of the lowest), ranged from 13 to45 mg/dl.
Comparison of the fluorometric system with
RID. The fluorometric system was compared
J.CLIN. MICROBIOL.
on February 7, 2020 by guest
http://jcm.asm.org/
W
0
° 2.2 'C
1-.
CN E 1.8
V
0
-4) IA
.)_
0 U-0 IL 0.6
.a2
00
*.0*
* 0 .
* 00
s_
.0* 0.0
0.2 1.0 14 1.8
RID
(mg/dl
OP2)
FIG. 1. Quantitation of C3: comparison of QIF andRID. Thestatisticalanalysis is shown in Table1.Line of
best fit:y= 45.7 +0.421x;n = 232; Sy.x= 19.2. Thehighest pointnotplottedbutincluded incalculationswas
137,368 (y, x).
with RIDin terms ofmeasuring C3 and C4. A
total of 232 serum samples, 91 ofwhich were
from
patients
withsystemic
lupus erthymatosis,were analyzed for C3. For the C4 evaluation, 202 serum samples were analyzed, 91 ofwhich
werefrom patients withsystemic lupus
erythma-tosis.
The results forC3 obtained bythe
fluoromet-ric assay wereplotted againstRID results (Fig. 1),using theRIDassay as thereference method. The data for C4 are
plotted
in Fig. 2. Thestatistical analyses of C3 and C4
by
linearre-gressionare summarized in Table 1.
For C3, the range of concentrations plotted varied from 25 to 368 mg/dl. The slope of the
regression line for C3 was 0.421, and the
inter-cept was45.7.
Considering
thataslope of1.0, anintercept of0, and minimal random error
indi-cate anideal bivariate relationship, the data for C3 show a
proportional
as well as a constantdifference
between the twoassay systems. Re-sults ofQIF were lower than those of RID. The random error was 19.2 mg/dl, asdefined by the standard deviation for distribution of the ob-servedy values above and below theregressionline
(syJ.)
The coefficient of correlation was 0.788.ForC4,theplottedvaluesrangedfrom 4 to 73
mg/dl(Fig. 2).Theline ofbest fit was y = 5.3 +
0.694x. Theslope of0.694indicates
proportional
difference,
and the intercept (5.3) relates tosomedegree ofconstant difference in C4
mea-surement between the two assay systems. At
levels of25 mg/dl, both methods showed good
agreement. The random error, sy., was6.2 mg/ dl; the correlation coefficient, which relates to
the fit ofthe values to the regression line, was
0.855. Three outliers were excluded from the
data (4
syJ.)
Considering the large number ofdata, the elimination of obvious outliers is
ap-propriate (4).
Precision of QIF and RID assays. (i) RID plates. The run-to-run precision of the RID
method for quantitating C3 and C4 was deter-minedby inclusion of high and low standards for
C3andC4 in eachassay.Thesestandards hadto
give ring diameters that fell within a predeter-minedrange tobe considered ashaving
accept-able
sensitivity
andprecision.Thecoefficientofvariation (CV) wascalculated from the
concen-trations (mg/dl) obtained for the high and low standards for C3 and C4. The CVfor C3, deter-minedby
using
thehigh standard in 61determi-nations, was 8.24%. The CV forC3, based on resultsof the low standard in 59determinations,
was4.27%. In termsoftherun-to-runprecision
of the C4assay,the CV for thehighstandard(58 determinations)was6.74%,and thatforthelow standard (56determinations) was8.76%.
(ii) QIF assay. The within-runprecision ofthe C3 assay was determined by evaluation of a
standard serumeight timeswithin a test run. The CV was 3.1%. In terms of the C4 within-run precision, theCV was7.4%, determined on the
on February 7, 2020 by guest
http://jcm.asm.org/
60 'a
E
%- 50
0
.-4
E 3o
0
.0 0
, 5~~~~~
~0~
a....ogo0,'0 ' '0^.'e,2
I10 /: " -t
I.S. . . .
l0 20 30 40 50 60 70
RID
(mg/di)
FIG. 2. Quantitation of C4: comparison ofQIF and RID. The statistical analysis is shown inTable 1.Line of best fit: y =5.3 + 0.694x; n= 202;sY.X =6.2. Three outlierswereexcluded from data(>4
syJ.)
basisof 10 determinations. Therun-to-run
preci-sion for the C3 and C4 determinations was
determined with a freshly thawed portion ofa
serumpool (stored at -70°C) in each run. The
CVswerecalculated from atotalof 38
determi-nationsfor each C3 and C4. the CVfor C3 was
9.3%, and that for C4was9.7%.
TimecomparisonofQIFandRID assays. The
quantitation of C3 and C4in 10 serum samples
by the QIF system was accomplished in 3 h,
includingdilution of thesera,performanceof the
assay (which includes four standards and two
dilutionsof the control serumpertestrun),and
readingand reportingtheresults.
The RID assay of C3 and C4 in 10 serum
specimens tookapproximately 1 h. After 18hof
incubation for C3 and22h forC4,theprecipitin
ringsareread,theconcentrationsarecalculated,
and the results are reported. The latter three
tasks took a total of 2 h; thus, the total time
needed to quantitate C3 and C4 in 10 serum
specimens byRIDwasapproximately3 h.
How-ever, the time from receipt of the specimen to
submission ofa report wouldbe aminimum of
21to25 h forRID,whereas the total time for the
QIF systemwould be approximately 3 h.
DISCUSSION
TheQIF and RID systems werecompared in
termsof thequantitationofcomplement compo-nents C3 and C4. A total of 232 clinical
speci-mens were analyzed for C3,and 202 specimens were analyzed for C4. Analysis of the data by
linear regression indicated a significant
propor-tional difference forboth C3 and C4 (slopes of
0.421 and 0.694, respectively) and some degree
ofconstantdifference.QIFvaluesfor C3and C4
tended to be lower than RID values. This was
especially true for C3 in concentrations above 160 mg/dl: the concentrations determined by
RID were almost twofold greater than those
determinedby QIF.The correlation coefficients
were 0.788 for C3 and 0.855 for C4, which
TABLE 1. Quantitation ofcomplementcomponents C3 and C4: statistical comparison of theQIFand RID assaysbylinearregression analysis
Component No.ofspecimens Concn QIFy RIDx Correlation
ComPonent
tested(mg/dl)
(mg/dl)
(mg/dl)
Slope Intercept Sy coefficientC3 232 25-368 114.6 163.6 0.421 45.7 19.2 0.788
C4 202 4-73 25.7 29.2 0.694 5.3 6.2 0.855
0 0 0
:
on February 7, 2020 by guest
http://jcm.asm.org/
COMPLEMENT COMPONENT QUANTITATION 275
indicates that the correlations were 62.1 and
73.1%, respectively.
For the determination of serum complement component C3, RID and QIF are designed to measure serum fractions 131C and 13A, 31A beingabreakdownproduct of
P1C.
Of the three antigenic determinants present onP1C,
only the A determinant is shared by13A.
It has beenfound that aged serum samples deplete more monospecific antibodies to the A determinant than are depleted by the amount of
PjC
in fresh serum(5,10).Davisetal. (5) have reported that when serum samples are held at 37°C for up to 7days, the ratio of the A determinant in aged versus fresh sera is 1.56. These researchers
found no appreciable difference when serum
samples were held at 5°C. An autoanalyzer system was used for the determination of
j1C
and
P1A.
The fact that the serum specimens assayedinourstudy were stored at -70°C untilassayed rulesoutthe possibility ofany
apprecia-ble breakdown of
P13C
fromthe time the speci-men was received to the time the assay wasperformed.
Bruver and Salkie (2) compared C3 values
obtained byusingtheHyland model PDQ
neph-elometerand the RIDmethod of Mancini et al.
(8). Forthe RIDmethod, the researcherschose ashortened
incubation
of16hat room tempera-ture, after which the diffusion rings weremea-sured.Twokitswereused for the comparison of nephelometry and RID. In the analysis of the
data by linear regression, theresearchers
report-ed higher values obtained by nephelometric analysis than by RID. However, we observed that values forC3 obtained by the fluorometric method werelowerthanthose obtained by RID.
The correlation coefficient reported by Bruver
and Salkie (nephelometric assay versus RID) was similar to the value we obtained for the
fluorometric assay versus RID, 0.79 (Fig. 1).
Whenmonitoring
patients
for C3 andC4,
boththefluorometric andRID assays werecapable of detecting a change in complement component
concentration.
In terms ofthe within-run
precision
ofQIF,
the CV for C3 was3.1%,
and that for C4 was7.4%. Values for within-run precision ofRID have beenreported by Stevensetal.(9).TheCV
for the run-to-run precision ofthe RIDassays, based on the concentrations (milligrams per
deciliter)
ofhighandlowstandards, rangedfrom 4.27to8.24%
forC3 and from 6.74 to8.76%
for C4. Intermsof the run-to-run reproducibilityofQIF, theCVs, based on a serumpool,were9.3 and
9.7%
for C3 andC4, respectively.
TheQIF methodis
relatively
easytoperform,
and valuesmay bereported withinthesame
day
that the specimen isreceived. TheQIF
systemprovidesahigh degree of
precision
in the deter-mination ofserumcomplement components C3andC4.
Both methods are suitable for
quantitating
patientserumcomplementcomponents in clini-cal laboratories. The QIF system,
however,
provides a reliable alternative method for thequantitation of C3 and C4. The results may be
reported on the same day that the specimen is received, whereas the RID method
requires
aminimum of 2 days before the results can be reported.
ACKNOWLEDGMENT
Wethank Helen Lees, Department of MedicalTechnology, State University of New York at Buffalo, Buffalo, for her guidance with the statisticalanalysis.
LITERATURE CITED
1. Appel, A. E., L. B. Sablay, R. A. Golden, P. Barland, A.I. Grayzel,and N. Bank. 1978.Theeffect of normaliza-tion of serum complement and anti-DNAantibodyonthe courseof lupusnephritis.Am.J. Med. 64:274-283. 2. Bruver, R. M., and M. L. Salkie. 1978. Acomparative
study of three approachestotheroutinequantitationof human serumproteins.Clin. Biochem. 11:112-116. 3. Casavant, C. H., A.C. Hart, and D. P. Stites. 1979.
Comparison ofasemiautomated fluorescent immunoas-say system and indirectimmunofluorescence for detection ofantinuclear antibodies in humanserum.J.Clin. Micro-biol. 10:712-718.
4. Cornbleet,P.J.,and N.Gochman. 1979. Incorrectleast squares regression coefficients in method comparison analysis. Clin. Chem. 25:432-438.
5. Davis,N.C., C.D.West,and M. Ho.1972.Effect of aging
ofserum onquantitation of complement component C3.
Clin. Chem. 18:1485-1487.
6. Fahey, J. L., and E. M. McKelvey. 1965. Quantitative determinations of serum immunoglobulins in antibody agarplates. J. Immunol. 94:84-90.
7. Garin,E.H.,W. H.Donnelly,S. T.Shulman,R. Fernan-dez, C. Finton,R. L.Williams,andG.A.Richard. 1979. Thesignificance of serialmeasurementsofserum comple-mentC3 and C4 components and DNAbinding capacityin patients with lupus nephritis. Clin. Nephrol. 12:148-155. 8. Mancini, G., A.0. Carbonara, and J.F.Heremans.1965. Immunochemical quantitation of antigens by single radial immunodiffusion.Immunochemistry 2:235-254. 9. Stevens, R.W., D. Elmendorf, M. Gourlay, E. Strobel,
and H. A.Gaafar. 1979. Applicationof
fluoroimmunoas-saytocerebrospinalfluidimmunoglobulinG and albumin. J.Clin. Microbiol.10:346-350.
10. West, C. D., S. Winter, J. Forristal, and N. C. Davis. 1968. Effect ofaging of serumonconsumption of antibody by P3IC-globulin determinants; evidence for circulating
breakdown products in glomerulonephritis. Clin. Exp.
Immunol.3:57-62.
VOL.