Bone stress evaluation in implants with Regular and Switching platforms

N/A
N/A
Protected

Academic year: 2021

Share "Bone stress evaluation in implants with Regular and Switching platforms"

Copied!
8
0
0

Loading.... (view fulltext now)

Full text

(1)

ABSTRACT Objective

Considering that the bone stress caused by the internal hexagon implant is lower in comparison with that caused by external hexagon implant, the aim of this study was to evaluate stress distribution of the peri-implant bone, by simulating the influence of the switching platform in external hexagon implants in comparison with regular platform in the internal hexagon implant.

Methods

TTwo mathematical models of an implant-supported central incisor were created: Regular (R), 4.5 x 11.5 mm internal hexagon implant and 4.5 mm abutment and Switching (S), 5.0 x 11.5 mm external hexagon implant and 4.1 mm abutment. The models were created using the SolidWorks 2010 (3Dtech, São Paulo, Brazil) program. Oblique forces (100 N) were applied to the palatine surface of the central incisor. The bone/implant interface was considered perfectly integrated. Maximum and minimum principal stress values were evaluated in the cortical and medullary bones. The numerical analysis was performed using the ANSYS Workbench 10.0 (Swanson Analysis System, Houston, Pa).

Results

For the cortical bone, the highest stress values were observed in the R (48.8 MPa), followed by the S (48.5 MPa). For the medullary bone, the highest stress values were observed in the S (3.66 MPa), followed by the regular (1.51 MPa).

Conclusion

External hexagon implant with switching platform showed a biomechanical performance similar to that of the internal hexagon implant with regular platform in the cortical bone analysis. Whereas, for the medullary bone, the switching platform model transmitted more stress than the regular model.

Indexing terms: Biomechanics. Dental implantation. Finite element analysis. Maxilla. Osseointegration. RESUMO

Objetivo

Avaliar a distribuição de tensão no osso peri-implantar de implante hexágono interno com plataforma regular e implante hexágono externo com plataforma switching.

Métodos

Foram confeccionados dois modelos matemáticos representativos de uma maxila parcial da região do elemento 11, sendo um modelo Regular, com implante hexágono interno (4,5 x 11,5 mm) e pilar 4,5 mm e outro modelo Switching, com implante hexágono externo (5,0 x 11,5 mm) e pilar 4,1 mm. Ambos os modelos continham uma coroa cimentada sobre o pilar do implante. Os modelos foram desenvolvidos pelo programa SolidWorks 2010 (3Dtech, São Paulo, Brasil). Forças oblíquas (100 N) foram aplicadas na face palatal das coroas. A interface osso-implante foi considerada perfeitamente integrada. Máxima e mínima tensão principal foram avaliadas no osso cortical e medular. A análise numérica foi realizada através do programa ANSYS Workbenck 10,0 (Swanson Analysis System, Houston, Pa).

Resultados

No osso cortical, os maiores valores de tensão foram observados para R (48,8 MPa), seguido de S (48,5 MPa). Para o osso medular, os maiores valores de tensão foram observados no modelo S (3,66 MPa), seguido por R (1,51 MPa).

Conclusão

O modelo switching apresentou comportamento biomecânico semelhante ao modelo de plataforma regular na análise do osso cortical. Enquanto que no osso medular, o modelo switching transmitiu mais tensão óssea do que o modelo Regular.

Termos de indexação: Biomecânica. Implantação dentária. Análise de elementos finitos. Maxila. Osseointegração.

1 Faculdade São Leopoldo Mandic, Curso de Odontologia, Programa de Pós-Graduação em Prótese Dental. Rua José Rocha Junqueira, 13, Swift, 13045-755,

Campinas, SP, Brasil. Correspondência para / Correspondence to: NC Gurgel-Juarez. E-mail: <[email protected]>.

2 Universidade Estadual Paulista Júlio de Mesquita Filho, Faculdade de Odontologia. Araçatuba, SP, Brasil.

Bone stress evaluation in implants with Regular and Switching

platforms

Avaliação da tensão óssea em implantes com plataforma Regular e Switching

Nália Cecília GURGEL-JUAREZ1 Fabiana Mantovani Gomes FRANÇA1 Sidney KINA1

Érika Oliveira de ALMEIDA2 Eduardo Passos ROCHA2

(2)

INTRODUCTION

The Branemark type of implant (Nobelpharma AB, Gothenburg, Sweden) is the most widely distributed on the market worldwide, and has been the most classical and most studied model since the 1960s. Researches conducted with these external hexagon implant have obtained favorable responses in the rehabilitation of edentulous

mandibles and with partial dental prostheses1-2.

One year after these external hexagon implants entered into function, a bone loss up to the level of the

first thread of the implant was observed3. This situation

was so common that in the first reports on criteria for

osseointegration, Albrektsson et al.4 and Smith & Zarb5

included peri-implant calciform resorption as one of the determinant factors of success.

This bone loss could also be observed with the use of the internal hexagon type of implants. In comparison with the external hexagon, the internal connection has presented better performance in laboratory tests and

those of structural integrity of the maxillary implant 6,

anti-rotational stability6, low rate of abutment screw loss7 and

lower stress transmitted to bone8.

In spite of the internal hexagon implants presenting these better mechanical properties with regard to the type of connection, they equally exhibit cup-shaped bone resorption around the head of the implant when they

enter into function9.

As the longevity of dental implants depends on integration between the implant components and hard

and soft tissues of the oral cavity10, various authors began

to study the causes of bone resorption.

The factors most frequently mentioned as possible

causes of this loss are: Occlusal overload11-18, contamination

of the microspace between the abutment and implant19,

re-establishment of biologic space19, the implant neck

design10,20-21, surgical trauma10, peri-implantitis10 and the

gingival biotype22.

Although there is still no consensus in the literature with regard to the factor triggering peri-implant resorption, it is of the utmost importance to know when the peri-implant bone loss attains a stable level, because preservation of the supporting bone is essential for the

esthetics of the soft tissues21,23. With the increasing

requirement by esthetically demanding patients, the search for more natural restorations has been a challenge to clinicians.

Within this context, absence or minimal bone loss would be ideal. An approach suggested by Lazzara &

Porter24 was to alter the horizontal relationship between

the implant diameters and those of the prosthetic component, thus introducting the platform switching concept. The principle of this technique consists of the prosthetic component diameter being smaller than the

implant diameter24, which has been studied a great deal

and widely disseminated in the literature.

Clinical, radiographic and histological studies have shown a reduction in peri-implant bone loss with the use

of platform switching19,21,25-26. Researches using the finite

element method have shown a more uniform distribution of force on the peri-implant bone with platform switching

in comparison with the traditional technique9,12-13,15-16.

Therefore, the aim of this study was to evaluate the different stress distributions on the bone around internal hexagon implants with traditional platforms, and external hexagon implants with platform switching.

METHODS

This study was approved by the Research Ethics Committee of the Araçatuba School of Dentistry (UNESP) under Protocol No. 2008/01845. After signing the Term of Free and Informed Consent, the patient was submitted to tomographic examination of the maxilla, from which a section of the right maxillary central incisor region, cortical and trabecular bone were obtained in dicom format. The mathematical models of the representation of this segment of the maxilla were fabricated using the software programs Mimics 11,11 (Materialise, Leuven, Belgium) and Solidworks (SolidWorks Corp., Concord, MA, USA).

To generate the geometrical models of the implants and prosthetic components of the SIN system (Sistema de Implantes, São Paulo, SP, Brazil), high definition photographs were taken with the aid of static and digital pachymeters (Mitutoyo Corp., Tokyo, Japan) to determine the measurements of the materials. The images were imported into the SolidWorks software program in .tif format for dimensioning, where the final models were made.

Two models were created: R, with an internal hexagon type of implant of the SIN system (Sistema de Implante, São Paulo, Brazil) Strong (4.5 x 11.5 mm) and regular platform (abutment diameter: 4,5 mm); S, with an external hexagon type of implant of the SIN system (Sistema de Implante, São Paulo, Brazil) Strong (5.0 x 11.5 mm) and platform switching (abutment diameter: 4.1 mm). The crown measurements were maintained constant

(3)

in both models, presenting a height of 13 mm, mesio distal width of 8.8 mm and vestibular-palatal distance of 7.1 mm.

The mathematical models consisted of a total crown made of IPS e-max Press (Ivoclar Vivadent, Schaan, Liechtenstein) cemented onto the preparable prosthetic abutment that was screw-retained on the implant, with representation of the cortical bone (1 mm) and medullary bone perfectly united to one another. The soft tissues in the region were not considered. The dimensions of all the structures, with the exception of the type of implant and prosthetic abutment were maintained constant. Simulation of crown cementation on the prosthetic abutments was performed with a 0.05 mm thick layer of Variolink II (Ivoclar Vivadent, Schaan, Liechtenstein) adhesive cement (Figures 1 and 2).

These models were exported to the ANSYS Workbench 10.0 (Swanson Analysis Inc. Houston, Pa, USA) software program, for recognition of the regions, determination of the mechanical properties, and generation of the finite element meshes.

The mechanical properties of the Modulus of Elasticity (E) and Poisson Coefficient () of each structure were used to consider the study homogeneous, isotropic and linearly elastic (Table 1). The mechanical properties refer to a bone quality type III, according to the

classification of Lekholm & Zarb28, as this type of bone is

more prevalent in the anterior region of the maxilla29. The

bone-implant contact interface was considered completely osseointegrated.

Table 1. Mechanical properties of Materials used in the study

Manufacturer 0.30

95.0 IPS e-max Press

Manufacturer 0.30 8.3 Variolink II Huang et al.14 0.35 110.0 Prosthetic abutment Huang et al.14 0.35 110.0 Abutment screw Huang et al.14 0.35 110.0 Implant Kao et al.27 0.30 1.6

Medullary bone type III

Huang et al.14 0.26 13.8 Cortical bone Bibliographic references Poisson coefficient

Modulus of elasticity (Gpa) Material

Manufacturer 0.30

95.0 IPS e-max Press

Manufacturer 0.30 8.3 Variolink II Huang et al.14 0.35 110.0 Prosthetic abutment Huang et al.14 0.35 110.0 Abutment screw Huang et al.14 0.35 110.0 Implant Kao et al.27 0.30 1.6

Medullary bone type III

Huang et al.14 0.26 13.8 Cortical bone Bibliographic references Poisson coefficient

Modulus of elasticity (Gpa) Material

A solid element for parabolic tetrahedron interpolation30 and a mesh composed of elements measuring 0.8mm were used. Model R had 167,899 knots and 105,185 elements; and Model S had 226,966 knots and 142,078 elements (Figure 3).

As the contour condition, there was fixation of all the knots in the mesial and distal regions of the model. All the knots were fixed to the 3 Cartesian axes X, Y and Z.

An oblique (45°) loading (C) of 100 N on the palatal surface of the maxillary central incisor was adopted (Figure 4).

The stress maps with maximum and minimum principal stress (max e min), in MPa, were obtained for the cortical and medullary bone. This criterion of analysis was selected due to the fact that bone is not a ductile material, but a friable material, in addition to presenting tensile stress values differing from the compression stress values.

Figure 1. Regular Models (a) and Switching (b).

Figure 2. Details of structures of Regular Models (a) and Switching (b): implant (1), prosthetic abutment (2), cement line (3), coping (4) and prosthetic crown (5).

(4)

Figure 3.Finite element mesh of Regular Models (a) and Switching (b).

Figure 4.Oblique load on the palatine surface of the maxillary central incisor.

RESULTS

Table 2 shows the numerical results obtained in the study. From a general aspect, the value of max on cortical bone was similar for the two models. On medullary bone, Model S presented higher max.

Table 2. Distribution of maximum principal stress (max) and minimum principal stress (min) on cortical and medullary bone in the regular (R) and switching (S) models. -0.53 MPa -0.53 MPa 3.66 MPa 48.5 MPa Switching (S) -1.0 MPa 8.05 MPa 1.51 MPa 48.8 MPa Regular (R)

αmaxMedullary Bone

αmaxCortical Bone

αmaxMedullary Bone

αmaxCortical Bone

Models -0.53 MPa -0.53 MPa 3.66 MPa 48.5 MPa Switching (S) -1.0 MPa 8.05 MPa 1.51 MPa 48.8 MPa Regular (R)

αmaxMedullary Bone

αmaxCortical Bone

αmaxMedullary Bone

αmaxCortical Bone

Models

In a specific manner for the cortical bone, Model R presented an max value (48.8 MPa) close to that of model S (48.5 MPa), with the difference in values found being around 0.6%. More divergent results were observed as regards the min values (8.05 MPa for Model R and -0.53 MPa for Model S). In addition to the second model having shown compression stress, there was a reduction of 93.41% in the value of min for this group in comparison with the first model (Table 2).

For the medullary bone, the max value in Model S was higher (3.66 Mpa) than the value observed in Model R (1.51 Mpa) by 142.38%. To the contrary, the minimum stress (min) for Model S was lower (-0.53 Mpa) when compared with the value observed for Model R (-1.0 Mpa) by 47%. In addition, it was observed that both min values demonstrated compression stress (Table 2).

In model R, higher stress concentration was observed in the proximal region of cortical bone with reference to the implant platform, represented by the warm colors in the stress maps (Figure 5). The same region of stress concentration was observed in the medullary bone, however, with greater extension towards the apical region, as may be observed in Figure 6.

In Model S, the highest stress concentration occurred on the palatine surface of cortical bone (Figure 5). Whereas on the medullary bone of this model, the highest concentration of stress was observed on the palatine and vestibular surfaces (Figure 6).

a

b

a

b

a

b

Figure 5. Stress Distribution (max) on cortical bone of Models R (a) and S (b).

a

b

a

b

a

b

(5)

DISCUSSION

The finite element method has been an important computational tool, widely used for biomechanical analysis of implants, implant supported dental prostheses and

peri-implant tissues. Quaresma et al.13 have pointed out

the advantage that virtual models have of allowing the evaluation of specific factors, without the influence of other variables.

As has occurred in other studies15,29-30, in the

present study some simplifications were admitted with reference to the characterization of the studied materials being considered homogeneous, isotropic and linearly

elastic. In the literature, there are studies29 that have used

more in-depth methods to determine the characterization of dental materials, such as the determination of possible existent non linearities, anisotrophy or orthotropy and heterogeneities.

In this type of method, certain simplifications are admitted, such as considering the studied materials being homogeneous, istotropic and linearly elastic, in order to make the processes of modeling and resolution possible. Nevertheless, some authors defend the realization of

non-linear, anisotropic and heterogeneous studies29.

With regard to the comparison between implants of different diameters, used in this study, this was also the methodology that has been used by various

authors9,13-14,16-17,25. According to the studies of Ding et

al.17, the differences in behavior between implants with

diameters of 4.1 and 4.8 mm (0.7 mm of difference between the platforms) were not notable. In the present study, the difference between the implant diameters was still less, to the order of 0.5 mm (implants with platforms of 4.5 and 5 mm).

This study considered perfect union of the bone-implant interface, with 100% osseointegration, in

accordance with the literature13,15,29.

The force of loading was 100N, as in the studies

in the literature13,15. This force with oblique load on the

maxillary central incisor in the palatine-vestibular direction simulated that which occurs physiologically, and was also

used by Saab et al.29, who affirmed that this provides

better conditions for analyzing the behavior of the studied implants, with bone responses within the physiological limits. Furthermore, the literature is unanimous in affirming that implants submitted to inclined forces transfer more

stress to the adjacent bone15,17-18.

When an implant is submitted to inclined loads in relation to its long axis, the highest stress generated is located on the cortical bone around the neck of the

implant12,13-18. In this study, this result was also observed, so

that for Model R, the concentration of this stress occurred specifically in the proximal regions. Whereas in Model S, the stress was more evident in the palatine region, suggesting a better behavior in an esthetic area when platform switching is used, because, less stress transmitted to the proximal bone may involve less bone loss in this region.

Analysis of this cortical bone showed that Groups R and S presented a similar behavior when the maximum principal stress (max) was verified in the models, since Group R presented slightly higher max (48.8 MPa) than Group S (48.5 MPa). However, one could speculate that the increase in implant diameter could result in less tension transmitted to the cortical bone, as various authors have

affirmed14,17-18.

Rodríguez-Ciurana et al.9 also found no significant

difference between the stresses presented by the external hexagon implant models with a diameter of 5 mm and internal hexagon with a diameter of 4.8 mm on cortical bone – dimensions close to those used in the present study; however, both models had a 4.1 mm abutment.

Maeda et al.12 observed higher stress concentration around

the regular implant when compared with the implant with platform switching, in his tridimensional models; this was similar to the effect that occurred in the S and R models.

Quaresma et al.13 found higher stress on the cortical bone

around the internal hexagon implant than the platform switching implant (characteristic of the implant design with internal conical connection used by the authors). However,

the values found in the study of these authors13 were more

discrepant, and not as close to the values of the present study - probably due to the internal conical connection used by the authors.

When the minimum principal tension (min) on cortical bone was analyzed, Group R presented tensile stress (8.05 MPa), whereas Group S presented compression stress (-0.53 MPa). Therefore, both max and min in Group R were tensile stresses, whereas in Group S tensile (max) and compression (min) stresses were found. These results

are similar to those found by Baggi et al.16, because their

implant model with platform switching presented tensile and compression stresses on the cortical bone.

The study of medullary bone showed that the maximum principal stress was higher for Group S (3.66 MPa) than for Group R (1.51 MPa). This means that the implant with a larger diameter (S: 5 mm) transmitted more

(6)

stress to the medullary bone than the implant with the 4.5 mm platform (R), which is in agreement with the findings

in the literature13,17. The explanation for this result is the

probable reduction in the quantity of medullary bone

when the implant diameter is increased17. According to

Holmgren et al.11, this result shows that the wider implant

is not necessarily the best choice because in cases with morphological limits, the stress distribution on the bone is unfavorable. However, when one takes the cortical bone into consideration, the larger implant diameter transfers

less bone stress11,14,17-18 and Ding et al.17 affirmed that

an implant with a wider cervical portion may dissipate masticatory forces better, thereby diminishing bone loss. Thus, from a biomechanical perspective, these authors suggest the choice of an implant with the maximum diameter allowed by the anatomy of the region.

From the point of view of the type of hexagon of each implant, this study showed that the internal connection generated 142% less stress on the medullary bone than that generated by the external connection,

which is in harmony with the results of Baggi et al.16, in

which the external hexagon generated around three times more bone stress than the internal hexagon.

In the studied models, the stress concentration on the medullary bone occurred in the cervical region. For Model S this stress was located on the vestibular and palatine surfaces, whereas for Model R stress was concentrated in the proximal region with greater extension in the apical direction. Without specifying the surfaces, some authors obtained stress concentration in the cervical

region of the medullary bone around the implant9,16-17.

However, for the cortical bone thickness of 1 mm (used in

this study), Okumura et al.18 reported stress concentration

on the apex of the implant; these authors found stress in the cervical region only for the simulated corticals of 0.5 and 0 mm. This difference possibly occurred because the

simulation used by these authors18 was of the posterior

maxilla, with different dimension from those of the model used in the present study, and modeling of the maxillary sinus. The behavior of apical stress propagation of Model R being more perceptible in Model S was also observed in

the study of Maeda et al.12.

It is known that the implant with an internal hexagon has a more favorable biomechanical performance

than that of the implant with an external hexagon6-8, in spite

of the external hexagon, classically, still being the platform most frequently used. There is also common accord in the literature that platform switching has diminished bone loss

around the neck of the implant9,12-13,15-16,19,21,24-26.

Therefore, it could be affirmed that within the limitations of this study, the use of the wide external hexagon implant with the platform switching principle obtained a similar performance to that of an internal hexagon implant with a regular platform, in the anterior maxilla, with regard to stress transmission to the cortical bone (which is the region were bone loss begins).

CONCLUSION

According to the methodology used, it could be concluded that for cortical bone, implants with the external hexagon type of connection, using the principle of platform switching, have a similar biomechanical performance to that of internal hexagon implants and platform diameter coinciding with that of the prosthetic platform.

Whereas, for medullary bone, the external hexagon implant model with platform switching transmitted higher bone tension than the internal hexagon implant model with a regular platform.

Collaborators

NCG JUAREZ conducted the scientific research, and performed the entire bibliographic survey and writing of the article. FMG França guided the research and participated in writing the article. S KINA was responsible for the idea of the work and participated actively in correction, carrying out the work and writing the article. ÉO Almeida fabricated all the tridimensional computational models, and together with NCG JUAREZ, performed the laboratory tests of the study. He participated actively in writing the article. EP ROCHA co-oriented the research, made the partnership between the institutions feasible for the development of the study. He guided the author ÉO ALMEIDA in fabrication of the computational models and helped to conduct and structure the entire development of the study. He participated in writing the article.

(7)

REFERENCES

1. Branemark P-I, Adell R, Breine U, Hansson BO, Lindström J, Ohlsson A. Intra-osseous anchorage of dental prosthesis. Experimental studies. Scand J Plast Reconstr Surg. 1969;3(2):81-100.

2. Jemt T, Chai J, Harnett J, Heath MR, Hutton JE, Johns RB. A 5-year prospective multicenter follow-up report on overdentures supported by ossseointegrated implants. Int J Oral Maxillofac Implants. 1996;11:291-8.

3. Adell R, Lekholm U, Rockler B, Branemark P-I, Lindhe J, Eriksson B, et al. Marginal tissue reactions at osseointegrated titanium fixtures. A 3 year longitudinal prospective study. Int J Oral Maxillofac Implants. 1986;15(1):30-52.

4. Albrektsson T, Zarb GA, Worthington P, Eriksson AR. The long-term efficacy of currently of currently used dental implants: a review and proposed criteria of success. Int J Oral Maxillofac Implants. 1986;1(1):11-25.

5. Smith DE, Zarb GA. Criteria for success of osseointegrated endosseous implants. J Prosthet Dent. 1989;62(5):567-72. doi: 10.1016/0022-3913(89)90081-4.

6. Möllersten L, Lockowandt P, Lindén L-A. Comparison of strength and failure mode of seven implant systems: an in vitro test. J Prosthet Dent. 1997;78(6):582-91. doi: 10.1016/S0022-3913(97)70009-X.

7. Krennmair G, Schmidinger S, Waldenberger O. Single-tooth replacement with the Frialit-2 system: a retrospective clinical analysis of 146 implants. Int J Oral Maxillofac Implants. 2002;17(1):78-85.

8. Bernardes SR, de Araujo CA, Neto AJ, Simamoto Junior P, das Neves FD. Photoelastic analysis of stress patterns from different implant-abutment interfaces. Int J Oral Maxillofac Implants. 2009;24(5):781-90.

9. Rodriguez-Ciurana X, Vela-Nebot X, Segala-TorreS M, Rodado-Alonso C, Mendez Blanco V, Mata-Buguelores. Biomechanical repercussions of bone resorption related to biologic width: a finite element analysis of three implant-abutment configurations. Int J Periodontics Restorative Dent. 2009;29(5):479-87. 10. Oh T-J, Yoon J, Misch CE, Wang H-L. The causes of early implant

bone loss: myth or science? J Periodontol. 2002;73(3):322-33. 11. Holmgren EP, Seckinger RJ, Kilgren LM, Mante F. Evaluating

parameters of osseo integrated dental implants using finite element analysis-a two dimensional comparative study examining the effects of implant shape, and load direction. J Oral Implantol. 1998;24(2):80-8. doi: 10.1563/1548-1336(1998)024<0080:EP OODI>2.3.CO;2.

12. Maeda Y, Miura J, Taki I, Sogo M. Biomechanical analysis on platform switching: is there any biomechanical rationale? Clin Oral Impl Res. 2007;18(5):581-84. doi: 10.1111/j.1600-0501.2007.01398.x.

13. Quaresma SET, Cury PR, Sendyk WR, Sendyk C. A finite element analysis of two different dental implants: stress distribution in the prosthesis, abutment, implant, and supporting bone. J Oral Implantol. 2008;34(1):1-6. doi: 10.1563/1548-1336(2008)34[1:AFEAOT]2.0.CO;2.

14. Huang HL, Hsu JT, Fuh LJ, Tu MG, Ko CC, Shen YW. Bone stress and interfacial sliding analysis of implant designs on an immediately loaded maxillary implant: a non-linear finite element study. J Dent. 2008;36(6):409-17. doi: 10.1016/j. jdent.2008.02.015.

15. Schrotenboer J, Tsao YP, Kinariwala V, Wang H L. Effect of microthreads and platform switching on crestal bone stress levels: a finite element analysis. J Periodontol. 2008;79(11):2166-72. doi: 10.1902/jop.2008.080178.

16. Baggi L, Cappelloni I, Di Girolamo M, Maceri F, Vairo G. The influence of implant diameter and length on stress distribution of osseointegrated implants related to crestal bone geometry: a three-dimensional finite element analysis. J Prosthet Dent. 2008;100(6):422-31. doi: 10.1016/S0022-3913(08)60259-0. 17. Ding X, Zhu X-H, Liao S-H, Zhang X-H, Chen H. Implant-bone

interface stress distribution in immediately loaded implants of different diameters: a three-dimensional finite element analysis. J Prosthodont. 2009;20:1-10.

18. Okumura N, Stegaroiu R, Kitamura E, Kurokawa K, Nomura S. Influence of maxillary cortical bone thickness, implant design and implant diameter on stress around implants: a three-dimensional finite element analysis. J Prosthodont Res. 2010. doi: 10.1016/j. jpor.2009.12.004.

19. Jung RE, Jones AA, Higginbottom FL, Wilson TG, Schoolfield J, Buser D, et al. The influence of non-matching implant and abutment diameters on radiographic crestal bone levels in dogs. J Periodontol. 2008;79(2):260-70. doi: 10.1902/ jop.2008.070132.

20. Jung YC, Han CH, Lee KW. A 1-year radiographic evaluation of marginal bone around dental implants. Int J Oral Maxillofac Implants. 1996;11(6):811-8.

21. Shin T-K, Han C-H, Heo S-J, Kin S, Chun H-J. Radiographic evaluation of marginal bone level around implants with different neck designs after 1 year. Int J Oral Maxillofac Implants. 2006;21(5):789-94.

22. Evans CD, Chen ST. Esthetic outcomes of immediate implant placements. Clin Oral Implants Res. 2008;19(1):73-80. DOI: 10.1111/j.1600-0501.2007.01413.x.

23. Tarnow D, Elian N, Fletcher P, Froum S, Magner A, Cho S-C, et al. Vertical Distance from the crest of bone to the height of the interproximal papilla between adjacent implants. J Periodontol. 2003;74(12):1785-8. doi: doi:10.1902/jop.2003.74.12.1785. 24. Lazzara RJ, Porter SS. Platform switching: a new concept in

implant dentistry for controlling postrestorative crestal bone levels. Int J Periodontics Restorative Dent. 2006;26(1):9-17.

(8)

25. Cappiello M, Luongo R, Di lorio D, Bugea C, Cocchetto R, Celletti R. Evaluation of peri-implant bone loss around platform-switched implants. Int J Periodontics Restorative Dent. 2008;28(4):347-55.

26. Guirado JLC, Ruiz AJO, Moreno GG, Mari LL, Gonzalez LAB. Immediate loading and immediate restoration in 105 expanded-platform implants via the Diem System after a 16-month follow-up period. Med Oral Patol Oral Cir Bucal. 2008;13(9):576-81. 27. Kao H-C, Gung Y-W, Chung T-F, Hsu M-L, Dent DM. The influence

of abutment angulation on micromotion level for immediately loaded dental implants: a 3-D finite element analysis. Int J Oral Maxillofac Implants. 2008;23(4):623-30.

28. Lekholm U, Zarb GA. Patient selection and preparation. In: Branemark PI, Zarb GA, Albrektsson T. Tissue-integrated prostheses: osseointegration in clinical dentistry. Chicago: Quintessence; 1985. p.199-209.

29. Saab XE, Griggs JA, Powers JM, Engelmeier RL. Effect of abutment angulation on the strain on the bone around an implant in the anterior maxilla: a finite element study. J Prosthet Dent. 2007;97(2):85-92. doi: 10.1016/j.prosdent.2006.12.002. 30. Cook RD, Malkus DS, Plesha ME. Concepts and applications

of finite element analysis: John Wiley e Sons. New York: [s.n.]; 2001.

Received on: 13/12/2010 Final version resubmitted on: 6/9/2011 Approved on: 13/1/2012

References

Related documents

PERFORMANCE of DIABETES SELF-CARE Cognitive Ability Time Management Ability to organize and prioritize Set Goals Reprioritize Biological Gender Female Emotional State

19% serve a county. Fourteen per cent of the centers provide service for adjoining states in addition to the states in which they are located; usually these adjoining states have

Standardization of herbal raw drugs include passport data of raw plant drugs, botanical authentification, microscopic &amp; molecular examination, identification of

The degree of resistance exhibited after 1, 10 and 20 subcultures in broth in the absence of strepto- mycin was tested by comparing the number of colonies which grew from the

In our study, consumption of high zinc biofortified wheat flour for 6 months compared to low zinc bioforti- fied wheat resulted in statistically significant reduction in days

The aim of the study was to assess the presence of pathogenic microbes on treatment tables in one outpatient teaching clinic and determine a simple behavioral model for

It was decided that with the presence of such significant red flag signs that she should undergo advanced imaging, in this case an MRI, that revealed an underlying malignancy, which

Also, both diabetic groups there were a positive immunoreactivity of the photoreceptor inner segment, and this was also seen among control ani- mals treated with a