0095-1137/81/061040-06$02.00/0
Evaluation of an Indirect
Hemagglutination
Test
for
Legionella
pneumophila
Serogroups
1
to
4
CHERYL A. YONKE,' HARLAN E. STIEFEL,2 DAVID L.WILSON,3 AND BERTTINA B.
WENTWORTH2*
DepartmentofMicrobiologyand PublicHealth,MichiganStateUniversity,EastLansing,
Michigan
48824'; Bureauof Disease Control andLaboratoryServices, MichiganDepartment ofPublicHealth,Lansing,Michigan 489092; and DifcoLaboratories, Detroit,Michigan482013
Received 10 December1980/Accepted 24February1981
Paralleltestingof895 seraby indirecthemagglutination andindirect fluores-cent-antibodytechniquesshowed 97.3%agreement.Althoughthe indirect hemag-glutination technique usually showed more
cross-reactivity
among serogroupsthan the indirect fluorescent-antibody technique with Formalin-fixed antigens
anda conjugate which detected primarily
immunoglobulin
Gantibodies,
heter-ologous serogroup reactionswere
significantly
lower thanhomologous
serogrouptiters and theetiological serogroup couldbe easily defined.Theindirect hemag-glutinationtechnique showed nocross-reactivity witha crudeextract of Esche-richiacoli013:K92:H4. Since the indirecthemagglutinationtechniquewasshown
to detect both immunoglobulin M and immunoglobulin G antibodies and was found to berapid,simple,andinexpensive,itappearstobeanexcellentalternative
tothe indirect
fluorescent-antibody
testforserodiagnosis
oflegionellosis.The indirect fluorescent-antibody (IFA) test
forlegionellosis (10, 15) hasbecome an
impor-tantdiagnostictool in the 3yearssinceits
intro-duction.
Performing
thetest on aroutinebasis,however, is becoming an increasingly complex
task; with the recent addition of the agents TATLOCK (13), WIGA (1), and NY 23 (4) to
the genus Legionella(proposeddesignationsL. micdadei [9], L. bozemanii, and L. dumoffii
[2], respectively), proper serodiagnosis of
Le-gionnaires disease involvesthe useof nine
sep-arate antigens. The use ofpolyvalent antigens
(7, 15) can simplify testing somewhat, buteven
this doesnotdiscountsome inherent disadvan-tages of thetechnique, i.e., it istechnically dif-ficult and expensive to perform. As a practical
alternative, Edson and associates (5) have
intro-ducedanindirecthemagglutination (IHA) tech-nique for legionellosis. The test is rapid, inex-pensive, and simple to perform and shows a 94.9% agreement with IFA serogroup 1 results. The IHA test has recently been expanded to includeL.pneumophila serogroups 2, 3, and 4, and thisstudywas madetoevaluate the
sensi-tivityandspecificityof themethod.
MATERIALS AND METHODS Sera forparallel testing ofIHAand IFA
tech-niques. Serawerereceived at theMichigan
Depart-ment ofPublic Health (Lansing, Mich.) for routine
IFAlegionellosistestingbetween MarchandOctober
1979.Ailserareceivedduringthisperiodwereincluded
inthestudy exceptforthose in whichthequantityof serum wasinsufficient for testing by bothtechniques. The sampletestedincludedatotalof895sera.There were single serum specimens from 483 patients and
pairedorserialserafrom 188patients.
Special sera selected for comparison of IHA andIFAtechniques. In additiontothe above-men-tioned, unselected sera, sera from 55 patients with
IFA-confirmed (fourfold or greater rise in titer) or presumptive (titer of -1:128) legionellosis were
se-lected for comparative studies. Single serum
speci-mens were available from 8patients, and paired or
serialsera wereavailablefrom47patients. Thesecases
represent 41% ofthe 134 cases oflegionellosis
diag-nosedby serology betweenOctober 1977 andAugust
1980. The onlyadditional criterionforinclusioninthe studywassufficient serum volumefortesting.
Sera for cross absorption andblocking fluid studies.The 13 sera withserogroup 1titers usedfor thesestudies(seeTable 6) were fromIFA-confirmed orpresumptivecases.Of the12serawithEscherichia colititers,4 werefromLegionnaires diseasepatients,
and 8werefrom patients without legionellosis.
Sera for sucrose density gradient fractiona-tion. Twelve sera selected from the 55 patients
de-scribed above were used. An additional three sera whichshowed fourfoldorgreater rises or titers of.-1:
128by IHA and were IFA negative werefractionated foranalysis.
Bacterial strains.Bacterial strains were grownin Rouxbottlesonenrichedchocolateagar (GCmedium base with heated sheep blood and vitamin supple-ments). The L. pneumophila Flint 1, Flint 2, and Detroit1strains weremaintained as stock culturesat the Michigan Department of Public Health. The L. 1040
on February 7, 2020 by guest
http://jcm.asm.org/
IHA FOR L. PNEUMOPHILA 1041 pneumophila Knoxville 1, Togus 1, Bloomington 2,
and Los Angeles 1 strains and E. coli 013:K92:H4
were obtained from the Centers for Disease Control
(Atlanta, Ga.).
IFAantigenpreparation. Bacterialgrowthwas
washed fromthesurfaceof the Rouxbottle after72h of incubation with 0.4% formalinized phosphate-buffered saline, filtered through a fine-mesh wire screentoremoveagardebris,andthenrefrigeratedat 4°Cfor1week.Thesuspensionwasthen washed three
timesinformalinizedphosphate-buffered saline, and
the opacity was standardized to 300 international
opacityunitsperml by comparisonwithWorld Health
Organization opacity standards. Suspensions of for-malinizedwholeorganisms,adjustedto2international opacityunitsperml,servedastheantigen.
IHAantigenpreparation. AllIHAantigenswere
suppliedasexperimental antigens byDifco
Laborato-ries(Detroit, Mich.).Themethodofpreparationwas
similar to that of Edson et al. (5) for other IHA
antigens. The antigens consisted ofturkey
erythro-cytesstabilizedbyglutaraldehydetreatmentand sen-sitized withboiled,sonicatedantigenfromagar-grown
isolatesbythebis-diazotized benzidinetechnique.
An-tigenswerepreparedfromthefollowingstrains: sero-group1 antigenas amixture ofFlint 1,Flint2, and
Detroit1;serogroup2fromTogus1;serogroup3 from
Bloomington2; andserogroup4fromLosAngeles1. Thepolyvalent antigenwasmadefromamixtureof
Knoxville1,Togus1, Bloomington 2,and LosAngeles 1.
IFA technique. A standard microtiter technique
with0.005-ml amounts ofantigenon12-well, acetone-resistantglassmicroscope slideswasemployed.
Con-jugatesweredilutedinphosphate-bufferedsalinewith 2% Tween 80. Rabbitanti-humanglobulin conjugate
waspurchasedfromBeckmanInstruments,Inc.
(Ful-lerton,Calif.).Anti-rabbitglobulin conjugate (Cappel Laboratories, Downington, Pa.)wasusedto determine titers on hyperimmune rabbit antisera. Anti-human
immunoglobulinM (IgM) conjugatewassupplied by
Burroughs Wellcome Co. (Research Triangle Park, N.C.).
IRAtechnique. TheIHA techniquewassimilar to
that ofEdson etal.(5). Unless otherwisenoted,sera werescreenedataninitial1:16dilutionagainst
poly-valent sensitized cellsand unsensitized controlcells.
Anyserumshowingareactionwith the testcellsbut notwiththecontrolcellswassubsequentlyscreened
withthe four monovalentantigens (serogroups1 to4)
andcorrespondingcontrolcells.Seraweretitrated to endpointinserial,twofold dilutions withappropriate
monovalentantigens.
Anyserumshowingagglutination ofboth
antigen-sensitizedand unsensitized controlcellswastreatedin the followingmanner. A 0.02-ml amountofpacked,
unsensitized (control) cells was added to0.12 mlof serum. After overnight incubation at 4°C, the cell suspension wascentrifuged at3,000rpmfor 15min, the absorbedserum wasremoved,and the IHA test
wasrepeated. Sera withhightitersofanti-cell anti-bodyrequiredtwoorthree successiveabsorptionsfor completeremoval of theseagglutinins.
Crossabsorption.Forabsorption,0.1mlofserum
was added to approximately 3.7 x 1010 organisms
which had been washed three times in phosphate-buffered saline to remove Formalin. The absorption mixture was incubated overnight at 4°C and centri-fuged at 3,000 rpm for 15 min, and the absorbed serum wasremoved.
Preparation of hyperimmune sera. Rabbits weighing >2 kg each were immunized with an initial intramuscular injection of1ml of antigen mixed with
anequal amount of Freund complete adjuvant, fol-lowed by five intravenous boosters of1ml of antigen
atdays 22, 24, 26, 28, and 30. Rabbits were exsangui-nated 9 days after the last booster.
Comparisonof immune sera preparedagainst IFA and IHA antigens. Rabbits were divided into four groups of two rabbits each and immunized as described above. Groups1and 3received injections of
fornalinizedIFAantigens at 16international opacity units per ml. Groups 2 and 4 received injections of IHA antigen(boiled and sonicated) at 16 international opacity units per ml. Groups1and 2 wereinoculated with the Flint1 strain of L.pneumophila. Groups 3 and4 were immunized with the Detroit 1strain of serogroup 1.Sera from two rabbits in each immuni-zation group were pooled, and IFA and IHA titers were thendetermined on each of the pools.
Blocking fluid preparation. Blocking fluids of the Flint 1 strain and E. coli 013:K92:H4 were pre-pared in the manner described by Wilkinson et al. (14). Sera were initially diluted in IHA diluent or
phosphate-buffered saline as a control and in both blocking fluids. Subsequent dilutions were made in diluent or phosphate-buffered saline, and IHA or IFA titers were determined.
Sucrose density gradients.Sera were fraction-atedon 10 to 50% sucrose density gradients (12) by centrifugation at 33,500 rpm for 17 h in a Spinco ultracentrifuge with anSW50.1 rotor. Ten0.5-mi frac-tions werecollected for analysis.
lmmunodiffusion.The presence of IgM or IgG or
both in each of the samples was confirmed with radial immunodiffusion kits (Meloy Laboratories, Spring-field, Va.). These kits employ the Fahey and Mc-Kelvey (6) techniqueto measureimmunoglobulins.
RESULTS
Parallel testing of 895 sera by the IHA
and IFA
techniques.
Results ofparallel
testingof 895 sera
by
theIHAandIFAtechniques
areshown inTable 1. There were 796nonreactive
TABLE 1. Resultsofparallel testingbythe IHAand IFAtechniqueson 895 sera
No.ofserainserogroup: Results oftesting
1 2 3 4
IHAa
andIFA- 796 879 885 858IHA+bandIFA+ 46 1 5 15
IHA-and IFA+ 14 0 2 2
IHA+andIFA- 39 15 3 20
a_ Titerof<1:16.
b+,Titer of
.1:16.
VOL.13,1981
on February 7, 2020 by guest
http://jcm.asm.org/
sera and 46 sera which reacted in both
tech-niques inserogroup 1. Therewere14sera
reac-tiveonly in the IFA technique whichwere
non-reactiveinthe IHAtechnique ata1:16dilution. There were 39 sera reactive only in the IHA
technique that failedto react in the IFA tech-niqueata1:16 dilution. Discrepanciesbetween
IFA and IHA reactivities withserogroups2, 3,
and 4 were due largely to the cross-reactive nature of the serogroup 1 antibodyand not to
serawithspecific serogroup2, 3,or4 antibody.
Agreement (percentage ofserapositive byboth
techniques plus percentage ofseranegative by
bothtechniques) was94.1, 98.3, 99.4, and 97.5%
forserogroups1,2, 3,and4, respectively. There was anoverall agreementforserogroups 1 to4
of 97.3%.
Among the 188 paired or serial sera, there
were nine instances of fourfold or greater
in-creases in titer and four discrepancies byboth
theIHAandIFAtechniques. Onepatientwith
afourfoldrisein IFA titer showedatwofoldrise
byIHA,anotherpatientshowed thereverse(i.e., afourfoldrisebyIHA andatwofoldrisebythe
IFAtechnique). Another patientwithagreater
than fourfoldincreasebythe IFAtechniquehad
a high stable titer (1:2,048) by IHA. Finally, serumfromapatientwith successive IFAtiters
of1:256and 1:128 inserogroup 1wasIHA
neg-ative.
Some of the diverse cross-reactivity among serogroups of the human antibody response
notedby Wilkinsonetal. (15) in the IFA tech-nique was observed with the IHA technique.
However, with Formalin-fixed antigens, the IFA technique didnotappeartobesocross-reactive
inthisstudy
(Table
2).When cross-reactivity wasseen,itwasfound
that theetiologicalserogroup could bestbe de-finedbycrossabsorption. Absorption ofa serum
with a serogroup-specific antigen removed the
cross-reacting antibody aswell as the
homolo-gous antibody. Absorption witha heterologous
serogroupantigen removedonlythe cross-react-TABLE 2. Diverse serogroup reactivity of the human
antibody response Convalescentserum titer
Sero-group an- Serum1 Serum2 Serum3 tigen
IFA IHA IFA IHA IFA IHA
1 64a 512 64 16 128 2,048
2 __b _c - 256 - 32
3 - - 512 512 -
-4 - 32 16 128 8 128
aReciprocalofhighest
dilution
showing reactivity. bIFA titer of <1:8.cIHA titer of <1:16.
ing antibody but didnotdecrease the
serogroup-specifictiter. Tencasesofserogroup 1 andone case of serogroup 4 were identified by cross
absorption. Two examplesareshown inTable3.
Patient 1appears tohaveserogroup 1
legionel-losis,and patient 2appearstohaveLegionnaires
disease ofserogroup4etiology.
Comparison of IHA and IFA techniques
onselected serafrom 55 patients. Of the 55
patients with IFA-confirmedorpresumptive
le-gionellosis, 54 were identified as serogroup 1,
and 1 was identified as serogroup 3. The IFA
technique detected fourfold or greater rises in titer in 42 patients,41of whomwereinserogroup
1 and 1 of whom was in serogroup 3. Ofthese
patients,37(36 inserogroup 1and1inserogroup
3) also showed significant rises in titer by the IHA technique (Table 4). In three of the
re-maining five patients, the IHA technique showed a rise from <1:16 to 1:16, and in two cases the IHA technique was nonreactive. Of
13 patients with presumptive legionellosis, 7
showed IHA titers of -1:128, 1 showed a rise
from <1:16to1:16, 2showed titers between 1:16
and 1:128, and 3 were nonreactive in the IHA
technique (Table 4).
TABLE 3. Crossabsorption to define etiology Serum IHAtiter of
Patient Serum serogroup:
1 2 3 4
1 Unabsorbed 256a - - 64
Absorbedwith Flint 1 _b - -
-(serogroup 1)
Absorbed with LosAn- 256 - -
-geles1(serogroup4)
2 Unabsorbed 256 - - 2,048
Absorbed with Flint1 - - - 2,048 (serogroup1)
Absorbed with LosAn- - - -
-geles1(serogroup 4)
Reciprocal of highest dilutionshowingreactivity. bTiter of<1:16.
TABLE 4. Comparison of IFAandIHAresults from
55patients with IFA-confirmed or presumptive legionellosis
IHA(no.of patients) IFA Pre-
~~Change
in SainIA Con-
Pr-titer
fromStto-Nea
firmedA tivebP negative' tivect 6 ters,
Confirmed 37 O 3 O 2
Presumptive 0 7 1 2 3
a Fourfold or greater rise in titer.
bTiter of -1:128. Titer of <1:16.
dReciprocalofhighest dilutionshowingreactivity. e Between 1:16 and 1:128.
on February 7, 2020 by guest
http://jcm.asm.org/
IHA FOR L. PNEUMOPHILA 1043
Itwasconsidered that differencesbetween the
resultsfrom the IHA and IFA techniquesmight
beexplained by: (i) measurement oftwo
differ-entantigen-antibodysystems,(üi)reactionswith
across-reactingantigen(s),or (iii)differencesin
immunoglobulin specificity. Toexplore each of
these possibilities, the studies discussed below
wereconducted.
Comparison of immune sera prepared
against
IFA and IHA antigens. Results ofIFA and IHA testing of hyperimmune rabbit
sera that were prepared with IFA and IHA
antigens areshowninTable5. Iftwodifferent
antigen-antibody
systems are involved, titerswithhomologousantigens might beexpected to
behigher than those with
heterologous
antigens.However, therewere no significant differences
intiters withIFAandIHAantigens, regardless
of theantigen usedtopreparethe antisera.
Blocking fluid andcross
absorption
stud-ies. Wilkinson and associates reported that a
crude extract of E. coli 013:K92:H4 can block
97% ofIFA-positive reactions withavariety of
gram-negativebacterial species and block6% of
L.pneumophila titers (14). ThisstrainofE.coli
wasobtained from H. W.Wilkinson, and
block-ing fluids of it and the Flint 1strain were
pre-pared.
Twelveserawith
anti-serogroup
1 IHAtiterswere tested by comparing titrations done with
blocking
fluid as diluent with titrations done withregular
test diluent (Table 6). Dilution inE. coli
blocking
fluid hadnosignificant
effectonanyofthe 12sera
tested,
whereasFlint1block-ingfluid causedafourfoldorgreaterdecrease in
all of theseratested.
Blocking fluid studies by theIFA technique
showed similar results. E. coli
blocking
fluid hadnoeffectonthe 13seratested.Dilution in Flint
1blocking fluid reducedserogroup 1titers
by
atleastfourfoldin100% of thecasestested.Twelve
serawith initial IFAE.colititerswere
similarly
tested
(Table 6).
Flint1blocking
fluid causednoTABLE 5. Comparisonofantiserapreparedagainst
IFA and IHAantigens
Titer'by:
tiommunia
Immunizing antigen
tiongroup IHA IFA
1 Flint1,IFA 4,096b 2,048c
2 Flint1,IHA 4,096 2,048c 3 Detroit1,IFA 4,096 2,048d
4 Detroit1, IHA 8,192 4,096d
a
Results
shown are theaverage valuesoftriplicate determinations.bReciprocal ofhighestdilutionshowingreactivity. cFlint1servedasthetestantigen.
dDetroit1servedasthetestantigen.
TABLE 6. Results of blocking fluid and cross-absorptionstudies
Test
IHA IHA IHA IHA IFA IFA IFA IFA IFA IFA IFA IFA
Initialserum titer to: Serogroup1
Serogroup1
Serogroup1
Serogroup1
Flint1
Flint1
Flint1
Flint1
E.coli E.coli E. coli E.coli
BF'orAbs.b used Flint1BF E.coli BF Flint1 Abs.
E.coli Abs. Flint1BF E.coli BF Flint1Abs. E. coli Abs. Flint1BF E.coli BF Flint1Abs. E.coli Abs.
Decrease in titer' 12/12
0/12 11/11 0/11 13/13 0/13 9/9 0/9 0/12 12/12 1/8
8/8
aBF, Blocking fluid used for initial
dilution
ofse-rum.
bAbs., Whole-cellantigen usedtoabsorb sera.
CFourfoldorgreater decrease in titer.
significant decreases in titer in any of the 12
sera,whereas E. coliblocking fluid significantly
decreased titersinallof thesera. There was one
patientwho showed a concurrent fourfold rise in
titer to both E. coli and the Flint 1 strain in
whichblocking was specific,i.e.,Flint 1 blocking
fluid reduced only the Flint 1 titer, and E. coli
blocking fluid reduced only the E. coli titer.
Titersin another patientwho showed a fourfold
rise in serogroup 1 titer to 1:128 by the IFA
techniqueand a maximumIHAtiter of 1:16 were
blocked by Flint1 blocking fluidbut not by E.
coli
blocking
fluid.Cross absorptionwasfelttobe a more
conclu-sivetestfor common antigenthan was the use
ofcrudeblocking fluidextract.Absorption of11
serathat hadserogroup 1 IHAtiters with
whole-cell E. coli antigen caused no significant
de-creasein any serogroup 1titers, whereas
absorp-tionwiththewhole-cell Flint 1 strain caused a
fourfoldor greaterdecrease inall titers (Table
6). Similarly, in the IFAtechnique, absorption
withwhole-cell E. coli did not change Flint 1
titers in nine sera, whereasabsorptionwith the
Flint 1 strain significantly decreased all of the
titers. Crossabsorptionstudiesofeightserawith
E.coliIFA titers
proved specific
forall
butoneserum.Thisserumhad aninitialE. colititerof 1:32, andabsorptionwith E. colidecreasedthe
titer to <1:8. Absorption with the whole-cell
Flint 1 strain decreased the titer fourfoldto 1:8.
Fractionation of sera on sucrose
gra-dients to determine the
immunoglobulin
class reactive in theIRAand IFAtests. A
total of 15 serawerefractionated
by
thesucrosedensity gradient
technique,
and the fractionswereanalyzed byboththe IFA and IHA
meth-ods (Table 7).Threeserafromcases
diagnosed
VOL.13,1981
on February 7, 2020 by guest
http://jcm.asm.org/
1044 YONKE ET AL.
TABLE 7. Immunoglobulin reactivity of15sera fractionatedon sucrosedensitygradients
No. of sera withreactivityin: Technique
IgMfractions IgG fractions
IFA O 12
IHA 13 il
TABLE 8. ComparisonofIHA and IFA resultson 10sucrosedensity gradientfractions
Immunodiffu-Fraction IHA IFAa IFA-Mb sion'
no.
IgM IgG
1 8d _e - _ _
2 32 - 2 +
-3 512 - 32 +
-4 128 - 8 +
-5 32 4 - - +
6 32 16 - - +
7 4 32 - - +
8 - 2 - - +
9 - _
-10 - -
-aAnti-humanglobulin
conjugate.
bAnti-humanIgM
conjugate.
'Presence (+) or absence (-) ofIgM and IgG in
eachfraction.
dReciprocalofhighestdilutionshowingreactivity.
e_,
Noreaction withundiluted
sample.by the IHAtechnique butnotbythe IFA
tech-nique showed reactivity in only the IgM
frac-tions. Tenserashowed
reactivity
intheIgMandIgGfractionsbythe IHAtestbut showed
reac-tivityinonlythe
IgG
fractionsbythe IFAtest.Another serum showed reactivity in the IgG
fractions by both techniques. The remaining serumshowedreactivityintheIgGfractionsby
the IFAtestbutwasnonreactive inallfractions
bythe IHAtest.Thisserumhadaninitial IFA titerof1:128(serogroup 1) andwasnonreactive by the IHA test.Itappearedthatthe standard
antiglobulin conjugate used in the IFA
tech-nique did not adequately detect IgM. To test
this hypothesis, the fractions of 10 sera which
showed the presence of IgMby theIHAtestbut
notby the IFAtestwerereexamined by the IFA
test with an IgM-specific conjugate. All sera
thenshowedreactivity in the IgM fractions (Ta-ble8).
DISCUSSION
Parallel testing of 895 sera showed that the
IHA and IFA testsyielded essentially the same resultsfor serogroups1to4.There was approx-imately 97% overall agreement between the two techniques,anddiscrepancies were observed in both directions. IHA-positive andIFA-negative
discrepancieswereassociatedwiththeinability
to detect IgM with the antiglobulin conjugate
used in thisstudy.Nagingtonandassociates(11)
reported that 3 of 22 Legionnaires disease pa-tientsshowed arisein IgM antibody withouta
concurrent rise in IgG antibody. Such cases would go undetected with a conjugate of this type. Polyvalent immunoglobulin conjugates
should be usedformaximumsensitivity(11, 14). Thisrequirement posesnoproblem intheIHA technique, asit wasconclusively showntodetect
IgMandIgG antibodies.
The problem ofIFA-positive and IHA-nega-tive sera is yettobe resolved. Onesuch serum
showedreactivityinonlytheIgGfractions. Con-tinued fractionation of sera ofthis naturemay provide an explanation for the differences be-tweenthe two techniques.
Correlation of clinical data with laboratory
dataprovides furtherinsightintothesensitivity
andspecificityof thetwotechniques.Twocases
of Legionnaires disease defined by the IHA which wereundiagnosed by the IFAtechnique hadclinical symptoms consistent with
legionel-losis and responded promptly to erythromycin
therapy. Among six patients diagnosed by the
IFA test but not by IHA, three had clinical
histories consistent with Legionnaires disease
andrespondedtotreatmentwitherythromycin;
one patient had symptoms consistent with the disease but responded well to treatment with
gentamicin and cephalothin, two antibiotics
which are of low efficacy in the treatment of legionellosis. Clinical histories of the two re-maining patients were not consistent with the
diagnosis made by IFA serology. Clinical data
were notavailable onthe remaining
discrepan-ciesdescribedinthis study.
Even when cross absorption was used in ad-ditiontothecruderblocking technique, the IHA test showed no significant cross-reactivity with E. coli. The IFA test showed evidence of cross-reactivity with E. coli in only one instance, but it should be noted that this serum had a low initial titer. Studies with 12 common gram-neg-ative bacterialantigens,including Haemophilus, showed nocross-reactivity in the IFA test with Legionella antigens (B. Wentworth, personal communication). Cherry and
associates
(3) found that of 400 bacterial strains tested, one strain ofPseudomonas fluorescens is antigeni-cally related to L.pneumophila serogroup 1.Although the IHA test showed more cross-reactivity among serogroups than did the IFA
testwith
Formalin-fixed
antigens, heterologous reactivity is usuallysignificantly lower than ho-mologous serogroup reactivity. The IHA test adequately defined the etiological serogroup inon February 7, 2020 by guest
http://jcm.asm.org/
most cases. If therewasanyquestionas tothe
etiological
serogroup, crossabsorption could beusedtodefine it.
Microagglutination andmicro-enzyme-linked
immunosorbentassayshave alsobeendescribed
for the detection of Legionella antibodies (8).
However, microagglutination is less sensitive
thantheIFAtest,and the
micro-enzyme-linked
immunosorbent assaytechnique, although
sen-sitive, ismoreexpensive and difficulttoperform
than theIHA
technique.
The IHAtechnique is rapid,
simple,
inexpen-sive, and well suited to
large-scale
studies androutineuse.Itprovidesasuitable alternativeto
the useof the morecomplicated and expensive
IFAtechnique.
LITERATURE CrIED
1.Bozeman, F.N.,J. W.Humphries,and J. M. Camp-bell. 1968. A new group of rickettsia-like agents re-covered from guineapigs.ActaVirol. (Engl.Ed.) 12: 87-93.
2. Brenner,D.J.,A. G.Steigerwalt,G. W.Gorman,R. E.Weaver, J. C.Feeley, C. Patton,B. M. Thoma-son,andK.R.Lewallen-Sasseville.1980.Legionella bozemaniispeciesnovaandLegionelladumofflispecies nova:classification of two additionalspeciesof
Legion-elaassociated with humanpneumonia. Curr. Micro-biol. 4:115-120.
3. Cherry, W.B.,B.Pittman,P. P.Harris,G. A.Hebert, B.M.Thomason, L.Thacker, and R. E. Weaver. 1978. Detection ofLegionnaires disease bacteria by directinmunofluorescent staining.J.Clin.Microbiol. 8:329-338.
4. Cordes, LG.,H. W.Wilkinson,G. W.Gorman,B. J. Fikes,and D. W. Fraser.1979.Atypical Legionella-like organisms: fastidious water associated bacteria pathogenictoman.Lancetii:927-930.
5. Edson,D.C.,H. E.Stiefel,B. B.Wentworth, andD. L Wilson. 1979. Prevalence of antibodiesto
Legion-naires disease: aseroepidemiologic survey of Michigan residentsusing thehemagglutinationtest. Ann. Intern. Med. 90:691-693.
6.Fahey,J. L, and E. M. McKelvey. 1965. Quantitative determination of serumiemmunoglobulinsin antibody agarplates. J. Immunol. 94:84-90.
7.Fallon, R. J., and W. H. Abraham. 1978. Detecting Legionnaires disease. Lancet ii:1318.
8. Farshy,C. E., G. C. Klein, and J. C. Feeley. 1978. Detection of antibodiesto Legionnairesdisease organ-ismby microagglutination and micro-enzyme-linked im-munosorbent assay tests. J.Clin. Microbiol. 7:327-331. 9. Hebert,G. A., A. G. Steigerwalt, and D. J. Brenner. 1980. Legionella micdadeispecies nova: classification ofa third species of Legionella associated with human pneumonia.Curr. Microbiol.3:255-257.
10. MeDade,J. E., C. C. Shepard,D. W. Fraser, T. F. Tsai, M. A.Redus,W. R.Dowdle, and the Labo-ratoryInvestigation Team. 1977. Legionnaires dis-ease: isolation of a bacteriumanddemonstrationof its role in other respiratory disease. N.Engl. J. Med. 297: 1197-1203.
11. Nagington,J.,and T.G.Wreghitt. 1979. The antibody response in Legionnaires disease. J. Hyg. 83:377-381. 12. Palmer, D. F., J. J. Calavero, K. Herrman, J. A.
Stewart, and K. W. Walls. 1974. Serodiagnosis of: toxoplasmosis, rubella, cytomegalic inclusion disease, herpessimplex, p. 101-110. Immunology Series no. 5. U.S. Department of Health, Education and Welfare, Atlanta, Ga.
13.Tatlock, H. 1944. Arickettsia-like organism recovered from guinea pigs. Proc. Soc. Exp. Biol. Med. 57:95-99. 14. Wilkinson,H. W., C. E.Farshy, B. J. Fikes, D. D.
Cruce,and L P.Yealy. 1979. Measure of immuno-globulin G-, M-, andA-specifictiters against Legionella pneumophila and inhibition of titersagainst nonspe-cific, gram-negative bacterialantigens in the indirect immunofluorescence test forlegionellosis.J.Clin. Mi-crobiol.10:685-689.
15. Wilkinson,H.W.,B. J.Fikes,and D. D.Cruce.1979. Indirect immunofluorescence test forserodiagnosis of Legionnairesdisease: evidence for serogroupdiversity ofLegionnairesdiseasebacterial antigens and for mul-tiple specificity of humanantibodies.J. Clin. Microbiol. 9:379-383.
13,