• No results found

COPPER AND SILVER IONIC WATER PURIFICATION SYSTEM

N/A
N/A
Protected

Academic year: 2021

Share "COPPER AND SILVER IONIC WATER PURIFICATION SYSTEM"

Copied!
12
0
0

Loading.... (view fulltext now)

Full text

(1)

P.O. BOX 703, BIGGERA WATERS, QUEENSLAND, 4214

PHONE: (07) 5574 4700 FAX: (07) 5594 0199

Email: [email protected]

www.aquamatics.com.au

COPPER AND SILVER

IONIC WATER

PURIFICATION SYSTEM

Installation & Running instructions

for the Mark 9 & Mark 10 models

(2)

COPPER & SILVER IONIC WATER PURIFICATION SYSTEM

Installation instructions for the Mk 9 and Mk 10 models.

HOW IT WORKS.

The ionic water purifier consists of two main parts, a control box and pairs of rods or electrodes. These electrodes are specially manufactured from an alloy of pure copper and silver metals. They are installed in a special electrode housing

assembly or “flow cell” according to the diagram and linked together so that they are electrically alternately positive and

negative. They are connected to the control box and to each other via ring terminals.

When water passes between the electrodes an electric current is able to flow through the water and generates positively charged copper and silver ions. Some of these ions do not reach the negatively charged electrodes but are carried into the pool where they are dispersed throughout the water. When Aquabrite is added to the pool these ions become activated and kill the algae and bacteria in the water. Aquabrite slowly releases oxygen into the water over a period of several days.

MOUNTING THE CONTROL CENTRE

The position of the pool pump and the power supply will determine where the control centre may be installed. Use a

vertical surface under suitable cover. The control centre must not be installed outside or in direct sunlight. If this

is impractical, the cable, which supplies the electrodes, may be extended by up to 20 meters using good quality cable. The piggy-back plug must be replaced if an extension cord is used, for safety reasons.The power supply lead to the control centre is fitted with a piggy-back mains plug. The pool pump should be plugged into the back of it for correct operation. The control centre must not be powered unless the pool pump is also running so they must share the same time-clock controlled power supply. The timer built in to the control centre is a count-down timer and is reset each time the filtration system is switched on by the time clock.

For example: When set for 3 HOURS the control centre will operate for 3 hours then shut down for the rest of the

filtration cycle. On commencement of each filter cycle thereafter the operation will be repeated until the setting is changed. SETTING THE CONTROLS

To check the timer setting press the TIME SET button once. The LED opposite one of the time positions will be illuminated. To change the timer setting press the TIME SET button and hold pressed until the LED begins flashing. Release and press

repeatedly until the LED next to the desired time is illuminated. Continuous operation is indicated by all LEDS flashing. The select sequence is “CONTINUOUS” 12 HOURS’ 6 HOURS’ 3 HOURS’ 1 1/2 HOURS’ 3/4 HOUR “CONTINUOUS” etc.

To set the new selected time press “NORMAL’.

CURRENT SET is normally set on 100% unless a unit is shared between two systems such as between a pool and a spa for example.

Test is used to check that normal polarity reversal is taking place across the electrodes. This evens the wear

on the electrodes and keeps them clean of calcium deposits. Press NORMAL to revert to normal display.

NORMAL DISPLAY. (LED = light emitting diode)

Polarity - only one LED on at a time but alternate every few minutes. Continuity - LED flashes approx. once per second when circuit is complete. Conductivity - continuously on when enough ionising current is flowing.

Freshly filled pools may not have enough conductivity and 10 Kg of calcium chloride should be added per 50,000 litres of pool water. The pH should be adjusted between 7.0 and 7.4 for optimum conditions. Aquabrite should be maintained to give a reading greater than 2 on the DPD “chlorine scale,” after 15 minutes waiting time.

(3)
(4)

1. D ril l 8 .5 m m h ol es in o ne e nd -c ap in t he p os iti on s in di ca te d be lo w . 2. In st al l t he e le ct ro de s. ( se e di ag ra m o n th e re ve rs e si de o f th e pa ck ) 3. A ss em bl e th e flo w -c el l ( as b el ow ) al ig n th e tw o en d-ca ps s o th at th e cl ip s lo ts a re in li ne o r th e cl ip s w ill n ot f it pr op er ly . 4. pl ac e a cl ip -( th e en d w ith ou t th e ho le in it )-in to t he c lip s lo t at t he e le ct ro de e nd o f th e flo w -c el l, pl ac e th e cl ip o n fla t su rf ac e an d pr es s do w n so t ha t th e cl ip ‘s na ps ’i nt o po si tio n. r ep ea t th is w ith t he o pp os ite c lip -t he n th e ot he r tw o (e ac h ce ll ha s fo ur c lip s) 5. To r em ov e a cl ip p ut a s cr ew dr iv er t hr ou gh t he h ol e in t he c lip s an d le ve r of f. D o no t us e ex ce ss iv e fo rc e or t he c lip m ay b e da m ag ed . W he n in st al le d ve rt ic al ly t he e le ct ro de t er m in al s sh ou ld be u nd er ne at h to p re ve nt p os si bl e da m ag e du e to flo od in g of t he e nd -c ap in t he e ve nt o f a le ak . F LO W C E LL E N D -C A P E LE C T R O D E P O S IT IO N S 1 P A IR O F E L E C T R O D E S 2 P A IR S O F E L E C T R O D E S C u /A g E L E C T R O D E T he M K 1 0 is s up pl ie d w ith a sp ec ia l d is tr ib ut io n rin g. 4 P A IR S O F E L E C T R O D E S

ve

rt

ic

al

h

o

ri

zo

n

ta

l

A

qu

am

at

ic

s I

on

ic

F

lo

w

-C

el

l

N

E

V

E

R

I

N

S

TA

L

L

T

H

IS

S

Y

S

T

E

M

B

E

T

W

E

E

N

T

H

E

P

U

M

P

A

N

D

F

IL

T

E

R

M

k

5

M

k

9

M

k

10

R IN G T E R M IN A L N U T S B R A S S W A S H E R R U B B E R W A S H E R S F L O W C E L L
(5)

CONTROL CENTRE FEATURES

Displays current flow in AMPS. The current flow is proportional to the release of copper and silver ions into the pool.

Polarity change - these TWO LED indicators reverse every 2 minutes, to show that a polarity change has occurred at the electrodes. Continuity indicates that the electrical circuit between the control centre and the electrodes is complete (flashes once per second) Conductivity indicates that there is sufficient current flowing for normal ionisation (only operates in upper range of current settings)

Over-range indicator is illuminated in the event of a short-circuit of the output terminals or the water conductivity is too high. This may occur if chemicals are added to the pool via the skimmer. The digital display reads zero amps when it is on.

Press once-display lamps indicate current as % of maximum. Press and hold until flashing-display indicates the setting of % current set when

previously programmed normally 100%. Press to revert to NORMAL display after programme adjustment-does this automatically after a few seconds.

Check for normal polarity reversal. The top two polarity lamps alternate for a few seconds. Press once display lamps

indicate time in hours set. Press and hold until flashing display indicates the present setting of time press once to change to a different setting . When all lamps are flashing the unit will run continuously. POLARITY CHANGE CONDUCTIVITY CONTINUITY NORMAL

DISPLAY CURRENTSET TIMESET TEST

11 2 3 Hours 6 Hours 12 Hours 3 4 Hours Hours OVER RANGE 120ma 180ma 60ma 240ma 300ma

Do notadd undiluted chemicals to the pool

Do not use copper-based algicides

Do not use bromine compounds

Do not use aluminium clarifiers

Do not use chlorine stabilisers

pH should not exceed 7.4

Use only Sodium bicarbonate to raise

the pH or total alkalinity

Oxidise the pool every 10 days in the

summer with Aquabrite

Check oxidiser with DPD tablet and

adjust before heavy bathing loads

General Running Instructions

For indoor use only Control Centre SILVER & COPPER IONIC WATER PURIFIER

AMPS

.com.au ®

PROGRAMME & DISPLAY BUTTONS

To programme Current or Time: Press appropriate button for 2 seconds until lamp flashes. Release then press again until required lamp is illuminated.

100% 80% 60% 40% 20%

CONTROL CENTRE FEATURES

Displays current flow in AMPS. The current flow is proportional to the release of copper and silver ions into the pool.

Polarity change - these TWO LED indicators reverse every 2 minutes, to show that a polarity change has occurred at the electrodes.

Continuity indicates that the electrical circuit between the control centre and the electrodes is complete (flashes once per second)

Conductivity indicates that there is sufficient current flowing for normal ionisation (only operates in upper range of current settings)

Over-range indicator is illuminated in the event of a short-circuit of the output terminals or the water conductivity is too high. This may occur if chemicals are added to the pool via the skimmer. The digital display reads zero amps when it is on.

(6)

TO SELECT THE OPERATING TIME CONTROL SETTINGS FOR

THE Mk 9 & Mk 10 AQUAMATICS IONIC WATER PURIFIER

AQUABRITE SYSTEM

1. Estimate the approximate volume of the pool in LITRES. (1 cubic metre is 1000 litres). 2. Measure the current reading on the display in AMPS.

3. Apply the formula below:

THE NUMBER OF HOURS/DAY REQUIRED = 1.5 x Volume of Pool in litres Current reading X 100,000 4. Measure the copper concentration in the pool water using the test kit provided.

If the copper reading is higher than 0.6 ppm (nominally the upper limit required) then select the next LOWER time setting. If the copper reading is lower than any discernable colour then calculate the hours as in 1 to 3 above.

In addition check the pH of the pool water. (See reverse of this sheet for additional information).

The pH should not exceed 7.4.

CHLORINE STABILISER MUST NOT BE ADDED TO AN AQUABRITE POOL.

DO NOT ATTEMPT TO CHANGE OR ROTATE THE ELECTRODES UNLESS THE POWER HAS BEEN DISCONNECTED. THE UNIT HAS AN AUTOMATIC CURRENT LIMIT CIRCUIT WHICH OPERATES IN THE EVENT OF A SHORT CIRCUIT AT THE ELECTRODES. THIS CAN HAPPEN IF CHEMICALS ARE ADDED VIA THE SKIMMER OR THE CONDUCTIVITY OF THE WATER HAS EXCEEDED THE NORMAL OPERATING LIMITS FOR A FRESH-WATER SYSTEM.

AUTO CURRENT LIMIT IS INDICATED ON THE FRONT PANEL BY THE LED NEXT TO THE 3/4 Hr OR 20% SETTING POSITION BEING ILLUMINATED.

THE LCD WILL INDICATE 0.00 AMPS WHILE THE CONDITION EXISTS. REFER ANY TECHNICAL PROBLEMS TO POOL AQUAMATICS

DO NOT ATTEMPT ANY ELECTRICAL REPAIRS.

THERE ARE NO USER - SERVICEABLE PARTS INSIDE

(7)

THE AQUAMATICS AOUABRITE SYSTEM

START -UP AND RUNNING PROCEDURE

START UP (newly filled pool)

I. Fill the pool to the top of the skimmer box then add either I Kg Calcium Hypochlorite or 6L Sodium Hypochlorite.

Then add all chemicals slowly through the skimmer to prevent stains.

2. Adjust the total alkalinity (T.A.) to within the range 80 -120 mg/L or ppm (as CaCO3) *1 (I Kg of bi-carb per day for

every 10,000 litres of pool water) This range should be 60 ppm for painted or fibreglass pools.

3. Adjust the total hardness (T.H.) within the range 100 -250 max mg/L or ppm (as Ca CO3) *2 4. Run the pool pump continuously in Filter mode for 12 hours.

5. Backwash the filter and adjust the water level in the pool to half-way up the skimmer plate. Adjust the pool pH to

within the range 6.8 - 7.0 for fibreglass lining and 7.0 - 7.4 for all others.

6. Add Ionic Starter Kit followed by one Kg. Pack of Aquabrite. Plug in Ionic Water Purification System with the pool

pump plugged into the piggyback plug and operate according to the following instructions.

RUNNING PROCEDURE

• Operate the pool pump for I Hour per day for each 10,000 L (MK5) of water and 20,000 L per hour for the Mk 9. Ensuring that the Ionic Water Purifier can operate for that period of time. This is the summer running time. For winter, use third of the summer settings.

• It Is very important to maintain the pH of the pool water within the range 7.0 -7.4 during this period. If the pH rises above 7.4 dilute acid must be added to the pool to bring the pool water back to the working range pH 7.0 -7.4 (Most Important).

• Measure the copper concentration (using the test kit provided) on a weekly basis. • Measure the Aquabrlte concentration (using a DPD test kit) on weekly basis.

• The demand for Aquabrite varies from pool to pool and it is necessary to measure the Aquabrite in the water by using DPD# I tablet test kit (like measuring chlorine) except that full colour development takes 15 minutes. Maintain a level of Aquabrite, which produces a reading of not less than 2 ppm on the chlorine scale.

• It is advisable to brush the walls to help with the curing of the pool surface. A suction cleaner will

aid this procedure. This applies to surfaces such as pebbles, quartz on and marblesheen. It Is not necessary with tiled, painted or fibreglass as these are less reactive surfaces and do not leach calcium at the same high rate.

Note

From time to time check the condition of the electrodes especially near the end of their expected life. They will form mud like coating on the surface. This Is normal - Do not clean off unless the electrodes have been allowed to dry out or the power supply has been disconnected for several weeks.

*1 To increase T.A. by 100 ppm add 170 grams of sodium bicarbonate (buffer) per 1000L of water **To increase T.H. by 100 p pm add 150 grams of calcium chloride dihydrate per 1000L of water

(8)

POOL AQUAMATICS PTY LTD.

OPERATING INSTRUCTIONS FOR THE AQUABRITE SYSTEM

1. DO NOT ALLOW THE pH OF THE WATER TO RISE ABOVE 7.4. (COPPER WILL NOT BE IN AN ACTIVE FORM)

2. DO NOT THROW UNDISSOLVED CHEMICALS INTO THE POOL. (STAINING OF THE POOL SURFACES

MAY OCCUR)

3. DO NOT USE ANY COMPOUNDS WHICH CONTAIN BROMINE. (BROMINE REMOVES SILVER FROM WATER) 4. DO NOT USE ANY COMPOUNDS WHICH CONTAIN ALUM. (ALUM REMOVES COPPER FROM WATER) 5. DO NOT USE ANY COMPOUNDS WHICH CONTAIN SODA ASH (SODIUM CARBONATE) & NEVER USE CAUSTIC SODA. (DISCOLOURATION OF THE INTERNAL POOL SURFACES MAY OCCUR)

6. CHECK THE CONCENTRATION OF COPPER (USING THE COPPER TEST KIT) ONCE A WEEK AND AFTER

PERIODS OF RAIN.

7. CHECK THE CONCENTRATION OF AQUABRITE (USING THE DPD TEST KIT) - ONCE A WEEK & ADJUST AS NECESSARY TO MAINTAIN THE READING ABOVE 2 ON THE CHLORINE SCALE. (AFTER 15 MINUTES USING A DPD #1 TABLET).

8. MAINTAIN A TOTAL ALKALINITY (T.A.) READING ABOVE 80 P PM BUT NOT EXCEEDING120 P PM. (USE SODIUM BICARBONATE). CHECK MONTHLY.

9. MAINTAIN A CALCIUM HARDNESS (C.H.) LEVEL ABOVE 100 P PM BUT NOT EXCEEDING 250 P PM. (USE CALCIUM CHLORIDE). CHECK MONTHLY.

10. REMOVE RUBBISH FROM THE SKIMMER BOX AND HAIR & LINT FILTERS DAILY & CHECK THE CONDITION OF THE IONIC ELECTRODES MONTHLY.

11. THIS SYSTEM DOES NOT REQUIRE POOL STABILISER. 12. AVOID THE USE OF CHELATING AGENTS.

(9)

ITEM

DESIRED VALUE

pH Paint, Fibreglass 6.8 - 7.0,

Pebble, Tile, Quartzon 7.2 - 7.4 max

• Only correct the pH 24 hours after the T A, (as below) is balanced.

• Add 500g (maximum per day) of dry acid mixed in a plastic watering can, to the surface of the pool with the filter pump running.

• Check the pH the following day (a false reading may be obtained if done on the same day) Fill the large tube with pool water to the pH line on the tube.

Add I drop of solution (antichlor) #4, swirl to mix.

Add 5 drops of solution (phenyl red) #2, again, swirl to mix.

Read the pH by comparing the colour in the tube against the scale of colours representing pH adjacent to the tube.

Aquabrite Pink

• Use the small test tube, fill to the top

• Add a DPD # I tablet and shake to dissolve, then leave it for 15 minutes before reading the colour against the chlorine scale (reacts faster with warmer water at spa water temperature).

• Colour should be pink, not yellow or colourless. If the colour is less than 2 (on the chlorine scale) then add a I Kg pack of AQUABRITE to each 50,000L of pool water.

Copper Test Yellow

• This test is normally only required once a month.

• Collect a sample of pool water in the empty tube of the Aquamatics Copper Test Kit provided. • Add I 0 drops of copper reagent.

• Leave for 2 minutes before reading the colour.

• Check the colour that must not be darker d1an the highest reading on the colour card in the test kit.

• Should the colour be darker than the highest reading on the colour card, the pool is being over-ionised and the ionic water purifier should be turned off until the copper reading drops to within the normal operating range. Measure the copper every week until the system settles down. Generally, operate the Ionic Water Purifier for 1 Hr per day for every 10,000L of water in the pool.

Total Alkalinity (TA) Paint,Fibreglass 60ppm,

Pebble tile QUartz on I00 -120 ppm

• This is a measurement of bicarbonate, (T A), in parts per million (PPM).

• Collect a sample of pool water in the larger test tube and fill to the lower line marked on the tube.

• Add 2 drops of blue dye marked #1, then add dropwise the clear solution marked #3, counting the number of drops it takes until the blue colour disappears (10 -12 drops indicates 100 to 120 PPM).

(10)

POLARITY CHANGE

CONDUCTIVITY CONTINUITY

NORMAL

DISPLAY CURRENTSET TIMESET TEST 11 2 3 Hours 6 Hours 12 Hours 34 Hours Hours OVER RANGE 120ma 180ma 60ma 240ma 300ma

NORMAL DISPLAY

POLARITY CHANGE lamps alternate every 2 minutes. CONTINUITY lamp flashes once per second.

CONDUCTIVITY lamp must be on continuously.

OVER RANGE lamp must not be on unless chemicals have just been added.

WATER CONDUCTIVITY TOO LOW

Check that “current set” is set to maximum 100%. The mineral concentration of the pool water is LOW. Add calcium chloride at 1.5 Kg per 10,000 litres of water. Check the Total Alkalinity and adjust by adding sodium bicarbonate (Buffer), 1Kg/day max. The CONDUCTIVITY lamp should be on after addition of the above chemicals.

OR

OR

POLARITY CHANGE CONDUCTIVITY CONTINUITY NORMAL

DISPLAY CURRENTSET TIMESET TEST 11 2 3 Hours 6 Hours 12 Hours 34 Hours Hours

T P CURRENT TIME P i t b tt f 2 d til

OVER RANGE 120ma 180ma 60ma 240ma 300ma POLARITY CHANGE CONDUCTIVITY CONTINUITY NORMAL

DISPLAY CURRENTSET TIMESET TEST 11 2 3 Hours 6 Hours 12 Hours 34 Hours Hours OVER RANGE 120ma 180ma 60ma 240ma 300ma POLARITY CHANGE CONDUCTIVITY CONTINUITY NORMAL

DISPLAY CURRENTSET TIMESET TEST 11 2 3 Hours 6 Hours 12 Hours 34 Hours Hours OVER RANGE 120ma 180ma 60ma 240ma 300ma

To programme Current or Time: Press appropriate button for 2 seconds until lamp flashes. Release then press again until required lamp is illuminated.

To programme Current or Time: Press appropriate button for 2 seconds until lamp flashes. Release then press again until required lamp is illuminated.

To programme Current or Time: Press appropriate button for 2 seconds until lamp flashes. Release then press again until required lamp is illuminated.

To programme Current or Time: Press appropriate button for 2 seconds until lamp flashes. Release then press again until required lamp is illuminated. 100% 80% 60% 40% 20% 100% 80% 60% 40% 20% 100% 80% 60% 40% 20% 100% 80% 60% 40% 20%

(11)

POLARITY CHANGE

CONDUCTIVITY CONTINUITY

NORMAL

DISPLAY CURRENTSET TIMESET TEST 11 2 3 Hours 6 Hours 12 Hours 34 Hours Hours OVER RANGE 120ma 180ma 60ma 240ma 300ma

CONDUCTIVITY TOO HIGH

The mineral content of the water is too high.

a) If the chemicals have just been added to the pool via the skimmer then the above display(s) is normal.

Check again after a few minutes.

b) Drain some water from the pool and replace it with fresh water until OVER RANGE lamp turns off.

CIRCUIT NOT CONTINUOUS

CONTINUITY and CONDUCTIVITY lamp off. Check low voltage cables and

connectors.Check there is water between electrodes.Electrodes may need replacing.

POLARITY CHANGE

CONDUCTIVITY CONTINUITY

NORMAL

DISPLAY CURRENTSET TIMESET TEST 11 2 3 Hours 6 Hours 12 Hours 34 Hours Hours

T P CURRENT TIME P i t b tt f 2 d til

OVER RANGE 120ma 180ma 60ma 240ma 300ma POLARITY CHANGE CONDUCTIVITY CONTINUITY NORMAL

DISPLAY CURRENTSET TIMESET TEST 11 2 3 Hours 6 Hours 12 Hours 34 Hours Hours OVER RANGE 120ma 180ma 60ma 240ma 300ma POLARITY CHANGE CONDUCTIVITY CONTINUITY NORMAL

DISPLAY CURRENTSET TIMESET TEST 11 2 3 Hours 6 Hours 12 Hours 34 Hours Hours OVER RANGE 120ma 180ma 60ma 240ma 300ma

To programme Current or Time: Press appropriate button for 2 seconds until lamp flashes. Release then press again until required lamp is illuminated.

To programme Current or Time: Press appropriate button for 2 seconds until lamp flashes. Release then press again until required lamp is illuminated.

To programme Current or Time: Press appropriate button for 2 seconds until lamp flashes. Release then press again until required lamp is illuminated.

To programme Current or Time: Press appropriate button for 2 seconds until lamp flashes. Release then press again until required lamp is illuminated.

OR

OR

100% 80% 60% 40% 20% 100% 80% 60% 40% 20% 100% 80% 60% 40% 20% 100% 80% 60% 40% 20%
(12)

References

Related documents

The University of North Carolina at Chapel Hill ("UNC-Chapel Hill") has moved to intervene in the above-captioned docket and files these Initial Comments, through its

This study was conducted to monitor the changes in the fatty acid composition of selected oils (i.e., RBD palm olein, canola, and sesame) and their combinations under

1) Press the RESET button on projector for 3 seconds or remote control TIMER button for 3 seconds and display the total lamp used timeb. • When accumulated operation time of the

2 Turn the headset on (press and hold the power button until the indicator LED flashes red/blue) The headset is now in..

Press and hold the Clock Adjust Button located on the back of the unit until the current time flashes on the display.. Press the or Button to adjust to correct time

Press and hold button 1 and 2 at the same time for a few seconds until alarm chirps 3 times and flashes lights 3 times to confirm. Select number of flashes

To program Automatic Passive Arming with Door Locks On, disarm the system, press and hold Button #3 for 10 seconds, until the unit chirps once. Release Button #3 then press and

Press the Tuning up/down button until ‘System’ shown on the display, then press the Select button to enter the setting.. Press the Tuning up/down button until ‘Time’ shown on