0095-1137/89/040716-05$02.00/0
CopyrightC 1989, AmericanSocietyfor Microbiology
Rapid
Bioluminescence Method for
Bacteriuria
Screening
MARIE T. PEZZLO,* VALERIE IGE, AMELIA P. WOOLARD, ELLENA M. PETERSON,
AND LUISM. DE LAMAZA
DepartmentofPathology, Division ofMedical Microbiology, University of California Irvine Medical Center,
101 City Drive, Orange, California 92668 Received 18August 1988/Accepted 3 January1989
Astudywasperformedtoevaluate the UTIscreen (Los AlamosDiagnostics, LosAlamos, N. Mex.),arapid
bioluminescencebacteriuriascreen. TheUTIscreenwascomparedwith threeotherrapidbacteriuriascreens: the Bac-T-Screen (Vitek Systems, Hazelwood, Mo.), an automated filtration device; the Chemstrip LN
(Boehringer Mannheim Diagnostics, BioDynamics, Indianapolis, Ind.), an enzyme dipstick; and the Gram stain. Asemiquantitative plate culturewasused asthereferencemethod. Ofthe 1,000 specimens tested,276
had colonycountsof >105CFU/ml bythe culturemethod. Of these, theUTIscreen detected 96% (265of276) using .5% ofthe integrated lightoutputofthe standard readingas a positive interpretive breakpoint, the
Bac-T-Screen detected 96% (266 of 276), the ChemstripLNdetected 90% (249of276), and the Gram stain detected 96% (264 of 276). Ofthe 214probable pathogensisolatedat >105 CFU/mI, theUTIscreendetected 95% (204 of 214),the Bac-T-Screendetected98%(210of214),theChemstripLNdetected 92% (198of214), and the Gramstain detected 98% (209of214). Thepredictivevalues ofnegativetestresultsat>105 CFU/ml for the UTIscreen, the Bac-T-Screen, the Chemstrip LN, and the Gram stain were 98, 97, 93, and 98%,
respectively. The overallspecificitiesat>105CFU/mlfortheUTIscreen,theBac-T-Screen,theChemstrip LN, and the Gram stainwere70, 48, 51,and69%, respectively.Therewere532specimenswithcolonycountsof >103 CFU/ml, and of these, the UTIscreen,theBac-T-Screen, theChemstrip LN, and the Gram staindetected 72, 81, 76,and73%,respectively. Ofthe 249probable pathogensisolatedat>103CFU/ml,theUTIscreen,the Bac-T-Screen,theChemstrip LN,and the Gram stain detected91, 95, 89,and93%, respectively. Theoverall specificities at >103CFU/ml forthese methodswere79, 55, 57, and78%, respectively. Thecostpertestfor detection was approximately $1.00to $1.20 for the UTIscreen, the Bac-T-Screen, and the Gram stain and approximately $0.50for theChemstripLN. Overall,theUTIscreenisrapidandeasytoperform;itssensitivity compared favorably with thoseof otherscreening methods;ithadahigher specificity thanthe Bac-T-Screen andChemstrip LN; and it allowed forbatchingofspecimens.
Urine specimens represent the majority of samples
re-ceived in the clinical microbiology laboratory for culture. Thelaboratoryfaces thechallenge of rapidly identifying both positive and negative specimens. Although the semiquanti-tative plate culture method allows for the isolation and enumeration ofmost infectious agents, it does not provide for same-day reporting of negative specimens. For this
reason, rapid urine screening tests have been developed.
These tests notonlyprovideforrapid reporting of negative specimensbutalsohavethepotential of reducingthecostof patientmanagement.
A numberofrapidurine screenshavebeendescribed(12). Theseincludemicroscopic, enzymatic, filtration, and photo-metric methods. The most commonly used microscopic method is the Gram stain. As a urine screen, it is rapid,
reliable, and correlates with colony countsof>105CFU/ml (13, 22, 23). However, becausethe acceptable sensitivity of thismethod isat105CFU/ml, low-level bacteriuriamaynot bedetected. Furthermore,theaccuracyisgreatly dependent on theexpertise of the reader. Enzyme dipsticks have also been usedasrapid bacteriuriascreens. Although thesetests
are easy to perform, the overall sensitivity of these rapid
enzymedipsticksistoolow(.90%)tobe usedaloneasurine
screens (15, 25). The first generation of semiautomated bacteriuria screens include photometric methods which
re-quire growth of the organism for detection; therefore test
results are delayed (1 to 13 h). The second generation of
* Correspondingauthor.
semiautomated urine screens include bioluminescence and
filtration methods. These systems are more rapid (1 to 15 min)than thegrowthdetection methodsand results of both
compare favorably(2, 5, 6, 20, 21, 24, 25).
The purpose of thisinvestigation wastoevaluatea
biolu-minescencemethod, theUTIscreen, andtocompareit with other rapidbacteriuriascreens: theBac-T-Screen, an auto-mated filtration method; the Chemstrip LN, an enzyme
dipstick method; and the Gram stain. In this study, these methods were evaluated at various colony counts in an
attempt to aid others in the selection ofa laboratory
ap-proach tourine screening.
MATERIALS ANDMETHODS
Specimens. Atotal of1,000 specimens which included 688 clean-voided and 312 catheterized urine specimens from both inpatients and outpatients submitted to the Medical Microbiology Laboratory at the University of California Irvine Medical Centerwere included in thisstudy. Patients receiving antimicrobial therapy were not excluded. Col-lected urinewas placed in asterile tube, refrigerated (4°C),
and processed within 8 h ofcollection.
Semiquantitative culture. A semiquantitative plate count
asdescribedby Clarridgeetal. (3)wasusedasthe reference
method. By using a calibrated platinum loop, a 0.001-ml sampleofawell-mixed urine specimenwasinoculated onto
a 5% sheep blood agar plate (BBL Microbiology Systems,
Cockeysville, Md.) and a biplate consisting ofMacConkey
agarand polymyxin B-nalidixic acid blood agar (Calscott,
Inc., Carson, Calif.). An additional 0.1 ml of urine was
716
on April 11, 2020 by guest
http://jcm.asm.org/
TABLE 1. Number and percentage of positive test results
Culture (CFU/ No.(%)positiveresultsby:
ml) (no.of
specimens) UTIscreen Bac-T-Screen ChemstripLN Gram stain
>105(276) 265 (96) 266 (96) 249 (90) 264(96) 1031-i1 (256) 116 (45) 164 (64) 156 (61) 124(48) 10'-103(175) 56 (32) 87 (50) 77 (44) 46(26) <10'(293) 44(15) 124 (42) 125 (43) 58(20)
inoculated ontoa5% sheepbloodagarplate. Cultureswere
incubated overnight at 35°C and examined for the number andtypes of organisms present. Organismsconsidered
con-taminants were diphtheroids, lactobacilli, viridans group
streptococci other than group D, and mixed cultures from
voided urine specimens.
UTIscreen. Urine specimenswere processed accordingto theinstructions of themanufacturer (Los Alamos Diagnos-tics, Los Alamos, N. Mex.). A sample of well-mixed urine (0.025 ml) was added to a tube containing dehydrated
somatic cell releasing agent. Additionally, 0.025 ml of the ATPstandard wasaddedtoaclean polystyrene tube (12 by
50 mm). The tubes were incubated for 15 min at room
temperature. Each tube was then placed into the specimen
well of the Luminometer 535 (Los Alamos Diagnostics), and bothluciferin-luciferasereagentandbacterial releasingagent
were added automatically. The integrated light output was
displayedonthe instrumentfront panelandalsorecordedas
partof theautomatic sequence. Inthisstudy,thedatawere
analyzed using 25% integrated lightoutput of the standard
as apositive test.
Bac-T-Screen. Urine specimens were processed with the
Bac-T-Screenaccordingto theinstructions ofthe manufac-turer (Vitek Systems, Hazelwood, Mo.). A 1-ml sample of well-mixed urine was added to the active barrel of the instrument.Thereagents,3 mlof urine diluent(14.5% acetic acid), 3ml of safraninOdye,andtwoadditions each of 3ml
ofdecolorizer(2.4%aceticacid),wereaddedautomatically. The filtercard was removed from the instrument whenthe
testcyclewas completed and placed in theDynadepthtest cardreader(VitekSystems). Apositivetestwasinterpreted as .4U above the negative control.
ChemstripLN.TheChemstripLN(BoehringerMannheim Diagnostics, BioDynamics, Indianapolis, Ind.) is a plastic
striptowhichareattachedreagentpapersforindicatingthe
presenceofleukocyteesteraseand nitrite in urine. The urine
specimenwasallowedtocometoroomtemperature priorto testing, and the plastic strip was dipped into the specimen
and immediately withdrawn to remove the excess urine.
Resultswerereadafter 2 min. Thecolor intensity ofthestrip wascoded withacolorguide provided by the manufacturer.
A positive leukocyte esterase was one that gave a purple
ranging fromatrace toa2+ intensity. Fornitrite, anypink was considered positive.
Gram stain. A sample (0.01 ml) ofwell-mixed uncentri-fuged urinewasGramstained and examinedfor thepresence
orabsenceofleukocytes andmicroorganisms. Thecriterion
fora positive Gram stain was the presenceofone ormore
bacterialcellsandone or moreleukocytesperoilimmersion field, which has been reportedtocorrelate with-10' CFU/ ml and pyuria, respectively (3).
Predictivevalue. Predictive valueswerecalculated by the method of Ransohoff and Feinstein (17). The sensitivity, specificity, and predictive values ofpositive and negative testresults werecalculatedasfollows: sensitivity = TP/(TP
+ FN); specificity = TN/(TN + FP); predictive value ofa
positive test = TP/(TP + FP); and predictive value of a
negative test = TN/(TN + FN), where TPistrue-positive,
TN is true-negative, FP is false-positive, and FN is false-negative.
Time and cost analysis. An analysis ofcost per test was
doneby calculatingthecostof materials routinely used and technical time incurred in ourlaboratory. The average cost
perspecimenwas calculated by includingboth positive and
negative screen results. The cost of detection included materialcostsandtechnicaltime. Technicaltimewas
calcu-lated based on current College of American Pathologists workload units (4) when available or the average time required toprocess 20specimens by atest method.
RESULTS
Distribution oftestresults. A total of 1,000 clean-voided and catheterized urine specimens were evaluated. There were 276 (27.6%) specimens with colony counts of >10' CFU/ml by the standard semiquantitative plate culture method. Of these, the UTIscreen detected 265 (96%)
com-paredwith 266(96%) bytheBac-T-Screen,249(90%) bythe Chemstrip LN,and 264 (96%) by the Gram stain(Table 1). Overall,226(82%)weredetectedbyall methods,andatleast
onescreening methodwas positive for each specimen. There were 256(25.6%) specimens with colonycountsof 103 to 105 CFU/ml. Of these, the UTIscreen detected 116 (45%), while the Bac-T-Screen, Chemstrip LN, and Gram stain detected 164(64%), 156(61%), and 124(48%),
respec-tively (Table 1). The remaining 468 specimens had <103 CFU/ml; of these, 175 had colony counts of 101 to 103 CFU/ml,and theremaining293specimenshadnodetectable
growth(<101CFU/ml). Thedetectionratesbyeachmethod for these levels of bacteriuriaare shown in Table 1.
The sensitivities, specificities, and predictive values for thetestmethodsatvariouscolonycountsareshown inTable 2. Inadditiontoeach methodhavingasensitivity ofatleast
90%at >105 CFU/mI, thepredictive values ofnegative test
results rangedfrom93 to98%. When the interpretive
break-TABLE 2. Sensitivities, specificities, andpredictivevaluesfortestmethods Results(%)with method atcolonycount
Parameter UTIscreen Bac-T-Screen ChemstripLN Gramstain
>10 >103 >101 >105 >103 >lo0 >105 >103 >101 >105 >103 >101
Sensitivity 96 72 62 96 81 73 90 76 68 96 73 61
Specificity 70 79 85 48 55 58 51 57 57 69 78 80
Predictive value
Positive 55 79 91 42 67 81 40 67 79 54 79 88
Negative 98 71 48 97 72 47 93 68 43 98 72 46
on April 11, 2020 by guest
http://jcm.asm.org/
TABLE 3. Probablepathogens at >105 CFU/ml detected by the test methods
No.of isolates detected by': Organism (no.ofisolates)
UTIscreen Bac-T-Screen ChemstripLN Gramstain
Escherichiacoli(108) 107 108 105 105
Candidaspp.(21) 18 20 18 21
Klebsiella spp. (19) 19 19 15 19
Enterococcus spp. (20) 17 19 19 19
Pseudomonas spp. (10) 8 8 7 9
Proteus mirabilis (11) 10 il 10 il
Enterobacter spp. (5) 5 5 5 5
Coagulase-negative 7 7 7 7
staphylococci (7)
Staphylococcus aureus (5) 5 5 4 5
Streptococcus agalactiae (3) 3 3 3 3
Serratiamarcescens (2) 2 2 2 2
Providencia stuartii (2) 2 2 2
Citrobacterspp. (1) 1 1 1 1
" The percentages of all isolatesdetected were asfollows: UTlscreen,95:Bac-T-Screen,98:Chemstrip LN,92; andGramstain,98.
pointwasdecreased for all organisms to >103 CFU/ml, the Chemstrip LN. The sensitivities for all methods at >101
sensitivities forthetest methods ranged from 81 to72%and CFU/ml rangedfrom84 to91%.
from 73 to 61% at >101 CFU/ml. The specificities and The UTIscreen and the Gram stain hadthehighest
spec-predictive values of positive test results were lower for the ificities for all organisms, which ranged from69% at
>105
Bac-T-Screen and the Chemstrip LN because of the high CFU/ml to85%at
>101
CFU/ml(Table 2). The specificitiesfalse-positiveresults by these methods. for theBac-T-Screen and the Chemstrip LN ranged from 48
Overall, there were 272probable pathogens isolatedfrom to 58% for colony counts of
>105
to>10'
CFU/ml. The157clean-voidedspecimens and 103 catheterizedspecimens. predictive value of a positive test increased by
approxi-Of these, 248 specimens had pure cultures of probable mately40%for each test method as the interpretive break-pathogens and the remaining 12 were catheterizedspecimens point of thereference method decreased from >105 to>101
with two probable pathogens. The most frequently isolated
CFU/ml,
whereas thepredictive
value of a negative testprobable pathogens included Escherichia coli (n = 130), decreased (Table 2).
Candida spp. (n = 30), Enterococcus spp. (n = 21), Klebsi- Costanalysis.The totalcostof
screening by
eachmethodellapneumoniae (n = 21),Pseudomonas spp. (n = 19), and was determined (Table 4). Although salaries differ from Proteus mirabilis (n = 16),accounting for themajority (79%) laboratory to laboratory, the cost per test was based on the
of isolates in this study. Of these, Pseudomonas spp., average salary ofthe technical personnelatour institution.
Enterococcus spp., and Candida spp. had the highest per- For
detection,
thecostof the testmethods rangedfrom$0.44centage offalse-negative results for all the test methods. fortheChemstripLN to
$1.17
fortheGram staincomparedThere were 214probable pathogens with colony counts of with
$1.63
for thereferenceculture method.However, when>105
CFU/ml, 35 at103
to105
CFU/ml, and 23 at101
to103
the average cost for both positive and negative specimensCFU/ml. wascalculated, the testmethods were more
expensive
thanOf the 214 probable pathogens at
>105
CFU/ml, the the reference method, ranging from$0.12
higher for theBac-T-Screen and the Gram stain detected 98%, the UTI UTIscreen to $0.49 higher for the Bac-T-Screen, with one
screen detected95%, and theChemstrip LN detected92%, exception, theChemstrip LN, whichwas
$0.20
lower.respectively (Table 3). There was no statistically significant
difference fordetection ofanyoftheprobable pathogens by DISCUSSION
the testmethods(P>0.1). When theinterpretive breakpoint
wasdecreased forprobable pathogensto
>103
CFU/ml,
the The purpose of urine screening is to improve patientsensitivities for the UTIscreen, the Bac-T-Screen, and the managementby rapidly reporting negativeresults and
iden-Gram stain were at least
90%
compared with85%
for the tifying those specimens which do not warrant culture.Cri-TABLE 4. Costanalysisofrapidurine screens at>105CFU/ml
Cost($)/specimen by:
Determination Chemstrip
Culture method UTIscreen Bac-T-Screen LN Gramstain
Costof supplies 0.75a 0.75 0.75 0.20 0.05
Costoftechnical time" 0.88 0.22 0.33 0.22 1.12
Cost ofdetection 1.63 0.97 1.08 0.44 1.17
Avgcost/specimen, including 1.63 1.75 2.12 1.43 1.97
positiveandnegative screenresult
Difference fromculture +0.12 +0.49 -0.20 +0.34
a Includes5%sheepblood agar andbiplate of MacConkeyagarandpolymyxin-B-nalidixic acid bloodagar.
bEstimatedat$13.00/hassuming 4.0, 1.0, 1.5, 1.0,and 5.1 minpersampleforculture,UTIscreen,Bac-T-Screen,Chemstrip LN,and Gramstain,respectively.
on April 11, 2020 by guest
http://jcm.asm.org/
teria for
selection
of a rapid urine screen should include accuracy, reproducibility,detection
time, ease of test per-formance, and low cost. Previousevaluations
ofmicroscopicand filtration methods have been favorable regarding sensi-tivities,
predictive
values of negative test results, and detec-tion times (10, 13, 14, 22, 23).During recent years, the criterion of
2105
CFU/ml estab-lished by Kass (7) as thedefinition
of a urinary tract infection has been challenged. Studies by Latham etal.(8) and Stamm etal.
(18) found.102
CFU/ml to be a better predictor of infection in symptomatic women. Lipsky et al. (9) and Musher et al. (11) recommended that2103
CFU/ml be used when evaluating voided urine specimens from men. Stark and Maki (19) found that.102
CFU/ml was a more valid index when patients were catheterized and had urinary symptoms or were immunosuppressed. Pfaller et al. (16) studied a random patient population and also found that lower count cultures most accurately identified urine speci-mens from infected patients. Because of the findings of these investigators, the rapid screens described here were evalu-ated at various colony counts.Previous
evaluations employing
bioluminescence assays for urine screening have demonstrated sensitivities of.85%at
>i04
CFU/ml,
withpredictive
values of negative testresults
of -95% (2, 5, 6, 20, 21, 24, 25). The overallsensitivityat
>105
CFU/ml
for the UTIscreen in this studywas 96% when a
25%
integrated light output was used as the positive breakpoint. At this interpretive breakpoint, theUTIscreen
was as sensitive as the Bac-T-Screen and the Gram stain.All
three methods had sensitivities of at least90% at
103
CFU/ml
for probable pathogens and 84% at 10'CFU/ml.
The UTIscreen missed Candida spp. and Enterococcus
spp. more frequently than the Bac-T-Screen and the Gram stain did. A possible explanation may be that some strains of bacteria and yeast had lower ATP levels in spite of high colony counts (21). Another explanation may be that the bacterial releasing agent did not sufficiently lyse the cell walls of these gram-positive organisms (24). These factors may have contributed to a lower UTIscreen sensitivity for these species, which were detected by the other screening methods.
The major
difference
between the semiautomated meth-ods, theUTIscreenand the Bac-T-Screen, is in the specific-ities of these methods. The specificity of the UTIscreen at>105
CFU/ml was higher than that of the Bac-T-Screen (70 versus 48%); however, the sensitivities were the same (96%). Although the specificities were increased in all in-stances by lowering the interpretive breakpoint to either103
or10'
CFU/ml,
the specificity of the Bac-T-Screen remained lower than that of theUTIscreen. A possible explanation for decreased specificities by the test methods is that low-level bacteriuria may not be detected by the culture method. Also, it has been reported that the culture method may have an error rate of as high as 50%(1). Another contributing factor may have been the presence of antimicrobial agents which inhibit microbial growth but do not inhibit detection by bioluminescence or filtration (6). The Bac-T-Screen detects leukocytes as well as bacteria (14). The trapping of leuko-cytes along with other cells (i.e., squamous epithelial cells) by the filter card probably accounts for the lower Bac-T-Screen specificity. This may also account for lower spec-ificities by the UTIscreen at the higher colony counts. If the somatic releasing agent does not destroy all of the somaticcell ATP, it will be detected by the luminometer. Macro-scopically bloody urine specimens have been reported to be
positive by bioluminescence (2). The specificities of the
UTIscreen
atvarious colonycounts are acceptable,consid-ering that the main purpose of a urine screen is to rapidly
identify
negative
specimens whilereliably detectingpositivespecimens.
In this study, the UTIscreen results were similar to the
Gram stain
results
regarding sensitivity, specificity, andpredictive
values at various colony counts. Although the Gram stain comparesfavorably with other urine screens andtest interpretation is subjective, it requires more labor time
to perform than the UTIscreen, therefore resulting in a higher cost per test. In addition, it can be a tedious proce-dure because mosturine specimens are negative. However, it can give a preliminary tentative identification of the
organism group and the presence ofleukocytes.
Both the UTIscreenand the Bac-T-Screen employ instru-mentation; therefore, interpretation of test results is objec-tive, whereas the Gram stain interpretation is subjective, requinng technical expertise. Another difference between the instrument methods is the volume of urine required to perform the tests; 0.025mlfor the UTIscreen compared with 1.0 ml for the Bac-T-Screen.
The ease of performance is an additional advantageof the
UTIscreen.
Once the 0.025 ml of urine is added to the tubecontaining the dehydrated somatic cell releasing agent and
the tube is allowed to incubate at room temperature for 10 min, technical time required to complete the test is minimal (15 s). Reagents are added and results are printed
automat-ically. Specimens can be easily batched, whereas the other
methods require individual handling. Cost is an important
consideration and the UTIscreen compared closely with the
culture method in this study when both positive and negative test results were considered.
In conclusion, the UTIscreen compared favorably with otherscreening methods with respect to sensitivity, predic-tive value of a negapredic-tive test, ease ofperformance, and cost. The advantages of this rapid urine screen include theability to batch test, objectiveinterpretation, higher specificity than the enzyme dipstick or the colorimetric filtration methods, and same-day results with cost similar to culture.
LITERATURE CITED
1. Albers, A. C., and R. D. Fletcher. 1983. Accuracy of calibrated-loop transfer. J. Clin. Microbiol. 18:40-42.
2. Bixier-Forell, E., M. A. Bertram, and D. A. Bruckner. 1985. Clinicalevaluation of three rapid methods for the detectionof significant bacteriuria. J. Clin. Microbiol. 22:62-67.
3. Clarridge, J. E., M. T. Pezzlo, andK. L. Vosti. 1987. Cumitech 2A, Laboratory diagnosis of urinary tractinfections. Coordinat-ing ed., A. L. Weissfeld. American Society for Microbiology, Washington, D.C.
4. College of American Pathologists WorkloadRecording Commit-tee. 1987. Manual for laboratory workload recording method, 1988 ed., p. 111-125. College of American Pathologists, Traverse City, Mich.
5. Drow, D. L., C. H. Baum, and G.Hirschfield. 1984. Comparison of the Lumac and Monolightsystems for detectionofbacteriuria by bioluminescence. J. Clin. Microbiol. 20:797-801.
6. Johnston, H. H., C. J. Mitchell, and G. D. W. Curtis. 1976. An automated test for thedetection of significant bacteriuria. Lan-cet ii:400-402.
7. Kass, E. H. 1956. Asymptomatic infections of theurinarytract. Trans. Assoc. Am. Phys. 69:56-63.
8. Latham, R. H., E. S. Wong, A. Larson, M. Coyle, and W. E. Stamm. 1985. Laboratory diagnosisof urinary tract infection in ambulatory women. J. Am. Med. Assoc. 254:3333-3335. 9. Lipsky, B. A., R. C. Ireton, S. D.Fihn, R. Hackett, and R. E.
Berger. 1987. Diagnosis of bacteriuria in men: specimen collec-tion and culture interpretation. J. Infect. Dis. 155:847-854.
on April 11, 2020 by guest
http://jcm.asm.org/
10. Murray, P. R., R. B. Smith, andT. C. McKinney. 1987. Clinical evaluation of three urine screening tests. J. Clin. Microbiol. 25:467-470.
11. Musher,D. M., S. B. Thorsteinsson, and V. M. Andriole. 1976. Quantitative urinalysis: diagnosing urinary tract infection in men. J. Am. Med. Assoc. 236:2069-2072.
12. Pezzlo,M. T. 1988. Detection ofurinary tract infections by rapid methods. Clin. Microbiol. Rev. 1:268-280.
13. Pezzlo,M. T., G. L. Tan, E. M. Peterson, and L. M. de la Maza. 1982.Screening of urine cultures by three automated systems. J. Clin. Microbiol. 15:468-474.
14. Pezzlo, M. T., M. A. Wetkowski, E. M. Peterson, and L. M. de la Maza. 1985. Detection ofbacteriuria and pyuria within two minutes. J.Clin. Microbiol. 21:578-581.
15. Pfaller, M. A., and F. P. Koontz.1985.Laboratoryevaluationof leukocyte esteraseand nitrite tests for thedetectionof bacteri-uria. J. Clin. Microbiol. 21:840-842.
16. Pfaller,M. A., B. Ringenberg, L. Rames, J. Hegeman, and F. Koontz. 1987. The usefulness of screening tests for pyuria in combination with culture in the diagnosis of urinary tract infection. Diagn. Microbiol. Infect. Dis.6:207-215.
17. Ransohoff, D. F., and A. R. Feinstein. 1979. Problems of spectrumbias in evaluating theefficiencyofdiagnostictests. N. Engl. J.Med. 299:926-930.
18. Stamm, W. E., G. W. Counts, K. R. Running, S. Fihn, M.
Turck, and K. K. Holmes. 1982. Diagnosis of coliform infection in acutedysuric women. N. Engl. J. Med. 307:463-468. 19. Stark, R. P., and D. G. Maki. 1984. Bacteriuria in the
catheter-ized patient: what quantitative level of bacteriuria is relevant? N. Engl. J. Med. 311:560-564.
20. Thore, A., A. Lundin, and S. Ansehn. 1983. Firefly luciferase ATP assay as a screening method for bacteriuria. J. Clin. Microbiol. 17:218-224.
21. Thore, A., S. A. A. Lundin, and S. Bergman. 1975. Detection of bacteriuria by luciferase assay of adenosine triphosphate. J. Clin. Microbiol. 1:1-8.
22. Tilton, R. E., and R. C. Tilton. 1980. Automated direct antimi-crobialsusceptibility testingofmicroscopically screened urine cultures. J. Clin. Microbiol. 11:157-161.
23. Washington, J. A., II, C. M. White, M. Laganiere, and L. H. Smith. 1981.Detection ofsignificant bacteriuria by microscopic examination of urine. Lab. Med. 12:294-296.
24. Welch, W. D., L. Thompson, M. Layman, and P. M. Southern, Jr. 1984. Evaluationof twobioluminescence-measuring instru-ments, the Turner design and Lumac systems, for the rapid screening of urinespecimens.J. Clin. Microbiol.20:1165-1170. 25. Wu, T. C., E. C. Williams, S. Y. Koo, and J. D. MacLowry. 1985. Evaluation ofthree bacteriuria screening methods in a
clinical research hospital.J. Clin. Microbiol. 21:796-799.