• No results found

Title - Sujet Automated External Defibrillators. Solicitation No. - N de l'invitation 0X /A. Client Reference No. - N de référence du client

N/A
N/A
Protected

Academic year: 2021

Share "Title - Sujet Automated External Defibrillators. Solicitation No. - N de l'invitation 0X /A. Client Reference No. - N de référence du client"

Copied!
7
0
0

Loading.... (view fulltext now)

Full text

(1)

RETOURNER LES SOUMISSIONS À:

Bid Receiving - PWGSC / Réception des soumissions - TPSGC

11 Laurier St./ 11 rue, Laurier Place du Portage, Phase III Core 0B2 / Noyau 0B2 Gatineau, Québec K1A 0S5 Bid Fax: (819) 997-9776

CCC No./N° CCC - FMS No./N° VME

SOLICITATION AMENDMENT

Time Zone

MODIFICATION DE L'INVITATION

02:00 PM

2016-02-02

Fuseau horaire

Eastern Standard Time EST

Destination: Other-Autre:

FAX No. - N° de FAX

(819) 956-3814

Issuing Office - Bureau de distribution

Scientific, Medical and Photographic Division / Division de l'équipement scientifique, des produits photographiques et pharmaceutiques

11 Laurier St./ 11 rue, Laurier 6B1, Place du Portage Gatineau, Québec K1A 0S5

indicated, all other terms and conditions of the Solicitation The referenced document is hereby revised; unless otherwise remain the same.

les modalités de l'invitation demeurent les mêmes.

Ce document est par la présente révisé; sauf indication contraire,

Instructions: Voir aux présentes Instructions: See Herein

Delivery Required - Livraison exigée Delivery Offered - Livraison proposée Vendor/Firm Name and Address

Comments - Commentaires

Raison sociale et adresse du fournisseur/de l'entrepreneur

Automated External Defibrillators

Solicitation No. - N° de l'invitation

0X001-150527/A

Client Reference No. - N° de référence du client

0X001-150527

GETS Reference No. - N° de référence de SEAG

PW-$$PV-899-68710

File No. - N° de dossier

pv899.0X001-150527

Solicitation Closes - L'invitation prend fin

at - à

on - le

F.O.B. - F.A.B. Plant-Usine:

Address Enquiries to: - Adresser toutes questions à:

Gauthier, Martin

Telephone No. - N° de téléphone

(613) 404-8642 ( )

Destination - of Goods, Services, and Construction: Destination - des biens, services et construction:

pv899

Buyer Id - Id de l'acheteur

Vendor/Firm Name and Address

Raison sociale et adresse du fournisseur/de l'entrepreneur

Facsimile No. - N° de télécopieur Telephone No. - N° de téléphone

Name and title of person authorized to sign on behalf of Vendor/Firm (type or print)

Nom et titre de la personne autorisée à signer au nom du fournisseur/ de l'entrepreneur (taper ou écrire en caractères d'imprimerie)

Signature Date

2016-01-14

Date

001

(2)

This amendment is raised for: A. Questions and Answers

Question no.1:

It’s my understanding that if the AED is deployed, it would most likely be used by a minimally trained

(BLS) or untrained responder. Is it fair to interpret that the customer is requesting a Public Access /

Regular workplace AED rather than a “professional rescuers AED” intended for use by minimally trained

responders or the lay-person?

Answer no.1:

Yes, a Public Access / Regular workplace AED similarly to the Philips model that was previously retained

Question no.2:

Would the customer prefer an AED with a superior battery warranty and device warranty than outlined in the RFP?

Answer no.2:

As long as the device can be delivered before the end of fiscal and at minimal cost, yes. (Newer models have 8 year warranties and should definitely be considered) The minimum lithium batteries warranty must be at least the amount stipulated in the SOW however, any device that exceed our basis specifications should be considered over devices that meet the basic requirements.

Question no.3:

Per the latest Heart & Stroke Guidelines, it’s “suggested” to use a CPR Feedback device with AED if

available. Is this something the customer is interested in having as part of the offering?

Answer no.3:

Could you please forward the latest H&S Guidelines?

To the best of my knowledge the CPR Feedback device is meant for CPR Training and is not a life saving

device, so the answer to your question would be no. Please correct me if I’m wrong.

As per the SOW, we do want:

ƒ All Download/Communication Cables and Software.

ƒ A minimum 5 year Warranty.

ƒ 1 hard-copy owner/operator manual in both official languages of Canada.

ƒ **Any audiovisual or training material provided must be in both official languages**

1 hard-copy of the AED maintenance instructions and schedule.”

(3)

Question no.4:

There is no mention in the RFP about the customer’s AED energy delivery preference. Is it important for the customer to have an AED that automatically escalates energy beyond 150J if the first shock is unsuccessful for the patient?

Answer no.4:

We would prefer this option if possible. However, we need delivery by end of fiscal year so want to make sure that this would not cause unnecessary delays.

Question no.5:

Is the customer accepting bid proposals that include a different manufacturer model that meets or

exceeds the “mandatory specifications” outlined in Annex B?

Answer no.5:

As long as there are no further delays, the device will be delivered before the end of fiscal and as long as the device are fully bilingual with an English/French voice selection option at AED start-up without the need of adjusting device default settings.

Must: Provide voice commands and screen messages, in either English and French, to guide the

responder through the defibrillation process step-by-step. Provide CPR coaching by voice prompts and/or audio cues, in either English and French.

Question no.6:

ANNEX A-REQUIREMENT/BASIS OF PAYMENT

Under this section, Item 06 the description states AED Trainer 3 but the item number referenced is for an AED International Wall Sign-Green.

Please confirm you are looking for AED Trainer 3 item number 861467 in your Request for Proposal.

Answer no.6:

We are requesting both, AED Trainer and wall signs to identify AED locations. Please provide your

working document; On our copy of the Statement of Work, item 6.0 refers to “6.0 PROCESS FOR

REQUESTING SERVICE DURING WARRANTY PERIOD”.

Question no.7:

Section 6.2.1 Installation

On-site installation must be provided and be carried out by a qualified service technician.

AED cabinet installation may be safer and definitely more cost effective if completed by a member of the building maintenance team at each site. The maintenance team will have knowledge of what is behind the walls where the AED cabinets will be mounted. Having your own specialist install the cabinet is also far more cost effective than hiring a third party contractor.

The installation of the cabinet is simple and we provide an easy to follow step by step instruction sheet. Installation consists of drilling 4 screws into the wall through the predrilled holes on the back of the

(4)

Would you consider the option of having a member of your maintenance team at each site provide the installation of the AED cabinet? If that is not an option, please define "qualified service technician."

Answer no.7:

We will have our maintenance staff or building management maintenance teams install the AEDS, as previously specified and confirmed.

Question no.8:

In reference to section 3.1.1 Installation you state that “On-site installation must be provided and be

carried out by a qualified service technician”

Given the multiple provinces and delivery locations, is it possible to reevaluate the installation requirement for this bid? We have not been required to install on similar AED tenders.

Answer no.8:

Installation has been removed completely from this requirement.

B. To delete: 1. On page 4 of 22

3.1.1 Installation: delete entirely 2. On page 8 of 22

6.2.1 Installation: delete entirely 3. On page 12 of 22

Delete:

ANNEX A

REQUIREMENT / BASIS OF PAYMENT

The Canada School of Public Service (CSPS) has a requirement for 27 Automated External Defibrillators (AED) Starter Kits as well as 1 (AED) Training Unit to CSPS offices across Canada including installation, manuals training, videos, promotional information and signage in English and French. The delivery is requested by March 31, 2016 across Canada.

Insert:

ANNEX A

REQUIREMENT / BASIS OF PAYMENT

The Canada School of Public Service (CSPS) has a requirement for 27 Automated External Defibrillators (AED) Starter Kits as well as 1 (AED) Training Unit to CSPS offices across Canada including manuals training, videos, promotional information and signage in English and French. The delivery is requested by March 31, 2016 across Canada.

(5)

4. On page 18 of 22 Delete: Annex C entirely

Insert:

ANNEX C DELIVERY ADDRESS

Vancouver

Canada School of Public Service Library Square

300 West Georgia Street, Suite 400 Vancouver (BC)

V6B 6C6

Contact: Charles Buchanan

Victoria

Canada School of Public Service P.L. James Place

1230 Government Street, Suite 539A Victoria (BC)

V8W 3M4

Contact: Per Sekhon

Whitehorse

Canada School of Public Service Elijah Smith Building

300 Main Street, Room 220 Whitehorse, YT

Y1A 2B5

(6)

Calgary

Canada School of Public Service Harry Hays Building

220 4th Avenue South East, Room: 832

Calgary (AB) T2G 4X3

Contact: Sharon Lake

Edmonton

Canada School of Public Service Canada Place

9700, Jasper Avenue, Suite 1110 Edmonton (AB)

T5J 4G3

Contact: Jessica Ward-King

Winnipeg

Canada School of Public Service 320, Donald Street, Suite 200 Winnipeg (Manitoba)

R3B 2H3

Contact: Janine Durand-Rondeau

Toronto

Canada School of Public Service Dominion Public Building 1 Front Street West Toronto (Ontario) 5th Floor, Room 5113

M5J 2X5

Contact: Christopher McHugh

Ottawa

Canada School of Public Service

Academie De-La-Salle

373 Sussex Drive, Room C201 Ottawa, ON

K1N 6Z2

(7)

Montréal

Canada School of Public Service Guy-Favreau au Complex 200, René-Lévesque blvd. West, 8th Floor, East Tower

Montréal (Québec) H2Z 1X4

Contact: Marieve Gauthier

Québec

Canada School of Public Service René-Nicolas Levasseur Building 94 Dalhousie street, 1st Floor Quebec (Qc)

G1K 4B8

Contact: Julie Auclair

Halifax

Canada School of Public Service Maritime Centre

1505 Barrington Street 17th North, suite 1769

Halifax (NS) B3J 3K5

Contact: Nancy Rogers

Moncton

Canada School of Public Service

Royal Trust Building

777 Main Street, room:100 Moncton (NB)

E1C 1E9

Contact: Jocelyne Mallet

Gatineau

Canada School of Public Service Asticou Center

373, Sussex Drive, room: C201 Ottawa (Ontario)

K1N 6Z2

References

Related documents

Annex B – Basis of Payment - B.1.3.1 OPTIONAL DESIGN CHANGE ELEMENTS – ADDITIONAL SECURITY FEATURES ON THE PR CARD - Can Canada confirm that Bidders are NOT required to

Column 2Column 3Column 4Column 5Column 6Column 7Column 8Column 9Column 10 Evaluation Criterion DescriptionCriterion TypeBid DeliverableBidder's Self AssessmentEvidence Location

INSTRUCTIONS TO BIDDERS: Bidders are to complete Sections B of the tables identified by the IU in this article as well as Tables 8 and 9. Bidders must provide a complete

Question: If the awarded consultant is precluded from bidding on the Environmental Monitoring Services scope (or others), can you please confirm the award of the PWTP will be

En effet, remplacer l’article 17.1.C.2.30 de la version française par : (voir aussi article 3 de la présente modification) 17.1.C.2.30: La pompe hydraulique principale, la

Perform all required PVT items identified in the Product Verification Plan (PD-3) on the first complete Decontamination System (including all components of the system), demonstrating

M1 (PB) The Bidder must demonstrate that the proposed resource has a minimum of one hundred and twenty (120) months of combined experience, prior to solicitation issuance date, as

After the incident, the staff member shall complete the AED (automatic external defibrillator) Treatment Incident Summary Form and submit it to the Superintendent