Fran Adar, Pauline Leary, and Thomas Kubic

With the increasing proliferation of counterfeit drug products, there is an incentive to screen

drugs for legitimacy. One method is to examine the tablet itself, which is usually a

destruc-tive operation. Another method that has been explored is to characterize the packaging,

which enables nondestructive screening of the product. Raman microscopy has been found

to be a useful tool, and it is often the tool of choice for these measurements. Raman

mapping of tablets enables not only the identification of components, but also the mapping

pattern can enable matching manufacturing sites of different tablets. The study of packaging

materials can provide an identification of coating and a depth-profile measurement of

thickness. In addition, unlike Fourier transform infrared spectroscopy, which can only

measure surface materials, Raman microscopy can provide spectra of dyes and pigments,

even if they are printed below a surface film.

Raman Microscopy for Detecting

Counterfeit Drugs — A Study of

the Tablets Versus the Packaging

Molecular Spectroscopy Workbench

I

n this column installment, we present results of tablet matching measurements and characterization of pack-aging. The examples show how these measurements aid in the detection of counterfeit drugs.

Raman Tablet Mapping

Two tablets of a well-known drug product were acquired, one from a pharmacy in the United States and one via the mail from an internet pharmacy. Initially, maps of the en-tire tablets were produced using a DuoScan Raman imag-ing system (Horiba Scientific) with a 10 objective and a

200-μm pixel. Although maps were acquired in reasonable amounts of time, there did not seem to be any variability in the data. That is, on the 200-μm scale, the tablets were homogeneous, at least for this product. So, smaller maps were collected to make chemical and textural compari-sons. This work was performed on an Evolution Raman microscope (Horiba Scientific), an 800-mm focal length system, using a 600-grooves/mm grating, and a 532-nm laser. Maps were acquired in the fingerprint region (200– 1800 cm-1) and the CH–OH region (2700–3700 cm-1). LabSpec multivariate analysis software (Horiba Scientific)

Electronically reprinted from June 2014 ®

(2)

was used to create the Raman maps, using either the classical least squares (CLS) algorithm when identifiable species could be seen in the data set, or the multivariate curve resolution (MCR) algorithm when the chemi-cal factors were not apparent. As the results will show, operating on the two data sets independently produced verification of the conclusions.

In reality, it was Witkowski of the Food and Drug Administration (FDA) Forensic Chemistry Center in Cincinnati, Ohio, who originally proposed the use of Raman map-ping for identifying chemical species and morphological distribution as a means of differentiating counterfeit from authentic tablets, and for dif-ferentiating counterfeits from vari-ous manufacturing sites (1). Since that work was done, developments in mapping hardware have improved

data collection times and multivari-ate software provides better quality images resulting from the use of scores mapping rather than intensity between cursors, which is subject to much more noise.

Before data collection, the tablets need to have the coating (usually opaque and pigmented) removed and the surface flattened. The maps shown in Figures 1 and 2 were ac-quired with step sizes between 5- and 15-μm, using the DuoScan Raman imaging system to distribute the laser intensity over each pixel in the map. The SWIFT imaging option (Horiba Scientific) was used to further reduce the acquisition times by eliminat-ing stop and start operations at each data point. After the acquisition of a map, the entire hyperspectral cube was baseline-flattened to improve the quality of the multivariate results.

Then we either surfed the file to iden-tify regions of homogeneous spectra that could be used with the CLS al-gorithm or the MCR alal-gorithm was invoked to “find” the spectral com-ponents in the file. Figures 1 and 2 show maps for each of the two tablets based on the fingerprint spectra and then the CH–OH spectra. The load-ings (“spectra”) used to produce the maps are also shown with reference spectra of the excipient presented for comparison. Because the identity of the active pharmaceutical ingredient (API) was not known, its spectrum could not be shown for comparison. However, APIs tend to be aromatic with large intensities above 1500 cm-1; the loadings shown in red in Figures 1 and 2 are presumed to be API and one can see that the red factors show the same spectra from the two tablets. That is, the counterfeit tablet did con-tain the active ingredient.

Note that for the internet sample, whose results are shown in Figure 2, the CH–OH map has more morpho-logical detail than the fingerprint map. Also, the use of the MCR al-gorithm revealed the presence of a hydrocarbon like magnesium (Mg) stearate. If we go back to the finger-print map and examine the spectrum in the region where Mg stearate ap-peared in the CH region, then that component can be added to the fac-tors, producing the map shown in Figure 3. Surprisingly, in this case the CH–OH region provided more infor-mation than the fingerprint region, as originally examined. We believe that this is because the fingerprint region

-100 -50 0 50 100 55 50 45 40 35 30 25 20 15 10 5 0 200 150 100 50 Intensity (counts) Raman shift (cm-1) Intensity (counts) Raman shift (cm-1) 500 2600 2800 3000 3200 3400 3600 3800 1000 1500 Y (µm) X (µm) 0 100 -100 -50 0 50 100 Y (µm) X (µm) 0 100

Figure 1: Left: Raman maps of tablet purchased from a pharmacy in the United States produced with fingerprint spectra (top) and CH–OH spectra (bottom). Right: Spectral loadings used to create the maps on the left. The red species is presumably the active drug, having bands near 1600 cm-1 and above 3000 cm-1 that indicate unsaturation. The reference spectrum shown in dark

green is that of a cellulose paper filter.

Surprisingly, in

this case the

CH–OH region

provided more

information than

the fingerprint

region, as originally

examined.

(3)

will be dominated by the unsaturated API, whereas in the CH region, all species have similar intensities. Package Matching

Examining the exterior packaging of drug products could provide a more rapid and nondestructive means of screening products for authenticity. The packaging of four samples was examined. To measure the inks and papers, it was necessary to use Raman spectroscopy instead of FT-IR be-cause in FT-IR the radiation will not penetrate any coating present on the paper. Figure 4 shows the spectra of the blue and blue-green pigments on each package, after baseline subtrac-tion. Band frequencies in the bottom spectrum are labeled to assign these spectra as resonance-enhanced. Under such circumstances, overtone and combination bands are strong, and CH stretches are usually nonex-istent. For example, 2  1343 ≈ 2677 cm-1, 1343 + 1531 ≈ 2877 cm-1, and 2  1531 cm-1 ≈ 3062 cm-1. (Note that the matches are usually not exact be-cause of anharmonicity of the vibra-tions.)

The three bottom spectra in Figure 4 appear to be quite similar. Although the bands in the top spectrum do coincide with bands in the other sam-ples, some of the spectral bands in the bottom three samples are absent; in particular, small bands at 1269 and 1487 cm-1 are missing. In the absence of information on the pigments, one can presume that the unsaturated cores of the molecule that provide

-100 -50 0 50 100 Y (µm) X (µm) 0 100 -100 -50 0 50 100 Y (µm) X (µm) 0 100 180 160 140 120 100 80 60 40 20 0 Intensity (counts) Raman shift (cm-1) 500 1000 1500 2000 250 200 150 100 50 Intensity (counts) Raman shift (cm-1) 2600 2800 3000 3200 3400 3600

Figure 2: Left: Raman maps of tablet purchased over the internet produced with fingerprint spectra (top) and CH–OH spectra (bottom). Right: Spectral loadings used to create the maps on the left. The red species is presumably the active drug, having bands near 1600 cm-1 and above

3000 cm-1 that indicate unsaturation. The reference spectrum shown in green is that of lactose.

Note that use of the MCR algorithm in the CH region identified the presence of a third species (shown in grey) whose spectrum is similar to that of Mg stearate.

-100 -50 0 50 100 Y (µm) X (µm) 0 100 250 200 150 100 50 Intensity (counts) Raman shift (cm-1) 500 1000 1500

Figure 3: Left: Raman maps of tablet purchased over the internet produced with fingerprint spectra. Right: Spectral loadings used to create the maps on the left. The red species is presumably the active drug, having bands near 1600 cm-1 and above 3000 cm-1 that indicate unsaturation.

Both excipient spectra (grey = hydrocarbon and blue = lactose) are heavily contaminated with the API probably because the materials do not segregate on this length scale. The reference spectra shown represent lactose in green, and Mg stearate in black.

To measure the inks

and papers, it was

necessary to use

Raman

spectros-copy instead of FT-IR

because in FT-IR the

radiation will not

penetrate any coating

present on the paper.

(4)

the color are quite similar, with some small (unsaturated) modifications which would account for small spec-tral differences.

Figure 5 shows spectra acquired from the noncolored regions of the

same packages. There is a one-to-one correspondence between Figures 3 and 4; that is, the top spectra on each figure come from the same sample, and likewise for the other spectra. Again, one can see that the top

spec-trum is an outlier.

The goal of these measurements was to determine if the packaging materials could be differentiated, and obviously that is documented. But since we have been looking at spectra for so long, we always like to challenge ourselves and see if we can guess what is present. Clearly, that is a frightening idea for the novice who is looking for quick answers to a prob-lem. So, at this point we want to show how current commercial products can alleviate the need for highly ex-perienced spectroscopists. We looked at the spectra of the white packaging and recognized the presence of calcite and polypropylene. Figure 6 shows the bottom spectrum of Figure 5 su-perimposed on our reference spectra of calcite and polypropylene. It is easy to see that most of the spectral fea-tures and intensity can be accounted for by these two species. However, the interesting thing was to determine what the BioRad KnowItAll software (2) (BioRad Laboratories) and data-base could do with this spectrum. Figure 7 shows the results of a mix-ture analysis performed in that soft-ware. The top part of the figure shows the components that were used to fit the spectrum; note that although we were aware of the calcite and poly-propylene we were not aware of the polystyrene until the mixture analy-sis pointed out its presence. Each of these spectra are shown, as well as the measured spectrum and the residual from the fit. Note that because of the tendency of polymers to orient, the relative intensities of the various bands will vary from measurement to measurement, so most of the features in the residual spectrum are not of concern. In addition, the weighting column at the top of the figure shows the relative contributions that were used to fit the measured spectrum. The bottom part of Figure 7 shows, in more detail, the measured spectrum superimposed on the composite spec-trum.

Summary

In the first part of this column in-stallment, we showed that Raman

5000 4000 3000 2000 1000 0 -2000 -3000 -4000 -5000 Intensity (counts) 500 682 749 1343 1269 1487 1531 2279 2677 2877 3061 1000 1500 2000 2500 3000 Raman shift (cm-1)

Figure 4: Raman spectra of green or blue inks from pharmaceutical packaging of a drug product and three possible counterfeits of that product. Spectra were excited with the 633-nm laser line and detected on an Evolution Raman microscope (Horiba) equipped with a 600-grooves/mm grating. All spectra exhibited a fluorescence background, presumably from the pigment itself; spectra were baseline-corrected for the display.

4000 2000 0 -8000 -10,000 Intensity (counts) 500 1000 1500 2000 2500 3000 3500 4000 Raman shift (cm-1)

Figure 5: Raman spectra of white regions from pharmaceutical packaging of a drug product and three possible counterfeits of that product. Spectra were excited with the 633-nm laser line and detected on an Evolution Raman microscope (Horiba) equipped with a 600-grooves/mm grating. Spectra were baseline-corrected for the display.

(5)

images constructed from XY maps of tablet surfaces can provide informa-tion both on the composiinforma-tion of the tablets and the components’ distribu-tion pattern, both of which can help determine if a sample is authentic or not. In addition, differences can aid in determining if counterfeit samples have a common origin. In the second

part of this column, we showed that spectra of the packaging material, which can be recorded without any sample destruction, can also aid in the identification of counterfeits. In addition, for those who want to be able to identify material from the spectra, we showed how the use of commercially available software can

provide identification, even for a mix-ture sample.

References

(1) F. Adar, E. Lee, and M. Witkowski, “Single Point Analysis and Raman Mapping of Dosage Formulation as a Means for Detecting and Sourcing Counterfeit Pharmaceuticals,” pre-sentation at Pittcon 2007. (2) Bio-Rad Laboratories, Inc.,

Philadel-phia, Pennsylvania, 2014. 6000 5000 4000 3000 2000 1000 0 -1000 Intensity (counts) 500 1000 1500 2000 2500 3000 3500 4000 Raman shift (cm-1)

Figure 6: Raman spectrum of white region of one of the packaging samples (top, red) superimposed on a reference spectrum of polypropylene (middle), and calcite (bottom).

2000 0 2000 0 4000 3800 3600 3400 3200 3000 2800 2600 2400 2200 2000 1800 1600 1400 1200 1000 800 600 400 Wavenumbers 200 0 Raman Table Plot HQI

23 937.58 N.A. Composite Spectrum H2C CH

H 2 C CH n n RAX #332; Polypropylene 5724WZ POLY(STYRENE), GENERAL PURPOSE*NATURAL Residual Spectrum Analytical Techniques QRX #74; CALCIUM CARBONATE, 85% CONCENTRATE N.A. 0.62 0.27 0.11 QRX 946

Weight Exclude DB ID Name Chemical Structure Spectrum <auto> (Raman)

– Composite Spectrum

Figure 7: Results of the mixture analysis in the KnowItAll software. The bottom of the figure shows the measured spectrum (in black) superimposed on the composite spectrum (in red) which is constructed using the pure spectra shown in the top of the figure

For more information on this topic, please visit: www.spectroscopyonline.com/adar

Fran Adar is the Principal Raman Applications Scientist for Horiba Scientific (Edison, New Jersey). She can be reached by e-mail at fran.adar@ horiba.com

Pauline Leary is the applications man-ager for the Americas for the chemical and biological product line at Smiths Detec-tion. Prior to holding her current position at Smiths, Pauline worked as a pharma-ceutical R&D scientist at Purdue Pharma in Ardsley, New York. Pauline holds a B.A. de-gree from Marist College, an M.S. dede-gree from John Jay College of Criminal Justice, and a PhD from the Graduate Center of the City University of New York.

Thomas A. Kubic, PhD, is an Associate Professor of Forensic Science and Chemistry at John Jay College and the Grad-uate Center, CUNY. He has been in the academic world for more than 15 years after completing 23 years in a municipal crime laboratory. His research interests are in the applications of light and electron mi-croscopy along with UV–vis and vibration spectrometry to the analysis of micro and ultramicro samples.

Posted with permission from the June 2014 issue of Spectroscopy ® www.spectroscopyonline.com. Copyright 2014, Advanstar Communications, Inc. All rights reserved.

References

Related documents

Results suggest that the probability of under-educated employment is higher among low skilled recent migrants and that the over-education risk is higher among high skilled

The paper is discussed for various techniques for sensor localization and various interpolation methods for variety of prediction methods used by various applications

Also, both diabetic groups there were a positive immunoreactivity of the photoreceptor inner segment, and this was also seen among control ani- mals treated with a

Field experiments were conducted at Ebonyi State University Research Farm during 2009 and 2010 farming seasons to evaluate the effect of intercropping maize with

Мөн БЗДүүргийн нохойн уушгины жижиг гуурсанцрын хучуур эсийн болон гөлгөр булчингийн ширхгийн гиперплази (4-р зураг), Чингэлтэй дүүргийн нохойн уушгинд том

19% serve a county. Fourteen per cent of the centers provide service for adjoining states in addition to the states in which they are located; usually these adjoining states have

Aptness of Candidates in the Pool to Serve as Role Models When presented with the candidate role model profiles, nine out of ten student participants found two or more in the pool