Le fabricant et nous même, nous réservons le droit de modifier le produit en vue d’une amélioration

Art-

2007-

SpO2 PULSE OXYMETER MD-300

MD-300 C

MD-300 D

MD-300 F

MD-300 Z

MD-300 A

MD-300 B

MD-300 I

Finger clip ADULT

Finger

clip

PEDIATRIC

Soft

tip

ADULT

Wrap

ADULT

->

PEDIATRIC

Wrap

PEDIATRIC

->

NEONATE

(2)

Pulse oxymètre portable MD-300C

Technical Parameters Display Format: LED

SPO2:35-99% in 2-digit 7-segment digital;

Pulse Ratio:30-240 BPM in 3-digit 7-segment digital; Pulse Wave: 6-segment bargraph

Battery Voltage Low Indication: LED signals.

* Power Requirements: Two AAA size 1.5V Alkali Battery; * Power Consumption: less than 40mA.

* Resolution: ±1% for SPO2 and ±1 BPM for Pulse Ratio.

* Measurement Accuracy: ±2% in stage of 70%-99% SPO2 and meaningless when stage being smaller than 70%. ±2 BPM for Pulse Ratio.

* Measurement Performance in Weak Filling Condition: Oxyhemoglobin and pulse value can be shown correctly when pulse-filling ratio is 0.5%.

* Interference Resistance Capacity against Ambient Light: Deviation is smaller than ±1% between value of Oxyhemoglobin measured in natural lighting indoor condition and present lighting sources and that measured in dark room.

Product Properties

Operation of the product is simple and convenient

* The product is small in volume, light in weight (total weight is about 50g including batteries) and convenient in carrying.

* Power consumption of the product is low and the two originally equipped two AAA batteries can be operated continuously for 30 hours.

* Low voltage warning will be indicated in visual window when battery voltage is so low that normal operation of the Oximeter might be influenced.

* The product will automatically be powered off when no signal is in the product for more than 8 seconds. Packing Dimension

15cm x 10.5cm x 4cm/pc/box Gross Weight

(3)

Le fabricant et nous même, nous réservons le droit de modifier le produit en vue d’une amélioration

Art-

2007-

SpO2 PULSE OXYMETER MD-300

Pulse oxymètre portable MD-300D

Technical Parameters Display Format: LCD SPO2:35-99% in 2-digit ;

Pulse Ratio:30-240 BPM in 3-digit; Pulse Wave: 6-segment bargraph Plethys Waveform

Battery Voltage Low Indication: .

* Power Requirements: Two AAA size 1.5V Alkali Battery; * Power Consumption: less than 40mA.

* Resolution: ±1% for SPO2 and ±1 BPM for Pulse Ratio.

* Measurement Accuracy: ±2% in stage of 70%-99% SPO2 and meaningless when stage being smaller than 70%. ±2 BPM for Pulse Ratio.

* Measurement Performance in Weak Filling Condition: Oxyhemoglobin and pulse value can be shown correctly when pulse-filling ratio is 0.5%.

* Interference Resistance Capacity against Ambient Light: Deviation is smaller than ±1% between value of Oxyhemoglobin measured in natural lighting indoor condition and present lighting sources and that measured in dark room.

Product Properties

Operation of the product is simple and convenient

* The product is small in volume, light in weight (total weight is about 50g including batteries) and convenient in carrying.

* Power consumption of the product is low and the two originally equipped two AAA batteries can be operated continuously for 30 hours.

* Low voltage warning will be indicated in visual window when battery voltage is so low that normal operation of the Oximeter might be influenced.

* The product will automatically be powered off when no signal is in the product for more than 8 seconds. Packing Dimension

15cm x 10.5cm x 4cm/pc/box Gross Weight

(4)

Pulse oxymètre portable MD-300C

Technical Parameters Display Format: LED

SPO2:35-99% in 2-digit 7-segment digital;

Pulse Ratio:30-240 BPM in 3-digit 7-segment digital; Pulse Wave: 6-segment bargraph

Battery Voltage Low Indication: LED signals.

* Power Requirements: Two AAA size 1.5V Alkali Battery; * Power Consumption: less than 40mA.

* Resolution: ±1% for SPO2 and ±1 BPM for Pulse Ratio.

* Measurement Accuracy: ±2% in stage of 70%-99% SPO2 and meaningless when stage being smaller than 70%. ±2 BPM for Pulse Ratio.

* Measurement Performance in Weak Filling Condition: Oxyhemoglobin and pulse value can be shown correctly when pulse-filling ratio is 0.5%.

* Interference Resistance Capacity against Ambient Light: Deviation is smaller than ±1% between value of Oxyhemoglobin measured in natural lighting indoor condition and present lighting sources and that measured in dark room.

Product Properties

Operation of the product is simple and convenient

* The product is small in volume, light in weight (total weight is about 50g including batteries) and convenient in carrying.

* Power consumption of the product is low and the two originally equipped two AAA batteries can be operated continuously for 30 hours.

* Low voltage warning will be indicated in visual window when battery voltage is so low that normal operation of the Oximeter might be influenced.

* The product will automatically be powered off when no signal is in the product for more than 8 seconds. Packing Dimension

15cm x 10.5cm x 4cm/pc/box Gross Weight

(5)

Le fabricant et nous même, nous réservons le droit de modifier le produit en vue d’une amélioration

Art-

2007-

SpO2 PULSE OXYMETER MD-300

Pulse oxymètre portable MD-300Z

Technical Parameters Display Format: OLED SPO2:70-99% in 2-digit 7;

Pulse Ratio:30-240 BPM in 3-digit ;

Battery Voltage Low Indication: LED signals.

* Power Requirements: Two AAA size 1.5V Alkali Battery; * Power Consumption: less than 40mA.

* Resolution: ±1% for SPO2 and ±1 BPM for Pulse Ratio.

* Measurement Accuracy: ±2% in stage of 70%-99% SPO2 and meaningless when stage being smaller than 70%. ±2 BPM for Pulse Ratio.

* Measurement Performance in Weak Filling Condition: Oxyhemoglobin and pulse value can be shown correctly when pulse-filling ratio is 0.5%.

* Interference Resistance Capacity against Ambient Light: Deviation is smaller than ±1% between value of Oxyhemoglobin measured in natural lighting indoor condition and present lighting sources and that measured in dark room.

Product Properties

Operation of the product is simple and convenient

* The product is small in volume, light in weight (total weight is about 50g including batteries) and convenient in carrying.

* Power consumption of the product is low and the two originally equipped two AAA batteries can be operated continuously for 30 hours.

* Low voltage warning will be indicated in visual window when battery voltage is so low that normal operation of the Oximeter might be influenced.

* The product will automatically be powered off when no signal is in the product for more than 8 seconds. * Size: 58mm(L), 32mm(W), 34mm(H)

Packing Dimension

15cm x 10.5cm x 4cm/pc/box Gross Weight

(6)

Pulse oxymètre portable MD-300A

Technical Specifications

[Patient Range]: Adult, Pediatric and Neonatal [SpO2]

Range 0%-100% Resolution 1%

Accuracy 70%-100%: ±2% 0%-69%:unspecified Method Dual wavelength LED

[Heart (Pulse) Rate] Range 25-254 bpm

Resolution 1bpm Accuracy ±2bpm or ±2% [Alarm]

Alarm SpO2, probe off Modes Audio, visual, information [Display]

Type Backlit LCD

Parameters SpO2, PR, Pleth waveform, Pleth bar.

Information SpO2 & PR trend, browse, Low power, error, time [Environmental]

Operating Temperature 0℃-50℃ Storage Temperature -10℃ - 60℃ Operating Humidity 15%RH - 95%RH Storage Humidity 10%RH - 95%RH [Classification]

Type of protection internally powered equipment Safety IEC Standard 61601-1-1 [Storage]

(7)

Le fabricant et nous même, nous réservons le droit de modifier le produit en vue d’une amélioration

Art-

2007-

SpO2 PULSE OXYMETER MD-300

Pulse oxymètre portable MD-300B

Technical Specifications

[Patient Range] Adult, Pediatric and Neonatal [SpO2]

Range 0%-100% Resolution 1%

Accuracy 70%-100%: ±2% 0%-69%:unspecified Method Dual wavelength LED

[Heart (Pulse) Rate] Range 25-254 bpm

Resolution 1bpm Accuracy ±2bpm or ±2% [Alarm]

Alarm SpO2, probe off Modes Audio, visual, information [Display]

Type 3-digit 7-segment LED

Parameters SpO2, PR, Pleth waveform, Pleth bar.

Information SpO2 & PR trend, browse, Low power, error, time [Environmental]

Operating Temperature 0℃-50℃ Storage Temperature -10℃ - 60℃ Operating Humidity 15%RH - 95%RH Storage Humidity 10%RH - 95%RH [Classification]

Type of protection internally powered equipment Safety IEC Standard 61601-1-1 [Storage]

Data records Store and replay 72 hours SPO2% and Pulse rate value [Mechanical]

Size 120 ×60 ×22mm Weight: 120 g [Battery]

Type 4 AAA Alkaline batteries or 4 AAA rechargeable batteries Operation time About 48 hours for normal operation

(8)

Technical Specifications

[Patient Range] Adult, Pediatric and Neonatal [SpO2]

Range 0%-100% Resolution 1%

Accuracy 70%-100%: ±2% 0%-69%:unspecified Method Dual wavelength LED

[Heart (Pulse) Rate] Range 25-254 bpm

Resolution 1bpm Accuracy ±2bpm or ±2% [Alarm]

Alarm SpO2, probe off Modes Audio, visual, information [Display]

Type LCD

Parameters SpO2, PR, Pleth waveform, Pleth bar. Information SpO2 & PR trend, browse,

Low power, error, time [Environmental]

Operating Temperature 0℃-50℃ Storage Temperature -10℃ - 60℃ Operating Humidity 15%RH - 95%RH Storage Humidity 10%RH - 95%RH [Classification]

Type of protection internally powered equipment Safety IEC Standard 61601-1-1 [Storage]

References

Related documents

analysis time frame Key Australian findings Key Victorian findings Distinguishing factors Economic Costs of Natural Disasters in Australia – BTE, 2001 EMA database

In the rules of English, the role of copula be is very important if there is no copula in sentences that should require a sentence in English, so the sentence can be

Put more directly, an increase of school size by 145 students, equivalent to the difference in average school size between the median state in the 1920-29 cohort and the median

We worked with Massachusetts Audubon's Wachusett Meadow wildlife sanctuary and Clinton elementary school to advance engineering skills in current Science, Technology, Engineering and

Uni- and multivariable Cox proportional hazard regression was used to examine the association between kidney function and outcome using three models: in the first

In light of the research gap, the need for alternative funding models to improve national water infrastructure management, and the necessity for improved access to drinking water for

Målet med studien er å se på hvilken betydning opplevelsen av sosiale relasjoner har for vellykket deltakelse i arbeidslivet og hvilken betydning sosiale arbeidsrelasjoner

There was a significant association between smoking behaviors among students and their age, gender, type of accommodation, field of study, income status, duration of study,