11 An Update on IOL PowerCalculation Formulas

ACCOMMODATING IOL (AIOL)

The accommodating rate function (M) defined as the accommodation amplitude increase per 1.0 mm forward movement of a plus AIOL may be expressed by the Lin’s M-formula

M = (Z/1336)P[2Dc+ZZ’P] (11)

Above formula is a general form for both phakia and aphakia and also for dual-optics AIOL. As shown in Fig. 3, for the thick-IOL single-optics case, the M value is higher

for convex-concave IOL configuration (having front and back surface power of P1 and P2) which has higher P1 for a given P1+P2, when P2<0 (concave). Depending on ocular conditions, M ranges from 0.5 to 2.5 (D/mm) for AIOL-power of 10 to 20 diopters. Greater details of single-optics and dual-single-optics AIOL may be found in other Chapters of Lin in this book.

Fig. 11.8: Accommodation rate (M) vs IOL-power in single-optics IOL at various configurations

CONCLUSION

The existing IOL-power (except Lin’s) based on 2-optics system could only apply to aphakia. For phakia or piggyback IOL, a 3-optics system based on Gaussian optics is required. The Lin’s new formulas provide more accurate calculations for both phakia and aphakia presented by:

• The S-formula (to include IOL thickness and types)

• The Seff-formula (to include the role of natural lens and primary-IOL)

• The personalized r2-formula (including corneal posterior surface power post-LASIK)

• The M-formula (for accommodating IOL efficiency)

• The Z2-formula (for conversion of IOL-power in 3-optics system of phakia or piggyback IOL).

BIBLIOGRAPHY

1. Atchison DA, Smith G. Optics of the human eye. Woburn, MA: Butterworth-Heinemann 2000;14-16,143-47,160-62.

2. Atchison DA. Optical design of intraocular lens. I. On-axis performance. Optom Vis Science 1989;66:492-506.

3. Binkhorst RD. The optical design of intraocular lens implants.

Ophthalmic Surg 1975;6:17-31.

4. Buschmann, W. Ultrasonic measurement\s of the axial length of the eye, Klin. Monatsbl. Augenheilkd. 1964;144:801-15.

5. Cheng ACK, Lam DSC. Keratometry for intraocular lens power calculation using Orbscan II in eyes with laser in situ keratomileusis. J Refract Surg 2005;21:365-68.

6. Connors R, III, Boseman P, III, Olson RJ. Accuracy and reproducibility of biometry using partial coherence interferometry. J Cataract Refract Surg 2002;28:235-8.

7. Coppens JE, van der Berg, JTTP, Bado CJ. Biometry of phakic intraocular lens using Scheimpflug photography. J Cataract Ref Surg 2005;31:1904-14.

8. Cumming JS, Slade SG, Chayet A. Clinical evaluation of the model AT-45 silicone Accommodating intraocular lens.

Ophthalmology 2001;108:2005-09.

9. Drews RC. Calculation of intraocular power, a program for Hewlett-Packard 97 calculator. Am Intraocular Implant Soc J 1977;3:209-12.

10. Drexler W, Findl O, Menapace R, et al. Partial coherence interferometry: a novel approach to biometry in cataract surgery. Am J Ophthalmol 1998;126:524-34.

11. Feiz V, Mannis MJ, Garcia-Ferrer F et al. Intraocular lens power calculation after LASIK for myopia and hyperopia.

Cornea 2001;20:792-97.

12. Eleftheriadis H. IOL Master biometry: refractive results of 100 consecutive cases. Br J Ophthalmol 2003;87(8):960-3.

13. Findl O, Drexler W, Menapace R, et al. Improved prediction of intraocular lens power using partial coherence interferometry. J Cataract Refract Surg 2001;27:861-7.

14. Fritz KJ. Intraocular lens power formulas. Am J Ophthalmol 1981;91:414.

15. Fyodorov SN, Galin MA, Linksz A. Calculation of the optical power intraocular lens. Invest Ophthalmol 1975;14:625-28.

16. Gantenbein C, Lang HM, Ruprecht KW, Georg T. First steps with the Zeiss IOL Master: A comparison between acoustic contact biometry and non-contact optical biometry, Klin Monatsbl Augenheilkd. 2003;220(5):309-14.

17. Gills JP, Denzl R. Minus-power intraocular lenses to correct refractive errors in myopic pseudophakia. J Cataract Refract Surg 1999;25:1205-08.

18. Habot-Wilner Z, Sachs D, Cahane M, et al. Refractive results with secondary piggyback implantation to correct pseudophakic refractive errors. J Cataract Refract Surg 2005;

31:2101-03.

19. H Eleftheriades. IOL Master Biometry refractive results of 100 consecutive cases. British Journal of Ophthalmology 2003;

87:960-63.

20. Kalogeropoulos C, Aspiotis M, Stefaniotou M, Psilas K.

Factors influencing the accuracy of the SRK formula in the intraocular less power calculation. Doc Ophthalmol 1994;85(3):223-42.

21. Kim JN, Lee DN, Joo CK. Measuring corneal power for intraocular lens power calculations after refractive surgery. J Cataract Refract Surg 2002;28(II):1932-38.

22. Ho A, Manns F, Pham T et al. Predicting the performance of accommodating intraocular lens using ray tracing. J Cataract Refract Surg 2006;32:129-36.

23. Hoffer KJ. Accuracy Ultrasound intraocular lens calculation.

Arch Ophthalmol 1981;99:1819-23.

24. Hoffer KJ. Intraocular Lens Calculation: The Problem of the Short Eye {Hoffer Formula}. Ophthalmic Surgery 1981;

12:269-72.

25. Hoffer KJ. The Effect of Axial Length on Posterior Chamber Lenses and Posterior Capsule Position. Current Concepts in Ophthalmic Surgery 1984;1:20-22.

26. Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg 1993;

19:700-712. ERRATA 1994;20:677 ERRATA: 2007;33:2-3.

27. Holladay JT. Refractive power calculation for intraocular lenses in the phakic eye. Am J Ophthalmol 1993;116:63-66.

28. Holladay JT. Optics and intraocular lens power calculation for phakic intraocular lenses. In: Horden DR. Lindstrom RL.

Davis EA, ed. Phakic intraocular lens: principles and practice.

Thorofare NJ: SLACK; 2004;59-61.

29. Kraff MC, Sanders DR, Lieberman HL. Determination of intraocular lens power: a comparison with and without ultrasound. Ophthalmic Surg 1978;9:81-84.

An Update on IOL Power Calculation Formulas

61

30. Kielhorn I, Rajan MS, Tesha PM, Subryan VR, Bell JA. Clinical assessment of the Zeiss IOL Master 1: J Cataract Refract Surg 2003;29(3):518-22.

31. Liang YS, Chen TT, Chi TC, Chan YC. Analysis of intraocular lens power calculation. J Am Intraocular Implant Soc 1985;

11(3):268-71.

32. Lin JT. Unified refractive-state analysis for customized vision correction. J Refract Surg 2004;20:398-400.

33. Lin JT. New formulas comparing accommodation in human lens and intraocular lens. J Refract Surg 2005;21:200-01.

34. Lin JT. A generalized refractive state theory and effective eye model. Chin J Optom and Ophthal 2005;7:1-6.

35. Lin JT. Correcting the cornea-power measurement for intraocular lens power calculation after myopic laser in situ keratomileusis. Am J Ophthalmol 2004;137:786-87.

36. Lin JT. Unified formulas for phakic-aphakic IOL and spectacle-cornea power conversion. J Refract Surg 2006;22:(in press).

37. Lin JT. The role of corneal power and anterior chamber depth on the accommodation in aphakic IOL. J Refract Surg 2005;21:765-66.

38. Lin JT. Error analysis of intraocular lens power in eye after LASIK. J Refract Surg 2006;22:127-28.

39. Lin JT. Analytic formulas for error factors in intraocular lens power calculation. J Cataract Refract Surg 2006;32:(in press).

40. Mcleod SD, Portney V, Ting A. A dual optic accommodating fordable intraocular lens. Br J Ophthalmol 2003;87:

1083-85.

41. Missotten T, Verkamme T, Blanckaert J, et al. Optical forumula to predict outcomes after implantation of accommodating intraocular lens. J Cataract Refract Surg 2004;30:2084-87.

42. Nawa Y, Ueda T, Nakatsuka M, et al. Accommodation obtained per 1.0 mm forward movement of a posterior chamber intraocular lens. J Cataract Refract Surg 2003;

29:2069-72.

43. Odenthal MTP, Eggink CA, Melles G et al. Clinical and theoretical results of intraocular lens power calculation for cataract surgery after photorefractive keratectomy for myopia. Arch Ophthalmol 2002;120:431-38.

44. Olsen T, Corydon L, Gimbel H. Intraocular lens power calculation with an improved anterior chamber depth prediction algorithm. J Cataract Refrqact Surg 1995;21:

313-19.

45. Olsen T. Theoretical approach to intraocular lens calculation using Gaussian optics. J Cataract Refract Surg 1987;13:

141-5.

46. Olsen T. Sources of error in intraocular lens power calculation.

J Cataract Refract Surg 1992;18:125-9.

47. Olsen T, Thim K, Corydon L. Accuracy of the newer generation intraocular lens power calculation formulas in long and short eyes. J Cataract Refract Surg 1991;17(2):

187-93.

48. Ouda B, Tawafik B, Derbala A, Youseif AB. Error correction of intraocular lens (IOL) power calculation. Biomed Instrum Technol 1999;33(5):438-45.

49. Pedrotti LS, Pedrotti FL. Optics and Vision. Liper Saddle River, NJ, Prentice Hall 1998;74-87:122-41.

50. Pedrotti LS, Pedrotti FL. Optics and Vision. Liper Saddle River, NJ, Prentice Hall 1998;74-87:92-95.

51. Pierscionek BK. Presbyopia-effect of refractive index. Chin Exp Optom 1990;73:23-30.

52. Preussner PR, Wahl J, Weitzel D. Topography-based intraocular lens power selection. J Cataract Refract Surg 2005;

31:525-33.

53. Rana A, Miller D, Magnante P. Understanding the accom-modating IOL. J. Cataract Refract Surg 2003;29:2284-87.

54. Rajan MS, Keilhorn I, Bell JA. Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations. Eye 2002;16:552–6.

55. Rosa N, Capasso L, Lanza M et al. Reliability of a new correcting factor in calculating intraocular lens power after refractive corneal surgery. J Cataract Refract Surg 2005;

31:1020-24.

56. Rose LT, Moshegov CN. Comparison of the Zeiss IOL Master and applanation A-scan ultrasound: biometry for intraocular lens calculation. Clin Experiment Ophthalmol 2003;31(2):

121-24.

57. Raj PS, Ilango B, Watson A : Measurement of axial length in the calculation of intraocular lens power. Eye 1998;

12(Pt2):227-9.

58. Shammas HJ. A comparison of immersion and contact techniques for axial length measurement. J Am Intra-Ocular Implant Soc 1984;10:444-7.

59. Sanders DR, Kraff MC. Improvement of intraocular lens power calculation using empirical data. Am Intraocular Implant Soc J 1980;6:263-67.

60. Sanders DR, Retzlaff J, Kraff MC. Comparison of empirically derived and theoretical aphakic refraction formulas. Arch Opththal 1983;10:965-67.

61. Sanders DR, Retzlaff J, Kraff MC. Comparison of the SRKII formulas and the other second generation formulas. J Cataract Refract Surg 1988;14:136-41.

62. Shammas HJ. Intraocular lens power calculations. Thorofare NJ: SLACK; 2004;59-61:64-65.

63. Stachs O, Schneider H, Steve J, Guthoff R. Potentially accommodating intraocular lenses-an in vitro and in vivo study using three-dimensional high-frequency ultrasound. J Refract Surg 2005;21:37-45.

64. Suto C, Hori S, Fukuyama E, Akura J. Adjusting intraocular lens power for sulcus fixation. J Cataract Refract Surg 2003;29(10):1913-7.

65. Telando A. Pseudo-accommodative cornea: a new concept for correction of presbyopia. J Refract Surg 2004;20(Suppl.):

S714-17.

66. Twa MD, Lemback RG, Bullimore MA, Roberts C, “A prospective randomized clinical trial of laser in situ keratomileusis with two different lasers.” Am J Ophthalmol 2005;140:173-83.

67. van der Heijde GL. The optical correction of unilateral aphakia.

Trans Amer Academy Ophthalmol Otolarynogol 1976;81:

80-88.

68. Verhulst E, Vrijghem JC. Accuracy of intraocular lens power calculations using the Zeiss IOL Master. A prospective study, Bull Soc Belge Ophthalmol 2001;(281):61-65.

69. Verhulst E, Vrijghem JC. Accuracy of intraocular lens power calculations using the Zeiss IOL Master. A prospective Study.

70. Vongthorgsri A, Phusitphoykai N, Tungsiriput T. Laser in situ keratomileusis for high myopia using a small ablation-zone and large aspheric transition ablation-zone. J Refract Surg 2004;20(Suppl):S669-673.

71. Wainstock MA. Ultrasonography: Its role in the success of intraocular implant surgery. Int Ophthalmol Clin 1979;19:

43-50.

72. Wallman J. How is emmetropization controlled? Results of research on experimental myopia. In: Tokoro T ed. Myopia updates: Proc. 6-th Inter. Conf. on Myopia, Tokyo: Spring-Verlag; 1998;13-30.

73. Wang XJ, Jin CP, Wang QM. Aberration with IOLs based on different optical design. Chinese J Optom and Ophthal (in Chinese). 2003;5:209-11.

74. Wildsoet CF. Structural correlates of myopia. In: Resenfield M, Gilmartin B (Eds). Myopia and Nearwork. Woburn, MA:

Butterworth-Heinemann; 1998;31-56.

12 The New IOL Formulas