This chapter presented a literature review of the dierent methods that will be incorporated in the design of three dehydration assessment devices. An introduction to existing hydration markers was given and the shortcomings of the markers were discussed. These shortcomings were due to the marker's inability to work as an ineld measuring device for dehydration. In addition to this the cost, portability and required technical expertise of each marker was given and it was seen to be in contrast with the objectives of this project. The theory behind BIA and SC impedance analysis was given as well as a brief insight to the workings of optical behaviour. Lastly with the collective knowledge of the dierent methods hypotheses were made postulating the abilities of the dierent methods.
Chapter 3
Device Design and Implementation
3.1 Device Descriptions
As mentioned in the previous chapter bioelectrical impedance analysis, stra- tum corneum impedance analysis and infrared spectrometry will be used to identify dehydration in a person. Here follows the description of the fabricated devices and how they will be implemented. The data sheets for the dierent components mentioned in this section can be found in the appendix.
3.1.1 BIA
Many BIA analysis devices exist and have been used as a standard device for several studies. An example of such a device is the Xitron Hydra used by Röthlingshöfer in Monitoring Change of Body Fluid during Physical Exercise using Bioimpedance Spectroscopy and Finite Element Simulations [15]. It was aimed to produce an inexpensive way to measure the bioimpedance of the body and it was thus opted to design a low current BIA device safe to use on infants. This was done through the use of a direct digital synthesizer (DDS) module, microcontroller and some analogue circuitry (discussed in more detail in Section 3.2.1). The microcontroller enables the DDS module to generate a sinusoidal signal which is lead to the skin through a coaxial cable attached to two disposable pre-gelled ECG electrodes. These electrodes are placed upon the ankles of the subject and the impedance is measured forthwith using the analogue to digital converters (ADC) of the microcontroller.
3.1.2 SC Analysis
There already exists a number of stratum corneum probes namely the Skicon 200 and Corneometer CM 825. As mentioned in the objectives of this project it is sought to use inexpensive technologies and it was thus decided to use the method incorporated by Hamed et al. [8]. Hamed et al. conducted a study using a double circular electrode comprising of copper, and showed that
Figure 3.1: Stratum corneum probe encapsulated in silicon with coaxial cable attached
hydration in the stratum corneum was indeed measurable and that it showed a good correlation in comparison to the Corneometer. It was thus decided that the probe used by Hamed et al. will be used for this project.
As described by Hamed et al. the probe consists of circular copper rings printed onto a circuit board. Figure 3.1 shows the probe with the inner elec- trode having a diameter of 6 mm and outer electrode having an inner diameter of 10 mm and outer diameter of 12 mm. These electrodes were relayed via a coaxial cable to the measuring hub described in Section 3.2.1.
3.1.3 Infrared Spectrometry
As mentioned in Section 2.5 infrared is a very advantageous spectrum to work with in the case of water concentration measurement. This is due to the high absorption capabilities of water in this spectrum. There do however exist other major absorbers in this spectrum which were taken into account and the wavelengths of interest were chosen in such a way that the absorption will be dominated by water. These wavelengths were chosen at 1300 nm and 1480 nm. These wavelengths were chosen purposefully so that the 1300 nm had an absorption coecient fairly smaller than that found at 1480 nm. This was done because it was believed that with the smaller absorption coecient the light will be able to travel further and hopefully propagate further into the skin before diminishing to an unmeasurable magnitude. It is also believed that at this wavelength a greater area of tissue will be observed compared to the 1480 nm wavelength. The 1480 nm wavelength was chosen as it was believed that it would be more sensitive to water concentrations given its high absorption coecient.
These two wavelengths are produced by using specially ordered infrared LEDs provided by Roithner Technik [35]. These are used in conjunction with a photodiode provided by the same manufacturer that measure the spectrum
Figure 3.2: Infrared probe concealed in silicon
between 1000 nm and 1600 nm. The LEDs and photodiode are mounted so that the sensing area of the photodiode and the light inducing area of the LEDs are in the same plane. This is to ensure that when pressed onto the skin all forms of light will be cuto from the photodiode and the only wavelengths it will measure will be that produced by the LEDs. Other forms of circuitry is needed to operate the LEDs and the photodiode but this is shown in Section 3.2.2.
An overview of the devices has been given with a summary with regards to design choices made. The following section describes the circuit design of the above mentioned devices in detail.