7 Technical data 30
7.3 Dimensions
Figure 7-2: SMARTPAT COND 7200
Dimensions [mm] Dimensions [inch]
a 170 / 200 6.69 / 7.87
b 143 / 173 5.63 / 6.81
c 60 / 90 2.36 / 3.54
d 40 / 70 1.57 / 2.76
e WS 27
f M28 x 1.5
g 35.9 1.41
h 22.5 0.88
i Ø20.5 Ø0.87
k VarioPin
l 25 0.98
m 2 0.08
7 TECHNICAL DATA
SMARTPAT COND 7200Figure 7-3: Sensor mounted with process connection 1 Sensor insertion length 40 mm / 1.57"
2 Sensor insertion length 70 mm / 2.76"
Dimensions [mm] Dimensions [inch]
a DN 50 2
DN 65 2 1/2
DN 80 3
DN 100 4
DN 125 5
DN 150 6
b 39 1.54
c 41 1.61
TECHNICAL DATA 7
SMARTPAT COND 7200
Figure 7-4: Process connection adapter 1 Varivent® DN 40...125
2 TriClamp DN 25/32/40 (1...1.5" / Ø50.5 mm) 3 TriClamp DN 50 (2" / Ø64 mm)
Dimensions [mm] Dimensions [inch]
a b a b
Varivent® DN 40...125 21.5 68 0.85 2.68
TriClamp DN 25/32/40 50.5 1.99
TriClamp DN 50 64 2.52
8 DESCRIPTION OF HART INTERFACE
SMARTPAT COND 72008.1 General description
The open HART® protocol, which can be used for free, is integrated into the sensor for communication.
Devices which support the HART® protocol are classified as either operating devices or field devices. When it comes to operating devices (Master), both manual control units (Secondary Master) and PC-supported workstations (Primary Master) are used in, for example, a control center.
HART® field devices include sensors, signal converters and actuators. The field devices range from 2-wire to intrinsically safe versions for use in hazardous areas.
The HART® data are superimposed over the analogue 4...20 mA signal via FSK modem. This way, all of the connected devices can communicate digitally with one another via the HART® protocol while simultaneously transmitting the analogue signals.
When it comes to the field devices and secondary masters, the FSK or HART® modem is integrated. If a PC is used, an external modem must be connected to the serial interface (USB interface). There are, however, other connection variants which can be seen in the following connection figures.
8.2 Software history
HART
®identification codes and revision numbers
INFORMATION!In the table below, "x" is a placeholder for possible multi-digit alphanumeric combinations, depending on the available version.
Release date ER version SW version HW version HART® Device Revision
DD Revision
2015-08-01 1.0.0_ 1.0.x 1.0.x 1 1
Manufacturer ID: 69 (0x45)
Device: 194 (0xC2)
Device Revision: 1
DD Revision: 1
HART® Universal Revision: 7
FC 475 system SW.Rev.: ≥ 3.8
DESCRIPTION OF HART INTERFACE 8
SMARTPAT COND 7200
8.3 Connection variants
The sensor is a 2-wire device with a passive 4...20 mA current output and HART® interface.
• Multidrop mode is supportedMultidrop mode is supportedMultidrop mode is supportedMultidrop mode is supported
In a multidrop communication system, up to 32 devices can be connected to a common transmission cable.
• Burst mode is not supportedBurst mode is not supportedBurst mode is not supportedBurst mode is not supported
In the burst mode a slave device transfers cyclic pre-defined response telegrams, to get a higher rate of data transfer.
There are two ways of using the HART® communication:
• as Point-to-Point connection and
• as Multidrop connection, with 2-wire connection.
CAUTION!
For installation in a multidrop communication system please consider the voltage drop for the 250 Ω loop resistance for HART® communication. The supply voltage has to be adjusted.
8 DESCRIPTION OF HART INTERFACE
SMARTPAT COND 72008.3.1 Point-to-Point connection - analogue / digital mode
Point-to-Point connection between the sensor and the HART® Master.
The current output of the sensor is passive.
Figure 8-1: Point- to-Point connection (only one primary and one secondary master can communicate with the slave simultaniously)
1 Primary master with e.g. PACTwareTM FDT/DTM 2 FSK modem
3 HART® signal 4 SMARTMAC 200 W 5 SMARTPAT Sensor
6 Secondary master with HART® DD
7 Power supply for devices (slaves) with passive current output 8 Load ≥ 250 Ω (Ohm)
DESCRIPTION OF HART INTERFACE 8
SMARTPAT COND 7200
8.4 Inputs/outputs and HART ® dynamic variables and device variables
Code = device variable code
The HART® dynamic variable PV can be switched between conductivity and resistivity. If the PV is mapped to conductivity, the SV is mapped to resistivity and vice versa.
8.5 Field Communicator 475 (FC 475)
The Field Communicator is a hand terminal from Emerson Process Management that is designed to configure HART® and Foundation Fieldbus devices. Device Descriptions (DDs) are used to integrate different devices into the Field Communicator.
8.5.1 Installation
The HART® Device Description for the sensor must be installed on the Field Communicator.
Otherwise only the functions of a generic DD are available to the user and the entire device control is not possible. A "Field Communicator Easy Upgrade Programming Utility" is required to install the DDs on the Field Communicator.
The Field Communicator must be equipped with a system card with "Easy Upgrade Option". For details consult the Field Communicator User’s Manual.
HART® dynamic variable
PV SV TV
Conductivity (Resistivity) Resistivity (Conductivity) Temperature
HART® device variable Code Type
Conductivity 0 linear
Resistivity 1 linear
Temperature 2 linear
8 DESCRIPTION OF HART INTERFACE
SMARTPAT COND 72008.6 Field Device Tool / Device Type Manager (FDT/DTM)
A Field Device Tool Container (FDT Container) is basically a PC program used to configure a field device via HART®. To adapt different devices, the FDT container uses a so-called Device Type Manager (DTM).
8.6.1 Installation
If the DTM for the sensor has not yet been installed on the FDT Container, setup is required and is available for download from the website or on CD-ROM. See the supplied documentation for information on how to install and set up the DTM.
8.7 Overview Basic-DD menu tree (for Field Communicator 475)
Main menu Submenu
1 Measuring value 1 Measuring value1 Measuring value
1 Measuring value 1 Conductivity 2 Resistivity
2 Quick setup2 Quick setup
2 Quick setup 1 TAG 2 Manual hold 3 I/O
4 Calibration 3 Logbook
3 Logbook3 Logbook
3 Logbook 1 Calibration logbook 2 Error logbook 4 Setup
4 Setup4 Setup
4 Setup 1 Process input
2 I/O 3 I/O HART® 4 Device 5 Service
5 Service5 Service
5 Service 1 Service calibration 2 Service parameter 3 Sensor lock
DESCRIPTION OF HART INTERFACE 8
SMARTPAT COND 7200
8.8 Basic-DD menu tree (details for settings)
1 Measuring value
1 Conductivity Display of the measured conductivity value on process control system, HART® handheld or SMARTMAC 200 W.
2 Resistivity Display of the measured resistivity value on process control system, HART® handheld or SMARTMAC 200 W.
3 Temperature Display of the measured temperature value in °C / °F on process control system, HART® handheld or SMARTMAC 200 W.
4 Measurand Display of the measured value on process control system, HART® handheld or SMARTMAC 200 W.
5 Loop current Display of the measured loop current in mA on process control system, HART® handheld or SMARTMAC 200 W.
6 Error Display of status messages and diagnostic information. For further information refer to Status messages and diagnostic information on page 27.
7 Condensed state Display of status information icon for sensor according to NAMUR NE 107.
2 Quick setup
1 TAG 1 TAG (set and display TAG)
2 Long Tag (set and display Long TAG) 3 Previous Long TAG (read only)
2 Manual hold 1 Yes
2 NoActivate or deactivate the manual hold function.
3 I/O 1 Measurand
Conductivity Resistivity
2 Measuring value at 4 mA (set and display) 3 Measuring value at 20 mA (set and display)
4 Time constant (set and display time constant for 4...20 mA update rate (1...60 seconds, 1 second as default))
4 Calibration 1 Product calibration 2 Calibration solution 3 Input cell constant
Start calibration procedure. For more information refer to Calibration on page 21.
3 Logbooks
1 Calibration logbook 1 Memory block 2 Calibration method 2 Error logbook 1 Memory block
2 Error counter 3 Error 4 Error time
8 DESCRIPTION OF HART INTERFACE
SMARTPAT COND 72004 Setup
1 Process input 1 Temperature 1 Temperature unit (set and display) 2 Temperature offset (set and display) 3 Date of offset (read only)
4 Temp. compensation (set and display) (off, linear and natural water. Default: linear) 5 Temp. coefficient (set and display)
(Only displayed if temp. compensation is set to linear.
Default setting: 2%/K)
6 Reference temp. (set and display)
(Only displayed if temp. compensation is set to linear.
Default setting: 25°C / 77°F) 2 Calibration 1 Product calibration
2 Calibration solution 3 Input cell constant
For further information refer to Calibration on page 21.
3 Cell constant Display cell constant (read only)
4 Maintenance interval Set and display maintenance interval in days (0...999 days);
Default setting: 0
5 Time remaining Display of the remaining maintenance time (read only)
6 Reset interval 1 No
2 Yes
Reset maintenance interval. Default setting: No
2 I/O 1 Measurand
Conductivity Resistivity
2 Meas. value at 4 mA (set and display) 3 Meas. value at 20 mA (set and display)
4 Time constant (set and display time constant for 4...20 mA, update rate 1…60 seconds) Default setting: 1 second
3 I/O HART 1 TAG 1 TAG (set and display TAG)
2 Long Tag (set and display Long TAG) 3 Previous Long TAG (read only)
2 Message Set and display HART® message (32 packed ASCII) 3 Polling address 1 Polling address (display and input possibility for polling
address and loop current mode) 2 Loop current (read only)
4 Device variables 1 PV 1 Conductivity
2 Max. conductivity
DESCRIPTION OF HART INTERFACE 8
SMARTPAT COND 7200
4 Device 1 Information
(read only) 1 Sensor information 1 Order code
2 Device name
3 Operating parameters (OP) 1 Commissioning date 2 Operating hours 3 OP time > max OP t.
4 Max. OP temperature 5 Max. temperature 1 6 Temp. compensation 7 Temp. coefficient 2 8 Reference temp 2 2 Commissioning Set date of commissioning
1 - Maximum reached temperature.
2 - Displayed only if linear temperature compensation is selected.
5 Service
1 Service calibration 1 Trimming at 4 mA (+/-) 2 Trimming at 20 mA (+/-)
2 Service parameter 1 Reset (Reboot sensor) 1 No
2 Yes 2 Load factory settings
(If you select YesYesYes, the following settings will be reset to Yes default: temperature unit, measurand (PV), I/O, calibration interval, 4...20 mA trimming, TAG and Long TAG. Polling address will be set to 0 and loop current will be enabled.
1 No2 Yes
3 Sensor lock 1 Lock sensor
If you select yesyesyes the setting is not reversible anymore and yes the sensor is invalid for use. The HART®communication is not possible anymore.
1 No2 Yes
KROHNE – Process instrumentation and measurement solutions
•
Flow•
Level•
Temperature•
Pressure•
Process Analysis•
ServicesPAT COND 7200 R01 en - Subject to change without notice.