An example of a method of CT dose measurement using with a dosimetry

phantom and MOSFET dosimeters

The following is an example of an empirical method that was used to measure dose from CT examinations in a paediatric phantom. In this example 4 banks of 5 dosimeters were utilised. Figure 13 illustrates a paediatric phantom with the MOSFET dosimeters located in specific organs and tissues. With only 20 dosimeters available the phantom is loaded and

Figure (13) Loading the phantom with MOSFET detectors and scan set up.

irradiated in order, locations 1-20, then 21-40 and so on. The obtained values (mV/mGy) are sent to the computer via wireless network and saved as an Excel file sheet. This was then repeated many times until all of the 167 or 271 dose locations had mV readings for

paediatric and adult ATOM phantoms, respectively. Once the data is gathered they are divided by the respective calibration factors (Table 11) for each MOSFET dosimeter in order to determine organ and tissues absorbed dose.

References

1. Winslow JF, Hyer DE, Fisher RF, Tien

CJ, Hintenlang DE. Construction of anthropomorphic phantoms for use in dosimetry studies. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS. 2009;10(3).

2. Green S, Palethorpe JE, Peach D,

Bradley DA. Performance assessment of patient dosimetry services and X-ray quality assurance instruments used in diagnostic radiology. Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine. 1999;50(1):137-52.

3. Rivera T. Thermoluminescence in

medical dosimetry. Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine. 2012;71 Suppl:30-4.

4. Bos AJJ. Theory of

thermoluminescence. Radiation Measurements. 2007;41:S45-S56.

5. Bos AJJ. High senstivity thermoluminescene dosimetry. Nuclear Instruments and Methods in Physics Research B. 2001;184(1):3-28.

6. Bos AJJ. On the energy conversion

in thermoluminescece dosimetry material. Radiation Measurements. 2001;33(1):737-44.

7. Kitis G, Furetta C, Prokic M, Prokic

V. Kinetic parameters of some tissue equivalent thermoluminescence materials. Journal of Physics D: Applied Physics. 2000;33(11):1252.

8. Kortov V. Materials for thermoluminescent dosimetry: Current status and future trends. Radiation Measurements. 2007;42 (4-5):576-81.

9. Luo LZ. Extensive fade study of

Harshaw LiF TLD materials. Radiation Measurements. 2008;43(2-6):365-70.

10. Moscovitch M, Horowitz YS.

Thermoluminescent materials for medical applications: LiF:Mg,Ti and LiF:Mg,Cu,P. Radiation Measurements. 2007;41:S71-S7.

11. Olko P. Advantages and

disadvantages of luminescence dosimetry. Radiation Measurements. 2010;45(3-6):506-11.

12. European Commission. European

protocol on dosimetry in mammography. 1996.

13. Gaona E, Nieto JA, Góngora

JAID, Arreola M, Enríquez JGF. TL dosimetry for quality control of CR mammography imaging systems. Radiation Effects and Defects in Solids. 2007;162(10-11):759-63.

14. Venables K, Miles EA, Aird EG, Hoskin

PJ, Group STM. The use of in vivo thermoluminescent dosimeters in the quality assurance programme for the START breast fractionation trial. Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology. 2004;71(3):303-10.

15. Triolo A, Brai M, Bartolotta A, Marrale M. Glow curve analysis of TLD-100H irradiated with radiation of different LET: Comparison between two theo- retical method. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectro- meters, Detectors and Associated Equipment. 2006;560(2):413-7.

16. ThermoScientific. High Sensitivity LIF: Mg, Cu, P Thermoluminescent Dosimetry Materials, Chips 2015 [cited 2016]. Available from: https:// www.thermofisher.com/order/catalog/ product/SCP18815#%2Flegacy=ther- moscientific.com.

17. Camargo-Mendoza RE, Poletti

ME, Costa AM, Caldas LVE. Measurement of some dosimetric parameters for two mammography systems using thermoluminescent dosimetry. Radiation Measurements. 2011;46(12):2086-9.

18. Izewska J, Rajan G. Radiation dosimeters. In: Podgorsak EB, editor. Radiation oncology physics: A handbook for teachers and students. Vienna: IAEA; 2005.

19. Collins P. Assessment of effective dose in computed tomography using an anthropomorphic phantom. Galway: National University of Irland; 2005.

20. Savva A. Personnel TLD monitors,

their calibration and response. Surry: University of Surrey; 2010.

21. Lisa D, Claire H, Helen W-F, Mamoon

H, Tomas K. Variations in dose response with x-ray energy of LiF:Mg,Cu,P thermoluminescence dosimeters: implications for clinical dosimetry. Physics in Medicine and Biology. 2004;49(17):3831.

22. Furetta C, Weng PS. Operational

Thermoluminescence Dosimetry. Singapore: World Scientific; 1998.

23. Costa AM, Barbi GL, Bertucci EC,

Ferreira H, Sansavino SZ, Colenci B, et al. In vivo dosimetry with thermo- luminescent dosimeters in external photon beam radiotherapy. Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine. 2010;68(4-5):760-2.

24. Shirazi A, Mahdavi SR, Khodadadee A,

Ghaffory M, Mesbahi A. Monte Carlo simulation of TLD response function: Scattered radiation fi eld application. Reports of Practical Oncology & Radiotherapy. 2008;13(1):23-8.

25. Sulieman A, Theodorou K, Vlychou M,

Topaltzikis T, Kanavou D, Fezoulidis I, et al. Radiation dose measurement and risk estimation for paediatric patients undergoing micturating cys- tourethrography. The British journal of radiology. 2007;80(957):731-7.

26. Tootell AK, Szczepura KR, Hogg

P. Optimising the number of thermoluminescent dosimeters required for the measurement of effective dose for computed tomography attenuation correction data in SPECT/CT myocardial perfusion imaging. Radiography. 2013;19(1):42-7.

27. Olgar T, Bor D, Berkmen G, Yazar

T. Patient and staff doses for some complex x-ray examinations. Journal of Radiological Protection. 2009;29(3):393.

28. M. Ali R, England A, McEntee MF,

Hogg P. A method for calculating effective lifetime risk of radiation- induced cancer from screening mammography. Radiography. 2015;21(4):298-303.

29. Sze SM. Semiconductor devices:

Physics and technology. 2nd ed: Wiley India Pvt. Limited; 2008.

30. Baliga BJ. Fundamentals of Power

Semiconductor Devices. USA: Springer Science + Business Media, LLC; 2008.

31. Tamma PK. Selecting P-channel

MOSFETs for Switching Applica- tions 2013 [cited 2016]. Available

from: http://www.infineon.com/dgdl/

Infineon+-+Application+Note+-+Pow- er+MOSFETs+-+OptiMOS+P-chan- nel.pdf?fileId=db3a304341e- 0aed001420380cc13101b.

32. Cygler J, Tanderup K, Beddar S,

Pérez-Calatayud J. In Vivo Dosimetry in Brachytherapy. In: Jack Venselaar, Ali S. Meigooni, Dimos Baltas, Hoskin PJ, editors. Comprehensive Brachytherapy. Imaging in Medical Diagnosis and Therapy: Taylor & Francis; 2012. p. 379-96.

33. Manninen A-L, Kotiaho A, Nikkinen J,

Nieminen MT. Validation of a MOSFET dosemeter system for determining the absorbed and effective radiation doses in diagnostic radiology. Radiation Protection Dosimetry. 2015; 164(3): 361–367.

34. Best Medical Canada. Operator’s

manual for the mobile MOSFET system. Wireless dosimetry system for use with mobile MOSFET Software Versions 2.2 and above; 2007 [cited 2016]. Available from: HEALTH CANADA:wireless dosimetry system

35. Koivisto JH, Wolff JE, Kiljunen T, Schulze D, Kortesniemi M. Characterization of MOSFET dosimeters for low-dose measurements in maxillofacial anthropomorphic phantoms. Journal of Applied Clinical Medical Physics. 2015; 16(4).

36. Soubra M, Cygler J, Mackay G.

Evaluation of a dual bias dual metal oxide‐silicon semiconductor field effect transistor detector as radiation dosimeter. Medical Physics. 1994; 21(4): 567-572.

37. Cheung T, Butson MJ, Yu P. Energy

dependence corrections to MOSFET dosimetric sensitivity. Australasian Physics & Engineering Sciences in Medicine. 2009; 32(1): 16-20.

38. Lavallee M-C, Gingras L, Beaulieu

L. Energy and integrated dose dependence of MOSFET dosimeter sensitivity for irradiation energies between 30kV and Co60. Medical Physics. 2006; 33(10): 3683-3689.

Part 2

Empirical research conducted