• No results found

3.3 Experimental aspects

3.3.4 Experimental setup fluorescence mode

XAS spectrum is normally measured in three modes: transmission, fluorescence or electron yield. The transmission mode is the most straightforward among the three. It simply detects the X-ray flux of the beam using ion chambers, both before (I0) and after (It) it passes through the sample.I0 and It thereby have a common energy (E) dependence,

the absorption coefficient µ(E) can be easily calculated by Equation 3.1. However, this measurement mode requires the samples to be homogeneous and have a constant thickness. And most important of all, the absorbing elements have to be concentrated enough such that the difference between I0 and It is significantly larger than the variation due to counting statistics [99].

Figure 3.8: Decay of the excited state: X-ray fluorescence (left) and Auger effect (right) [100].

3.3. Experimental aspects 57 the electron yield mode, where the emitted fluorescent X-rays or the emitted electrons due to X-ray absorption are detected respectively to calculate theµ(E) value. The emissions of both fluorescence X-rays and Auger electrons are caused by the decay of the core hole after its finite life time. The core hole is filled by an electron at the higher orbital within∼10−15 seconds after the photoelectric event described in Section 3.1.1 occurs. The transition of the electron leads to two types of phenomena as sketched in Figure 3.8. In X-ray fluorescence, an X-ray with a well-defined energy is emitted, which is named according to the levels the electron transmits from and to. For example, a Kα fluorescence line is created by the electron that drops from the L to the K shell of an atom. In the Auger effect, a second electron is emitted into the continuum when the core hole is filled. Both of the effects are related to the X-ray absorption of the sample in some manner, thus the two corresponding XAS measurement modes, fluorescence and electron yield, are found. The likelihood of the two mechanisms to occur in a decay are energy dependent, X-ray fluorescence and Auger effect dominate the high (>3 keV ) and low energy (<3 keV) regions, respectively [101]. Considering most of the samples in this thesis are dilute (absorbing atoms < 1 at. %) and the absorption edge selected in this thesis is In K-edge (27.94 keV [114]), the X-ray fluorescence mode was setup in the Australian synchrotron XAS beamline. Herein, we intuitively treat µ(E) ∝ If/I0, where If is the monitored intensity of a fluorescence

line that is associated with the absorption process [100]. The actual calculation of µ(E) utilizingIf is certainly more complicated, details of which can be found in References [100] and [101].

Figure 3.9: A schematic of the fluorescence mode XAS measurement setup in the XAS beamline of the Australian Synchrotron.

In order to detect the X-ray fluorescence signal and measure the supplemental X-ray fluxes, the experimental section of the XAS beamline in Australian Synchrotron is config- ured as in Figure 3.9. All the components are in an interlocked and radiation-protected hutch, into which the monochromatic X-ray beam enters. The incoming X-ray intensity (I0) is firstly measured by an ion chamber filled with Helium before interacting with the sample. The sample is mounted on a vibration minimized cryostat and maintained at a

temperature of 18 K to reduce thermal disorder, with a tilt of 45◦ respect to the incoming X-ray beam. When the X-ray beam hits the sample, the fluorescence yield is normally highest in a direction perpendicular to the incident beam, where the 10× 10 pixel-array Ge solid state detector is positioned. The flux of the transmitted radiation is detected in the second ion chamber, noted asI1, mainly used for sample alignment in the fluorescence mode. For the energy calibration in the subsequent data analysis, the XAS spectrum of the reference sample is simultaneously measured in a transmission mode utilizing the third ion chamber (with an X-ray intensity ofI2). In this thesis, a uniform In foil was the reference sample material used for the XAS measurements. Multiple scans (3 - 6 times) were carried out for each sample and averaged afterward to maximize the signal to noise ratio.

With an experimental setup as in Figure 3.9, the Ge fluorescence detector not only receives the characteristic X-ray signal but also a significant amount of noise;

(1) a peak due to the elastic scattering of the incident X-ray beam with the sample,

(2) a Compton (inelastic) scattered radiation from the weakly bound electrons,

(3) ,,, etc. lines from all the elements with edges below the incident photon

energy in the sample, kapton tape, sample substrate, contamination in the cryostat, etc.

Figure 3.10 shows a spectrum of the recorded X-ray intensity in an actual experiment, where the In fluorescent line is present together with large Compton and elastic scat- tered X-ray signals. The fluorescence X-rays from the atoms other than In are also visible in the low energy channels.

The sample preparation procedure described previously is for the purpose of reducing this noise, but in most cases, additional approaches are necessary to isolate the desired fluorescent line from the background. The Si or Ge solid state detectors have the advantage of being able to record a wide energy range X-ray fluorescence spectrum. The X-ray signals from individual origins can be identified and quantified as shown in Figure 3.10. With the detector resolution of ∼ 200 eV, a region of interest can be set on the fluorescence spectrum to isolate the desired fluorescence signal (e.g. the In in Figure 3.10) from the background. One of the drawbacks of this type of detector is that determining the energy of each photon takes the detector a finite amount of time, during which another X-ray hitting the detector is not registered as a separate X-ray. This “deadtime” effectively limits the count rate. It is common to couple independent detectors to form multiple-element detectors to improve the signal to noise ratio. The one in the XAS beamline of Australian Synchrotron, for instance, is a 100 element detector in a 10× 10 array.

This signal isolating approach is sufficient for samples with concentrated absorbers, but less so for the dilute sample whose region of interest is heavily contaminated by the tail of the Compton scattering signal, as in Figure 3.10. The situation can be improved by introducing a Z −1 or Z −2 filter between the sample and the detector, where Z is

3.4. Data processing 59