• No results found

The pH sensors - transmitters type CH 221 and CH 228 (I or T) are shipped and stocked without filling solution in the reference electrode's chamber.

Before operation, the reference electrode must be filled with the reference electrolyte solution type BF 121 (3M KCl / sat AgCl), that comes with the sensor.

Unscrew and remove the electrodes' plastic protection guard from the lower part of the sensor.

Avoid touching the glass membrane.

Unscrew carefully the liquid junction module at the lower part of the reference electrode.

Ensure that the sealing o-ring at the end of the thread of the liquid junction module is slightly lubricated (with sili-cone grease)

Discard the old filling solution (if any) and rinse the reference electrode's chamber first with distilled or de-ionized water and then with a small volume of the refer-ence electrolyte.

Fill the reference electrode's chamber with the reference electrolyte solution, type BF 121.

Place the liquid junction module to the reference elec-trode, screw carefully until it stops and tighten by hand.

Avoid to touch the glass electrode and the glass fibres of the liquid junction.

Place (screw) the electrodes' plastic protection guard at the lower part of the sensor.

For the type "T" sensors - transmitters, place the plastic protection guard to perform the calibration procedure.

Then remove the protection guard and fix the sensor -transmitter into the thread of the "T" adaptor (see chapter 4.4, page 14)

The filling solution of the reference electrode should be replaced every 2 - 4 months, depending on the specific application.

Ò

4.3. Installation of the type "I" sensor - transmitter

The type "I" sensor - transmitter is designed for easy installation in open-surface basins and reservoirs or closed tanks.

For proper performance of the pH sensor, (and the whole system of course) the installer has to consider the following :

- The location of the pH sensor - transmitter in the tank is of critical importance.

- The pH and the reference electrode (at the lower part of the transmitter) should be immersed in a depth of 10 to 80 cm below water surface. Special care must be taken that the water level is not allowed to drop below the measuring electrodes or to raise above the upper end of the sensor - transmitter.

- Water must flow adequately around the electrodes and be bubble free. When air bubbles adhere to the electrodes' sensing tips, the instrument indicates erratic pH values.

- The sensor's location on the tank-wall has to be easily accessible and the transmitter must be easily removable, for convenient maintenance, cleaning and recalibration.

For installation of the type "I" pH sensor - transmitter on the tank wall, it is recommended to use a metal frame as a sensor holder (see fig. 7 - 2).

For the construction of the holding frame may be used a rectangular cut out profile (approx.

dimensions 20 x 30 mm) from aluminium, galvanised iron or stainless steel.

1

1. pH sensor - transmitter 2. Sensor's metal holding frame 3. Plastic stop- ring

4. Plastic holders for 40 mm PVC rigid tubes

5. Upper permitted water level 6. Lower permitted water level

Connection of the electrodes' cleaning system, type SCL 20 7. Solenoid valve

8. Electrodes' cleaning head, type SCB 21 I

9. Clean water inlet

10. Plastic tube (PE - 6mm), connect-ing the cleanconnect-ing head with the solenoid valve

11. Holders of the 6mm plastic tube

Figure 7 : Installation of the CH 221 I sensors - transmitters in open tanks

To install the CH 221 I pH sensor - transmitter on the tank wall, proceed as follows :

- Fix the two 40mm plastic tube - holders on the metal frame, 40 - 50 cm apart from each other (fig. 7 - 4).

- Weld the stop ring on the sensor body in the proper position (fig. 7 - 3), using special adhesive for PVC - rigid material.

The metal holding frame may be fixed on the tank wall with 2 - 3 screws, as shown at fig. 7.

Then you must connect (if available) the automatic electrodes' cleaning system type SCL 20.

If the pH sensor - transmitter has been ordered with the type SCL 20 electrodes' cleaning system, the cleaning head and the 6mm tube holders are already fixed on the sensor and the steps 1 - 3 must be omitted.

1. Unscrew and remove the electrodes' plastic protection guard from the lower part of the sensor.

(see fig 5 - 11).

2. Using a special adhesive for PVC - rigid material, fix the type SCB 21 I electrodes' cleaning head (fig. 7 - 8) at the lower end of the plastic protection guard, so that two (at least) of the water jets are directed to the glass surface and the glass fibres of the reference electrode.

CAUTION ! Do not block from excess glue the small holes (nozzles) at the inner side of the cleaning head.

Place (screw) the electrodes' plastic protection guard at its original place at the lower part of the sensor.

3. Using a special adhesive for PVC - rigid material, fix the two plastic holders of the 6 mm tube on the sensor's body, aligned with the water inlet of the cleaning head, 40 - 60 cm apart from each other (fig. 7 - 11).

4. Fix the solenoid valve (fig. 7 - 7) at the outer side of the tank's wall, or at a proper place close to the sensor.

5. Install and connect the control panel of the type SCL 20 electrodes' cleaning system (see chapter 8.2).

6. Connect the 6mm water tube (fig. 7 - 10) with the cleaning head and the solenoid valve.

7. Connect the clean water supply (fig. 7 - 9) to the input of the solenoid valve. For proper function of the automatic cleaning system, the water pressure at the input of the solenoid valve should be 2 - 4 bar (for output to 6 / 4 mm tube, length 2 - 2.5 m). If the pressure of the clean water network is higher, it is recommended to connect a control valve in series with the solenoid valve.

8. Connect the solenoid valve to the 230 Vac output of the SCL 20 control unit.

CAUTION ! Do not activate the cleaning system when the sensor is not immersed into the water. The pressure of the water jet in air is considerably higher than that in water and may damage the glass electrode.

For installation of the CH 221 I sensor - transmitter alone on the tank wall (i.e. without the automatic electrodes' cleaning system) the following procedure is recommended :

- Fix two rings of inner diameter slightly more than 40mm on the tank-wall, 40 to 60 cm apart from each other, aligned vertically. The rings are to be placed in such way to allow the transmitter to move freely through them.

- Weld the stop ring on the transmitter in the proper position, using special adhesive for PVC-rigid material.

After installation, attach the waterproof plug of the CL 200 interconnection cable to the socket at the top of the transmitter :

a. Unscrew the cable gland of the waterproof plastic cover.

b. Confirm that both O-rings, at the upper part of the sensor and inside the cable gland, are lubricated properly (silicon grease is recommended).

c. Connect and secure (by its own ring screw) the metal plug on the transmitter.

d. Screw the waterproof plastic cover on the sensor, by hand, tightly.

e. Finally, tighten the cable gland.

4.4. Installation of the type "T" sensor - transmitter

The CH 221 T sensor - transmitter is designed for installation at pipe networks, connected in series on the flow or bypass.

For proper performance of the pH sensor - trans-mitter, (and the whole system of course) the installer has to consider the following:

- The pH sensor - transmitter has to be placed in a vertical position (a slope of up to 30° is acceptable), in a location where the water pressure is kept below 2 bar, the maximum temperature is up to 60 °C and the flow velocity is up to 30 cm/sec.

It is recommended to install the sensor (if possible) close to a free outlet, so that the pressure is kept as low as possible and eliminate the risk of momentary pressure increase (kickbacks) over the sensors' speci-fied range.

- Water must flow adequately around the electrode and be bubble free. When air bubbles adhere to the electrode's sensing tip, the instrument may indicate erratic pH values.

- When the water flow is interrupted, some water must remain into the "T" type sensor's Figure 8 : Installation of the type CH 221 T sensors at pipe network.

- The sensor's location should be easily disconnected from the flow (using control valves before and after the "T" type holder) and should be easily accessible, so that the sensor - transmitter can be easily removed for convenient maintenance, cleaning and recalibra-tion.

The following method for fixing the pH sensor - transmitter on pipe-networks is recommended:

1. Fix the "T" shaped PVC adaptor, supplied along with the pH transmitter, on the water pipeline (or on a parallel line connected to it ), using special adhesive for PVC - rigid material.

If the type SCL 20 electrodes' cleaning system is to be used, install the "T" shaped adaptor that comes with the cleaning system (with the cleaning head fixed on it).

2. Confirm that the O-ring at the lower part of the sensor body is lubricated properly (silicon grease is recommended).

3. Screw the lower part of the transmitter (electrode side) into the thread of the "T" adaptor.

4. Attach the waterproof plug of the CL 200 interconnection cable to the socket at the top of the transmitter :

a. Unscrew the cable gland of the waterproof plastic cover.

b. Confirm that both O-rings, at the upper part of the sensor and inside the cable gland, are lubricated properly (silicon grease is recommended).

c. Connect and secure (by its own ring screw) the metal plug on the transmitter.

d. Screw the waterproof plastic cover on the sensor, by hand, tightly.

e. Finally, tighten the cable gland.

For connecting the automatic electrodes' cleaning system type SCL 20 (if available), proceed as follows :

1. Fix the solenoid valve at a proper position close to the sensor.

2. Install and connect the control panel of the type SCL 20 electrodes' cleaning system (see chapter 8.2).

3. Connect the 6mm water tube with the cleaning head and the solenoid valve.

4. Connect the clean water supply to the input of the solenoid valve. For proper function of the automatic cleaning system, the water pressure at the input of the solenoid valve should be 2 - 4 bar higher than the pressure of the water into the pipe network (for output to 6 / 4 mm tube, length 2 - 2.5 m). If the pressure of the clean water network is higher, it is recommended to connect a control valve in series with the solenoid valve.

5. Connect the solenoid valve to the 230 Vac output of the SCL 20 control unit.

CAUTION ! Do not activate the cleaning system when the sensor is not immersed into the water. The pressure of the water jet in air is considerably higher than that in water and may damage the glass electrode.

5. ELECTRICAL CONNECTION

- The notes and warnings in these operation instructions must be strictly adhered to ! - Maintenance work when the unit is connected to mains may only be carried out by

qualified personnel !

- Any faults in the instrument may be remedied with the aid of the instructions given to this users manual without requiring intervention in the instrument itself.

- Interventions in and modifications to the instrument are impermissible and will void the warranty.

- If faults cannot be remedied, the instrument must be removed from service and secured to prevent accidental start - up.

- Repairs may be performed at the manufacturer's works or by authorised service only.

After fixing the instrument, electrical connections of mains, switching contacts, transmitter and signal lines are to be made :

- to the terminals on the rear panel of the type PH 222 (panel mounting) instrument - see page 6, fig. 3 (11, 12, 13 and 14)

or

- to the terminals on the terminals' board, under the lower non transparent cover, of the type PH 222 / W (wall mounting) instrument - see fig. 9, page 16.

1. Connect mains 230 Vac to terminals 11 (P1) and 12 (P2) (11: L, 12 : N).

It is recommended that the panel mount instrument is connected to mains via an external fuse (0.5 A) and an external switch. If there are more than one measuring and control instruments on the panel, a common fuse may be used, but it is suggested to use individual switches for every instrument (or group of instruments).

If the mains suffers frequent overvoltage problems ( > 270V ) or voltage fluctuations ( > ±15% ), it is recommended to use a mains voltage stabiliser (or a UPS) or at least, for lower cost, a mains voltage guard that will interrupt the power supply to the instrument if the input voltage exceeds 260 Vac.

The instrument is equipped with an internal voltage stabiliser and protection from momentary overvoltage - 275V varistor -, but if the mains voltage fluctuates more than ±15% the system may not function properly. In case that the mains voltage exceeds 270 Vac, a permanent damage may occur to the instrument.

2. Connect the sensor's interconnection cable (type CL 200) to terminals S1 - S5 (1 - 4), according to the following colour code :

S1 (1) : green (or yell/grn) neg. power supply -5V S3 (2) : gray (or black) analogue ground

S4 (3) : brown pH signal (± 700 mV for 0 to 14 pH) S5 (4) : pink (or red) pos. power supply +5V

(For your convenience the colour code is printed on the rear panel of the instrument as

3. Connect the "external commands" transfer cable to terminals 7 - 8 (O1 - O2) & 9 - 10 (O3 - O4) according to the indications on the rear panel.

7 - 8 (O1 - O2) : output 1 (controlled by the low limit) 9 - 10 (O3 - O4) : output 2 (controlled by the low limit)

Both outputs are normally open relay contacts and they are specified for currents up to 5A / 250 Vac (1250VA). If the application requirements are higher, an appropriate power relay must be employed. Each output of the PH 222 / W instrument is protected via an internal 6A fast-blowing fuse.

4. Connect the "signal transfer" cable (to recorders, PLC’s or auxiliary control units) to terminals 5 (+Vo) and 6 (-Vo), according to the indications on the instrument’s rear panel.

5 (+Vo) : pH output signal (+100mV /pH , 0 to 1400 mV) 6 (-Vo) : Analogue ground of voltage output

WARNING ! USE ONLY SCREENED CABLE for the transfer of the analog output signal.

Shield must be connected either to terminal 6 of the PH 222 instrument (analog gnd) or to sh of the PH222/W instrument.

DO NOT CONNECT terminals 5 and 6 (analogue voltage output) to power earth ground (PE) directly or indirectly, through connection to external devices. If this type of connection cannot be avoided (because the negative analog input of the external device is grounded), we suggest to use an analog signal isolator as the ISO 30 (see ch. 9, page 25)

Figure 9 : Electrical connections terminals' board of the PH 222 / W instrument.

pH - CONTROLLER MODEL : PH 222 /W

XENON SCIENTIFIC INSTRUMENTS Co 15 Delphon str., Chalandri, Athens 152 33, GREECE

LINE POWER FUSE 0.1 A

OUT 1 : Low limit (alkali pump) OUT 2 : Hi limit (acid pump) DO NOT CONNECT the mV analog output to line power earth (PE).

pH mV

0 0

7 700 14 1400

6. INSTRUMENT's CALIBRATION

6.1. Calibration of the pH measurement

For proper function, the instrument must be calibrated with the pH sensor - transmitter in use.

The calibration procedure should be performed in the following cases : - At start up.

- After electrolyte replacement to the reference electrode.

- At regular intervals during normal operation (every 1 to 3 months, depending on the application) The PH 222 controller has a two point calibration capability - i.e. asymmetry (ÄpH ) and slope (mV/pH). For the calibration of the instrument, two standard solutions are required:

1. A neutral buffer solution (pH = 7.00 ±0.5).

2. An acid (pH=4.00) or an alkaline (pH=9 or 10) buffer solution.

It is recommended that the temperature of the buffer solutions is close - within ±10 °C - to the temperature of the solution to be monitored.

For the calibration of the instrument, the following procedure must be carried out:

- Connect the instrument with the pH sensor - transmitter and to mains (see ch. 5, page 15).

- Ensure that both the relay outputs are not connected to any device under power.

- Ensure that the reference electrode is filled with the reference electrolyte solution (see chapter 4.2).

- Rinse the electrodes with distilled or deionized water to remove any salt deposits.

- If the sensor is going to be used for the first time, after a period of service or if the electrode is just refilled, the electrodes must be reconditioned. Immerse the electrodes in clean water and let the sensor to operate for at least 2 hours before performing any calibration.

- Immerse the lower part of the sensor - transmitter (the pH electrode, the reference electrode and the temperature sensor as well) into the neutral buffer solution (pH=7).

- Set the selector switch (2) on the front panel (fig. 2) to "meas" and wait for a while (5 - 10 minutes) until the displayed value remains constant.

- Using a small screwdriver, adjust the asymmetry - ÄpH - pot (fig. 2 - 3) for a reading equal to the pH value of the buffer solution (e.g. 7.00).

- After asymmetry calibration, rinse the electrodes with distilled or deionized water and wipe away adhering drops on them with a soft tissue.

- Immerse the lower part of the transmitter (the pH electrode, the reference electrode and the temperature sensor as well) into an acid (e.g. pH 4) or an alkaline (e.g. pH 10) buffer solution.

Note ! It is a good practice to calibrate the instrument with a buffer solution suitable to the desired pH level i.e. acid buffer for acid solutions and vice versa.

- Wait for a while (5 - 10 minutes) until the displayed value remains constant.

- Using a small screwdriver, adjust the slope - mV/pH - pot (fig. 2 - 4) for a reading equal to the pH value of the buffer solution (e.g. 4.00).

6.2. Setting the limit values

The PH 222 controller features two switching outputs (normally open relay contacts), that can be used as a control signal to external devices.

Both set points can be adjusted accurately at any value within the measuring range and can function the one as an lower and the other as a upper limit to the control process. Moreover the instrument offers continuously adjustable settings for hysteresis (dead band) on control activa-tion.

The low limit activates the relay of output 1 if the measured pH value drops below the pre-defined value (set low - dead band) and the <OUT 1> led indicator (fig. 2 - 6) turns on. When the measured pH value rises above the pre-defined value (set low + dead band), the relay of output 1 is deactivated and the <OUT 1> led indicator turns off.

The hi limit activates the relay of output 2 if the measured pH value rises above the pre-defined value (set hi + dead band) and the <OUT 2> led indicator (fig. 2 - 9) turns on. When the measured pH value drops below the pre-defined value (set hi - dead band), the relay of output 2 is deactivated and the <OUT 2> led indicator turns off.

To set the low limit, switch the selector (2) on the front panel (fig. 2) to the "set LO" position and, with a small screwdriver, adjust the SET LOW LIMIT pot (fig. 2 - 5) for indication of the desired value.

To set the hi limit, switch the selector (2) on the front panel (fig. 2) to the "set HI" position and, with a small screwdriver, adjust the SET HI LIMIT pot (fig. 2 - 8) for indication of the desired value.

To set the hysteresis (dead band) of the control activation for the low limit, adjust the low limit's

"dead band" regulator (fig. 2 - 7) to the desired value on the dial printed on the front panel of the instrument.

To set the hysteresis (dead band) of the control activation for the hi limit, adjust the hi limit's

"dead band" regulator (fig. 2 - 10) to the desired value on the dial printed on the front panel of the instrument.

Hysteresis values can be set from 0 to 1 pH units and define the distance from the set point where the control activation or deactivation takes place.

Practically the hysteresis splits each set point in two control points (upper = [set + dead band]

and lower = [set - dead band] ), which are placed symmetrically above and below it. The control command is activated (or deactivated) when the measured pH reaches the one value and the opposite occurs when reaching the other.

7. ELECTRODES' MAINTENANCE AND CLEANING

For proper function of the instrument, the pH sensor has to undergo inspection and maintenance frequently, since it is the most sensitive and critical part of the whole system. It is worth to say that the pH sensor is the most frequent reason for improper function of the system.

The most common causes for malfunction of the pH sensor are faulty installation and poor maintenance.

The robust construction and the electrodes' design of the types CH 221 sensors - transmitters, provide, with proper maintenance, extra long service life (several years) and reliable operation under conditions that conventional pH sensors fail to operate.

The electrodes' maintenance must be carried out as follows :

- Every 1 to 4 days (depending on the application), the pH sensor - transmitter has to be removed from the basin and washed with clean water - especially the pH and the reference electrode.

If there are deposits on the glass membrane or on the reference's liquid junction, remove them by immersing the electrodes in dilute hydrochloric acid or, if the deposits consist of

If there are deposits on the glass membrane or on the reference's liquid junction, remove them by immersing the electrodes in dilute hydrochloric acid or, if the deposits consist of

Related documents