SECTION 8: QUALITY ASSURANCE AND QUALITY
8.2 What are the on-going QA test requirements in Part 75 for units reporting
Year-round reporting of emissions data is required for all Acid Rain Program units, all CAIR SO2 Trading Program units, and for certain NOx Budget Program and CAIR NOx Trading
Program units (see Section 8.5, below). For CEMS, the on-going QA test requirements for year- round reporters are summarized in Table 17. Table 17 shows that routine QA testing of CEMS is required at four basic frequencies:
C Daily; C Weekly; C Quarterly; and
C Semiannual or Annual.
Calibration error checks of all monitors and interference checks of flow monitors are required daily. For Hg CEMS with converters, system integrity checks are required weekly. Gas monitor linearity checks, flow-to-load ratio tests, and leak checks (for DP-type flow monitors) are required quarterly. RATAs are required either semiannually or annually, depending on the type of
monitor and/or the results of the tests (see Section 8.6, below).
For Appendix D fuel flowmeters, the basic frequency for the required accuracy tests is annual. For Appendix E systems, NOx emission testing is required once every five years, in order
to develop new correlation curves.
Table 18 summarizes the on-going QA requirements for sorbent trap monitoring systems.
Table 17: On-Going QA Test Requirements for Year-Round Reporters
Perform this type of QA test....
On these continuous monitoring systems....
At this frequency....
With these qualifications and exceptions....
Calibration error test Gas and flow monitors Daily C Calibrations are not required when the unit is not in operation.
Interference check Flow monitors Daily C Check is not required when the unit is not in operation.
System integrity check (single-level)
Hg CEMS with converters Weeklya C Not required if daily calibrations are
done with a NIST-traceable source of oxidized Hg
Linearity check Gas monitors Quarterly C Required only in “QA operating quarters”b and only on the range(s)
used during the quarter---but no less than once a year
C 168 operating hour grace period available
C Not required if SO2 or NOx span is
< 30 ppm
C For Hg monitors, you may perform a 3-level system integrity check using oxidized Hg standards, in lieu of this test
System integrity check (3-level)
Hg CEMS with converters Quarterly C For Hg monitors, you may perform a linearity check using elemental Hg standards, in lieu of this test Flow-to-load ratio or
gross heat rate test
Flow monitors Quarterly C Required only in “QA operating quarters”
C Non load-based units are exempted C Complex configurations may be
exempted by petition under §75.66 Leak check Differential pressure-type
flow monitors
Quarterly C Required only in QA operating quarters
C 168 operating hour grace period available
Table 17 (cont’d)
Perform this type of QA test....
On these continuous monitoring systems....
At this frequency....
With these qualifications and exceptions....
RATA and Bias test
Gas and flow monitors
(Bias test applies to SO2, NOx,
Hg, and flow monitoring systems, only)
Semiannual or Annualc
C Not required for SO2 monitors if the
unit exclusively burns very low sulfur fuel, or burns higher-sulfur fuel for < 480 hours per year
C 720 operating hour grace period available
C For Hg monitoring systems, the RATA frequency is always annual
Flowmeter Accuracy test Fuel flowmeter systems
Once every four “fuel flowmeter QA operating quarters”d
C The optional “fuel flow-to-load ratio” or “gross heat rate” test in Appendix D, section 2.1.7 may be used to extend the interval between flowmeter accuracy tests to up to 20 quarters
Primary element visual inspection
Orifice, nozzle, and venturi- type fuel flowmeters that are certified by design
Once every 3 years
(12 calendar quarters)
C The optional fuel flow-to-load ratio or gross heat rate test may be used to extend the interval between visual inspections to up to 20 quarters
NOx emission rate
testing
Appendix E systems Once every 5 years (20 calendar quarters)
---
a “Weekly” means once every 168 unit operating hours
b That is, a quarter with at least 168 hours of unit operation
c Depending on the % relative accuracy obtained in the previous test, the next RATA is required either
“semiannually” (within 2 QA operating quarters) or “annually” (within 4 QA operating quarters), not to exceed 8 calendar quarters between successive tests.
d
Table 18. Quality Assurance/Quality Control Criteria for Sorbent Trap Monitoring Systems
QA/QC Test or
Specification Acceptance Criteria Frequency
Consequences if Not Met
Pre-test leak check #4% of target sampling rate Prior to sampling
Sampling shall not commence until the leak check is passed Post-test leak check #4% of average sampling rate After sampling Sample
invalidateda Ratio of stack gas flow
rate to sample flow rate
Maintain within ± 25% of initial ratio from first hour of data collection period
Every hour throughout data collection period
Case-by-case evaluation
Sorbent trap section 2 breakthrough
#5% of Section 1 Hg mass Every sample Sample invalidateda Paired sorbent trap
agreement
#10% Relative Deviation (RD)
Every sample Sample invalidateda Spike recovery study Average recovery between
85% and 115% for each of the 3 spike concentration levels
Prior to analyzing field samples and prior to use of new sorbent media
Field samples shall not be analyzed until the percent recovery criteria has been met
Multipoint
analyzer calibration
Each analyzer reading within ± 10% of true value and r2 $0.99
On the day of analysis, before analyzing any samples
Recalibrate until successful
Analysis of independent calibration standard
Within ± 10% of true value Following daily calibration, prior to analyzing field samples
Recalibrate and repeat independent standard analysis until successful
Spike recovery from section 3 of sorbent trap
75-125% of spike amount Every sample Sample invalidateda RATA RA #20.0%
or Mean difference # 1.0 :g/dscm for low emitters
For initial certification and annually thereafter
Data from the system are invalidated until a RATA is passed
Dry gas meter calibration
(At 3 orifice settings initially, and 1 setting thereafter)
Calibration factor (Y) within ± 5% of average value from the initial (3-point)
calibration
Prior to initial use and at least quarterly thereafter
Recalibrate the meter at three orifice settings to determine a new value of Y
Temperature sensor calibration
Absolute temperature measured by sensor within ± 1.5% of a reference sensor
Prior to initial use and at least quarterly thereafter
Recalibrate. Sensor may not be used until specification is met.
Table 18 (cont’d)
QA/QC Test or
Specification Acceptance Criteria Frequency
Consequences if Not Met
Barometer calibration Absolute pressure measured by instrument within ± 10 mm Hg of reading with a mercury barometer
Prior to initial use and at least quarterly thereafter
Recalibrate. Instrument may not be used until specification is met
a
And data from the pair of sorbent traps are also invalidated