5 Hardware and Software Statuses, Spare Parts Lists
5.3 ME402 Hardware Configurations and their Components
5.3.2 Hardware Configuration 02
The main board C40-1925 has been replaced by the new version C40-2165 which, however, is not backward-compatible with HW00 and HW01, and the power manager board C40-1927 has been replaced by the new version C40-2169 (also not backward-compatible with HW00 and HW01). Apart from a newly added electronic “version recognition” function, these two modules do not provide any new functions. The connector plugs of choke L6 have been modified. This also necessitated a new version II of the monopolar board (C40-2224), which differs from ver-sion I only with regard to plug connector X4, and a new verver-sion II of the monopolar socket module, also fitted with a new connector.
The bipolar board I (C40-1930) has been replaced with the bipolar board II (C40-2155). Board II has a new impedance monitor offering a higher resolution for higher tissue resistances – a change necessitated by the introduction of the Auto Start/Stop functions. Moreover, the bipolar board II supports the use of the bipolar port expander, which allows connecting two instru-ments (for bipolar cutting and bipolar coagulation) to the bipolar output B2 in the Martin (“m”) version. This is controlled via the instrument recognition function that is available from
SW1.336.
To ensure that the old hardware configuration can be distinguished from the new one, an elec-tronic module recognition function has been integrated for the major boards. This function also allows displaying the hardware versions and serial numbers of the various modules for servic-ing purposes.
Serial numbers: ME402x0201 050600 to ME402x0201 071499
Lowest software version required: 1.336
Spare Parts List for HW02
No. Designation Item/Order No.
1 Housing top, preassembled 08-012-00-50
2 Housing bottom, preassembled 08-012-00-51
3 Power cord connector assembly 08-024-00-60
4 Fuse holder for power cord connector assembly 08-034-00-01 5 Footswitch socket for double-pedal footswitch w. cable 08-010-01-03 6 Footswitch socket for single-pedal footswitch w. cable 08-010-01-04 7 Connecting socket for maxium® Argon Beamer w. cable 08-001-00-62 8 Serial interfaces connector assembly w. cable 08-001-00-63
9 Equipotential bonding connector 08-027-00-01
10 Loudspeaker w. cable 08-020-00-21
11 Fan w. cable 08-020-00-22
12 Housing foot 08-012-00-52
13 Printed circuit board guide 08-020-00-23
14 Plastic optical fiber 08-018-00-58
15 Internal connecting cable (mains connector socket to power
elec-tronics boards) 08-041-00-51
16 High-voltage cable to monopolar board 08-024-00-66
17 Insulating cover for power manager board 08-016-01-01
18 Set of PE cables 08-032-00-20
20 Front frame w. MENU membrane button 08-010-01-05
21 Front glass w. gaskets 08-010-01-06
22 ON/OFF button w. cable 08-008-00-44
23 Connecting cable for display screen 08-001-00-64
24 Connecting cable for front board 08-001-00-65
25 Connecting cable for activation indicators (LEDs) 08-001-00-66 26 Connecting cable for NE status indicators (LEDs) 08-001-00-67 27 Connecting cable for rotary switch backlight 08-001-00-68
28 Pressure plate for screen cable 08-001-00-69
30 Rotary switch unit, complete 08-006-00-21
31 Rotary switch 08-006-00-22
32 Rotary button 08-006-00-23
33 Hub for rotary button 08-024-00-61
34 Backlight ring for rotary switch 08-003-00-33
35 Radial bearing socket for rotary switch 08-025-00-05
36 Fixing bracket 08-024-00-63
No. Designation Item/Order No.
40 Socket panel II for Martin accessories (“m” version), complete 08-003-01-26 41 Socket module carrier, “m” version, with membrane foil and
channel selector buttons
08-003-01-01
42 Light distributor, “m” version 08-021-00-03
43 Monopolar socket module II, “m” version (M1 and M2) 08-022-01-30 44 Bipolar connecting socket (B1), small, “m” version, w. cable and
flange 08-003-00-24
45 Bipolar connecting socket for instruments (B2), “m” version, w.
cable 08-003-00-25
46 NE connecting socket, “m”/“e” versions, w. cable and insulating
cup 08-024-00-34
47 Optical guide for NE status indicator, “m”/“e” versions 08-021-00-01 48 Optical guide for activation indicators 08-021-00-02 49 Insulating cover for monopolar HF socket module 08-016-01-02
50 Socket panel II for international and US-market accessories, “i”
version, complete 08-003-01-27
51 Socket module carrier, “i” version, with membrane foil and
chan-nel selector buttons 08-003-01-02
52 Light distributor, “i” version 08-021-00-04
53 Monopolar socket module II, “i” version (M1 and M2) 08-022-01-31 54 Bipolar connecting socket (B1), single, “i” version, w. cable 08-003-00-28 55 Bipolar connecting socket for instruments (B2), “i” version, w.
cable 08-003-00-29
56 NE connecting socket, “i” version, w. cable 08-024-00-62 57 Optical guide for NE status indicator, “i” version 08-021-00-01 58 Optical guide for activation indicators 08-021-00-02 59 Insulating cover for monopolar HF socket module 08-016-01-02
60 Socket panel II for Erbe accessories, “e” version, complete 08-003-01-30 61 Socket module carrier, “e” version, with membrane foil and
channel selector buttons 08-003-01-03
62 Light distributor for “e” version socket module 08-021-00-05 63 Monopolar socket module II, “e” version (M1 and M2) 08-022-01-32 64 Bipolar combination connecting socket (B1), “e”/“i” versions, w.
cable 08-003-00-31
65 Bipolar combination socket (B2), “e”/“i” versions, w. cable 08-003-00-40 66 NE connecting socket, “e”/“m” versions, w. cable and insulating 08-024-00-34
No. Designation Item/Order No.
70 Main board II, C40-2165, w/o guide rails 08-024-00-65
71 Controller board II, C40-2217 08-005-00-12
72 Power manager board II, C40-2169, w/o insulating cover 08-027-01-17
73 Monopolar board II, C40-2224 08-022-01-33
74 Bipolar board II, C40-2155 08-003-00-34
75 TFT screen module I, w. fluorescent tubes 08-038-00-14
76 Front board II, C40-2175-10 08-010-01-13
77 Activation indicator board II, C40-2173-11 08-027-00-18 78 NE status indicator board II, C40-2173-12 08-027-00-19 79 Rotary switch backlight board II, C40-2175-11 08-006-00-27 80 Socket illumination board II, C40-2173-14 08-003-00-51
90 ME402 small parts kit 08-018-00-57
T6.3A fuse cartridge for mains connector socket 08-034-00-12
M1.6A fuse cartridge 08-034-00-28
M0.125A fuse cartridge 08-034-00-62
Protective cover for bipolar socket B2, w. opening for 8mm plug 08-027-00-20 Current data carrier for software update*
Bottom Part of Unit, HW02
00-51 00-22
00-23
00-65 00-12 00-52
08-022-01-33 08-003-00-3408-018-00-5708-024-00-66 08-041-00-51 08-016-01-01 08-020-00-21 08-032-00-20 08-024-00-60 08-027-00-01
08-027-01-17 08-001-00-63 08-001-00-62 08-010-01-04
08-010-01-03
530-030 10 530-030 10
530-030 10
530-030 10
913046 lt1057 0406
lt 2355 0523
lt 2355 0523
913046 lt1057 0406 913046
lt1057 0406
STS 0550 'LF' 1278636
F1 M0,125A F2 M1,6A B82625-B2602-M1 6A / 250V~ 2 x 0,18mH
V 15487
prim 240V/50-60Hz 1-5sek 15V/0,5 VA 7-9 4µ7 M SD 4µ7 M SD
STS 06061348633 'LF' STS 06051295638 'LF'
STS 0547 'LF' 1278637
C40-2165 0606/029 VXX
X15 X16 VDC
GND GT2X18
X17 GT1
73 16 14 74 15 17 10 18 3 72 9 8 7 5 6
2 70 13 11 71 12
Main board II: 08-024-00-65, spare part no. 70 (C40-2165)
F1 M0,125A F2 M1,6A
X13
X12
X9
X14 X2 X21 X1
prim 240V/50-60Hz 1-5sek 15V/0,5 VA 7-9 4µ7 M SD 4µ7 M SD
STS 06051295638 'LF' STS 0547 'LF' 1278637
X15 X16 GND X18 X17
F1 F2
to X1 at connector module
X5 controller board connector X7connector monopolar board X6 bipolar board connector
to X4 and X5 at monopolar board to X3 at
Monopolar Board
X4 Power Manager board connector 2 X3 Power Manager board connector 1
to X2 at bipolar board
Spare boards are delivered without guide rails. Therefore, these must be transferred from the old board to the new one.
The main board does not provide any adjusting positions. If it is replaced, the setting of the power manager board must be checked for correct-ness, however.
Controller board II: 08-005-00-12, spare part no. 71 (C40-2217)
X2
X4 X10
X5 X8
X6
STS
0
848M08 BR PÜ
16 387 939
X11 X9
X500
X1 Board connector
The first version (V00) does not in-clude a backup battery yet – and therefore no clock. This also means that service functions involving time management are not available. More-over, version V00 cannot be upgraded in this respect.
Power manager board II: 08-027-01-17, spare part no. 72 (C40-2169)
101 C40-2169 0406/034 VXX
ME402 Powermanager 2
This spare part is shipped without pro-tective cover and corresponding mounts. These as well as the plastic optical fiber must be transferred from the old part to the new part.
Note that the power manager board may never be plugged into the main board if the unit has already been connected to the mains!
The spare part is delivered pre-adjusted. Its setting must be checked during installation and corrected if necessary. For procedure, see section 10.2.1.
Monopolar board II: 08-022-01-33, spare part no. 73 (C40-2224)
STS
The only difference between versions I and II consists in the design of con-nector X4.
This spare part comes fully adjusted, so no readjustment is required during installation. However, it is necessary to check the optical receiver located on the power manager board for cor-rect setting. For procedure, see section 10.2.1.
Bipolar board II: 08-003-00-34, spare part no. 74 (C40-2155)
C40-2155 3805/045 VXX
X4 X3 X2
GND GND
X1 Board connector
This spare part comes fully adjusted, so no readjustment is necessary dur-ing installation.
However, as the optical link must be interrupted during board replacement, correct setting must subsequently be checked on the power manager board in acc. with section 10.2.1.
Front Part, HW02
08-010-01-05 08-001-00-6908-001-00-68
08-008-00-44 08-001-00-64 08-001-00-65 08-021-00-03 08-027-00-18 08-021-00-02 08-003-01-01
08-022-01-30
High Voltage CAUTION RISK OF ELECTRIC SHO
CKLECTRI HE ECT T CONNEDIS
C POWER BEFORE SERVICEHigh Voltage CAUTION RISK OF ELECTRIC SHO
CKCTRILEHE ECT T CONNEDIS
C POWER BEFORE SERVICE
C40-2175
Rear side of front part with m-version socket module
C40-2225
08-021-00-01 08-027-00-19 08-021-00-01 08-003-00-29
08-003-00-28
08-024-00-62
08-021-00-01 08-027-00-19 08-021-00-04 08-027-00-18 08-021-00-02
08-003-01-02 08-022-01-31
08-021-00-02 08-027-00-19 08-021-00-05 08-027-00-18 08-021-00-02
08-003-01-03 08-022-01-32
Socket module
“m” version 08-003-01-26 Socket module
“i” version 08-003-01-27 Socket module
“e” version 08-003-01-30
Front board II: 08-010-01-13, spare part no. 76 (C40-2175-10)
ME402 Frontplatine 2 Xnummeriert.dsf 16.01.08
The spacer on the rear side of the board is not included in the spare part delivery and therefore must be trans-ferred from the old board to the new board.
Screen module I: 08-038-00-14, spare part no. 75
Hig
C POWER BEFORE SERVICE
M ODE
AINS RESCE LUO DE F THO COLD CA
NT LOC LOW EASE FOL PS. PL LAM
This spare part is delivered complete with two fluorescent tubes.
Note that the screen module is not an
“exchange part” (i.e. not available as a repaired part).
Activation indicator board II 08-027-00-18, spare part no. 77 (C40-2173-11)
X4
X10
C40-2173-11
To facilitate installation, the plug and the socket of the connection between the activation indica-tor board and the socket illumination board have been interchanged. In all other respects, board II is identical with its predecessor.
Socket illumination board II: 08-003-00-51, spare part no. 80 (C40-2173-14)
X3 C40-2173-14
To facilitate installation, the plug and the socket of the connection between the activation indica-tor board and the socket illumination board have been interchanged. In all other respects, board II is identical with its predecessor.
NE status indicator board II: 08-027-00-19, spare part no. 78 (C40-2173-12)
C40-2173-12
X1
Version II is connection- and function-compatible with version I.
Rotary switch backlight board II: 08-006-00-27, spare part no. 79 (C40-2175-11)
C40-2175-11
X11
Version II is connection- and function-compatible with version I.
Rotary switch assembly I: 80-006-00-21, spare part no. 30
30
08-006-00-23
08-024-00-61
08-003-00-33
08-006-00-22 08-024-00-63 08-018-00-57
08-025-00-05 08-006-00-21
31 36 35 90 34 33 32
The rotary switch unit is availa-ble either complete or in individ-ual components. The rotary but-ton is simply plugged in place and can be removed with little effort.
When replacing the rotary switch, it is also necessary to remove the backlight board, which must subsequently be reinserted correctly into the opt-ical guide.