• No results found

6 Using the IndraMotion Service Tool

6.3 Standard IMST Toolbar Icons

6.4.3 Oscilloscope Functionality

Overview

The Oscilloscope in the IMST is a Java applet designed to detect user-defined control or drive signals and to display them in a scaled grid. These signals are configured using the options available to the right and below the grid.

The oscilloscope is available only if the relevant Internet Explorer settings are made (refer to chapter 5.2 "Browser Settings" on page 39 for details). The current user has to have viewing and modifica‐

tion rights as well to use the oscilloscope.

The oscilloscope applet window contains the following areas:

● Main graphical display area with x-y grid

● Signal details on detected recordings just below the x-y grid

● Tabs to configure the measurement and for signal triggers

Using the IndraMotion Service Tool

● Tabs for scaling and color selection

● Active measurement to configure signals

Fig.6-15: Initial applet for the oscilloscope function

Increase the viewing area of the oscilloscope by hiding the naviga‐

tion tree structure. Click the "Tree icon" to hide the naviga‐

tion tree structure. Click the "Tree icon" again to display the tree structure.

Graphical Display Area

The signal display area contains a x-y grid with the following properties:

● The x-scale (horizontal) and the y-scale (vertical) are calculated dynami‐

cally based on the minimum and maximum values of the signals and the time period intended for measurement

● The y-measurement is displayed from minimum to maximum signal value.

Resizing the Explorer window dynamically scales the detected recordings and increments the x-y units

● Signal scaling can be increased or decreased using the scaling options

● Signal units are displayed in the upper left and lower right of the display area

Using the IndraMotion Service Tool

Fig.6-16: Graphical display area for the oscilloscope function

Active Measurement

The "Active Measurement" tab is used to configure the source device and the signals to be detected. The tab can contain up to four buttons depending on the trigger type selected in "Configure Measurement/Trigger".

A. Used to select the source and signals to be detected

B. Displays the current recordings stored in the devices parameters C. Only visible if the trigger type is set to "Trigger Manually"

D. Starts the oscilloscope and detects recordings based on signal settings and trigger type

Fig.6-17: Active Measurement

Proceed as follows to select a source device and signals to be detected:

1. From the "Active Measurement" tab, select the Signal button to open the Configure Signal window shown in fig. 6-18 "Signal configuration" on page 58.

2. Select a device (control or drive) from the drop-down list (1.).

Using the IndraMotion Service Tool

For controls, also select the parameter type C (control) or A (axis) from a second drop-down list (2.).

3. Signals are selected by highlighting a parameter in the left window and clicking on the right-directional arrows (3.). To remove signals from the list, highlight the respective parameter in the right window and click on the left-directional arrows.

Signals can also be selected or removed in both windows by double-click‐

ing on the parameter.

A. The drives support four channels and the IndraMotion MLC/

IndraLogic XLC control supports 16 channels B. Indicates the number of channels that can be added Fig.6-18: Signal configuration

4. Click OK (4.) to complete the selection of signals.

5. Then, select the trigger type and configure the measurement settings.

Refer to chapter "Configuring Measurement and Signal Trigger" on page 58 for details.

6. Once all settings are configured, click on Start to begin the detection.

For manual trigger types, select the Trigger Manually button after the Start button.

Configuring Measurement and Signal Trigger

Configuring measurement

The "Confg Measurement" tab is used to configure the number of measure‐

ments to be taken and the recording period in ms. The actual recording period is write-protected and is calculated by multiplying the "Number of Samples" with the "Time Period".

Using the IndraMotion Service Tool

Fig.6-19: Configuring measurement Signal Trigger/Manual Trigger

The "Signal Trigger" tab is used to configure the settings for a "signal trigger"

or a "manual trigger". The Configure button is used to select the trigger type, pretrigger and source parameter to be analyzed.

A. Tab changes to "Signal Trigger" when the trigger type is set by pressing the Configure.

B. Tab changes to "Manual Trigger" when the trigger type is set by pressing the Configure.

Fig.6-20: Signal and manual trigger configuration

Click on the Configure button proceed as follows to configure a trigger:

1. Select either the Manual Trigger or the Signal Trigger button.

A signal trigger requires additional configuration starting at step 3.

2. Enter a value in percent for the Pre Trigger and Post Trigger fields.

3. Click on the Select button and select the parameter to be used as trigger.

4. Select the Trigger Edge to be used for the selected parameter.

5. Enter a value in the Trigger Value field that is used to start the Oscillo‐

scope.

The fields in the "Trigger Value" section change based on the selected parameter in step 3.

Using the IndraMotion Service Tool

Fig.6-21: Configuring trigger

6. Click OK to complete the trigger configuration.

Numerical parameters The selection of a numerical parameter (float or integer) displays the current value, units and trigger value. The trigger value is defined in the "Trigger Val‐

ue" field.

Binary or hexadecimal parameters Selecting a binary or hexadecimal parameter allows the setting of a binary trig‐

ger by defining a trigger mask. Selecting the Bit Trigger checkbox allows to define the transition of a single bit. The value is determined by selecting the relevant checkboxes or radio buttons.

Fig.6-22: Configuring a binary parameter trigger

Detecting Recording with the Oscilloscope

Once all relevant signals and triggers are configured, the signals are ready for detection. Proceed as follows to detect the recordings from the selection sig‐

nals.

Using the IndraMotion Service Tool

1. Click on Start in the "Active Measurement" tab.

If the signal trigger is set to "Manual Trigger", click on the Trigger Man‐

ually button to detect and display the recordings.

A signal trigger detects the recordings when the conditions configured in the "Signal Trigger" tab are met.

2. Once all recordings are detected and displayed, the recordings can now be edited using the tools available on the "Scaling/Color" tabs and the measured signals specified just below the recordings.

Editing Oscilloscope Signals

Overview

Once all oscilloscope recordings are detected and displayed, the recordings can be edited as follows:

● Display and hide recordings

● Scale and shift the recordings to the x-axis

● Zoom in the x-axis using an incremental zoom function or selecting a spe‐

cific area using the cursor function

● Display the recordings using the following options:

– Interpolated measuring points – Real measuring curve

– Single measuring points

● Change color of each recording

Measured Signals (Displaying Recordings)

Detected signal recordings are shown or hidden by enabling or disabling the checkbox on the left next to each signal. The currently selected signal is high‐

lighted in the table and displayed with a recording printed in bold. The recording of the remaining visible signals is more detailed. The measuring units on the y-axis is adjusted to the selected recording. The x-y-axis always measures time in ms.

Fig.6-23: Displaying measured signals

Number of supported channels The number of signals that can be measured at a time depends on the selected device. IndraMotion MLC and IndraLogic XLC always support 16 channels.

IndraDrive drives support four channels.

Recording Color Each recording along with its measuring unit is assigned to a matching color that can be modified by selecting the "Color" tab on the right next to the graph‐

ical display area. Ensure that each recording has a unique color assigned.

Additional columns when using os‐

cilloscope cursor function The measured signal table displays additional columns when using the cursor function of the oscilloscope available in the "Scaling" tab. If this cursor function is used, the "Y Cursor Diff" and "X Cursor Diff" columns appear in the table.

Using the IndraMotion Service Tool

Fig.6-24: Additional columns when using the cursor function Y-Scaling

The oscilloscope recordings are scaled sing the options available on the "Scal‐

ing" tab. Only the currently selected recording in the "Signal" table is scaled.

Fig.6-25: Scaling on the y-axis

Clicking the minus button ( - ) decreases the increment of the y-axis causing the current signal to extend in y-direction.

Clicking the plus button ( + ) decreases the increment of the y-axis causing the current signal to reduce in y-direction.

Clicking the up arrows (↑) shifts the active signal up by 10 %.

Clicking the down arrow (↓) shifts the active signal down by 10 %.

The last two buttons with a curve image are used to change the amplitude of all signals as follows:

● The first amplitude button from the left changes the amplitude of all signals to use the full range of the display (i.e. the signals are stretched along the y-axis)

● The second amplitude button changes the amplitude of all the signals based on the largest signal (i.e. the signals are compressed along the y-axis)

Zoom

The buttons in the "Zoom" section of the "Scaling" tab are used to zoom in and out and affect all the recordings in the display at once.

Using the IndraMotion Service Tool

Fig.6-26: Zoom function for recordings in the x-Axis

Zoom out Clicking the minus button (-) decreases the increment of the x-axis causing all the recordings to extend in x-direction.

Switching off zoom The one-to-one button (1:1) becomes active when a zoom button (+/-) is clicked.

This returns the recordings to their original size.

Zoom in Clicking the plus button ( + ) increases the increment of the x-axis causing all recordings to be decreased in x-direction.

Showing cursor

The Show cursor button is used to select a specific curve segment by creating a frame around the segment with two sets of perpendicular lines. The cross section of all four lines is the important point. Each line can be positioned by clicking and dragging each line to the desired position until a box is created around the relevant area. The mouse pointer changes to a two-headed arrow line (↕) when positioned on a line.

Using the IndraMotion Service Tool

Fig.6-27: Defining a frame with "Show cursor"

Zooming between cursors

Clicking on the Zoom between X cursors button enlarges the display to show the center area between all four cursor lines.

Fig.6-28: Display after zooming in with "Show Cursor"

Curve Display

There are three options to display recordings using the curve display buttons:

● Interpolated measuring points

● Real measuring curves

Using the IndraMotion Service Tool

● Single measuring points

Interpolated measuring points Clicking on the I button displays the current recordings using interpolated measuring points.

Fig.6-29: Displaying interpolated measuring points

Real measuring curves Clicking on the R button displays the current recordings real measuring curves.

Fig.6-30: Displaying real measuring curves

Single measuring points Clicking on the S button displays the current recordings using single measuring points.

Fig.6-31: displaying single measuring points

Using the IndraMotion Service Tool

Color

The "Color" tab is used to select or change the color for the display background, grid and the individual recordings. The number of available signal colors de‐

pends on the selected source.

● IndraDrives support four colors

● IndraMotion MLC and IndraLogic XLC always support 16 colors

The number displayed in each color rectangle corresponds to a recording start‐

ing at 1 for the first recording in the list as shown in the following figure:

Fig.6-32: Color selection

Changing a signal color The color of the background, grid or recording is changed by selecting the rel‐

evant color button and selecting a new color from the "Select Color" window.

Fig.6-33: Changing a color Using the IndraMotion Service Tool

6.5 "Hardware" folder

6.5.1 Overview

The Hardware folder displays a read-only overview of all supported compo‐

nents currently connected to the control and configured in the IndraWorks project.

Fig.6-34: Main hardware component page

The main hardware component page displays one control and the total number of drives connected and supported.

Related documents