• No results found

Special Conditions of Use

PH10 AND PH12 SMART SENSORS CSA Entity Approval

Schneider Electric Systems USA, Inc.

38 Neponset Avenue

Foxboro, Massachusetts 02035-2037 United States of America http://www.schneider-electric.com

SIZE

A

CAGE CODE

1W0E8

DWG NO.

4S201FB REV A

SCALE

NONE WEIGHT SHEET 2

PH10 AND PH12 SMART SENSORS CSA Entity Approval

INTRINSICALLY SAFE / SECURITE INTRINSEQUE NON - INCENDIVE / NON - INCENDIE

CSA Certified Associated Apparatus (Barrier)

Any CSA Approved Transmitter with Intrinsically

Safe Sensor Output Parameters or Non Incendive

Sensor Output Parameters.

PH10 Sensor

CSA Certified Associated Apparatus (Barrier)

Any CSA Approved Transmitter with Intrinsically Safe Sensor

Output Parameters or Non Incendive Sensor Output

Parameters.

PH12 Sensor

Schneider Electric Systems USA, Inc.

38 Neponset Avenue

Foxboro, Massachusetts 02035-2037 United States of America http://www.schneider-electric.com

INTRINSICALLY SAFE MARKINGS IS CL I, DIV 1, GP ABCD INSTALL PER DRAWING 4S201FB

Sensor Entity Parameters Ui/Vmax = 11V Ii/Imax = 50 mA Pi/Pmax = 400 mW Li = Negligible Ci = 1.2 nF

NON INCENDIVE , FIELD MARKINGS CL I,DIV 2, GP ABCD

T4...T3 Ta = -20°C to 85°C T5 Ta = -20°C to 75°C T6 Ta = -20°C to 60°C INSTALL PERDRAWING 4S201FB

Sensor Entity Parameters Ui/Vmax = 11V

MODEL: Model Code per Sales Order

STYLE: Style Designation “A”

ORIGIN: Manufacturing Origin Code (Ex. 2A1720) 2A = First two digits: Area of manufacture 17 = Next two digits year of manufacture

20 = last two digits week of manufacture. A two-character numeric code between 01 and 52

SN: = Sensor Serial Number Zone B: Is the location of certification markings.

Zone C: Is the location of certification company logo or CE logo (if required).

Zone D: Is the plant of manufacture location.

Schneider Electric Systems USA, Inc.

38 Neponset Avenue

Foxboro, Massachusetts 02035-2037 United States of America http://www.schneider-electric.com

SIZE

A

CAGE CODE

1W0E8

DWG NO.

4S201FB REV A

SCALE

NONE WEIGHT SHEET 4

Intrinsically Safe Notes

1. Figure noted above must not be modified without prior CSA approval.

2. Barrier must be CSA certified and must be installed in accordance with manufacturer's instructions.

3. Maximum nonhazardous area voltage must not exceed 250 V.

4. Install in accordance with Canadian Electrical Code (Part 1), ANSI/ISA-PR12.06.01, or NEC NFPA 70.

5. Total resistance between Intrinsic Safety Ground and Earth Ground must be less than 1 Ohm.

6. Entity parameters must meet the following requirements:

Uo or Voc ≤ Ui or Vmax, Io or Isc ≤ Ii or Imax, Po ≤ Pi or Pmax, Co ≥ Ci + Ccable Lo ≥ Li + Lcable

Nonincendive Field Wiring Notes:

1. Install in accordance with Canadian Electrical Code (Part 1), ANSI/ISA-PR12.06.01, or NEC NFPA 70.

2. Nonincendive field wiring installation:

a. The Nonincendive Field Wiring Circuit Concept allows interconnection of Nonincendive Field Wiring Apparatus with Associated Nonincendive Field Wiring Apparatus or Associated Intrinsically Safe Apparatus.

Schneider Electric Systems USA, Inc.

38 Neponset Avenue

Foxboro, Massachusetts 02035-2037 United States of America http://www.schneider-electric.com COMPONENT REPLACEMENT OR SUBSTITUTION HAZARD

Do not substitute components. Substitution of components may impair intrinsic safety rating.

Failure to follow these instructions can result in death, serious injury, or equipment damage

AVERTISSEMENT REMPLACEMENT DES COMPOSANTS OU RISQUE DE SUBSTITUTION

Ne substituez pas les composants. La substitution de composants peut compromettre la classification Sécurité I ntrinsèque de l’équipement.

Le non-respect de ces instructions peut entraîner la mort, des blessures graves ou l’endommagement des équipements.

WARNING COMPONENT REPLACEMENT OR SUBSTITUTION HAZARD

Do not substitute components. Substitution of components may impair suitability for Class I, Division 2 Failure to follow these instructions can result in death, serious injury, or equipment damage.

AVERTISSEMENT REMPLACEMENT DES COMPOSANTS OU RISQUE DE SUBSTITUTION

Ne substituez pas les composants. La substitution de composants peut rendre cet équipement impropre à l’utilisation en zone Classe I Division 2.

Le non-respect de ces instructions peut entraîner la mort, des blessures graves ou l’endommagement des équipements.

WARNING HAZARD OF ELECTRIC SHOCK, EXPLOSION, OR ARC FLASH

To help prevent ignition of flammable or combustible atmospheres, disconnect power before servicing.

Failure to follow these instructions will result in death or serious injury

AVERTISSEMENT RISQUE D’ÉLECTROCUTION, D’EXPLOSION OU D’ARC ELECTRIQUE

Afin d’éviter l'inflammation d'atmosphères combustibles ou inflammables, débrancher l'alimentation avant tout entretien.

Le non-respect de ces instructions entraînera la mort ou des blessures graves

Schneider Electric Systems USA, Inc.

38 Neponset Avenue

Foxboro, Massachusetts 02035-2037 United States of America http://www.schneider-electric.com POTENTIAL GENERATION OF ELECTROSTATIC CHARGE

The PH10-*S or PH12-*S sensor is non-conducting and may generate an ignition-capable level of electrostatic charge under certain extreme conditions. Ensure that the equipment is not installed in a location where it may be subjected to external conditions which might cause a build-up of

electrostatic charge on nonconducting surfaces. Additionally, cleaning of the PH10-*S or PH12-*S sensor should only be done with a damp cloth.

Failure to follow these instructions can result in death, serious injury, or equipment damage.

WARNING HAZARD OF ESCAPING CHEMICALS

When installing or removing sensors, wear appropriate protective clothing including protective goggles.

Escaping chemicals can cause severe injury including blindness.

Failure to follow these instructions can result in death, serious injury, or equipment damage.

WARNING HAZARD OF PERSONAL INJURY OR EQUIPMENT DAMAGE

Use care when connecting and disconnecting high-pressure service connections. Use protective gloves and follow the recommended procedures to avoid injury to personnel or damage to equipment. When processing hazardous liquids, follow the recommended procedures.

Failure to follow these instructions can result in death, serious injury, or equipment damage.

WARNING HAZARD OF PERSONAL INJURY OR EQUIPMENT DAMAGE

In addition to the pressure and temperature limits of the sensor, the sensor mounting accessories also have pressure and temperature limits. The specifications for the mounting accessories may be greater or less than the sensor specifications. Always use the lesser of the specification limits when designing the installation of a sensor with accessories.

Failure to follow these instructions can result in death, serious injury, or equipment damage.

CAUTION HAZARD OF EQUIPMENT DAMAGE

Use care when handling sensitive sensor components such as glass electrodes.

Failure to follow these instructions can result in equipment damage.

ATTENTION RISQUE D’ENDOMMAGEMENT DE L’EQUIPEMENT

Manipuler avec précaution les composants fragiles du capteur tels que les électrodes en verre.

Le non-respect de ces instructions peut entrainer un endommagement de l’équipement.

Schneider Electric Systems USA, Inc.

38 Neponset Avenue

Foxboro, Massachusetts 02035-2037 United States of America http://www.schneider-electric.com

SIZE

A

CAGE CODE

1W0E8

DWG NO.

4S201FB REV A

SCALE

NONE WEIGHT SHEET 7

Specific conditions of use:

1. The PH10 or PH12 Smart sensor is only certified for use in ambient temperatures marked on the sensor housing and should not be used outside this range.

2. The maximum process temperature must not exceed the ambient temperatures marked on the sensor.

3. All field wiring must be rated for 85°C or greater.

4. The PH10 or PH12 Smart sensor contains no user-replaceable parts and is not intended to be repaired by the user. Repair of the PH10 or PH12 Smart sensor is to be carried out by the manufacturer, or their approved agents, in accordance with the applicable code of practice.

5. The surface resistivity of the PH10 or PH12 Smart sensor's enclosure may be greater than one gigaohms. Care should be taken to avoid electrostatic charge build-up. The PH10 or PH12 Smart sensor's enclosure must not be rubbed or cleaned with solvents or a dry cloth.

6. FM Installation Instructions for PH10 and PH12 Smart Sensors

REVISIONS HISTORY ELEC-P

LTR DESCRIPTION CODE REVISED BY

APPROVED BY DATE

SIMILAR TO - Schneider Electric Systems USA, Inc.

38 Neponset Avenue Foxboro, Massachusetts 02035-2037 United States of America http://www.schneider-electric.com

This drawing and specifications herein are the property of Schneider Electric Systems USA, Inc. and must not be loaned or otherwise transferred to any third party or reproduced, copied, or used in connection with the

manufacture, purchase, or sale of items without written consent of Schneider Electric Systems USA, Inc.

FIRST USED IN MI 611-206 DESIGNED FOR PH10-*S

UNLESS OTHERWISE

PRODUCT LINE ANALYTICAL

E. O. NUMBER USDM0091

ENG. SERVICES MGR. DATE

PRODUCT ENGINEER

24-MAR-2017 COMPUTER TITLE:

SIZE

A ISSUED PER ERN2015000163 31-MAR-2015

B UPDATED TO INCLUDE PH12, UPDATED AGENCY MARKINGS AND ENTITY PARAMETERS PER ERN 2017000175.

CC:7 CET

ALL CHANGES REQUIRE PRODUCT SAFETY ENGINEER

APPROVAL

Schneider Electric Systems USA, Inc.

38 Neponset Avenue

Foxboro, Massachusetts 02035-2037 United States of America http://www.schneider-electric.com

SIZE

A

CAGE CODE

1W0E8

DWG NO.

4S201EY REV B

SCALE

NONE WEIGHT SHEET 2

PH10 AND PH12 SMART SENSORS