RU-Config Update
3.4 Precautions to be Observed During the Use of Output Options
IIf an error message of the error code 50001 appears at CL-AP startup, check the settings for the following output options:
IR346 DICOM CR STORAGE
IR346 DICOM IM PROCESSED (processed image transmission)
IR346 DICOM PRINT
IR346 DMS NET-WORK IN-OUT
IR346 LOCAL PRINT
The checks to be performed are detailed below:
Option key Check
IR346 DICOM CR STORAGE
IR346 DICOM IM PROCESSED
IR346 DICOM PRINT
Verify that the option key is installed, and then check whether SCU setup is properly completed for this equipment. The procedure is described below:
1. With the Service Utility, open the “Setup Configuration Item” window.
2. Open “NETWORK CONFIGURATION” and then
“THIS HOST (IIP)”.
3. Check that an AE whose “Role” is SCU is registered. If it is not registered, register it now.
For detailed registration procedure
descriptions, see “OE: Connecting the CL to Other Equipment”.
IR346 DICOM CR STORAGE OE3.
IR346 DICOM IM PROCESSED OE4.
IR346 DICOM PRINT OE6.
IR346 DMS
NETWORK IN-OUT Verify that the option key is installed.
IR346 LOCAL PRINT
Verify that the option key is installed, and then check whether the PEI01A board is installed in the PC. Also, check that DIP switch setup is properly completed.
For DIP switch settings and other checkout procedures, see “CL + Local Printer (E-i/f Connection)” in
“OE: Connecting the CL to Other Equipment”.
3.4 Precautions to be Observed During the Use of Output Options
CR-IR346CL Service Manual
010-054-20E
11.30.2009 FM5622 (ITC)
MT-24
NOTE
Windows 2000 screens are used herein, which are the same as those displayed by Windows XP or Windows Vista.
4. Troubleshooting (for DELL PC)
The table below summarizes various troubles that may arise with the CL. In accordance with the encountered trouble, perform a procedure indicated in the “Check Method/Remedies” column.
Symptom Probable Causes Check Method/Remedies
The PC (Windows) does not
start up. It is conceivable that the PC’s memory may be
defective. Check the PC’s memory. If it is defective, replace it.
For the memory check procedure, see “4.1 Checking the Memory”.
The CL-AP (application)
does not start up. It is conceivable that the data stored on the hard disk
may be partly damaged. Check the hard disk. If any error is revealed by the check, perform a procedure for achieving error recovery or replace the hard disk.
For the hard disk check procedure, see “4.2 Checking the Hard Disk”.
Output system options are not installed or not properly set up.
[Displayed error code: 50001 20110]
Install or properly set up output system options.
For details, see “3.4 Precautions to be Observed During the Use of Output Options”.
The dummy node is not set up.
[Displayed error code: 30151 20110] Choose the Service Utility’s “Setup Configuration Item”, “NETWORK
CONFIG”, “HIS HOST (IIP)” to check for node setup. If no node is set up, set DICOM Print SCU as a dummy node.
The DICOM Print option key need not be installed.
For the setup procedure, see “2.1.4 Setup of Information on this Equipment for the DICOM Print Function” in
“OE5: CL+FN-PS551/DRYPIX/FM-DP L (DICOM Print)”.
The connection to the network is not established.
[Displayed error code: 30151 20110] Connect the CL to the network.
The image database is not initialized.
[Displayed error code: 50001 20001 30107 30151 20110]
With the Service Utility, initialize the image database.
For the initialization procedure, see “12. Initializing the Image Database” in “RI: Reinstalling the Software”.
(Continued to the next page)
The MAC address is not acquired.
[Displayed error code: 20111 20110
(Error 20111 is recorded only in the application’s event log.)]
Reinstall the network driver and then update it.
For the reinstallation procedure, see “4.7 Reinstalling the Network Driver”.
For the software version update procedure, see “4.8 Installing the 3Com Network Driver (Ver. A10)”.
If the error recurs after the network driver is updated, replace the PC’s motherboard. Before motherboard replacement, acquire the MAC address manually and start up the CL-AP.
For the MAC address acquisition procedure, see “3.14 MAC Address Manual Input” in “MU: Maintenance Utility”.
4. Troubleshooting (for DELL PC)
CR-IR346CL Service Manual
010-054-20E
MT-25
Symptom Probable Causes Check Method/Remedies
The CL-AP (application)
does not start up. The selector resetup procedure is not completed.
[Displayed error code: 31099] After the host name for the built-in type equipment is changed, use the selector setup of the Service Utility to set the built-in type equipment host name again.
For the setup procedure, see “12. Setting the Selectors – Selector Setting” in “MU: Maintenance Utility”.
The LCD flickers. The refresh frequency setting is improper. Correct the refresh frequency setting.
For the refresh frequency setup procedure, see “4.4 Resetting the Refresh Frequency”.
The LCD touch panel does
not sound. The screen contact position correction procedure was
performed twice in succession. With the utility tool for the touch panel, perform the touch panel sound setup procedure again.
For the setup procedure, see “4.5 Resetting the Touch Panel Sound”.
For the driver software change procedure, see “4.6 Changing the Barcode Reader/Magnetic Card Reader Driver Software”.
The BCR/MCR does not
normally operate. The BCR/MCR is connected to an improper USB port. Connect the BCR/MCR to a recommended USB port.
For the reconnection procedure, see “4.3 USB Ports Recommended for Barcode Reader/Magnetic Card Reader Mounting (for GX150 Only)”.
For details, see “3.2 Typical Procedure for Achieving Recovery from an LP (Local Printer) Printout Generation Failure”.
Patient data cannot be
For the driver software change procedure, see under the preceding symptom, which is entitled “The local printer does not generate a film output”.
The communication with the RU or other equipment
For the update procedure, see “4.8 Installing the 3Com Network Driver (Ver. A10)”.
4. Troubleshooting (for DELL PC)
CR-IR346CL Service Manual
010-054-21E