Purpose
This menu item enables you to configure the balance, i.e., adapt the balance to your individual requirements by selecting from a list of parameter options in a menu. You can block access to this menu by assigning a password.
Features
The simple weighing function is available at all times. You can select one from each of the following application groups. This means a
number of combinations are possible.
Application 1 (basic settings) – Toggle weight units – Counting
– Weighing in percent – Animal weighing
(averaging) – Calculation – Recalculation – Density determination – Differential weighing
– Air buoyancy correction and air density determination
– Diameter determination Application 2 (control functions) – Checkweighing
– Time-controlled functions Application 3 (data records) – Totalizing
– Formulation – Statistics
In addition, you can assign 2 extra functions to each of the soft keys, in some cases (depending on the Setup configuration):
– Second tare memory – Identification codes – Manual storage in app.
3 memory (M+ key) – Changing the resolution – Product data memory – SQmin function*
– DKD uncertainty of measurement*
Auto-start application when the balance is switched on
Factory settings:
only application parameters
Factory Settings for the Parameters The factory settings are identified by the symbol “ο” in the list starting on page 35.
Preparation
Display available application parameters:
§ Select the Setup menu:
press the o key
> SETUP is displayed
§ Select parameters: repeatedly press the qand Osoft keys
If you have already assigned a password:
> The password prompt is displayed:
$ If access is blocked by a password: enter the password using the numeric/alpha-betic keys
$ If the last character of the password is a letter: conclude input by pressing a
§ Confirm your password and have the application parameters displayed: press the lsoft key
> The application menu is displayed:
$ To select the next group:
press the qsoft key (down arrow)
$ To select the previous item of a group: press the Qsoft key (up arrow)
$ To select the next sub-item within a group: press the Osoft key (right arrow)
$ To select the previous group: press the osoft key (left arrow)
$ To confirm: press the lsoft key
Extra Functions
§ Exit the Setup menu:
press the oosoft key
> Restart your application
§ Print parameter settings:
– When the balance/scale functions are displayed, press r
> Printout (example)
Texts with more than 20 characters are truncated
SETUP
Application
---Application 1 (basi Off Application 2 (cont Off Application 3 (data Off Extra function (F4) Off Extra function (F5) Off Auto-start app. whe Off
---Application Parameters (Overview) οfactory settings
√user-defined setting(s)
Setup – Application 1 ο Off (weighing only) Application (Basic settings)
parameters Toggle wt. units Weight unit 1 ο Grams /g
Kilograms /kg*
Carats /ct Pounds /lb*
Ounces /oz*
Troy ounces /ozt*
Hong Kong taels /tlh*
Singapore taels /tls*
Taiwanese taels /tlt*
Grains /GN*
Pennyweights /dwt*
# Milligrams /mg Parts per pound //lb*
Chinese taels /tlc*
Momme /mom*
Austrian carats /K*
Tola /tol*
Baht /bat*
Mesghal /MS*
Display accuracy 1 ο All digits Fewer for weight changes Last digit off*
Weight unit 2 # Grams /g
Kilograms /kg*
Carat s/ct Pounds /lb*
Ounces /oz*
Troy ounces /ozt*
Hong Kong taels /tlh*
Singapore taels /tls*
Taiwanese taels /tlt*
Grains /GN*
Pennyweights /dwt*
ο Milligrams /mg Parts per pound //lb*
Chinese taels /tlc*
Momme /mom*
Austrian carats /K*
Tola /tol*
Baht /bat*
Mesghal /MS*
Display accuracy 2 ο All digits Fewer digits for weight change Last digit off*
For counting through air buoyancy correction, see the following pages Applications 2 and 3 see following pages
Factory setting Factory setting
Setup – Application 1 Counting Accuracy – Average ο Display accuracy Application (Basic settings) piece weight calculation + 1 decimal place
parameters (resolution) + 2 decimal places
Average piece Off
weight updating Manual ο Automatic Percent weighing Weight storage accuracy ο Display accuracy
(resolution) + 1 decimal places + 2 decimal places Decimal places for None
readout in percent 1 decimal place ο 2 decimal places
3 decimal places 4 decimal places 5 decimal places 6 decimal places 7 decimal places Display calculated value ο Residue
Loss Ratio 1 (DR) Ratio 2 (OR) Animal weighing* Animal activity Calm
(averaging) Normal
Active
0.1% of the animal/object 0.2% of the animal/object 0.5% of the animal/object 1% of the animal/object 2% of the animal/object ο 5% of the animal/object 10% of the animal/object 20% of the animal/object 50% of the animal/object 100% of the animal/object
Start Manual mode
ο Automatic mode Minimum load for automatic None
storage 10 digits
20 digits 50 digits ο 100 digits
200 digits 500 digits 1000 digits Decimal places in None
result display 1 decimal place ο 2 decimal place 3 decimal places 4 decimal places 5 decimal places 6 decimal places 7 decimal places
Printout None
ο Average weight only Average and calculated value Recalculation*
Factory setting
Setup – Application 1 Calculation Decimal places in None
Application (Basic settings) calculated result 1 decimal place
parameters ο 2 decimal places
3 decimal places 4 decimal places 5 decimal places 6 decimal places 7 decimal places
Density Method Density of liquid
ο Buoyancy Displacement Pycnometer
Liquid causing ο Water
buoyancy Ethanol
User-definable
Decimal places None
for disp. of vol./ 1 decimal place
density ο 2 decimal places
3 decimal places 4 decimal places
Printout ο None
All data
Differential weighing Weighing sequence* Individual weighing
Consecutive individual weighing ο Combined weighing
Serial weighing
Tare weighing No
ο Yes
Result with None
decimal point 1 decimal place
ο 2 decimal places 3 decimal places 4 decimal places 5 decimal places 6 decimal places 7 decimal places
Autosave ο Off
values On, first value at stability
On, last value at stability On, value bet.
70 – 130% at stabil.
Minimum load None
for autosave 10 digits
ο 20 digits 50 digits 100 digits 200 digits 500 digits 1000 digits
Save statistics ο No
Yes Air buoyancy correction: See next page
Applications 2 and 3: see following pages
Factory setting
Setup – Application 1 Differential weighing Generate printout None
Application (Basic settings) ο Automatic after backweighing
parameters Auto. after initial weigh.
and backweighing
Automatic after tare-, initial-and backweigh.
Include sample ID in text line ο No Yes
Wg.seqkey No
ο Yes Clear sample after individual ο No weight, result + and unload Yes
Last residual weight is stored ο No as initial weight (ashing) Yes Air buoyancy correction Air density determination ο No Yes Change steel/aluminum ο No
references Yes
Diameter determination Application 2 ο Off
(Control functions) Checkweighing* Activation of port lines ο Within checkweighing range Always on
Stability and checkweighing range At stability
Stability + checkweigh.
range on
Type of checkweighing input ο Target, min., max. weight Minimum, maximum weight Target, min. in %, max. in % Weight display mode ο Absolute value
Difference from the target Automatic printout Yes
of OK values ο No
Time-controlled Function after time interval Beep (acoustic signal)
functions Lock in readout
ο Automatic printout of values Store value in applicat. 3 memory (totalizing, formulation, statistics) Automatic function restart ο On
Off
Storage mode ο Without stability After stability After higher stability Print then tare Off
ο On Factory setting Factory setting
Setup – Application 3 ο Off Application (Data records)
parameters Totalizing* Automatic storage ο Off
On, first value at stability On, last value at stability On, value bet. 70% – 130%
at stability Minimum load for None automatic storage 10 digits
ο 20 digits 50 digits 100 digits 200 digits 500 digits 1000 digits Source of data for auto storage ο Application 1
Application 2 Evaluated values ο Net
Calculated Net + calculated
Evaluation mode, MRfunction ο Intermediate evaluation, print Final evaluation, print Intermediate evaluation, display+print
Final evaluation, display+print M+/M– function, then tare ο Off
On Printout of individual No
components ο Yes
Formulation* Automatic storage ο Off
On, first value at stability Minimum load for None
automatic storage 10 digits ο 20 digits 50 digits 100 digits 200 digits 500 digits 1000 digits Source of data for ο Application 1 automatic storage Application 2 Evaluated values ο Net
Calculated Net + calculated
Evaluation mode, MRfunction ο Intermediate evaluation, print Final evaluation, print Printout of individual No
components ο Yes
Factory setting Factory setting
Setup – Application 3 Statistics Automatic storage o Off
Application (Data records) On, first value at stability
parameters On, last value at stability
On, value 70% – 130%
at stability
Minimum load for None
automatic storage 10 digits o 20 digits 50 digits 100 digits 200 digits 500 digits 1000 digits Source of data for o Application 1 automatic storage Application 2
Evaluated values o Net
Calculated Net + calculated
Evaluation mode, o Intermediate evaluation, print MRfunction Final evaluation, print
Intermediate evaluation, display + print Final evaluation, display + print M+/M– function, then tare o Off
On Printout of individual No
components o Yes
Extra Functions – for other extra functions, see next page
Factory setting Factory setting
Setup – Extra Functions ο Off Application F4 key
parameters Extra Functions 2nd tare memory Container tare weight ο No
F5 key Yes
Automatic printout Net value Tare/preset tare ο Off
Identification Printout Automatic, if configured
codes Once after pressing r
(IDs) if configured
ο Each time the print key (r) is pressed Once for M+ function (application 3 memory) Manual storage
in app. 3
memory M+ (totalizing, formulation, statistics) Changing the resolution Product data memory
SQmin Function Display o Text display
Bargraph
Print in o Off
GLP Header On
DKD uncertainty Display1) U (absolute)
of measurement U* (relative)
PG (process accuracy)
Print1) U (absolute)
U* (relative)
PA (process accuracy) Process accuracy Input: 1–100,
5 decimal places
Auto-start On
application when ο Off power goes on
Factory settings No
only for Yes
application-parameters
Factory setting