Setting the Application Parameters ( Application )

In document Sartorius ME and SE Series (Page 34-42)

Purpose

This menu item enables you to configure the balance, i.e., adapt the balance to your individual requirements by selecting from a list of parameter options in a menu. You can block access to this menu by assigning a password.

Features

The simple weighing function is available at all times. You can select one from each of the following application groups. This means a

number of combinations are possible.

Application 1 (basic settings) – Toggle weight units – Counting

– Weighing in percent – Animal weighing

(averaging) – Calculation – Recalculation – Density determination – Differential weighing

– Air buoyancy correction and air density determination

– Diameter determination Application 2 (control functions) – Checkweighing

– Time-controlled functions Application 3 (data records) – Totalizing

– Formulation – Statistics

In addition, you can assign 2 extra functions to each of the soft keys, in some cases (depending on the Setup configuration):

– Second tare memory – Identification codes – Manual storage in app.

3 memory (M+ key) – Changing the resolution – Product data memory – SQmin function*

– DKD uncertainty of measurement*

Auto-start application when the balance is switched on

Factory settings:

only application parameters

Factory Settings for the Parameters The factory settings are identified by the symbol “ο” in the list starting on page 35.

Preparation

Display available application parameters:

§ Select the Setup menu:

press the o key

> SETUP is displayed

§ Select parameters: repeatedly press the qand Osoft keys

If you have already assigned a password:

> The password prompt is displayed:

$ If access is blocked by a password: enter the password using the numeric/alpha-betic keys

$ If the last character of the password is a letter: conclude input by pressing a

§ Confirm your password and have the application parameters displayed: press the lsoft key

> The application menu is displayed:

$ To select the next group:

press the qsoft key (down arrow)

$ To select the previous item of a group: press the Qsoft key (up arrow)

$ To select the next sub-item within a group: press the Osoft key (right arrow)

$ To select the previous group: press the osoft key (left arrow)

$ To confirm: press the lsoft key

Extra Functions

§ Exit the Setup menu:

press the oosoft key

> Restart your application

§ Print parameter settings:

– When the balance/scale functions are displayed, press r

> Printout (example)

Texts with more than 20 characters are truncated

SETUP

Application

---Application 1 (basi Off Application 2 (cont Off Application 3 (data Off Extra function (F4) Off Extra function (F5) Off Auto-start app. whe Off

---Application Parameters (Overview) οfactory settings

√user-defined setting(s)

Setup – Application 1 ο Off (weighing only) Application (Basic settings)

parameters Toggle wt. units Weight unit 1 ο Grams /g

Kilograms /kg*

Carats /ct Pounds /lb*

Ounces /oz*

Troy ounces /ozt*

Hong Kong taels /tlh*

Singapore taels /tls*

Taiwanese taels /tlt*

Grains /GN*

Pennyweights /dwt*

# Milligrams /mg Parts per pound //lb*

Chinese taels /tlc*

Momme /mom*

Austrian carats /K*

Tola /tol*

Baht /bat*

Mesghal /MS*

Display accuracy 1 ο All digits Fewer for weight changes Last digit off*

Weight unit 2 # Grams /g

Kilograms /kg*

Carat s/ct Pounds /lb*

Ounces /oz*

Troy ounces /ozt*

Hong Kong taels /tlh*

Singapore taels /tls*

Taiwanese taels /tlt*

Grains /GN*

Pennyweights /dwt*

ο Milligrams /mg Parts per pound //lb*

Chinese taels /tlc*

Momme /mom*

Austrian carats /K*

Tola /tol*

Baht /bat*

Mesghal /MS*

Display accuracy 2 ο All digits Fewer digits for weight change Last digit off*

For counting through air buoyancy correction, see the following pages Applications 2 and 3 see following pages

Factory setting Factory setting

Setup – Application 1 Counting Accuracy – Average ο Display accuracy Application (Basic settings) piece weight calculation + 1 decimal place

parameters (resolution) + 2 decimal places

Average piece Off

weight updating Manual ο Automatic Percent weighing Weight storage accuracy ο Display accuracy

(resolution) + 1 decimal places + 2 decimal places Decimal places for None

readout in percent 1 decimal place ο 2 decimal places

3 decimal places 4 decimal places 5 decimal places 6 decimal places 7 decimal places Display calculated value ο Residue

Loss Ratio 1 (DR) Ratio 2 (OR) Animal weighing* Animal activity Calm

(averaging) Normal

Active

0.1% of the animal/object 0.2% of the animal/object 0.5% of the animal/object 1% of the animal/object 2% of the animal/object ο 5% of the animal/object 10% of the animal/object 20% of the animal/object 50% of the animal/object 100% of the animal/object

Start Manual mode

ο Automatic mode Minimum load for automatic None

storage 10 digits

20 digits 50 digits ο 100 digits

200 digits 500 digits 1000 digits Decimal places in None

result display 1 decimal place ο 2 decimal place 3 decimal places 4 decimal places 5 decimal places 6 decimal places 7 decimal places

Printout None

ο Average weight only Average and calculated value Recalculation*

Factory setting

Setup – Application 1 Calculation Decimal places in None

Application (Basic settings) calculated result 1 decimal place

parameters ο 2 decimal places

3 decimal places 4 decimal places 5 decimal places 6 decimal places 7 decimal places

Density Method Density of liquid

ο Buoyancy Displacement Pycnometer

Liquid causing ο Water

buoyancy Ethanol

User-definable

Decimal places None

for disp. of vol./ 1 decimal place

density ο 2 decimal places

3 decimal places 4 decimal places

Printout ο None

All data

Differential weighing Weighing sequence* Individual weighing

Consecutive individual weighing ο Combined weighing

Serial weighing

Tare weighing No

ο Yes

Result with None

decimal point 1 decimal place

ο 2 decimal places 3 decimal places 4 decimal places 5 decimal places 6 decimal places 7 decimal places

Autosave ο Off

values On, first value at stability

On, last value at stability On, value bet.

70 – 130% at stabil.

Minimum load None

for autosave 10 digits

ο 20 digits 50 digits 100 digits 200 digits 500 digits 1000 digits

Save statistics ο No

Yes Air buoyancy correction: See next page

Applications 2 and 3: see following pages

Factory setting

Setup – Application 1 Differential weighing Generate printout None

Application (Basic settings) ο Automatic after backweighing

parameters Auto. after initial weigh.

and backweighing

Automatic after tare-, initial-and backweigh.

Include sample ID in text line ο No Yes

Wg.seqkey No

ο Yes Clear sample after individual ο No weight, result + and unload Yes

Last residual weight is stored ο No as initial weight (ashing) Yes Air buoyancy correction Air density determination ο No Yes Change steel/aluminum ο No

references Yes

Diameter determination Application 2 ο Off

(Control functions) Checkweighing* Activation of port lines ο Within checkweighing range Always on

Stability and checkweighing range At stability

Stability + checkweigh.

range on

Type of checkweighing input ο Target, min., max. weight Minimum, maximum weight Target, min. in %, max. in % Weight display mode ο Absolute value

Difference from the target Automatic printout Yes

of OK values ο No

Time-controlled Function after time interval Beep (acoustic signal)

functions Lock in readout

ο Automatic printout of values Store value in applicat. 3 memory (totalizing, formulation, statistics) Automatic function restart ο On

Off

Storage mode ο Without stability After stability After higher stability Print then tare Off

ο On Factory setting Factory setting

Setup – Application 3 ο Off Application (Data records)

parameters Totalizing* Automatic storage ο Off

On, first value at stability On, last value at stability On, value bet. 70% – 130%

at stability Minimum load for None automatic storage 10 digits

ο 20 digits 50 digits 100 digits 200 digits 500 digits 1000 digits Source of data for auto storage ο Application 1

Application 2 Evaluated values ο Net

Calculated Net + calculated

Evaluation mode, MRfunction ο Intermediate evaluation, print Final evaluation, print Intermediate evaluation, display+print

Final evaluation, display+print M+/M– function, then tare ο Off

On Printout of individual No

components ο Yes

Formulation* Automatic storage ο Off

On, first value at stability Minimum load for None

automatic storage 10 digits ο 20 digits 50 digits 100 digits 200 digits 500 digits 1000 digits Source of data for ο Application 1 automatic storage Application 2 Evaluated values ο Net

Calculated Net + calculated

Evaluation mode, MRfunction ο Intermediate evaluation, print Final evaluation, print Printout of individual No

components ο Yes

Factory setting Factory setting

Setup – Application 3 Statistics Automatic storage o Off

Application (Data records) On, first value at stability

parameters On, last value at stability

On, value 70% – 130%

at stability

Minimum load for None

automatic storage 10 digits o 20 digits 50 digits 100 digits 200 digits 500 digits 1000 digits Source of data for o Application 1 automatic storage Application 2

Evaluated values o Net

Calculated Net + calculated

Evaluation mode, o Intermediate evaluation, print MRfunction Final evaluation, print

Intermediate evaluation, display + print Final evaluation, display + print M+/M– function, then tare o Off

On Printout of individual No

components o Yes

Extra Functions – for other extra functions, see next page

Factory setting Factory setting

Setup – Extra Functions ο Off Application F4 key

parameters Extra Functions 2nd tare memory Container tare weight ο No

F5 key Yes

Automatic printout Net value Tare/preset tare ο Off

Identification Printout Automatic, if configured

codes Once after pressing r

(IDs) if configured

ο Each time the print key (r) is pressed Once for M+ function (application 3 memory) Manual storage

in app. 3

memory M+ (totalizing, formulation, statistics) Changing the resolution Product data memory

SQmin Function Display o Text display

Bargraph

Print in o Off

GLP Header On

DKD uncertainty Display1) U (absolute)

of measurement U* (relative)

PG (process accuracy)

Print1) U (absolute)

U* (relative)

PA (process accuracy) Process accuracy Input: 1–100,

5 decimal places

Auto-start On

application when ο Off power goes on

Factory settings No

only for Yes

application-parameters

Factory setting