This thesis has thus far examined haptic feedback methodologies for augmenting tool-tissue interactions and for informing the operating surgeon if they begin to exert excessive force. This chapter examines the case in which tool-tissue contact is to be avoided altogether. “Active constraints” [6] also known as “virtual fixtures” [106] were introduced in Chapter 2. Active constraints exploit synergistic human-robot control to guide a surgical tool to or away from pre-defined anatomical regions [167][108]. This principle has been adopted extensively in orthopaedic applications for the precise milling of bone for implant preparation [105]. Active constraints have mainly been applied to grounded, cooperatively-controlled robotic systems, which permit haptic feedback to the surgeon. Examples include the original Acrobot™ (Active Constraint Robot, see Figure 6.1), the subsequent Sculptor RGA (Stanmore Implants, Elstree, UK.), RIO® MAKOplasty® (Stryker Corp., MI, USA) and the

Steady-Hand robot for Otorhinolaryngology applications [108].

Figure 6.1: The Acrobot™ surgical system developed at Imperial College (©2002 reproduced from Jakopec et al with permission from Springer) [105].

150

Figure 6.2: The active constraint principle for orthopaedic surgery from Jakopec et al ©2003 IEEE [107].

These grounded robotic platforms use force control to guide the operator along or away from the active constraint. This principle, developed by Davies et al [107] is highlighted in Figure 6.2. RI, RII and RIII refer to three separate regions. RI is a safe region in which the robot does not impede the motion of the surgical tool. RII is a region of impedance; the robot actively resists motion in the direction normal to the pre-defined boundary. The final region, RIII, is the “no-go” boundary at which point the robot becomes very stiff to prevent penetration of the surgical instrument tip in to an unsafe region of anatomy. The application of active constraints to ungrounded, hand-held instruments requires an alternative approach.

Hand-held surgical devices with active constraints must implement position-based control on the tool tip rather than force-based constraints because of the absence of a grounding frame. The Navio PFS (BlueBelt Technologies, Plymouth, MN, USA) and the device developed by Brisson et al [51] are examples of hand-held tools that have been devised for precise milling of pre-defined 3D geometries in knee replacement surgery. These devices are optically- tracked and their rotating burrs are retracted when the constraint is encountered to allow intelligent sculpting of a bone surface. Active constraints have also been implemented on

151

Micron [48][49]. In this work, the instrument tip’s position is constrained so as to not deviate from a pre-programmed trajectory. This research also features ‘soft’ positional constraints that effectively scale the tool tip position relative to the hand motion of the surgeon. The application of active constraints significantly improves tool tip accuracy in micromanipulation tasks; however there is no force-feedback to the operator that indicates tool interaction with the active constraint.

The novel contribution of this chapter is the fusion of position-based active constraints and force-feedback in to an ungrounded, hand-held device. The motion compensation scheme provides a position constraint on the surgical tool whilst the force-feedback allows for tactile interaction with the active constraint. This technique could allow for surgical tools to be guided through safe channels with less cognitive burdening of the operator. It is envisaged that such a device has potential applications in neurosurgery, otorhinolaryngology and paediatric surgery, in which surgeons must operate around delicate anatomical areas within tight spatial constraints and with restricted vision. This chapter includes bench tests that verify the motion compensation and force-feedback aspects of the proposed methodology. User studies also demonstrate the combined impact of motion compensation and force- feedback in allowing human perception of an active constraint with an ungrounded device.

6.1 Motion Compensation Device with Force-Feedback

The device is designed to be held as a stylus by the operator with a surgical tool protruding from the tip as shown in Figure 6.3. The surgical tool can be actuated along its axis so as to retract into the stylus as it encounters an active constraint. A force-feedback display exerts forces on to the operator’s fingertip when the active constraint is penetrated. An Optotrak Certus® (Northern Digital Inc, Ontario, Canada) optical tracking system is used in order to locate the device’s position and orientation with respect to the active constraint. The optical tracking system is capable of achieving sub-millimetre resolution and operates at acquisition speeds in excess of 1kHz.

152

Figure 6.3 A photograph of the hand-held device illustrating its key features of motion compensation and force-feedback.

6.1.1 Instrument Tracking and Active Constraints

Four optical markers are attached to the casing of the device and are tracked by the optical tracking system. A rigid body coordinate system is defined at the origin of the optical markers (Fmarkers) which is known with respect to a world coordinate system (Fworld) defined

by the Optotrak Certus®. Rigid transformations are computed in order to obtain coordinate

systems at the base of the surgical tool (Fbase) and at the uncompensated tool tip (Ftip). These

transformation computations are as follows:

 𝑊𝑀𝑇 is the transformation from the F

world to Fmarkers which is provided by the optical

tracking system as the tool is manipulated by the user.  𝑀𝐵𝑇 is the transformation from the F

markers to Fbase which is fixed by the design of the

device.

 𝐵𝑇𝑇 is the transformation from the Fbase to Ftip which is fixed by the design of the

153