VP-01 Preparation for Alignment and Verification
3. Verify that the Data bits, Parity and Stop bits are configured according to the
CELL-DYN SETUP menu.
4.
In the CELL-DYN SETUP menu, verifythe CTS enabled setting:
1.
The test session window named sertest1 is now active.Note
If the message "Unable to open COM1.
Please check your port settings"
displays, there may be another device assigned to the COM1 port, such as the modem. Contact the local area IS
administrator for assistance with resolving device interrupt conflicts.
2.
Compare the results to the interface specification. The screen is shown for example only. Transmit specimen data as many times as desired.Transmit Data
1. From the MAIN Menu, press
[DATALOG], followed by [TRANSMIT DATA].
2. Enter starting and ending sequence numbers of data to be transmitted.
3.
Compare the results to theCELL-DYN 1800 Interface Specification.
Note
The screen below is shown for example only.
4.
If desired, click on File then Print to obtain a printout. When testing is complete, click on File then Save. After saving, click on Call, then Disconnect, then exit the window by clicking the X.Close out the Hyper Terminal window by clicking the X.
VP-14 Hard Disk Drive Setup and Verification
Purpose Provide instructions to verify/modify CMOS Settings, partition and format replacement hard disk drives
Module CELL-DYN 1800
Materials Required
DK, Drive, Hard, IDE (8200570701)
DKS (C) SRV ABBOTT CD3200 (8938070001) CELL-DYN 1800 Installation Disk
Phillips Screwdriver
Time 00:30 min
Action Steps Reference
Preparation Note
If replacing an existing customer hard disk drive, print copies of all customer settings, quality control data files, and the data log if the drive is accessible. It may not be possible to access this data if the hard disk drive is corrupt or defective.
1. Power OFF the system.
2.
Verify jumper setting is set to Master according to Jumper Configuration Table on hard disk drive.Install Hard Drive
1.
Install the new hard drive using the existing hardware and cables (refer to RR-F1.04 Hard Disk Drive).Access CMOS Setup Utility Screen
1.
Power ON the analyzer and press the Delete key, when prompted, to display the CMOS Setup Utility Screen.Verify CMOS Settings:
Standard CMOS Setup
1. Use the up/down arrow keys to select the STANDARD CMOS SETUP.
2. Press the Enter key.
3. Set Pri Master to Auto.
4.
Set 32 Bit Mode to ON.Note
Pri Slave, Sec Master and Sec Slave show Not Installed.
5.
Press ESC to exit the screen.Verify CMOS Settings:
Advanced CMOS Setup
1. Use the up/down arrow keys to select the ADVANCED CMOS SETUP.
2. Press the Enter key.
3. Set Quick Boot to Enabled.
4. Set 1st Boot Device to Floppy.
5. Set 2nd Boot Device to IDE-0.
6. Set 3rd Boot Device to Disabled.
7. Set 4th Boot Device to Disabled.
8.
Press ESC to exit the screen.Verify CMOS Settings:
Advanced Chipset Setup
1. Use the up/down arrow keys to select the ADVANCED CHIPSET SETUP.
2. Press the Enter key.
3. Set LCD CRT Selection to Simultaneous.
4. Set LCD Type to Type 8.
5.
Press ESC to exit the screen.Verify CMOS Settings:
Power Management Setup
1. Use the up/down arrow keys to select the POWER MANAGEMENT SETUP.
2. Press the Enter key.
3. Set Power Management/APM to Disabled.
4. Set CPU Critical Temperature to 60°C/140°F.
5.
Press ESC to exit the screen.Verify CMOS Settings:
Peripheral Setup
1. Use the up/down arrow keys to select the PERIPHERAL SETUP.
2. Press the Enter key.
3. Set OnBoard FDC to Enabled.
4.
Press ESC to exit the screen.Save CMOS Settings and Exit CMOS Setup
1. Place the DKS (C) SRV ABBOTT CD3200 (8938070001) (service disk) into the floppy drive.
2.
Press F10.The message Save current settings and exit (Y/N)? displays.
3.
Type Y and press Enter.Note
Saves the CMOS settings, exits the CMOS Setup Utility and restarts the computer.
Create DOS Partition 1. At the A:\> prompt, type FDISK and press Enter.
2. At the FDISK Options menu, enter choice 1 and press Enter.
3.
At the Create DOS Partition or Logical DOS Drive menu enter choice 1 and press Enter.Note
If you get the message Primary DOS Partition already exists, press Esc to exit. Then go to Format Hard Drive, Step 1.
The message Do you wish to use the maximum available size for a Primary DOS Partition and make the partition active (Y/N)? displays.
4.
Type Y, and press Enter.Note
Uses the maximum size for the hard drive DOS partition and makes it active.
5.
When the screen displays System will now restart. Insert DOS system diskette in drive A:. Press any key when ready, press Enter.The system reboots and displays the A:\ prompt.
Format Hard Drive
1.
At the A:\> prompt, type FORMAT C: /S and press Enter.The message WARNING: ALL DATA ON NON-REMOVABLE DISK DRIVE C: WILL BE LOST! Proceed with Format (Y/N)? displays.
2.
Type Y, and press Enter.The message Volume label (11 characters, ENTER for none)? displays.
3.
Press Enter for None.Install DOS and System Software
1.
At the A:\> prompt, type INSTALL and press Enter.Note
Installs DOS and system files.
2. When the system installation is complete, remove the service disk and insert the CELL-DYN 1800 Installation disk into the floppy drive.
3.
Type Install and press Enter.Note
When the installation is complete, the message "Power off the instrument and remove diskette. Power on the instrument." displays.
4.
Remove the CELL-DYN 1800 Installation disk and power OFF the instrument.Verification 1. Power ON the instrument
2.
Verify that the system boots up properly to the application software.Note
During initial power up after a hard disk drive installation, DO NOT turn off the instrument while the Data Log is being created. This corrupts the Data Log.
3.
If printouts are available, re-enter the data from the printouts. If not, perform instrument calibration procedure and verify control recovery.VP-15 Software Installation/Upgrades
Purpose Install or Upgrade Software Module CELL-DYN 1800 Materials Required Disk, installation (8938144001) Time 00:10 min
Action Steps Reference
Software
Installation/Upgrade
1. With the CELL-DYN instrument ON, go to the DIAGNOSTICS Menu.
2. Press ESC on the PC keyboard. A warning message displays on the screen.
3. Press Alt and x (simultaneously). The following DOS prompt displays:
C:\CD1800>.
4. Insert the CELL-DYN 1800 Installation disk into Drive A.
5. Type a: and press Enter.
6.
Type install and press Enter.Note
When the installation is complete, the message "Power off the instrument and remove diskette. Power on the instrument." displays.