• No results found

1166.pdf

N/A
N/A
Protected

Academic year: 2020

Share "1166.pdf"

Copied!
94
0
0

Loading.... (view fulltext now)

Full text

(1)

Page

TABLE OF CONTENTS... i

LIST OF FIGURES ... v

LIST OF TABLES ... vi

ACKNOWLEDGMENTS... vii

ABSTRACT ... viii

L INTRODUCTION A. Background on Incineration... 1

B. Scope of Project ... 3

C. Literature Review ... 4

11. METHODS AND MATERIALS A. Description of Incinerator ... 9

B. Waste Combustion Process ... 9

C. Incineration Procedure ... 12

D. Description of Animals Used for Test Burns ... 13

E. 35s Sulfate Used for Test Burns ... 13

F. 35s Methionine Used for Test Burns... 14

G. Overview of Burns ... 14

H. Breakdown of the 18 Burns ... 15

I. Burn Sequence... 15

K. Sampling Methods 1. Description of Ash ... 16

2. Rough Ash Sampling... 16

3. Ground Ash Sampling ... 17

(2)

J. Liquid Scintillation Counting

1. Theory of Operation ... 18

2. Problems with Quench ... 19

3. Quench Curves and Standards ... 20

4. Ash Settling ... 22

5. Measurement of Rough Ash with the Liquid Scintillation Counter ... 27

6. Measurement of Ground Ash with the Liquid Scintillation Counter ... 28

7. Measurement of Extract Fluid with the Liquid Scintillation Counter ... 28

K. Windowless Gas Flow Proportional Counter 1. Theory of Operation ... 30

2. Characteristic Curve for Proportional Counter ... 31

3. 35s Extract Fluid Standard for the Proportional Counter ... 33

4. Measurement of Extract Fluid with the Proportional Counter ... 35

5. Ash Sample Size for Proportional Counter .... 35

6. Measurement of Ash with the Proportional Counter ... 37

III. RESULTS A. Statistical Analysis of the Data... 41

B. Problems Encountered in Rough Ash Measurement ... 41

C. Mass Reduction Efficiency of the Incinerator ... 43

D. Calculation of Retention Efficiency ... 46

(3)

F. Effects of Glass on Retention Efficiency... 52

G. Comparison of Sampling Methods ... 54

H. Determination of Extraction Procedure Efficiency ... 54

I. Differences Between Detectors ... 61

IV. CONCLUSIONS ... 64

V. RECOMMENDATIONS ... 67

VI. LIST OF REFERENCES ... 68

VII. APPENDICES A. Extraction Procedure ... 70

B. Procedure for Using the Proportional Counter ... 72

C. Set-up for the Proportional Counter ... 73

D. Calculation of Activity for Ash Using Data Obtained from the Proportional Counter ... 74

E. Calculation of Activity for Extraction Fluid Using a Proportional Counter ... 76

F. Calculation of Retention Efficiency ... 77

G. Retention Efficiencies 1. Rentention Efficiencies for Different Chemical Forms of 35s ... 78

2. Retention Efficiencies for Burns with Glass and without Glass ... 79

3. Retention Efficiencies for the Different Detectors ... 80

4. Mass Reduction Efficiencies... 80

(4)

2. Data on the Calculation of the Activity of the Ash Using the Liquid Scintillation

Counter ... 82

3. Data on the Calculation of the Activity of the Extract Fluid Using the Liquid

Scintillation Counter ... 83

4. Data on the Calculation of the Activity of the Extract Fluid Using the Proportional

Counter ... 84

Calculation of the Retention Efficiencies

1. Data for the Calculation of Retention

Efficiency Using the Ash and the

Proportional Counter ... 85

2. Data for the Calculation of Retention

Efficiency Using the Ash Measured with

a Liquid Scintillation Counter ... 86

3. Data for the Calculation of Retention Efficiency Using Extract Fluid and the

Liquid Scintillation Counter ... 87

4. Data for the Calculation of Retention

Efficiency Using the Extract Fluid and the

(5)

Page 1. Controlled Air, Dual Chamber, Pathological

Waste Incinerator... 10

2. Quench Curve for Ground Ash ... 21

3. Quench Curve for Extraction Fluid and Carbon Tetrachloride ... 23

4. Settling Pattern of Ash over Time Using Ash from Burn 3 ... 24

5. Settling Pattern for Extraction Fluid from Burn 3 ... 25

6. SIE and CPM for Several Ash Quantities ... 29

7. Plateau Curve for 35S Ash Standard ... 32

8. Relationship between the Activity Measured and the Amount of 35S Standard Added ... 34

9. Count Rate vs. Ash Weight for Ash with Standard Added... 38

10. Absolute Efficiency vs. Weight for Ash with Standard Added ... 39

11. Comparison of Rough Ash and Ground Ash Using Ash from Burn 17 ... AA 12. Mass Reduction of Incinerated Wastes ... 45

13. Retention Efficiency Measured by Proportional Counter ... 55

(6)

LIST OF TABLES

Page

1. Analysis of Variance ... 42

2. Mass Reduction of Incinerated Wastes ... 47

3. Retention Efficiencies for Ash and Extract Fluid ... 48

4. Effects of Chemical Form and Glass on

Retention Efficiencies ... 50 5. Effects of Chemical Form on Retention Efficiencies ... 5 1

6. Effects of Glass on Retention Efficiencies ... 53

7. Activities of Ash and Extract Fluid from Consecutive

(7)

ACKNOWLEDGMENTS

I wish to thank the following people in the Health and Safety Branch

at NIEHS for their help and advice:

Chris Hunt

Valeria Shropshire

Scott Merkle Patricia Deese Diane Wilson

Thanks to my committee members Dr. Philip Hamrick, Dr. James

Watson and Dr. Douglas Crawford-Brown for their guidance, suggestions and constructive criticism.

A special thanks to Dr. Philip Hamrick, William Fitzgerald and

Thomas Smith for their guidance and expert advice. It was a pleasure and a honor working with them.

(8)

WILLIAM E. DAVIDSON. The Measurement of 35S Retention in

Incinerator Ash. (Under the Direction of Dr. Philip E. Hamrick and Dr. James E. Watson, Jr.)

ABSTRACT

The two main objectives of this study were to measure

quantitatively the retention of 35$ in residual incinerator ash and to

determine the optimal ash sampling method for use at NIEHS. In

pursuing these objectives, 12 test burns with known activity and 6

control burns were performed using a dual chamber, starved-air,

pathological waste incinerator.

The two variables examined for their affects on the retention

efficiency of 35s in the ash were the type of vial incinerated (glass vs.

polyethylene) and the chemical form of 35s (methionine vs. sulfate).

The three sample forms tested were ash taken directly from the incinerator, ash ground in a jar mill grinder and extraction fluid obtained using EPA's Extraction Procedure and Toxicity Test Method

1310. Samples from all 18 burns were counted using a liquid

scintillation counter and a gas flow proportional counter. Because the data collected on the rough ash were considered to be biased and

inconsistent, they were not used for any calculations of retention or mass reduction

A statistically significant difference in the mean 35s retention

was found between the methionine and sulfate forms of 35s. The

(9)

for methionine was 1.51%. Ash from burns containing polyethylene and glass vials had a mean retention efficiency of 8.22% which was

significantly higher than the 3.84% for burns with polyethylene vials only.

It was concluded that until further research is done on the

extraction procedure it is best for NIEHS to use the gas flow

proportional counter and ground ash for routine sample counting to

(10)

Background on Incineration

The National Institute of Environmental Health Sciences

(NIEHS) is a biomedical research facility. Like other research

facilities NIEHS uses radioactive materials in much of its research

and must dispose of the radioactive materials once they become a

waste.

Burial in a low-level radioactive waste site in the past was the

most widely used form of disposal for low-level radioactive wastes. However, due to the increased cost of landfiUing low-level

radioactive wastes, many facilities are turning to incineration, which

has been shown to be an economically feasible alternative to landfill

disposal (Gregory and MaiUe, 1975; Glauberman and Loysen, 1963: Machis and Geyer, 1952).

The incineration process thermally oxidizes the organic

constituents in the waste stream and converts all the wastes into a

stabilized non-combustible form while greatly reducing the mass of

the wastes. Most wastes can be reduced to less than 10% of their

original mass (Glauberman and Loysen, 1963). The degree to which

the mass of the wastes is reduced affects the disposal cost. Reducing wastes to the smallest mass possible results in lower disposal costs.

The residual ash from the waste must be disposed of in an

appropriate land burial facility. The license granted by the Nuclear Regulatory Commission allows NIEHS to dispose of residual ash in a

municipal landfill if it contains a concentration of one or more

(11)

limits in 10 CFR 20 Appendix B then the ash must be shipped to a

low-level radioactive waste landfill which usually results in an large

increase in shipping and burial costs.

According to the U.S. Atomic Energy Commission (1970) the principal

benefits arising from incinerating combustible radioactive solid

wastes are:

1 - Volume Reduction - The reduction of wastes through incineration reduces the space needed for land burial. This results in less

material needed for packaging and less land needed for burial

2- Conversion of the Wastes to Noncombustlble State - Spark induced fires are common in normal landfill operations. Since the ash is non-combustible, it does not represent a

fire hazard.

3- Improved Lone-Term Environmental Safetv Conditions - Since there is a smaller volume of material, it is easier to monitor and maintain the site. The ash is also less likely to leach into

the ground water.

4- Retrievabilitv - Until a final solution is found for the disposal of radioactive wastes, it is necessary to maintain retrievability..

Incinerated wastes would be easier to retrieve and rehandle.

ͣ

S- Incineration Would Minimize Land Requirements for Burvine

Radioactive wastes - In recent years landburial has become increasing less acceptable for waste disposal and siting new landburial facilities has become very difficult. Because of the large reduction in the mass of the wastes through incineration,

(12)

1969. The three most common radioisotopes incinerated at NIEHS

are l^^C, 3h and 35s. Since 1980, the use of 35s has increased by a

factor of 10 while the use of I'^C and 3h has increased only slightly.

The fate of 35s was investigated in a previous study at NIEHS by

measuring activities in refractory ash and gaseous effluent (Hamrick

et al.. 1989).

The study by Hamrick et al. (1989) concentrated mainly on the

measurement of radioactivity in the gaseous effluent. However, some ash samples were collected and counted using a liquid

scintillation counter. The activities accounted for in both the ash and

the gaseous effluent summed to less than 50% of the total in 3 of the

11 burns and less than 95% in 9 of the 11 burns. This resulted in

doubts about the accuracy of the method for measuring the ash.

The present study investigated 33s levels in the residual ash

only. The 35s concentration in the residual ash was measured using

three sampling methods; ground ash, rough ash and extraction fluid obtained using the EPA's Extraction Procedure and Toxicity Test

Method 1310. The activity of the samples was determined using a

liquid scintillation counter and a gas flow proportional counter.

In the present study, two forms of 35s, a organic (methionine)

and an inorganic (sulfate), were examined for their effects on the

retention of 35s in the residual ash under normal operating

conditions at NIEHS. The 35s was added to several scintillation vials

(13)

be found in the waste stream at NIEHS.

By determining the extent to which different chemical forms of

a radioisotope are retained in the ash, NIEHS can make better

economic decisions concerning their disposal. Wastes containing chemical forms of radioisotopes, retained in ash in high quantities compared to other chemical forms, could be separated and held for

decay before being incinerated. Disposal costs could then be reduced

by storing the wastes containing the isotope forms which were retained in higher amounts until they had decayed to a lower

activity. At this activity they could be incinerated and disposed of in

a municipal landfill. Wastes containing isotope forms which are

retained, in the ash, in lower amounts could be incinerated without

storage and disposed of in a municipal landfill.

Incineration of glass may also affect the retention of

radioactivity. Instead of being released out the stack as gaseous

effluent, the radioisotopes may adhere to the glass and remain in the

residual ash thus making a higher concentration of radioactivity in

the ash. The effects of glass on the retention of 35s in the residual

ash and on the mass reduction of wastes were examined in this study.

Literature Review

The use of radioisotopes for research in the United States has increased significantly since 1945 (Machis and Geyer, 1952).

Because incineration is a practical and economical means of waste

(14)

burial costs (Glauberman and Loysen, 1963). A 95% mass reduction of wastes is common for incinerators (Glauberman and Loysen, 1963; Classic et al., 1985). Some of the major factors considered in

determining the feasibility of incineration are activity, half-life,

maximum permissible air concentration of the isotope and the size of

the animal carcass incinerated (Gregory and Maille, 1975).

With the increase in incineration of radioactive wastes comes the need to determine the fate of the radioisotopes. No significant difference has been found, in previous studies, (Bush and Hundal,

1972; Hamrick et al., 1989) in the retention of 35$ between chemical

forms. However, the tests were not usually representative of normal

operating conditions. In the Bush and Hundal study (1972), the animals and activity were put in stainless steel containers to be

incinerated and were not mixed with the rest of the wastes. In the

Hamrick et al. study (1989) the waste stream that was burned

during the tests was not the same as the typical waste stream at that installation.

During incineration the radioisotopes are deposited, in various proportions, in the gaseous effluent, residual ash and on the interior of the stack (Machis and Geyer, 1952). Bush and Hundal (1972) reported that the fate of incinerated radioactive material was dependent upon several factors including:

(15)

Bush and Hundal reported in 1972 that the retention of 35s in

ash for two inorganic forms of 35$ ranged from 10 to 90% for 35s sodium sulfate and from 40 to 77% for potassium sulfate. This

difference was not found to be significant. No significant difference in retention was detected for various chemical and physical forms of

32p and I25i. Samples on the order of a few [iCi or less were used.

These samples were placed in small metal cans, covered with screens

and put into an incinerator with non-radioactive waste.

A study performed by Classic et al. (1985) reported a 0%

retention of 35s in the ash for Guinea pigs which were placed in

stainless steel containers for incineration and injected with labeled

microspheres.

Katsikis et al. (1984) reported that no activity remained in the ash when ^'^C, 3H and 35s were incinerated with other pathological

wastes. They believed, a priori, that complete combustion would

result in all l^C, 3h and 35$ going up the stack as CO2 , H2O and SO2.

No details were given on the testing methods.

In 1952, a study at The Johns Hopkins University (Machis and

Geyer, 1952) found that 20 to 100% of the radionuclide 32p remained

in the ash after incineration. The reasons given for the wide

variability in the results were the difficulty in mixing the samples due to the range in particle size and the limitation on the sample size

(16)

Knapp (1984) studied the fate of I'^C and 3h during incineration

at NIEHS. No activity was found on the refractory surface and less than 0.1 % was found in the ash. It was assumed, therefore, that

99.9% of the l^^C and 3h went out the stack as off-gases.

A study was conducted at NIEHS (Hamrick et al., 1989) to

measure the amount of 35s that was contained in stack effluent and

in the refractory ash. Approximately 2.3% of the 35s methionine and

4.3% of the 35s sulfate was measured in the ash. However, only a

total of 89% of the 35s methionine and 37% of the 35s sulfate was

detected in the ash and gaseous effluent. The aqueous 35s samples

were placed in plastic jugs and the plastic jugs were placed in boxes

for incineration.

The studies presented here give ranges of 35s retention in ash

from 0 to 90%. The main difference between the studies seems to be

the testing procedures. There is wide variability between the testing

procedures of the studies. These differences in the testing

procedures may account for the variability in the retention of 35s.

Because of previous problems at NIEHS with measuring the ash directly in a liquid scintillation counter, an extraction procedure developed by the EPA was tested in this study. The extraction procedure was designed to quantify the metals, pesticides and

herbicides present in the solid wastes (Kimmell, 1986). This is done

by removing the substance of interest from the solid waste into a

(17)

(Ham et al.. 1978). *

For this study, it was thought that by removing the 35s from

the ash into the extraction fluid the total amount of 35$ originally in

(18)

A controlled air, dual chamber pathological waste incinerator

was used for this study (Figure 1). The incinerator is owned and

operated by NIEHS and is permitted to burn up to 125 ft3/hr or 350

Ib/hr of type IV waste. Type IV waste consists mainly of organics

which contain up to 85% moisture and have a heating value of 1000

Btu/lb. The incinerator is also licensed by the Nuclear Regulatory

Commission (NRC) to burn radioactive wastes. The lower chamber is heated by four 350,000 Btu/hr natural gas burners and the upper

chamber has one 1.5 x 10^ Btu/hr burner with a burner blower for

creating turbulence which supplies the excess oxygen and mixing

needed for complete combustion. The upper chamber of the

incinerator has been modified to allow for a 2 second residence time

to ensure that the wastes are completely combusted.

Waste Combustion Process

The incineration process utilizes thermal decomposition to

destroy the organic part of the waste and to reduce its total volume

(Oppelt, 1987). Under optimal operating conditions nearly all

combustable wastes will be in the form of residual ash or gaseous

effluent consisting of CO2 and H2O. It is not known if the incinerator

used for this study was operating under optimal conditions.

Wastes are pushed into the lower chamber via a hydraulic ram

charger. The lower chamber is normally operated at a temperature

of between 1200 - 1800 degrees F with 50 - 80% excess air. Due to

(19)

Figure I

Controlled Air. Dual Chamber, Pathological

Waste Incinerator

STACK

ROOF

DOOR

FAN

UPPER

BURNER

UPPER

CHAMBER

/,^

\

LOVER

CHAMBER

LOVER BURNERS

AIR

litff ft

0 0 0 0 0 0 0

RAM

LOADER

CONTHOL PANEL

(20)

#

This results in the formation of volatile hydrocarbons. CO and

combustion products. The secondary chamber normally has

100 - 200% excess oxygen and a temperature of 1400 - 2000

degrees F. The high temperature, high turbulence, and long

residence time of the secondary chamber results in the complete combustion of the CO and volatile hydrocarbons from the primary chamber. A 100% combustion efficiency can be obtained under

optimum operating conditions in which only water vapor (H2O) and

CO2, and in this case SO2, compounds would be discharged via the

stack. All other non-combustible material would remain as residual

ash.

2 C (waste) + O2 (air)---i>' CO (product of incomplete

combustion)

2 CO + O2 (air)---^ CO2 (product of complete

combustion)

2 H2 (waste or air) + O2 (air) ~^ H2O (water vapor)

(21)

Incineration Procedure

For each burn, five cardboard boxes were filled with 65 - 75 kg

of used rat bedding (wood shavings), 800 scintillation vials (Packard

Instrument Co., Inc., Meridan, CT) containing scintillation fluid (Ready

Safe, Beckman Instruments, Inc. FuUerton, CA) and approximately 50

sheets of lab bench paper (ATD American Co., Wyncote, PA). There

were also 5-12 boxes of assorted size containing a total of 75 - 85

kg of mouse and rat carcasses. These items are typically found in the

waste stream at NIEHS in variable amounts. The decision on the type

and quantity of each material incinerated during the test burns was

based on comments made by the incinerator operator on what was in

a typical NIEHS waste stream.

For each of the spiked burns 35s, in aqueous form, was pipetted

into 80 - 120 out of a total of 800 vials. The boxes were loaded into

the incinerator at a rate of 1 - 2 every 7 minutes, alternating between boxes containing animal carcasses and boxes containing

scintillation vials, animal bedding, and lab bench paper. There were

no boxes burned until the incinerator temperature was 1300 - 1400

degrees F in the primary chamber and 1800 - 1900 degrees F in the

secondary chamber. The incinerator continued burning at these

temperatures for 6 hours after the last box was loaded, at which time

it automatically shut itself down. The incinerator was allowed to cool

down over night after each burn. After cooling, the ash was swept from the bottom of the refractory using a straw broom and a flat-headed shovel. The ash was collected in a bucket and weighed on a

triple beam balance (Ohaus heavy duty solution balance, Florham

(22)

#

Description of Animals Used in Test Burns

NIEHS conducts much of its research using laboratory mice and

rats. After the animals are sacrificed, they are put into boxes which

are placed in large freezers for storage. The animals were taken from these freezers to the incinerator. The boxes, which varied in

size, usually contained between 50 and 100 mice or between 2 and

30 rats in each box. Each box was labeled with information about the

radionuclides, if any, contained in the animals. The labels on the

boxes used for this study were checked to ensure that only

uncontaminated rats and mice were burned.

35s Sulfate Used for Test Burns

The 5 mCi of 35s sulfate solution used in the spiked 35s sulfate

burns was obtained in a 1 ml volume from New England Nuclear

Research Products, Inc. (E.I. duPont deNemours and Co., Bilerica, MA).

Six 100 |il aliquots of 35s sulfate (0.5 mCi each) were pipetted into

six 100 ml volumetric flasks. The flasks were each filled with 100

ml of distilled water and mixed by inverting them several times. Six

25 III aliquots were pipetted from each flask and placed into 7 ml

scintillation vials to be counted by a liquid scintillation counter (LSC).

All sample vials to be counted in the LSC were filled with liquid

scintillation cocktail (Biofluor, NEN Research Products, Boston, MA)

and mixed using a Vortex Genie (Scientific Industries, Inc.) before

being placed in the LSC. The activity of each vial was calculated by

the LSC and the average activity of the 6 vials was used to calculate

(23)

remaining flasks. Since the activities of the flasks were all measured

at the same time before the first burn with 35s sulfate, the decay was

accounted for in each remaining burn with 35s sulfate. A 1 ml

pipette was used to transfer the 35s sulfate to 100 of the 800

scintillation vials incinerated in each of the six burns.

3^S Mgthionmg Usgd for Tgst Burns

The 35s methionine used in this study was recycled from a lab

at NIEHS. Ten 100 |il aliquots of 35s methionine were pipetted into 7

ml scintillation vials and counted via the LSC. The average activity,

calculated by the LSC in units of dpm/ml, was used as the activity of

the 35s methionine. The activity of the 35s was corrected for decay

before each burn. A 1 ml pipette was used to transfer the 35s

methionine into 80 - 120 scintillation vials for each of the six burns.

Overview of Burns

Each test burn was similar to normal burns at NIEHS except

that the materials incinerated during the spiked test burns contained

a known amount of 35s and no other known radioactivity. Because

the activity in the wastes before a burn and the activity of the ash

after the burn were known, a determination could be made of the

retention efficiency of the 35s in the ash.

A total of 18 trial burns were performed for this study. The

decision to perform only 18 burns was limited primarily by time

restrictions on the use of the incinerator since it was usually only

available for tests burns once a week. Because the incinerator was

(24)

used for this study, six control burns were designed into the schedule

to monitor the background from residual 35s, I'^C and 3h.

Control burns were the same as the spiked burns except that

no 35s was added to the wastes before incineration. The mean

activity of the 6 control burns was subtracted from the activity of

each spiked burn to reduce the affects of residual 35s, l^C, and 3h

from previous spiked burns and normal NIEHS burns on the

calculation of retention.

Breakdown of the 18 Burns

3 control burns with glass vials 3 control burns without glass vials

3 burns with glass, with 35s sulfate 3 burns without glass, with 35s sulfate

3 burns with glass, with 35s methionine

3 burns without glass, with 35s methionine

Burn Sequence 1- control burn with glass

2- burn without glass, with 35s methionine

3- burn with glass, with 35s methionine

4- control burn without glass

5- burn without glass, with 35s sulfate 6- burn with glass, with 35s sulfate

7-control burn with glass

8- burn with glass, with 35s methionine

9- burn without glass, with 35s methionine

10- control burn without glass

11- burn without glass, with 35s sulfate

12- burn with glass, with 35s sulfate

13-control burn with glass

14- burn with glass, with 35s methionine

15- burn without glass, with 35s methionine

16-control burn without glass

(25)

#

17- burn without glass, with 35s sulfate 18- burn with glass, with 35s sulfate

Sampling Methods

Description of Ash

The ash contains products of combustion and uncombusted

material. Ash from burns in which no glass was incinerated

consisted of a mixture of powdery ash and bone fragments. This

type of ash was easily swept out of the incinerator. The burns in

which glass vials were incinerated yielded ash that contained bone fragments, pow.dery ash and various size chunks of molten glass and bone. Most of the glass vials were melted beyond recognition and had combined together into large chunks. However, the shape of some of the glass vials was not distorted. A flat-headed shovel had to be used to remove some of the glass chunks which had adhered to

the refractory surface.

Three ash sampling methods were tested and compared in this

project. The first two methods involve measuring the activity in the

ash by measuring the ash itself. The third method utilizes EPA's

Extraction Procedure (EP) and Toxicity Test Method 1310.

Rough Ash Sampling

After the ash from the incinerator was collected in a stainless

steel bucket, a 100 ml beaker was used to collect several samples

from different parts of the bucket. This was done in an attempt to

(26)

ash in the bucket. These samples were mixed together to form a

sample totalling about 500g. The rough ash is that which is taken directly from the incinerator in an unprocessed form. Since NIEHS

presently measures the ash in this form, it was evaluated in this

study for consistency between samples and its consistency with the

other methods.

Ground Ash Sampling

Using the same procedure for collecting ash from the bucket,

approximately 500g of ash was placed in a jar mill grinder (Model

753RM, US Stoneware, Pahwah, NJ) and ground for one to three hours. The majority of the ash obtained from the jar mill after this

grinding process was in a fine powder form. When the ash samples contained glass, some of the larger chunks of glass were not

completely ground. All ash from the jar mill was sifted through a

1/16 inch screen to remove large particles.

Extract Fluid Sampling

The 24 hour extraction procedure (EPA method 1310) was

designed to simulate the leaching conditions of a sanitary landfill for

a period of 3 to 10 years (Kimmell, 1986). The procedure, given in

Appendix A, was designed to remove several chemicals and metals from incinerator ash. During the extraction procedure, lOOg of ash is

added to distilled water and acetic acid and mixed for 24 hours. The

objective is to remove the chemicals or metals from the ash into the

water and acetic acid mixture (extract fluid) for quantification. In

(27)

#

remove 35s from the ash and into the extract fluid. Since the extract

fluid was clear, it would result in much less quenching which would

require less extrapolation by the liquid scintillation counter (LSC) for

determination of activity. Also, because the fluid would not be

expected to settle-out, as is the case with particulate ash, no settling

problems would need to be remedied.

Liquid Scintillation Counting

Theory of Operation

In liquid scintillation counting the sample is placed in the liquid scintillation fluid contained in a vial. For ideal conditions, this results in 4tt counting geometry and very little sample self

absorption. The vials are automatically moved to a light tight

enclosure between two photomultiplier tubes (Knoll, 1979). The low

energy beta particles from the sample excite the fluor which is part

of the liquid scintillation fluid. The fluor emits de-excitation energy

in the form of light photons (UTP, 1982). As the photons leave the

vial, they are detected by the two photomultiplier tubes (PMT) which

measure the light photons in coincidence. Therefore, electronic noise

pulses and background radiation, which occur randomly, would not

be counted as a pulse. Because most betas have enough energy to

produce more than one photon during an interaction with the fluor, it

is likely that both PMT's will simultaneously receive light photons

from one beta interaction (UTP, 1982). Only 2 photons reaching the

PMT's simultaneously (in coincidence) will be registered as a pulse

(28)

reach the PMT's in coincidence, it is assumed that only photons counted in coincidence represent a beta particle from the source.

However, several problems arise when counting ash samples

using a LSC. The ash, for the most part, is not soluble even after being shaken vigorously. Immediately after shaking, the ash begins to settle to the bottom of the vial. The amount of settling-out on the bottom of the vial increases with time, causing self absorption of the betas and reducing the count rate measured by the LSC compared to

the actual activity of the source in the vial measured. The quotient

of the count rate (cpm), measured by the LSC, divided by the activity

(dpm) of the sample is known as the counting efficiency. Because the

ash is continually settling, the first vial, in a series of vials with equal

quantities of ash and activity, would be expected to have a much

higher count rate than the last vial.

Problems with Quench

A problem inherent in the LSC process is referred to as quench. It results when a sample introduced into the scintillation fluid

reduces the amount of light reaching the PM tubes (FCnoU, 1979).

Different materials cause different amounts of quench. High

quenching materials absorb a large amount of the light scintillations,

resulting in an underestimation of the activity of the sample by the

LSC. The amount of quench is measured by the LSC and is called the

index of the external standard (SIE). The LSC measures the SIE by

placing a I37cs pellet near the vial before counting. The high energy

photons from the I37(;;s pellet interact with the scintillation fluid via

(29)

m

physically the same as beta particles and therefore cause the

emission of light scintillations through de-excitation of the fluor. The

scintillations are summed to produce an energy spectrum. The LSC

compares the energy spectrum of the scintillations produced with no

quenching present to the energy spectrum of the scintillations produced from the test sample. If quenching is present, the

spectrum will be shifted toward the lower energies because more

light is being absorbed by the sample and less light is reaching the

PMT. The higher the amount of quench, the more the spectrum will shift toward the lower energies (UTP, 1982).

Quench Curves and Standards

The LSC uses the SIE (quench) and the efficiency to produce a

quench curve. The efficiency was calculated by using eight vials,

each filled with 0.06iiCi (lOOuD of 35s standard solution. Increasing

amounts of ground ash from control burn 1, used as the quenching

agent, were added to 7 of the vials in increments of O.lg ranging

from O.lg to 0.7g. No ash was added to the remaining vial so that it

could be used as a reference point. With the activity of the 35s

samples manually programed into the machine, the LSC generated a

quench curve by plotting efficiency versus SIE (shown in Figure 2).

The LSC calculated the counting efficiency by dividing the count rate

(cpm) of each vial by the activity (dpm) of the 35$ added. The

quench curve was stored in the memory of the LSC as a specific

program. The activities of the ash samples were computed by using

(30)

Figure 2

Quench Curve for Ground Ash

>« u c

UJ

100

80

-

60-

40-

20-400

(31)

sample and located the corresponding counting efficiency on the quench curve. The count rate (cpm) measured by the LSC was

divided by the counting efficiency in order to calculate the actual

activity of the sample (dpm).

^ , ^ , ,, , Count Rate Measured by Detector (cpm) Activity of the Sample (dpm) =

-%

Absolute Efficiency of Detector

A separate quench curve was needed for measuring the extract fluid. The procedure used to calculate the activity of an unknown

extract fluid sample was basically the same as the procedure for the

unknown ash sample. However, the extract fluid was used as the quenching agent instead of the ash. The extract fluid's ability to quench, even in large quantities, was very poor. As a result Carbon Tetrachloride (CCI4), which is a good quenching agent, was used in its place. Eight vials, each containing 0.06 )iCi of 35s standard were prepared, and CCI4 was added to six of them. Amounts of CCI4

ranging from 25 to 150 |il in increments of 25 |il were added to these six vials. Figure 3 shows the resultant quench curve.

Ash Settling

Figure 4 shows that the apparent activity of the ground ash measured by the LSC decreases with time and Figure 5 shows that

the activity of the extract fluid remains constant over time indicating no settling of the fluid. In order to account for the settling of the ash in the vial before and during counting, the initial count rate at 0

(32)

>« o e

tit

Figure 3

Quench Curve for Extraction Fluid

and Carbon Tetrachloride

100

500 600 700

(33)

#

2000

Figure 4

Settling Pattern of Ash over Time Using Ash from Burn 3

1500

m

<

o

a

Q

1000-

500-100

Time Elapsed Since Mixing (In Minutes)

(34)

^

Figure 5

ͣͣ

2''

•IP*

I

K

lU

i

a.

a

Settling Pattern for Extraction Fluid

from Burn 3

<:uuu

-

15001000

-

50

0-«

•*• •. • * .*•••

—1---' 1---'---1---ͣ>---1

(35)

the best estimation of the actual activity of the ash. To estimate the

initial count rate, six scintillation vials, each containing 0.2g of ash

from the same burn, were shaken using a Vortex-Genie and loaded

into the LSC to be counted immediately. After the 6 vials were

counted they were again shaken, but this time there was a 5 minute

interval between mixing and the initiation of the counting process of

the LSC. This procedure was continued several more times with the

time interval between mixing and count initiation increasing each

time. The time interval between mixing and count initiation was

added to the amount of time elapsed from count initiation until the

LSC counted the vial (which is given on the LSC printout). By using

this method, a count rate was known for each 2-3 minute time

interval from 0 minutes up to 100 minutes. Normally, a total of 30

-40 data points were obtained.

The plots of count rate versus time made from this data

showed the trends in the count rate of the ash and extract fluid over

time. It was thought if the settling trends were similar between ash

samples, then a general equation could be used to describe the

trends. And if the settling trends could be described with an equation, then the count rate could be "corrected", using the

equation, for the effects of time on the count rate. Therefore, the

count rate for all the vials would appear as if they were measured at

0 minutes. This procedure was used for ash samples from burns 2, 3,

5, 6, 11, 12, 17 & 18 in order to try to model the general settling

patterns of the ash to be used for all the ash samples.

A Cricket Graph graphics program (Cricket Software, Malvern,

(36)

curve, for each ash sample, was different. Therefore, a general

equation would not adequately correct for the effects of time in all ash samples. It was subjectively determined through trial and error that the data from the burns were best fit with a logarithmic curve.

Exponential, polynomial, and simple curves were also tried, using the graphics program. Once the settling curve was generated by the

program, it was extrapolated to 0 minutes to find the initial count rate immediately after mixing. The activity of the ash was calculated

from this count rate for each vial. This method of accounting for the

settling-out of the ash was applied to all 18 burns.

For burns 1, 4, 7, 8, 9, 10, 13, 14, 15. & 16, ten vials, each

containing 0.2g of ash from each burn, were counted resulting in 10

data points. The more in depth settling curves, previously

mentioned, were not empirically determined for these ash samples

because of the length of time involved in obtaining the extra data

points. The 10 data points obtained for each ash sample were fit

with a logarithmic curve using the Cricket Graph program. The

activity of each ash sample was taken to be the activity extrapolated

to time 0 on the curve.

Measurement of Rough Ash with the Liquid Scintillation Counter

A small spatula was used to transfer ash from a 100 ml beaker

to the scintillation vials. To remain consistent with the procedure at

NIEHS, an attempt was made to sample only the smallest particles of

ash. The larger particles did not mix well and self-absorption within

the particle would cause an underestimation of the activity. Because

(37)

beaker, samples were taken from that area. In each vial 0.2g of ash were placed along with approximately 7 ml of scintillation fluid.

Measurement of Ground Ash with the Liquid Scintillation Counter

Five 7 ml polyethylene scintillation vials were filled with ash in increments of O.lg starting with O.lg and finishing with 0.5g. A

Mettler balance (Model P-1200, E.H. Sergant and Co.) was used for

weighing the samples. Figure 6 shows that the quench increases (SIE

decreases) as the quantity of ash increases, and that 0.2g of ash

resulted in the highest count rate. A higher count rate (cpm) would

require less extrapolation for estimation of the sample activity (dpm)

by the LSC. Similarly, a low amount of quench would also require less extrapolation by the LSC for determination of the sample

activity. Therefore, an ash sample weight was desired that had a

high count rate but not a large amount of quench relative to the

other ash weights. Based on these two stipulations, a subjective decision was made to use 0.2g of ash as the sample weight of ash to

be measured in the LSC.

Measurement of Extract Fluid with the Liquid Scintillation Counter To measure the activity of the samples of extract fluid from each burn, ten 7 ml vials were each filled with 1 ml extract fluid.

The remaining portion of the 7 ml vials were filled with liquid

(38)

%

Q.

o

Figure 6

SIE and CPM for Several Ash Quantities

2000 -I---r 400

1900

18001700

1600

-1500

1400

300

UJ

200

100

-Q--- cpm

ͣ

^--- SIE

Grams of Ash

(39)

#

Windowless Gas Flow Proportional Counter

Theory of Operation

In an effort to validate the data collected with the LSC, an

alternate beta counting technique was used to measure the activity

in the samples. A 2tt windowless gas flow proportional counter (PC)

was used for this alternative counting technique.

The PC is well suited for measuring the low energy betas from

35s. Since the sample is placed in the active volume of the detector,

problems with beta particles being stopped or losing energy from window absorption are eliminated. This allows maximum efficiency to be achieved. However, the efficiency of the PC is limited to about 50% by its 2tt geometry configuration (Knoll, 1979).

Beta particles emitted from a source result in the formation of

ion pairs through ionization in the active volume of the detector.

Since each ion is charged, it will be attracted, by coulombic forces, to

any particle with an opposite charge and repelled from a particle with a similar charge. Voltage is applied to the anode causing it to

become positively charged. The walls of the detector, therefore,

become negatively charged. The negatively charged electrons are

swept toward the anode, and the positively charged ions are swept toward the wall of the detector. As the electrons are moving toward the anode, they collide with the neutral gas molecules causing more ion pairs to be formed. These secondary collisions spread in a

cascading fashion toward the anode until all the electrons are

collected, resulting in a pulse. A high voltage is needed to increase

(40)

Figure 7

Plateau Curve for 35S Ash Standard

0.

o

1400-

/

ͣ

^^ /

1200 -

/^^

1000800

-y*

600-/

y

400

-4

/

200-0- ---1---11---•--—r— -1----1 • 1

1600 1800 2000 2200 2400 2600

(41)

The operating voltage for this study was set near the middle of the

plateau at 2350 volts.

35s Extract Fluid Standard for the Proportional Counter

A 35$ standard solution was made by pipetting 0.07 iiCi of 35s

standard into a 20 ml volumetric flask. A 20 ml aliquot of distilled

water measured in a 25 ml graduated cylinder was added to the

flask making the concentration 0.0035 |iCi/ml. After sealing the

flask with a glass stopper, the contents were mixed by inverting the

flask several times. Five 0.5 ml aliquot were counted with a LSC.

The activity of the 35s standard solution was calculated to be 7568

dpm/ml. This activity was checked by counting five 0.5 ml aliquots,

from the flask, using the LSC. The mean activity was 7579 dpm/ml

(7492. 7372, 7594, 7890 and 7548 dpm/ml). Using a 0.5 ml pipette,

an aliquot of the standard solution was pipetted onto each aluminum

planchette (Nuclear Chicago. Inc.) in increments of 0.5 ml starting

with 0.5 ml and ending with 2.5 ml. The planchettes were heated on

a hot plate until the fluid had evaporated. Each planchette was

counted for 10 minutes using the PC. Figure 8 shows the linear

relationship between the ml of standard added and the activity of

the standard measured. This means the quantity of fluid counted

doesn't affect the count rate of the sample. In order to test for the

possibility of 35s being lost through evaporation, a piece of 4.25 cm

Whatman filter paper (Whatman International Ltd., Maidstone.

(42)

Figure 8

Relationship between the Activity Measured

and the Amount of S35 Standard Added

20000

15000

i

10000-5000

(43)

paper was placed in a scintillation vial and counted with a LSC. No activity above background was detected.

Measurement of Extract Fluid with the Proportional Counter Extraction fluid samples were pipetted onto aluminum

planchettes. A 0.5 ml pipette was used to transfer a 2 ml aliquot of

extraction fluid to each planchette. The planchettes were heated on a

hot plate to decrease evaporation time. After complete evaporation

of the extract fluid, each planchette was counted for 15 minutes.

Appendix E shows the method used for calculating the activity per

gram of ash from the counts per ml of extract fluid.

Ash Sample Size for Proportional Counter

In order to make a decision on the amount of ash that was to

be put in the planchettes for counting, two main problems were

confronted:

1- Self absorption of low energy betas by the ash.

2- Limited accuracy in measuring small amounts of ash.

The problem with self absorption of low energy beta particles can be

substantial. In order to measure the amount of self-absorption for

the several ash weights, the thickness of the ash at each weight was

needed. The first step in determining the thickness of the ash at a

specified weight was to calculate the density of the ash. To do this, a 25 ml graduated cylinder was filled to the 25 ml mark with ash and weighed. The weight of the empty graduated cylinder was

(44)

f

#

#

ash to determine the weight of the ash. The following method was

used to calculate the thickness of the ash:

Density

grams of ash _ 9

25 ml of ash ml

,, , weight of ash (g)

Volume =--- = ml density of ash (g/ml)

volume (ml)

Height = --- = thickness of the ash in cm

n r^

grams of ash thickness of ash (cm)

0.1 0.003

0.2 0.006

0.3 0.009

0.4 0.012

0.5 0.015

The average ash density calculated in this study was 1.1 g/ml.

The average range of beta particles emitted from 35s in ash of this

density is 0.0033 cm (USDHEW, 1970). Shown in the table above are

the five ash weights tested and their calculated thicknesses. As the

ash weight increases, the amount of self-absorption also increases

because the ash sample is thicker and absorbs more betas before

they reach the detector. The lowest weight was set at 0.1 g because

(45)

f

#

#

highest weight was set at 0.5g because at this amount of ash the

count rate becomes very low, indicating much self-absorption of the

betas (Figure 9) which results in low detector efficiency (Figure 10).

A final decision to use O.lg of ash as the sample size was made

because of the high detection efficiency at this weight resulting from

the least amount of sample self-absorption (see Figure 10). It was

thought that the possible errors in measuring O.lg of ash would

cause less error in the calculation of activity than the low efficiencies

associated with measuring the higher quantities of ash. Figures 9

and 10 also show the count rate and efficiency of the standard

solution (O.Og of ash). This represents the maximum efficiency and

count rate possible because there is no self-absorption from the ash.

The efficiency at O.Og also indicates the efficiency for the extract

fluid.

For sampling counting a 0.5 ml aliquot of distilled water was

added to each ash sample in the planchette to facilitate the spreading

of the ash into a smooth layer on the planchette. The ash was spread

evenly over the planchette using a small spatula. Because of possible

beta absorption by the water, the samples were gently heated on a

hot plate for 15 minutes until the water was evaporated.

Measurement of Ash with the Proportional Counter

Ground ash samples from each burn were measured into 3

planchettes in aliquots of O.lg. A 0.5 ml pipette was used to add 0.5

ml of distilled water on each planchette. This was done to ensure

that the ash was in a uniform layer in the planchette. The

(46)

9

Figure 9

Count Rate vs. Ash Weight for Ash

with Standard Added

2000

CL

a

1800-

1600-

1400-

1200-

1000-

800-600 ---1---1---1---1---1---1---1---1---1 I---r—

0.0 0.1 0.2 0.3 0.4 0.5 0.6

Grams of Ash

(47)

#

u c

3

o

Figure 10

Absolute Efficiency vs. Weight for Ash

with Standard Added 0.5

0.4-0.3

0.2

0.1

3784 dpm of standard was added to each planchet

—'---1---"---1---'---1---'---1—

0.0 0.1 0.2 0.3 0.4

Grams of Ash

—I---'—

0.5 0.6

(48)

f

completely evaporated and then counted for 15 minutes.

Appendix D contains the equations used for calculating the sample

activity and accounting for decay.

#

(49)

f

#

#

RESULTS

Statistical Analysis of the Data

Statistical procedures were performed on the data using Systat

and Mystat statistical programs (Systat Inc., Evanston, ID for the

personal computer. The mean, standard deviation, Students t-test

statistic, and analysis of variance were calculated for the 35s ash

retention efficiencies.

A Students t-test was used to compare the percentage of

radioactivity extracted from ash with glass to ash without glass

during the extraction procedure. The Students t-test compares the

means of two sample populations. Based on the t statistic a decision

can be made, depending on the confidence level, on the significance

of differences in the means.

An analysis of variance (ANOVA) was made on four groups of

data to test for significant variation between the group means

compared to among the groups (Remington and Schork, 1985). In

Table 1, the results of the analysis of variance are listed for the four

groups of data which are as follows:

-glass vs. no glass

-sulfate vs. methionine

-proportional counter vs. liquid scintillation counter

-ash vs. extract fluid

Problems Encountered in Rough Ash Measurement

There are several sources of bias associated with measuring

(50)

Table 1

Analysis of Variance

Variance within Sample Groups

Source p-value

Type of Vial 0.012

Chemical Form <0.001 Sampling Method 0.444

Type of Detector 0.669

Variance between Sample Groups

Source p-value

Type of Vial vs.

Samping Method 0.917

Type of Vial vs.

Chemical Form 0.018

Type of Vial vs.

Type of Detector 0.803

Chemical Form vs.

Sampling Method 0.511

Chemical Form vs.

Type of Detector 0.776

Sampling Method vs.

(51)

#

ash is supposed to sample only the smallest ash particles. This

sample would not be representative of the entire sample because not

all of the ash would be sampled. Second, the ash particle size

between samples may vary greatly between samples depending on

the person measuring the ash samples. Samples with different

particle sizes have different settling rates and would therefore have

different count rates. Third, most of the particles sampled are large

enough that they settle out to the bottom of the vial immediately

after mixing. As a result, the count rates for rough ash samples are

consistently much lower than count rates of ground ash. The

calculated dpm for ground ash are frequently more than twice those

of the rough ash from the same burn (see Figure 11). It was

therefore decided to persue no further study of the rough ash and to

eliminate its use in determination of retention efficiency.

Mass Reduction Efficiencv of the Incinerator

The mass reduction indicates the extent to which the wastes

incinerated are combusted. The mass reduction efficiency (MRE) for

the wastes for the 18 burns was calculated using the following

equation:

Mass of ash after incineration

|v|p^ = I _ --- X 100

Mass of wastes incinerated

In Figure 12, the mass reduction efficiencies are shown for each

(52)

Figure 11

Comparison of Rough Ash and Ground Ash

Using Ash from Burn 17

16000 -I---z---r 16000

-14000

-12000

-10000

-8000

-6000

-4000

14000

12000

10000

2000 0 2 4 6 8 10 12

Time Elapsed Since Mixing (in Minutes)

(53)

Figure 12

Mass Reduction of Incinerated Wastes

c

o

u 3

•D

«>

CC

contains glass

92.5-90.0

12 3 4 5 6 7 8 9 10 1112 13 14 15 16 17 18

(54)

#

#

glass was incinerated consistently had lower MRE's with a mean of

92.4% compared to 96.5% for the 9 burns containing no glass vials

(see Table 2). An ANOVA table computed on the MRE data showed

that this difference is statistically significant (p<0.0001).

Since polyethylene is organic, it is combusted and released as a

gaseous effluent. The glass is not an organic and is not combusted,

and since temperatures in the incinerator never reach the boiling

point of glass, it would be expected to remain in the residual ash.

The extra weight from the glass is probably responsible for the lower

MRE's for the burns with glass.

Calculation of Retention Efficiencv

The quantity of 35$ remaining in the ash after a burn was

measured by the retention efficiency. The retention efficiency was

calculated by dividing the activity in the ash after each burn by the

activity in the wastes before each burn and multiplying by a factor

of 100. The complete calculation of the retention efficiency is given

in Appendix F.

Table 3 gives the retention efficiencies for the 18 burns. The

control burns have a retention efficiency of 0 because no activity was

added to the wastes that were incinerated. The retention efficiency

for burn 18 seems to be much higher than for any other burn. This

may have resulted from the 35s in the ash being more concentrated

in the ash that was sampled compared to the rest of the ash. The 35s

may not have been distributed evenly through out the residual ash,

and the ash samples collected from burn 18 may have been from the

(55)

#

Table 2

Mass Reduction of Incinerated Wastes

#

Burn #

kg of wastes

incinerated

kg of residual

ash

Mass reduction efficiency (%)

1 * 69.3 5.20

92.5 1

2- 71.1 3.64 97.7

3* 72.9 5.10 93.0

4- 72.5 2.84 96.1

5- 72.0 2.57 96.5

6* 72.9 5.36 92.7

7* 71.6 6.30 91.3

8* 67.5 5.80 91.5

9- 73.0 2.72 96.3

10- 68.0 2.23 96.8

11- 71.5 2.70 96.2

12* 73.8 4.80 93.5

13* 70.6 4.90 93.1

14* 68.4 5.71 91.7

15- 68.4 2.04 97.0

16- 68.0 2.99 95.6

17- 68.8 2.26 96.8

1 18* 68.8 5.44

92.1 1

contains glass

(56)

Table 3

Retention Efficiencies for Ash and Extract Fluid

Proportional Counter Liquid Scintillation Counter

Burn #

ancJ Code Ash% Extract fluid

Burn #

and Code Ash% Extract fluid %

1 CG 0.00 0.00 1 CG 0.00 0.00

2 M 1.90 1.63 2 M 1.16 2.05

3 MG 2.25 1.14 3 MG 2.70 1.84

4 C 0.00 0.00 4 C 0.00 0.00 Codes

5 S 4.92 2.77 5 S 3.22 5.78

6 SG 9.32 5.01 6 SG 8.16 11.66 C: Control Burn

7 CG 0.00 0.00 7 CG 0.00 0.00 M: Methionine

8 iVIG 2.18 1.88 8 MG 2.10 2.99 S: Sulfate

9 M 2.01 2.18 9 M 1.54 2.90 G: Glass

10 C 0.00 0.00 10 C 0.00 0.00

11 S 7.89 6.32 11 S 3.02 11.95

12 SG 11.14 6.38 12 SG 5.27 12.20

13 CG 0.00 0.00 13 CG 0.00 0.00

14 MG 0.50 0.44 14 MG 0.18 1.25

15 M 0.28 0.28 15 M 0.21 0.27

16 C 0.00 0.00 16 C 0.00 0.00

17 S 8.76 5.47 17 S 4.60 11.04

18 SG 29.64 21.70 18 SG 16.33 40.10

45^

(57)

w^ere burned on days w^hen test burns Tvere not being performed,

residual 35s, I'^C or 3h from the NIEHS wastes may have been in the

ash in large amounts. If this was the case, inadequate cleaning before burn 18 may not have removed all the ash from the refractory and it may have combined with the ash from burn 18 resulting in a seemingly much higher retention efficiency.

Effects of Chemical Form on Retention Efficiencv

The 6 burns with 35s sulfate yielded a higher mean retention efficiency than the mean of the 6 burns with 35s methionine. As

listed in Table 1, the p value of 0.00001 indicates that there is a statistically significant difference in the retention efficiencies between the two chemical forms of 35s. In Table 4 the retention

efficiencies are broken down by chemical form, detector type,

sampling method and the presence of glass. The sulfate form of 35s

has a much higher retention efficiency in all categories. Table 5 also

shows the mean retention efficiencies. The overall mean retention

efficiency was 10.54% for sulfate and 1.51% for methionine.

Included in the calculation of these efficiencies were the efficiencies

calculated with both detectors and efficiencies determined by measuring ground ash and extract fluid. A total of four mean

retention efficiency values (Appendix G) were averaged together to

produce the overall mean retention efficiency for each chemical form

of 35s. If burn 18 were eliminated, the mean retention efficiency for

the 4 remaining burns containing 35s sulfate would be 7.35%

(58)

#

0

Table 4

Effects of Chemical Form and Glass

on Retention Efficiencies

Ash Extract Fluid

£Q J^Q E£ LSQ

Sulfate

Glass 16.70 9.95 11.03 21.3(

No glass 7.19 3.61 4.80 9.62

Methionine

Glass 1.65 1.64 1.15 2.12

(59)

#

Table 5

Effects of Chemical Form on Retention Efficiencies

Ash Extract Fluid

Sulfate 11.95 6.78 7.94 15.50

Methionine 1.52 1.30 1.26 1.94

Mean retention efficiency using sulfate: 10.54%

Mean retention efficiency using methionine: 1.51%

(60)

#

During incineration, sulfate decomposes to SO2, and the

methionine decomposes to SO2 and NOj which are eliminated as gases

out the stack. Methionine, being an organic compound, could be

combusted to form H2O and CO2 along with SO2 and NOj and vented

up the stack. Sulfate is an inorganic compound and would not be

combusted to H2O and CO2, which may explain why more of the

sulfate is retained in the ash. The difference between the retention

efficiencies of sulfate and methionine may be due to the extent to

which they are converted to these gases and other combustion

products. Even though there are significant differences between the

mean retention efficiencies of the sulfate and methionine from of 35s,

it should be noted that the standard deviations are very large.

Effects of Glass on Retention Efficiencv

The ash and extract fluid from the burns with glass have

consistently higher retention efficiencies than the burns without

glass. Statistical data on the retention efficiencies are listed in

Table 6. The mean retention efficiency for ash with glass is 8.22%

and for ash without glass is 3.84%. There was a statistically

significant difference (p=0.012) between retention efficiencies from

burns with glass vials and burns without glass vials. Table 4 shows

the differences between burns with glass and without glass for both

chemical forms of 35s. An analysis of variance (Table 1) resulted in a

p value of 0.018, indicating that the presence of glass and the

chemical form of 35s have a statistically significant affect upon each

(61)

#

Table 6

Effects of Glass on Retention Efficiencies

Ash Extract Fluid

EQ LSQ PC LSC

Glass 9.23 5.80 6.09 11.75

Without Glass 4.29 2.28 3.11 5.69

Mean retention efficiency for burns with glass: 8.22%

(62)

«b

Comparison of Sampling Methods

The measurement of ash and extract fluid with the PC consistently resulted in higher retention efficiencies for the ash samples (Figure 13). In contrast, the measurement of ash and extract fluid with the LSC resulted in consistently higher retention efficiencies for the extract fluid samples (Figure 14). The mean retention efficiency using both detectors for ash was 5.39% and the

extract fluid was 6.66%. The analysis of variance in Table 1 shows

that there is no significant difference between the two sampling

methods. The reason that no statistically significant difference was

found between retention in ash and extract fluid may be due to the

large standard deviations (Appendix G) and may also be caused by

using the mean retention efficiency of both detectors combined. The

data in Table 1 indicate that the sampling method and the type of

detector have a statistically significant effect (p=0.046) upon each

other. This is probably because both sampling methods were not

averaged together to calculate the mean for each detector as was

previously done.

Determination of Extraction Procedure Efficiencv

In order to compare the activity of the extract fluid (dpm/ml)

to the activity of the ash (dpm/g) the activity of the extract fluid was

converted into units of dpm/g of ash (see Appendix E).

One extraction was performed on each of the 18 burns to

determine the activity of the ash from that burn (see Appendix H).

(63)

#

u e UJ c o c 4> 45 40- 35- 30-25 20- 15- 10- 5-Figure 13

Retention Efficiency Measured by

Proportional Counter

5

ͣ

" Li "

ͣ

^.1.. MM. HI

M

I - l^^S^^n—"T"

12 3 4 5 6 7 8 9 10 1112 13 14 15 16 17 18

GG GGG GGG G

Burn *

Ash

Extract Fluid

M. Methionine

5: Sulfate

(64)

#

ui c o

on

Figure 14

Retention Efficiency Measured by

Liquid Scintillation Counter

B^

1 23456789 10

G G G G G

Burn ͣ

12 13 14 15 16 17 18

G G G G

ͣ

Ash

B Extract Fluid

M: Methionine

S: Sulfate

(65)

determined. This was done by performing consecutive extractions on

the same ash sample. Ash from burns 12, 13, 17 and 18 was used

for these additional extractions.

The extraction procedure, listed in Appendix A, was used for

each extraction. After each 24 hour mixing period, the ash was

allowed to settle for about 20 minutes. The extract fluid was

siphoned off the top by applying suction through a narrow glass

tube. The leftover ash slurry was filtered through Watman paper so

that only the ash remained. This ash was reused for the next

extraction. Samples of the extract fluid and ash were collected and

counted after each extraction. Only Ig of ash was collected after each

extraction so not to greatly affect the remaining extractions.

The column titled "net activity of ash" in Table 7 shows the

activity removed during several extractions on the same ash sample.

The initial activity of the ash is listed as extraction #0 for each burn.

The activity of the ash after each extraction is also given. It appears

in burns 11 and 12 that there is approximately the same amount of

activity remaining in the ash after each extraction. Therefore, the

mean activity measured in the ash, after each extraction, represents

the amount of activity that could not be removed from the ash by

the extractions. The mean background activity was calculated from

four extractions on ash from control burn 10. All activities were

determined using the liquid scintillation counter.

To calculate the percent of the total activity expected to be in

the extraction fluid, the activity of the ash initially and the average

activity of the ash after each of the extractions was determined using

Figure

Figure 4 shows that the apparent activity of the ground ash measured by the LSC decreases with time and Figure 5 shows that the activity of the extract fluid remains constant over time indicating no settling of the fluid
Table 3 gives the retention efficiencies for the 18 burns. The control burns have a retention efficiency of 0 because no activity was added to the wastes that were incinerated

References

Related documents

The semi-Markov process and regenerative point technique is adopted to derive expressions for various measures of system effectiveness such as mean time to system failure,

The results were sensitive to: the remission rates maintained in responders to 'usual care' and to 5 mg/kg infliximab, the rate of remission induced by adalimumab in non-responders to

The Twin-arginine translocation (Tat) pathway of Escherichia coli has great potential for the export of biopharmaceuticals to the periplasm due to its ability to

Samples were then freeze dried, ground to homogeneity and a small size fraction utilizing a blender, and then packed densely into 24-mL liquid scintillation vials for gamma

Come learn what it would be like to attend different types of institutions: your local community college, a two-year residential college, a four-year in-state public school,

In order to isolate the forage grain fraction (which is then subject to flattening) from the grain that is arriving after threshing by the combine harvesters on

In addition, the findings further reveal that none of the studies investigate the mediating effect of the nature of innovation (incremental and radical) on the relationship