• No results found

Serum Creatinine Measurement Specificity

N/A
N/A
Protected

Academic year: 2022

Share "Serum Creatinine Measurement Specificity"

Copied!
16
0
0

Loading.... (view fulltext now)

Full text

(1)

Serum Creatinine

Measurement Specificity

NKDEP Lab Working Group

Neil Greenberg, PhD

July, 2007, San Diego, CA

(2)

Problem Summary

From: Myers, et al, Recommendations for Improving Serum Creatinine Measurement, Clinical Chemistry 52:1, 5–18 (2006)

• Endogenous and exogenous interfering substances

contribute to lack of analytical specificity of Jaffe (alkaline picrate) creatinine methods

• Interfering substances in serum and/or plasma, particularly proteins, can lead to overestimation of 15%–25% with

various Jaffe methods

• Interference from glucose and ketoacids particularly

important in diabetics who are at high-risk for CKD.

(3)

Problem Summary

From: Myers, et al, Recommendations for Improving Serum Creatinine Measurement, Clinical Chemistry 52:1, 5–18 (2006)

• Most Jaffe interference studies dated 10-30 years ago

• Several modifications to Jaffe method reported to improve specificity (e.g. use of kinetic assays or early read times)

• An offset of 21 µmol/L (0.234 mg/dL) was needed to further correct “compensated” Jaffe method for non-creatinine Jaffe- reacting chromogens (Junge W, et al. Clin Chim Acta 2004;

344:137-48)

¾ …even with low imprecision and assay standardized to IDMS reference measurement procedure

¾ If analytical non-specificity bias remains, errors in estimating

GFR will occur.

(4)

Problem Summary (cont)

From: Myers, et al, Recommendations for Improving Serum Creatinine Measurement, Clinical Chemistry 52:1, 5–18 (2006)

• Although enzymatic creatinine methods are reported to have fewer interferences than Jaffe methods, reports are published of various interferents with enzymatic methods

• HPLC methods have greater analytical specificity than conventional methods.

¾Sample de-proteinization combined with selectivity of mobile-phase conditions make it unlikely that many substances will interfere

• GC-IDMS is method of choice for establishing true

concentration of creatinine in serum due to excellent

specificity and relative SD (0.3%)

(5)

Creatinine Analytical Performance Goals

From: Myers, et al, Recommendations for Improving Serum Creatinine Measurement, Clinical Chemistry 52:1, 5–18 (2006)

CVi CVg Goal Level CVa Goal Basis CVa Goal

Bias Goal Basis Bias Goal

TE Goal*

4.30% 12.90

%

Minimum

Acceptable (0.75 CVi) 3.20% 0.375 (CVi2_ CVg2)1/2 5.10% 11.40%

Desirable

(0.5 CVi) 2.20% 0.25 (CVi2_ CVg2)1/2 3.40% 7.60%

Optimum

(0.25 CVi) 1.10% 0.125 (CVi2_ CVg2)1/2 1.70% 3.80%

Performance Goals Based on Biological Variability, creat = 1.0 – 1.5 mg/dL

Issue:

No goal/limit has been defined for sample-dependent random bias (specificity) performance.

* TE (total error) goal calculation: Bias Goal + (1.96 X CVa goal)

(6)

Creatinine Analytical Performance Goals - Specificity Problem/Issue:

For serum/plasma creatinine measurements, especially in context of using serum creatinine values for determination of GFR with estimating equations, what are appropriate acceptance criteria for individual sample random biases due to

interferents and other sample-dependent factors?

(7)

(Enzymatic) CREAT BIAS vs HPLC

-0.50 -0.40 -0.30 -0.20 -0.10 0.00 0.10 0.20 0.30 0.40

0.00 0.50 1.00 1.50 2.00 2.50

HPLC CREAT (mg/dL)

Series1

2002 OCD Internal Study, Rate Jaffe and (enzymatic) method compared to HPLC, 410 random patient samples, creatinine range = 0.4 – 2.0 mg/dL

(8)

Example Rate JAFFE CREAT Bias vs. HPLC

-0 . 3 0 -0 . 2 0 -0 . 10 0 . 0 0 0 . 10 0 . 2 0 0 . 3 0 0 . 4 0 0 . 5 0 0 . 6 0

0 . 0 0 0 . 5 0 1. 0 0 1. 5 0 2 . 0 0 2 . 5 0

HPLC CREAT (mg/dL)

Series1

2002 OCD Internal Study, Rate Jaffe and (enzymatic) compared to HPLC, 410 random patient samples, creatinine range = 0.4 – 2.0 mg/dL

(9)

0.6 0.4

0.2 0.0

-0.2 -0.4

350 300 250 200 150 100 50 0

C4

Frequency

Mean Bias = 0.07 mg/dL Bias SD = 0.07 mg/dL

HPLC creat range = 0.4 – 2.0 mg/dL

creatinine bias (mg/dL)

0

(enzymatic) Creatinine method

Distribution of Biases (mg/dL) vs. HPLC

(10)

350 300 250 200 150 100 50 0

Frequency

Example Rate Jaffe Creatinine Method

Distribution of biases (mg/dL) vs HPLC

Mean Bias = 0.15 mg/dL Bias SD = 0.10 mg/dL

HPLC creat range = 0.4 – 2.0 mg/dL

0

(11)

Creatinine Analytical Performance Goals - Specificity Statistical Model for Total Error including Random Bias

Lawton WH, Sylvester EA, Young-Ferraro BJ. Statistical comparison of multiple analytic procedures: application to clinical chemistry.

Technometrics 21:397-409 (1979)

%TE (total error) goal =

%FB Goal + 1.96 X SQRT[%CV

A2

goal + %CV

RB2

goal]

Where…

%FB = allowable (systematic) calibration %bias

%CV

A

= allowable analytical imprecision, CV%

%CV

RB

= allowable random bias (non specificity), CV%

(12)

Creatinine Analytical Performance Goals - Specificity

Rationale for Proposed Goal – Random Bias (non-specificity)

Following the statistical model defined by Lawton, et al…

%TE = %FB + 1.96 X SQRT[%CVA2 + %CVRB2] Given that (from biological variability), %CVB-G = 12.9%

(where %CVB-G = % biological variation among individuals), allow that

%TE can be as large as 12.9%.

Example 1: Where allowable TE= 12.9%, %FB = 5% and %CVA =3%…

• 12.9% = 5% + 1.96 X SQRT[(3%)2 + %CVRB2]

• then allowable %CV = 2.7%

(13)

Creatinine Analytical Performance Goals - Specificity

Rationale for Proposed Goal – Random Bias (non-specificity)

Following the statistical model defined by Lawton, et al…

%TE = %FB + 1.96 X SQRT[%CVA2 + %CVRB2]

Example 2: Where allowable TE= 12.9%, %FB = 1% and %CVA =3%…

• 12.9% = 1% + 1.96 X SQRT[(3%)2 + %CVRB2]

• then allowable %CVRB = 5.3

Example 3: Where allowable TE= 11%, %FB = 1% and %CVA =3%…

• 11% = 1% + 1.96 X SQRT[(3%)2 + %CVRB2]

• then allowable %CVRB = 4.1%

(14)

Creatinine Analytical Performance Goals - Specificity

Proposed rationale and approach for

development of acceptance criteria – specificity

• If systematic bias can be controlled to within ±1% from IDMS reference target, allowable random bias (%CVRB) can be ~± 5% without

compromising “minimally acceptable” TE% goals

• Defined criteria should be stated as…

¾ When a creatinine method is calibrated to be traceable to the IDMS reference method, for the indication of screening patients for CKD (using MDRD or other estimating equations), random sample-

dependent bias should be within ± 2*X% for 95% of samples tested, where X = %CVRB Goal.

¾ Possible Criteria:

ƒ 5% “minimally acceptable”;

ƒ 3% “desirable”

(15)

0.6 0.4

0.2 0.0

-0.2 -0.4

350 300 250 200 150 100 50 0

C4

Frequency

Enzymatic creatinine method

Distribution of Biases (mg/dL) vs. HPLC

Adj. Mean Bias = 0 mg/dL Bias SD = 0.07 mg/dL

HPLC creat range = 0.4 – 2.0 mg/dL

0

Mean value = 1.1 mg/dL with bias compensation; Proposed Allowable Random Bias = 5%

+5%

-5%

Failure Rate (bias >10%) = ~5%

(16)

350 300 250 200 150 100 50 0

Frequency

Example Rate Jaffe Creatinine Method

Distribution of biases (mg/dL) vs HPLC

Adj. Mean Bias = 0 mg/dL Bias SD = 0.10 mg/dL

HPLC creat range = 0.4 – 2.0 mg/dL

0

Mean value = 1.1 mg/dL with bias compensation; Proposed Allowable Random Bias = 5%

-5% +5%

Failure Rate (bias >10%) = ~18%

References

Related documents

With Analysis of Variance, the empirical results, Table 2, presented the notable effects of age on the correlations between Learning Activities, Religious Activities, Leisure

The feasibility of two types of nanoporous silica materials, including mesoporous silica (MCM-41 and SBA-15 with 2D hexagonal pores) and 3D macroporous silica (3DOM silica), as

51 Figure 4-2: Tailored welded patch for indirect hot-stamping schematics: side flange weld failure (red) of a U-channel due to localized weld shear before final hot

Scientific advice and participation in the international special exhibition at the Federal State Museum for Prehistory Halle/Saale “Krieg – eine archäologische

Rats were also fed with 5000mg/kg extract The toxicity in the animals was carried out by assessing the effects on biochemical parameters, body weight and histopathological study

For the reconstruction of muon tracks with the Micromegas modules the signals on the copper readout strips have to be recorded and evaluated.. The following chapter describes

Although 3-week treatment of NB-360 caused a significant but reversible reduction of density of total dendritic spines, persistent spines (persisting ⩾ 7 days) and