CHAPTER 2: LIQUID VISCOSITY MEASUREMENT

Objective

Calculate viscosity (dynamic or absolute, and kinematic) and determine how this property varies with changes in temperature for a constant-composition multi-component hydrocarbon mixture.

Introduction

When shear stress is applied to a fluid it will suffer a continuous deformation, which is referred to as “flow”. Fluid viscosity is the property, which indicates the fluid’s resistance to flow due to an unbalanced shearing force (see Figure 2-1). Fluids characterized as Newtonian fluids exhibit constant values for viscosity with variations in the shear rate, everything else being constant. As a first approximation, the liquids found in petroleum reservoirs (hydrocarbons and water) are considered to be Newtonian. Newton’s Law of viscosity can be written as follows:

         y v   ...(2-1)

Where  is the shear stress applied to the fluid,  is the absolute viscosity of a Newtonian liquid, and

v/y

is the velocity gradient (flow or deformation).

The variation in viscosity of liquid petroleum is due to the variation in composition, which results from reductions in pressure and the resultant liberation of gases or vapors. A study of such variation is beyond the scope of this laboratory. However, changes in temperature are another important source of variation of viscosity for constant composition hydrocarbon mixtures. This is the objective of this study. Thus, the purpose of this laboratory exercise is to determine how the viscosity of a multi-component hydrocarbon varies with temperature.

(2)

Figure 2-1. Flow Between Two Plates (Modified from Bear, 1972)

There is a multitude of methods for measuring viscosity. Each method often has its own relative viscosity scale, which is good only for that particular method. For instance, two such units are Saybolt (SSU) or the Redwood scale. Viscosity can be absolute (dynamic) or kinematic. Absolute viscosity is commonly used for equations describing the flow of fluids in porous media. It is related to the kinematic viscosity through density:

 

  * ...(2-1)

Where  is kinematic viscosity of liquid and  is density of the liquid. Common units for absolute viscosity are illustrated in Table 2-1.

Table 2-1. Units of Dynamic (Absolute) Viscosity

Units System S.I. c.g.s Field

Units N*s/m2 Dyne*s/cm2 = Poise cP = Poise/100

The kinematic viscosity is usually given in centi-Stokes (cSt). Multiplying cSt by density in g/cc, the dynamic (absolute) viscosity is obtained in cP.

(3)

Figure 2-2 presents a Cannon-Fenske viscometer. This instrument is used to measure kinematic viscosity, which will be used (and density measurements) to compute dynamic (absolute) viscosity.

Figure 2-2. Cannon-Fenske Viscometer

The instructions for using this apparatus are (Cannon-Fenske Viscometer Instrument Manual):

1. The viscometer should be cleaned with a suitable solvent and dried in a stream of clean, filtered N2.

2. The instrument should be periodically cleaned with chromic acid to remove any possible traces of organic deposits.

3. If a possibility of lint, dust, or other solid material is present in the liquid sample, this may be removed by filtering through sintered glass filter or fine mesh screen

4. To introduce sample into viscometer, invert viscometer, immerse tube “A” into liquid and apply suction to “I”, which causes the sample to rise to etched line “E”. Turn the viscometer to normal position and wipe tube “A” clean.

5. Insert the viscometer into a holder and place in constant temperature bath. Allow some minutes for viscometer to reach equilibrium at desired temperature (10 min for 100oF; the

(4)

7. Apply suction to tube “A” and bring sample into bulb “B” a short distance above mark “C”.

8. The efflux time is measured by allowing the sample to flow freely through mark “C”, measuring the time for the meniscus to pass from “C” to “E”.

9. To repeat efflux time measurement, repeat steps 7 and 8.

10. The kinematic viscosity is calculated by multiplying the efflux time by the viscometer constant.

Every Cannon-Fenske tube has its own calibrated constant, which is multiplied by the recorded time in second to obtain kinematic viscosity in cSt, that is:

t k *

where k is the viscometer constant in cSt/seconds and t is time in seconds. It is particularly important to remember the following points:

 Use the correct viscometer size. The best results will be obtained operating near the center of the viscometer range. For instance, if the fluid has a viscosity of 100 cSt, then use a tube size 300 (range 50-250). Table 2-2 shows the tube size with their respective viscosity ranges and approximated constants.

 Avoid overlapping the viscometer. Using either too much or too little sample will produce inaccurate values of viscosity. The volume that fills the tube from E to the inlet of the A tube will approximately half fill the bulb marked H.

Use the correct calibration factor, k. The constant shown in Table 2-2 are only approximated. Each viscometer tube must be calibrated regularly with standard fluid to find the correct constant.

 After loading the sample, wait at least 10 minutes before making measurements. Some time is required to allow the sample to equilibrate at the temperature of the water bath and to allow air bubbles to segregate.

(5)

 Use special methods for opaque samples. For high viscosity oil samples, which are sometimes quite opaque, it is necessary to use auxiliary illumination to improve judgments of the passage of the interface past the starting and ending marks.

Table 2-2. Recommended Viscosity Ranges for The Cannon-Fenske Routine Viscometers (Cannon-Fenske Viscometer Instrument Manual)

Size cSt / Second Approximate constant Centistokes Range 25 0.002 0.5 to 2 50 0.004 0.8 to 4 75 0.008 1.6 to 8 100 0.015 3 to5 150 0.035 7 to 35 200 0.1 20 to 100 300 0.25 50 to 250 350 0.5 100 to 500 400 1.2 240 to 1200 450 2.5 500 to 2500 500 8 1600 to 8000 600 20 4000 to 20000 Laboratory Experiments

 Place fluid sample in a Cannon-Fenske tube as explained in this chapter and put it in a controlled temperature bath.

 Measure the time that takes to the sample to go from mark “C” to mark “E” for four or five different temperatures (same as those of density measurements).

References

1. Cannon-Fenske Viscometer Instrument Manual.

References

Related documents

It is estimated that between 8% and 50% of beginning teachers in Queensland leave the profession in the first five years 1. National workforce data indicate that heavy workload

Another important lever for poverty reduction through environmental policy is the mitigation of global climate change by making efforts to reduce the emission of greenhouse

It only states that the monks could purchase a new roaster for $35,000 which would increase production, but there is really no supporting evidence that the monks or Father Daniel

To summarize; in this lesson we have discussed about the kinematic viscosity of bitumen and its measurement because in the previous lesson we have talked about the absolute

All oil shows decrease in viscosity with increase in temperature The dynamic viscosity calculated in terms of Kinematic Viscosity .The temperature dependent

This section shows each major section of your engine broken down into individual part numbers along with an associated illustration for identifying the parts you require. The

The idea of amplitude modulation is extended to the case where the modulating signal is complex, and the ‘carrier’ is an orthonormal complex exponential, that is a complex

Auto-ignition temperature : No data available Decomposition temperature : No data available Viscosity, dynamic : No data available Viscosity, kinematic : No data available