Centrifuge 5417 C / 5417 R

Centrifuge 5417 C / 5417 R

 5417 C

 5417 C from Serial no.

from Serial no. 13,000 upwa

13,000 upwards / 5417 R from Seri

rds / 5417 R from Serial no. 7,000 upw

al no. 7,000 upwards

ards

Service Manual 

Service Manual 

 English

 English

(2)
(3)
(4)

 Disassembly / Assembly

 Disassembly / Assembly 4

4

Safety 1

Safety 1

(blank)

(blank)

PA

PA

CT

CT

informati

informati

on

on

sheets

sheets

2

2

 Error lists / T

 Error lists / Troubleshooting

roubleshooting 3

3

(blank)

(blank)

Alignment

Alignment

/

/

Adjustment

Adjustment

5

5

 Diagrams

 Diagrams 6 

Spare

Spare parts

parts list

list 7 

T

Technic

echnical

al data

data 8

8

 Maintenance

(5)
(6)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards

Contents

1

Safety 

Using this manual ... ... 1-1 Warranty ... ... 1-1 User profile ... ... 1-1 Warning signs and hazard icons ... ... 1-2  Danger levels ... ... ... 1-1 Safety instructions on the device ... ... 1-2 Safety precautions ... ... 1-3

 2

PACT information (blank)

 3

Error lists / Troubleshooting

 Error list ... ... 3-1 Warnings and directions ... ... 3-8  Errors from the service programs ... ... 3-8 Imbalance measuring system:... ... 3-8 Customer service displays ... ... 3-9

1 Temperatures (5417 R only) ... ... 3-9 2 Condenser fan (5417 R only) ... ... 3-9 3 FC parameters... ... 3-10 4 FC parameters... ... 3-10 5 Rotor recognition... ... 3-10 6 Imbalance determination... ... ... 3-11 7 Last error message... ... 3-11 8 Version... ... 3-11 9 Current error message ... ... 3-12

(7)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards

Contents

 Description of service programs ... ... 3-13 Program overview... ... 3-13 General description ...3-14 Setting the serial interface...3-14  Voltage version ... ... 3-15

Display and keypad check ...3-15 Lid function test... ... ...3-15 Initializing the EEPROM... ... 3-16 First initialization ... ... 3-16 Follow-up initializations ... ... 3-16 Switching the aggregates / temperature sensor test (5417 R only) ... 3-17 Cycle program ... ... 3-17 Rotor recognition test / rotor pulses...3-18 Frequency converter (FC) test ... ... 3-19 Imbalance calibration ...3-20 Determination of acceleration and braking times... ... 3-21 Overspeed cut-off test...3-22  Viewing the error table... ... ... 3-22

Display number of starts / number of runs ... ... 3-22  Viewing the values stored in the EEPROM... ... 3-23

 4

Disassembly / Assembly 

5417 R only!... ... 4-1 Centrifuge upper part (5417 C only)... ... 4-1 Upper part of housing (5417 C only) ... 4-2 Upper part of housing (5417 R only) ... 4-3  Front cover (5417 R only) ... ... 4-4  Lid... ... 4-4  Lid latch (5417 C only) ... ... 4-4  Lid latch (5417 R only) ... ... 4-5  Motor (5417 C only) ... ... 4-6   Motor (5417 R only) ... ... 4-6   Frequency converter PCB (5417 C only) ... ... 4-7   Frequency converter PCB (5417 R only) ... ... 4-8  Keypad (5417 C only) ... ... 4-9  Keypad (5417 R only) ... ... 4-9  Brake resistors (5417 C only) ... ... 4-10   Brake resistors (5417 R only) ... ... 4-11  Rotor chamber (5417 C only) ... ... 4-12

(8)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards

Contents

 5

Alignment / Adjustment (blank)

6

Diagrams

Overview 5417 C electronics ... ... 5417 810.937  Overview 5417 R electronics ... ... 5408 810.983 Control ... ... 5408 810.908

7

Spare parts list 

Spare parts list ... ... 7-1  Recommended spare parts 5417 C – First equipment ... ... 7-13  Recommended spare parts 5417 R – First equipment ... ... 7-15

 8

Technical data

Technical data 5417C (from S/N 13,000 onwards) ... ... 8-1 Technical Data 5417 R (from S/N 7,000 onwards) ... 8-2

 9

Maintenance

Overview maintenance ... ... 9-1 Operation Qualification/Additional maintenance ... 9-2  Description of centrifuge mainte nance... 9-3  Additional maintenance for centrifuges... ... 9-8  Description of the device test for centr ifuges ... ... 9-9

(9)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards

Contents

 Hardware description – General construction ... ... 12-1 Frequency converter assembly group...12-1 Display/operation assembly group...12-1 Imbalance monitoring assembly group ... ... 12-1 Three-phase asynchronous motor ... ... 12-1 Lid latch with query keys...12-1  Various cables ... ... 12-1

Refrigerating aggregate, condenser fan, solenoid valve ... 12-1  Functional description of the assembly groups ... ... 12-2  Power supply... ... 12-2  Frequency converter assembly group ... ... 12-2 Input power supply unit with PFC control ... ... 12-2 Switching power unit for the voltage supply for the control assembly groups ... 12-2 Power unit for motor drive and brake chopper ... ... 12-3 Computer core with interface to the operating unit PCB ... 12-3  Display/operation assembly group ... ... 12-3  Rotational speed determination and rotor recognition ... 12-4  Active mechanism ... ... 12-4  Imbalance determination ... ... 12-4 Temperature determination (Ref  rigerated Centrifuge 5417 R only) ... 12-5 Rotor chamber temperature ... ... 12-5 Condenser temperature...12-5  Lid magnet selection and lid query ... ... 12-5

Converter interface ...12-5 Condenser fan regulation (Refrigerated Centrifuge 5417 R only) ... 12-6 

Power supply

for refrigerating aggregate, condenser fan, solenoid valve (5417 R only) ... 12-6 Refrigerating aggregate...12-6 Condenser fan ... ... 12-6 Solenoid valve ...12-6  PC interface... ... ... 12-6 

(10)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards

1 Safety 

Using this manual 

 Read this service manual before commencing work on the device.

 Please also follow the safety instructions of the corresponding operating manual.

Make sure that you have the current versions of the service and operating manuals. Check this using the version numbers under www.eppendorf-support.com and www.eppendorf.com.

Warranty 

Liability in the case of incorrect maintenance, servicing or testing

The Eppendorf service par tner commissioned with carrying out maintenance, servicing or testing is responsible for ensuring that all work is performed in a professional manner.

 Maintenance, servicing or testing work must not alter the condition of the device such that its safety is reduced in comparison to its original condition.

Only use accessories and original spare par ts recommended by Eppendorf for maintenance, servicing or testing as well as measuring and testing devices recommended by Eppendorf.

User profile

Requirements of specialists entrusted with maintenance, servicing or testing work The following prerequisites are required for all maintenance, servicing or testing work:

 Service training provided by Eppendorf AG for the product to be maintained, repaired or tested.

 Successful participation in service training measures with Eppendorf AG certification.

 Specialist training with knowledge of applicable local and international standards.

  Ability to assess delegated work.

(11)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards

1 Safety 

Warning signs and hazard icons

 Danger levels

The danger level is part of a safety note and defines the potential consequences if the information contained therein is not observed.

Safety instructions on the device

Biological hazard Danger of explosion

Risk of electric shock Risk of being crushed

General danger Material damage

Notice

DANGER Will lead to serious injuries or death. WARNING May lead to serious injuries or death. CAUTION May lead to minor to moderate injuries.  ATTENTION May lead to material damage.

Notice Provides useful information.

Depiction Meaning Location

WARNING

General hazard point. Follow the operating manual.

(12)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards

1 Safety 

Safety precautions

DANGER

Electric shock

 Switch off the device and disconnect the power plug before opening the device, performing work on electrical connections or starting maintenance or cleaning work.

DANGER

Risk of explosion.

 Do not operate the device in rooms where work is being carried out with explosive substances.

 Do not use this device to process any explosive or highly reactive substances.

 Do not use this device to process any substances which could create an explosive atmosphere.

WARNING

Infection by contaminated material.

 Inform yourself about any potential contamination risks before starting work with the device.

 Ensure that the device is decontaminated. Check the respective decontamination certificate.

 Wear protective clothing (protective gloves, protective glasses) when working with the device.

WARNING

Centrifuge lid can crush. Keep hands clear.

 When opening or closing the device lid, do not reach between the lid and device or into the latching mechanism of the lid.

 Always open the centrifuge lid completely to prevent it from falling.

WARNING

Injuries from sharp-edged components

Sharp-edged components are exposed when the device is open.

 Wear protective gloves and exercise proper caution when working on the device.

WARNING

Risk from incorrect supply voltage

 Only connect the device to power sources that match the electrical specifi cations on the device's nameplate.

 Only use sockets with protective earth conductor.

WARNING

Risk of injury from chemically or mechanically damaged accessories. Even minor scratches and cracks can result in serious internal material damage.

 Protect all parts from mechanical damage.

 Check accessories regularly.

(13)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards

1 Safety 

NOTICE

If a fault occurs, objects in the immediate vicinity of the devices could get damaged.

 In accordance with the recommendations of EN 61010-2-020, leave a safety distance of 30 cm clear around the device during operation.

NOTICE

Damage to electronic components from condensation.

After the device has been moved from a cool to a warmer environment, formation of condensation can occur inside the device.

 Wait for at least three hours before connecting the device to the mains power supply.

 Alternatively: let the device run for half an hour j ust before a short transport.

NOTICE

Damage to electronic components can result from electrostatic discharge.

 Store and transport vulnerable components or modules only in antistatic or conducting packaging.

 Discharge yourself before removing the component or module from its packaging. For example, use a grounding bracelet, conducting shoes, or conducting base under the device.

 Hold the component or module only by the edges.

 Do not touch protruding pins or conductor tracks.

 Do not rub the component or module against objects, especially plastics.

NOTICE

Refrigerating system malfunction or damage to centrifuge.

 Repairs to the refrigerating system must only be carried out by a refrigeration expert.

 Repairs, e.g., replacement of spare parts, are carried out at your own risk.

 Observe all national and local safety regulations.

NOTICE

Risk of damage to compressor and refrigerating system by tilting the device

The compressor and refrigerating system may be damaged by switching on the device when in a tilted position or by switching it on immediately after tilting.

 Allow the device to stand in an upright position for at least 4 hours before switching it on.

 Never turn the device upside down.

NOTICE

Damage as a result of incorrect packing.

Eppendorf accepts no warranty or liability for damage caused by incorrect packing.

(14)

Centrifuge 5417 C / 5417 R 

5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards 

(15)

PACT-System

Product News

Application News

Commercial Info

Technical Info

Mögliche Beschädigungen an Bedienteilen der Centrifugen 5417 C

Durch Rückmeldungen über unsere Service Info Line (SIL) aus den Märkten wurden wir auf einen möglichen Fehler an neuen Bedienteilen der Centrifugen 5417C aufmerksam gemacht.

Beim Abziehen der Schutzfolie verblieben offensichtlich Kleberreste auf der Oberfläche des Display. In vielen Fällen ließen sich diese Rückstände mit handelsüblichem Waschbenzin entfernen. In einigen Fällen war jedoch die Oberfläche des Display bereits angegriffen, so dass das Bedienteil ausgewechselt werden musste.

 Als Verursacher dieser Beschädigungen wurde ein Silikon-Lösungsmittel in einem Glasreiniger ermittelt. Der entsprechende Fertigungsprozess wurde bereits geändert.

Sollten Sie ein Display vorfinden, bei dem sich die Kleberreste nicht rückstandslos mit Waschbenzin entfernen lassen, berichten Sie uns diesen Fall bitte unter Angabe der Gerätenummer per SIL. Wir werden das Bedienteil umgehend austauschen.

Possible damage to control panels of the Centrifuge 5417 C

Due to feedback obtained from the markets through our Service Info Line (SIL) , a possible error involving the new control panels of the Centrifuge 5417C has been brought to our attention.

When the protective foil is pulled off, it seems that adhesive residue remains on the surface of the display. In many cases this residue can be removed with commercially available petroleum ether. However, in some cases the surface of the display had already been damaged, meaning that the control panel needed to be exchanged.

 A silicone solvent in a glass cleaner has been determined to be the cause of the damage. The production process involved has already been changed.

(16)

PACT-System

Product News

Application News

Commercial Info

Technical Info

New serial number for all centrifuges

Starting with the beginning of January 2008 all newly produced centrifuges carry a

redesigned nameplate. Additionally, there is a new consistent format of the serial number.

Redesigned nameplate for centrifuges with the new 12-digit serial number  (example):

Please note that, depending on the type of instrument, it can take some time until the first

devices with the new nameplate and the new serial number reach the market.

(17)

Centrifuge 5417 C / 5417 R 

5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards 

(18)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 Error list 

 Display  Error Cause Test/Solution

 Error 1

Centrifuge brakes.

1. Error in measurement

system for rotational

speed. No rotor

recognized.

2. Magnet assignment of

the rotor coding is

defective.

3. Only two magnets

rec-ognized (half rotational

speed).

4. Incorrect voltage

ver-sion has been set.

5. A loose rotor has been

recognized (rotor may

have tightened during

acceleration).

Use another rotor. If the

error does not recur, check

the coding ring of the rotor.

Check whether the

mag-nets are complete and in

good condition.

Check the rotational speed

sensor with customer

ser-vice program 8:

– Four pulses per rotation;

test with service program

8 (SE 8) possible.

– Rotor ratio, test with

ser-vice program 8 or

cus-tomer service display 5.

 Error 2

Centrifuge does not start or

brakes.

Imbalance idle value is

out-side the permitted limits, or

the measuring value is

incorrect.

Incorrect idle value must be

applied for at least three

seconds.

1. Sensor value < 0.5 V or

> 4.5 V; sensor is

defective.

2. Idle value is ± 220 mV

outside the stored idle

value (SE15).

Customer service program 6

shows the stored idle value

and the current idle value.

Test: Current idle value

changes when the motor

axle moves. If it does not

change, the sensor is

defec-tive or the magnet bracket is

missing.

When the centrifuge is

sta-tionary, the current idle

value must be within the

lim-its of ± 25 diglim-its (equivalent

to ± 220 mV) of the stored

(19)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 Error 3

No rotor pulses.

No rotor pulses after

start-ing. Centrifuge brakes for

ten seconds.

No rotor pulses during run.

Centrifuge brakes and is

locked for eight minutes.

No rotor; measuring system

for rotational speed is

defective or drive is

defec-tive. Error remains active

after a power supply

inter-ruption. Keep the centrifuge

switched on. After an

eight-minute blocking time, Error

3 is deleted and the lid

opening is released. Press

the Open key.

Check the state of the rotor

coding. Check the function

of the rotational speed

sen-sor by using SE 8.

 Error 4

(not occupied)

 Error 5

Drive switches off;

centri-fuge decelerates without

braking.

No report from lid by at

least one limit switch during

run.

Reports of lid positions:

5417 R:

1. Lid is closed –

microswitch in the lid

latch.

2. Lid is locked –

micro-switch in the lid latch.

5417 C:

1. Lid is closed – Reed

contact in the lid latch,

activated by the lid

magnet.

2. Lid is locked –

micro-switch in the lid latch.

Test of the lid signals with

service program 5 or SE 4.

(20)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 Error 6 

Centrifuge does not start, or

decelerates without braking.

 Additional error messages

of the frequency converter

(FC) only in service displays

as last number of display 9:

0 – No error

1 – Undervoltage

2 – Overvoltage

4 – Overtemperature

Converter

7 – Overcurrent converter

8 – EEPROM FC error

10 – No release

40 – Short-circuit in power

unit

FA – Reset by watchdog

FF – Serial interface time

out

Error in the frequency

con-verter; possible thermal

overload.

1. Not ready after

switching on.

2. Error detected in

fre-quency converter

(Par. 23).

3. Error when braking.

4. Frequency converter

carried out a RESET.

 Allow centrifuge to cool for

five minutes, then restart.

If necessary, check the

function of the frequency

converter (FC) with service

program 9 (SE 9).

Check the state of the brake

resistors thoroughly before

replacing a converter.

+ Repair work on the con-verter is not permitted! 

 Error 7 

Centrifuge brakes.

1. Overspeed monitoring

is active; maximum

rotational speed of the

rotor has been

exceeded.

2. There are regulator

deviations of ± 800 rpm

from the nominal value.

Check the motor. Check the

frequency converter with

SE 9.

 Display  Error Cause Test/Solution

power off w/o exiting a service ro ram

(21)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 Error 8

Centrifuge brakes.

Non-compliance with

accel-eration ramps.

Ramp control has detected

a non-permissible deviation

during acceleration.

The rotor may not be

cor-rectly fastened.

The voltage version is not

set correctly. Check with

customer service display 8.

 Adjust the voltage version

using SE 2.

Check the acceleration with

SE 11.

 Error 9  and  Error 10 

EEPROM error.

Writing error (Error 9) or

reading error (Error 10) of

the EEPROM. Error is

dis-played and the standard

settings are set. If there is

an error when reading the

setting for imbalance, the

centrifuge can no longer

start (Error 2!).

Re-initialize the EEPROM

with SE 5. Alternatively,

replace the operating unit.

 Error 11

No further actions possible.

Watchdog error on the

Operation/display PCB.

Replace the operating unit.

 Error 12  and  Error 13

Communication error of the

PC interface.

Centrifuge can be operated

manually.

Error 12, error during

trans-mission.

Error 13, reception error.

Parity error, incorrect code

number, check sum error

Check the data exchange

with the interface tester.

Check the cable

connec-tions.

 Error 14

No further actions possible.

EPROM check sum error

after mains-on.

Error program memory,

EPROM defective.

Replace the operating unit.

(22)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 rotor  (error 15)

Centrifuge switches off and

decelerates with braking.

Imbalance error at

n > 3,000 rpm.

The centrifuge was pushed

or the rotor was not

fas-tened.

Check installation of the

centrifuge, even surface,

stable position.

Check the motor and rotor.

Undamaged rotor sit and

motor axle. Check whether

the rotor is correctly

fas-tened and positioned.

 Error 16  and Error

17 

Centrifuge switches off and

decelerates without braking.

Communication error with

the frequency converter

(FC).

Error 16:

1. Error during request to

frequency converter.

2. Error after telegram

repetition.

Error 17:

Error during transmission of

select telegram to FC:

1. Queue overflow FC

telegrams.

2. Error after three

repeti-tions of a select

tele-gram.

Check the cable connection

between the operating unit

and the converter PCB.

Replace the converter or the

operating unit.

(23)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 Error 18

(5417R only)

Centrifuge switches off,

braked deceleration.

Refrig-erating aggregate off.

Overtemperature of the

rotor chamber:

1. After the temperature

regulation the interval

of ± 5 °C is exceeded.

2. Chamber temperature

is more than 10 °C

higher than the start

temperature or the

nominal temperature.

3. Chamber temperature

> 50 °C.

Check the cooling system

with SE 6.

 Error 19

(5417R only)

Centrifuge switches off,

braked deceleration, the

refrigeration aggregate is

switched off, the fan keeps

running.

Excess temperature of the

condenser; measured

tem-perature higher than

+ 58 °C.

 Allow the refrigeration

aggregate to cool.

Check the function of the

fan.

Clean the condenser, the

fan opening must not be

blocked.

 Error 20

(5417R only)

Centrifuge switches off,

deceleration with braking,

the refrigeration aggregate

is switched off.

Temperature sensor in the

rotor chamber is defective

(NTC).

Check the chamber sensor

with SE 6.

Replace the sensor.

 Error 21

Centrifuge switches off,

deceleration with braking.

Temperature sensor on

con-denser is defective (NTC).

The measuring value is not

between – 8 °C and

+ 83 °C.

Replace the condenser

sen-sor.

(24)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 Error 22 (5417R only)

Fan is possibly controlled

with full power.

Error in the fan control.

There is no 50(60) Hz

sig-nals for the fan control and

mains-off recognition:

1. During run and

stand-by, the fan is controlled

with full power, error

display.

2. After mains-on.

Replace the converter PCB.

 Error 22 (only  5417C)

Broken mains detection

Replace the converter PCB.

 Error 23

Lid is blocked until the

fre-quency converter detects a

standstill.

Irregularities during braking.

Check connections of the

brake resistors. Check cable

between operating unit and

converter.

Check function of converter

with SE 9.

 Int 

(Error 25)

Mains interrupted during a

run.

Check power supply: mains

connection cable, mains

input module in centrifuge,

mains plug.

(25)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

Warnings and directives

+ Warnings are not stored in the EEPROM! 

(Warning messages=Wxx).

clEAn ro W28

 After the 200th/201st/202nd run with the installed rotor, the maintenance

advice "clean rotor" is emitted.

SPEED W29

The rotational speed set is higher than the nominal rotational speed of the rotor

in use.

The centrifuge starts, but stops running after recognizing the rotor.

 InbAL W30 

There is a non-permissible rotational speed of n < 2,000 rpm.

The centrifuge brakes.

 Error 37 

The voltage version of the frequency converter does not correspond to that of

the centrifuge.

The check is carried out after switching on the centrifuge.

 Errors from the service programs

 Imbalance measuring system:

 Error 32

Hardware error: imbalance measuring system.

 Error 33

Measuring values outside the measuring range, too little space between sensor and

magnet.

 Error 34

Incorrect measuring series:

1. Mean value not within limits or distance too short.

2. Scattering run 1-3 too great.

3. Scattering run 4-6 too great.

Possible causes: The distance between the sensor and the magnet is too large or is set

incorrectly.

 Error 35

Imbalance sensor idle value exceeds the setting limits of 145 to 195.

 Error 36 

EEPROM not yet initialized; call up service program 5.

(26)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

Customer service displays

For Centrifuge 5417 C from serial no. 13.000 onwards and for Centrifuge 5417 R from serial no. 7.000

onwards.

If the SOFT and SPEED-PLUS keys are pressed simultaneously, the centrifuge switches over to the

ser-vice display mode. This does not depend on the current centrifuge operational mode. The centrifugation

runs remain unaffected. The displays are run through by pressing the SOFT key. During the service

dis-play, a SOFT switchover is not possible.

In the case of error, the display automatically jumps to "current error message". It is then possible to

switch to different displays.

The display "current error message" shows the current error and the error location. For Error 6, the

con-verter error number is also shown.

1 Temperatures (5417 R only)

1

m

T

y

y

y

z

n

z

z

m

Solenoid valve, A = open, – = closed

T

Regulator value (0 to 3, parameter for production)

yy.y

Actual temperature rotor chamber sensor

zz.z

Corrected nominal value = regulator value

n

Motor heating switched on in standby

 2 Condenser fan (5417 R only)

2

x

x

x

K

 

n

y

xx

Condenser temperature, refrigeration aggregate switches off at 75 °C

yy

Display of the regulator level, 0 to 16

K

Refrigeration aggregate E = switched on, A = switched off

n

Solenoid valve is opened.

(27)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 3 FC parameters

3

x

x

x

x

x

|

|

|

|

1 2

3 4

xxxxx

Rotational field speed

FC status

1 – FC is ready.

2 – Go-ahead is given.

3 – Timing switched on (power unit (FC) is selected).

4 – Collective error message (short-term indication, if there is a converter error,

the current error message is displayed with Function 9).

 4 FC parameters

4

y

y

y

z

n

x

x

yyy

Open lid:

software version FC

Closed lid: intermediate circuit voltage

z

FC mode 0/1/3/7, parameter for production

n

FC temperature warning, from approx. 5 °C below switch-off value

xx

Temperature frequency converter, switch-off at approx. 75 °C

 5 Rotor recognition

5

x

x

x

y

u

v

xxx

Rotor relation

y

Rotor number

u, v

lid switch

0 = open, 1 = closed

 Rotor no. Rotor model Rotor relation 5417 C 5417 R

R

0

F

45-30-11

85–132

14,000

rpm

14,000

rpm

R 0

FA 45-30-11

85–132

14,000 rpm

14,000 rpm

R 1

A-12-11

142–195

10,000 rpm

Rotor too large

R

1

A-8-11

142–195

10,000

rpm

10,500

rpm

R 1

F 45-48-PCR

142–195

10,000 rpm

10,500 rpm

(28)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

6 Imbalance determination When the lid is open

6

x

x

x

y

y

y

 

xxx

Current imbalance idle value during standstill 170

±

25 (-->145-195)

yyy

Stored idle value

When the lid is closed 

6

x

x

y

y

z

z

xx

Max. imbalance that occurred during a run up to 3,000 rpm

yy

Max. imbalance that occurred during a run from n Š 3,000 rpm

zz

Imbalance switch-off limit for the rotor in use.

7 Last error message

7

x

x

y

y

z

z

xx

Last error

yy

Last but one error message

 8 Version

8

x

x

x

x

y

y

 

(29)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 9 Current error message

9

E

x

x

y

z

z

xx

Error number

y

Number for locating the cause

zz

FC error number, if the converter switches off with error 6

FC error number:0

– No error detected in FC

1

– Undervoltage

2

– Overvoltage

4

– Excess temperature converter

7

– Overcurrent converter

8

– Error

EEPROM

FC

10

– No

release

40

– Short-circuit in power unit

FA

– Reset by watchdog

(30)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 Description of service programs

Centrifuge 5417 C from serial no. 13.000 onwards and Centrifuge 5417 R from serial no. 7.000 onwards.

 Program overview 

Program no.Program function

1

Setting the serial address

2

Setting the voltage version

3

Checking the display and the keypad

4

Testing the lid function / lid limit switches

5

EEPROM initializing

6

Test refrigeration aggregate / fan / temperature sensors

7

Cycle test

8

Rotor recognition

9

Test frequency converter

10

Calibration of imbalance sensor

11

Determination of the acceleration and braking times

12

Overspeed cut-off test

13

Viewing the error table

14

Display of the number of starts / number of runs / operating hours

15

Display of the switching off values for imbalance / sensor idle value / voltage version

16–19

Test program for production.

(31)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

General description

The service programs are independent of the normal operating modes. It is possible to access program

selection by pressing the SOFT and TIME-PLUS keys together and simultaneously pressing the mains

switch.

Display:

S

E

x

x

xx

Service program number

The required program number can be set by pressing the TIME-PLUS or TIME-MINUS key. The lid is

opened by pressing the OPEN key. All programs are started by pressing the START key. Errors which are

displayed can be deleted by pressing the START key.

If "Error 36" appears in the display immediately after the service programs have been called up, the

EEPROM must be initialized.

The service mode can only be quit by switching off the centrifuge.

Setting the serial interface START Program 1

b

U

S

x

0 = Disable

1 = RS232

2 = Reserved for CAN bus

Press the START key to access the setting of the device address.

 A

d

r

x

x

xx

Device address 20...29 (Hex)

By pressing the TIME-PLUS key, the address to be set is continuously counted upwards.

START

To confirm the selected address.

(32)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

Voltage version START program 2

S

P

x

x

x

Start SE 2 The version which has been set is displayed.

SPEED-PLUS buttonSelection of the voltage version 230 – 120 – 100

SOFT buttonValue is stored in the EEPROM – display 'OK'.

STOP buttonEnd of program

 Display and keypad check  START program 3

1. Display checkThe sound is on for 1.5 seconds; all segments are displayed

2. Keypad checkDisplay is deleted

press any key"1" is displayed for each key

The STOP key is the last key to be pressed; this is also the end of the program.

 Lid function test  START program 4

L

I

d

x1

x2

x1 – bolt limit switch

1 = Lid closed 0 = Lid open

x2 – lid limit switch

1 = Lid locked 0 = Lid unlocked

2)

3)

4)

(33)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 Initializing the EEPROM 

The set nominal values and the operating conditions are stored in the EEPROM after each run of the

centrifuge. The limits for switching off (imbalance) and statistical data are also stored in the EEPROM.

 First initialization

Error 36 is indicated if no identification is found in the EEPROM when entering the ser vice mode.

The error can be erased with the START button.

START program 5

I

n

I

T

0

Presetting rotational speed = 14,000 rpm, rcf = 20,800, radius = 89 mm,

time = 10 min.,

temp. = 18 °C, imbalance idle value = 170,

imbalance = 17,

interface adr. = 20H, inactive,

all tables are erased

SOFT button

initialization is carried out,

STOP button

no initialization, return to program selection

4

1

4

0

0

0

1

0

+  Always carry out imbalance calibration after the first initialization! 

 Follow-up initializations START program 5

I

n

I

T

1

STOP buttonno initialization, return to program selection

SOFT button

initialization is carried out, calibration data for imbalance,

voltage version and interface setting remain intact,

all other data is deleted.

Possible errors during initialization: Error 9, EEPROM cannot be inscribed correctly.

(34)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

Switching the aggregates / temperature sensor test (5417 R only) START program 6 

 A

x

U

y

L

z

s

s

Switching the aggregatesTEMP_PLUSrefrigeration aggregate x = 1 agg. on / x = 0 = agg. off

SPEED-MINUS solenoid valve y = 1 valve open / y = 0 valve closed

SPEED-PLUS

condenser fan Z = 1 fan on / Z = 0 fan off

ss – fan step (0..16), the step can be changed by means

of the TIME-PLUS and MINUS keys

FC key

·

 To switch over to the temperature display

 y

y

y

x

x

x

xx

Condenser temperature

yyy

Rotor chamber temperature

STOP

End

Cycle program START program 7 

This program is used for a continuous centrifuge test. The SHORT, SOFT and FC functions are available.

Specifications:

5417 C: 4 runs of 1 minute each; interval of 1 minute; 10-minute interval after the 4th run.

5417 R: 4 runs of 10 minutes each; interval of 1 minute; 10-minute interval after the 4th run.

P

A

U

S

E

x

x

The long interval can be set gradually with the SOFT button in limits of between 1 minute and

20 minutes. (Standard: 10 minutes).

The nominal values of the run can be changed after the service program has started. The operating

6) 7)

A toggle

U toggle

L

(yy.y)

(xx.x)

(35)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

START program 7 

4

1

4

0

0

0

1

0

Display of the preset nominal values. Alterations are possible.

START

To begin the cycle run

Display after each run:

P

A

U

S

E

x

x

xx

Interval time, in minutes or seconds

 An individual run can be aborted by pressing the STOP key. The lid can be opened during the interval.

The cycle program can be quit by pressing the SHORT key.

 Rotor recognition test / rotor pulses START program 8

r

o

1

2

3

4

Slowly turn the rotor by hand; a digit is displayed for each pulse (4 pulses per rotation!)

Press the SPEED_PLUS key: Calculation of rotor ratio and rotor number

r

o

x

x

x

z

xxx

Rotor ratio

z

Rotor number 0, 1, 2

SE 8 is ended via STOP.

8)

(36)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 Frequency converter (FC) test 

Both lid limit switches must be closed during this test, otherwise the frequency converter is not released.

Safety precautions:

 Do not insert a rotor if the test is car ried out with an open centrifuge or without the centri-fuge lid! 

START program 9

 After the program has started, an indication that the centrifuge is ready to use must appear in the

dis-play.

TIME-PLUS keyRun with approx. 120 rpm

TIME-MINUS keyBraking

F

U

b

x

x

x

z

F

U

E

r

6

y

y

 

b

Converter ready

z

1 = FC started

xxx

Rotational field frequency

yy

FC error code in case of error 1 – Undervoltage

2 – Overvoltage

4 – Overtemperature converter

7 – Overcurrent converter

8 – Error EEPROM FC

10 – No release

40 – Short-circuit in power unit

FA – Reset by watchdog

FF – Serial interface time out

STOP key or SHORT key: the FC is switched off; centrifuge returns to program selection.

(37)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 3 Error lists / Troubleshooting

 Imbalance calibration

General information:

Imbalance calibration requires two calibration weights. Two reaction tubes of 1.5 ml or 2.0 ml are filled

and balanced with the corresponding material (e.g. quartz sand).

The tube for the lower calibration value must have a total weight of 2.3 g ± 0.1 g. The total weight of the

vessel for the upper calibration value is 3.8 g ± 0.1 g.

If the imbalance calibration is carried out with fixed-angle rotor F 45-30-11 or FA 45-30-11 (rotor number

R0), the determined limits are converted for the swing-bucket rotor and the PCR rotor (rotor number R1)

as well as the 24-position rotor (rotor number R2).

However, if the calibration is carried out with a different rotor than the 30-position rotor, the determined

values are valid for this rotor only; the values cannot be carried over to another rotor!

The imbalance calibration is carried out without rotor lid.

Install a flyweight of 2.3 g (lower value) into the rotor.

Close the centrifuge lid.

START program 10 

Three runs are carried out and the measured value imbalance value is displayed,

The lid automatically opens after the third run.

Install a flyweight of 3.8 g (upper value).

Close the centrifuge lid, Three runs are repeated.

Calculation and display of the switch-off value (it is stored in the EEPROM).

Start

test

 run. In the display "Inb

 AL

" is displayed. The service function is leaving.

r

u

n

y

x

x

y

Run

number

xx

Measured imbalance value

E

n

d

x

x

xx

Calculated switch-off value

START

Test run with the 3.8 g flyweight; the centrifuge must display "I

nb

 AL" and then switch off

OPEN

To open the centrifuge lid.

SHORT

To abort the calibration in progress; return after the machine has come to a standstill.

Error displays:

Error 33

Measuring values outside the measuring range; too little distance between sensor and

magnet.

Error 34

Incorrect measuring series (deviation is too large or the space between the measuring

(38)

Centrifuge 5417 C / 5417 R

Centrifuge 5417 C / 5417 R

 5417 C from seria

 5417 C from serial no. 13,000 onw

l no. 13,000 onwards / 5417 R from se

ards / 5417 R from serial no. 7,000 onw

rial no. 7,000 onwards

ards

 3

 3 Error lists / T

Error lists / Troubleshooting

roubleshooting

 Determination of acceler

 Determination of acceleration and braking tation and braking t imesimes

Load the rotor completely.

Load the rotor completely.

Close the lid.

Close the lid.

START program 11 START program 11

3

3

0

0

1

1

4

4

0

0

0

0

0

0

1

1

0

0

S

ST

TA

AR

RT

T

T

To

o b

beeg

giin t

n thhe

e m

meeaassuurriinng

g rruunn..

 After the rotor has been reco

 After the rotor has been recognized, the measuring rotational

gnized, the measuring rotational speed is calculated b

speed is calculated by the

y the

program = 97 % of the max. rotational speed of the rotor used.

program = 97 % of the max. rotational speed of the rotor used.

Display of the acceleration time in seconds (x10)

Display of the acceleration time in seconds (x10), if the

, if the measuring rotational speed was reached

measuring rotational speed was reached

H

H

L

L

x

x

x

x

x

x

ST

STO

OP

P

T

To

o st

star

art

t of

of th

the

e m

mea

eassur

urem

emen

ent

t of

of th

the

e br

brak

akin

ing

g ti

tim

mee

Display of the braking time in seconds (x10), if the centrifuge has

Display of the braking time in seconds (x10), if the centrifuge has come to a standstill.

come to a standstill.

b

b

L

L

x

x

x

x

x

x

SH

SHOR

ORT

T

T

To r

o ret

etur

urn t

n to t

o thhe p

e pro

rogr

graam s

m seele

lect

ctio

ionn

 Acceleration and

 Acceleration and braking times

braking times depend on

depend on the rot

the rotor model,

or model, load and

load and voltage

voltage version. The

version. The average

average values

values

list

listed below are valid for 30-position rotor F 45-30-11 with

ed below are valid for 30-position rotor F 45-30-11 with the maximum load:

the maximum load:

V

Voollttaagge ve veersrsiioonn AAcccceelleerraattiioon tn tiimme (e (sseecc)) BBrraakkiinng tg tiimme (e (sseecc))

22330

0

V

V

77..55––1100..55

99..00––1111..55

11115

5

V

V

1100..00––1133..55

1111..00––1155..00

11000

0

V

V

1100..00––1133..55

1111..00––1155..00

S

So

offt

t rraam

mp

p

aap

pp

prro

oxx.

. 660

0 sseecc

aap

pp

prro

oxx.

. 660

0 sseecc

11)

(39)

Centrifuge 5417 C / 5417 R

Centrifuge 5417 C / 5417 R

 5417 C from seria

 5417 C from serial no. 13,000 onw

l no. 13,000 onwards / 5417 R from se

ards / 5417 R from serial no. 7,000 onw

rial no. 7,000 onwards

ards

 3

 3 Error lists / T

Error lists / Troubleshooting

roubleshooting

Overspeed cut-off test  Overspeed cut-off test 

Close the lid

Close the lid

START program 12 START program 12

 A rotational speed abo

 A rotational speed above the cut-off rotational speed is set b

ve the cut-off rotational speed is set by the software. If the centrifuge is OK, it

y the software. If the centrifuge is OK, it

switches itself off via Error 7

switches itself off via Error 7 when the cut-off speed has been reached and then decelerates with brak-

when the cut-off speed has been reached and then decelerates with

brak-ing.

ing.

 If there is no cut-off, the run m

 If there is no cut-off, the run m ust be immediately ust be immediately stopped by hand! stopped by hand! 

The cut-off rotational speeds are 300 rpm above the nominal speed of the rotor in use. The cut-off rotational speeds are 300 rpm above the nominal speed of the rotor in use.

--> SHORT: Leaving the Ser

--> SHORT: Leaving the Service-Functio

vice-Functionn

Viewing the error table Viewing the error table

The error that occurred first during a run

The error that occurred first during a run is stored.

is stored.

START program 13 START program 13

Display of the error number and the number of errors during a run

Display of the error number and the number of errors during a run

S

SO

OF

FT

T

T

To

o ttrraannssffeer

r tto

o tthhe

e sstteep

p m

mo

od

dee..

S

SH

HO

OR

RT

T

T

To c

o co

onnttiinnuue s

e stteep

p--b

byy--sstteep

p..

S

ST

TO

OP

P

T

To

o aab

bo

orrt

t tthhe

e p

prro

og

grraam

m..

E

E

r

r

x

x

x

x

y

y

y

y

 

 

xxxx

E

Errrro

or

r nnuum

mb

beerr

yyyyyy

N

Nuum

mb

beer

r o

of

f eerrrro

orrss

 Display of number of star

 Display of number of starts / number of runsts / number of runs START program 14

START program 14

x

x

x

x

x

x

x

x

x

x

E

E

A

A

 

 

xxxxxxxx

N

Nuum

mb

beer

r o

of

f ssttaarrttss

S

ST

TA

AR

RT

T

··

 next display

 next display

x

x

x

x

x

x

x

x

x

x

C

C

L

L

xxxxxxxx

N

Nuum

mb

beer

r o

of

f cceennttrriiffuug

ge

e rruunnss

START

START

··

 next display

 next display

x

x

x

x

x

x

x

x

x

x

h

h

12) 12) 13) 13) 14) 14)

(40)

Centrifuge 5417 C / 5417 R

Centrifuge 5417 C / 5417 R

 5417 C from seria

 5417 C from serial no. 13,000 onw

l no. 13,000 onwards / 5417 R from se

ards / 5417 R from serial no. 7,000 onw

rial no. 7,000 onwards

ards

 3

 3 Error lists / T

Error lists / Troubleshooting

roubleshooting

Viewing the values stored in the EEPROM  Viewing the values stored in the EEPROM  START program 15

START program 15

, then

, then continue switching all displays with

continue switching all displays with ST

START

ART

x

x

x

x

x

x

U

U

y

y

 

 

yy

R

Ro

otto

or

r nnuum

mb

beerr

xxxxxx

S

Sw

wiittcchh--o

off

ff vvaalluuees

s ffo

or

r aalll

l rro

otto

orrs

s ((nno

o.

. 00,

, 11,

, 22))

x

x

x

x

x

x

U

U

rr

xxx

xxx

Idle

Idle value

value imbalance

imbalance sensor

sensor –

– measurement

measurement at

at start

start of

of imbalance

imbalance calibration.

calibration.

x

x

x

x

x

x

S

S

P

P

xxxxx

x

V

Vo

ollttaag

ge

e vveerrssiio

onn

 y

 y

y

y

X

X

XX

XX

P0

P0, P

, P1,

1, PL

PL =

=T

Teest

st w

wor

ord

ds f

s for

or cche

heccki

king

ng th

the t

e teest

st pr

pro

oce

cess

ss du

duri

ring

ng tthe

he p

pro

rod

duc

ucti

tio

on;

n;

not for Service!

not for Service!

15)

15)

(0, 1, 2)

(0, 1, 2)

(displays each consecutively)

(41)

Centrifuge 5417 C / 5417 R

Centrifuge 5417 C / 5417 R

 5417 C from seria

 5417 C from serial no. 13,000 onw

l no. 13,000 onwards / 5417 R from se

ards / 5417 R from serial no. 7,000 onw

rial no. 7,000 onwards

ards

 3

(42)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 4 Disassembly / Assembly 

 5417 R only! 

 Do not place the device on its side. If the centrifuge has been tilted, leave the device in its  normal position for at least four hours in order to prevent dam age to the compressor! 

 5417C and 5417R:

 Each time the PCB, motor or imbalance angle is replaced or disassembled, align the  imbalance sensor for all available rotors! 

Centrifuge upper part (5417 C only)

– Switch off the centrifuge and disconnect the mains plug. – Open the lid.

– Disassemble the rotor.

– Remove the motor cap (undo the four Phillips screws). – Unscrew the two rear feet of the device.

– Unscrew the two front feet of the device. When doing so, hold together the upper and lower parts. – For further disassembling, place the upper part of the centrifuge on its right-hand side.

(43)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 4 Disassembly / Assembly 

Upper part of housing (5417 C only)

– Switch off the centrifuge and disconnect the mains plug. – Open the lid.

Precondition: Remove the lid. – Disassemble the rotor.

– Remove the Phillips screws (on the top side of the housing).

 Remove

 Phillips screws

 Remove

 Phillips screws

– Remove the dummy plug (situated below the keyboard) and press out the keypad.

– Disconnect the ground cable and the plug connections to the PCB of the operating unit. – Remove the housing by pulling it upwards.

 Assemble in reverse order.

(44)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 4 Disassembly / Assembly 

Upper part of housing (5417 R only)

– Switch off the centrifuge and disconnect the mains plug. – Open the lid.

Precondition: Remove lid. – Disassemble the rotor.

– Undo the two screws (top, rear) on the upper part of the housing.

– Remove the front cover by undoing the two screws on the bottom side of the housing (if necessary, move the centrifuge slightly over the table edge) (ball-pin fastening).

– Disconnect the ground cable and the plug connections to the PCB. – Remove the plastic button for emergency release.

– Unscrew the cover on the rear. – Undo the screws at the front.

   R  e   m   o   v   e   s   c   r   e   w   s    R  e   m   o   v   e   s   c   r   e   w   s

(45)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 4 Disassembly / Assembly 

 Front cover (5417 R only)

– Switch off the centrifuge and disconnect the mains plug.

– Remove the front cover by undoing the two screws on the bottom side of the housing (if necessary, move the centrifuge slightly over the table edge) (ball-pin fastening).

– Disconnect the ground cable and the plug connections to PCB.

 Assemble in reverse order.

 Lid 

– Loosen the four Phillips screws on the inside of the lid. – Remove the upper part of the lid.

– Unhook the torsion springs.

– Remove the snap ring from the left-hand side of the axle.

– Remove the axle from its mounting by slightly moving it to the side. When doing so, pay attention to the bushes.

– Remove the lid plate.

 Assemble in reverse order.

 Lid latch (5417 C only)

– Switch off the centrifuge and disconnect the mains plug.

Precondition: Remove the lid, disassemble the upper part of the housing. – Undo the two Phillips screws (at the front of the top side).

 Remove Phillips  screws

 Remove

(46)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 4 Disassembly / Assembly 

– Pull out the lid latch towards the front. – Loosen the plug connections.

 Assemble in reverse order.

– Start the centrifuge.

 Lid latch (5417 R only)

– Switch off the centrifuge and disconnect the mains plug.

– Precondition: Remove lid, remove the front cover, disassemble the upper part of the device. – Loosen the two Phillips screws (at the front of the top side).

 Remove Phillips screws  Remove Phillips screws

– Pull out the lid latch downwards.

– Loosen plug connection from frequency converter PCB.

 Assemble in reverse order.

(47)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 4 Disassembly / Assembly 

 Motor (5417 C only)

– Switch off the centrifuge and disconnect the mains plug. Precondition: Remove the upper par t of the centrifuge.

– Disconnect the plug connection and the ground wire.

– Loosen the motor mountings (three screws on the bottom of the base plate). – Take out the motor.

When replacing the oscillating metals (they may be replaced as a complete set of three only)  install the torsion protection.

 Assemble in reverse order.

– Start the centrifuge.

 Motor (5417 R only)

– Switch off the centrifuge and disconnect the mains plug. – Open the lid.

– Disassemble the rotor. – Unscrew the motor cap.

Precondition: Remove the front panel.

– Loosen the plug connections on the frequency converter PCB. – Place the device on its side.

– Remove the tray for condensation water by pulling it out towards the front. – Loosen the motor fastening (three screws).

(48)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 4 Disassembly / Assembly 

– Remove the motor by pulling it downwards. – Unscrew the motor plate (three screws).

When replacing the oscillating metals (they may be replaced as a complete set of three only   install the torsion protection.

 Assemble in reverse order.

– Start the centrifuge.

 After installing the new motor, check the function and rotation dir ection using ser vice function 9! The sleeve must firmly fit over the edge of the rotor chamber, otherwise it may  slip off during operation! 

 5417C and 5417R:

 Each time the PCB, motor or imbalance angle is replaced or disassembled, align the  imbalance sensor for all available rotors! 

 Frequency converter PCB (5417 C only)

– Switch off the centrifuge and disconnect the mains plug. Precondition: Remove the upper part of the centrifuge.

– Loosen plug connections on the frequency converter PCB.

– Remove three screws on the top side (two on the left-hand side, one on the right-hand side).

  e   s   c   r   e   w   s  Remove screws

(49)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 4 Disassembly / Assembly 

– Undo four screws (on the bottom of the base plate). – Take out the frequency converter PCB.

 Assemble in reverse order.

– Start the centrifuge.

 Frequency converter PCB (5417 R only)

– Switch off the centrifuge and disconnect the mains plug. Precondition: Remove the front cover.

– Loosen the plug connections on the LP frequency converter. – Loosen the connection between housing and ground.

– Remove the mounting plate from the front cover (two screws and two ball pins).

 Remove  screws  Remove  screws  Remove  ball pins  Remove  ball pins

– Loosen three outer screws (frequency converter PCB, loading side) and four countersunk screws from the rear.

 Assemble in reverse order.

(50)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 4 Disassembly / Assembly 

Operating unit (5417 C only)

– Switch off the centrifuge and disconnect the mains plug.

– Remove the dummy plug (situated below the keypad) and press out the operating unit. – Disconnect the ground cable and plug connections to the PCB of the operating unit.

 Assemble in reverse order.

– Start the centrifuge.

Operating unit (5417 R only)

– Switch off the centrifuge and disconnect the mains plug. Precondition: Remove the front cover.

– Loosen the connection between housing and ground. – Remove the plug connection (flat-band cable).

– Press out the operating unit from the left-hand side (spring plates).

(51)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 4 Disassembly / Assembly 

 Brake resistors (5417 C only)

– Switch off the centrifuge and disconnect the mains plug. Precondition: Remove the upper par t of the centrifuge.

– Loosen the plug connection (white cable).

 Loosen screws

– Loosen four screws.

 Assemble in reverse order.

(52)

Centrifuge 5417 C / 5417 R

 5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards

 4 Disassembly / Assembly 

 Brake resistors (5417 R only)

– Switch off the centrifuge and disconnect the mains plug.

Precondition: Remove the lid, remove the front cover, take off the upper part of the housing. – Loosen the plug connection (white cable).

– Remove the brake resistors incl. plate (two nuts) from the side panel.

   R  e   m   o   v   e   n   u    t  s

 Assemble in reverse order.

References

Related documents

Before your scan begins, you’ll be asked to provide information about your medical history and any previous experience with X-ray contrast (“dye”).. Preparation for a CT scan

The Nominating Committee suggests for Chairman of District IV, the name of

Vaccination with Brucella recombinant DnaK and SurA proteins induces protection against Brucella abortus infection in BALB/c mice. Cloeckaert A, Bowden

Clause 8 amends the Health Insurance (Approved Scheme) Regulations 1971 in consequence of the amendment of section 26 of the Health Insurance Act 1971 in clause 4. Clause 9 amends

Finally, a through thickness crack depth was modeled, and the stress intensity factor was obtained for several crack lengths using the Virtual Crack Closure Technique (VCCT).

Four eyes (9.1%), needed pars plana vitrectomy for the treatment of redetachment associated with proliferative vitreoretinopathy and scleral buckle revision surgery was

The proposed framework utilises colour deconvolution and the novel CAR network solely based on a single counter- stain channel with superpixels to identify the epithelial re- gions

The Office of Government Commerce (OGC) state that a framework is “an agreement or other arrangement between one of more contracting authorities and one or more economic