Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards
Contents
1
Safety
Using this manual ... ... 1-1 Warranty ... ... 1-1 User profile ... ... 1-1 Warning signs and hazard icons ... ... 1-2 Danger levels ... ... ... 1-1 Safety instructions on the device ... ... 1-2 Safety precautions ... ... 1-3
2
PACT information (blank)
3
Error lists / Troubleshooting
Error list ... ... 3-1 Warnings and directions ... ... 3-8 Errors from the service programs ... ... 3-8 Imbalance measuring system:... ... 3-8 Customer service displays ... ... 3-9
1 Temperatures (5417 R only) ... ... 3-9 2 Condenser fan (5417 R only) ... ... 3-9 3 FC parameters... ... 3-10 4 FC parameters... ... 3-10 5 Rotor recognition... ... 3-10 6 Imbalance determination... ... ... 3-11 7 Last error message... ... 3-11 8 Version... ... 3-11 9 Current error message ... ... 3-12
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards
Contents
Description of service programs ... ... 3-13 Program overview... ... 3-13 General description ...3-14 Setting the serial interface...3-14 Voltage version ... ... 3-15
Display and keypad check ...3-15 Lid function test... ... ...3-15 Initializing the EEPROM... ... 3-16 First initialization ... ... 3-16 Follow-up initializations ... ... 3-16 Switching the aggregates / temperature sensor test (5417 R only) ... 3-17 Cycle program ... ... 3-17 Rotor recognition test / rotor pulses...3-18 Frequency converter (FC) test ... ... 3-19 Imbalance calibration ...3-20 Determination of acceleration and braking times... ... 3-21 Overspeed cut-off test...3-22 Viewing the error table... ... ... 3-22
Display number of starts / number of runs ... ... 3-22 Viewing the values stored in the EEPROM... ... 3-23
4
Disassembly / Assembly
5417 R only!... ... 4-1 Centrifuge upper part (5417 C only)... ... 4-1 Upper part of housing (5417 C only) ... 4-2 Upper part of housing (5417 R only) ... 4-3 Front cover (5417 R only) ... ... 4-4 Lid... ... 4-4 Lid latch (5417 C only) ... ... 4-4 Lid latch (5417 R only) ... ... 4-5 Motor (5417 C only) ... ... 4-6 Motor (5417 R only) ... ... 4-6 Frequency converter PCB (5417 C only) ... ... 4-7 Frequency converter PCB (5417 R only) ... ... 4-8 Keypad (5417 C only) ... ... 4-9 Keypad (5417 R only) ... ... 4-9 Brake resistors (5417 C only) ... ... 4-10 Brake resistors (5417 R only) ... ... 4-11 Rotor chamber (5417 C only) ... ... 4-12
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards
Contents
5
Alignment / Adjustment (blank)
6
Diagrams
Overview 5417 C electronics ... ... 5417 810.937 Overview 5417 R electronics ... ... 5408 810.983 Control ... ... 5408 810.908
7
Spare parts list
Spare parts list ... ... 7-1 Recommended spare parts 5417 C – First equipment ... ... 7-13 Recommended spare parts 5417 R – First equipment ... ... 7-15
8
Technical data
Technical data 5417C (from S/N 13,000 onwards) ... ... 8-1 Technical Data 5417 R (from S/N 7,000 onwards) ... 8-2
9
Maintenance
Overview maintenance ... ... 9-1 Operation Qualification/Additional maintenance ... 9-2 Description of centrifuge mainte nance... 9-3 Additional maintenance for centrifuges... ... 9-8 Description of the device test for centr ifuges ... ... 9-9
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards
Contents
Hardware description – General construction ... ... 12-1 Frequency converter assembly group...12-1 Display/operation assembly group...12-1 Imbalance monitoring assembly group ... ... 12-1 Three-phase asynchronous motor ... ... 12-1 Lid latch with query keys...12-1 Various cables ... ... 12-1
Refrigerating aggregate, condenser fan, solenoid valve ... 12-1 Functional description of the assembly groups ... ... 12-2 Power supply... ... 12-2 Frequency converter assembly group ... ... 12-2 Input power supply unit with PFC control ... ... 12-2 Switching power unit for the voltage supply for the control assembly groups ... 12-2 Power unit for motor drive and brake chopper ... ... 12-3 Computer core with interface to the operating unit PCB ... 12-3 Display/operation assembly group ... ... 12-3 Rotational speed determination and rotor recognition ... 12-4 Active mechanism ... ... 12-4 Imbalance determination ... ... 12-4 Temperature determination (Ref rigerated Centrifuge 5417 R only) ... 12-5 Rotor chamber temperature ... ... 12-5 Condenser temperature...12-5 Lid magnet selection and lid query ... ... 12-5
Converter interface ...12-5 Condenser fan regulation (Refrigerated Centrifuge 5417 R only) ... 12-6
Power supply
for refrigerating aggregate, condenser fan, solenoid valve (5417 R only) ... 12-6 Refrigerating aggregate...12-6 Condenser fan ... ... 12-6 Solenoid valve ...12-6 PC interface... ... ... 12-6
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards
1 Safety
Using this manual
Read this service manual before commencing work on the device.
Please also follow the safety instructions of the corresponding operating manual.
Make sure that you have the current versions of the service and operating manuals. Check this using the version numbers under www.eppendorf-support.com and www.eppendorf.com.
Warranty
Liability in the case of incorrect maintenance, servicing or testing
The Eppendorf service par tner commissioned with carrying out maintenance, servicing or testing is responsible for ensuring that all work is performed in a professional manner.
Maintenance, servicing or testing work must not alter the condition of the device such that its safety is reduced in comparison to its original condition.
Only use accessories and original spare par ts recommended by Eppendorf for maintenance, servicing or testing as well as measuring and testing devices recommended by Eppendorf.
User profile
Requirements of specialists entrusted with maintenance, servicing or testing work The following prerequisites are required for all maintenance, servicing or testing work:
Service training provided by Eppendorf AG for the product to be maintained, repaired or tested.
Successful participation in service training measures with Eppendorf AG certification.
Specialist training with knowledge of applicable local and international standards.
Ability to assess delegated work.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards
1 Safety
Warning signs and hazard icons
Danger levels
The danger level is part of a safety note and defines the potential consequences if the information contained therein is not observed.
Safety instructions on the device
Biological hazard Danger of explosion
Risk of electric shock Risk of being crushed
General danger Material damage
Notice
DANGER Will lead to serious injuries or death. WARNING May lead to serious injuries or death. CAUTION May lead to minor to moderate injuries. ATTENTION May lead to material damage.
Notice Provides useful information.
Depiction Meaning Location
WARNING
General hazard point. Follow the operating manual.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards
1 Safety
Safety precautions
DANGER
Electric shock
Switch off the device and disconnect the power plug before opening the device, performing work on electrical connections or starting maintenance or cleaning work.
DANGER
Risk of explosion.
Do not operate the device in rooms where work is being carried out with explosive substances.
Do not use this device to process any explosive or highly reactive substances.
Do not use this device to process any substances which could create an explosive atmosphere.
WARNING
Infection by contaminated material.
Inform yourself about any potential contamination risks before starting work with the device.
Ensure that the device is decontaminated. Check the respective decontamination certificate.
Wear protective clothing (protective gloves, protective glasses) when working with the device.
WARNING
Centrifuge lid can crush. Keep hands clear.
When opening or closing the device lid, do not reach between the lid and device or into the latching mechanism of the lid.
Always open the centrifuge lid completely to prevent it from falling.
WARNING
Injuries from sharp-edged components
Sharp-edged components are exposed when the device is open.
Wear protective gloves and exercise proper caution when working on the device.
WARNING
Risk from incorrect supply voltage
Only connect the device to power sources that match the electrical specifi cations on the device's nameplate.
Only use sockets with protective earth conductor.
WARNING
Risk of injury from chemically or mechanically damaged accessories. Even minor scratches and cracks can result in serious internal material damage.
Protect all parts from mechanical damage.
Check accessories regularly.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards/ 5417 R from serial no. 7,000 onwards
1 Safety
NOTICE
If a fault occurs, objects in the immediate vicinity of the devices could get damaged.
In accordance with the recommendations of EN 61010-2-020, leave a safety distance of 30 cm clear around the device during operation.
NOTICE
Damage to electronic components from condensation.
After the device has been moved from a cool to a warmer environment, formation of condensation can occur inside the device.
Wait for at least three hours before connecting the device to the mains power supply.
Alternatively: let the device run for half an hour j ust before a short transport.
NOTICE
Damage to electronic components can result from electrostatic discharge.
Store and transport vulnerable components or modules only in antistatic or conducting packaging.
Discharge yourself before removing the component or module from its packaging. For example, use a grounding bracelet, conducting shoes, or conducting base under the device.
Hold the component or module only by the edges.
Do not touch protruding pins or conductor tracks.
Do not rub the component or module against objects, especially plastics.
NOTICE
Refrigerating system malfunction or damage to centrifuge.
Repairs to the refrigerating system must only be carried out by a refrigeration expert.
Repairs, e.g., replacement of spare parts, are carried out at your own risk.
Observe all national and local safety regulations.
NOTICE
Risk of damage to compressor and refrigerating system by tilting the device
The compressor and refrigerating system may be damaged by switching on the device when in a tilted position or by switching it on immediately after tilting.
Allow the device to stand in an upright position for at least 4 hours before switching it on.
Never turn the device upside down.
NOTICE
Damage as a result of incorrect packing.
Eppendorf accepts no warranty or liability for damage caused by incorrect packing.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
PACT-System
Product News
Application News
Commercial Info
Technical Info
Mögliche Beschädigungen an Bedienteilen der Centrifugen 5417 C
Durch Rückmeldungen über unsere Service Info Line (SIL) aus den Märkten wurden wir auf einen möglichen Fehler an neuen Bedienteilen der Centrifugen 5417C aufmerksam gemacht.
Beim Abziehen der Schutzfolie verblieben offensichtlich Kleberreste auf der Oberfläche des Display. In vielen Fällen ließen sich diese Rückstände mit handelsüblichem Waschbenzin entfernen. In einigen Fällen war jedoch die Oberfläche des Display bereits angegriffen, so dass das Bedienteil ausgewechselt werden musste.
Als Verursacher dieser Beschädigungen wurde ein Silikon-Lösungsmittel in einem Glasreiniger ermittelt. Der entsprechende Fertigungsprozess wurde bereits geändert.
Sollten Sie ein Display vorfinden, bei dem sich die Kleberreste nicht rückstandslos mit Waschbenzin entfernen lassen, berichten Sie uns diesen Fall bitte unter Angabe der Gerätenummer per SIL. Wir werden das Bedienteil umgehend austauschen.
Possible damage to control panels of the Centrifuge 5417 C
Due to feedback obtained from the markets through our Service Info Line (SIL) , a possible error involving the new control panels of the Centrifuge 5417C has been brought to our attention.
When the protective foil is pulled off, it seems that adhesive residue remains on the surface of the display. In many cases this residue can be removed with commercially available petroleum ether. However, in some cases the surface of the display had already been damaged, meaning that the control panel needed to be exchanged.
A silicone solvent in a glass cleaner has been determined to be the cause of the damage. The production process involved has already been changed.
PACT-System
Product News
Application News
Commercial Info
Technical Info
New serial number for all centrifuges
Starting with the beginning of January 2008 all newly produced centrifuges carry a
redesigned nameplate. Additionally, there is a new consistent format of the serial number.
Redesigned nameplate for centrifuges with the new 12-digit serial number (example):
Please note that, depending on the type of instrument, it can take some time until the first
devices with the new nameplate and the new serial number reach the market.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Error list
Display Error Cause Test/Solution
Error 1
Centrifuge brakes.
1. Error in measurement
system for rotational
speed. No rotor
recognized.
2. Magnet assignment of
the rotor coding is
defective.
3. Only two magnets
rec-ognized (half rotational
speed).
4. Incorrect voltage
ver-sion has been set.
5. A loose rotor has been
recognized (rotor may
have tightened during
acceleration).
Use another rotor. If the
error does not recur, check
the coding ring of the rotor.
Check whether the
mag-nets are complete and in
good condition.
Check the rotational speed
sensor with customer
ser-vice program 8:
– Four pulses per rotation;
test with service program
8 (SE 8) possible.
– Rotor ratio, test with
ser-vice program 8 or
cus-tomer service display 5.
Error 2
Centrifuge does not start or
brakes.
Imbalance idle value is
out-side the permitted limits, or
the measuring value is
incorrect.
Incorrect idle value must be
applied for at least three
seconds.
1. Sensor value < 0.5 V or
> 4.5 V; sensor is
defective.
2. Idle value is ± 220 mV
outside the stored idle
value (SE15).
Customer service program 6
shows the stored idle value
and the current idle value.
Test: Current idle value
changes when the motor
axle moves. If it does not
change, the sensor is
defec-tive or the magnet bracket is
missing.
When the centrifuge is
sta-tionary, the current idle
value must be within the
lim-its of ± 25 diglim-its (equivalent
to ± 220 mV) of the stored
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Error 3
No rotor pulses.
No rotor pulses after
start-ing. Centrifuge brakes for
ten seconds.
No rotor pulses during run.
Centrifuge brakes and is
locked for eight minutes.
No rotor; measuring system
for rotational speed is
defective or drive is
defec-tive. Error remains active
after a power supply
inter-ruption. Keep the centrifuge
switched on. After an
eight-minute blocking time, Error
3 is deleted and the lid
opening is released. Press
the Open key.
Check the state of the rotor
coding. Check the function
of the rotational speed
sen-sor by using SE 8.
Error 4
(not occupied)
Error 5
Drive switches off;
centri-fuge decelerates without
braking.
No report from lid by at
least one limit switch during
run.
Reports of lid positions:
5417 R:
1. Lid is closed –
microswitch in the lid
latch.
2. Lid is locked –
micro-switch in the lid latch.
5417 C:
1. Lid is closed – Reed
contact in the lid latch,
activated by the lid
magnet.
2. Lid is locked –
micro-switch in the lid latch.
Test of the lid signals with
service program 5 or SE 4.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Error 6
Centrifuge does not start, or
decelerates without braking.
Additional error messages
of the frequency converter
(FC) only in service displays
as last number of display 9:
0 – No error
1 – Undervoltage
2 – Overvoltage
4 – Overtemperature
Converter
7 – Overcurrent converter
8 – EEPROM FC error
10 – No release
40 – Short-circuit in power
unit
FA – Reset by watchdog
FF – Serial interface time
out
Error in the frequency
con-verter; possible thermal
overload.
1. Not ready after
switching on.
2. Error detected in
fre-quency converter
(Par. 23).
3. Error when braking.
4. Frequency converter
carried out a RESET.
Allow centrifuge to cool for
five minutes, then restart.
If necessary, check the
function of the frequency
converter (FC) with service
program 9 (SE 9).
Check the state of the brake
resistors thoroughly before
replacing a converter.
+ Repair work on the con-verter is not permitted!
Error 7
Centrifuge brakes.
1. Overspeed monitoring
is active; maximum
rotational speed of the
rotor has been
exceeded.
2. There are regulator
deviations of ± 800 rpm
from the nominal value.
Check the motor. Check the
frequency converter with
SE 9.
Display Error Cause Test/Solution
power off w/o exiting a service ro ram
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Error 8
Centrifuge brakes.
Non-compliance with
accel-eration ramps.
Ramp control has detected
a non-permissible deviation
during acceleration.
The rotor may not be
cor-rectly fastened.
The voltage version is not
set correctly. Check with
customer service display 8.
Adjust the voltage version
using SE 2.
Check the acceleration with
SE 11.
Error 9 and Error 10
EEPROM error.
Writing error (Error 9) or
reading error (Error 10) of
the EEPROM. Error is
dis-played and the standard
settings are set. If there is
an error when reading the
setting for imbalance, the
centrifuge can no longer
start (Error 2!).
Re-initialize the EEPROM
with SE 5. Alternatively,
replace the operating unit.
Error 11
No further actions possible.
Watchdog error on the
Operation/display PCB.
Replace the operating unit.
Error 12 and Error 13
Communication error of the
PC interface.
Centrifuge can be operated
manually.
Error 12, error during
trans-mission.
Error 13, reception error.
Parity error, incorrect code
number, check sum error
Check the data exchange
with the interface tester.
Check the cable
connec-tions.
Error 14
No further actions possible.
EPROM check sum error
after mains-on.
Error program memory,
EPROM defective.
Replace the operating unit.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
rotor (error 15)
Centrifuge switches off and
decelerates with braking.
Imbalance error at
n > 3,000 rpm.
The centrifuge was pushed
or the rotor was not
fas-tened.
Check installation of the
centrifuge, even surface,
stable position.
Check the motor and rotor.
Undamaged rotor sit and
motor axle. Check whether
the rotor is correctly
fas-tened and positioned.
Error 16 and Error
17
Centrifuge switches off and
decelerates without braking.
Communication error with
the frequency converter
(FC).
Error 16:
1. Error during request to
frequency converter.
2. Error after telegram
repetition.
Error 17:
Error during transmission of
select telegram to FC:
1. Queue overflow FC
telegrams.
2. Error after three
repeti-tions of a select
tele-gram.
Check the cable connection
between the operating unit
and the converter PCB.
Replace the converter or the
operating unit.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Error 18
(5417R only)
Centrifuge switches off,
braked deceleration.
Refrig-erating aggregate off.
Overtemperature of the
rotor chamber:
1. After the temperature
regulation the interval
of ± 5 °C is exceeded.
2. Chamber temperature
is more than 10 °C
higher than the start
temperature or the
nominal temperature.
3. Chamber temperature
> 50 °C.
Check the cooling system
with SE 6.
Error 19
(5417R only)
Centrifuge switches off,
braked deceleration, the
refrigeration aggregate is
switched off, the fan keeps
running.
Excess temperature of the
condenser; measured
tem-perature higher than
+ 58 °C.
Allow the refrigeration
aggregate to cool.
Check the function of the
fan.
Clean the condenser, the
fan opening must not be
blocked.
Error 20
(5417R only)
Centrifuge switches off,
deceleration with braking,
the refrigeration aggregate
is switched off.
Temperature sensor in the
rotor chamber is defective
(NTC).
Check the chamber sensor
with SE 6.
Replace the sensor.
Error 21
Centrifuge switches off,
deceleration with braking.
Temperature sensor on
con-denser is defective (NTC).
The measuring value is not
between – 8 °C and
+ 83 °C.
Replace the condenser
sen-sor.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Error 22 (5417R only)
Fan is possibly controlled
with full power.
Error in the fan control.
There is no 50(60) Hz
sig-nals for the fan control and
mains-off recognition:
1. During run and
stand-by, the fan is controlled
with full power, error
display.
2. After mains-on.
Replace the converter PCB.
Error 22 (only 5417C)
Broken mains detection
Replace the converter PCB.
Error 23
Lid is blocked until the
fre-quency converter detects a
standstill.
Irregularities during braking.
Check connections of the
brake resistors. Check cable
between operating unit and
converter.
Check function of converter
with SE 9.
Int
(Error 25)
Mains interrupted during a
run.
Check power supply: mains
connection cable, mains
input module in centrifuge,
mains plug.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Warnings and directives
+ Warnings are not stored in the EEPROM!
(Warning messages=Wxx).
clEAn ro W28
After the 200th/201st/202nd run with the installed rotor, the maintenance
advice "clean rotor" is emitted.
SPEED W29
The rotational speed set is higher than the nominal rotational speed of the rotor
in use.
The centrifuge starts, but stops running after recognizing the rotor.
InbAL W30
There is a non-permissible rotational speed of n < 2,000 rpm.
The centrifuge brakes.
Error 37
The voltage version of the frequency converter does not correspond to that of
the centrifuge.
The check is carried out after switching on the centrifuge.
Errors from the service programs
Imbalance measuring system:
Error 32
Hardware error: imbalance measuring system.
Error 33
Measuring values outside the measuring range, too little space between sensor and
magnet.
Error 34
Incorrect measuring series:
1. Mean value not within limits or distance too short.
2. Scattering run 1-3 too great.
3. Scattering run 4-6 too great.
Possible causes: The distance between the sensor and the magnet is too large or is set
incorrectly.
Error 35
Imbalance sensor idle value exceeds the setting limits of 145 to 195.
Error 36EEPROM not yet initialized; call up service program 5.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Customer service displays
For Centrifuge 5417 C from serial no. 13.000 onwards and for Centrifuge 5417 R from serial no. 7.000
onwards.
If the SOFT and SPEED-PLUS keys are pressed simultaneously, the centrifuge switches over to the
ser-vice display mode. This does not depend on the current centrifuge operational mode. The centrifugation
runs remain unaffected. The displays are run through by pressing the SOFT key. During the service
dis-play, a SOFT switchover is not possible.
In the case of error, the display automatically jumps to "current error message". It is then possible to
switch to different displays.
The display "current error message" shows the current error and the error location. For Error 6, the
con-verter error number is also shown.
1 Temperatures (5417 R only)
1
m
T
y
y
y
z
nz
z
m
Solenoid valve, A = open, – = closed
T
Regulator value (0 to 3, parameter for production)
yy.y
Actual temperature rotor chamber sensor
zz.z
Corrected nominal value = regulator value
n
Motor heating switched on in standby
2 Condenser fan (5417 R only)2
x
x
x
K
n
y
y
xx
Condenser temperature, refrigeration aggregate switches off at 75 °C
yy
Display of the regulator level, 0 to 16
K
Refrigeration aggregate E = switched on, A = switched off
n
Solenoid valve is opened.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
3 FC parameters
3
x
x
x
x
x
|
|
|
|
1 2
3 4
xxxxx
Rotational field speed
FC status
1 – FC is ready.
2 – Go-ahead is given.
3 – Timing switched on (power unit (FC) is selected).
4 – Collective error message (short-term indication, if there is a converter error,
the current error message is displayed with Function 9).
4 FC parameters
4
y
y
y
z
nx
x
yyy
Open lid:
software version FC
Closed lid: intermediate circuit voltage
z
FC mode 0/1/3/7, parameter for production
n
FC temperature warning, from approx. 5 °C below switch-off value
xx
Temperature frequency converter, switch-off at approx. 75 °C
5 Rotor recognition
5
x
x
x
y
u
v
xxx
Rotor relation
y
Rotor number
u, v
lid switch
0 = open, 1 = closed
Rotor no. Rotor model Rotor relation 5417 C 5417 R
R
0
F
45-30-11
85–132
14,000
rpm
14,000
rpm
R 0
FA 45-30-11
85–132
14,000 rpm
14,000 rpm
R 1
A-12-11
142–195
10,000 rpm
Rotor too large
R
1
A-8-11
142–195
10,000
rpm
10,500
rpm
R 1
F 45-48-PCR
142–195
10,000 rpm
10,500 rpm
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
6 Imbalance determination When the lid is open
6
x
x
x
y
y
y
xxx
Current imbalance idle value during standstill 170
±25 (-->145-195)
yyy
Stored idle value
When the lid is closed
6
x
x
y
y
z
z
xx
Max. imbalance that occurred during a run up to 3,000 rpm
yy
Max. imbalance that occurred during a run from n Š 3,000 rpm
zz
Imbalance switch-off limit for the rotor in use.
7 Last error message
7
x
x
y
y
z
z
xx
Last error
yy
Last but one error message
8 Version
8
x
x
x
x
y
y
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
9 Current error message
9
E
x
x
y
z
z
xx
Error number
y
Number for locating the cause
zz
FC error number, if the converter switches off with error 6
FC error number:0
– No error detected in FC
1
– Undervoltage
2
– Overvoltage
4
– Excess temperature converter
7
– Overcurrent converter
8
– Error
EEPROM
FC
10
– No
release
40
– Short-circuit in power unit
FA
– Reset by watchdog
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Description of service programs
Centrifuge 5417 C from serial no. 13.000 onwards and Centrifuge 5417 R from serial no. 7.000 onwards.
Program overview
Program no.Program function
1
Setting the serial address
2
Setting the voltage version
3
Checking the display and the keypad
4
Testing the lid function / lid limit switches
5
EEPROM initializing
6
Test refrigeration aggregate / fan / temperature sensors
7
Cycle test
8
Rotor recognition
9
Test frequency converter
10
Calibration of imbalance sensor
11
Determination of the acceleration and braking times
12
Overspeed cut-off test
13
Viewing the error table
14
Display of the number of starts / number of runs / operating hours
15
Display of the switching off values for imbalance / sensor idle value / voltage version
16–19
Test program for production.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
General description
The service programs are independent of the normal operating modes. It is possible to access program
selection by pressing the SOFT and TIME-PLUS keys together and simultaneously pressing the mains
switch.
Display:
S
E
x
x
xx
Service program number
The required program number can be set by pressing the TIME-PLUS or TIME-MINUS key. The lid is
opened by pressing the OPEN key. All programs are started by pressing the START key. Errors which are
displayed can be deleted by pressing the START key.
If "Error 36" appears in the display immediately after the service programs have been called up, the
EEPROM must be initialized.
The service mode can only be quit by switching off the centrifuge.
Setting the serial interface START Program 1
b
U
S
x
0 = Disable
1 = RS232
2 = Reserved for CAN bus
Press the START key to access the setting of the device address.
A
d
r
x
x
xx
Device address 20...29 (Hex)
By pressing the TIME-PLUS key, the address to be set is continuously counted upwards.
START
To confirm the selected address.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Voltage version START program 2
S
P
x
x
x
Start SE 2 The version which has been set is displayed.
SPEED-PLUS buttonSelection of the voltage version 230 – 120 – 100
SOFT buttonValue is stored in the EEPROM – display 'OK'.
STOP buttonEnd of program
Display and keypad check START program 3
1. Display checkThe sound is on for 1.5 seconds; all segments are displayed
2. Keypad checkDisplay is deleted
press any key"1" is displayed for each key
The STOP key is the last key to be pressed; this is also the end of the program.
Lid function test START program 4
L
I
d
x1
x2
x1 – bolt limit switch
1 = Lid closed 0 = Lid open
x2 – lid limit switch
1 = Lid locked 0 = Lid unlocked
2)
3)
4)
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Initializing the EEPROM
The set nominal values and the operating conditions are stored in the EEPROM after each run of the
centrifuge. The limits for switching off (imbalance) and statistical data are also stored in the EEPROM.
First initialization
Error 36 is indicated if no identification is found in the EEPROM when entering the ser vice mode.
The error can be erased with the START button.
START program 5
I
n
I
T
0
Presetting rotational speed = 14,000 rpm, rcf = 20,800, radius = 89 mm,
time = 10 min.,
temp. = 18 °C, imbalance idle value = 170,
imbalance = 17,
interface adr. = 20H, inactive,
all tables are erased
SOFT button
initialization is carried out,
STOP button
no initialization, return to program selection
4
1
4
0
0
0
1
0
+ Always carry out imbalance calibration after the first initialization!
Follow-up initializations START program 5
I
n
I
T
1
STOP buttonno initialization, return to program selection
SOFT button
initialization is carried out, calibration data for imbalance,
voltage version and interface setting remain intact,
all other data is deleted.
Possible errors during initialization: Error 9, EEPROM cannot be inscribed correctly.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Switching the aggregates / temperature sensor test (5417 R only) START program 6
A
x
U
y
L
z
s
s
Switching the aggregatesTEMP_PLUSrefrigeration aggregate x = 1 agg. on / x = 0 = agg. off
SPEED-MINUS solenoid valve y = 1 valve open / y = 0 valve closed
SPEED-PLUS
condenser fan Z = 1 fan on / Z = 0 fan off
ss – fan step (0..16), the step can be changed by means
of the TIME-PLUS and MINUS keys
FC key
·To switch over to the temperature display
y
y
y
x
x
x
xx
Condenser temperature
yyy
Rotor chamber temperature
STOP
End
Cycle program START program 7
This program is used for a continuous centrifuge test. The SHORT, SOFT and FC functions are available.
Specifications:
5417 C: 4 runs of 1 minute each; interval of 1 minute; 10-minute interval after the 4th run.
5417 R: 4 runs of 10 minutes each; interval of 1 minute; 10-minute interval after the 4th run.
P
A
U
S
E
x
x
The long interval can be set gradually with the SOFT button in limits of between 1 minute and
20 minutes. (Standard: 10 minutes).
The nominal values of the run can be changed after the service program has started. The operating
6) 7)
A toggle
U toggle
L
(yy.y)
(xx.x)
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
START program 7
4
1
4
0
0
0
1
0
Display of the preset nominal values. Alterations are possible.
START
To begin the cycle run
Display after each run:
P
A
U
S
E
x
x
xx
Interval time, in minutes or seconds
An individual run can be aborted by pressing the STOP key. The lid can be opened during the interval.
The cycle program can be quit by pressing the SHORT key.
Rotor recognition test / rotor pulses START program 8
r
o
1
2
3
4
Slowly turn the rotor by hand; a digit is displayed for each pulse (4 pulses per rotation!)
Press the SPEED_PLUS key: Calculation of rotor ratio and rotor number
r
o
x
x
x
z
xxx
Rotor ratio
z
Rotor number 0, 1, 2
SE 8 is ended via STOP.
8)
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Frequency converter (FC) test
Both lid limit switches must be closed during this test, otherwise the frequency converter is not released.
Safety precautions:
Do not insert a rotor if the test is car ried out with an open centrifuge or without the centri-fuge lid!
START program 9
After the program has started, an indication that the centrifuge is ready to use must appear in the
dis-play.
TIME-PLUS keyRun with approx. 120 rpm
TIME-MINUS keyBraking
F
U
b
x
x
x
z
F
U
E
r
6
y
y
b
Converter ready
z
1 = FC started
xxx
Rotational field frequency
yy
FC error code in case of error 1 – Undervoltage
2 – Overvoltage
4 – Overtemperature converter
7 – Overcurrent converter
8 – Error EEPROM FC
10 – No release
40 – Short-circuit in power unit
FA – Reset by watchdog
FF – Serial interface time out
STOP key or SHORT key: the FC is switched off; centrifuge returns to program selection.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
3 Error lists / Troubleshooting
Imbalance calibration
General information:
Imbalance calibration requires two calibration weights. Two reaction tubes of 1.5 ml or 2.0 ml are filled
and balanced with the corresponding material (e.g. quartz sand).
The tube for the lower calibration value must have a total weight of 2.3 g ± 0.1 g. The total weight of the
vessel for the upper calibration value is 3.8 g ± 0.1 g.
If the imbalance calibration is carried out with fixed-angle rotor F 45-30-11 or FA 45-30-11 (rotor number
R0), the determined limits are converted for the swing-bucket rotor and the PCR rotor (rotor number R1)
as well as the 24-position rotor (rotor number R2).
However, if the calibration is carried out with a different rotor than the 30-position rotor, the determined
values are valid for this rotor only; the values cannot be carried over to another rotor!
The imbalance calibration is carried out without rotor lid.
•
Install a flyweight of 2.3 g (lower value) into the rotor.
•Close the centrifuge lid.
START program 10
–
Three runs are carried out and the measured value imbalance value is displayed,
–
The lid automatically opens after the third run.
–
Install a flyweight of 3.8 g (upper value).
–
Close the centrifuge lid, Three runs are repeated.
–
Calculation and display of the switch-off value (it is stored in the EEPROM).
–
Start
testrun. In the display "Inb
AL" is displayed. The service function is leaving.
r
u
n
y
x
x
y
Run
number
xx
Measured imbalance value
E
n
d
x
x
xx
Calculated switch-off value
START
Test run with the 3.8 g flyweight; the centrifuge must display "I
nbAL" and then switch off
OPEN
To open the centrifuge lid.
SHORT
To abort the calibration in progress; return after the machine has come to a standstill.
Error displays:
Error 33
Measuring values outside the measuring range; too little distance between sensor and
magnet.
Error 34
Incorrect measuring series (deviation is too large or the space between the measuring
Centrifuge 5417 C / 5417 R
Centrifuge 5417 C / 5417 R
5417 C from seria
5417 C from serial no. 13,000 onw
l no. 13,000 onwards / 5417 R from se
ards / 5417 R from serial no. 7,000 onw
rial no. 7,000 onwards
ards
3
3 Error lists / T
Error lists / Troubleshooting
roubleshooting
Determination of acceler
Determination of acceleration and braking tation and braking t imesimes
•
•
Load the rotor completely.
Load the rotor completely.
••
Close the lid.
Close the lid.
START program 11 START program 11
3
3
0
0
1
1
4
4
0
0
0
0
0
0
1
1
0
0
S
ST
TA
AR
RT
T
T
To
o b
beeg
giin t
n thhe
e m
meeaassuurriinng
g rruunn..
After the rotor has been reco
After the rotor has been recognized, the measuring rotational
gnized, the measuring rotational speed is calculated b
speed is calculated by the
y the
program = 97 % of the max. rotational speed of the rotor used.
program = 97 % of the max. rotational speed of the rotor used.
Display of the acceleration time in seconds (x10)
Display of the acceleration time in seconds (x10), if the
, if the measuring rotational speed was reached
measuring rotational speed was reached
H
H
L
L
x
x
x
x
x
x
ST
STO
OP
P
T
To
o st
star
art
t of
of th
the
e m
mea
eassur
urem
emen
ent
t of
of th
the
e br
brak
akin
ing
g ti
tim
mee
Display of the braking time in seconds (x10), if the centrifuge has
Display of the braking time in seconds (x10), if the centrifuge has come to a standstill.
come to a standstill.
b
b
L
L
x
x
x
x
x
x
SH
SHOR
ORT
T
T
To r
o ret
etur
urn t
n to t
o thhe p
e pro
rogr
graam s
m seele
lect
ctio
ionn
Acceleration and
Acceleration and braking times
braking times depend on
depend on the rot
the rotor model,
or model, load and
load and voltage
voltage version. The
version. The average
average values
values
list
listed below are valid for 30-position rotor F 45-30-11 with
ed below are valid for 30-position rotor F 45-30-11 with the maximum load:
the maximum load:
V
Voollttaagge ve veersrsiioonn AAcccceelleerraattiioon tn tiimme (e (sseecc)) BBrraakkiinng tg tiimme (e (sseecc))
22330
0
V
V
77..55––1100..55
99..00––1111..55
11115
5
V
V
1100..00––1133..55
1111..00––1155..00
11000
0
V
V
1100..00––1133..55
1111..00––1155..00
S
So
offt
t rraam
mp
p
aap
pp
prro
oxx.
. 660
0 sseecc
aap
pp
prro
oxx.
. 660
0 sseecc
11)
Centrifuge 5417 C / 5417 R
Centrifuge 5417 C / 5417 R
5417 C from seria
5417 C from serial no. 13,000 onw
l no. 13,000 onwards / 5417 R from se
ards / 5417 R from serial no. 7,000 onw
rial no. 7,000 onwards
ards
3
3 Error lists / T
Error lists / Troubleshooting
roubleshooting
Overspeed cut-off test Overspeed cut-off test
•
•
Close the lid
Close the lid
START program 12 START program 12
A rotational speed abo
A rotational speed above the cut-off rotational speed is set b
ve the cut-off rotational speed is set by the software. If the centrifuge is OK, it
y the software. If the centrifuge is OK, it
switches itself off via Error 7
switches itself off via Error 7 when the cut-off speed has been reached and then decelerates with brak-
when the cut-off speed has been reached and then decelerates with
brak-ing.
ing.
If there is no cut-off, the run m
If there is no cut-off, the run m ust be immediately ust be immediately stopped by hand! stopped by hand!
The cut-off rotational speeds are 300 rpm above the nominal speed of the rotor in use. The cut-off rotational speeds are 300 rpm above the nominal speed of the rotor in use.
--> SHORT: Leaving the Ser
--> SHORT: Leaving the Service-Functio
vice-Functionn
Viewing the error table Viewing the error table
The error that occurred first during a run
The error that occurred first during a run is stored.
is stored.
START program 13 START program 13
Display of the error number and the number of errors during a run
Display of the error number and the number of errors during a run
S
SO
OF
FT
T
T
To
o ttrraannssffeer
r tto
o tthhe
e sstteep
p m
mo
od
dee..
S
SH
HO
OR
RT
T
T
To c
o co
onnttiinnuue s
e stteep
p--b
byy--sstteep
p..
S
ST
TO
OP
P
T
To
o aab
bo
orrt
t tthhe
e p
prro
og
grraam
m..
E
E
r
r
x
x
x
x
y
y
y
y
xxxx
E
Errrro
or
r nnuum
mb
beerr
yyyyyy
N
Nuum
mb
beer
r o
of
f eerrrro
orrss
Display of number of star
Display of number of starts / number of runsts / number of runs START program 14
START program 14
x
x
x
x
x
x
x
x
x
x
E
E
A
A
xxxxxxxx
N
Nuum
mb
beer
r o
of
f ssttaarrttss
S
ST
TA
AR
RT
T
··next display
next display
x
x
x
x
x
x
x
x
x
x
C
C
L
L
xxxxxxxx
N
Nuum
mb
beer
r o
of
f cceennttrriiffuug
ge
e rruunnss
START
START
··next display
next display
x
x
x
x
x
x
x
x
x
x
h
h
12) 12) 13) 13) 14) 14)Centrifuge 5417 C / 5417 R
Centrifuge 5417 C / 5417 R
5417 C from seria
5417 C from serial no. 13,000 onw
l no. 13,000 onwards / 5417 R from se
ards / 5417 R from serial no. 7,000 onw
rial no. 7,000 onwards
ards
3
3 Error lists / T
Error lists / Troubleshooting
roubleshooting
Viewing the values stored in the EEPROM Viewing the values stored in the EEPROM START program 15
START program 15
, then
, then continue switching all displays with
continue switching all displays with ST
START
ART
x
x
x
x
x
x
U
U
y
y
yy
R
Ro
otto
or
r nnuum
mb
beerr
xxxxxx
S
Sw
wiittcchh--o
off
ff vvaalluuees
s ffo
or
r aalll
l rro
otto
orrs
s ((nno
o.
. 00,
, 11,
, 22))
x
x
x
x
x
x
U
U
rr
xxx
xxx
Idle
Idle value
value imbalance
imbalance sensor
sensor –
– measurement
measurement at
at start
start of
of imbalance
imbalance calibration.
calibration.
x
x
x
x
x
x
S
S
P
P
xxxxx
x
V
Vo
ollttaag
ge
e vveerrssiio
onn
y
y
y
y
X
X
X
X
XX
XX
P0
P0, P
, P1,
1, PL
PL =
=T
Teest
st w
wor
ord
ds f
s for
or cche
heccki
king
ng th
the t
e teest
st pr
pro
oce
cess
ss du
duri
ring
ng tthe
he p
pro
rod
duc
ucti
tio
on;
n;
not for Service!
not for Service!
15)
15)
(0, 1, 2)
(0, 1, 2)
(displays each consecutively)
Centrifuge 5417 C / 5417 R
Centrifuge 5417 C / 5417 R
5417 C from seria
5417 C from serial no. 13,000 onw
l no. 13,000 onwards / 5417 R from se
ards / 5417 R from serial no. 7,000 onw
rial no. 7,000 onwards
ards
3
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
4 Disassembly / Assembly
5417 R only!
Do not place the device on its side. If the centrifuge has been tilted, leave the device in its normal position for at least four hours in order to prevent dam age to the compressor!
5417C and 5417R:
Each time the PCB, motor or imbalance angle is replaced or disassembled, align the imbalance sensor for all available rotors!
Centrifuge upper part (5417 C only)
– Switch off the centrifuge and disconnect the mains plug. – Open the lid.
– Disassemble the rotor.
– Remove the motor cap (undo the four Phillips screws). – Unscrew the two rear feet of the device.
– Unscrew the two front feet of the device. When doing so, hold together the upper and lower parts. – For further disassembling, place the upper part of the centrifuge on its right-hand side.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
4 Disassembly / Assembly
Upper part of housing (5417 C only)
– Switch off the centrifuge and disconnect the mains plug. – Open the lid.
Precondition: Remove the lid. – Disassemble the rotor.
– Remove the Phillips screws (on the top side of the housing).
Remove
Phillips screws
Remove
Phillips screws
– Remove the dummy plug (situated below the keyboard) and press out the keypad.
– Disconnect the ground cable and the plug connections to the PCB of the operating unit. – Remove the housing by pulling it upwards.
Assemble in reverse order.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
4 Disassembly / Assembly
Upper part of housing (5417 R only)
– Switch off the centrifuge and disconnect the mains plug. – Open the lid.
Precondition: Remove lid. – Disassemble the rotor.
– Undo the two screws (top, rear) on the upper part of the housing.
– Remove the front cover by undoing the two screws on the bottom side of the housing (if necessary, move the centrifuge slightly over the table edge) (ball-pin fastening).
– Disconnect the ground cable and the plug connections to the PCB. – Remove the plastic button for emergency release.
– Unscrew the cover on the rear. – Undo the screws at the front.
R e m o v e s c r e w s R e m o v e s c r e w s
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
4 Disassembly / Assembly
Front cover (5417 R only)
– Switch off the centrifuge and disconnect the mains plug.
– Remove the front cover by undoing the two screws on the bottom side of the housing (if necessary, move the centrifuge slightly over the table edge) (ball-pin fastening).
– Disconnect the ground cable and the plug connections to PCB.
Assemble in reverse order.
Lid
– Loosen the four Phillips screws on the inside of the lid. – Remove the upper part of the lid.
– Unhook the torsion springs.
– Remove the snap ring from the left-hand side of the axle.
– Remove the axle from its mounting by slightly moving it to the side. When doing so, pay attention to the bushes.
– Remove the lid plate.
Assemble in reverse order.
Lid latch (5417 C only)
– Switch off the centrifuge and disconnect the mains plug.
Precondition: Remove the lid, disassemble the upper part of the housing. – Undo the two Phillips screws (at the front of the top side).
Remove Phillips screws
Remove
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
4 Disassembly / Assembly
– Pull out the lid latch towards the front. – Loosen the plug connections.
Assemble in reverse order.
– Start the centrifuge.
Lid latch (5417 R only)
– Switch off the centrifuge and disconnect the mains plug.
– Precondition: Remove lid, remove the front cover, disassemble the upper part of the device. – Loosen the two Phillips screws (at the front of the top side).
Remove Phillips screws Remove Phillips screws
– Pull out the lid latch downwards.
– Loosen plug connection from frequency converter PCB.
Assemble in reverse order.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
4 Disassembly / Assembly
Motor (5417 C only)
– Switch off the centrifuge and disconnect the mains plug. Precondition: Remove the upper par t of the centrifuge.
– Disconnect the plug connection and the ground wire.
– Loosen the motor mountings (three screws on the bottom of the base plate). – Take out the motor.
When replacing the oscillating metals (they may be replaced as a complete set of three only) install the torsion protection.
Assemble in reverse order.
– Start the centrifuge.
Motor (5417 R only)
– Switch off the centrifuge and disconnect the mains plug. – Open the lid.
– Disassemble the rotor. – Unscrew the motor cap.
Precondition: Remove the front panel.
– Loosen the plug connections on the frequency converter PCB. – Place the device on its side.
– Remove the tray for condensation water by pulling it out towards the front. – Loosen the motor fastening (three screws).
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
4 Disassembly / Assembly
– Remove the motor by pulling it downwards. – Unscrew the motor plate (three screws).
When replacing the oscillating metals (they may be replaced as a complete set of three only install the torsion protection.
Assemble in reverse order.
– Start the centrifuge.
After installing the new motor, check the function and rotation dir ection using ser vice function 9! The sleeve must firmly fit over the edge of the rotor chamber, otherwise it may slip off during operation!
5417C and 5417R:
Each time the PCB, motor or imbalance angle is replaced or disassembled, align the imbalance sensor for all available rotors!
Frequency converter PCB (5417 C only)
– Switch off the centrifuge and disconnect the mains plug. Precondition: Remove the upper part of the centrifuge.
– Loosen plug connections on the frequency converter PCB.
– Remove three screws on the top side (two on the left-hand side, one on the right-hand side).
e s c r e w s Remove screws
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
4 Disassembly / Assembly
– Undo four screws (on the bottom of the base plate). – Take out the frequency converter PCB.
Assemble in reverse order.
– Start the centrifuge.
Frequency converter PCB (5417 R only)
– Switch off the centrifuge and disconnect the mains plug. Precondition: Remove the front cover.
– Loosen the plug connections on the LP frequency converter. – Loosen the connection between housing and ground.
– Remove the mounting plate from the front cover (two screws and two ball pins).
Remove screws Remove screws Remove ball pins Remove ball pins
– Loosen three outer screws (frequency converter PCB, loading side) and four countersunk screws from the rear.
Assemble in reverse order.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
4 Disassembly / Assembly
Operating unit (5417 C only)
– Switch off the centrifuge and disconnect the mains plug.
– Remove the dummy plug (situated below the keypad) and press out the operating unit. – Disconnect the ground cable and plug connections to the PCB of the operating unit.
Assemble in reverse order.
– Start the centrifuge.
Operating unit (5417 R only)
– Switch off the centrifuge and disconnect the mains plug. Precondition: Remove the front cover.
– Loosen the connection between housing and ground. – Remove the plug connection (flat-band cable).
– Press out the operating unit from the left-hand side (spring plates).
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
4 Disassembly / Assembly
Brake resistors (5417 C only)
– Switch off the centrifuge and disconnect the mains plug. Precondition: Remove the upper par t of the centrifuge.
– Loosen the plug connection (white cable).
Loosen screws
– Loosen four screws.
Assemble in reverse order.
Centrifuge 5417 C / 5417 R
5417 C from serial no. 13,000 onwards / 5417 R from serial no. 7,000 onwards
4 Disassembly / Assembly
Brake resistors (5417 R only)
– Switch off the centrifuge and disconnect the mains plug.
Precondition: Remove the lid, remove the front cover, take off the upper part of the housing. – Loosen the plug connection (white cable).
– Remove the brake resistors incl. plate (two nuts) from the side panel.
R e m o v e n u t s
Assemble in reverse order.