These particles have something in common
Algae
Chromosomes Blood cells
Protozoa
Certain parameters of these particles can be measured with a flow
cytometer
Which parameters can be measured?
the relative size (Forward Scatter - FSC)
the granularity or complexity (Side Scatter
-SSC)
Characteristics of FSC and SSC
Coherent lightsource (488 nm) Forward scatter Cell size (488 nm) Side scatter Granularity (488 nm) Side scatter (SSC)measured in 90° direction to the excitation light
proportional to cell „complexity“ or granularity
Forward scatter (FSC)
measured along the axis of the incoming light
proportional the the cell size / cell surface (only true for perfect round cells)
An example of light scatter:
Forward scatter Side scatter Granulocytes Debris Lymphocytes MonocytesFluorescence
λ=488 nm
Excitation light
λ=530 nm
Emission light
The fluorochrome molecule absorbes the energy of the incoming light It releases the absorbed energy by:
vibration and dissipated heat
Fluorescence intensity
FITC
FITC
101 102 103 104
Relative fluorescence intensity
Number of Events FITC FITC FITC F IT C FITC FITC FITC FITC
Parts of a flow cytometer
• Fluidics
– Provide a constant stream of sheath
– Transport the sample to the interrogation point – Arrange and focus the cells to the laser intercept
• Optics
– Focus the excitation light – Collect the emitted light
•
Electronics
– Convert the optical signals into electronic signals – Send the signals to the analysis computer
• Computer
– Display data graphically – Control instrument settings
What a flowcytometer is
Very basically, a flow cytometer is an
automated fluorescence microscope (in
fact, that is how the first prototype
instruments looked like).
Like a microscope, some adjustments have to
be made to optimally illuminate and collect
the light.
The basic microscope
In a standard
microscope, the
operator uses the
XY-stage to screen the
sample and detect
cells of interest.
The automated Microscope
Waste
Detector
& Counter
This primitive diagram shows the principle: Cells are
passing the microscope objective, and an electronic circuit decides whether the cells is fluorescent or not. This is how a flow cytometer works!
Basic fluidics of the FACSAria
Plenum
CuvetteWaste
Sheath
Fluidics Cart
Pressure
Sample tubeHydrodynamic focussing in the cuvette
Sample Sample 1 Sheath Sheath Sample pressure low, small core stream. Good for DNA analysis High sample pressure, broader core stream. Bad for DNA analysisSummary
• Pressure (= Sheath Pressure) drives the sheath buffer through the
cuvette, and the higher pressure in the sample tube
(= Sample Differential) delivers the sample to the cuvette.
• In the cuvette the principle of hydrodynamic focussing arranges the
cells like pearls on a string before they arrive at the laser interception
point for analysis
• Hydrodynamic focussing cannot separate cell aggregates! Flow
cytrometry is a technique that requires single cell suspensions
Basic optics
• Somehow the light from the laser(s) must be directed to the
cuvette to illuminate the cells.
• At the same time, the emitted light must be collected to
analyse the signals from the cells.
Basic optics
A system of prisms and
lenses directs the laser light
to the interrogation point in
the cuvette
Basic Optics
The emitted light induced from each laser is focussed onto
separate glass fibers.
Optical filters
460 500 540 460 500 540 460 500 540 SP 500 SP 500 LP 500 LP 500 BP500/80BP500/80 Longpass Shortpass BandpassOctagon Detection System
PerCP-Cy5.5
FITC
695/40 655 LPSSC
PE
PE-Cy7
Summary
• Excitation light is steered with prisms and lenses to the
interception point
• Emitted light is collected using lenses and is split up
with dichroic mirrors and filters
Tasks for the electronical system
Convert the optical signals into electonic signals
(voltage pulses)
Digitise the data
Analyse Height (H), Width (W) and Area (A) of
the pulse
How a voltage pulse from the PMT is generated
Voltage Laser Laser Laser Laser Laser Laser t t t Voltage Voltage 1. 2. 3.Height, Area, and Width
Time (µs) Voltage Pulse area(A) Pulse H e ight (H) Pulse Width (W) 40 0Digitization
The pulses go into a digitizing system that scans the pulses with a rate of 10 MHz, which corresponds to a sample each 0,1 µs.
The pulse height of each slice is converted to a 14 bit number.
10 MHz scanning
Parameter Calculation
After the pulse has been digitized, the system can calculate the area and the width and apply the compensation to all signals.
Basically, the system works vice versa like your CD player at home. It generates the
Threshold
The threshold defines the minimal signal intensity which has to be
surpassed on a certain channel. All signals with a lower intensity are not displayed and not recorded for later analysis.
Summary
During passing the laser voltage pulses are generated
at the PMT
Amplifiers enhance the signals
The electronics digitizes the pulse using 10MHz
sampling
Only signals passing the desired threshold(s) are
analysed and recorded
The data are finally passed to the analysis computer
connected to the cytometer
Instrument settings
the exact values for PMT voltages and thresholds are depending on the applications (type of cells, staining methods) and the specific instrument.
Displaying the data in a linear fashion or using the logarithmic form is also depending on the application.
the exact values for PMT voltages and thresholds are depending on the applications (type of cells, staining methods) and the specific instrument.
Displaying the data in a linear fashion or using the logarithmic form is also depending on the application.
Workstation
• The connected workstation is designed for instrument control, data acquisition, -storage and -analysis.
• OS is Windows2000 Professional running on a IBM-compatible computer platform.
•Software
• DiVa application: Instrument connectivity, Data-acquisition and analysis system
• DiVa Data Manager: Backup and Restore the database.
• The connected workstation is designed for instrument control, data acquisition, -storage and -analysis.
• OS is Windows2000 Professional running on a IBM-compatible computer platform.
•Software
• DiVa application: Instrument connectivity, Data-acquisition and analysis system
Data saving
All data are saved directly into a special database. Every plot is connected with its corresponding datafile. All tubes carry a copy of the instrument setting that was active during
acquisition.
Due to this, there are no special save commands in the software. Every action is recorded in the database. When you quit and re-start the software, it will open the last
experiment exactly at the position you left it.
All data are saved directly into a special database. Every plot is connected with its corresponding datafile. All tubes carry a copy of the instrument setting that was active during
acquisition.
Due to this, there are no special save commands in the software. Every action is recorded in the database. When you quit and re-start the software, it will open the last
Visualization of data
Visualization of data
FSC SSC FL1 FL2 30 60 638 840 100 160 245 85 300 650 160 720 Listmode file 2) Dotplot - two parameter are plotted onX and Y Event 1 Event 2 Event 3 400 800 1000 0 840 85 245 638 FL1-H FL2-H 0 200 400 600 800 1000 200 600
Enough theory of flow!
Let`s have a look at an example from real life
Enough theory of flow!
Example: Determine the percentage of CD3, CD4, and CD8
populations from whole blood
Material
• Mouse splenocytes
Method
• Three-colour immunofluorescence
Preparation
• Staining of freshly isolated splenocytes
Stainings
• Isotype controls
• Single-colour stainings for FITC, PE, PerCP und CD3-APC to determine suitable instrument settings
Material
• Mouse splenocytes
Method
• Three-colour immunofluorescence
Preparation
• Staining of freshly isolated splenocytes
Stainings
• Isotype controls
• Single-colour stainings for FITC, PE, PerCP und CD3-APC to determine suitable instrument settings
Proper adjustment of FSC and SSC voltage
• FSC und SSC are optimally
adjusted when the population of interest (i.e. Lymphocytes) can be resolved from all other
populations
• The threshold on FSC is
adjusted so that most of the
debris is excluded from the data acquisition.
• FSC und SSC are optimally adjusted when the population of interest (i.e. Lymphocytes) can be resolved from all other
populations
• The threshold on FSC is adjusted so that most of the
debris is excluded from the data acquisition.
Parameters (I)
• FSC and SSC
are depending on cell type and cell state (activated, resting)
depend on the preparation method (Ficoll, LW, LNW, fixation method etc.)
are normally used to define the population of interest for further analysis
• FSC and SSC
are depending on cell type and cell state (activated, resting)
depend on the preparation method (Ficoll, LW, LNW, fixation method etc.)
are normally used to define the population of interest for further analysis
Parameters
• Fluorescence channels (FL1, FL2, FL3, FLX)
depending on the specific staining (conjugate) antibodies, propidium iodide for DNA-labelling, etc.)
most of the time fluorescence serves as marker for the statistical analysis
• Fluorescence channels (FL1, FL2, FL3, FLX)
depending on the specific staining (conjugate) antibodies, propidium iodide for DNA-labelling, etc.)
most of the time fluorescence serves as marker for the statistical analysis
Defining the population of interest
About „Gating“
• selectively analyse defined cell populations
• Gates can be set manually or automatically by software • multidimensional gating with hierarchical gates
• too narrow gates may lead to the loss of cell populations • too wide gates enhance the number of unwanted cells
• during analysis of the desired cell population the cells in the gate are considered to be the 100%
• selectively analyse defined cell populations
• Gates can be set manually or automatically by software
• multidimensional gating with hierarchical gates
• too narrow gates may lead to the loss of cell populations
• too wide gates enhance the number of unwanted cells
• during analysis of the desired cell population the cells in the gate are considered to be the 100%
Adjusting the fluorescence settings
A) Adjusting PMT voltages Sample: Isotype control
• The observed fluorescence is considere to be unspecific background fluorescence,
• Setup is done „gated“ on the lymphocyte population
• Try to put the background into the first decade (only a rule of thumb!)
B) Defining quadrants
Traditionally, a „Quadrant“ is set to define the possible four populations in two-colour experiment. Later we will see that quadrants are not the appropriate way for multicolour analyses.
Theory of quadrant analysis
FL1-H FL2-H FITC + PE FITC PE negative Q2 Q4 Q3 Q1Real life:
FITC-fluorescence overspill
650nm 700nm 500nm 550nm 600nm Relative Int ensität Wellenlänge (nm) FL1 530/30 FL2 585/42FITC Compensation
Detektor - … % Signal FL1 530/30 FL2 585/42 650nm 700nm 500nm 550nm 600nm Relative Int ensität Wellenlänge (nm)FITC Compensation
Lowering the FITC-population is achieved by...
... Subtracting a percentage of
FITC-intensity from the affected PE-channel ... 650nm 700nm 500nm 550nm 600nm FL1 530/30 585/42FL2 R e la ti v e I n te ns
ity … because 25% of the
FITC-signal are actually detected in the PE
channel ...
PE-fluorescence overspill
650nm 700nm 500nm 600nm Wavelength (nm) 550nm Rel a tiv e Intensity FL3 größer 650 FL1 530/30 FL2 585/42Automatic Multicolour
Compensation
•
Multicolour compensation with more than three colours can
become very time-consuming because each channel has to be
compensated against each other.
• Automatic compensation offers the possibility to run
single-color controls and let the software calculate all overspills.
• Mathematical calculation results in the correct spillover
values for all channels. However, to the user the visual data
may look undercompensated. This will be discussed in detail
during the training course.
Summary
What we have seen:
• the emission spectra of common fluorochromes (FITC, PE)
• the spectral overlap of fluorochromes into neighbouring channels
depending on the emission spectra and filtersets
• how spectral overlap can lead to misinterpretation of multicolour stainings
• How compensation can correct the spectral overlap of fluorochromes
What we have seen:
• the emission spectra of common fluorochromes (FITC, PE)
• the spectral overlap of fluorochromes into neighbouring channels
depending on the emission spectra and filtersets
• how spectral overlap can lead to misinterpretation of multicolour stainings • How compensation can correct the spectral overlap of fluorochromes