No.
2B730-670E
OPERATION MANUAL
FOR
DIAGNOSTIC ULTRASOUND SYSTEM
MODEL SSA-530A
[FUNDAMENTALS]
(2B730-670E)
IMPORTANT!
Read and understand this manual before operating
the equipment. After reading, keep this manual in an
easily accessible place.
Introduction
This operation manual describes the operating procedures for the diagnostic ultrasound system SSA-530A. To ensure safe and correct operation of the system, carefully read and understand the manual before operating the system.
IMPORTANT!
1. No part of this manual may be copied or reprinted, in whole or in part,
without written permission.
2. The contents of this manual are subject to change without prior notice
and without our legal obligation.
REVISION RECORD
REV. DATE REASON PAGE SER. DOC.
(MM/YY) /AUTHOR CHANGED No. PRODUCT.
Intellectual Property
Availability of This Software and Related Documents is
Restricted.
The software used for this system is licensed to Toshiba Medical Systems Corporation by a Licensor.
(1) The software and related documents must be used only for this system.
(2) The intellectual property of this software and related documents is not assigned to you. (3) You must not copy the software or documents, nor modify the software in whole or in part. (4) You must not recompile or reassemble the software.
(5) You must not assign, disclose, transfer, or sublicense the software or documents to a third party.
(6) The software is subject to the U.S. and Japanese Export Administration Laws and Regulations and you must not export or re-export the software in whole or in part unless properly authorized by the U.S. or Japanese government.
(7) The information in the documents, or programs in the software are subject to change without notice.
Organization of the Operation
Manuals
1. Notation
Conventions
In this operation manual, the following words are used in addition to the signal words related to the safety precautions (refer to "Safety Precautions"). Please read this operation manual before using the system.
NOTE: Indicates information of interest to users of system as to exceptional conditions or operating procedures.
2. Operation
Manuals
A TOSHIBA service person or instructor will explain the basic operating procedures for this system at the time of delivery. However, read this operation manual carefully before using the system in order to understand the detailed operating procedures, functions, performance, and maintenance procedures. The organization of the documents supplied with this system is shown below:
Main unit operation manual
Transducer operation manuals
Describes detailed system information on preparation, operating procedures, maintenance checks, and functions.
Describe the operating and sterilization procedures for transducers.
NOTE: For certain applications, the following manuals are available in English:
• 2B730-672E (Application volume) • 2B730-674E (Acoustic power data)
Table of Contents
Introduction
...C-1Intellectual Property
... L-1Organization of the Operation Manuals
...U-1Safety Precautions
...S-1Important Information
...W-1Outline of the SSA-530A
1.
Intended Use and Specifications
... 1-11.1 Intended
Use
... 1-11.2 Specifications
... 1-12. System
Configuration
... 2-12.1 Standard
Configuration
... 2-12.2
List of Optional Devices
... 2-12.3
List of Available Transducers
... 2-23. System
Overview
... 3-13.1
Name of Each Part
... 3-13.2 Main
Panel
... 3-23.3 Side
Panel
... 3-43.4 Rear
Panel
... 3-53.5 Power
Panel
... 3-6Operation
4.
Preparation for Examination
... 4-14.1
Moving the System
... 4-14.2
Connecting the Transducer
... 4-25.
Connecting the Power Cable
and Protective Earth
... 5-16. Power
ON/OFF
... 6-16.1 Power
ON
... 6-16.2 Power
OFF
... 6-16.3
Preparation for Use During an
Operation or for Emergency Cases
... 6-27.
Checks Before and After Use
... 7-17.1
Check Before Turning ON the Power
... 7-17.2
Check After Turning ON the Power
... 7-28.
Basic Screen and Menu
... 8-18.1
Display of Various Data Items
... 8-18.2
Display and Operation of the Menu
... 8-29.
Entering the Patient ID
... 9-19.1
Entering and Saving Data on the
Patient ID Registration Screen
... 9-210. Selecting
Presets
... 10-110.1 Introduction
... 10-110.2
Procedure for Selecting Presets
... 10-111.
Display and Operation in Each Mode
... 11-111.1 B
Mode
... 11-111.2 M
Mode
... 11-512. Cine
Function
... 12-112.1 Introduction
... 12-112.2 Cine
Playback
... 12-113.
Display of Reference Signals
... 13-113.1 Display
Procedures
... 13-113.2
Adjustment of Reference Signals
... 13-214.
Body Mark Entry and Deletion
... 14-114.1
Display of a Body Mark
... 14-114.2
Selection of a Body Mark
... 14-114.3
Movement of the Transducer Mark
... 14-214.4
Movement of the Body Mark
... 14-214.5
Deletion of the Body Mark
... 14-215.
Entry and Deletion of Comments
... 15-115.1
Entry from the Keyboard
... 15-115.3
Entry of the Arrow Mark
... 15-215.4
Entry of Gender Symbols ( , )
... 15-215.5
Moving the Selected Comment
... 15-315.6
Correcting the Comments
... 15-315.7
Deletion of the Selected Comment
... 15-315.8
Deletion of All Comments
... 15-415.9
Initial Cursor Display Position Setting
... 15-416.
Needle Mark Display
... 16-116.1 Display
Method
... 16-316.2
Adjustment of the Needle Mark Angle
... 16-416.3
Display/Hiding of Biopsy Range Marks
... 16-416.4
Selection of the Biopsy Guide Size
... 16-416.5
Selection of the Biopsy Adaptor
... 16-4Measurements
17. Measurement
... 17-117.1 Basic
Operation
... 17-117.2 B-mode
Measurements
... 17-217.3 M-mode
Measurements
... 17-3General Information
18.
Output/Saving of Image Data
... 18-118.1
Data Archive to a CD-R
... 18-119. Other
Function
... 19-119.1
Beep Tone Adjustment
... 19-119.2
Adjustment of the Panel Backlight
... 19-120. Using
MI/TI
... 20-120.1
Basic Knowledge of MI/TI
... 20-120.2
MI/TI Display Description
... 20-320.3
Parameters Affecting the MI/TI Values
... 20-320.4
Operating Procedures for MI/TI
... 20-420.5
References for MI/TI
... 20-521. Maintenance
Check
... 21-121.1
Maintenance Checks to Be Carried
Out by Customers
... 21-121.2
Maintenance Checks to Be Carried
Out by Service Personnel
... 21-321.3
Consumable Parts and Parts
Requiring Periodic Replacement
... 21-322.
Checks Before the System
Is Judged Defective
... 22-123. Accuracy
of
Measurement
... 23-124. Conformance
Standards
... 24-126.
Guidance and Manufacturer's
Declaration
... 26-1Safety Precautions
1. Meaning of Signal Words
In this operation manual, the signal words
DANGER
,WARNING
, andCAUTION
are used regarding safety and other important instructions. The signal words and their meanings are defined as follows. Please understand their meanings clearly before reading this manual.Signal word Meaning
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.CAUTION
Indicates a potentially hazardous situation which, if not avoided,may result in minor or moderate injury.
CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in property damage.
2. Meaning of Safety Symbols
Symbol Description Type-BF applied part
ECG : Type-BF equipment
Others : Type-BF when Type-BF applied part is connected. All ultrasound transducers can be connected to this system are Type-BF applied part.
3. Safety
Precautions
Please observe the following precautions to ensure patient and operator safety when using this system.
DANGER:
Do not use flammable gasses such as anesthetics, or flammable liquids such as ethanol, near this product, because there is danger of explosion.WARNING:
1. Be sure to connect the power plug to a two-pin (with dual earth terminal) outlet meeting the ratings indicated on the rating nameplate. If this type of outlet is not available, contact your TOSHIBA representative.2. Information concerning the terminal to which the potential equalization conductor is to be connected ( ):
To use another medical device in combination with this system, an equipotential wire for connecting to an equipotential bus must be supplied. For more information, contact your TOSHIBA representative.
* Be sure to connect the potential-equalization lead wire before inserting the equipment power plug into the receptacle. Also, be sure to remove the equipment power plug from the
receptacle before disconnecting the wire to avoid electrical shock.
3. Information concerning the functional earth terminal ( ):
Use of the functional earth terminal is not allowed in Europe. The system should be connected according to local requirements. 4. Connect the earth conductor only before turning ON the system.
Disconnect the grounding cable only after turning OFF the system. Otherwise, electric shock may result.
5. Do not connect this system to outlets with the same circuit breakers and fuses that control current to devices such as life-support systems. If this system malfunctions and generates an overcurrent, or when there is an instantaneous current at power ON, the circuit breakers and fuses of the building’s supply circuit may be tripped.
6. Do not allow this system or other equipment to come into contact with the patient. If this system or other equipment is defective, the patient may receive an electric shock.
7. Do not connect to the system transducers other than those specified by TOSHIBA, to prevent accidents such as burns, electrical shock, or fire.
WARNING:
8. Do not subject the transducers to knocks. Use of defective transducers may cause an electric shock.9. Do not remove the covers or panels of the system, or electric shock may result.
10. Prolonged and repeated use of keyboards can result in hand or arm nerve disorders in some individuals. Observe the local institution work safety/health regulations on keyboard use. 11. When the optional ECG cable is used, the following measures are
required.
(1) To prevent electric shock, check the following items before operation.
• There must be no abnormality of the ECG cable. • The ECG cable must be connected correctly.
(2) Connect the ECG cable to the system before attaching the pick-up of the ECG to the patient. If the connector of the ECG cable comes into contact with live parts, the patient may receive an electric shock.
(3) Do not place the ECG electrodes in direct contact with the patient's heart. Cardiac arrest may occur.
* The ECG electrodes are of the BF type and are not designed to be placed in direct contact with the heart. (4) To reduce the risk of electric shock or burns, use only the
cable and patient leads supplied with the reference signal unit .
(5) The ECG electrodes used should be commercially available products provided by the customer.
12. Before using equipment such as an electric scalpel, high-frequency therapy equipment, a defibrillator, etc., remove the ECG electrodes from the patient. In addition, do not allow ultrasound transducers to come into contact with the patient. The patient may be burned or receive an electric shock. 13. Prepare a backup system when an urgent examination is to be
performed, for example during an operation or for an emergency case. If a severe problem should occur with the system, the normal operation of the system may not be recovered by resetting the system and the examination may not be able to continue.
CAUTION:
1. Precautions concerning clinical examination techniques (1) This system must be used only by medical personnel fullytrained in clinical examination techniques.
(2) This operation manual does not describe clinical examination techniques. Selection of the proper clinical examination technique must be based on specialized training and clinical experience.
2. Malfunctions due to radiowaves
(1) Use of radiowave-emitting devices in the proximity of this kind of medical electronic system may interfere with its operation. Do not bring or use devices which generate radio waves, such as cellular telephones, transceivers, and radio controlled toys, in the room where the system is installed.
(2) If a user brings a device which generates radio waves near the system, they must be instructed to immediately turn OFF the device. This is necessary to ensure the proper operation of the system.
3. Precautions concerning installation and movement of the system (1) Be sure to install the system on a level floor and lock the
casters. If this is not done, the system may move, injuring the patient.
(2) Do not push the system from the side. If the system is pushed from the side, it may fall down and cause injury.
(3) When the system is moved over a sloped surface, it must be moved slowly by two persons. Otherwise, the system may slide unexpectedly and cause a serious injury.
(4) Do not sit on the system.
The system may move, causing you to lose your balance and fall.
(5) Do not place any objects on top of the monitor. They may fall, causing injury.
(6) Confirm that the peripheral units are secured before moving the system. Otherwise, the peripheral units may fall and cause injury.
(7) If the ECG cable is connected, disconnect it before moving the system. Otherwise, the cable may get caught in the casters or someone may trip over the cable and injury may result.
(8) When the system is moved over a step, be careful not to allow the system to fall. When holding the system at the bottom to help move it over a step, take special care to prevent hand injuries.
CAUTION:
4. To prevent electric shock, do not connect the peripheral units (video printer, VCR, etc.) to an external outlet (except for Deskjet series printers). Peripheral units should be connected to the service outlet of the system. For the connection procedures, contact yourTOSHIBA representative.
In addition, be sure to use the cable supplied with the system to connect a Deskjet series printer. If a different cable is used, there is a risk of electric shock.
5. Before cleaning the system, be sure to disconnect the power cable from the outlet. If the system is defective, there is a risk of electric shock.
6. Do not allow fluids such as water to contact the system or peripheral devices. Electric shock may result.
7. Terminals and switches on the system rear panel
(1) Only equipment that conforms to the safety standards must be connected to the ETHERNET terminal on the rear panel. Otherwise, smoke or an electric shock may result. (2) Turn OFF the system power before connecting or
disconnecting the cable from the ETHERNET terminal. If the cable is connected or disconnected with the system power ON, the system may not operate properly.
(3) When a HUB is used, turn ON the system power only after tuning ON the power to the HUB.
(4) Do not press the RESET switch when the system is operating normally. Data loss or damage to the system may result. 8. In normal diagnostic ultrasound mode, there is no danger of a
low-temperature burn; however, keeping the transducer on the same region of the patient may cause such a burn.
9. If there is any possibility of secondary infection, cover the entire system with a sterile cover. To obtain a sterile cover, contact your TOSHIBA representative.
CAUTION:
10. Precautions regarding patient information entry(1) Before examining a new patient, press to delete the patient information and data recorded in the image memory for the previous patient. Otherwise, the new data may be confused with the data of the previous patient.
(2) If the patient information is modified by pressing PATIENT ID , the
patient information for the data already recorded in the image memory is not modified (the data recorded in the image
memory is not deleted). Thus, if the data recorded in the image memory is that for the previous patient, images for the different patients will have the same patient information, and this may cause confusion.
It a patient ID is to be changed, press to delete the patient information and data recorded in the image memory for the previous patient. Otherwise, the data belonging to the current patient may be confused with the data for the previous patient. (3) Confirm that the patient to be examined is the same as the
patient ID. There is a possibility of recording images with incorrect patient IDs.
(4) Enter the patient's height and weight correctly in the patient ID registration screen. If the entered height and weight are not correct, the BSA will be calculated incorrectly.
(5) Enter the patient information including LMP and IVF correctly in the patient ID registration screen. Otherwise, the fetal growth data will be calculated incorrectly.
11. Do not turn OFF the power supply of the system during printing, data saving, or invoking. Doing so may cause these processes to not operate correctly. In addition, doing so during access to a CD-R or an HDD may damage these media.
12. When discarding any part of this system, follow all applicable local regulations.
CAUTION: 1. Do not use this system in locations subject to intense electric or magnetic fields
(near transformers, for example). In such locations, the monitor will be adversely affected.
2. Do not use this system near devices generating high frequencies (such as medical telemeters and cordless telephones). The system can malfunction or adversely affect such devices.
3. To prevent damage to the system, do not use it in: • Locations where it may be exposed to direct sunlight
• Locations where it may be subject to sudden changes in temperature • Locations that are dusty
• Locations where it may be subjected to vibrations • Locations near heat generators
• Locations where the humidity may be high
• Locations where the air filter of the system is blocked (for example, near a wall or another unit)
4. Turn ON the system only after the power has been OFF for more than 5 seconds. If the system is turned ON immediately after being turned OFF, the system may malfunction.
5. Turn OFF the system subswitch or stop transmission by FREEZE switch before connecting or disconnecting a transducer. If a transducer is connected or disconnected with an image displayed, the system and/or the transducer may malfunction.
6. The user can record ID registration and user registrations (with a function to permit user-specific items such as the hospital name to be registered). Be sure to back up such information on external storage media. Data stored in the system may be lost due to improper operation or an accident.
7. Move the system by the handle. If the user pushes or pulls other sections, the system may be damaged.
8. Do not press or use force on the main panel. It may damage the system. 9. If this system is used in a small room, it may cause the room temperature to
rise. Proper ventilation must be provided.
10. Do not connect or disconnect a printer from the Deskjet series when the system power is ON. The printer may become defective.
11. The air filter at the left of the system must be cleaned once every 6 months. If the air filter is clogged, the internal temperature will rise, shortening the service life of the system. For inspection and cleaning by service personnel, contact your TOSHIBA representative.
12. Deterioration of electrical and mechanical safety characteristics (such as generation of a leakage current or deformation/abrasion of mechanical parts) and of image sensitivity and resolution may occur over a period of time. A maintenance service contract is recommended to avoid accidents and erroneous diagnosis and to maintain system performance.
13. Manage the system properly to prevent it from being infected with computer viruses.
4. Warning
Labels
Various warning labels are attached to this system in order to call the user's attention to potential hazards.
* The symbol on the warning labels attached to the system indicates safety precautions. The warning labels use the same signal words as used in the descriptions in the operation manuals.
* Detailed information about the warning labels is given in the main body of this operation manual. Read the operation manual carefully before using the system.
The name, appearance, indication, and location of each warning label is as follows. <<System to which pictogram labels are attached>>
<1> <2>
<3>
No. Label Meaning
(a) Cautions against sitting on the system. <1>
(a) (b)
(b) Precautions regarding the registration of panel switches
(a) Urges caution related to handling of the transducers.
For handling of the transducers, refer to the transducers’ operation manual.
<2>
(a) (b) (b) Cautions that the system must not be used around flammable gasses. (a) Cautions that the system must be placed on a
horizontal surface. <3>
(b) Cautions that the cover must not be removed in order to prevent electric shock.
<<System to which labels containing pictogram and characters are attached>>
<1> <2>
<3>
No. Label Meaning
(a) Cautions against sitting on the system. <1>
(b) Precautions regarding the registration of panel switches
(a) Cautions that the system must not be used around flammable gasses.
<2>
(b) Urges caution related to handling of the transducers.
For handling of the transducers, refer to the transducers’ operation manual.
(a) Cautions that the system must be placed on a horizontal surface.
<3>
(b) Cautions that the cover must not be removed in order to prevent electric shock.
Important Information
1. The responsibility for maintenance and management of the product after delivery resides with the customer who has purchased the product.
2. The warranty does not cover the following items, even during the warranty period: (1) Damage or loss due to misuse or abuse.
(2) Damage or loss caused by Acts of God such as fires, earthquakes, floods, lightning, etc.
(3) Damage or loss caused by failure to meet the specified conditions for this system, such as inadequate power supply, improper installation, or unacceptable
environmental conditions.
(4) Damage or loss due to mobile use in a vehicle which is not authorized by TOSHIBA. (5) Damage or loss due to use outside the territory in which the system was originally sold. (6) Damage or loss involving system purchased from a source other than TOSHIBA or its
authorized distributors or agents.
3. This system shall not be used by persons other than fully qualified and certified medical personnel.
4. Do not make changes or modifications to the software or hardware of this product. 5. In no event shall TOSHIBA be liable for problems, damage, or loss caused by relocation,
modification, or repair performed by personnel other than those designated by TOSHIBA. 6. The purpose of this system is to provide physicians with data for clinical diagnosis.
The responsibility for diagnostic procedures lies with the physicians involved. TOSHIBA shall not be liable for the results of diagnostic procedures.
7. Important data must be backed up on external recording media such as clinical records, notebooks, or CD-R.
8. TOSHIBA shall not be liable for loss of data stored in the memory of this system caused by operator error or accidents.
9. This manual contains warnings regarding foreseeable potential dangers. Be alert at all times to dangers other than those indicated.
10. TOSHIBA shall not be liable for damage or loss that results from negligence or from ignoring the precautions and operating instructions contained in this operation manual. 11. On the occasion of change of the administrator or manager for this system, be sure to hand
over this operation manual.
12. When disposing of this system, contact your TOSHIBA representative. Do not dispose of this system without consulting TOSHIBA representative first. TOSHIBA does not assume
1. Intended Use and Specifications
1.1 Intended
Use
(1) This system provides high-quality ultrasound images in B mode and M mode. (2) Transducers have their own characteristic applications. However, this system
allows transducers to be used for regions other than those specified as applicable, because this system conforms to Real Time Display of Thermal and Mechanical Output Indices on Diagnostic Ultrasound Equipment. For the transducers that can be used with this system and their applications, refer to subsection 2.3 "List of Available Transducers".
1.2 Specifications
(1) Power
(a) Line voltage : 100 VAC ±10%
120 VAC ±10%
100 VAC to 127 VAC ±10% 220 VAC to 240 VAC ±10% (b) Line frequency : 47 Hz to 63 Hz
(c) Power consumption : Maximum 600 VA (including external devices) (100 V, 120 V, and 100 V to 127 V system) Maximum 580 VA (including external devices) (220 V to 240 V system)
(2) Operating environmental conditions
(a) Ambient temperature : 10°C to 35°C
(b) Relative humidity : 35% to 85% (no condensation) (c) Atmospheric pressure : 700 hPa to 1060 hPa
(3) Storage and transportation conditions (a) Ambient temperature : -10°C to 55°C
(b) Relative humidity : 30% to 85% (no condensation) (c) Atmospheric pressure : 700 hPa to 1060 hPa
NOTE: The line voltage differs depending on the area.
(4) External dimensions and mass (not including optional units)
(a) External dimensions : 420 mm wide × 630 mm deep × 1260 or 1300 mm high
2. System Configuration
2.1 Standard
Configuration
(1) Main unit (2) Monitor (3) Accessories
• Transducer cable hanger • Transducer holder • Operation manual
2.2 List of Optional Devices
The following list is accurate and complete at the time of printing. Due to constant developments, certain products may change or be no longer available at a later stage. For an up-to-date overview of available options, please consult your TOSHIBA
representative.
The following optional devices are available with this system.
No. Item Model
1 Reference signal unit UJUR-530A
2 VCR installation kit UZRI-530A
3 DICOM kit USDI-530A
4 Footswitch UZFS002A
5 VCR SVO-9500MD (SONY/NTSC)
SVO-9500MDP (SONY/PAL)
6 Black-and-white printer UP-895MD (SONY)
P91W (MITSUBISHI) TP-8010 (MITSUBISHI)
7 Printer Deskjet 5650/5652 (HP)
* It may not be possible to use some of the peripheral devices listed above depending on the power conditions of the country. For details, contact your Toshiba
2.3 List of Available Transducers
Transducer name Indications for use
PLF-308P Abdominal, intraoperative, pediatric
PLQ-805A Peripheral vascular, small organ, musculoskeletal PLQ-1203A Peripheral vascular, small organ, musculoskeletal PVQ-375A Abdominal, fetal, pediatric
PVQ-381A Abdominal, fetal, pediatric cardiac, neonatal cephalic PVQ-641V Endocavitary
PVQ-662A Abdominal, pediatric, neonatal cephalic PVF-738F Abdominal, small organ, intraoperative PVF-738H Abdominal, small organ, intraoperative PVF-745V Abdominal, small organ, intraoperative PVF-715R Endorectal
3. System Overview
3.1 Name of Each Part
Main panel
Transducer connector
Monitor
Transducer cable hanger Transducer holder CD-R drive Side panel Caster Handle Rear panel Power panel Subswitch
* Before disconnecting the power cable of the system from the commercial power supply, turn OFF the sub switch, wait until the screen display disappears, and then turn OFF the main switch on the power panel.
3.2 Main
Panel
PATIENT ID A B C A / B ACOUSTIC POWER GAIN B PRESET TRANSDUCER STANDBY HDD B S 2 3 4 5 6 7 8 9 0 1 Q W E R T Y U I O P A S D F G H J K L Z X C V B N M Shift DelCtrl Alt Alt Ins
F-POSI THI /FREQ IP <1> <2> <3> <4> <5> <6> <7> <8> <9> <10> <12> <11> <13> <14> <15> <16> <17> <18> <19> <20> <21> <22> <23> <24> <25> <26> <27> <28>
No. Name Function
<1> ACOUSTIC POWER Adjusts the ultrasonic beam output.
<2> NEW PATIENT Deletes the data for the previous patient, including the ID and measured values, and resets the system.
<3> PATIENT ID The patient ID registration screen appears. <4> PRESET Invokes the registered initial settings (presets). <5> TRANSDUCER Selects a transducer.
<6> STANDBY/HDD The STANDBY LED blinks when the main switch is ON and the subswitch is OFF (standby status).
The HDD LED blinks when the HDD is being accessed. <7> STC Adjusts the ultrasound echo reception sensitivity according
to the depth from the body surface. <8> Keyboard Used to enter characters and symbols
Shift + alphabetic character keys;
Lowercase letters can be entered.
Shift + keys other than alphabetic character keys
(such as );
The symbol in the upper row can be entered. <9> MENU Displays a menu according to the system status. <10> Body mark Permits entry of body marks.
No. Name Function
<11> Multifunctional dial Used for editing measurements. Also used to adjust the angle of the needle mark, input auto annotation, and rotate the transducer mark on the body mark.
<12> IP Adjusts the image quality according to the registered image conditions.
<13> M+B Displays B-mode and M-mode simultaneously.
<14> M Displays M-mode images in full screen mode.
<15> B single Displays a standard (single) B-mode image. <16> B dual Displays two B-mode images side by side.
<17> THI/FREQ Switches the transmission/reception frequencies. When a transducer that supports THI mode is used, this switch is also used to set the system in THI mode.
<18> DEPTH/ZOOM Adjusts the viewing depth for display of the ultrasound images. Every time this dial is pressed, the function is switched between DEPTH and ZOOM.
<19> F-POSI Adjusts the focus position.
<20> OUTPUT B Controls assigned output device (such as VCR or black-and-white printer).
<21> OUTPUT A Controls assigned output device (such as VCR or black-and-white printer).
<22> CINE Sets the trackball to image memory control mode. <23> SET Determines the cursor position for measurement, and
determines the selected items, etc.
<24> Trackball Moves the cursor or mark during image movement or measurement.
<25> GAIN Adjusts the sensitivity of black/white images. Also used to play back the cine loop.
<26> NEXT Moves the cursor to the next position during measurement, selected items, etc.
<27> STILL REVIEW No function is registered to this switch at the time of shipment.
<28> FREEZE Freezes and unfreezes the image.
NOTE: User function switches
User function switches are switches to which other functions can be registered. These include switches to which functions have already been assigned and switches to which no functions are registered, such as the switch. For details, contact your TOSHIBA representative.
* To check the functions registered to the user function switches, select [Panel View] from the [Other] menu.
3.3 Side
Panel
NOTE: The ECG input is optional.
GAIN ECG gain
FOOT SW PATIENT A B ECG ECG position
Input terminal for ECG lead
For connection of the footswitches
3.4 Rear
Panel
CAUTION:
1. Only equipment that conforms to the safety standards must be connected to the ETHERNET terminal on the rear panel. Otherwise, smoke or an electric shock may result.2. Turn OFF the system power before connecting or
disconnecting the cable from the ETHERNET terminal. If the cable is connected or disconnected with the system power ON, the system may not operate properly.
3. When a HUB is used, turn ON the system power only after tuning ON the power to the HUB.
OUT REMOTE <1> <2> <3> <4> VIDEO RESET IN ETHERNET <5> <6> SERIAL PARALLEL
No. Name Functions
<1> VIDEO External video device input/output terminals
<2> PARALLEL Parallel ports
<3> REMOTE Printer connection terminals
<4> RESET System reset switch
<5> SERIAL Serial ports
3.5 Power
Panel
NOTE: The voltage and current indications of the AC line differ depending on the
area.
<1> <2> <3>
<4>
FUSE 250V T4.0A
100V TOTAL MAX. 264VA (ISOLATED/SWITCHED)
<5>
100V 50/60Hz 600VA
<6>
No. Name Functions
<1> FUSE Fuse
<2> Main switch This is a circuit breaker.
<3> Functional earth terminal Terminal for the functional earth cable.
<4> Equipotential terminal Terminal for the potential equalization conductor.
<5> Outlet Connector for peripheral devices.
3.6 Symbols
This system uses the following symbols. For safety symbols, refer to page S-1.
Symbol Description Functional grounding
Equipotentiality
Main switch OFF (The AC power is turned OFF.) Main switch ON (The AC power is turned ON.)
Subswitch OFF (The power to electronic circuits is turned OFF: Standby status)
Subswitch ON (The power to electronic circuits is turned ON.)
TRANSDUCER A Transducer connector A
TRANSDUCER B Transducer connector B
ECG Electrocardiogram (option) FOOT SW Footswitch
AC (Alternating current)
4. Preparation for Examination
4.1 Moving the System
CAUTION:
1. Be sure to install the system on a level floor and lock the casters. If this is not done, the system may move, injuring the patient.2. Do not push the system from the side. If the system is pushed from the side, it may fall down and cause injury. 3. When the system is moved over a sloped surface, it must be
moved slowly by two persons. Otherwise, the system may slide unexpectedly and cause a serious injury.
4. Do not sit on the system.
The system may move, causing you to lose your balance and fall.
5. Do not place any objects on top of the monitor. They may fall, causing injury.
6. Confirm that the peripheral units are secured before moving the system. Otherwise, the peripheral units may fall and cause injury.
7. If the ECG cable is connected, disconnect it before moving the system. Otherwise, the cable may get caught in the casters or someone may trip over the cable and injury may result.
8. When the system is moved over a step, be careful not to allow the system to fall. When holding the system at the bottom to help move it over a step, take special care to prevent hand injuries.
CAUTION: Observe the following precautions before moving the system to prevent
system malfunctions.
1. Turn OFF the subswitch and the main switch on the power panel on the rear of the system.
2. Store the transducer in the transducer holder and hang the cable on the transducer cable hanger.
3. When moving the system, use the handle. Do not push weak sections such as the panel.
(1) Unlock the four casters.
(2) Move the system using the handle.
4.2 Connecting the Transducer
CAUTION: Turn OFF the system subswitch or stop transmission by FREEZE switch
before connecting or disconnecting a transducer. If a transducer is connected or disconnected with an image displayed, the system and/or the transducer may malfunction.
(1) Connect the transducer and turn the lock handle clockwise to lock the connector. (2) When disconnecting the transducer, turn
the lock handle counterclockwise to unlock the connector and then hold the connector and pull it straight out.
Lock Lock handle
5. Connecting the Power Cable and
Protective Earth
WARNING:
1. Be sure to connect the power plug to a two-pin (with dual earth terminal) outlet meeting the ratings indicated on the rating nameplate. If this type of outlet is not available, contact your TOSHIBA representative.2. Information concerning the terminal to which the potential equalization conductor is to be connected ( ):
To use another medical device in combination with this system, an equipotential wire for connecting to an equipotential bus must be supplied. For more information, contact your TOSHIBA representative. * Be sure to connect the potential-equalization lead wire
before inserting the equipment power plug into the receptacle. Also, be sure to remove the equipment power plug from the receptacle before disconnecting the wire to avoid electrical shock.
3. Information concerning the functional earth terminal ( ):
Use of the functional earth terminal is not allowed in Europe. The system should be connected according the local requirements.
4. Connect the earth conductor only before turning ON the system. Disconnect the grounding cable only after turning OFF the system. Otherwise, electric shock may result.
5. Do not connect this system to outlets with the same circuit breakers and fuses that control current to devices such as life-support systems. If this system malfunctions and generates an overcurrent, or when there is an
instantaneous current at power ON, the circuit breakers and fuses of the building’s supply circuit may be tripped.
Connect the power plug to an outlet for medical equipment. By doing this, the protective earth line is connected.
6. Power ON/OFF
6.1 Power
ON
(1) Check the items below before turning the power ON. • The power cable should be connected.
• One or more transducer should be connected.
(2) Turn ON the main switch on the power panel on the rear of the system. The STANDBY LED blinks.
(3) Turn ON the subswitch. Some time after the setup screen is displayed, the B-mode screen appears.
(4) If required, adjust the brightness ( ) and contrast ( ) using the corresponding knobs on the monitor.
(a) Set the contrast ( ) to minimum. The image disappears.
(b) Adjust the brightness ( ) so that the brightness level of the screen is the same as in the examination room.
(c) Then adjust the contrast ( ) to set the maximum luminance.
NOTE: When this system is started up or another transducer is selected, you
may hear a click. This does not indicate abnormality.
6.2 Power
OFF
CAUTION: Do not turn the system power OFF while a CD-R disk or HDD is being
accessed. If the system power is turned OFF during access to the CD-R disk or HDD, the disk may be damaged and recorded data may be lost. (1) Before turning OFF the system, confirm that the CD-R or HDD is not being
accessed.
(2) Turn OFF the subswitch. The STANDBY LED blinks.
(3) When the system is not going to be used for an extended period of time, turn OFF the main switch on the power panel and disconnect the power plug.
6.3 Preparation for Use During an Operation or for Emergency
Cases
WARNING:
Prepare a backup system when an urgent examination is to be performed, for example during an operation or for an emergency case. If a severe problem should occur with the system, the normal operation of the system may not be recovered by resetting the system and the examination may not be able to continue.6.3.1 Preparation of a backup system
Prepare a backup system when an urgent examination is to be performed, for example during an operation or for an emergency case.
6.3.2 Power OFF/ON in the case of system failure
When any of the following abnormalities occurs with the system, the system may be able to recover from the abnormality by pressing the RESET switch.
• An error message is displayed and does not disappear. • The screen display is abnormal.
• The system operations are disabled. <System OFF/ON procedure>
Follow the steps below to reset the system. (1) Press the main switch on the power panel.
(2) Wait at least 5 seconds, and then turn ON the main switch. The system is restarted.
7. Checks Before and After Use
In the interests of safety, it is the user's responsibility to carry out the following checks before and after using the system.
7.1 Checks Before Turning ON the Power
Before turning ON the power, perform the following checks.
No. Check item Check column
1 The temperature, humidity, and atmospheric pressure should meet the
conditions of use.
2 There should be no condensation.
3 There should be no deformation, damage, or stains to the system and peripheral units.
* If any stains are present, perform cleaning referring to subsection 21.1.1 "Cleaning the system".
4 There should not backlash or loose screws in the casters, monitor, panel,
etc.
5 Caster locks should function correctly.
6 There should be no damage to cables (power cable, ECG cable etc.) and
no looseness in the connectors.
7 There should be no damage or stains to the transducer and transducer cables.
* If any stains are present, perform cleaning, disinfection, or sterilization referring to the operation manual provided with the transducer.
8 No clips etc. are placed on the main panel.
9 There should be no obstacles near the movable sections and air filter of the system.
7.2 Checks After Turning ON the Power
After turning ON the power, perform the following checks.
No. Check item Check column
1 There should be no abnormal sound, unusual smells, or overheating.
2 No error message is displayed.
3 There should be no obviously abnormal noise, discontinuous display, or dark areas for B-mode images.
4 There should be no abnormality in the date and time.
5 The acoustic lens surface of the transducer should not be unusually hot.
(Perform check by hand.)
6 Switches and knobs on the panel should function normally.
7 There should be no abnormality in the color or display position of images
8. Basic Screen and Menu
8.1 Display of Various Data Items
In addition to ultrasound images, various data items are displayed on the monitor. This section describes the screen display common to all modes. Refer to section 11 "Display and Operation in Each Mode" for the display specific to each mode.
Date Time Scanning direction Heart rate Preset name Patient ID Hospital name Focus mark (1) MI (2) TIS (3) TIB (4) TIC TOSHIBA 123456789012 Abdominal 2004/03/16 ABCDEF 999 9.9C9.9 100% PM 07:05:43 0 10 5 9.9M 999 fps BG99 DR99 B frequency Current
frame rateB gain Dynamicrange value
0:00:00 VCR Acoustic power Transducer frequency IP MI 9.9 TIS 9.9 TIB 9.9 TIC 9.9 Image Process (1) MI (Mechanical Index)
This is a measure of the mechanical effects such as cavitation. (2) TIS (Thermal Index Soft Tissue)
This is a measure of the calculated/estimated rise in temperature of soft tissues. (3) TIB (Thermal Index Bone)
This is a measure of the calculated/estimated rise in temperature of the cranial bone of fetuses/infants.
(4) TIC (Thermal Index Cranial)
This is a measure of the calculated/estimated rise in temperature of the cranial bone of children/adults.
8.2 Display and Operation of the Menu
The functions not available on the panel are displayed in the menu. The menu should be displayed when required.
(1) Press
MENU . The menu appropriate for the current display mode is displayed.
Functions that cannot be selected or switched in the menu are displayed in gray. (a) Movement of the menu display position
The menu can be moved to the left or right by operating the trackball to the left or right.
(b) Display of other menus
Move the cursor to the top of the menu and then move the trackball slightly to the left or right, depending on the position in which the menu is displayed. A pull-down menu, which can be used to select other menus, is displayed. Select the menu to be displayed and press SET .
B B M Other Focal Ptn Density DR Time Smth Axial Smth LateralSmth Mid Filter E.Enhance AGC PostProcess Reject L/R U/D Rotation Scan Rang High Frame Needle
Top of the menu
(3) Press SET on the main panel.
• For an item with only ON and OFF options
The item toggles between ON and OFF each time SET is pressed. • For an item with several options
When SET is pressed or the trackball is slightly moved laterally, a pull-down menu appears next to the menu. Select the desired item by operating the trackball and then press SET .
mark on the right of an item indicates that other pull-down menus or pages exist.
(4) To close the menu, press
9. Entering the Patient ID
CAUTION:
Observe the following precautions to avoid erroneous diagnosis.1. Before examining a new patient, press to delete the patient information and data recorded in the image memory for the previous patient. Otherwise, the new data may be confused with the data of the previous patient.
2. If the patient information is modified by pressing PATIENT ID , the
patient information for the data already recorded in the image memory is not modified (the data recorded in the image memory is not deleted). Thus, if the data recorded in the image memory is that for the previous patient, images for the different patients will have the same patient information, and this may cause confusion.
It a patient ID is to be changed, press to delete the patient information and data recorded in the image memory for the previous patient. Otherwise, the data belonging to the current patient may be confused with the data for the previous patient.
3. Confirm that the patient to be examined is the same as the patient ID. There is a possibility of recording images with incorrect patient IDs.
4. Enter the patient's height and weight correctly in the patient ID registration screen. If the entered height and weight are not correct, the BSA will be calculated incorrectly.
5. Enter the patient information including LMP and IVF
correctly in the patient ID registration screen. Otherwise, the fetal growth data will be calculated incorrectly.
9.1 Entering and Saving Data on the Patient ID Registration
Screen
(1) Press to initialize the system. (2) Press PATIENT
ID . The patient ID registration screen appears.
Name Date of Birth
Sex Male Female Unknown
Age Height Weight BSA cm kg m2 ADULT
Blood Pressure High/Low mmHg
Comment
ABC ID List ID Save
Close Patient Data Input
/
2004/03/16
Used: 10 Free: 1990 Total: 2000
AM06:54:47 ID Number
LMP PRV EDD IVF CLN
(YYYY/MM/DD)
(3) Enter the appropriate patient information in the fields.
• In the "Date of Birth" field, enter the date using the same format as the current date displayed at the top right of the screen. When the date of birth is entered, the age is calculated automatically and displayed in the "Age" field. It is also possible to type a value directly in the "Age" field.
• When the height and weight are entered, the BSA (body surface area) field is automatically filled with a calculated value.
• Depending on the preset menu setting, the BP fields and options for obstetric data may not be displayed.
• When one of [LMP], [PRV], [EDD], [IVF], or [CLN] is selected and the Enter key is pressed, the screen for entering the date and the number of weeks appears. On this screen, enter the date.
(4) After the data has been entered, press [ID Save]. A confirmation dialog is displayed. Select [OK] to save the data.
NOTE: (1) Units of height and weight
The units of height and weight differ depending on the setting of "Entry System" in the preset menu.
(a) Metric ... "cm" and "kg"
(b) English ... "feet", "inch", "lb", or "cm", "kg" (2) Formula for body surface area
The formula differs depending on the setting of "Calculation" in the preset menu. (a) Oriental • ADULT : BSA = WT0.425 * HT0.725 * 73.58 / 10000 • PEDI 12-14 : BSA = WT0.423 * HT0.651 * 102.65 / 10000 • PEDI 6-11 : BSA = WT0.444 * HT0.663 * 88.83 / 10000 • PEDI 1-5 : BSA = WT0.423 * HT0.362 * 381.89 / 10000 • PEDI 0 : BSA = WT0.473 * HT0.655 * 95.68 / 10000 (b) Occidental • ADULT : BSA = WT0.425 * HT0.725 * 71.84 / 10000 HT : Height (cm) WT : Weight (kg)
9.2 Sorting and Calling a Patient ID
(1) Press [ID List] on the patient ID registration screen. The list of patient IDs saved in the system is displayed in ascending order of ID number.
ID: Name:
Sort key
Select Delete Delete All Used:
Free: Total: Close 2000 1997 3 3 2 1 No. ID Name 111 123 222 AAA BBB CCC ID List EDD EDD (YYYY/MM/DD)
(2) When the button for "ID", "Name", or "EDD" is selected in the [Sort Key] field, the patient ID numbers are sorted according to the selected condition.
(3) Select the desired patient data item and select [Select] or move the cursor to the desired patient data and press SET twice. The patient data is displayed on the Patient ID Registration screen.
9.3 Deleting Saved Patient IDs
(1) To delete a patient ID
(a) Using the trackball, select the patient ID to be deleted. (b) Press [Delete].
(c) Select [OK] on the deletion confirmation screen. The patient ID data is deleted.
(2) To delete all the saved patient IDs (a) Press [Delete All].
(b) Select [OK] on the deletion confirmation screen. All the saved patient ID data is deleted.
10. Selecting Presets
10.1 Introduction
In this system, 12 types of initial settings can be saved for examinations. These initial settings are called presets. The examination presets can be selected by pressing the PRESET switch on the main panel.
10.2 Procedure for Selecting Presets
When A or B is pressed, the corresponding preset is selected.
* When C is pressed, the Exam Select menu is displayed. The desired preset can
be selected from the displayed list by using the trackball and SET .
Exam Select ABDOMINAL1 ABDOMINAL2 OB/GY1 S-PARTS1 S-PARTS2 PV Cardiac 11 12 PRESET OB/GY2 PROSTATE 10 Display example
11. Display and Operation in Each
Mode
11.1 B Mode
(1) Entering B single mode Press B .
(2) Entering B dual display mode Press DUAL .
(3) Adjustment on the panel PATIENT ID A B C A / B ACOUSTIC POWER GAIN B PRESET TRANSDUCER STANDBY HDD B S 2 3 4 5 6 7 8 9 0 1 Q W E R T Y U I O P A S D F G H J K L Z X C V B N M Shift Del
Ctrl Alt Alt Ins
F-POSI THI /FREQ IP <3> <5> <2>, <6> <4> <1> <8> <7>
Adjustment items Procedures
<1> Sensitivity of the entire image
GAIN
2D
Gain
Use the GAIN dial of the palm controller.
<2> Field depth
DEPTH ZOOM
Use the DEPTH/ZOOM dial.
* When the ZOOM LED is lit, hold down the dial to light the DEPTH LED.
<3> Sensitivity according to the depth Use the STC slide controls.
<4> Calls up registered image-quality conditions.
IP
Use IP to select the desired image-quality conditions.
<5> Switches the focus position.
F.POSI
Adjustment items Procedures <6> Magnification, reduction, and
movement of the image (PAN/ZOOM)
DEPTH ZOOM
Use the DEPTH/ZOOM dial and the trackball. (1) When the DEPTH LED is lit, hold down the dial
to light the ZOOM LED.
(2) When the dial is turned, the image is magnified/reduced. When the trackball is operated, the image moves.
(3) When the dial is held down again, the system quits PAN/ZOOM and the DEPTH function is selected.
<7> Switches the
transmission/reception frequencies or switches to THI mode
THI/FREQ
Switches the transmission/reception frequencies. When a transducer that supports THI mode is used, this switch is also used to set the system in THI mode.
<8> Freeze the image (to a still image).
FRZ
Press FRZ . When it is pressed again, the
(4) Adjustment using the menu When
MENU is pressed, the B menu is displayed.
Automatically sets the focus of the ultrasound beam.
Enhances the edge of the image.
Adjusts the image smoothness in the lateral direction. Adjusts the image smoothness in the axial direction. Adjusts the image smoothness.
Adjust the dynamic range. Increases image sharpness.
Displays the image right-left reversed.
Rotates the image in 90 steps.°
Displays the image top-bottom inverted. B Focal Ptn Density DR Time Smth Axial Smth LateralSmth Mid Filter E.Enhance AGC PostProcess Reject L/R U/D Rotation Needle Scan Rang High Frame
Displays the needle mark for biopsy procedures. Selects the filter type.
Suppresses comet-tail and reduces noise. Adjusts the image display intensity. Reduces noise at the low brightness level.
Sets the field angle (field width). Sets the High Frame function ON/OFF.
11.2 M Mode
(1) Entering M mode Press M . (2) Entering M + B mode Press B M+B .(3) Setting the M mark
(a) Enter M + B mode. The M mark is displayed on the B-mode image. (b) Move the M mark display position using the trackball.
(4) Adjustment on the panel Refer to page 11-2.
(5) Adjustment using the menu When
MENU is pressed, the M menu is displayed.
Adjusts the image smoothness in the axial direction. Enhances the edge of the image.
Adjusts the sensitivity of the M-mode image. Adjusts the dynamic range.
Adjusts the sweep speed of the M-mode image.
Turns ON/OFF the Time mark display for the M image area.
Suppresses comet-tail and reduces noise. Adjusts the image display intensity.
Selects the Max Sampling pattern. M M Speed M DR M Gain Cmp M E.Enhance M AxialSmth M Time Mark M AGC M Post Proc M Max Smple M Time Smth
12. Cine Function
12.1 Introduction
CAUTION:
When performing examination of a new patient, press to delete the recorded data in the image memory. Otherwise, the new data may be confused with the data of the previous patient.When an image is frozen, the images immediately before the frozen image can be played back and edited. This function is called Cine. Cine images are cleared by turning OFF the power or unfreezing the frozen image.
12.2 Cine Playback
(1) Press FRZ to freeze the image.
(2) When the trackball is operated, the system enters frame-advance playback mode. • Operate the trackball to the right for forward playback.
• Operate the trackball to the left for reverse playback.
* For M-mode images, the trackball operation is opposite to the above. (3) When the trackball dial (the dial around the trackball) is turned during
frame-advance playback, the system enters continuous playback mode. The playback speed can be changed using the dial.
• Operate the dial clockwise to increase the playback speed.
• Operate the dial counterclockwise to decrease the playback speed.
When the trackball is operated, the system returns to frame-advance playback mode.
(4) Continuous playback edit function (setting of the playback range)
It is possible to set the start frame and end frame for continuous playback in frame-advance playback mode.
(a) Operate the trackball to display the start frame of the playback range and press SET .
(b) Operate the trackball to display the end frame and press NEXT .
(c) Turn the trackball dial. The images between the specified start and end frames are played back continuously.
13. Display of Reference Signals
Electrocardiographic (ECG) reference signals can be displayed during B-mode display, B Dual mode display, M-mode display, and M+B mode display. This function is optional.
WARNING:
1. To prevent electric shock, check the following items before operation.• There must be no abnormality of the ECG cable. • The ECG cable must be connected correctly.
2. Connect the ECG cable to the system before attaching the pick-up of the ECG to the patient. If the connector of the ECG cable is in contact with live parts, the patient may suffer an electric shock.
3. Do not place the ECG electrodes in direct contact with the patient's heart. Cardiac arrest may occur.
* The ECG electrodes are of the BF type and are not designed to be placed in direct contact with the heart. 4. To reduce the risk of electric shock or burns, use only the
cable and patient leads supplied with the reference signal unit .
5. The ECG electrodes used should be commercially available products provided by the customer.
6. Before using equipment such as an electric scalpel, high-frequency therapy equipment, a defibrillator, etc., remove the ECG electrodes from the patient. In addition, do not allow ultrasound transducers to come into contact with the patient. The patient may be burned or receive an electric shock.
13.1 Display Procedures
(1) Turn OFF the power of the system and connect the pick-up cable etc. to the side panel.
(2) Turn ON the power of the system. (3) Attach the ECG electrodes to the patient.
(4) Set the mode to M mode in which reference signals can be displayed.
* If the signals need to be displayed in B-mode and B dual mode, it is necessary to preset the display setting to ON.
13.2 Adjustment of Reference Signals
(1) Adjustment of the ECG waveform
Use the ECG dial on the side panel to adjust the ECG waveform. • GAIN: Adjusts the sensitivity.
• POSI: Adjusts the display position. (2) Adjustment using the menu
When [ECG] is selected from the [Other] menu, the following menu is displayed. The settings can be changed using the trackball and SET .
ECG Sync Mod Sync+Real CH1 Delay Interval
Performs ECG-gating at the delay time set.
Displays ECG-gating images and live images simultaneously (for dual-display mode).
Sets ECG synchronization mode.
Sets the delay time from detection of the R wave to refereshing of the B screen.
Sets the interval before the B screen is updated in internal gating mode.
14. Body Mark Entry and Deletion
A body mark, which indicates the anatomical region for the current examination, can be displayed on the screen.
14.1 Display of a Body Mark
(1) Press the body mark switch on the main panel. The preset body mark for the current examination region is displayed.
14.2 Selection of a Body Mark
(1) To display another body mark for the same examination region Press SET and then NEXT . The body mark is switched. (2) To display a body mark for another examination region
(a) Press
MENU . The body mark menu is displayed.
(b) Move the cursor to the upper part of the menu and move the trackball slightly to the left or right, depending on the position in which the menu is displayed. A pull-down menu is displayed.
(c) Select the desired examination region and press SET .
(d) Select the body mark to be displayed and press SET . The selected body mark is displayed on the screen.
14.3 Movement of the Transducer Mark
(1) Operate the trackball to move the transducer mark. (2) The transducer mark rotates as the multifunctional
dial is turned.
Transducer mark
14.4 Movement of the Body Mark
(1) Select [Mark Move] in the menu. (2) Press SET .
(3) The body mark moves as the trackball is operated.
(4) When SET is pressed again, the body mark position is set.
14.5 Deletion of the Body Mark
(1) Select [Mark Disp] in the menu.
(2) Press SET . The body mark disappears.
* While the body mark is displayed, set the body mark switch on the panel to OFF, and turn ON/OFF the body mark switch again. The body mark disappears.