Copyright 01977 AmericanSociety forMicrobiology Printed inU.S.A.
Center for Disease Control
Diagnostic Immunology
Proficiency Testing Program Results
for 1976
ROGER N. TAYLOR,'* KAREN M. FULFORD,' VINCENT A. PRZYBYSZEWSKI,l AND
VICTORIA POPE2
DiagnosticImmunologySection, ProficiencyTestingBranch,'andVenereal DiseaseSerology Laboratory, Bacterial Immunology Branch,1 Center forDiseaseControl, Atlanta,Georgia 30333
Received forpublication4May1977
Over 900 laboratories participated inthe Diagnostic Immunology portionof
the 1976 Proficiency Testing Program, which was provided by the Center for
DiseaseControl under the authority of the Clinical Laboratories Improvement
Act of 1967. One hundredspecimens prepared bythe Centerfor DiseaseControl
for analysis were distributed on a quarterly schedule or in special surveys.
Feedback from participating laboratories included over 37,500 qualitative and
33,000 quantitative responses, which were analyzed to determine individual
laboratory proficiency levels. Inaddition, informationsuppliedby participants
in each survey helped to delineate trends in testingprotocols. The specimens
chosen for analysis called for a broad range of tests commonly performed in
diagnostic immunology laboratories, including those for rubella antibodies,
hepatitis B surface antigen, bacterial antibodies, rheumatoid factor,
immuno-globulins and other serum-specific proteins, and carcinoembryonic antigen. A
summary of the dataanalysisisprovidedsothat the laboratories can improve
theiroverall performance levels.
The Center forDiseaseControl (CDC)
Profi-ciency Testing Program was
developed
toim-plement the Clinical Laboratories
Improve-mentActof1967 (4).Its function is tomeasure
laboratory performance with theintentof
eval-uating and improving that performance. The
program'sobjectivesareaccomplished by
iden-tifying deficiencies, evaluating methods, and
disseminating information.
Input tothisprogram isgainedinthe
follow-ing manner. CDC prepares specimens and
sends themtolicensed laboratories andtosome
nonlicensedparticipants(special study and
ref-erence laboratories) with the instruction that
they be tested in the manner routine to that
laboratory. This means that regular laboratory
staff membersareexpectedtoprocess the
speci-mens in the same manner and with the same
methods that they use for routine specimens.
Participation in the program is mandatory for laboratories that provide interstate testing ser-vices.
This report contains a summary of data
ob-tained from the 1976 Proficiency Testing
Pro-gramand discusses the implications of the
re-ported observations. Table 1lists types of tests
that the laboratories were requested to
per-form and tabulates the response levels of the
morethan 900 laboratories participating in the
Diagnostic Immunology portion of the
Profi-ciency Testing Program. Eachlaboratory was
asked to carry out onlythe tests that its staff
performs routinely.
Results ofspecificsurveys areprovided
else-where (5,6, 11,12,15, 16, 21-24).The purpose of
thisreport is todelineate the overall trends and
changes that becomeapparent when all of the
1976testdata wereevaluated.
MATERIALS AND METHODS
Most of the sera or plasma used for specimen preparationswerepurchasedfrom commercial sup-pliers on government contract. Other sources do-natedpools of humanserumofknown reactivityin
speciftictests.
Preliminary testing foracceptability of the sera wasdoneby the Diagnostic Immunology Section of the ProficiencyTestingBranch, by anappropriate CDCspecialty laboratory, orboth. HepatitisB sur-face antigen (HBsAg) reactivity for all sera was determined by the DiagnosticImmunology Section ortheVirology Section, Phoenix Laboratories Divi-sion.Onlysera negative for HBsAg by radioimmu-noassay wereusedfor specimens other than those to betested for HBsAg. Specimens obtainedasplasma weredefibrinated with calcium chloride or throm-bin.
Details ofspecimen preparationfor each survey areincludedinthesummaryanalyses(5, 6, 11, 12, 15, 16, 21-24). Briefly, specimens were adjustedto thedesired titer, filtered through sterile membrane filters, and dispensed into appropriate vials or 224
on February 7, 2020 by guest
http://jcm.asm.org/
CDC IMMUNOLOGY PROFICIENCY TESTING PROGRAM 225
TABLE 1. Summary of CDC Diagnostic Immunology Proficiency Testing Program 1976
No. of No.of Mean no. Determination chal-
sur-
of labsre-Determnatioycl
sponding/
veys/yr
surveyRubellaantibody ... 8 4 283
Streptococcal
antibod-ies
Antistreptolysin 0 6 2 354
Anti-deoxyribonucle-ase ... 6 2 23
Multiple
streptococ-cal antbodies ... 6 2 98
Rheumatoid factor .... 5 2 437
C-reactive protein .... 2 1 377
Infectious
mononucleo-sisserology ... 5 2 459
Antinuclear antibodies 4 2 337 Rickettsial antibodies. 1 1 359
Salmonellagroup D
an-tibody ... 1 1 365
Brucellaantibody... 2 1 393
Tularemiaantibody... 2 1 221
Toxoplasma antibody . 2 1 144
Immunoglobulin quan-titation
IgG... 7 2 333
IgA... 7 2 333
IgM ... 7 2 333 Complement
C3... 5 1 304
C4... 5 1 164
a-1-Antitrypsin ... 5 1 245
a-2-Macroglobulin .... 5 1 47
Haptoglobin... 5 1 213
Transferrin ... 5 1 82
Ceruloplasmin... 5 1 115
Carcinoembryonic
an-tigen ... 5 1 125 Syphilis... 40a 4a 404 HBsAg... 10 2 264
aDoes notinclude a replacement shipment of 10 samples
necessitatedby losses and delays in the mail.
tubes. Theadequacyofsampleswasconfirmed inde-pendently by the Diagnostic Immunology Section, by other CDClaboratories,andbyreference l1bora-tories. Anongoingquality control program used by theDiagnosticImmunologySectionensuresthat all specimenssatisfy preestablishedcriteria for
steril-ity, antibodytiter and stability, andbetween-vial
variability.
Each specimenshipmentwaspackagedin
accord-ancewithpostal regulationsandincluded
appropri-ate instructions and report forms. Completed
re-ports weretobepostmarkedwithin2 weeks of the initialshipping date. Responseswerecompiledand graded, and individualperformance rankings were reported to participants within 3 to 4weeks after responses werereceived. The acceptable responses weredetermined from referencelaboratoryresults. Overall-response data, which were evaluated and
compiledinsummaryanalysisorpublishedas
sepa-rate reports (18-20), were later sent toall partici-pants.
RESULTS
Rubella results were relatively constant
throughoutthe year (21-24). Geometric mean
titersfor samplessent outinconsecutive
quar-ters werenot significantly different (Table 2).
Slightchanges in methodologywerenoted. The
percentage of laboratories usingkaolin serum
treatment decreased slightly from 59.8 in the
firstquarterto 53.9 inthe fourth. Increases of
33.8 to 39.1% and 7.8 to 10.0% were noted in
those usingheparin-manganous chlorideserum
treatment and trypsinized human 0 cells,
re-spectively. Higher geometricmean titerswere
obtained with trypsinized human cells than
withother cells. Titers were about 24% lower
when chick cellswereused and 31% when
Ab-bott duracytes were used. Laboratories using
thekaolinserumtreatmenthadapercentageof
results outside the acceptable limits almost
twice aslargeasthose using the
heparin-man-ganouschlorideordextransulfateserum
treat-ment (Table 3).
The first-quarter shipment included four
specimens intendedtocomparethe variationin
test results for streptococcal antibodies
ob-tained withlyophilizedserum versuswhole
se-rumofhuman and horse antisera (21). Results
for thelyophilized specimens varied lessamong
respondents than those for whole sera.
Al-though the geometricmean antistreptolysin 0
titers and thepercentageofpositive qualitative
antistreptolysin 0 results were higher for the
horse serum specimens than for the human
serum specimens, the percentage of positive
multiple-enzymetestresultswas lower for the
horseserum(Table 4). Quality controltestshad
shown that the agglutination reaction ofthe
multiple-enzyme test was weaker with the
horse serumthan with the human serum, and
therefore the test withhorse seurm had to be
performed more carefullyfor accurate results.
Table 5 shows the participants' performance
with thestreptococcal antibodytests. The
aver-agepercentageofresponsesoutside the
accept-ablerange islargerfor the antistreptolysin 0
testthan forthe other twotests,but thehigher
TABLE 2. Resultsofrubellahemagglutination inhibitionfor repeatsamples
Geometricmean titer Sam- Survey Hepa- Dex- Kaolin
ples ea e-Koi
rin tran
A II, May 28 24 17
A II, May 29 23 17
A III, Aug. 26 25 15
B III, Aug. 56 49 50
B IV, Nov. 60 64 49
VOL. 6, 1977
on February 7, 2020 by guest
http://jcm.asm.org/
TAYLOR ET AL.
TABLE 3. 1976rubellaproficiency testingresults
Reference Accepta- No. ofN. Resultsoutside Sampleno. laboratoryrblerange patii acceptable
XG pants
~range
%Heparin-MnCl2 or dextran sulfate
S6-001 8 .16 109 8
S6-002 111 64-256 108 20
S6-011 29 16-64 124 9
S6-012 29 16-64 124 9
S6-021 95 32-256 125 9
S6-022 35 16-128 122 11
S6-031 106 64-128 140! 49
S6-032 131 64-256 140 44
(20)b
Kaolin
S6-001 8 .16 155 I 19
S6-002 111 64-256 157 39
S6-011 29 16-64 148 36
S6-012 29 16-64 148 33
S6-021 95 32-256 159 12
S6-022 35 16-128 159 42
S6-031 106 64-128 157 69
S6-032 131 64-256 157 62
(39)b atG,Geometricmean.
bAverage.
TABLE 4. Results ofstreptococcalantibody tests
Antistreptolysin 0 Multiple
Serum enzyme
SrmGatiter| SDGb | %tive tive)
Human, whole 344 1.77 75 93 Human,lyophilized 205 1.50 68 85 Horse, whole 320 1.67 89 68 Horse, lyophilized 307 1.59 89 67
a'XG,Geometric mean.
bGeometricstandard deviation-antilog of the standard
deviationof thelog of titers.
percentages areresultsofsubstantially smaller
acceptable ranges for the antistreptolysin 0
tests. The number of laboratories reporting
anti-deoxyribonuclease Bandmultiple-enzyme
resultsis small but growing.
Because therehad beenexcessive variability
intherheumatoid factor test results from
pre-vious surveys, three specimens calibrated in
international
units (IU) for rheumatoid factorwere included in the second-quarter shipment
(19, 22). Results were reported in the usual
manner butwere then converted to IU by
com-paring results of the first two specimens
against the one containing 250 IU of
rheuma-toid factor per ml. The percentage of results
within one dilution of the median titer
in-creaseddramaticallyafterstandardization
(Ta-ble 6). Table 7 shows performance achieved
without the benefit ofstandardization in 1976.
Proceduresand distributions of results were
similar for the bacterial agglutination tests
(brucella, tularemia, salmonella group D, and
Weil-Felix-ProteusOX19)(21-24). Different
an-tigen sources apparently caused the greatest
variation in titers obtained. Geometric mean
titers ofslide test results were lower than of
tube test results, except for those from the test
for tularemia.Approximately 80% of the
partic-ipantsbased theirresultsonthe final dilution
of serum (including the antigen volume), and
about 65% used 2 + as the end-point reaction.
Titers of80 to 160 were considered significant
by mostlaboratories, but 70 to 80% considered a
changeintiterbetween acute and convalescent
sera to be more definitive. Table 8 shows the
performancewith these tests.
Few participant laboratories had difficulty distinguishing between negative and positive
samplesfor antinuclear antibodies by indirect
immunofluorescence methods; however, titers
were widely distributed (23, 24). In two
sur-veys, useofratliver cells resultedinthe
high-est geometric mean titers in these tests,
whereas the lowest titers were obtained with
TABLE 5. 1976streptococcalantibody proficiency testing results
Reference Results
Accepta- No.ofpar- outside
Sample no. laboratory blerange ticipants acceptablerange (%)
Antistreptoly-sin0
S6-003 180 160-250 340 22
S6-004 276 160-250 335 66
S6-005 287 250-333 344 38
S6-006 297 166-500 343 10
S6-033 443 256-625 364 49
S6-034 170 125-250 364 31
Anti-deoxyri- bonu-clease B
S6-003 254 166-680 12 17
S6-004 274 166-1360 13 23
S6-005 400 240-1360 13 15
S6-006 409 166-1360 13 8
S6-033 741 680-960 33 36
S6-034 202 170-240 31 26
(21J)b
Multiple enzymes
S6-003 156 .400 26 4
S6-004 113 s400 26 0
S6-005 115 c400 24 4
S6-006 155 .333 24 12
S6-033 342 200-500 25 44
S6-034 200 100-400 23 43
1
~~~~~(18)b
atG,Geometricmean. bAverage.
on February 7, 2020 by guest
http://jcm.asm.org/
CDC IMMUNOLOGY PROFICIENCY TESTING PROGRAM 227
TABLE 6. Effect of standardization on rheumatoid factortestresults Laboratories(%)reporting results equal to:
Test Median Median + 1tube Median + 2 tubes
Std" NotStdb Std Not Std Std Not Std
Slide 54.2 15.4 93.7 50.3 97.9 74.1
Tube 51.0 33.2 91.2 78.1 98.5 91.8
Total 41.0 27.1 91.0 61.3 97.6 75.1
aStd, Titers compared to reference preparation.
bNot Std, Raw titers reported.
TABLE 7. 1976 rheumatoid factor proficiency testing results
Reference
Results
laboratory Acceptable
No. ofpar-outside
Sampleno. XZGaorqual- c
itative rane ticipants acceptable
sult range
Slide test
S6-013 Posb 1-640 73 3
S6-014 Pos 1-640 72 4
S6-015 Pos 1-2560 72 0
S6-023 Pos orNegcPosorNeg 44 0
S6-024 Pos Pos 79 38
(9)d
Latex tube
S6-013 346 160-640 171 22
S6-014 320 160-640 171 22
S6-015 691 640-2560 171 43
S6-023 254 Neg-320 146 11
S6-024 226 Neg-640 165 10
(22)d
at,Geometric mean.
bPos,Positive.
cNeg, Negative.
dAverage.
human leukocytes. Latex agglutination tests,
especially the Hylandtest, producedmore
neg-ative results on the lower-titered specimens
than did indirect immunofluorescence tests.
Only ninelaboratoriesreported usinga
peroxi-dase method fordetectingantinuclear
antibod-ies. Results of the antinuclear antibody and
toxoplasmatests areshown in Table 9.
Variables in the toxoplasma indirect
immu-nofluorescence results thatwerepointedoutin
the summary analysis include source of anti-gen, source of conjugate, and types of
micro-scopeocular andlightsourceused (23).
Table 10shows the infectious mononucleosis
test results. The percentage of results outside
the acceptable limits is greaterfor the oxcell
hemolysin tests than it is for the heterophile
tests. But, considering that the acceptable
ranges for the hemolytic testare smaller, the
performance is actuallybetter with theoxcell
hemolysin test. A detailed evaluation of tests
for infectious mononucleosis has been
pub-TABLE 8. 1976bacterial agglutininsproficiency testing results
Reference Aze ble lio Span Results Refe
n.abrencey
AcceptableNo. of par- outsideac-Sample no. laboratory..
Ga range ticipants ceptable
range(%) Salmonella
Slide
S6-010 80 -160 300 17
Tube
S6-010 184 80-320 133 42
Weil-Felix Slide
S6-018 359 160-1280 293 15 Tube
S6-018 538 160-1280 129 9
Brucella Slide
S6-027 320 40-640 275 2
S6-028 226 40-640 271 1
Tube
S6-027 160 40-640 168 2
S6-028 180 40-640 168 5
Tularemia Slide
S6-035 160 160 138 60
S6-036 <20 <20 84 21
Tube
S6-035 135 80-160 131 26
S6-036 c40 c40 93 4
a
kG,
Geometricmean.lished which compares sensitivity, specificity,
reproducibility, and cost of the mostfrequently
used tests (14).
Table 11 shows the immunoglobulin
quanti-tation results. Immunoglobulins (IgG, IgA,
IgM)arequantitatedby single radial
immuno-diffusion in most laboratories(22). Other
meth-ods usedbyparticipantsinclude automated
im-munoprecipitation, fluorometry,
electroimmu-noprecipitation, and lasarnephelometry. Most
results werereported inmilligramsper
decili-ter, but a few were in IU per milliliter. Results
(milligrams per deciliter) varied somewhat
among manufacturers of antiserum or plates. Usually more than half of the results were
outside the acceptable limits. The need for
standardization of this test isobvious.
VOL. 6, 1977
on February 7, 2020 by guest
http://jcm.asm.org/
TABLE 9. Antinuclearantibodies andtoxoplasma proficiency testingresults
Reference No. of
Results
Samleno.laoraorAccepta-
paric- outside SamPle no. laboraty blerange pants acceptableXGa pants range (%
Antinuclear antibody
S6-029 <20 <20 129 22
S6-030 212 80-512 130 28
S6-039 58 32-160 229 27
S6-040 <20 <20 161 6 (21)b
Toxoplasma Indirect
im-
munoflu-orescence
S6-037 108 32-256 113 24
S6-038 14 C32 106 37
(31)b Indirect
he- magglu-tination
S6-037 91 62-128 33 36
S6-038 45 32-64 31 23
(29)b
aXG, Geometricmean.
bAverage.
TABLE 10. 1976infectious mononucleosis proficiency testingresults
Reference No.of Results Sampleno. labOra- blerange
partici-
outside
toryXG Pants acceptable
_______ range (%)
Presumptive heterophile
S6-007 83 28-224 189 10
S6-008 <7 <7 117 71
S6-009 83 28-224 189 9
S6-025 22 <32 146 17
S6-026 96 10-224 175 9
(23)b
Oxcell
hemoly-sin
S6-007 135 80-160 24 33
S6-008 <10 <10 21 24
S6-009 135 80-160 24 33
S6-025 16 10-20 26 50
S6-026 80 40-160 28 21
(32)b
aXC,Geometricmean.
bAverage.
Because carcinoembryonic antigen (CEA) de-terminations are being performed in an in-creasing number of laboratories, a special
pro-ficiencytesting survey for CEA determinations
wasconductedinJune 1976 (20). Participation
was voluntary, and performance was evaluated
to assess inter- and intralaboratory variation
and toprovidebase-line data to help determine
whether CEA should be included in routine
surveys. Most of the participating laboratories
used the commercially available CEA-Roche
procedure, but a fewusedamethod in whicha
second antibody (anti-human globulin) is used
to precipitate the anti-CEA and bound CEA.
Nearly 25%of the results
reported
for thesam-plesinthissurvey were
incorrectly
categorized
(normal, intermediate, elevated, or indicative
of metastasis). Differences in positivity
be-tweensamplesofnormal poolserum (1.1 ngof
CEA/ml)andserumwith aCEAconcentration
of 5.5 ng/dl were not detected in 16% ofthe
responses. Results reported for samples with
CEA levels -20 ng/ml showed that the direct
method produced significantly higher values
than the indirect method on either whole or
dilutVdplasma.
Third-generation tests for HBsAg are now
being usedin morethan95%ofthe laboratories
in the program,compared with68% a year ago
(18) (Fig. 1). Failure to detect HBsAg in
profi-ciencytesting samples has mostoften been
by
respondents using a second-generation test
only.False-positive resultsare most
frequently
obtained whenrespondents failtoconfirm
posi-tiveradioimmunoassay results. Table12shows
theresultsofthe hepatitisB surveys.
Table13shows the resultsofthreecommonly
usedtests fromthe syphilis serologyprogram.
Satisfactory performance percentages were
highforthese and for otherserologicaltests.
DISCUSSION
The Diagnostic Immunology portion of the
CDC
Proficiency
Testing Program
is intended TABLE 11. 1976immunoglobulin quantitationproficiency testing
resultsReference No. of Results Sampleno.
tory.
labora-Ga0 Acceptablepatc-outside
range pants
acceptable
pnsrange
(%Immuno-globulin G
S6-019 1,198b 1,120-1,300b 279 80
S6-020 802 750-850 279 87
(83)c
Immuno-globulin
A
S6-019 280 265-312 278 73
S6-020 124 80-150 277 46
Immuno-globulin
S6-019 442 330-532 278 54
S6-020 98 84-135 278 52
1
~~~~~~~~~(53)C
aXG,
Geometricmean.bAllunitsinmilligramsperdeciliter.
cAverage.
on February 7, 2020 by guest
http://jcm.asm.org/
CDC IMMUNOLOGY PROFICIENCY TESTING PROGRAM 229
110
100
90
so
c 70 3rdGeneration
2 60
L 50 nGet
O 40
2'30
20
10
1stGeneration
0
1971 1972 1973 1974 1975 1976
FIG. 1. Tests used by participants in the CDC Proficiency TestingProgram for HBsAg. Multiple tests, Number of tests used expressed as the percentage in excess of 100% of the laboratories:
no.oftests -no.of laboratories
Multipletests=
100no.
of laboratoriesTABLE 12. 1976hepatitis B proficiency testing resultsa
Acceptablere- No. ofpar- Incorrect Sample no.
Aceta
re-pn.
op responsesL6-001 Neg 265 0
L6-002 Pos 265 12
L6-003 Pos 265 1
L6-004 Pos 265 12
L6-005 Pos 265 37
L6-006 Pos 262 8
L6-007 Neg 262 3
L6-008 Pos 262 0.4
L6-009 Pos 262 0.4
L6-010 Pos 262 9
(8)C
aL6-002 and L6-004 are
duplicates.
L6-005-HBsAglevelnearlower limitdetectable by
radioim-munoassay.
bPos, Positive;Neg, negative.
CAverage.
to measure laboratory proficiency for a broad
spectrumofserological
procedures.
Onecanseefrom the list in Table 1 that the
samples
pro-vided encompass most of the tests commonly
used in aserology laboratory. Fourmajor
fac-tors are considered when the
samples
for thediagnosticimmunology proficiency testing
pro-gram aredecidedupon: (i)thenumberof
labo-ratoriesthatprovidetherelevanttest, (ii)the
numberoftimesthoselaboratories
perform
thetest in ayear, (iii) theimportance ofthedisease
ordiseasestowhich thetestresultsare
related,
and (iv)the
degree
ofdifficulty
thatisusually
encountered in obtaining useful results. The
final samplemix isoften affected
by
otherfac-tors such as the availability of materials or
other resources and the regulatory
require-ments. Atpresent,the fourfactorslisted above
aresubjectively evaluated and
applied,
butaneffort is being made to
weight
themnumeri-callysothatamoreobjective
approach
toplan-ning sample composition can be achieved.
Ef-forts arealso being madeto remove or modify
the otherconstraints.
The distribution oftests used and the
per-formancelevels for rubella testing
changed
lit-tle from previous years. There was some shift
towarduseof therecommendedmethodologies
(9, 10). It is strongly recommended that the
standardized methods be used.
They
werede-veloped to minimize variation in results while
maintaining sensitivity and specificity. Poor
performance has
repeatedly
been correlatedwith failure to adhereto the
established
stan-dards (21-24). The manuals
describing
thesemethods can be obtained from the
Diagnostic
ImmunologyTraining Branch attheCDC.
Because the CDC selects reference
laborato-ries only from among those that use
recom-mended methodologies, participant
laborato-ries using cells or serum treatments that are
knownto
produce
lowerormorevariabletitersare morelikelyto report
unacceptable
results.Laboratories using kaolin serum treatment
hadalmost twice aslargeapercentageof
unac-ceptable results as did those
using
therecom-mended serumtreatments. Previoussummary
analyses have shown that a similarsituation
exists when something other than fresh chick
ortrypsin human0 cellsareused
(23, 24).
To investigate the effects of
using
animalVOL. 6, 1977
on February 7, 2020 by guest
http://jcm.asm.org/
TABLE 13. 1976syphilisproficiencytesting results
FTA-ABSa VDRL° RPRC
Acceptable
ceepartici-
No-t.
of % In-result correctpants
2+ 126 9
B-2 + 126 21
2+-4+ 126 1
B-2+ 126 25
N 126 6
2+-3+ 126 6
N 126 9
2+-3+ 126 12
2+-3+ 126 0
2+-3+ 126 6
4+ 145 0
N 145 10
4+ 145 0
2+-4+ 145 0
3+-4+ 145 6
N 145 4
3+-4+ 145 2
4+ 145 0
N-B 145 14
2+-4+ 145 7
1+-3+ 144 58
4+ 144 1
3+-4+ 144 2
N 144 8
2+-3+ 144 12
3+-4+ 144 6
2+-3+ 144 13
4+ 144 3
N 144 9
4+ 144 1
4+ 154 2
4+ 154 0.6
4+ 154 0.6
N 154 7
4+ 154 2
2+ 154 19
4+ 154 1
2 + 154 19
N 154 7
4+ 154 1
(8)d
Acceptable No.o % In-result partici- correct
pants N-Wo N Rl-R2 N N N-Wo N N-Wo Rl-R2 N-Wo Rl-R2 N Wo-1 Wo N-Wo N N-Wo Wo-Rl N N N Wo Wo-Rl N N Wo N Wo-Rl N Ri-R2 Wo Ri Wo N RI N Wo N N Wo 165 2 165 11 165 5 165 7 165 1 165 3 165 4 165 3 165 5 165 5 201 2 201 1 201 3 201 16 201 1 201 2 201 1 201 6 201 2 201 14 195 4 195 30 195 2 195 0 195 42 195 22 195 3 195 4 195 2 195 1 205 26 205 23 205 35 205 2 205 24 205 3 205 27 205 3 205 0.5 205 36 (iO)d
AcetableAcceptbe partici-No. of %In
In-result pants correct
pants
N-Ri 131 0
N 131 5
R2-R4 131 2
N 131 4
N 131 2
N-Ri 131 2
N 131 2
N-Ri 131 2
R2-R4 131 2
N-Ri 131 1
R2-R4 153 4
N 153 4
R1-R2 153 4
Ri 153 21
N-Ri 153 2
N 153 1
N-Ri 153 3
R1-R2 153 3
N 153 3
N 153 10
N 164 4
R1-R2 164 2
R2 164 10
N 164 1
N-Ri 164 1
Ri 164 16
N 164 2
R2 164 10
N 164 1
R2-R4 164 3
R1-R2 172 4
R2-R4 172 3
Ri 172 48
N 172 2
R2-R4 172 3
N 172 5
R1-R2 172 5
N 172 5
N 172 1
Ri 172 48
(6)d
aFTA-ABS, Fluorescent treponemalantibody-absorptiontest.
bVDRL, VeneralDiseaseResearch Laboratorytest.
c RPR, RapidPlasmaReagintest.
dAverage.
serum versushumanserumandlyophilized se- Basedonproficiency testingexperience, itis
rum versus whole serum in a proficiencytest recommended that the anti-deoxyribonuclease
for streptococcal antibodies, combinations of Bandmultiple-enzymetests notbe usedasthe
each were included in the first shipment for onlyscreentodetect thepresenceof
antistrep-1976. The results revealed that, despite addi- tococcal antibodies. These tests are valuable
tional sources oferrorintroduced by lyophiliz- adjuncts to the antistreptolysin 0 test but
ing andreconstituting, therewaslessvariation should not be used in lieu ofit, especially for
in the results reported from lyophilized sam- measuringantibodiesintheseraof infants and
ples. The study also revealed that, when ani- olderpersonsinwhich the titersarefrequently
mal sera are used in proficiency testing, the nearthethreshold of thetests. Small laborato-resultsshouldbeinterpreted with caution. riesthatused onlythe multiple-enzymetest to Sampleno. T6-001 T6-002 T6-003 T6-004 T6-005 T6-006 T6-007 T6-008 T6-009 T6-010 T6-011 T6-012 T6-013 T6-014 T6-015 T6-016 T6-017 T6-018 T6-019 T6-020 T6-021 T6-022 T6-023 T6-024 T6-025 T6-026 T6-027 T6-028 T6-029 T6-030 T6-031 T6-032 T6-033 T6-034 T6-035 T6-036 T6-037 T6-038 T6-039 T6-040
230
TAYLOR ET AL.on February 7, 2020 by guest
http://jcm.asm.org/
CDC IMMUNOLOGY PROFICIENCY TESTING PROGRAM 231
screen for antistreptococcal antibodies seemed
to have the most difficulty obtaining
satisfac-tory test results.
The largest source of variation in the
bacte-rial agglutination tests was the antigen (23). This reaffirms the necessity for using
standard-ized antigens when they are available, as for
the brucellatest (7), and emphasizes the need
todevelop standardantigens for the other
bac-terial agglutination tests. Recently, the CDC
reevaluated the reference antigen
recom-mended for use in the standardized brucella
agglutination tubetest and determinedthat it
should be diluted 1:50 instead of 1:100 to
con-form to the U.S. Department of Agriculture
standard (3, 7). The higher concentration
ap-pears to increase the reproducibility of results
byminimizing prozone and other erratic
reac-tionswhileproducingsharper end points.
Man-ufacturers have been notified of this change
and have been encouraged to use the same
brucella strain nowused by the National
Ani-mal Disease Laboratory.
Questions regardingthe sensitivity and
use-fulness of the slide agglutinationtests for
bru-cella have arisen many times, because many
laboratoriesusethemto screen or todetermine
titer. The CDC isconducting astudyinwhich
humanserum willbeusedto comparethetube
and slide tests. Results will bereported when
the studyiscompleted.
Standardization problems alsoexist withthe
antinuclear antibody tests. The various
sub-strates
produce
widely
varyingdegrees ofsensi-tivity. Because sensitivity and specificity are
usually inversely related, the latter probably
also varies widely. Until physical and
proce-dural standardsareformally developed, the
sit-uationcould be improvedby laboratories
volun-tarily using such reliabletests,substrates, and
reagents asthose listedby the CDC (2). To be
practical, standardization ofantinuclear
anti-body
and toxoplasma indirectimmunofluores-cence testsshould include specificationsof the
microscope oculars and the lightsources to be
used.
As a result of the poor performance in the
special survey, the CEA test will be
incorpo-rated into the routine proficiency testing
pro-gram for 1977 (20). Many of the aberrant CEA
results could have been eliminated by strict
adherence to the test
protocol
and toquality
control procedures.
It waspreviouslyreported(18) that the most
reliable and economicalHBsAgresultsare
ob-tained by laboratories that usea
single
third-generation test
(preferably
radioimmunoas-say) andconfirm the
specificity
ofpositive
reac-tionsby another method
(preferably
neutraliza-tion). Subsequentresults of
proficiency
testing
surveys indicate that many laboratories have
achieved good results by confirming positive
third-generation results with counterelectro-phoresis or some other second-generation test
andby reserving the neutralizationprocedures
forsamples containing antigen at levels below
thosedetectableby thesecond-generationtests.
By the end of 1976 more than 95% of the
laboratoriesin the Proficiency Testing Program
were using a third-generation test for HBsAg.
In light of their demonstrated superiority,
there can be little doubt that the increase in
numbers of laboratories using them represents
significant laboratoryimprovement. The rapid
shift toward third-generation tests is
gratify-ing, but, as their sensitivity increases, it
be-comes moreimportant to verify positive results
and thus eliminate false positives.
Therheumatoid factorstandardization
profi-ciency testing study (19) demonstrates that,
whereas a proficiency testing program can be
instrumental in improving laboratories by
evaluatingtestsandmethodologies(14), it can
also encourage standardization and estimate
*the amount of improvement thatcould be
ex-pected from standardization. To emphasize fur-ther the need for and value of standardization,
areference sampleforrheumatoid factorwillbe
included with future proficiency testing
sam-ples until otherstandardsarereadily available
and commonly used. The National Committee
for Clinical Laboratory Standards has
estab-lished a committee to study this problem and
developappropriate standards, but, until their
standards are available, better comparability
couldbe achieved by relatingtestresultstothe
international referencepreparation (1).
Although commercialstandards for
quantita-tive immunoglobulin determinations are
sup-posedly calibrated against the World Health
Organization Reference Preparation for
immu-noglobulins with valuesstatedinIUper
milli-liter, most laboratories continue to report in
milligrams per deciliter. These latter values
vary among manufacturers, as evidenced by
ourproficiency testing results and
by
thevar-ious conversion ratiosofmilligramsper
decili-ter toIUpermilliliter
reported
by themanufac-turers themselves. Some of the variation in
immunoglobulin results could be avoided
by
using a reference preparation andreportingthe
results in IU per milliliter rather than
milli-grams per deciliter. If thelatter valuesare
pre-ferred,IUpermillilitercanbe convertedto
mil-ligrams per deciliter by the factors listed
by
Rowe etal. (13).
The major areas of
difficulty
with variousserological tests for
syphilis
occur with serahaving minimal levels of
reactivity.
Rarely
areproblems encounteredwith
clearly
reactiveandVOL. 6, 1977
on February 7, 2020 by guest
http://jcm.asm.org/
nonreactive sera. Minimally reactive sera are
included to provide a morecritical evaluation of
laboratoryperformance. Withsuchserajudged
reactive by competent reference laboratories,
participants frequently report nonreactive
re-sults especially with theRapid Plasma Reagin
18-mmcircle cardtest. Asimilar problem has
been observed with the fluorescent
trepone-mal antibody-absorption test. Several factors
areinvolved, butamajor
problem
is feltto beinexperience in test performance, since
well-standardizedreagents are
readily
available. Atestaddedtothe1977group,theReagin Screen
test, willprobably havesomeofthesame
vari-ation inresults becauseitscriticalareaisalso
the "minimally reactive" range. Acceptable
performance levels of the variouslaboratories
were high in syphilis serology in 1976. The
major change for the year was dropping the
automated fluorescent
treponemal
antibody
testfrom the programbecause equipment and
reagents are no longermanufactured. Any
re-sults sent by participating laboratories were
graded against the reference laboratories'
re-sults for the fluorescenttreponemal
antibody-absorption test.
Boththe artandscienceof proficiencytesting
are in their infancy. There is great need for
morebasic researchintolaboratory evaluation
methodology.Although theCDCandother
pro-ficiency testing agencies have a great deal of
experience withproficiency testing,some
ques-tions remain unanswered. Possible
mecha-nisms for achieving improved laboratory
per-formance by methods other than the traditional
approaches have been reviewed inthis paper,
butother potential roles of proficiency testing
needtobeexplored.
LITERATURE CITED
1. Anderson, S. G., M. W. Bentzon, V. Houba, and P. Krag. 1970. International reference preparation of rheumatoid arthritis serum. Bull. W.H.O. 42:311-318.
2. Cavallaro, J. J., D. F. Palmer, and P. E. Bigazzi. 1976. Immunofluorescence detection of autoimmune dis-eases. Immunology series no. 7. Center for Disease Control, Atlanta, Ga.
3. Center forDiseaseControl. 1976.Bureauof Laborato-ries, current item 248. Recommended dilution of USDAbrucella reference antigen changed. Center for Disease Control, Atlanta, Ga.
4. Department of Health, Education, and Welfare. 1968.
42CFR part 74. Clinical laboratories; notice of pro-posed rule making, p. 15297-15303. Fed. Reg., vol. 33,
no. 201.
5. Larsen, S. A., V. Pope, and R. N. Taylor. 1976. Profi-ciency testing summary analysis.Syphilisserology I and IA. Center for Disease Control, Atlanta, Ga.
6. Larsen, S. A., V. Pope, and R. N. Taylor. 1976. Profi-ciency testing summary analysis. Syphilis serology II. Center for Disease Control, Atlanta, Ga.
7. National Animal Disease Laboratory. 1965. NADL diagnostic reagents manual 65D. Standard aggluti-nation testprocedures for the diagnosis of brucellosis.
U.S. Department ofAgriculture Research Service, Ames, Iowa.
8. Palmer, D. F., J. J. Cavallaro, and R. H.Gait. 1973. Laboratory diagnosis by serologic methods. Course no.9440-C. Center for DiseaseControl, Atlanta, Ga. 9. Palmer, D. F., K. L. Herrmann, and J. J. Cavallaro.
1974. Modified rubella hemagglutination-inhibition test, p. 57-83. In D. F. Palmer, J. J. Cavallaro, K. Herrmann, J. A. Stewart, and K. W. Walls (ed.), Serodiagnosis of: toxoplasmosis, rubella, cytomegalic inclusion disease, and herpes simplex. Center for Disease Control, Atlanta, Ga.
10. Palmer, D. F., K. L. Herrmann, R. E. Lincoln, M. V. Hearn, and J. M. Fuller. 1970. Standardized rubella hemagglutination-inhibition test. Immunology series
no. 2. Procedural guide. Center for Disease Control, Atlanta, Ga.
11. Pope, V., R.N.Taylor, and J. C. Feeley. 1977. Profi-ciency testing summaryanalysis. Syphilis serology III 1976.Center forDiseaseControl, Atlanta, Ga.
12. Pope, V., R. N. Taylor, and J. C. Feeley.1977. Profi-ciency testing summaryanalysis. Syphilisserology IV1976.Center for DiseaseControl, Atlanta, Ga.
13. Rowe, D. S., B. Grab, and S. G. Anderson. 1972. An international reference preparation for human serum immunoglobulins G, A, and M: content of immuno-globulins byweight. Bull. W.H.O. 46:67-79.
14. Taylor,R.N.,H.Ehrhard,and S. Marcus. 1976. Evalu-ation of tests forinfectious mononucleosis through proficiency testing. Health Lab. Sci. 13:34-44.
15. Taylor, R. N., and K. M. Fulford. 1976. Proficiency testing summary analysis. Diagnostic immunology 1976I.Hepatitis B. Center for DiseaseControl, At-lanta,Ga.
16. Taylor, R. N., and K. M. Fulford. 1976. Proficiency testing summary analysis. Diagnostic immunology 1976II.Hepatitis B. Center for DiseaseControl, At-lanta,Ga.
17. Taylor, R. N., and K. M. Fulford. 1976. Proficiency testing summaryanalysis. Serumspecificproteins. Center for DiseaseControl, Atlanta,Ga.
18. Taylor,R. N.,and K. M. Fulford. 1976. Results of the Center for Disease Control proficiency testing pro-gram for the detection ofhepatitis B surface anti-gen.J. Clin. Microbiol. 4:32-39.
19. Taylor, R. N., K. M. Fulford, and W. L.Jones. 1977. Reduction of variation inresults of rheumatoid
fac-tortestsbyuseof a serum reference preparation. J. Clin.Microbiol. 5:42-45.
20. Taylor, R. N., K. M. Fulford, and A. Y. Huong. 1977. Results ofa nationwide proficiency test for carci-noembryonicantigen. J. Clin. Microbiol. 6:433-438.
21. Taylor, R. N., K. M. Fulford, and V. A. Przybyszewski.
1976. Proficiency testing summary analysis. Diag-nostic immunology 1976. I. Rubella, streptococcal antibodies,infectiousmononucleosis,salmonella (ty-phoid0).CenterforDiseaseControl, Atlanta, Ga.
22. Taylor, R. N., K. M. Fulford, and V. A. Przybyszewski.
1976. Proficiency testing summary analysis. Diag-nosticimmunology1976.II.Rubella,rheumatoid fac-tor,C-reactiveprotein, rickettsial antibodies (Weil-Felix), quantitative immunoglobulins. Center for Disease Control, Atlanta, Ga.
23. Taylor, R. N., K. M. Fulford, and V. A. Przybyszewski.
1976. Proficiency testing summary analysis.
Diag-nostic immunology 1976. III. Rubella, rheumatoid factor, infectious mononucleosis, brucella, antinu-clear antibodies. Center for Disease Control, At-lanta,Ga.
24. Taylor, R. N., K. M. Fulford, and V. A. Przybyszewski.
1977. Proficiency testing summary analysis.
Diag-nostic immunology 1976. IV. Rubella, streptococcal antibodies,tularemia, toxoplasma, antinuclear anti-bodies. Center for Disease Control, Atlanta, Ga.