RAB Lighting Inc.
Xiao Xiang,15921313292,gary.xiao@rabweb.com
Prepared By
Prepared By
Approved By
Wangzun Zhu
Kevin Jia
Prepared For
Room 6A33, No.1388, Wuzhong road,Shanghai ,China
DLF1906101
DLF1906101-1a
Deliver Co., Ltd.
Block 11, 78 Keling Road, SSTP, Suzhou, China
0512-66801950,kevin.jia@szdeliver.com
Project Number
Report Number
Photometric Test Report
Relevant Standards
RANSI C82.77:2014
RIES LM-79-2008
This report shall not be reproduced, except in full, without written approval of Deliver Co.,Ltd. This report must not be used by the client to claim product certification, approval, or endorsement by NVLAP.
Test Date
Issue Date
2019/6/3
2019/6/3
The results contained in this report pertain only to the tested sample.
Doc No.: DLFLAB-ZY-01-28
DLC Technical Requirements v4.4
Test value
10195
standard
97.0
Premium
116.4
129.2
78.9
25.00%
120V
6.10%
25.00%
277V
7.67%
0.873
120V
0.998
0.873
277V
0.950
3810
74
100.00%
120
277
0.659
0.293
78.9
77.0
ANSI C82.77:2014
Input Voltage (V)
IES LM-79-2008
(Goniophotometer - Section 4.2)
Wrost Case
(THD & PF - section 4.3)
Non-Wrost Case
Allowable CCTs* (K)
(Integrating Sphere - Section 4.1)
IES LM-79-2008
5700
Minimum CRI
(Integrating Sphere - Section 4.1)
IES LM-79-2008
CIE 13.3-1995
63
1.0
Test Summary
Outdoor - Architectural Flood and Spot Luminaires
Requirement Category
(Test Data Source)
Test Method
Requirements
Luminaire Output (lm)
(Goniophotometer - Section 4.2)
IES LM-79-2008
1000
Zonal Lumen Requirement (0°-90°)
(Goniophotometer - Section 4.2)
IES LM-79-2008
≥82%
Minimum Luminaire Efficacy (lm/W)
(Goniophotometer - Section 4.2)
IES LM-79-2008
Power (Input Wattage) (W)
(Goniophotometer - Section 4.2)
IES LM-79-2008
Power Factor
(THD & PF - Section 4.3)
ANSI C82.77:2014
Total Harmonic Distortion (A%)
(THD & PF - Section 4.3)
Wrost Case
ANSI C82.77:2014
ANSI C82.77:2014
(THD & PF - section 4.3)
Non-Wrost Case
Input Current (A)
Power (Input Wattage - W)
IES LM-79-2008
IES LM-79-2008
(Goniophotometer - Section 4.2)
Wrost Case
(Goniophotometer - Section 4.2)
Wrost Case
2.0
Test List
Test Item
Test Date
Sample No.
1
2019/6/3
A1
2
2019/6/3
A1
3
2019/6/3
A1
Remark(If any)
3.0 Production Description
Luminaire Description:
Electrical Specification:
Photos of Luminaire Characteristics
THD and PF Test
FXLED78SFN
1、This report shall not be used by the client to claim product endorsement by NVLAP,
NIST or any agency of the US government.
2、The results reported herein have been performed in accordance with the laboratory’s
terms of accreditation. This report shall not be reproduced except in full without the written
approval of the Laboratory. The results in this report apply to the test sample(s) mentioned
above at the time of the testing period only and are not to be used to indicate applicability
to other similar products. This report does not imply that the product(s) has met the criteria
for certification.
FXLED78SFN
120V-277V,50/60HZ
Test
Model Number
Integrating Sphere Test
FXLED78SFN
Goniophotometer Test
FXLED78SFN
Doc No.: DLFLAB-ZY-01-28
4.0 LM-79 Measurement and Test Results
Test Method
4.1 Integrating Sphere Test
Model No.
FXLED78SFN
Sample ID.
The voltage of an AC power supply (RMS voltage) or DC power supply (instantaneous
voltage) applied to the device under test shall be regulated to within ±0.2 percent under
load.
The sample was measured using 4π geometry and operated at rated voltage and was
stabilized before measurement. Chromaticity coordinates, correlated color temperature
and color rendering index were calculated from the spectral radiant flux measurements
taken at 1 nm intervals over the range of 380 to 780 nm.
Voltage (V)
120.02
CCT (K)
3810
CRI
74
Duv
3.1E-03
Test Result
The samples were tested according to the IES LM-79-2008.
Photometric paramters were measured using an integrating sphere, a spectroradiometer
and software. The ambient temperature condition inside the sphere was maintained at 25º
C ± 1º C.
The sample measurements were made using a spectroradiometer connected by a fiber
optic cable and detector through the detector port of the integrating sphere.
4.1 Integrating Sphere Test
Doc No.: DLFLAB-ZY-01-28
The samples were tested according to the IES LM-79-2008.
Voltage
(Vac)
Frequency
(Hz)
Current
(A)
Power (W)
Power
Factor
120.05
60
0.659
78.9
0.998
Horizontal
Spread
Vertical
Spread
Horizontal
Spread
Vertical
Spread
10195
119.6
103.6
91.6
69.4
100.00%
Luminous
Efficacy
(lm/W)
Flux
(lm)
Test Method
The ambient temperature shall be maintained at 25° C ± 1° C, measured at a point not more than 1
m from the sample and at the same height as the sample.
The voltage of an AC power supply (RMS voltage)or DC power supply (instantaneous voltage)
applied to the device under test shall be regulated to within ±0.2 percent under load.
The samples were operated at rated voltage and was stabilized before measurement. Luminous
flux, luminaire efficacy, zonal lumen were calculated from the software taken at 0.5
overtical
intervals and 10
ohorizontal intervals.
Field Angle(10%)
Beam Angle(50%)
Orientatio
n
Light Down
Test Conditions
Test Result
129.2
FXLED78SFN
90
Sample ID.
Stabilization time
(Min.)
Zonal Lumen
Requirement
(0°-90°)
4.0 LM-79 Measurement and Test Results
4.2 Goniophotometer Test
Photometric paramters were measured using a type C goniophotometer and software.
Light Distrubtion Curve
Isolux Plot
4.2 Goniophotometer Test
Doc No.: DLFLAB-ZY-01-28
Zonal Lumen Summary
ZONAL LUMEN SUMMARY
Zonal (lm)
Total (lm)
522.78
0 - 10
522.78
1511.29
0 - 20
2034.07
2361.46
0 - 30
4395.53
2722.92
0 - 40
7118.45
2070.73
0 - 50
9189.17
820.57
0 - 60
10009.74
150.22
0 - 70
10159.95
28.20
0 - 80
10188.15
6.65
0 - 90
10194.80
0.00
0 - 100 10194.80
0.00
0 - 110 10194.80
0.00
0 - 120 10194.80
0.00
0 - 130 10194.80
0.00
0 - 140 10194.80
0.00
0 - 150 10194.80
0.00
0 - 160 10194.80
0.00
0 - 170 10194.80
0.00
0 - 180 10194.80
80-90
90-100
100-110
110-120
170-180
120-130
130-140
140-150
150-160
160-170
40-50
50-60
98.18%
99.66%
70-80
90.14%
19.95%
43.12%
69.82%
0-10
10-20
20-30
30-40
4.2 Goniophotometer Test
100.00%
100.00%
100.00%
100.00%
100.00%
100.00%
99.93%
100.00%
100.00%
100.00%
100.00%
60-70
Percent
5.13%
Doc No.: DLFLAB-ZY-01-28
4.3 Goniophotometer Test
Axial Candela
DEG.
HOR.
DEG.
VERT.
4.3 Goniophotometer Test
Characteristics
NEMA Type
6 H x 6 V
Maximum Candela
6254.166
Maximum Candela Angle
25.5 H 0 V
Horizontal Beam Angle (50%)
91.6
Vertical Beam Angle (50%)
69.4
Horizontal Field Angle (10%)
119.6
Vertical Field Angle (10%)
103.6
Lumens Per Lamp
N.A. (absolute)
Total Lamp Lumens
N.A. (absolute)
Beam Lumens
7787
Beam Efficiency
N.A.
Field Lumens
9835
Field Efficiency
N.A.
Spill Lumens
359
Luminaire Lumens
10195
Total Efficiency
N.A.
Total Luminaire Watts
78.9368
Ballast Factor
1
ISOCANDELA CURVES
Doc No.: DLFLAB-ZY-01-28
LUMEN TABULATION
0 1 3 5 7 9 11 13 15 17 20 23 26 29 33 38 43 48 55 65 75 85 90 Total 90 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 85 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 75 0.1 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.3 0.3 0.2 0.2 0.2 0.1 0.1 0 0 0 0 65 0.5 0.9 1 1 1 1 0.9 0.9 0.9 1.1 1.3 1.2 1.3 1.3 1.2 1 0.7 0.6 0.3 0.1 0 0 0 55 1.48 * 2.99 * 3.00 * 2.97 * 2.9 2.8 2.7 2.7 2.7 3.3 3.8 3.7 3.9 4 3.6 2.6 1.5 1.3 0.5 0.1 0 0 0 47.5 2.40 * 4.83 * 4.83 * 4.78 * 4.68 * 4.55 * 4.46 * 4.40 * 4.31 * 5.23 * 6.02 * 5.72 * 6.20 * 6.28 * 5.65 * 4.15 * 2.3 1.6 0.6 0.1 0 0 0 42.5 3.94 * 7.90 * 7.88 * 7.81 * 7.68 * 7.55 * 7.45 * 7.35 * 7.19 * 8.67 * 9.93 * 9.45 * 10.29 * 10.48 * 9.45 * 7.21 * 4.26 * 2.8 0.9 0.1 0 0 0 37.5 4.67 * 9.35 * 9.32 * 9.24 * 9.13 * 9.01 * 8.88 * 8.71 * 8.50 * 10.36 * 12.09 * 11.63 * 12.73 * 13.05 * 12.00 * 9.46 * 5.76 * 3.9 1.3 0.2 0 0 0 33 4.83 * 9.66 * 9.62 * 9.53 * 9.44 * 9.36 * 9.25 * 9.07 * 8.89 * 10.97 * 12.97 * 12.55 * 13.88 * 14.43 * 13.58 * 10.93 * 6.66 * 4.51 * 1.5 0.2 0 0 0 29 4.75 * 9.50 * 9.45 * 9.35 * 9.25 * 9.16 * 9.04 * 8.90 * 8.82 * 10.97 * 12.97 * 12.61 * 14.08 * 14.77 * 14.15 * 11.55 * 7.08 * 4.87 * 1.6 0.2 0 0 0 25.5 4.50 * 9.00 * 8.96 * 8.88 * 8.80 * 8.71 * 8.57 * 8.44 * 8.37 * 10.41 * 12.34 * 12.06 * 13.52 * 14.26 * 13.70 * 11.29 * 7.01 * 4.87 * 1.7 0.2 0 0 0 22.5 4.83 * 9.65 * 9.61 * 9.53 * 9.43 * 9.33 * 9.21 * 9.09 * 8.99 * 11.16 * 13.34 * 13.14 * 14.83 * 15.66 * 14.95 * 12.43 * 7.79 * 5.43 * 1.9 0.2 0 0 0 19.5 4.18 * 8.35 * 8.32 * 8.26 * 8.19 * 8.10 * 8.01 * 7.91 * 7.83 * 9.74 * 11.72 * 11.66 * 13.21 * 13.88 * 13.30 * 11.19 * 7.05 * 4.93 * 1.8 0.2 0 0 0 17 3.38 * 6.75 * 6.74 * 6.71 * 6.67 * 6.62 * 6.56 * 6.49 * 6.43 * 8.05 * 9.71 * 9.69 * 10.94 * 11.50 * 11.15 * 9.47 * 6.01 * 4.23 * 1.5 0.2 0 0 0 15 3.39 * 6.78 * 6.77 * 6.77 * 6.74 * 6.68 * 6.63 * 6.58 * 6.54 * 8.21 * 9.94 * 9.91 * 11.16 * 11.79 * 11.54 * 9.86 * 6.31 * 4.48 * 1.6 0.2 0 0 0 13 3.38 * 6.76 * 6.76 * 6.77 * 6.75 * 6.71 * 6.67 * 6.62 * 6.59 * 8.27 * 10.07 * 10.04 * 11.33 * 12.06 * 11.83 * 10.16 * 6.57 * 4.69 * 1.7 0.2 0 0 0 11 3.36 * 6.73 * 6.73 * 6.72 * 6.71 * 6.70 * 6.68 * 6.65 * 6.61 * 8.28 * 10.08 * 10.11 * 11.50 * 12.27 * 12.03 * 10.38 * 6.78 * 4.88 * 1.7 0.2 0 0 0 9 3.34 * 6.69 * 6.69 * 6.68 * 6.67 * 6.68 * 6.68 * 6.66 * 6.62 * 8.26 * 10.05 * 10.15 * 11.63 * 12.42 * 12.17 * 10.54 * 6.95 * 5.03 * 1.8 0.2 0 0 0 7 3.32 * 6.64 * 6.65 * 6.65 * 6.65 * 6.67 * 6.67 * 6.66 * 6.63 * 8.26 * 10.03 * 10.17 * 11.68 * 12.49 * 12.27 * 10.67 * 7.08 * 5.14 * 1.8 0.2 0 0 0 5 3.32 * 6.64 * 6.66 * 6.67 * 6.67 * 6.67 * 6.68 * 6.67 * 6.64 * 8.27 * 10.03 * 10.18 * 11.71 * 12.54 * 12.34 * 10.76 * 7.17 * 5.23 * 1.8 0.2 0 0 0 3 3.35 * 6.70 * 6.71 * 6.71 * 6.71 * 6.70 * 6.70 * 6.68 * 6.67 * 8.32 * 10.06 * 10.20 * 11.73 * 12.57 * 12.39 * 10.80 * 7.22 * 5.27 * 1.9 0.2 0 0 0 1 1.69 * 3.38 * 3.38 * 3.38 * 3.37 * 3.36 * 3.36 * 3.35 * 3.35 * 4.18 * 5.05 * 5.11 * 5.88 * 6.29 * 6.20 * 5.41 * 3.61 * 2.64 * 0.9 0.1 0 0 0 0Doc No.: DLFLAB-ZY-01-28
FXLED78SFN
Power
Factor
Non-Wrost Case
The total harmonic distortion shall be measured to the 40th order.
The ambient temperature condition was maintained at 25º C ± 1º C. The sample measurements
were made using a digital power meter and power supply. The sample was operated at rated
voltage and was stabilized before measurement. The total harmonic distortion were calculated.
Test Results
Voltage
(Vac)
Power
Factor
Voltage
(Vac)
Current
Wattage
Wrost Case
THD
0.950
7.67%
25.3
The samples were tested according to the ANSI C82.77:2002.
6.10%
4.0 LM-79 Measurement and Test Results
4.3 THD and PF Test
A1
53.0
Sample ID.
Humidity
(%RH)
Model No.
Temperature
(°C)
Test Method
THD
120.02
0.998
277.05
0.293
77.0
Doc No.: DLFLAB-ZY-01-28