Error analysis and solutions

In document BC-5300vet (Page 196-200)

Errors related to the drive board are identified and handled by the host control software in the form of error codes. Note that only some of the error codes reported to the host control software from the dive board can be handled automatically, while some need to be judged with the help of the users or service engineers. When error is reported, the error No. and error names are provided by host control software. The recommended solutions may appear on the screen as the software help information.

When you check for errors, first ensure the board is well fixed and all the error-related connections are properly connected. For example, when “Sample probe action failed” is

error causes are found and new components are replaced, you need to replace the drive board.

The error codes, error names and recommended solutions are listed in the following table.

Table 5-66 Troubleshooting common errors Error

code Error name Error feature

Recommended action

0100 Valve output failed

FPGA does not allow the MCU write data into the valve-pump data register within 50ms, i.e. the serial output status indication bit in the register (which is used to judge the status of valves and pumps of FPGA) is always 1. It might be FPGA error or MCU read register error.

Replace drive board

0101

Valve NO. out of limit (greater than 40 or equal to 0)

Download Valve NO. out of limit (greater than 40 or equal to 0)

Communication abnormal. Please try again. 0201 X motor failed to move to target position

X motor does not move to the pre-set target position after action

1. X motor error, action is interfered. 2. X motor target position photocoupler error. 0202 X motor failed to move to home position after initialization

X motor is not at the home position after initialization 1. X motor error, action is interfered. 2. X motor initial position photocoupler error. 0203 X motor failed to leave home position after initialization

X motor failed to leave home position when initializing

1. X motor error, action is interfered. 2. X motor initial position photocoupler error. 0204 Adjusting X motor to target position failed

X motor does not move to target position after the adjusting command is executed.

1. X motor error, action is interfered. 2. X motor target position

position failed is executed 2. X motor initial position photocoupler error. 0206 Adjusting X motor to leave home position failed

X motor does not leave home position after the adjusting command is executed 1. X motor error, action is interfered. 2. X motor initial position photocoupler error. 0207 X motor adjustment error

X motor adjustment is out of limit [1, 9]

Re-send the X motor adjusting command

0208 X motor adjusting steps out of limit

X motor adjusting steps at the target position are out of limit [-20, 20]

Adjust the X motor again

0209 X motor adjusting end position error

The position in the X motor adjusting end command disagrees with the pre-send adjusting position

Re-send the X motor adjusting command 0211 Y motor failed to move to target position

Y motor does not move to the pre-set target position after action

1. Y motor error, action is interfered. 2. Y motor target position photocoupler error. 0212 Y motor failed to initialize to upper position

Y motor is not at the upper position after initialization 1. Y motor error, action is interfered. 2. Y motor upper position photocoupler error. 0213 Y motor failed to initialize to lower position

Y motor does not move to lower position when initializing

1. Y motor error, action is interfered. 2. Y motor lower position

photocoupler error.

0216

Adjusting Y motor to leave upper position failed

Y motor does not leave upper position after the adjusting command is executed 1. Y motor error, action is interfered. 2. Y motor upper position photocoupler error. 0217 Y motor adjustment error

Y motor adjustment is out of limit [1, 6]

Re-send the Y motor adjusting command

0218 Y motor adjusting steps out of limit

Y motor adjusting steps at the target position are out of limit [-63,63]

Adjust the Y motor again

0219 Y motor adjusting end position error

The position in the Y motor adjusting end command disagrees with the pre-send adjusting position

Re-send the Y motor adjusting command 0220 Sample probe assembly is working

X motor or Y motor action is not finished.

Send the sample probe command after the sample probe action is finished

0221 Cannot adjust sample probe

Can not send the command to move

the sample probe upward/downward/leftward/rightward

by one step before sending the adjusting command.

You can adjust the sample probe only after the adjusting command is sent.

0222 X motor action overtime

The action time is longer than the download time.

0223 X motor action forbidden

X motor is not at the photocoupler position or Y motor is not at the upper photocoupler position

Move the Y motor to upper photocoupler position

0224

X motor does not match current position

X motor current position does not

match the photocoupler position Initialize the X motor

0225 Y motor action overtime

The action time is longer than the download time.

0226 X motor action forbidden

X motor is not at the photocoupler position

Move the X motor to the photocoupler position

0228

Y motor failed to pierce to lower position

Y motor is not at the lower position

after piercing Pierce again

0229

Y motor failed to pierce to upper position

Y motor failed to back to upper

position when piercing Pierce again

0230 X motor

photocoupler error

X motor photocoupler might be damaged Replace X motor photocoupler 0231 Sample probe RBC bath photocoupler error

RBC bath photocoupler might be damaged Replace RBC bath photocoupler 0232 Sample probe WBC bath photocoupler error

WBC bath photocoupler might be damaged Replace WBC bath photocoupler 0233 Sample probe DIFF bath photocoupler error

DIFF bath photocoupler might be damaged

Replace DIFF bath photocoupler

0234 Y motor

photocoupler error

Y motor photocoupler might be damaged Replace X motor photocoupler 0235 Sample probe upper photocoupler error

Upper photocoupler might be damaged

Replace the upper photocoupler

0236

Sample probe lower

photocoupler error

Lower photocoupler might be damaged

Replace the lower photocoupler

0300 Sampling syringe is working

Sampling syringe does not finish the previous action when it is requested for initializing, resetting, aspirating or dispensing.

Command is sent early while the sampling syringe is still working.

1. Sampling syringe photocoupler error