Chapter 2 Experimental Methods
2.5 Experimental Techniques, UHV Chambers and Instrumentation
2.5.1 Temperature Programmed Desorption (TPD)
2.5.1.1 TPD Theory
Temperature-programmed desorption (TPD), also known as thermal desorption
spectroscopy (TDS) is a basic surface science technique that records a thermal desorption
spectrum by monitoring the partial pressure of desorbing species as a function of
temperature. In this technique, a sample, placed in vacuum, is dosed with reactant
molecules and then heated in a controlled manner by using a linear heating rate. Then the
desorbing species are detected by a mass spectrometer. The temperature at which
molecules desorb from a surface depends on the energy with which they are adsorbed,
and the intensity of the signal detected by the mass spectrometer is proportional to the
rate of desorption of the adsorbate. Desorption is an activated process that has Arrhenius
behavior, where the rate constant(ππππ ) of desorption depends on the temperature and is
given by:
ππππ = π΄ ππ₯π (βπΈπππ
π π )
(2.3)
Where πΈπππ is the activation energy for desorption, R is the gas constant, T is the
temperature and A is the pre-exponential factor. During a TPD experiment, as the
temperature increases the rate of desorption increases. However, because the coverage
starts to decrease at the same time, the net result is the appearance of a peak as shown in
Figure 2-6. Typically a linear heating rate is used: T(t) = T0 + π½t, where T is the final
Figure 2-6 Origin of peak features during a TPD experiment
A qualitative TPD profile analysis can be used to gain a lot of information about a
surface reaction. For example:
(i)TPD can be used to determine the desorption kinetics of a reaction The rate of desorption is usually expressed by a rate law of nth order 9
ππππ = β ππ
ππ‘ = ππππ ππ
(2.4)
where ππππ is rate constant for the desorption process, π is the adsorbate coverage and n is the kinetic order of desorption. By substituting the value of ππππ from equation 2.3
equation 2.4 becomes: ππππ = βππ ππ‘ = π΄ exp ( βπΈπππ π π ) ππ (2.5)
which is known as Polanyi-Wigner equation. When the heating rate (π½) is constant, a relationship can be derived for the desorption rate as a function of temperature
(ππππ ) which is described in the following. As described above, typically a linear heating
π = π0+ π½π‘ (2.6)
Taking the first derivative we get,
ππ ππ‘ = π½ β ππ‘ ππ = 1 π½ (2.7)
The desorption rate as a function of temperature ( ππππ ) be written as, ππ ππ= ππ ππ‘ ππ‘ ππ (2.8)
Substituting the value of ππππ‘ from equation 2.7 we get, ππ ππ= 1 π½ ππ ππ‘ (2.9)
Using the value of ππππ‘ from equation 2.5 the above expression becomes, ππ ππ = 1 π½ π΄ exp ( βπΈπππ π π ) ππ (2.10)
At the maximum peak temperature (when T = Tp) the derivative of the rate of
desorption( ππππ‘) becomes zero and the equation 2.10 can be solved assuming that Edes is
independent of coverage, i.e.
π2π ππ‘2 = 0 π€βππ π = ππ As a result, π2π ππ2 = 0 π€βππ π = ππ βπ2π ππ2 = πππβ1 ππ ππ+ ππ πΈπππ π ππ2 = 0 (2.11)
Inserting the values of ππ
ππ from equation 2.10 the above expression becomes:
πππβ11 π½ π΄ exp ( βπΈπππ π π ) ππ+ ππ πΈπππ π ππ2 = 0 (2.12)
Solving the above equation for πΈπ ππππ π2 we get, πΈπππ π ππ2 = 1 π½π΄ exp ( βπΈπππ π π ) πππβ1 (2.13)
For a first order reaction (n=1) the above equation becomes:
πΈπππ π ππ2 = 1 π½π΄ exp ( βπΈπππ π π ) (2.14)
which implies that for a first order process, Tp is independent of the coverage,
while a second order process depends on coverage and Tp decreases with increase in
coverage. If lateral interactions between the adsorbates are ignored, for a second-order
desorption process the peak shape is nearly symmetric, while first-order desorption peak
shapes are asymmetric. Thus the peak shape gives information about the reaction order of
the desorption process. The pre-exponential factor has the units of frequency, and is
given by: π΄ =ππ Ρ ( ππ‘ πππ) (2.15)
where Ρ is plankβs constant over 2Ο, and ππ‘ and π
ππ are the partition functions of the initial and transition state respectively. If the two partition functions are equal, for
example molecules desorb directly from their adsorption site, then π΄ is in the order of molecular vibrational frequencies, ~1013 s-1. If the partition functions are different, e.g. mobile adsorbates diffuse on the surface, A can be significantly different, in this example
smaller.
Also, the peak temperature gives information about the nature of bonding of the
adsorbates to the sample. The peak at lower temperature corresponds to weakly bound
species and the peak at higher temperature corresponds to strongly bound species to the
(ii) Calculating Activation Energy of Desorption
In 1962, Redhead derived a relationship between desorption activation energy
(Edes) and Tp by assuming that A and Edes is independent of coverage and the desorption
kinetics are first order. The Redhead equation10 is given by,
πΈπππ = π ππ[ππ (π΄ππ
π½ ) β 3.64]
(2.16)
This equation is widely used in surface science to calculate the heat of desorption of the
adsorbates by knowing the heating rate, pre-exponential factor and peak temperature. For
a given heating rate, the peak temperature can be obtained from a TPD spectrum and it is
usually assumed that the magnitude of A is in the order of molecular vibrational
frequencies ~ 1013 s-1.
(iii) Relative Coverage Measurement from TPD profile
The integrated area under any desorption profile is proportional to the amount
originally adsorbed, thus it can be used to determine relative coverages of an adsorbate.
(iv) Decomposition Mechanism and Reaction products
TPD may also be used to determine the decomposition mechanisms and reaction
products, by sequentially monitoring several masses. Since mass spectrometric detection
is used, the sensitivity of the technique is good with attainable detection limits below
0.1% of a monolayer of the adsorbate.
2.5.1.2 TPD Chamber
TPD experiments are carried out in the TPD chamber as described below. This
system consists of a 12-inch diameter bell jar chamber, pumped by a combination of a
mechanical pump, turbomolecular pump and an ion pump (250 L/s). The system achieves
quadruple mass spectrometer for TPD experiments, a double-pass cylindrical mirror
analyzer with a built in electron gun for Auger spectroscopy analysis, a X-ray source for
X-ray photoelectron spectroscopy, an ion gun for argon sputtering, respective sources for
sample dosing (leak valves and Knudsen source) and preparation and an ion gauge for
measuring the background pressure of the chamber. Gases are introduced into the
chamber by a gas handling line through leak valves. There are three leak valves attached
to the chamber, two leak valves are used for backfilling the chamber and the other one
has a direct dosing tube attached to it. While dosing reactants through a dosing tube, the
sample position is adjusted such that the dosing tube is directly pointed towards it which
helps to keep the background pressure low. It has been observed that the dosing tubes
enhance the dosing rate by a factor of ~10 or higher in comparison to background dosing.
Therefore reactant gases which pump slowly from the UHV chamber are generally
introduced through a dosing tube, while for introducing gases such as oxygen and argon
for sample cleaning, backfilling is used. A Knudsen source is also attached to the
chamber for dosing compounds with very low or high vapor pressure.
A Kurt J. Lesker XYZ precision sample manipulator is used in this system. The
manipulator can be moved in X, Y and Z directions and also can be rotated through 360Β°
around the Z axis. A Pd (111) single crystal is mounted on a rotatable feedthrough
attached to the manipulator. A cooling system is attached to the manipulator so that the
sample can be cooled to ~80 K by contact with a liquid-nitrogen filled reservoir. Sample
can be cooled by pumping liquid-nitrogen from a Dewar by a mechanical pump through a
capillary tube which is attached to the sample manipulator. The sample is isolated from
DC power supply. The temperature of the sample is measured by a K-type thermocouple
junction spot-welded either to the side or to the back of the sample.
Figure 2-7 (a) Schematics of the TPD-XPS chamber 2 (b) XYZ precision manipulator 2 (c) Sample mounting
(a)
2.5.1.3 Instrumentation for the TPD Experiments
A Dycor model M100M is used for the TPD experiments. During TPD data
collection the sample (front face) is placed very close to the mass spectrometer. Data
collection was controlled by a PC with a data acquisition card by using a Lab View
program. The linear heating rate was controlled by using a proportional-integral-
derivative (PID) control script in Lab View. The computer collects the temperature
readings (in millivolts) from the thermocouple attached to the sample, which passes
through a pre-amplifier gain of 100, and collects the intensity of various masses from the
control electronics of the mass spectrometer. The Lab View script combines temperature
and mass information into one file which can be imported to Origin7 as an ASCII file
where the X column is temperature (in millivolts) and the 5 different masses are collected
in five different Y columns. The temperature in millivolt is converted to Kelvin by using
a script written in Lab View.