• No results found

2.1 Converted aperture metric

The  converted  aperture  metric  is  based  on  the  distances  between  the  MLC  leaves  both   parallel   and   opposed   the   MLC   direction   (Götstedt  et   al,   2015)   (figure   1).   The   metric   gives,   for   a   MLC   opening,   a   complexity   score   value   between   0   (non-­‐complex)   and   1   (complex).  The  distances  are  measured  every  5  mm.  

 

     

Figure 1. For the converted aperture metric the distances between MLC leaves are measured both parallel (green solid line) and opposed (red dashed line) the MLC direction. The MLC leaves are 5 mm wide.

 

The  conversion  function,  f,  penalizes  smaller  distances  compared  to  larger  distances  (eq.  

1).    

𝑓(𝑥) = 1 −𝑒−𝑥    (𝐸𝑞. 1)    

With  the  distances  (mm)  and  the  equivalent  field  size  as  input  arguments  the  function   derives  a  value  between  0  (complex)  and  1  (non-­‐complex).  The  complexity  score  is  then   given  by  the  following  equation  (Götstedt  et  al,  2015):  

 

𝐶𝑜𝑛𝑣𝑒𝑟𝑡𝑒𝑑  𝑎𝑝𝑒𝑟𝑡𝑢𝑟𝑒  𝑚𝑒𝑡𝑟𝑖𝑐 = 1 − 𝑓 𝑑! ∙ 𝑓 𝑎!"    (𝐸𝑞. 2)    

where  di  is  the  measured  distances  and  aeq  is  the  equivalent  square  field  size.  To  get  a   higher  score  for  increasing  complexity,  and  a  lower  score  for  decreasing  complexity,  the  

score  is  subtracted  from  one.  The  calculations  were  performed  in  an  in-­‐house  MatLab® software.  

2.2 Edge area metric

The  edge  area  metric  depends  on  the  length  of  the  edge  of  the  MLC  opening  (Götstedt  et   al,   2015).   As   for   the   converted   aperture   metric,   the   edge   area   metric   calculates   a   complexity   score   based   on   the   MLC   opening.   The   MLC   opening   is   divided   into   two   different  areas.  The  first  area,  Ae,  includes  5  mm  on  both  sides  of  the  MLC  edge  (figure   2).  The  second  area,  Ao  (mm2),  are  defined  as  the  rest  of  the  area  of  the  MLC  opening.  

The  edge  area  metric  is  given  by  the  following  equation:  

 

𝐸𝑑𝑔𝑒  𝑎𝑟𝑒𝑎  𝑚𝑒𝑡𝑟𝑖𝑐 = 𝐴!

𝐴!+ 𝐴!      (𝐸𝑞. 3)    

As  for  the  converted  area  metric,  a  higher  score,  between  zero  and  one  means  that  the   field  is  more  complex.    

Figure 2. Schematic graph of the areas for the edge area metric. Area Ae (green and light green) includes 5 mm on both sides of the edges of the MLC leaves. The remaining area, Ao,

is marked as white. The MLC leves are 5 mm wide.

2.3 Pencil Beam Convolution Algorithm (PBC)

The   pencil   beam   convolution   algorithm   (PBC)   is   implemented   in   Eclipse   TPS   (Varian   medical  systems,  2010)  but  similar  algorithms  are  also  implemented  in  other  treatment   planning  systems.      

The   absorbed   dose   is   calculated   by   convolving   total   energy   released   by   unit   mass   (TERMA)   with   a   pencil   beam   kernel.   The   pencil   beam   kernels   describe   the   dose   deposition  from  all  photons  and  electrons  emerging  from  an  in-­‐finite  part  of  the  beam.    

 

In  every  voxel  the  TERMA  is  calculated  based  on  the  beam  model.  This  model  describes   the  fluence  from  the  accelerator  and  is  calculated  once  and  for  every  unique  accelerator.  

It  is  based  on  measured  parameters  like  MLC  transmission  factors  and  dose  profiles.  

 

The  irradiated  volume  is  divided  into  finite  voxels.  The  voxel  size  is  defined  by  the  grid   size  chosen  by  the  user.  The  dose  calculation  resolution  i.e.  grid  size  can  be  chosen  to  be   0.125   cm,   0.25   cm,   0.5   cm   or   1.0   cm   for   PBC.   Reduced   grid   size   improves   the   dose   calculation  resolution  and  vice  versa.  The  convolving  process  takes  place  in  each  voxel.  

The  total  dose  in  a  voxel  is  calculated  by  superposition  of  the  smaller  dose  depositions.  

 

2.4 Anisotropy Analytical Algorithm (AAA)

The  anisotropic  analytical  algorithm  (AAA)  is  a  3D  pencil  beam  convolution  algorithm   developed  by  Varian  medical  systems  (Varian  medical  systems,  2010).  The  grid  size  can   be  chosen  between  0.1  cm  and  0.5  cm.        

 

The  dose  calculation  model  is  divided  into  a  beam  model  and  a  dose  calculation  model.  

The  first  one  describes  the  beam  in  the  phase  space  plane  by  different  source  models.  

The   source   models   are   the   primary-­‐,   secondary-­‐,   wedge   scattering-­‐,   and   electron   contamination  source  model.    

 

Monoenergetic   pencil   beam   kernels   are   transformed   to   polyenergetic   pencil   beam   kernels.   They   are   also   scaled   based   on   the   electron   density   (heterogenities)   along   the   central  axis  of  the  kernel  and  in  six  lateral  directions.  

 

The  beam  is  divided  into  finite  beamlets.  Every  beamlet  is  convolved  with  a  kernel.  This   is  done  for  every  source.  The  total  energy  in  every  voxel  is  given  by  superposition  of  the   doses  from  the  different  source  models.    

 

2.5 Acuros XB (AXB)

The   acuros   XB   algorithm   is   implemented   in   the   Eclipse   TPS   (Varian   medical   systems,   2010).   AXB   is   considered   to   be   a   fast   algorithm   compared   to   MC   even   though   the   computing  time  compared  to  other  clinical  calculation  algorithms  is  somewhat  longer.  

The  grid  sizes  available  to  be  chosen  are  between  0.1  cm  and  0.3  cm.  The  beam  model  is   the  same  as  for  the  anisotropic  analytical  algorithm  (AAA).  

 

Acuros   XB   calculates   the   absorbed   dose   by   solving   the   linear   Boltzmann   transport   equation   (LBTE)   (Vassiliev  et   al,   2010).   The   LBTE   describes   macroscopically   how   ionizing  particles  for  example  photons  and  electrons  interact  with  different  matter.  

 

LBTE  is  solved  in  Eclipse  by  numerical  methods  (Varian  medical  systems,  2010).  In  the   calculations,   limitations   in   accuracy   are   induced   because   of   the   discrete   variables   in   angle,  energy  and  space.  Acuros  XB  calculates  absorbed  doses  in  heterogenic  material  in   the  same  order  of  accuracy  as  MC  calculations  (Fogliata,  2011).  The  long  computing  time   compared  to  other  calculation  algorithm  is  a  disadvantage  (Vassiliev  et  al,  2010).  

 

2.6 Collapsed cone convolution (CC)

The   collapsed   cone   convolution   (CC)   algorithm   is   implemented   in   the   Oncentra   TPS   (Nucletron).  The  grid  size  can  be  chosen  between  0.1  cm  and  0.5  cm.  The  beam  model  is   made  up  of  two  models:  one  that  describes  the  primary  fluence  and  another  that  takes   into  account  the  fluence  of  head-­‐scatter  components.  These  are  stored  in  separate  2D-­‐  

matrices.  

 

CC  uses  analytical  point  kernels  to  describe  the  dose  deposition  from  primary  photons   in   the   medium   (Ahnesjö,   1989).   A   point   kernel   gives   the   dose   deposition   distribution   from  primary  photons  and  scattered  photons.  The  absorbed  dose  is  calculated  by  a  3D   convolution/superposition   method   like   the   AAA   and   PBC   algorithms.   The   deposited   energy,  within  a  solid  angle  Ω,  is  transported  to  voxels  positioned  along  a  line  inside  a   cylindrical   coordinate   system.   Every   line   defines   an   axis   of   a   cone.   The   line   is   passing   thought   the   middle   point   of   a   3D   voxel   defined   in   a   cartesian   coordinate   system.   The   total   energy   deposited   along   the   line   is   transported   to   this   voxel.   That   means   that   parallel  lines,  one  for  every  cartesian  voxel,  gives  the  total  absorbed  dose  in  the  whole   radiated  volume.  

 

CC  is  an  algorithm  that  effectively  takes  into  account  heterogeneities  due  to  the  use  of  a   cylindrical  coordinate  system.  

Related documents