• No results found

THE IMPACT OF DIFFERENT DOSE CALCULATION ALGORITHMS AND GRID SIZES ON APERTURE-BASED COMPLEXITY METRICS

N/A
N/A
Protected

Academic year: 2022

Share "THE IMPACT OF DIFFERENT DOSE CALCULATION ALGORITHMS AND GRID SIZES ON APERTURE-BASED COMPLEXITY METRICS"

Copied!
41
0
0

Loading.... (view fulltext now)

Full text

(1)

DEPARTMENT OF RADIATION PHYSICS

THE IMPACT OF DIFFERENT DOSE

CALCULATION ALGORITHMS AND GRID SIZES ON APERTURE-BASED

COMPLEXITY METRICS

M.sc. Thesis, Gothenburg spring 2015

Axel Larsson

Supervisors: Anna Bäcka Ph.D. medical physics

Anna Karlsson Hauera Ph.D. medical physics

a) Department of therapeutic radiation physics, Sahlgrenska University Hospital, Gothenburg

Email: [email protected] Tel: +46 70 616 48 07

(2)

The author is thankful for the great support from his supervisors Anna Bäck and Anna Karlsson Hauer. Special thanks to Magnus Gustavsson, Sahlgrenska University Hospital, for

his support with the film measurement technique and Sebastian Sarudis, Sahlgrenska University Hospital, for the help with the collapsed cone calculations.

Gothenburg, June 2015

(3)

Abstract  

Background: The   use   of   the   IMRT   and   VMAT   technique   enables,   in   many   cases,   an   increased   absorbed   dose   to   the   tumor,   with   a   more   conform   dose   distribution,   compared   to   conventional   radiation   therapy.  Invers  treatment  planning  is  used  to  create  IMRT  and  VMAT   plans.   The   treatment   plan   is   made   up   of   several,   often   small   or   irregular,   MLC   openings.   Small   and   irregular   MLC   openings   increase   the  complexity  in  dose  calculation  and  dose  delivery,  mainly  due  to  the   absence   if   charge   particle   equilibrium   (CPE).     Götstedt  et   al   (2015)   developed   two   different   aperture-­‐based   complexity   metrics:    

converted  aperture  metric  and  edge  area  metric,  and  evaluated  them   for   30   different   static   MLC   openings.   These   metrics   provide   a   complexity   score   for   each   MLC   opening.   The   metrics   were   evaluated   for  the  anisotropy  analytical  algorithm  (AAA)  and  0.25  cm  calculation   grid  size  using  three  EPID  measurements  and  one  film  measurement.  

A  correlation  between  the  complexity  scores  for  both  metrics  and  the   difference   between   calculated   and   measured   dose   distribution   was   found,  for  both  EPID-­‐  and  film  measurements.    

 

In   this   this   study   it   was   investigated   how   the   two   earlier   described   metrics   correlate   with   the   difference   between   calculated   and   measured   dose   when   the   dose   calculation   is   made   using   the   dose   calculation   algorithms   pencil   beam   convolution   (PBC),   acuros   XB   advanced   calculation   algorithm   (AXB)   and   the   collapsed   cone   algorithm   (CC).   This   study   is   based   on   film   measurements.   The   previous   film   measurement   made   by   Götstedt   et   al   (2015)   was   complemented  by  two  further  film  measurements  in  order  to  take  into   account   the   precision   in   the   measurement   procedure.   The   impact   of   different   dose   calculation   resolutions   on   the   correlation   was   also   investigated.  

 

Method: The   30   different   MLC   openings   described   in   Götstedt  et   al   (2015)   were   used.   These   MLC   openings   were   measured   on   two   separate  occasions  using  a  6  MV  photon  beam  and  Gafchromic®  EBT3   film.   The   films   were   calibrated,   scanned   and   analysed   in   RIT113   (Radiological   imaging   technology).   In   the   analysis,   measured   and   calculated  absorbed  dose  was  compared.  The  percentage  of  pixels  that   did   not   deviate   (calculated   vs   measured)   more   than   5   %   and   3   %   normalized   to   the   calculated   maximum   absorbed   dose   in   each   of   the  

(4)

Results: High   correlation   coefficients   i.e.   pearsons´s   r-­‐values   were   found  for  each  metric  and  all  the  dose  calculation  algorithms  tested.  A   lower  correlation  coefficient  was  found  for  Acuros  XB.  A  higher  dose   calculation   resolution   generated   a   better   match   between   calculated   and   measured   dose   distribution.   The   complexity   scored   for   both   the   metrics   correlated   with   the   evaluation   results   for   all   examined   calculation   resolutions.   The   calculated   dose   differences   for   the   different  MLC  openings  were  dependent  on  the  examined  calculation   algorithm.   The   PBC   algorithm   performed   generally   worse   than   other   algorithms   and   AAA   performed   generally   better   than   the   other   algorithms.  

 

Conclusion: The   conclusions   from   Götstedt   et   al   (2015)   were   confirmed   for   AAA.   The   average   spread   of   the   dose   difference   pass   rate   values   from   the   three   different   film   measurements,   i.e.   the   standard   deviation,   was   generally   within   3   %   for   the   5   %   dose   difference  pass  rate  criterion  and  4  %  for  the  3  %  dose  difference  pass   rate  criterion.  The  scores  for  the  converted  aperture  and  the  edge  area   metric   correlate   with   the   dose   difference   between   measured   and   calculated  dose  distribution,  also  for  the  PBC,  AXB  and  CC  algorithm,  as   well   as   for   the   AAA   calculations   with   different   dose   calculation   resolution.  A  slightly  lower  r-­‐value  was  found  for  AXB  compared  to  the   other  algorithms.    

(5)

Abbreviations

AAA  –  Anisotropic  analytical  algorithm   AXB  –  Acuros  XB    

CC  –  Collapsed  cone  convolution   CPE  –  Charge  particle  equilibrium  

DICOM  –  Digital  imaging  and  communications  in  medicine     EPID  –  Electronic  portal  imaging  device  

IMRT  –  Intensity-­‐modulated  radiation  therapy   LBTE  –  Linear  Boltzmann  transport  equation   MC  –  Monte  carlo  

MLC  –  Multi  leaf  collimator   MU  –  Monitor  unit  

OAR  –  Organs  at  risk   OD  –  Optical  density  

PBC  –  Pencil  beam  convolution   QA  –  Quality  assurance  

QC  –  Quality  control   SD  –  Standard  deviation   SNR  –  Signal  to  noise  ratio  

TERMA  –  Total  energy  released  per  unit  mass   TPS  –  Treatment  planning  system  

VMAT  –  Volumetric-­‐modulated  arc  therapy  

(6)

List of Contents

1. Introduction ... 1

1.1 Aim ... 3

2. Theory ... 4

2.1 Converted aperture metric ... 4

2.2 Edge area metric ... 5

2.3 Pencil Beam Convolution Algorithm (PBC) ... 5

2.4 Anisotropy Analytical Algorithm (AAA) ... 6

2.5 Acuros XB (AXB) ... 6

2.6 Collapsed cone convolution (CC) ... 7

3. Material and methods ... 8

3.1 The Different MLC Openings ... 8

3.2 Film Measurement ... 9

3.3 Calculation of Dose Distributions ... 11

3.4 Evolution of the Dose Difference ... 12

4. Results ... 14

4.1 Precision of the Film Measurement Procedure ... 14

4.2 Different Dose Calculation Algorithms ... 19

4.3. Different Grid Size ... 23

5. Discussion ... 26

6. Conclusions ... 29

7. References ... 30

Appendix ... 32

(7)

1. Introduction

Radiation  therapy  plays  an  important  role  in  the  treatment  of  cancer,  and  it  is  used  as  a   treatment   for   more   than   half   of   all   the   cancer   patients   world   wide   (IAEA,   2004).   The   radiation   treatment   design   is   simulated   in   a   treatment   planning   system   (TPS)   prior   delivery  to  patient.  The  dose  distribution  is  calculated  and  optimized  in  the  TPS.  

 

Intensity-­‐modulated   radiation   therapy   (IMRT)   is   a   radiation   technique   where   the   fluence  of  the  beam  is  non-­‐uniform  (Kahn,  2010).  With  IMRT  the  fluence  in  a  beam  at  a   static   gantry   angle   can   be   adjusted   during   radiation   delivery   based   on   user-­‐defined   constrains.  This  enables  higher  and  more  conformal  absorbed  dose  distributions  to  the   tumor   with   less   absorbed   dose   to   OAR.   Another   radiation   technique   is   the   volumetric   modulated  arc  therapy  (VMAT).  For  VMAT  the  treatment  is  delivered  at  the  same  time   as   the   gantry   is   rotating   around   the   patients   and   with   a   continuously   changing   MLC   shape,  dose  rate  and  angle  speed  of  the  gantry  (O´Daniel  et  al,  2012).    

 

VMAT  has  the  potential  to  deliver  a  conformal  dose  distribution  with  less  monitor  units   (MU)   and   shorter   treatment   time   compared   to   IMRT   technique   (Younge  et   al,   2012).  

Shorter   treatment   time   reduces   the   risk   of   patient   movement   during   treatment   and   gives  the  patient  an  increased  comfort.    

 

Unlike  conventional  radiation  therapy,  IMRT  and  VMAT  often  have  a  greater  ability  to   increase   the   dose   to   the   target   with   a   reduced   dose   to   healthy   tissue   (McNiven  et   al,   2010).  However,  the  characteristics  of  the  IMRT/VMAT  treatment  make  higher  demands   on  treatment  planning,  dose  delivery  and  quality  assurance.  

 

An  IMRT/VMAT  treatment  plan  is  created  using  an  optimization  algorithm  in  the  TPS.  

This  optimization  procedure  creates  a  treatment  plan  based  on  user-­‐defined  constrains   that   tells   the   system   what   the   prescribed   absorbed   dose   should   be   for   the   target   and   maximum   permissible   absorbed   dose   for   the   OAR.   This   generates   a   treatment   plan   composed  of  unique  MLC  openings  with  different  sizes  and  shapes  (Kahn,  2010).  

 

During   optimizing   of   VMAT   or   IMRT   plans   the   TPS   creating   more   small   and   irregular   shaped   MLC   openings   compared   to   conventional   treatment   planning     (Younge  et   al,   2012).  Small  or  irregular  MLC  openings  make  it  more  difficult  for  the  TPS  to  calculate  an   accurate  dose  distribution  due  to  regions  lacking  charge  particle  equilibrium  (CPE).  Fog   et   al   (2011)   showed   that   apertures   consisting   of   small   subfields   or   large   fields   containing   isolated   MLC   leaves   could   give   rise   to   significant   dose   calculation   errors.  

Small  MLC  openings  also  lead  to  a  more  pronounced  dependence  on  the  position  of  the   MLC  leafs  during  delivery  and  the  MLC  modeling  in  the  TPS  (LoSasso  et  al,  1998).    

 

(8)

dose   delivery   from   the   treatment   machine   are   in   tolerance.   This   is   done   by   quality   assurance   (QA)   (Nelms  et   al,   2011).   The   QA   includes   IMRT   quality   control   (QC).   A   common   way   for   IMRT   QC   is   to   compare   the   TPS   calculated   dose   distribution   with   corresponding  measured  dose  distribution  in  a  phantom.  Common  QC  methods  that  are   in   use   nowadays   have   been   found   to   miss   relevant   clinical   differences   between   calculated   and   delivered   dose   distributions   (Nelms  et   al,   2011;   Götstedt   et   al,   2015;  

Nilsson  et  al,  2013).  

 

By   taking   advantage   of   complexity   metrics   that   describes   the   calculation   and   delivery   complexity   by   calculating   a   complexity   score   in   the   dose   calculation   process,   complex   MLC  openings  can  be  avoided  before  the  treatment  plans  are  approved  (Götstedt  et  al,   2015).  Complexity  metrics  can  also  be  used  as  a  supplement  and  simplification  of  the  QA   process  by  giving  a  signal  to  the  physicist  of  which  QA  method  that  may  be  required  and   which  treatment  plans  that  need  detailed  examination.  QA  measurements  require  a  lot   of  staff  time  and  demands  machine  time.  

 

Younge  et  al  (2012)  showed  that  highly  complex  MLC  openings  are  not  always  needed   to   create   a   clinically   accepted   VMAT   plan,   instead   it   is   an   unwanted   effect   of   the   optimizing  process  in  the  TPS.  Oliver  et  al  (2011)  concluded  that  decreasing  the  MU:s   and   creating   MLC   openings   with   increased   opening   area   should   lead   to   more   stable   treatment  plans  with  preserved  quality.  

 

Complexity   metrics   can   also   be   used   in   the   optimization   procedure   to   force   the   optimizer  to  create  treatment  plans  with  less  complex  apertures  (Younge  et  al,  2012).  

Younge  at   el   (2012)   developed   a   function   that   was   integrated   in   the   optimization   process.   The   function   penalized   small   and   irregular   MLC   openings. The   study   verified   that   penalizing   small   and   irregular   MLC   openings   could   give   an   increased   agreement   between  calculated  and  measured  absorbed  dose  with  negligible  changes  in  the  planed   dose  to  target  and  OAR.  

 

The  two  aperture-­‐based  metrics  evaluated  in  this  study  is  the  converted  aperture  metric   based  on  the  distance  between  the  MLC  leave  positions  and  the  edge  area  metric  based   on  the  circumference  of  the  MLC  opening  in  relation  to  the  area  of  the  opening.  These   metrics  were  developed  in  a  previously  study  (Götstedt  et  al,  2015).  The  study  examined   the   correlation   between   the   metric   scores   and   the   difference   between   calculated   and   measured  absorbed  dose.  The  evaluations  were  performed  with  three  electronic  portal   imaging   device   (EPID)   measurements   and   one   film   measurement   and   the   calculations   were  performed  with  the  anisotropic  analytical  algorithm  (AAA). The  results  indicated  a   correlation  between  the  metrics  and  the  percentage  of  pixels  that  did  not  deviate  more   than  3  %  or  5  %  normalized  to  the  maximum  calculated  absorbed  dose.  

 

The  correlations  were  calculated  with  pearsons´s  r-­‐value.  The  pearsons´s  r-­‐value  gives   the  linear  dependence  i.e.  correlation  between  two  variables  by  giving  a  value  between    

(9)

-­‐1   and   1   (Djurfeldt  et   al,   2010).   -­‐1   och   1   means   a   negative   or   positive   correlation   respectively.  A  value  close  to  0  means  that  no  correlation  exists.  

 

This  project  is  a  complement  and  an  extension  of  the  study  by  Götstedt  et  al  (2015)  for   the   converted   aperture   and   the   edge   are   metric.   The   influence   on   the   correlation   between   the   metric   scores   and   the   difference   between   measured   and   calculated   dose   distribution  were  investigated  when  different  dose  calculation  algorithms  and  grid  sizes   were  used.  

 

1.1 Aim

§ Reproduce   the   film   measurements   from   Götstedt  el   at   (2015)   to   study   the   precision  for  the  measurement  procedure.  

 

§ Evaluate  the  correlation  between  the  two  different  complexity  metrics,  edge  area   and  converted  aperture  metric,  and  the  dose  difference  (i.e.  difference  between   measured   and   calculated   dose   distributions)   between   measured   and   calculated   dose   distributions   for   the   pencil   beam   convolution   (PBC),   collapsed   cone   (CC)   and  acuros  XB  (AXB)  algorithms.  

 

§ Study  the  impact  of  the  dose  calculation  grid  size  on  the  correlation  between  the   dose  differences  and  the  two  complexity  metrics  when  using  the  AAA  algorithm.  

(10)

2. Theory

2.1 Converted aperture metric

The  converted  aperture  metric  is  based  on  the  distances  between  the  MLC  leaves  both   parallel   and   opposed   the   MLC   direction   (Götstedt  et   al,   2015)   (figure   1).   The   metric   gives,   for   a   MLC   opening,   a   complexity   score   value   between   0   (non-­‐complex)   and   1   (complex).  The  distances  are  measured  every  5  mm.  

 

     

Figure 1. For the converted aperture metric the distances between MLC leaves are measured both parallel (green solid line) and opposed (red dashed line) the MLC direction. The MLC leaves are 5 mm wide.

 

The  conversion  function,  f,  penalizes  smaller  distances  compared  to  larger  distances  (eq.  

1).    

𝑓(𝑥) = 1 −𝑒−𝑥    (𝐸𝑞. 1)    

With  the  distances  (mm)  and  the  equivalent  field  size  as  input  arguments  the  function   derives  a  value  between  0  (complex)  and  1  (non-­‐complex).  The  complexity  score  is  then   given  by  the  following  equation  (Götstedt  et  al,  2015):  

 

𝐶𝑜𝑛𝑣𝑒𝑟𝑡𝑒𝑑  𝑎𝑝𝑒𝑟𝑡𝑢𝑟𝑒  𝑚𝑒𝑡𝑟𝑖𝑐 = 1 − 𝑓 𝑑! ∙ 𝑓 𝑎!"    (𝐸𝑞. 2)    

where  di  is  the  measured  distances  and  aeq  is  the  equivalent  square  field  size.  To  get  a   higher  score  for  increasing  complexity,  and  a  lower  score  for  decreasing  complexity,  the  

(11)

score  is  subtracted  from  one.  The  calculations  were  performed  in  an  in-­‐house  MatLab® software.  

2.2 Edge area metric

The  edge  area  metric  depends  on  the  length  of  the  edge  of  the  MLC  opening  (Götstedt  et   al,   2015).   As   for   the   converted   aperture   metric,   the   edge   area   metric   calculates   a   complexity   score   based   on   the   MLC   opening.   The   MLC   opening   is   divided   into   two   different  areas.  The  first  area,  Ae,  includes  5  mm  on  both  sides  of  the  MLC  edge  (figure   2).  The  second  area,  Ao  (mm2),  are  defined  as  the  rest  of  the  area  of  the  MLC  opening.  

The  edge  area  metric  is  given  by  the  following  equation:  

 

𝐸𝑑𝑔𝑒  𝑎𝑟𝑒𝑎  𝑚𝑒𝑡𝑟𝑖𝑐 = 𝐴!

𝐴!+ 𝐴!      (𝐸𝑞. 3)    

As  for  the  converted  area  metric,  a  higher  score,  between  zero  and  one  means  that  the   field  is  more  complex.    

Figure 2. Schematic graph of the areas for the edge area metric. Area Ae (green and light green) includes 5 mm on both sides of the edges of the MLC leaves. The remaining area, Ao,

is marked as white. The MLC leves are 5 mm wide.

2.3 Pencil Beam Convolution Algorithm (PBC)

The   pencil   beam   convolution   algorithm   (PBC)   is   implemented   in   Eclipse   TPS   (Varian   medical  systems,  2010)  but  similar  algorithms  are  also  implemented  in  other  treatment   planning  systems.      

(12)

The   absorbed   dose   is   calculated   by   convolving   total   energy   released   by   unit   mass   (TERMA)   with   a   pencil   beam   kernel.   The   pencil   beam   kernels   describe   the   dose   deposition  from  all  photons  and  electrons  emerging  from  an  in-­‐finite  part  of  the  beam.    

 

In  every  voxel  the  TERMA  is  calculated  based  on  the  beam  model.  This  model  describes   the  fluence  from  the  accelerator  and  is  calculated  once  and  for  every  unique  accelerator.  

It  is  based  on  measured  parameters  like  MLC  transmission  factors  and  dose  profiles.  

 

The  irradiated  volume  is  divided  into  finite  voxels.  The  voxel  size  is  defined  by  the  grid   size  chosen  by  the  user.  The  dose  calculation  resolution  i.e.  grid  size  can  be  chosen  to  be   0.125   cm,   0.25   cm,   0.5   cm   or   1.0   cm   for   PBC.   Reduced   grid   size   improves   the   dose   calculation  resolution  and  vice  versa.  The  convolving  process  takes  place  in  each  voxel.  

The  total  dose  in  a  voxel  is  calculated  by  superposition  of  the  smaller  dose  depositions.  

 

2.4 Anisotropy Analytical Algorithm (AAA)

The  anisotropic  analytical  algorithm  (AAA)  is  a  3D  pencil  beam  convolution  algorithm   developed  by  Varian  medical  systems  (Varian  medical  systems,  2010).  The  grid  size  can   be  chosen  between  0.1  cm  and  0.5  cm.        

 

The  dose  calculation  model  is  divided  into  a  beam  model  and  a  dose  calculation  model.  

The  first  one  describes  the  beam  in  the  phase  space  plane  by  different  source  models.  

The   source   models   are   the   primary-­‐,   secondary-­‐,   wedge   scattering-­‐,   and   electron   contamination  source  model.    

 

Monoenergetic   pencil   beam   kernels   are   transformed   to   polyenergetic   pencil   beam   kernels.   They   are   also   scaled   based   on   the   electron   density   (heterogenities)   along   the   central  axis  of  the  kernel  and  in  six  lateral  directions.  

 

The  beam  is  divided  into  finite  beamlets.  Every  beamlet  is  convolved  with  a  kernel.  This   is  done  for  every  source.  The  total  energy  in  every  voxel  is  given  by  superposition  of  the   doses  from  the  different  source  models.    

 

2.5 Acuros XB (AXB)

The   acuros   XB   algorithm   is   implemented   in   the   Eclipse   TPS   (Varian   medical   systems,   2010).   AXB   is   considered   to   be   a   fast   algorithm   compared   to   MC   even   though   the   computing  time  compared  to  other  clinical  calculation  algorithms  is  somewhat  longer.  

The  grid  sizes  available  to  be  chosen  are  between  0.1  cm  and  0.3  cm.  The  beam  model  is   the  same  as  for  the  anisotropic  analytical  algorithm  (AAA).  

 

(13)

Acuros   XB   calculates   the   absorbed   dose   by   solving   the   linear   Boltzmann   transport   equation   (LBTE)   (Vassiliev  et   al,   2010).   The   LBTE   describes   macroscopically   how   ionizing  particles  for  example  photons  and  electrons  interact  with  different  matter.  

 

LBTE  is  solved  in  Eclipse  by  numerical  methods  (Varian  medical  systems,  2010).  In  the   calculations,   limitations   in   accuracy   are   induced   because   of   the   discrete   variables   in   angle,  energy  and  space.  Acuros  XB  calculates  absorbed  doses  in  heterogenic  material  in   the  same  order  of  accuracy  as  MC  calculations  (Fogliata,  2011).  The  long  computing  time   compared  to  other  calculation  algorithm  is  a  disadvantage  (Vassiliev  et  al,  2010).  

 

2.6 Collapsed cone convolution (CC)

The   collapsed   cone   convolution   (CC)   algorithm   is   implemented   in   the   Oncentra   TPS   (Nucletron).  The  grid  size  can  be  chosen  between  0.1  cm  and  0.5  cm.  The  beam  model  is   made  up  of  two  models:  one  that  describes  the  primary  fluence  and  another  that  takes   into  account  the  fluence  of  head-­‐scatter  components.  These  are  stored  in  separate  2D-­‐  

matrices.  

 

CC  uses  analytical  point  kernels  to  describe  the  dose  deposition  from  primary  photons   in   the   medium   (Ahnesjö,   1989).   A   point   kernel   gives   the   dose   deposition   distribution   from  primary  photons  and  scattered  photons.  The  absorbed  dose  is  calculated  by  a  3D   convolution/superposition   method   like   the   AAA   and   PBC   algorithms.   The   deposited   energy,  within  a  solid  angle  Ω,  is  transported  to  voxels  positioned  along  a  line  inside  a   cylindrical   coordinate   system.   Every   line   defines   an   axis   of   a   cone.   The   line   is   passing   thought   the   middle   point   of   a   3D   voxel   defined   in   a   cartesian   coordinate   system.   The   total   energy   deposited   along   the   line   is   transported   to   this   voxel.   That   means   that   parallel  lines,  one  for  every  cartesian  voxel,  gives  the  total  absorbed  dose  in  the  whole   radiated  volume.  

 

CC  is  an  algorithm  that  effectively  takes  into  account  heterogeneities  due  to  the  use  of  a   cylindrical  coordinate  system.  

(14)

3. Material and methods

3.1 The Different MLC Openings

30  MLC  openings  of  varying  complexity  were  used  in  this  study.  The  MLC  openings  were   described  in  a  previous  study  (Götstedt  et  al,  2015).  The  MLC  openings  have  different   shapes   and   sizes   and   were   created   in   the   EclipseTM   treatment   planning   system   (v.  

11.0.47,   Varian   Medical   Systems).   The   MLC   openings   were   divided   into   six   series,   A-­‐F   (figure   3).   In   each   series   the   six   openings   were   numbered   1-­‐5.   Increased   number   indicates  increased  complexity  within  the  series.  The  circumference  and  area  of  series  A   are  decreasing,  while  the  shape  of  the  MLC  opening  is  retaining  a  square  shape.  The  jaws   of  series  A  form  a  10  cm  x  10  cm  large  area.  In  series  B  the  MLC  opening  area  is  slightly   decreasing   with   a   considerable   increased   circumference.   Series   D   simulates   small   subfields.   Series   D   and   E   have   a   constant   MLC   opening   area,   with   an   increasing   circumference.   Series   C   and   F   have   a   constant   circumference,   with   a   decreasing   MLC   opening  area.  

Figure 3: 30 different MLC openings described by Götstedt et al (2015) were used in this study. The MLC openings were divided into five series A-F. Increased number

(15)

3.2 Film Measurement

Film  dosimetry  was  carried  out  using  Gafchromic®  EBT3  film.  70  films  were  prepared:  

60   films   for   the   measurement   of   the   30   MLC   openings   in   two   different   occasions   and   further  10  films  for  the  calibration.    

 

Two   film   measurements   were   performed   in   this   study.   Each   film   measurement   was   done  on  three  subsequent  days.  During  day  one  all  the  40  films  were  scanned  with  an   Epson  Expression  1650  Pro  scanner  to  get  information  about  the  unique  features,  such   as   irregularities   and   transparency,   in   each   pixel   for   each   film.   For   each   scan   the   films   were   placed   in   the   same   reproducible   way   in   the   scanner   and   four   scans   were   performed.   The   first   scan   was   for   heating   the   film   and   this   scan   was   disregarded   in   further  analysis.  The  other  three  directly  subsuquent  scans,  that  were  made  for  each  of   the   films   during   a   day   of   irradiation   were   averaged   to   increase   signal   to   noise   ratio   (SNR).   This   was   made   in   an   in-­‐house   developed   program   written   in   Matlab©   (Mathworks).    

   

On  the  same  day  as  the  scanning,  all  films  were  irradiated  to  a  uniform  absorbed  dose  of   2  Gy.  It  was  done  to  obtain  the  radiation  sensitivity  of  each  film,  which  was  a  part  of  the   calibration.  A  6  MeV  photon  beam  was  used  to  deliver  the  dose  (Cliniac®  iX  linac,  Varian   medical  systems).  The  films  were  placed,  perpendicular  to  the  beam,  at  10  cm  depth  in  a   30   cm   ×   35   cm   solid-­‐water   phantom.   To   ensure   complete   backscatter   7   cm   of   solid   water  was  placed  under  the  film  (figure  4).  Two  films  were  placed  on  top  of  each  other   and  irradiated  at  the  same  time.  The  gantry  angle  was  set  to  0  degrees,  the  field  size  to   40  cm  ×  40  cm  and  the  source  to  skin  distance  (SSD)  to  90  cm.  The  treatment  table  was   placed  in  90  or  270  degrees  during  the  delivery  of  the  first-­‐half  of  the  dose.  Then,  the   table  was  rotated  180  degrees  before  the  second  half  of  the  dose  was  delivered.  This  was   done  to  give  the  films  as  homogeneous  absorbed  dose  as  possible  by  taking  into  account   possible  skewness  in  the  beam  from  the  accelerator.  The  number  of  MUs  to  be  delivered   was  calculated  according  to  equation  4:  

 

     𝑀𝑈 = 𝐷 (1.172

120 )

×  𝑀!"#$%      (𝐸𝑞. 4)    

 

D  (Gy)  is  the  absorbed  dose  to  deliver  and  !.!"#!"#    (Gy/MU)  is  the  reference  output.  Mratio  is   given  by  equation  5.  

𝑀!"#$% =𝑀 ∙ 𝐾!"

𝑀!"#      (𝐸𝑞. 5)    

𝑀(C)   is   the   averaged   value   of   three   output   measurements.   The   Mratio   was   calculated  

(16)

ionizing  chamber  at  10  cm  depth  in  a  solid-­‐water  phantom  with  the  dimensions  20  cm  ×   20   cm.   The   field   size   was   10   cm   ×   10   cm   and   the   SSD   was   90   cm.   The   linac   was   calibrated  to  deliver  1  Gy  at  10  depth  in  water  for  a  field  opening  of  10  cm  ×  10  cm  at   SSD   90   cm.   In   each   irradiation   120   MU   was   delivered.  Mref  is   the   reference   value.  

Pressure  (mmHg)  and  temperature  (oC)  were  measured  at  each  occurence  to  give  the   pressure-­‐  and  temperature  factor  KTP.  

 

   

Day  two,  more  than  18  hours  after  the  first  irradiation,  the  films  where  scanned  again.  

The  scan  was  performed  in  the  same  way  as  previously  describe.  Later  on  the  same  day,   day  2,  the  measurement  started  by  irradiating  each  of  the  30  films  with  one  of  the  30   different   MLC   openings.   The   number   of   MUs   for   the   MLC   openings   were   calculated   to   deliver  about  2  Gy  in  the  center  of  the  MLC  opening  at  10  cm  depth  with  a  SSD  of  90  cm.  

The   dose   and   MU   calculations   were   carried   out   in   EclipseTM.   The   calculated   MUs   are   listed  in  table  5,  in  appendix.  Each  of  the  films  was  placed  in  the  same  reproducible  way   during  irradiation.    

 

The  double  exposure  method  was  used  to  calibrate  the  films  to  interconnect  the  optical   density   with   absorbed   dose.   The   principles   are   the   same   as   described   by   Zhu  et   al   (1997).  A  calibration  curve  was  calculated  according  to  equation  6.  The  purpose  of  the   calibration   procedure   was   to   fit   a   polynome   by   tuning   the   parameters   A,   B   and   C   to   correlate   the   absorbed   dose   with   the   optical   density.   Parameter   A   and   B   applies   for   every   batch   and   C   gives   the   sensitivity   in   each   pixel   of   every   film.   The   parameter   C   is   produced  from  the  2  Gy  exposure  on  day  one.  

 

𝑂𝐷 =𝐴×𝑑𝑜𝑠𝑒

𝐵×𝑑𝑜𝑠𝑒+ 𝐶×𝑑𝑜𝑠𝑒        (𝐸𝑞. 6)    

Figure 4. A photo of the linac and the couch. The gantry angel was at 0 degrees and the couch at 90 or 270 degrees angle during day one. The couch is positioned at 270 degrees in the photo (left). The 40 films were placed at 10 cm depth in a solid-water phantom with 10 cm solid-water on top of the films (right).

(17)

On  day  two,  after  the  second  scan  and  after  the  delivery  of  the  30  MLC  openings,  10  films   for  the  calibration  were  irradiated  to  a  predetermined  absorbed  dose.  One  film  was  left   unirradiated.  The  MUs  were  calculated  as  previously  describe  according  to  equation  4   and  are  listed  in  table  1.  

     

 

On   day   three,   more   than   18   hours   after   the   second   irradiation,   all   the   40   films   were   scanned  again.  The  optical  density  for  the  10  individual  films  irradiated  with  different   absorbed  doses  was  measured  from  a  2.5  cm  ×  2.5  cm  region-­‐of-­‐interest  (ROI)  in  the   center   of   the   10   films.   An   in-­‐house   developed   software   in   Matlab©   was   used.   The   parameters  A  and  B  were  calculated  in  Excel  (Microsoft  corporation)  and  then  applied   to  the  films  with  an  in-­‐house  developed  software  in  Matlab©.  

3.3 Calculation of Dose Distributions

The  dose  distributions  from  the  different  MLC  openings  were  calculated  with  the  AAA   (v.10.0.28),   PBC   (v.10.0.28)   and   AXB   (v.11.0.31)   algorithms   in   EclipseTM   (v.11.0.47,   Varian  medical  systems)  and  with  the  CC  (v.4.0)  in  Oncentra©  (v.5.3,  Nucletron).    

 

The   phantom   used   in   the   TPS   was   the   same   as   used   by   Götstedt  et   al   (2015).   The   dimensions   of   the   phantom   were   30   cm   ×   30   cm   ×   30   cm   (figure   5).   The   phantom   material   was   set   to   water   in   the   TPS.   The   isocenter   for   the   individual   MLC   openinges   were  placed  at  10  cm  depth  centrally  in  the  phantom.  The  chosen  dose  calculation  grid   sizes   for   the   AAA   algorithm   were   0.125   cm,   0.25   cm   and   0.5   cm   and   for   the   other   algorithms   the   grid   size   was   set   to   0.25   cm.   The   calculated   coronal   absorbed   dose   distributions  at  10  cm  depth,  for  the  individual  MLC  openings,  were  exported  to  RIT113   (v.5.4)  for  evaluation.  The  dose  distributions  exported  from  the  TPS  had  the  dimensions   22.59  cm  ×  28.39  cm  and  the  resolution  in-­‐plane  was  0.056  cm  ×  0.056  cm.  Thus,  the   resolution  of  the  exported  data  was  considerably  higher  than  the  resolution  of  any  of  the   calculated  grid  sizes.  

Film Dose (Gy) #MU0

1   0   0  

2   0.28   26  

3   0.56   52  

4   0.75   77  

5   1.00   103  

6   1.50   155  

7   2.01   207  

8   3.00   310  

9   3.00   413  

10   5.00   517  

Table 1. Absorbed dose with corresponding number of MU used for the calibration of the films.

(18)

 

3.4 Evolution of the Dose Difference

The   film   measurements   were   evaluated   pixel-­‐by-­‐pixel   in   RIT113   (v.5.4,   Radiological   Imaging  technology  INC).  The  measured  absorbed  dose  data  was  individually  imported   and  registrered  with  the  corresponding  calculated  absorbed  dose  data  (DICOM-­‐RD)  for   each  MLC  opening.  When  the  films  were  imported,  the  calibration  curve  was  applied  to   convert  the  blackening  of  the  film  to  absorbed  dose.  A  median  9  pixels  ×  9  pixels  filter   was  applied  for  noise  reduction  of  the  measured  dose  data.    

 

The  registration  between  the  measured  and  the  calculated  dose  distributions  was  done   with  a  four-­‐point-­‐match.  The  registration  points  was  saved  and  used  when  matching  the   same   measured   data   with   another   corresponding   calculated   dose   distribution.   New   templates  were  created  for  the  films  in  the  second  measurement.  

 

Orthogonal  dose  profiles,  horizontally  and  vertically,  were  used  to  verify  the  registration   (figure  6).  If  the  dose  profiles  did  not  match  symmetrically,  the  points  in  the  four-­‐point-­‐

match   were   relocated   in   accordance   with   the   result   of   the   profile   evaluations.   The   normalization   was   performed   in   the   central   part   of   the   individual   MLC   opening.   For   D2:D5  a  mean  value  of  the  individual  openings  was  used  for  the  normalization.  In  the  B-­‐

series   and   C-­‐series   the   central   part   in   the   largest   open   part   of   the   MLC   opening   was   used.  Normalization  was  done  with  a  0.5  cm  ×  0.5  cm  ROI  for  all  openings,  except  A5,  D5   and  F5  there  a  0.3  cm  ×  0.3  cm  ROI  was  used  because  of  the  small  MLC  openings,  and  to   avoid  placing  the  normalization  ROI  in  the  sharp  absorbed  dose  gradients.  The  median   value  within  the  ROI  was  used  for  normalization.  A  dose  cut-­‐off  threshold  of  10  %  was   used  to  exclude  pixels  that  had  received  less  than  10  %  of  the  maximum  calculated  dose   in  the  analysis.  Dose  difference  pass  rate  was  used  as  a  measure  of  agreement  between   Figure 5. The 30 cm × 30 cm × 30 cm phantom used in the calculations of the absorbed dose distributions. The A1 MLC opening is visualized with the corresponding calculated 3D dose distribution in EclipseTM. The phantom material was set to water in the TPS.

(19)

the  measured  and  calculated  dose  distribution.  The  dose  difference  pass  rate  gives  the   precentage  of  evaluated  pixels  that  do  not  deviate  more  than  a  selected  precentage.  In   the   evaluation   of   the   dose   differences   in   this   study   and   in   the   study   by   Götstedt  et   al   (2015)  a  3  %  and  a  5  %  dose  difference  pass  rate  criterion  was  used.    

Figure 6. Dose difference analyzes in RIT113. The calculated and the measured coronal dose planes in colored scale (left), the vertical and the horizontal absorbed dose profiles (upper middle and upper right) and the dose difference between the calculated and measured dose data visualized in 3D and 2D (lower middle and lower right).

(20)

4. Results

4.1 Precision of the Film Measurement Procedure

Pass   rate   values   calculated   for   the   different   MLC   openings   differed   depening   on   the   investigated  algorithm.  Pass  rate  values  for  the  criterion  of  5  %  and  3  %  are  shown  in   figure   7   and   8   respectively.   The   SD   was   calculated   from   the   three   film   measurements   and  is  shown  in  the  figures  as  error  bars.  The  pass  rate  values  decreased  with  a  more   complex   MLC   opening.   PBC   showed   in   generall   a   lower   pass   rate   value   regerdless   complexity  compered  to  the  other  algorithms  and  AAA  showed  in  generall  a  higher  pass   rate  value  regardless  complexity.  When  a  part  of  the  field  area  is  complex  (series  C)  it   seems   to   be   not   so   complex   for   the   algorithms,   which   also   emerges   for   Götstedt  et   al   (2015).  

Figure 7. The measured pass rates for the 5 % criterion for the 30 different MLC openings. The pass rates were evaluated for the AAA, PBC, CC and AXB algorithm. The error bars correspond to 1 SD.

0   50   100  

1   2   3   4   5  

5  %-­‐  dd  paserate  (%)  

MLC  opening  

       Series  A  

0   50   100  

1   2   3   4   5  

5  %-­‐  dd  pass  rate  (%)  

MLC  opening  

     Series  B  

0   50   100  

1   2   3   4   5  

5  %-­‐  dd  pass  rate  (%)  

MLC  opening  

Series  C  

0   50   100  

1   2   3   4   5  

5  %-­‐dd  pass  rate  

MLC  opening  

   Series  E  

0   50   100  

1   2   3   4   5  

5  %-­‐  dd  pass  rate  (%)  

MLC  opening  

           Series  F  

0   50   100  

1   2   3   4   5  

5  %-­‐  dd  pass  rate  (%)  

MLC  opening  

Series  D  

1,#87,19# 2,#87,36# 3,#86,01# 4,#83,63# 5,#80,78#

1,#78,11# 2,#76,49#

3,#70,99#

4,#64,77#

5,#53,58#

1,#85,77# 2,#81,55#

3,#74,13#

4,#65,82#

5,#60,63#

1,#81,95# 2,#80,92# 3,#77,40# 4,#74,44#

5,#68,09#

1# 1,5# 2# 2,5# 3# 3,5# 4# 4,5# 5#

E.serien#

AAA# PBC# AcurosXB# CC#

(21)

0   50   100  

1   2   3   4   5  

3  %-­‐  dd  pass  rate  (%)  

MLC  opening  

     Series  A  

0   50   100  

1   2   3   4   5  

3  %-­‐  dd  pass  rate  (%)  

MLC  opening  

     Series  B  

0   50   100  

1   2   3   4   5  

3  %-­‐  dd  pass  rate  (%)  

MLC  opening  

 Series  C  

0   50   100  

1   2   3   4   5  

3  %-­‐  dd  pass  rate  (%)  

MLC  opening  

   Series  D  

0   50   100  

1   2   3   4   5  

3  %-­‐  dd  pass  rate  (%)  

MLC  opening  

   Series  E  

0   50   100  

1   2   3   4   5  

3  %-­‐  dd  pass  rate  (%)  

MLC  opening  

         Series  F  

1,#87,19# 2,#87,36# 3,#86,01# 4,#83,63# 5,#80,78#

1,#78,11# 2,#76,49#

3,#70,99#

4,#64,77#

5,#53,58#

1,#85,77# 2,#81,55#

3,#74,13#

4,#65,82#

5,#60,63#

1,#81,95# 2,#80,92# 3,#77,40# 4,#74,44#

5,#68,09#

1# 1,5# 2# 2,5# 3# 3,5# 4# 4,5# 5#

E.serien#

AAA# PBC# AcurosXB# CC#

     

Figure 8. The measured pass rates for the 3 % criterion for the 30 different MLC openings. The pass rates were evaluated for the AAA, PBC, CC and AXB algorithm. The error bars correspond to 1 SD.

(22)

The   dose   differences   between   measured   and   calculated   absorbed   dose   distributions   were  larger  than  3  %  or  5  %  exclusively  close  to  the  edges  of  the  MLC  opening.  In  figure   9  the  colored  dose  differens  maps  for  the  deviations  between  measured  and  calculated   dose  distributions  for  the  B:4  MLC  opening  are  visualized.  

AAA PBC AXB CC

B:4

Figure 9. Colored dose difference maps from the comparison of the absorbed dose distribution from a film measurement and the calculated dose distributions from the investigated dose calculation algorithms. The color scale gives the precentage dose differens.

(23)

0   20   40   60   80   100  

0,00   0,20   0,40   0,60   0,80   1,00  

5  %-­‐  dd  pass  rate  (%)  

Edge  area  metric  

r  =  -­‐0,84  

0   20   40   60   80   100  

0,00   0,20   0,40   0,60   0,80   1,00  

3  %-­‐  dd  pass  rate  (%)  

Edge  area  metric  

r  =  -­‐0,90  

0   20   40   60   80   100  

0,00   0,20   0,40   0,60   0,80   1,00  

3  %-­‐  dd  pass  rate  (%)  

Converted  aperture  metric  

r  =  -­‐0,85      

0   20   40   60   80   100  

0,00   0,20   0,40   0,60   0,80   1,00  

5  %-­‐  dd  pass  rate  (%)  

Converted  aperture  metric  

r  =  -­‐0,80  

!"!!!!!!!

!10,00!!!!!

!20,00!!!!!

!30,00!!!!!

!40,00!!!!!

!50,00!!!!!

!60,00!!!!!

!70,00!!!!!

!80,00!!!!!

!90,00!!!!!

0,00! 0,10! 0,20! 0,30! 0,40! 0,50! 0,60! 0,70! 0,80! 0,90! 1,00!

5"%"pass"rate"

Converted"Aperture"Metric"

Pencil"Beam"

Serie!A! Serie!B! Serie!C! Serie!D! Serie!E! Serie!F!

r"="90,89"

The   results   for   AAA   from   Götstedt  et   al   (2015)   was   successfully   reproduced   for   the   investigated  complexity  metrics.  Pass  rates,  with  corresponding  standard  deviation  for   the   evaluated   dose   differences   between   the   measured   and   calculated   absorbed   dose   distributions  for  the  30  MLC  openings  were  plotted  against  the  scores  of  the  complexity   metrics   (figure   10).     The   SD   was   calculated   from   the   three   film   measurements.   The   standard   deviation,   was   most   often   within   3   %   for   the   5   %   dose   difference   pass   rate   criteria   and   4   %   for   the   3   %   dose   difference   pass   rate   criteria.   A   higher   pearsons´s   r-­‐

value  was  found  for  all  combinations  of  metrics  and  pass  rate  criteria  compared  to  the   film  measurement  by  Götstedt  et  al  (2015)  (table  2).    

Figure 10. The pass rates for the dose difference evaluation of the measured and calculated dose distributions versus the complexity metric scores for AAA. A dose difference criterion of 5 % (left) and 3% (right) was used. The error bars correspond to 1 SD.

Anisotropic Analytical Algorithm

0,01,$$76,45$$$$$

0,10,$$65,45$$$$$

0,25,$$61,69$$$$$

0,40,$$51,74$$$$$

0,60,$$36,63$$$$$

0,02,$$74,86$$$$$

0,07,$$69,93$$$$$

0,13,$$61,77$$$$$

0,18,$$54,79$$$$$

0,31,$$38,46$$$$$

0,01,$79,28$0,02,$78,29$0,04,$77,30$

0,06,$74,48$0,10,$73,60$

0,01,$$74,99$$$$$

0,05,$$65,69$$$$$

0,11,$$56,65$$$$$

0,17,$$50,86$$$$$

0,27,$$45,74$$$$$

0,01,$$76,45$$$$$

0,10,$$65,45$$$$$

0,25,$$61,69$$$$$

0,40,$$51,74$$$$$

0,60,$$36,63$$$$$

0,02,$$74,86$$$$$

0,07,$$69,93$$$$$

0,13,$$61,77$$$$$

0,18,$$54,79$$$$$

0,31,$$38,46$$$$$

0,01,$79,28$0,02,$78,29$0,04,$77,30$

0,06,$74,48$0,10,$73,60$

0,01,$$74,99$$$$$

0,05,$$65,69$$$$$

0,11,$$56,65$$$$$

0,17,$$50,86$$$$$

0,27,$$45,74$$$$$

0,02,$$78,11$$$$$0,03,$$76,49$$$$$

0,07,$$70,99$$$$$

0,11,$$64,77$$$$$

0,19,$$53,58$$$$$

0,00,$$84,21$$$$$

0,01,$$81,37$$$$$

0,04,$$76,88$$$$$

0,11,$$67,40$$$$$

0,35,$$47,91$$$$$

0,02,$78,10666667$0,03,$76,49$

0,07,$70,99$

0,11,$64,77$

0,19,$53,58$

0,00,$$84,21$$$$$

0,01,$$81,37$$$$$

0,04,$$76,88$$$$$

0,11,$$67,40$$$$$

0,35,$$47,91$$$$$

Pencil'Beam'

Series$A$ Series$B$ Series$C$ Series$D$ Series$E$ Series$F$

r'='-0,89'

.

(24)

Table 2. Persons´s r-values from the evaluation of the pass rates for the dose difference evaluation of the measured and calculated dose distributions versus the complexity metric scores for AAA. The persons´s r-values in the first column are from one film measurement performed by Götstedt et al (2015) and the persons´s r-values in the second column are from all three film measurements.

  Götstedt et al (2015) This study

Metric r (5 %) r (3 %) r (5 %) r (3 %)

Conv.   -­‐0.76   -­‐0.78   -­‐0.80   -­‐0.85  

Edge.   -­‐0.79   -­‐0.83   -­‐0.84   -­‐0.90  

(25)

4.2 Different Dose Calculation Algorithms

Dose  difference  pass  rates  when  comparing  the  measured  and  calculated  absorbed  dose   distributions   was   studied   individually   for   PBC   (figure   11),   AXB   (figure   12)   and   CC   (figure   13).   Pearson’s   r-­‐values   were   calculated   for   all   the   individual   linear   fits   of   the   dose  difference  pass  rates  and  the  complexity  metrics  scores  for  the  30  MLC  openings   (table   3).   The   SD   was   calculated   from   the   pass   rate   values   from   the   three   film   measurements.   High   pearsons´s   r-­‐values   were   found   for   all   combinations   of   metrics,   pass  rate  criteria  and  algorithms.

0   20   40   60   80   100  

0,00   0,20   0,40   0,60   0,80   1,00  

5  %-­‐  dd  pass  rate  (%)  

Converted  aperture  metric  

r  =  -­‐0,89  

0   20   40   60   80   100  

0,00   0,20   0,40   0,60   0,80   1,00  

3  %-­‐  dd  pass  rate  (%)  

Converted  aperture  metric  

r  =  -­‐0,93  

0   20   40   60   80   100  

0,00   0,20   0,40   0,60   0,80   1,00  

3  %-­‐  dd  pass  rate  (%)  

Edge  area  metric    

r  =  -­‐0,95  

0   20   40   60   80   100  

0,00   0,20   0,40   0,60   0,80   1,00  

5  %-­‐  dd  pass  rate  (%)  

Edge  area  metric  

r  =  -­‐0,94  

Pencil Beam Convolution

Figure 11. The pass rates for the dose difference evaluation of the measured and calculated dose distributions versus the complexity metric scores for PBC. A dose difference criterion of 5 % (left) and 3 % (right) was used. The error bars correspond to 1 SD.

0,01,$$76,45$$$$$

0,10,$$65,45$$$$$

0,25,$$61,69$$$$$

0,40,$$51,74$$$$$

0,60,$$36,63$$$$$

0,02,$$74,86$$$$$

0,07,$$69,93$$$$$

0,13,$$61,77$$$$$

0,18,$$54,79$$$$$

0,31,$$38,46$$$$$

0,01,$79,28$0,02,$78,29$0,04,$77,30$0,06,$74,48$0,10,$73,60$

0,01,$$74,99$$$$$

0,05,$$65,69$$$$$

0,11,$$56,65$$$$$

0,17,$$50,86$$$$$

0,27,$$45,74$$$$$

0,01,$$76,45$$$$$

0,10,$$65,45$$$$$

0,25,$$61,69$$$$$

0,40,$$51,74$$$$$

0,60,$$36,63$$$$$

0,02,$$74,86$$$$$

0,07,$$69,93$$$$$

0,13,$$61,77$$$$$

0,18,$$54,79$$$$$

0,31,$$38,46$$$$$

0,01,$79,28$0,02,$78,29$0,04,$77,30$0,06,$74,48$0,10,$73,60$

0,01,$$74,99$$$$$

0,05,$$65,69$$$$$

0,11,$$56,65$$$$$

0,17,$$50,86$$$$$

0,27,$$45,74$$$$$

0,02,$$78,11$$$$$0,03,$$76,49$$$$$

0,07,$$70,99$$$$$

0,11,$$64,77$$$$$

0,19,$$53,58$$$$$

0,00,$$84,21$$$$$

0,01,$$81,37$$$$$

0,04,$$76,88$$$$$

0,11,$$67,40$$$$$

0,35,$$47,91$$$$$

0,02,$78,10666667$0,03,$76,49$

0,07,$70,99$

0,11,$64,77$

0,19,$53,58$

0,00,$$84,21$$$$$

0,01,$$81,37$$$$$

0,04,$$76,88$$$$$

0,11,$$67,40$$$$$

0,35,$$47,91$$$$$

Pencil'Beam'

Series$A$ Series$B$ Series$C$ Series$D$ Series$E$ Series$F$

r'='-0,89'

(26)

Acuros XB

0   20   40   60   80   100  

0,00   0,20   0,40   0,60   0,80   1,00  

5  %-­‐  dd  pass  rate  (%)  

Converted  aperture  metric  

r  =  -­‐0,52  

0   20   40   60   80   100  

0,00   0,20   0,40   0,60   0,80   1,00  

3  %-­‐  dd  pass  rate  (%)  

Converted  aperture  metric  

r  =  -­‐0,70  

0   20   40   60   80   100  

0,00   0,20   0,40   0,60   0,80   1,00  

5  %-­‐  dd  pass  rate  (%)  

Edge  area  metric  

r  =  -­‐0,71  

0   20   40   60   80   100  

0,00   0,20   0,40   0,60   0,80   1,00  

3  %-­‐  dd  pass  rate  (%)  

Edge  area  metric  

r  =  -­‐0,81  

Figure 12. The pass rates for the dose difference evaluation of the measured and calculated dose distributions versus the complexity metric scores for AXB. A dose difference criterion of 5 % (left) and 3% (right) was used. The error bars correspond to 1 SD.

!"!!!!!!!

!10,00!!!!!

!20,00!!!!!

!30,00!!!!!

!40,00!!!!!

!50,00!!!!!

!60,00!!!!!

!70,00!!!!!

!80,00!!!!!

!90,00!!!!!

0,00! 0,10! 0,20! 0,30! 0,40! 0,50! 0,60! 0,70! 0,80! 0,90! 1,00!

5"%"pass"rate"

Converted"Aperture"Metric"

Pencil"Beam"

Serie!A! Serie!B! Serie!C! Serie!D! Serie!E! Serie!F!

r"="90,89"

0,01,$$76,45$$$$$

0,10,$$65,45$$$$$

0,25,$$61,69$$$$$

0,40,$$51,74$$$$$

0,60,$$36,63$$$$$

0,02,$$74,86$$$$$

0,07,$$69,93$$$$$

0,13,$$61,77$$$$$

0,18,$$54,79$$$$$

0,31,$$38,46$$$$$

0,01,$79,28$0,02,$78,29$0,04,$77,30$

0,06,$74,48$0,10,$73,60$

0,01,$$74,99$$$$$

0,05,$$65,69$$$$$

0,11,$$56,65$$$$$

0,17,$$50,86$$$$$

0,27,$$45,74$$$$$

0,01,$$76,45$$$$$

0,10,$$65,45$$$$$

0,25,$$61,69$$$$$

0,40,$$51,74$$$$$

0,60,$$36,63$$$$$

0,02,$$74,86$$$$$

0,07,$$69,93$$$$$

0,13,$$61,77$$$$$

0,18,$$54,79$$$$$

0,31,$$38,46$$$$$

0,01,$79,28$0,02,$78,29$0,04,$77,30$

0,06,$74,48$0,10,$73,60$

0,01,$$74,99$$$$$

0,05,$$65,69$$$$$

0,11,$$56,65$$$$$

0,17,$$50,86$$$$$

0,27,$$45,74$$$$$

0,02,$$78,11$$$$$0,03,$$76,49$$$$$

0,07,$$70,99$$$$$

0,11,$$64,77$$$$$

0,19,$$53,58$$$$$

0,00,$$84,21$$$$$

0,01,$$81,37$$$$$

0,04,$$76,88$$$$$

0,11,$$67,40$$$$$

0,35,$$47,91$$$$$

0,02,$78,10666667$0,03,$76,49$

0,07,$70,99$

0,11,$64,77$

0,19,$53,58$

0,00,$$84,21$$$$$

0,01,$$81,37$$$$$

0,04,$$76,88$$$$$

0,11,$$67,40$$$$$

0,35,$$47,91$$$$$

0,00$ 0,10$ 0,20$ 0,30$ 0,40$ 0,50$ 0,60$ 0,70$ 0,80$ 0,90$ 1,00$

Pencil'Beam'

Series$A$ Series$B$ Series$C$ Series$D$ Series$E$ Series$F$

r'='-0,89'

References

Related documents

It is well known that obesity is an independent risk fac- tor for cardiovascular disease (CVD) and one of the main causes of the increased risk of diseases such as dyslipi-

Preliminary phytochemical screening of the acetone, chloroform, ethanol, petroleum ether and aqueous extracts of Cassia alata leaf were tested to identify the

A, dorsal aspect of male thallus with rows of antheridia; B, apical portion of female thallus with stalk arising from apical sinus and interrupted further up, at its tip

ния величин , измеренные на первой по ходу движения оси экипажа ( среднее по трем макси -. мумам с предварительным

R ESUME : Dans le but de décrire le profil épidémiologique des intoxications accidentelles isolées liées aux aliments, nous rapportons les résultats de

A stylistic survey of Malaysian mosque architecture shows four distinct trends: 1) Traditional style, which covers most of the mosques that were built before the

Die beschriebenen Assoziationen von SOX9 mit fortgeschrittener TNM-Kategorie und schlechter Prognose sowohl beim Magenkarzinom als auch bei anderen gastrointestinalen Tumoren [31,

e-ISSN: 2348-6848 p-ISSN: 2348-795X Volume 05 Issue 06..