I – Product Description/Function

The CNSolution 9310 On-line Cyanide Analyzer is designed to measure and control cyanide used in hydrometallurgical leaching of gold and silver from ores. Most gold occurs at very low concentrations in ores: less than 10 grams/ton (0.001%). Hydrometallurgical extraction is the only economically viable method of extracting gold from such ore. Leaching solutions typically contain cyanide in concentrations ranging from 50 ppm to 2,000 ppm. Figure 1 depicts the cyanidation process and points in the process where CN 9310 analyzers may be deployed. Controlling the cyanide concentration of the leach solution minimizes cyanide consumption and ensures a sufficient concentration is present for efficient extraction of gold. On-line cyanide monitoring enables mill operators to optimize the process and significantly reduce operating costs. A 10 percent decrease in cyanide-related costs (consumption and detoxification) can potentially save an average milling operation hundreds of thousands of dollars per year.

CNSolution

TM

9310

On-line Cyanide Analyzer

Product Release Synopsis

1 3 4 5 2 + – Ore Extraction Milling Leaching Recovery

Detoxification Tailings Pond

Electrowinning Smelting Carbon Cyanide Addition Cyanide Recycle Gold Eluate Dore Cyanide Addition Leaching Cyanide Recycle Detoxification

Effluent Discharge / Tailings The CNSolution 9310 supports measurement and control of cyanide in multiple cyanidation unit operations.

(2)

Hydrometallurgical gold milling operations have multiple potential applications for CNSolution 9310 cyanide analyzers. Instruments may be deployed in unit operations and locations other than the leaching circuit. A brief description of these applications is provided below.

Laboratory Analysis

Some gold mills will have laboratories on-site or located nearby to help support their operations. The chemical composition of ores is key factor in determining the processing steps and conditions necessary to extract precious metals efficiently and economically. Metallurgists conduct bench top bottle roll tests on milled ore samples to determine the optimum cyanide concentration for recovery of gold. A CNSolution 9310 can be used as a bench top analyzer to perform these tests. Grab samples are also taken on a non-routine basis from the leaching tanks and tested in the laboratory as a process control check.

Detoxification

The International Cyanide Management Code defines good practices and guidelines governing the use, treatment, and disposal of cyanide by hydrometallurgical milling operations. The code set a limit of 50mg/L in tailings solution discharges. To comply with this limit spent cyanide solutions from the leaching process must be detoxified by one of several processes. Detoxification involves the use of strong oxidizing agents to destroy cyanide and reduce its concentration below 50 ppm. The efficiency of the detoxification process can be monitored and controlled by installing CNSolution 9310 analyzers to measure the cyanide concentrations in influent and effluent solutions. Optimization of the detoxification process can significantly reduce the use and associated expense of oxidizing agents in this unit operation.

Tailings

Following detoxification to a concentration below 50ppm tailings solutions are discharged into a tailings impoundment pond. Exposure to natural ultraviolet light further degrades cyanide levels in tailings ponds. The CNSolution 9310 analyzer is used to measure the cyanide concentration in tailings pond solutions and in cyanide recycle circuits to verify that concentrations are suitable for their intended use.

For instance, cyanide recycle solutions that will be used for flotation of copper sulfides (chalcopyrite) must be below 0.5 ppm.

II – Principle of Operation

The CNSolution 9310 (Figure 2) measures free cyanide in precious metal leaching and tailings solutions by the gas-diffusion amperometry technique in U.S. EPA Method OIA-1677 and ASTM D 6888-09. The principle of operation is depicted in Figure 3 and described below. In operation, a filtered sample is drawn into the CNSolution 9310 to fill a fixed volume loop. A base reagent is continuously pumped through one side of a gas diffusion membrane and out through the flow-cell of an amperometric detector. Sample in the loop is injected into an acidic carrier streat. The acidic conditions convert the cyanide ion into hydrogen cyanide which diffuses across the hydrophobic membrane into the base reagent where it converts back to the cyanide ion and enters the flow-cell of the amperometric detector. Cyanide ions react with a silver electrode and generate a current proportional to the cyanide concentration. The detector response for each injection is displayed in real-time as a peak on the touch-screen display and can be output to a Supervisory Control and Data Acquisition (SCADA) system.

(3)

CNSolution™ 9310 On-line Cyanide Analyzer – Product Release Synopsis

III – Product Specifications

Design and performance characteristics of the CNSolution 9310 On-line Cyanide Analyzer are summarized in Table 1. Three measurement ranges are cited reflecting deployment of the CNSolution 9310 in different locations and applications in the cyanidation process.

Operating Principle FIA by gas diffusion amperometry

Measurement Technique Amperometric detection – Silver electrode

Measurent Ranges 0.2 to 50 / 2.0 to 500 / 20 to 2000 ppm CN

Reference Methods USEPA OIA-1677 / ASTM D 6888-09 (Available CN)

Calibration 2 point calibration

Measurement Accuracy ± 5% at 50-ppm

Sample Introduction Continous on-line fill-and-spill sampling system

Sampling Interval User programmable

Analysis Time < 3 minutes

Operating Environment 5 - 45 ˚C, up to 90% humidity (non-condensing)

Operator Interface Windows® CE-based, color touch-screen display

Reagents Required Water, NaOH, HCl, CN-calibration standards

Power Requirements 24 VDC

Output Relays 2 (system alarm, sample alarm)

Analog Output 2 4-20mA (user-configurable concentration)

Data Export To PC via Ethernet, or using USB memory stick

Instrument Enclosure NEMA 4X / IEC Class IP-56

External Dimensions 48.3 cm H x 31.1 cm W x 31.1 cm D (19” H x 12.25” W x 12.25” D) Weight 11 kg (24 lbs) Certifications CE Base Acid Carrier Acid Waste Detector CN Waste Gas Diffusion Membrane Valve Mixing Coil Micro-Pump Micro-Pump Micro-Pump CNSolution 9310 On-line Cyanide Analyzer

1 2

3 Available Cyanide Analysis by Gas-diffusion Amperometry

(USEPA Method OIA-1677 / ASTM D 6888-09)

Membrane HCN HCN HCN CN- CN- CN- Gas Diffusion Acidification CN- + H+ HCN (g) 1 2 Amperometric Detection 2 CN- + Ag Ag(CN)- + e- at + 0.0 V2 3

Figure 3. CNSolutionTM 9310 Operating Principle

Figure 2. CNSolutionTM 9310

On-line Cyanide Analyzer

(4)

IV – Technical / Competitive Advantages

The CNSolution 9310 was developed to overcome problems and limitations associated with other cyanide measurement techniques (i.e.; titration, electrodes, etc.). Some of the key technical and competitive advantages of the CNSolution 9310 analyzer are noted below.

Superior Accuracy

The gas-diffusion amperometry technique employed in the CNSolution 9310 has been demonstrated to be free of interferences from copper and metallic sulfides in precious metal leaching solutions. Copper complexes with cyanide reducing the cyanide available for leaching. Titration methods commonly used for process control do a poor job of estimating the amount of cyanide available for leaching when copper is present. The CNSolution 9310 responds quantitatively to CN-, plus zinc, copper, cadmium, and

silver cyanide complexes over the entire instrument calibration / measurement range (0.2 to 2,000ppm). Measurement accuracy is ± 5% at 50ppm.

Verifiable Results

The automatic calibration function of the CNSolution 9310 and programmed analysis of independent check standards help verify the analyzer is reporting accurate results. Results of on-line measurements obtained with the CNSolution 9310 can also be verified by testing grab samples on a laboratory instrument running the same methodology (OIA-1677 / ASTM D 6888) without applying a correction factor to account for the variance between different analytical techniques / methods.

Multiple Measurement Ranges

The CNSolution 9310 supports three measurement ranges; 0.2 to 50, 2.0 to 500, and 20 to 2,000ppm enabling the analyzer to be deployed in multiple locations and applications in the cyanidation process. These measurement ranges enable continuous monitoring and control of cyanide concentrations in leaching and detoxification operations.

Fast Analysis Time

The gas-diffusion amperometry flow injection analysis technique provides results within two minutes of sample injection. Cyanide concentration and peak traces are displayed in real-time on the CNSolution 9310 color touch-screen display. Fast analysis time and continuous on-line monitoring supports tighter control of cyanide concentration and associated operating costs.

V – Market Opportunity

The target market for the CNSolution 9310 On-line Cyanide Analyzer are facilities that use

hydrometallurgical cyanidation to extract gold and silver from ores. There are approximately 875 licensed commercial gold and silver operations worldwide. 460 of these operations use cyanide for gold and silver extraction.

Countries in North and South America reporting gold production levels in excess of tens of thousands of kilograms include; Argentina, Bolivia, Brazil, Canada, Chile, Colombia, Mexico, Peru, and the United States. Countries in the Asia-Pacific region reporting gold production levels in excess of thousands of kilograms include; Australia, China, Indonesia, Japan, New Zealand, New Guinea, Philippines, and Vietnam.

(5)

CNSolution™ 9310 On-line Cyanide Analyzer – Product Release Synopsis

P.O. Box 9010

College Station, Texas 77842-9010

Tel: (979) 690-1711 • Fax: (979) 690-0440 • www.oico.com

VI – Product Launch Plan

The official launch of the CNSolution 9310 On-line Cyanide Analyzer will be at two conferences and tradeshows in the first quarter of 2013. The Society of Mining and Metallurgical Engineers Conference and Exhibition (SME 2013) will be held in Denver, February 24 – 27, 2013. The Pittsburgh Conference on Analytical Chemistry (PittCon 2013) will be held March 18 – 21, 2013.

The CNSolution 9310 will be on display at these exhibitions and technical presentations will be delivered in the technical sessions of both conferences.

The first print advertisements for the CNSolution 9310 will appear in the February issue of Mining Engineering magazine and the official program of SME 2013. Additional ads will be placed in Mining Engineering and other appropriate U.S. and International publications going forward.

Product Literature & Sales Tools

An assortment of product literature and sales tools have been prepared to support the sales process. A CNSolution 9310 product brochure with analyzer specifications (Publication # 3765) is available along with an application note (Publication # 3789), Lock-Out Bid Specifications (Publication # 3770), and a FABS – Features, Advantages and Benefits document (Publication # 3771).

Figure

Figure 1. Precious Metal Cyanidation Process
Table 1. CNSolution TM  9310 Specifications

References

Related documents

Thus this study concluded that preeclampsia patients had increased levels of serum lactate dehydrogenase levels when compared with antenatal patients with normal blood

As yet, effects of restricted placental growth and/or function on spontaneous ambulatory 105.. activity (SAA) have not been reported; nor has the effect of IUGR or

At the processing plant, with the technique of cyanide leaching, gold is extracted, in this method ores are crushed and soaked in the solution of cyanide

Overexpression of NEAT1 promoted proliferation and invasion while inhibited apoptosis in HCC cells, which was opposite to the effect of NEAT1 knockdown.. Also, AKT2 was increased in

Licensed under Creative Common Page 364 The level of clinical patient satisfaction at the ASRI Foundation was obtained by distributing. questionnaires to 100 patient

 Cyanide compounds: calcium, mercuric, ferric, potassium, sodium, silver, zinc o Keep away from acids as they generate highly toxic hydrogen cyanide gas.  Sulfide

Labor demand for the fixed cost activity rises in proportion to labor demand for the variable cost activity in this case, so the fixed/variable distinction has no bearing on

employment, and trade—has led to an emphasis on these issues as alternative measures of “cost.” Jorgenson and Wilcoxen (1990), for example, focus on the growth consequences

 Comparative  proteomics  of  glycosomes  from  bloodstream  form   and  procyclic  culture  form  Trypanosoma  brucei  brucei..  Purification  and

 Question: Will scheduling a follow-up appointment within seven days for heart failure patients decrease 30-day readmission rates..  Population: All heart

Keywords : Interorganizational Networks; Global Production Networks; Supply Chains; Supply Chain Risks Management; Employer-Employee Issues; Labor Relations; Risk;

En este sentido, recordamos cómo El Ce- rro del Castillo se encontraría en una zona po- tencialmente importante para la explotación so- bre todo de hierro y de otros metales en menor

Through the ranking analysis, it was found that the five greatest barriers to the adoption of EEIMs were: (1) the lack of subsidies and rebates on energy efficient equipment,

Using several different models to estimate the difference between test score levels of students attending charter schools versus those from traditional public schools in the

Although source-based taxation remains theoretically valid in today’s globalized world, the way source is defined under the permanent establishment threshold needs to be

Our in vivo and in vitro study demonstrated that SCAGP markedly inhibited PM and cell viability of MGC-803 cells, induced cell apoptosis and cell cycle arrest in G0/G1 phase, as

Figure India 2015 Calcium Cyanide Sales Market Share Table Global 2011-2016E Calcium Cyanide Sales by Type Table Different Types Calcium Cyanide Product Interview Price Table

The key research question is ‘to what extent is it possible to change the direction of design and technology education in the classroom through highlighting creativity?’ This

Coronal 4-D flow velocity map gives a comprehensive overview of the vascular anomalies and shows multiple sites where flow is quantified: scimitar vein (asterisk), systemic

Graph 1: Effect of different exposure time of sodium hypochlorite treatment at surface sterilization of Aloe vera.. Ethanol is a powerful sterilizing agent but also highly

It is not compulsory to have an SCR, but if you choose to opt out of the scheme, then you will need to complete a form (available on our website or at reception) and bring it back