• No results found

Hybrid Analytical Platform Based on Field-

N/A
N/A
Protected

Academic year: 2020

Share "Hybrid Analytical Platform Based on Field-"

Copied!
13
0
0

Loading.... (view fulltext now)

Full text

(1)

Article

1

Hybrid Analytical Platform Based on

Field-2

Asymmetric Ion Mobility Spectrometry, Infrared

3

Sensing, and Luminescence-Based Oxygen Sensing

4

for Exhaled Breath Analysis

5

L. Tamina Hagemann, Stefan Repp, Boris Mizaikoff*

6

Institute of Analytical and Bioanalytical Chemistry (IABC), Ulm University, Albert-Einstein-Allee 11, 89081

7

Ulm, Germany; [email protected].

8

* Correspondence: [email protected]; Tel.: +49-731-50-22750

9

Received: date; Accepted: date; Published: date

10

Abstract: The reliable online analysis of volatile compounds in exhaled breath remains a challenge

11

as a plethora of molecules occur in different concentration ranges (i.e. ppt to %), and need to be

12

detected against an extremely complex background matrix. While this complexity is commonly

13

addressed by hyphenating a specific analytical technique with appropriate preconcentration and/or

14

preseparation strategies prior to detection, we herein propose the combination of three analytical

15

tools based on truly orthogonal measurement principles as an alternative solution: field-asymmetric

16

ion mobility spectrometry (FAIMS), Fourier-transform infrared (FTIR) spectroscopy-based sensors

17

utilizing substrate-integrated hollow waveguides (iHWG), and luminescence sensing (LS). These

18

three tools have been integrated into a single compact analytical platform suitable for online exhaled

19

breath analysis. The analytical performance of this prototype system was tested via artificial breath

20

samples containing nitrogen (N2), oxygen (O2), carbon dioxide (CO2) and acetone as a model volatile

21

organic compound (VOC) commonly present and detected in breath. Functionality of the combined

22

system was demonstrated by detecting these analytes in their respectively breath-relevant

23

concentration range and mutually independent of each other generating orthogonal yet correlated

24

analytical signals. Finally, adaptation of the system towards the analysis of real breath samples

25

during future studies is discussed.

26

Keywords: exhaled breath analysis; field-asymmetric ion mobility spectrometry; FAIMS;

Fourier-27

transform infrared spectroscopy; FTIR; luminescence sensing; infrared sensors; hyphenated

28

techniques; hybrid techniques; acetone; carbon dioxide; oxygen

29

30

1. Introduction

31

Breath contains a wide variety of molecules in largely different concentration ranges - from ppt

32

to percent - that are potentially useful for therapy monitoring and elucidation of metabolic pathways.

33

The analysis of such a complex sample by a single analytical techniques is almost impossible. Hence,

34

the combination of orthogonal analytical tools appears to be a viable strategy addressing this issue.

35

To date, predominantly preconcentration, e.g. via solid-phase microextraction (SPME) fibers and

36

needle trap devices (NTD), and/or preseparation schemes, e.g. gas chromatography (GC) or

37

multicapillary (MCC) columns are implemented for addressing trace concentrations, and for

38

reducing the sample complexity. By combining preseparation schemes with FID[1], mass

39

spectrometers, (e.g. TOF-MS[2]) or ion mobility based detectors, e.g. IMS[3] or DMS[4] potent

40

analytical tools have resulted. However, MS-based equipment - while being able to detect a very wide

41

variety of analytes - tends to be costly, bulky and frequently not suitable for online analysis. Also, if

42

only one type of detector is used, potentially useful analytes (i.e. biomarkers) that are not sensitive to

43

the selected detector type remain undetected.

44

(2)

Therefore, the integration of orthogonal detection schemes into a single hybrid analytical

45

platform is the next logical step. Only a few research groups have selected this path for exhaled breath

46

analysis. The probably most commonly selected approach is the use of electronic noses[5–10], i.e.

47

arrays of different colorimetric[8] or metal oxide sensors[9] individually responding to different types

48

of molecules. While these sensors arrays offer portable and rapidly responding breath detection

49

capabilities, specific biomarker identification and inter-device comparability remain challenging[11].

50

Vaks et al.[12] and Shorter et al.[13] both combined light sources emitting different wavelengths or

51

even wavelength regimes (i.e. subTHz, THz, IR) in order to broaden the scope of addressable analytes

52

in breath. However, even if these light sources complemented each other, hence providing

53

orthogonality to some extent the basic detection mechanism was essentially similar. Hence, molecules

54

not responding to the respective detection scheme (here, sufficient light absorption in the selected

55

wavelength regimes) will remain undetected. Consequently, truly orthogonal methods are based on

56

different physical principles generating the analytical signals, yet applied to the same sample. This

57

approach has already been proposed[14,15] and put into practice[10,16–20] by various research

58

groups. For example, Covington et al.[10] applied an eNose and GC-IMS to the same breath samples,

59

whereas Williams et al.[19] parallely used non-dispersive infrared analysis and PTR-TOF-MS on the

60

same sample set. It is important to notice though that all above-mentioned groups except Monks et

61

al.[20] applied different analytical methods as standalone-techniques, i.e. the used analytical devices

62

were not integrated into a single setup. This entails extensive sample handling – and potentially

63

associated handling errors – and extended analysis times, e.g. required for separate sample injection

64

and limits application at the patient bedside. Furthermore, with the exception of Williams et al.[19]

65

offline breath analysis was performed frequently involving gas bags or sample storage, and thus

66

taking the risk of cross-contamination and sample degradation.

67

Only few groups have developed hybrid analytical devices that enable online breath analysis

68

based on truly orthogonal principles integrated in a single sensing platform. Tiele et al.[21] published

69

a portable device for CO2 and O2 detection that additionally measured temperature and pressure.

70

Miekisch et al.[22,23] presented a multidimensional sensing platform including hemodynamic

71

monitoring as well as comprehensive breath monitoring via capnometry, spirometry and

PTR-TOF-72

MS, all being integrated into the same online monitoring platform. While Miekisch et al. analyzed

73

human breath, our research team has focused on exhaled mouse breath analysis within a mouse

74

intensive care unit (MICU) at the Institute of Anesthesiologic Pathophysiology and Method

75

Development (IAPMD) at Ulm University Medical Center, which requires as an additional challenge

76

the analysis of exceedingly (i.e. few hundreds of microliters) small breath sample volumes[24–26]. In

77

order to gain metabolic insights, 12CO2, 13CO2 and O2 concentrations as well as the respiratory quotient

78

(RQ) were evaluated using various analytical tools (iHWG-FTIR spectroscopy, interband cascade

79

laser based tunable diode laser absorption spectroscopy (TDLAS) and LS), which were all adapted to

80

the challengingly small breath volumes exhaled by a mouse or any comparable small animal model.

81

Besides these already quantifiable analytes in mouse breath, the detection of additional volatile

82

compounds such as acetone and H2S is currently in development for therapy monitoring and to aid

83

in understanding the underlying metabolism of traumatized mice.

84

Hence, the present study aims at extending the scope of addressable analytes in mouse breath

85

beyond CO2 and O2 by combining FTIR and LS with FAIMS serving as truly orthogonal analytical

86

methods. The detection principles of iHWG based FTIR spectroscopy[27], LS[28] and FAIMS[29] have

87

been described in detail elsewhere. O2 detection via LS was necessary, as O2 is neither IR active nor

88

does it give rise to a FAIMS signal. Furthermore, CO2 could not have been detected by the

89

luminescence sensor and is not ionizable by the 63Ni FAIMS ionization source, and hence, not

90

detectable by FAIMS. In turn, it provides a signal via IR spectroscopy/sensing techniques. Last but

91

not least, the luminescence sensor does not respond to VOCs, and the sensitivity of the selected IR

92

approach would not have allowed for VOC detection at the breath-relevant ppt or ppb concentration

93

range, even though a wide variety of breath-relevant VOCs are IR-active. Hence, integrating FAIMS

94

(3)

Synthetic breath samples containing N2, O2, CO2 and acetone as an exemplary breath VOC were

96

prepared and analyzed to demonstrate functionality of the developed hybrid prototype. The

97

presented data proves the feasibility of the integration of FAIMS, FTIR and LS into a single analytical

98

platform for simultaneous online analysis of O2, CO2 and acetone as a breath VOC representative. It

99

was shown that the detection of all analytes was possible in the respective breath-relevant

100

concentration range, and that FAIMS, FTIR and LS signals were independent of one another, yet

101

correlated as determined at the same time within the same sample.

102

103

2. Materials and Methods

104

2.1 Hybrid Analytical Platform

105

2.1.1 Gas Sample Preparation

106

A stock gas mixture of 2.33 ppm acetone in synthetic air (± 0.23 ppm, MTI Industriegase,

Neu-107

Ulm, Germany) was diluted down by synthetic air (produced with 20.5 vol.% O2 grade 5.0, remains

108

N2 grade 5.0, H2O ≤ 5 ppmv, NO+NO2 ≤ 0.1 ppmv, low molecular weight hydrocarbons CnHm <

109

0.1 ppmv, by MTI Industriegase, Neu-Ulm, Germany) and CO2 (technical grade (DIN EN ISO 14175),

110

≥ 99.8 vol-%, N2 ≤ 1000 ppmv, H2O ≤ 120 ppmv, MTI Industriegase, Neu-Ulm, Germany) to eight

111

samples, containing acetone concentrations between 0 and 20 ppb and a background concentration

112

of 3, 4 or 5 % CO2 and 19.6 ± 0.5 % O2 (concentrations given here are volumetric concentrations). The

113

acetone, air and CO2 flow were regulated by mass flow controllers (Bronkhorst El Flow Prestige,

FG-114

201CV-RBD-11-K-DA-000, 80 mL/min full scale capactiy for acetone; FG-201CV-ABD-11-V-DA-000,

115

3000 mL/min full scale capacity for synthetic air; Vögtlin red-y smart series, type GSC-A9KS-BB22,

116

200 mL/min full scale capacity for CO2). For cleaning and drying purposes, air and CO2 were filtered

117

through active charcoal (# 20626, Restek, Bad Homburg, Germany), molecular sieve (5Å pore size,

118

# 8475.2, Carl Roth GmbH & Co KG, Karlsruhe, Germany) and sintered glass filter elements

119

(VitraPor®, 40-100 µm, 4-5.5 µm, 1.5 µm). The dew point of air and CO2 was measured to be -39.8°C

120

(humidity sensor SF52-2-X-T1-B, Michell Instruments, Ely, UK), corresponding to a water content of

121

192 ppm. The acetone sample gas was neither VOC filtered nor dried, since this would have caused

122

analyte loss. The water content in the acetone gas cylinder was assumed to be negligible due to the

123

dilution of acetone sample gas in comparatively big volumes of CO2/air.

124

Acetone and CO2 were mixed first, by leading their flow through a filter with 0.5 μm pore size

125

(SS-2TF-05, Swagelok, Reutlingen, Germany) to induce turbulences for homogeneous mixing. The

126

combined acetone/CO2 flow was then combined with the air flow. A schematic of the gas mixing unit

127

is displayed in Figure 1 (left half) in section 2.1.2 together with the hybrid FAIMS-FTIR-LS sensing

128

platform.

129

2.1.2 Hybrid FAIMS-FTIR-LS Platform and Concentration-Dependent Measurements

130

The hybrid analytical platform is displayed in Figure 1. Gas samples were provided by the gas

131

mixing unit displayed in the left half of Figure 1 and described in the previous section. The sample

132

flow produced by the gas mixing unit was constantly kept at 2200 mL/min. The relief valve

(SS-133

RL3S4, Swagelok, Reutlingen, Germany) between the gas mixing unit and the FTIR/O2 sensor unit

134

was adjusted so that the flow reaching the FTIR/O2 sensor unit was 400 ± 10 mL/min and the flow

135

through the FAIMS PAD was 1800 ± 30 mL/min. These flows were regularly checked on with a digital

136

flow meter (ADM1000, J&W Scientific, Folsom, CA, USA) at the outlet of the O2 sensor and with the

137

flow sensor integrated in the FAIMS PAD, respectively. To minimize analyte adsorption along the

138

tubing walls, perfluoroalkoxy alkane (PFA) tubings (1/8’’ and 1/4’’ outer diameter, Swagelok,

139

Reutlingen, Germany) and heated (41 °C) Sulfinert tubings (#29242, Restek, Bad Homburg, Germany)

140

(4)

142

Figure 1. Experimental setup comprising the gas mixing unit and the hybrid analytical platform.

143

Numbers in blue are gas flows in mL/min.

144

Before starting a measurement series, a hold time was adopted until flow and pressure had

145

stabilized in the FAIMS device (1800 ± 30 mL/min, 0.800 ± 0.020 barg) to ensure reproducibility of the

146

FAIMS data. In case the flow and pressure varied beyond the given limits, the needle valve at the

147

exhaust of the FAIMS as well as the relief valve between FAIMS and FTIR were adjusted until flow

148

and pressure had stabilized for at least ten minutes in the range defined above.

149

Each measurement series included eight acetone/CO2/air gas samples. Prior to the analysis of an

150

acetone/CO2/air mixture, one sample containing pure air and one sample containing only air and CO2

151

were recorded (see Table 1). During the pure air sample, the FTIR background was recorded, and the

152

according FAIMS spectrum was used to ensure that the system had entirely cleaned down after the

153

previous sample. The CO2/air measurement, on the other hand, served as a background spectrum for

154

FAIMS. Before analysis, the respective sample gas was led through the setup for at least two minutes

155

to ensure a constant analyte concentration in the whole setup and during the entire measurement.

156

Table 1: Overview on the measurement protocol within the hybrid setup.

order injected sample FAIMS iHWG-FTIR LS

1 pure air verifying system

cleanliness

background recording

-

2 CO2/air background

recording

- -

3 acetone/CO2/air acetone signal recording

CO2 signal recording

O2 signal recording 4 repeated for all further samples of a measurement series in random order

157

After recording a blank as the first sample of every measurement series, the remaining samples

158

were analyzed in a random sample order that was different in each measurement series. The CO2

159

concentration was constant (3, 4 or 5 %) within one measurement series. Three measurement series

160

were recorded per CO2 concentration. For all air, CO2/air and all acetone/CO2/air samples, five FAIMS

161

spectra (~19 min) and five FTIR spectra (~ 3.5 min) were successively recorded, while the sample was

162

continuously flowing through the hybrid setup. Simultaneously, the O2 concentration was

163

continuously monitored for the duration of the FAIMS measurements.

164

(5)

2.2 Details on the Individual Analytical Methods

166

2.2.1 Field-Asymmetric Ion Mobility Spectrometry

167

FAIMS data were recorded with an OEM FAIMS PAD (Owlstone Inc., Cambridge, UK), using

168

the Lonestar software (version 4.912, Owlstone Inc., Cambridge, UK). After ionization by a 63Ni

169

ionization source, analytes were detected by the FAIMS sensor (gap size 37 µm; RF waveform:

170

267 ± 2 V maximum peak-to-peak voltage, 26 MHz ± 26 Hz RF, 25 % Duty Cycle, 51 steps;

171

compensation voltage (CV) from -6 to +6 V (512 steps, ~4.5 s per full CV scan), flow 1800 ± 30 mL/min;

172

sensor temperature: 60 °C). The sample gas was continuously flowing through the spectrometer at

173

1800 ± 30 mL/min as the data were recorded. The pressure could be regulated via the needle valve

174

(SS-2MG-MH, Swagelok, Reutlingen, Germany) at the FAIMS outlet and was set to 0.800 ± 0.020 barg.

175

A membrane filter at the inlet of the FAIMS device (polytetrafluoroethylene (PTFE) membrane, 1 µm

176

pore size), heated to 100 °C to avoid analyte accumulation in the filter, prevented particle

177

introduction into the FAIMS PAD. In order to avoid charge build up, the intersweep delay between

178

two subsequent recordings was set to 1500 ms. The obtained FAIMS spectra, also called dispersion

179

plots, displayed the ion current on the detector (z axis) in dependence on the CV (x axis) and the

180

percentage of the dispersion field (DF) which was scanned by varying the peak-to-peak-voltage

181

between 0 and 267 V stepwise.

182

2.2.2 Substrate-Integrated Hollow Waveguide Coupled Fourier-Transform Infrared Spectroscopy

183

CO2 concentrations were monitored via iHWG coupled FTIR spectroscopy. The setup and gas

184

cell have been described in detail elsewhere[30]. Light from an ALPHA FTIR spectrometer (Bruker

185

Optik GmbH, Ettlingen, Germany) was coupled into an iHWG (aluminum, 7.5 cm optical path

186

length, 4x4 mm internal cross-section, produced by fine mechanical workshop West, Ulm University,

187

Ulm, Germany) and then onto the internal detector of the spectrometer via two gold-coated off-axis

188

parabolic mirrors (Thorlabs, MPD254254-90-M01, 2″ RFL). Using the software OPUS (version 7.2,

189

Bruker Optik GmbH, Ettlingen, Germany), IR spectra were recorded in the spectral range from 4000

190

to 400 cm-1 at a spectral resolution of 2 cm-1, with 20 averaged scans, and at a flow rate of

191

400 ± 10 mL/min. The Fourier transformation was done in OPUS, using the Blackman-Harris 3-term

192

apodization function. In order to exclude CO2 from ambient air from the optical absorption paths, the

193

entire IR setup was housed in a plastic bag which was purged with synthetic air for at least 15 minutes

194

prior to as well as during each measurement series.

195

2.2.3 Oxygen Sensing

196

A flow-through O2 sensor, detecting O2 based on luminescence quenching, (FireStingO2, Pyro

197

Science GmbH, Aachen, Germany)[28] was used for monitoring the O2 concentration, supported by

198

the Software FireSting Logger (version 2.365, PyroScience GmbH, Aachen, Germany). One data point

199

per second was recorded.

200

201

2.3 Data Processing

202

2.3.1 Field-Asymmetric Ion Mobility Spectrometry

203

Since a direct import of the FAIMS data (.dfm format) into Matlab was not possible, FAIMS data

204

were exported from the Lonestar software as text files and then imported into Matlab (R2018A, The

205

Mathworks Inc., Natick, MA, USA). For baseline correction, the average of the five repetitions of the

206

FAIMS dispersion plot of an CO2/air sample was substracted from the average of the five repetitions

207

of the subsequent acetone CO2/air sample. Acetone monomer and dimer peak volumes were

208

approximated by respectively summing all intensity values in selected regions of the dispersion plot

209

(monomer: 68 to 72 % DF, -2.75 to -1.95 V CV; dimer: 46 to 50 % DF, -0.35 to +0.45 V CV). These

210

(6)

compromise between achievable signal height and freedom from interferences with other spectral

212

components. The so obtained monomer and dimer peak volumes were then added together to obtain

213

the total acetone signal (from now on, only called “acetone signal”). Singly integrating the monomer

214

or the dimer peak would have distorted the FAIMS data evaluation: while the monomer peak was

215

very faint or even invisible at higher acetone concentrations, its contribution to the total acetone signal

216

at higher concentrations would not have been negligible.

217

After normalization with the mean acetone signal at the maximum measured acetone

218

concentration (20 ppb), the signal was averaged and the standard deviation was calculated. The

219

normalized and averaged acetone signal was plotted against the acetone concentration and an

220

asymptotic fit (y=A-B·Cx) was applied. Following IUPAC regulations[31], the concentration at the

221

limit of detection (LOD) and at the limit of quantification (LOQ) was estimated by inserting the signal

222

intensity at the LOD and the LOQ (µB + 3.29·σB and µB + 10·σB, respectively, with average normalized

223

signal intensity of the blank µB and its according standard deviation σB) into the inverse of the

224

calibration function (x=ln((A-Y)/B)/lnC).

225

2.3.2 Fourier-Transform Infrared Spectroscopy

226

IR data were imported from OPUS into Origin Pro 2017G. An exemplary spectrum is shown in

227

Figure A1 in the Appendix. For baseline correction, each IR spectrum was shifted by the median of

228

the data set, since the latter suitably represented the baseline. The area under the baseline-corrected

229

IR peak at 2360 cm-1 between 2200 and 2450 cm-1 was averaged for the five repetitions recorded in a

230

row for each sample. The so obtained CO2 signal was then normalized by division by the overall

231

maximum CO2 signal and the normalized signal was averaged for the three repetitions of the

232

measurement series recorded for each CO2 concentration (3, 4 and 5 % CO2). The according standard

233

deviation was calculated.

234

2.3.3 Oxygen Sensing

235

For each measurement, the O2 concentrations directly output by the FireSting Logger software

236

was averaged for the time span between 5 and 15 min after starting the O2 measurement. O2

237

concentrations recorded between 0 and 5 min were not included in the average, because the O2

238

concentration reached an equilibrium after approximately 5 min (see Figure A2 in the Appendix).

239

The so obtained O2 signal was then normalized by division by the overall mean O2 signal; the

240

normalized signal was averaged for the three repetitions of the measurements series recorded for

241

each CO2 concentration and the standard deviation was calculated.

242

243

3. Results and Discussion

244

3.1 FAIMS Results

245

As mentioned above, FAIMS dispersion plots of pure air and of CO2/air were recorded before

246

recording an acetone/CO2/air containing sample (for further detail also see Table 1 in Section 2.1.2).

247

Figure 2 exemplarily shows a dispersion plot for each sample type collected in positive mode. The

248

dispersion plot of pure air (Figure 2a) mainly showed the reactant ion peak (RIP), which, in positive

249

detection mode, appears due to the formation of ionized clusters of water molecules present in the

250

carrier gas[32]. The faint vertical signal in Figure 2a at around 0 V CV was approximately constant

251

for all recorded dispersion plots. It could not be erased throughout the whole project and was likely

252

to be caused by substances emitted from the tubings and the FAIMS device itself. The CO2/air

253

dispersion plot (Figure 2b) also mainly showed the RIP. No clear analyte peak appeared, since CO2

254

is not ionizable by the 63Ni source. The faint additional trace at around -65 % DF and -3 V CV

255

assumably occurred because of contaminations from the CO2 gas bottle that could not be entirely

256

removed by the used filters. The acetone/CO2/air dispersion plot (Figure 2c) showed an intensity

257

decrease of the RIP as well as the appearance of two main additional peaks. Generally, once an

258

(7)

carrier gas clusters are replaced by the analyte molecules. Hence, the RIP intensity decreases and a

260

monomer and/or dimer peak appear, respectively. The tentative assignment of monomer and dimer

261

peak, as it is indicated in Figure 2c, was based on the concentration-dependent behavior of both

262

peaks: while the monomer peak intensity showed an intensity maximum at lower concentrations, the

263

dimer peak constantly increased with increasing acetone concentration, as an additional water

264

molecule in each monomer cluster was replaced by a second acetone molecule, thus forming a dimer

265

cluster. The relative position of monomer and dimer peak also was in accordance with our

266

expectations and thus substantiated our peak assignment: the lighter, less bulky and hence more

267

mobile monomer cluster gave rise to a peak at a lower CV than the less mobile dimer cluster. The

268

exact origin of the faint feature between monomer and dimer peak in Figures 2c and 2d

269

(~50 %DF, -0.5 V CV) is unknown, but its potential effect is commented on in section 3.2. In order to

270

obtain the net monomer and dimer signal, Figure 2b was subtracted from Figure 2c for background

271

substraction. The resulting data is shown in Figure 2d. The z axis of Figure 2d was varied compared

272

to Figures 2a to 2c in order to make the monomer and dimer peak more clearly visible. At the position

273

where the RIP appeared in Figure 2a to 2c, the signal intensity was negative in Figure 2d, since the

274

RIP intensity decreased while acetone was present in the FAIMS sensing region.

275

276

Figure 2. FAIMS dispersion plots. (a) pure air sample (b) CO2/air sample (c) acetone/CO2/air sample

(d) background substracted acetone/CO2/air sample ((c) minus (b)). CO2 and acetone concentration of

(8)

3.2 Co-Dependencies of Acetone, CO2 and O2 Signal

277

The normalized total acetone signal, composed of monomer and dimer peak volume, was

278

plotted against the acetone concentration, shown for 3, 4 and 5 % CO2 in Figure 3a. Due to saturation

279

of the FAIMS detector, the acetone signal converged towards a maximum value for higher acetone

280

concentrations. Thus, an asymptotic fit was applied. It is obvious from Figure 3a, that the acetone

281

signal was statistically identical, regardless if the CO2 concentration was 3, 4 or 5 %. Likewise, the

282

according analytical figures of merit, i.e. LOD, LOQ, R2 and parameters of the asymptotic fit, did not

283

depend on the CO2 concentration (see Table 2). Hence, the CO2 concentration did not have any effect

284

on the FAIMS results. Reversely, Figure 3b and 3c reveal, that the acetone concentration did neither

285

affect the CO2 nor the O2 signal. Also, the O2 signal did not change depending on the CO2

286

concentration, but stayed constant irrespective if 3, 4 or 5 % CO2 were present. In conclusion, no

287

mutual co-dependencies of the acetone, CO2 and O2 signal were detected.

288

Table 2: Analytical figures of merit of the concentration-dependent FAIMS measurements of acetone. No statistical difference between fit parameters A, B and C of the asymptotic fit (equation y = A - B·Cx) at 3, 4 or 5 % CO2. R2, concentration at LOD and concentration at LOQ varied, yet with

no clear trend visible depending on the CO2 content. This indicates independence of the acetone

signal from the CO2 concentration.

3 % CO2 4 % CO2 5 % CO2

fit parameter A 0.998 ± 0.019 0.964 ± 0.025 0.989 ± 0.023 fit parameter B 1.024 ± 0.025 0.962 ± 0.038 0.997 ± 0.034 fit parameter C 0.803 ± 0.012 0.765 ± 0.022 0.772 ± 0.019

R2 > 0.995 0.989 0.992

LOD [ppt] 145 78 56

LOQ [ppt] 358 405 165

289

The FAIMS error bars shown in Figure 3a are relatively big compared to the FTIR and LS error

290

bars in Figures 3b and 3c, respectively. Several different sources have presumably contributed to the

291

acetone signal variance. First, three slight features apart from RIP, monomer and dimer peak were

292

visible in the dispersion plots of the acetone/CO2/air sample (see Figure 3c at ~65 % DF / -3 V CV, at

293

~50 % DF / -0.5 V CV and at ~75 % DF / +0.5 V CV). As mentioned before, these possibly appeared

294

due to contaminations from the CO2 gas bottle and due to evaporations from the tubings and the

295

FAIMS itself. Even if they do not seem to have fundamentally impacted the obtained data, these

296

contaminations might still have competed with acetone for the ionization energy in the FAIMS

297

Figure 3: No mutual signal co-dependencies of acetone, CO2 and O2 were detected. All displayed error bars are

1σ error bars. (a) Acetone signals recorded with FAIMS depend on the acetone concentration (asymptotic fit y = A – B·Cx), yet is independent of the CO2 content. (b) CO2 signals recorded by iHWG-FTIR only vary

depending on the CO2 concentration. (c) O2 signals recorded by LS are neither influenced by the acetone nor by

(9)

ionization region, therefore possibly altering the acetone signal intensity and increasing the

298

associated error bars. Furthermore, it cannot be excluded that slight humidity variations occured,

299

additionally enhancing the variance of the acetone signal. Finally, the saturation of the FAIMS

300

detector at higher acetone concentrations can be assumed to also have made a contribution to the

301

signal variance.

302

3.3 Towards Real Breath Analysis

303

It is our goal to further develop the hybrid FAIMS-FTIR-LS platform towards online analysis of

304

mouse breath. Already with the current setup, the detection of the main breath components CO2, O2

305

and acetone, as a breath VOC representative, was possible in breath-relevant concentrations.

306

Especially the fact that breath VOC detection is possible down to LODs and LOQs in the low to

307

medium ppt range with this hybrid setup, makes it a promising tool for real breath analysis, since

308

breath VOCs most often occur in ppt to ppb concentrations[33]. Furthermore, O2, CO2 and acetone

309

signal were found to be mutually independent. This underlines the excellent orthogonality of FAIMS,

310

FTIR spectroscopy and LS, making their combination especially suitable for a complex matrix like

311

exhaled breath: simply by selecting a suitable combination of analytical methods, a first – at least

312

virtual – “preseparation” of the sample components has been undertaken, thus already simplifying

313

the analytical task.

314

Nevertheless, the hybrid setup and the experiments conducted with it need to be further evolved

315

before online analysis of real mouse breath is possible. First, unlike in our model samples, of course

316

more than one VOC is present in real breath. All these breath VOCs will compete for the FAIMS

317

ionization energy and therefore cause co-dependencies of their signals. To prevent this, preseparation

318

based on a GC or an MCC column will be integrated into the hybrid setup, enabling the VOCs to

319

reach the ionization region one by one. Since the contaminations discussed above (see Figure 2c) will

320

also be separated from the analytes via the GC or MCC column, the FAIMS signal variance may

321

additionally benefit from the preseparation scheme. Furthermore, alkanes, as an important class of

322

breath VOCs[34], cannot be detected with the current setup, because they are not ionized by the 63Ni

323

ionization source. This problem could be overcome by taking advantage of the modular flexibility of

324

the FTIR detection unit: extending the optical path length of the iHWG and replacing the FTIR

325

spectrometer by a more intense light source like a tunable quantum cascade laser, the LOD/LOQ for

326

alkane detection via FTIR could be shifted to breath-relevant concentrations. Moreover, the samples

327

tested until now only contained minimal amounts of water, whereas real breath is oversaturated with

328

humidity. Since the FAIMS detection mechanism is based on ionized water clusters, changes in

329

humidity have a major effect on the FAIMS signal intensity. Here, chemometric data treatment in

330

dependence of the present water level or experimentally filtering out the humidity by a condenser as

331

proposed by Maiti et al.[35], which is explicitly suitable for dehumidifying breath without significant

332

VOC loss, could be possible solutions.

333

4. Conclusions

334

A compact hybrid sensing platform enabling orthogonal analysis of gas/vapor phase samples based

335

on FAIMS, FTIR and LS was presented, and its utility online analysis of synthetic breath samples

336

containing acetone, CO2 and O2 was demonstrated. It was shown that the signals of these compounds

337

were independent of one another, and that all three components could be detected at their respective

338

breath-relevant concentrations. The LOQ of acetone could even be lowered to the medium ppt

339

concentration range, which renders the method a promising approach for the potential analysis of

340

trace level breath VOCs. Yet, challenges according to nonetheless integrating additional analyte

341

preconcentration/-separation strategies and dealing with high humidity levels will need to be

342

resolved prior to the useful analysis of real-world exhaled breath samples, and will be addressed

343

during future studies.

344

Abbreviations: Å Angström, barg unit for gauge pressure in bar (pressure in bar exceeding atmospheric

345

pressure), CO2 carbon dioxide, CV compensation voltage, DF dispersion field, FAIMS field-asymmetric ion

(10)

mobility spectrometry, FTIR Fourier-transform infrared spectroscopy, GC gas chromatography, IABC Institute

347

for Analytical and Bioanalytical Chemistry, IAPMD Institute for Anesthesiological Pathophysiology and

348

Method Development, iHWG substrate-integrated hollow waveguide, min minutes, LS luminescence sensing,

349

MCC multicapillary column, µL microliter, MICU mouse intensive care unit, N2 nitrogen, O2 oxygen, OEM

350

original equipment manufacturer, PFA perfluoroalkoxy alkane, ppb parts per billion, ppm parts per million, ppt

351

parts per trillion, PTFE polytetrafluoroethylene, RIP reactant ion peak, RQ respiratory quotient, TDLAS tunable

352

diode laser absorption spectroscopy, THz Terahertz, VOC volatile organic compound, °C degree Celsius.

353

Acknowledgments: This work was supported by the Research training group PULMOSENS at Ulm University

354

(GRK 2203) and by the German National Academic Foundation (Studienstiftung des Deutschen Volkes).

355

Author Contributions: Conceptualization: B.M., L.T.H.; Design, establishment and optimization of setup:

356

L.T.H.; Design of experiment: L.T.H.; Final data acquisition: S.R.; Data evaluation: S.R., L.T.H.; Writing original

357

draft: L.T.H.; Writing-review and editing: B.M., S.R.; Funding acquisition: B.M., L.T.H.; All authors read and

358

approved the final manuscript.

359

Conflicts of Interest: The authors declare no conflict of interest. The founding sponsors had no role in the design

360

of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript, and in the

361

decision to publish the results.

362

Appendix

363

The primary signals of CO2 and O2 recorded by FTIR spectroscopy and LS, respectively, are

364

shown in Figure A1 and A2.

365

366

Figure A1: IR spectrum of 4 % CO2. Acetone theoretically also is IR active, but is not detected here

367

due to its extremely low concentrations in the ppb range.

368

369

Figure A2: O2 concentration as detected by the luminescence sensor.

370

References

371

1. Mueller, W.; Schubert, J.; Benzing, A.; Geiger, K. Method for analysis of exhaled air by microwave energy

372

desorption coupled with gas chromatography-flame ionization detection-mass spectrometry. J.

373

Chromatogr. B Biomed. Appl.1998, 716, 27–38.

374

1000 2000

3000 4000

0.0 0.5 1.0 1.5 2.0 2.5

CO

2

Abs

or

b

ance [1

]

Wavenumber [cm-1]

0 200 400 600 800 18.10

18.12 18.14 18.16 18.18 18.20

Oxyg

en

co

ncentration [%

]

(11)

2. Smolinska, A.; Klaassen, E.M.M.; Dallinga, J.W.; Van De Kant, K.D.G.; Jobsis, Q.; Moonen, E.J.C.; Van

375

Schayck, O.C.P.; Dompeling, E.; Van Schooten, F.J. Profiling of volatile organic compounds in exhaled

376

breath as a strategy to find early predictive signatures of asthma in children. PLoS One2014, 9, e95668.

377

3. Hüppe, T.; Lorenz, D.; Maurer, F.; Albrecht, F.W.; Schnauber, K.; Wolf, B.; Sessler, D.I.; Volk, T.; Fink, T.;

378

Kreuer, S. Exhalation of volatile organic compounds during hemorrhagic shock and reperfusion in rats:

379

An exploratory trial. J. Breath Res.2016, 10, 16016.

380

4. Basanta, M.; Koimtzis, T. Sampling and analysis of exhaled breath on human subjects with thermal

381

desorption- gas chromatography- differential mobility spectrometry. Int. J. Ion Mobil. Spectrom.2006, 9,

382

45–49.

383

5. Di Natale, C.; Macagnano, A.; Martinelli, E.; Paolesse, R.; D’Arcangelo, G.; Roscioni, C.; Finazzi-Agrò,

384

A.; D’Amico, A. Lung cancer identification by the analysis of breath by means of an array of

non-385

selective gas sensors. Biosens. Bioelectron.2003, 18, 1209–1218.

386

6. Machado, R.F.; Laskowski, D.; Deffenderfer, O.; Burch, T.; Zheng, S.; Mazzone, P.J.; Mekhail, T.;

387

Jennings, C.; Stoller, J.K.; Pyle, J.; et al. Detection of lung cancer by sensor array analyses of exhaled

388

breath. Am. J. Respir. Crit. Care Med.2005, 171, 1286–1291.

389

7. Peng, G.; Tisch, U.; Adams, O.; Hakim, M.; Shehada, N.; Broza, Y.Y.; Billan, S.; Abdah-Bortnyak, R.;

390

Kuten, A.; Haick, H. Diagnosing lung cancer in exhaled breath using gold nanoparticles. Nat.

391

Nanotechnol.2009, 4, 669–673.

392

8. Mazzone, P.J.; Wang, X.F.; Xu, Y.; Mekhail, T.; Beukemann, M.C.; Na, J.; Kemling, J.W.; Suslick, K.S.;

393

Sasidhar, M. Exhaled Breath Analysis with a Colorimetric Sensor Array for the Identification and

394

Characterization of Lung Cancer. J. Thorac. Oncol.2012, 7, 137–142.

395

9. De Vries, R.; Brinkman, P.; Van Der Schee, M.P.; Fens, N.; Dijkers, E.; Bootsma, S.K.; De Jongh, F.H.C.;

396

Sterk, P.J. Integration of electronic nose technology with spirometry: Validation of a new approach for

397

exhaled breath analysis. J. Breath Res.2015, 9, 46001.

398

10. Tiele, A.; Wicaksono, A.; Kansara, J.; Arasaradnam, R.P. Breath Analysis Using eNose and Ion Mobility

399

Technology to Diagnose Inflammatory Bowel Disease — A Pilot Study. Biosensors2019, 9, 55–70.

400

11. Fens, N.; van der Schee, M.P.; Brinkman, P.; Sterk, P.J. Exhaled breath analysis by electronic nose in

401

airways disease. Established issues and key questions. Clin. Exp. Allergy2013, 43, 705–715.

402

12. Vaks, V.L.; Domracheva, E.G.; Pripolzin, S.I.; Chernyaeva, M.B. Multifrequency high precise

subTHz-403

THz-IR spectroscopy for exhaled breath research. Terahertz Emit. Receiv. Appl. VII2016, 9934, 99340E–1

404

to 99340E–10.

405

13. Shorter, J.H.; Nelson, D.D.; McManus, J.B.; Zahniser, M.S.; Milton, D.K. Multicomponent Breath Analysis

406

With Infrared Absorption Using Room-Temperature Quantum Cascade Lasers. NIH Public Access2010,

407

10, 76–84.

408

14. Ibrahim, W.; Wilde, M.; Cordell, R.; Salman, D.; Ruszkiewicz, D.; Bryant, L.; Richardson, M.; Free, R.C.;

409

Zhao, B.; Yousuf, A.; et al. Assessment of breath volatile organic compounds in acute cardiorespiratory

410

breathlessness: A protocol describing a prospective real-world observational study. BMJ Open2019, 9,

411

1–13.

412

15. Inside the Breath Cluster https://www.insidebreath.com/en/, accessed on April 30th 2019.

413

16. Dharmawardana, N.; Ooi, E.; Goddard, T.; Woods, C.; Watson, D.; Yazbek, R. Characterization of breath

414

volatile organic compounds in patients with upper aerodigestive mucosal squamous cell carcinoma.

415

IABR Breath Summit 2018.

416

(12)

early diagnosis of post-operative complications following colorectal cancer and segmental resection.

418

IABR Breath Summit 2018.

419

18. Shimouchi, A.; Jinno, N.; Taniguchi, K.; Okumura, N.; Miyamoto, Y. Volatile compounds in exhaled air

420

associated with scores of general physical and mental conditions. IABR Breath Summit 2018.

421

19. Williams, J.; Stönner, C.; Wicker, J.; Krauter, N.; Derstroff, B.; Bourtsoukidis, E.; Klüpfel, T.; Kramer, S.

422

Cinema audiences reproducibly vary the chemical composition of air during films, by broadcasting

423

scene specific emissions on breath. Sci. Rep.2016, 6, 1–10.

424

20. Wilde, M.J.; Cordell, R.L.; Salman, D.; Zhao, B.; Ibrahim, W.; Bryant, L.; Ruszkiewicz, D.; Singapuri, A.;

425

Free, R.C.; Gaillard, E.A.; et al. Breath analysis by two-dimensional gas chromatography with dual flame

426

ionisation and mass spectrometric detection – method optimisation and integration within a large-scale

427

clinical study. J. Chromatogr. A2019, 1594, 160–172.

428

21. Tiele, A.; Covington, J. Design and development of an end-tidal breath capture device. IABR Breath

429

Summit 2018.

430

22. Sukul, P.; Trefz, P.; Kamysek, S.; Schubert, J.K.; Miekisch, W. Instant effects of changing body positions

431

on compositions of exhaled breath. J. Breath Res.2015, 9, 47105.

432

23. Sukul, P.; Schubert, J.K.; Kamysek, S.; Trefz, P.; Miekisch, W. Applied upper-airway resistance instantly

433

affects breath components: A unique insight into pulmonary medicine. J. Breath Res.2017, 11, 47108.

434

24. Fortes, P.R.; Wilk, A.; Seichter, F.; Cajlakovic, M.; Koestler, S.; Ribitsch, V.; Wachter, U.; Vogt, J.;

435

Radermacher, P.; Carter, C.; et al. Combined sensing platform for advanced diagnostics in exhaled

436

mouse breath. Front. Biol. Detect. From Nanosensors to Syst. V2013, 8570, 85700Q–1 to 85700Q–9.

437

25. Seichter, F.; Tütüncü, E.; Hagemann, L.T.; Vogt, J.; Wachter, U.; Gröger, M.; Kress, S.; Radermacher, P.;

438

Mizaikoff, B. Online monitoring of carbon dioxide and oxygen in exhaled mouse breath via

substrate-439

integrated hollow waveguide Fourier-transform infrared- luminescence spectroscopy. J. Breath Res.2018,

440

12, 036018.

441

26. Tütüncü, E.; Nägele, M.; Becker, S.; Fischer, M.; Koeth, J.; Wolf, C.; Köstler, S.; Ribitsch, V.; Teuber, A.;

442

Gröger, M.; et al. Advanced Photonic Sensors Based on Interband Cascade Lasers for Real-Time Mouse

443

Breath Analysis. ACS Sensors2018, 3, 1743–1749.

444

27. Wilk, A.; Chance Carter, J.; Chrisp, M.; Manuel, A.M.; Mirkarimi, P.; Alameda, J.B.; Mizaikoff, B.

445

Substrate-integrated hollow waveguides: A new level of integration in mid-infrared gas sensing. Anal.

446

Chem.2013, 85, 11205–11210.

447

28. Seichter, F.; Vogt, J.; Radermacher, P.; Mizaikoff, B. Nonlinear calibration transfer based on hierarchical

448

Bayesian models and Lagrange Multipliers: Error bounds of estimates via Monte Carlo – Markov Chain

449

sampling. Anal. Chim. Acta2017, 951, 32–45.

450

29. Covington, J.A.; van der Schee, M.P.; Edge, A.S.L.; Boyle, B.; Savage, R.S.; Arasaradnam, R.P. The

451

application of FAIMS gas analysis in medical diagnostics. Analyst2015, 140, 6775–6781.

452

30. Hagemann, L.T.; McCartney, M.M.; Fung, A.G.; Peirano, D.J.; Davis, C.E.; Mizaikoff, B. Portable

453

combination of Fourier transform infrared spectroscopy and differential mobility spectrometry for

454

advanced vapor phase analysis. Analyst2018, 143, 5683–5691.

455

31. Inczédy, A. IUPAC Compendium of Analytical Nomenclature: Definitve Rules. Chapter 18.4.3.7; 3rd ed.; 1997;

456

32. Eiceman, G.A.; Karpas, Z. Ion Mobility Spectrometry; 2nd ed.; Taylor & Francis Group LLC: Boca Raton,

457

FL, USA, 2005;

458

33. Phillips, M. Method for the collection and assay of volatile organic compounds in breath. Anal. Biochem.

459

(13)

34. Cheepsattayakorn, A.; Cheepsattayakorn, R. Breath tests in respiratory and critical care medicine: From

461

research to practice in current perspectives. Biomed Res. Int.2013, 2013, 702896.

462

35. Maiti, K.S.; Lewton, M.; Fill, E.; Apolonski, A. Sensitive spectroscopic breath analysis by water

463

condensation. J. Breath Res.2018, 12, 046003.

464

Figure

Table 1: Overview on the measurement protocol within the hybrid setup.
Figure 2. FAIMS dispersion plots. (a) pure air sample (b) CO2/air sample (c) acetone/CO2/air sample (d) background substracted acetone/CO2/air sample ((c) minus (b))
Table 2: Analytical figures of merit of the concentration-dependent FAIMS measurements of acetone
Figure A1: IR spectrum of 4 % CO2. Acetone theoretically also is IR active, but is not detected here due to its extremely low concentrations in the ppb range

References

Related documents

Interviewer: It sounds like, before coaching, you didn‟t necessarily talk to a lot other people about your studies, your school work, your classes and what goals you‟re working

The results of the study using multiple linear regression analysis showed that the three dimensions of tax fairness that affect MSMEs compliance are general fairness,

In order to show the performance of both proposed techniques, i.e., CP selection and symbol time randomization, in cyclic feature suppression, CAF of an OFDM signal is given in

Even 1 year later, when a pivotal ( “ phase II ” ) trial already begun to enroll human partici- pants [ 18 ], comprehensive reviews of the literature did not report any

The Terms of Reference (Annex B) require a review and collation of the approaches to aid allocation being used by different agencies, including the use of formal aid allocation

Specially designed for the MINI Cooper SE, the MINI Electric interior includes door armrests and knee rolls in Carbon Black, decorative door trims in elegant Piano Black