DNA IQ
TM
Casework Pro Kit
for Maxwell
®
16
A Validation Study
The search for a alternative DNA
extraction and purification
method …
In comparison with EZ1 / M48 (Qiagen):
Sufficient DNA quantity and quality
Usable for all kinds of different forensic biological
stains relevant to various carrier materials
First Validation Study
Maxwell
®16, Promega:
DNA IQTM Casework Pro Kit
(without new extraction buffer)
and
Blood LEV DNA Kit
BioRobot EZ1, Qiagen:
EZ1 DNA Investigator Kit
VS.
All tests were performed according to the manufacturer’s instructions with an incubation step of 70°(IQ Kit) / 56°C (Blood / Investigator Kit)
Sensitivity Study
blood, diluted 1:10, 1 µl on a swab, air-dried
5 samples / series
all tests were carried out in parallel
Quantification with RT-PCR (Quantifiler Human DNA Quantification Kit, AB)
each sample was quantified twice
RT-PCR set up was performed by an automatic liquid handler all samples were quantified within one RT-PCR run
elution volumes were adapted
Sensitivity Study
Results
General remarks:
yielded values only comparable within one series
results only give tendencies
but these tendencies could be confirmed by repeat testing
Average DNA Amount:
Blood Kit > Investigator > IQ Kit
Inhibition study
1 µl diluted blood (1:10) on different papers and adhesive tape -
air-dried- Swabs are the most common stains in our lab
stamps, paper from cigarette butts, paper towels and newspaper
adhesive tape (10cm long/2cm broad), swabbed with extraction buffer Once more:
all tests were carried out in parallel and within one RT-PCR run Quantifiler Human DNA Quantification Kit, AB
each sample was quantified twice
RT-PCR set up was done by a automatic liquid handler elution volumes were adapted
Results: Inhibition study
Number of successful investigated stains
Maxwell
EZ1
Blood Kit
IQ Pro Kit
Investigator
Cellulose 5 / 5* 0 / 5 5 / 5*
Newspaper 1 / 5 1 / 5 5 / 5
Stamps 1 / 6 not tested 6 / 6
Cig.-butts 2 / 4 not tested 4 / 4
Adhesive tape 5 / 5** not tested 5 / 5**
Average amount of yielded DNA: * Blood Kit > Investigator (~ 60%)
Conclusions
Maxwell / Blood Kit:
all kinds of biological stains on swabs
and adhesive tapes
EZ1 / Investigator Kit:
stains on all other carrier materials
Since most stains were swabs and tapes –
Second validation study
Blood LEV DNA Kit
DNA IQ
TMCasework
Pro Kit
(with the new extraction buffer)
VS.
All tests were performed according to the manufacturer's
instructions with an incubation step
of 70
°
(IQ Kit) / 56
°
C (Blood Kit) for > 2 hour.
Investigator Kit/EZ1 tested in parallel as “positive control” for
inhibition study.
Sensitivity study
1 µl diluted blood (1:10 and 1:100)
6 samples for each dilution step
Two different dilution series: blood directly added to buffer
air dried blood stains (1:10), on swabs Once more:
all tests were carried out in parallel and within one RT-PCR run (Quantifiler Human DNA Quantification Kit, AB)
each sample was quantified twice
RT-PCR set up was performed by an automatic liquid handler elution volumes were adapted
Results: Sensitivity study
1:10
Blood Kit
<
IQ Pro Kit <
EZ1
1:100
Blood Kit
<
IQ Pro Kit <
EZ1
Swabs
Blood Kit
<
IQ Pro Kit =
EZ1
Test repeated 2 weeks later showed
comparable results!
• Self designed lot test with dried blood stains on swabs
The 19 different lots of cartridges tested in this study revealed
“
remarkable differences in the amount of yielded DNA and
obtained profiles.”
9 lots:
no signal at all for all nine tested STR loci
6 lots:
profiles with more less allelic or locus drop outs
4 lots:
“full profiles” all alleles of the 9 loci were detected
Conclusion:
Lot test (Blood Kit)
1µl blood 1:10 and 1µl blood 1:10 on a swab, air-dried
1 µl blood (1:10)
Lot Number: 11223 9908 / 10882 / 9968 10754
DNA amount: 52pg/µl 42pg/µl 25pg/µl
1 µl blood (1:10) on a swab, air-dried
Lot Number: 11223 9908 / 10882 / 9968 10754
Conclusions
2005:
each new lot was checked with a self designed test system
before its use in routine casework
Qiagen improved the quality checks of the different reagents
comparable quality, only small differences were observed
2012:
Possible differences between the lots in use complicate a
comparison between the different kits and methods!
Inhibition study
1 µl diluted blood (1:10) and 1cm
2paper
different papers: stamps, paper from cigarette butts, cellulose 5 samples for each kind of paper Once more:
all tests were carried out in parallel and within one RT-PCR run (Quantifiler Human DNA Quantification Kit, AB)
each sample was quantified twice
RT-PCR set up was performed by an automatic liquid handler elution volumes were adapted
Results: Inhibition study
Number of successful investigated stains
Maxwell
EZ1
Blood Kit
IQ Pro Kit
Investigator
Cellulose 5 / 5 5 / 5 5 / 5
Stamps 0 / 5 5 / 5 5 / 5
Cig.-butts 0 / 5 5 / 5 5 / 5
Average amount of yielded DNA:
Cellulose: Blood Kit (84%) < Investigator (90%) < IQ (100%)
Silver bullet –
New extraction buffer for both kits
1µl blood 1:10 on a swab, air-dried
Maxwell
Blood Kit Blood Kit IQ Pro Kit
corresponding buffer new extraction buffer new extraction buffer
Summary
Extraction strategy was not changed!
Maxwell / Blood Kit:
all kinds of biological stains on swabs
(with new extraction buffer)