Regulations Concerning
Airborne Particle Counting
Airborne Particle Counting
Particle Counting
Applications
Applications
Two types of activities performedwith an airborne particle counter:
Environment
Classification focuses primarily on
the environment
Environment
Monitoring focuses on process,
people and the environment
People Process
People as a Contamination Source Shed 5 to 10 million skin cells each day Shed 5 to 10 million skin cells each day 2000 to 5000 microorganisms/cm2 on skin
Classification:
FS209 and ISO 14644-1
FED STD 209EISO 14644-1
FS209 and ISO 14644-1
Classification is the process of qualifying theEnglish Metric 1
ISO Class
2
cleanroom environment by the number of particles using a standard method
D i l ifi i f di
1 M1.5
10 M2.5
3
4
Determine classification of room according to standards
“ISO Class 5” or “Class 100”
Performed on a regular basis but not frequently 100 M3.5
1,000 M4.5
5
6
6 months, yearly or ???
Standards define minimum number of sample points
Usually based on area of cleanroom or clean zone
10,000 M5.5
100,000 M6.5
7
8
Standards define minimum amount of air to be sampled
Minimum volumes for statistically valid samples
Environmental Monitoring
Determine stability of room or zone over time; develop trend data Executed on a scheduled basis: daily weekly monthly
Minimum number of sample points defined by QA
Executed on a scheduled basis: daily, weekly, monthly
– Area of cleanroom or clean zone – Activity
– Risk to product
Frequency and volume of sample points defined by QA
– Frequent enough to show control – Frequent enough to provide
meaningful trend information
Process Monitoring
Determine readiness of room to carry out designated task Performed whenever relevant activity occurs – daily work Minimum number of sample points
defined by QA
A f l l
Performed whenever relevant activity occurs daily work
Area of cleanroom or clean zone Activity
Risk to product
Frequency and volume of sample points defined by QA
F t h t h t l
Frequent enough to show control Frequent enough to manage
financial risk of product contamination
N d i d b l i b b i k!
contamination
Differences
Classification Monitoring
Frequency 6 months or annual Daily, weekly, monthly or continuous
Number of positions By formula By need for data Number of positions By formula By need for data
Sample volume By formula By need for data
Pass/Fail criteria By table By need for trend info or control
or control
Reporting format By standard In form needed for rapid understanding
Differences
Classification Monitoring Distribution of counts in a room or zone Uniform or homogenous Unique at each sample positionClassification Standards for Airborne Particles
Classification Standards for Airborne Particles
General Air Monitoring Standards
– ISO 14644-1
• Classification of air cleanliness
– ISO 14644-2
• Specifications for testing and monitoring to prove continued compliance with ISO 14644-1
– ISO 14644-3ISO 14644 3
• Guidance on instrumentation to be used for testing for
compliance with ISO 14644-1
1999
G
l St
d
d
Electronics
General Standard
for all Industries
Electronics • Semiconductor • Flat Panel • Circuit Board • Optical • MEMS/Nanomachines Life Sciences • Pharmaceutical • Biotechnology • Medical Devices • Hospitals/Pharmacies Electronics Laboratory Other Aerospace • Launch Vehicles • Satellites • Commercial/Military Aircraft
Laboratories Life Sciences
Aerospace
• Analytical Laboratories • Universities
Other
• Nuclear
• Photographic, X-ray films • Automobile Painting
ISO 14644
ISO 14644 consists of the following parts, under the general title Cleanrooms
and associated controlled environments: and associated controlled environments:
Part 1: Classification of air cleanliness by particle concentration
Part 2: Monitoring to provide evidence of cleanroom performance by airborne Part 2: Monitoring to provide evidence of cleanroom performance by airborne
particle cleanliness.
Part 3: Test methods
Part 4: Design construction and start up Part 4: Design, construction and start-up Part 5: Operations
Part 6: Vocabulary
Part 7: Separative devices (clean air hoods gloveboxes isolators and mini Part 7: Separative devices (clean air hoods, gloveboxes, isolators, and
mini-environments)
Part 8: Classification of airborne molecular contamination
Part 9: Classification of surface cleanliness by particle concentration Part 9: Classification of surface cleanliness by particle concentration Part 10: Classification of surface cleanliness by chemical concentration
Classification Standard:
ISO 14644 1 1999
ISO 14644-1:1999
• Defines cleanroom classes in a zone • Defines cleanroom classes in a zone
• Establishes minimum sampling volumes p g
– Purpose: Gather a sample volume with theoretically at least 20 particles for a statistically valid sample
• Establishes minimum number of points to classify area, based on statistical criteria
– Gather from a valid number of locations for a representative sample of the total air volume
Classification Standard:
ISO 14644 1 1999 Li it
Cl N b f P ti l C bi M t b Mi t Si
ISO 14644-1:1999 Limits
Class Number of Particles per Cubic Meter by Micrometer Size0.1 m 0.2 m 0.3 m 0.5 m 1 m 5 m ISO 1 10 2 ISO 1 10 2 ISO 2 100 24 10 4 ISO 3 1,000 237 102 35 8 ISO 4 10,000 2,370 1,020 352 83 ISO 5 100,000 23,700 10,200 3,520 832 29 ISO 6 1 000 000 237 000 102 000 35 200 8 320 293 FS 209E Class 100 ISO 6 1,000,000 237,000 102,000 35,200 8,320 293 ISO 7 352,000 83,200 2,930 ISO 8 3,520,000 832,000 29,300 ISO 9 35,200,000 8,320,000 293,000
Classification Standard:
ISO 14644 1 1999
ISO 14644-1:1999
2.0 liter Minimum sample volume
Minimum sample time 1 minute Minimum number of
locations
1
with at least 3 samples total
Note: Typical sample volume may be larger than minimum listed above
especially for smaller size particles in very clean areas (better than ISO Class 5 or FS 209E Class 100)
Classification Standard:
ISO 14644 1 1999
Minimum Sample Volume
ISO 14644-1:1999
Minimum Sample Volume (in liters)
ISO Class 5, 0.5 microns = (20/3520) x 1000 = 5.6 liters ISO Class 5 5 microns = (20/29) x 1000 = 690 liters
ISO 14644-2:2000
Continued compliance
Continued compliance
(re-qualification)
Schedule of Tests to Demonstrate Continuing Compliance
Test Parameter Class Maximum Time Test Procedure
Particle Count ≤ ISO 5 6 Months ISO 14644-1
Test Parameter Class
Interval Test Procedure
Air Pressure
All Classes 12 Months ISO 14644-1
Test Annex A
ISO 6, 7, 8, 9 12 Months
Airflow All Classes 12 Months ISO 14644-1
A B4
Difference All Classes 12 Months Annex B5
Example
ISO 14644 1 C l l ti
ISO 14644-1 Calculations
Freeze Dryer 1 Vial Washing System 5 m Freeze Dryer 2 8 m 5 m Freeze Dryer 3 5Calculations for Number of Points: Area of clean zone = 80 m²
Take the SQRT (80) = 8.94
Rounding up to next integer = 9 sample positions
4 m
Example
ISO 14644 1 C l l ti
ISO 14644-1 Calculations
Freeze Dryer 1 Vial Washing System System 1 2 3 4 5 6 7 Freeze Dryer 2Calculations for Number of Points: Area of clean zone = 80 m²
8
Freeze Dryer 3
Area of clean zone = 80 m² Take the SQRT (80) = 8.94
Rounding up to next integer = 9 sample positions
Example
ISO 14644 1 C l l ti
1 2 3 4 5 6 7 8 9 10ISO 14644-1 Calculations
Freeze Dryer 1 Vial Washing System F System Freeze Dryer 2• Need to adjust for equipment in room.
• Under ISO 14644-1, if you sample at 10 or more
iti id th dd d l l ti f
Freeze Dryer 3
positions, you can avoid the added calculation of the UCL (Upper Confidence Limit). Calculation of the UCL is only mandated when the number of positions used is between 2 and 9.
• Best to sample near potential problem spots which
are near entrances and exits and near operator positions.
Example
ISO 14644 1 C l l ti
ISO 14644-1 Calculations
Freeze Dryer 1 Vial Washing System 1 2 8 9 10 Freeze 2 3 4 5 6 7 11 12 Dryer 2• Need to adjust for equipment in room.
• Under ISO 14644-1, if you sample at 10 or more
positions, you can avoid the added calculation of
12
13
Freeze Dryer 3
the UCL (Upper Confidence Limit). Calculation of the UCL is only mandated when the number of positions used is between 2 and 9.
• Best to sample near potential problem spots
which are near entrances and exits and near
14
Example
ISO 14644 1 C l l ti
ISO 14644-1 Calculations
1. Average the sample data values at each position
1. Average the sample data values at each position
2. Normalize the average to “number of particles per cubic meter” 3. Compare normalized value to the target class limit; normalized
value at each and every sample point must be less than the value at each and every sample point must be less than the limit for the given size and target room classification
If the number of points sampled is more than 1 but less than 10 If the number of points sampled is more than 1 but less than 10, then the UCL factor must be applied:
a) Calculate the standard deviation b) Use Student’s T factor from tables b) Use Student’s T-factor from tables c) Calculate UCL
d) Compare to classification limit; UCL must not exceed the applicable limit
Probable Revisions to
ISO 14644-1, -2
Revisions to
ISO 14644 1:1999
ISO 14644-1:1999,
14644-2:2000
Revision committee has met
in Zurich in the beginning of September 2012; b WEBEX i J l 2013
by WEBEX in July 2013:
in Reno October 2013
• Revision process still ongoing
• Revision process still ongoing
• Likely to be end of 2013 before new draft(s)
submitted for vote vote is recorded
submitted for vote, vote is recorded,
Forthcoming Revisions to
ISO 14644 1
ISO 14644-1
1. Frequency of re-certification for “continued
1. Frequency of re certification for continued
compliance”:
Now:
• If zone is ≤ ISO Class 5, every 6 months • If zone is ≥ ISO Class 6, every 12 months
Proposed:
• Remove re-certification periodRemove re certification period
• Re-certification timeframe will be set by regulatory or advisory committees for a particular industry
• Or: 1 year if not monitoring !! • Or: 1 year if not monitoring !!
Forthcoming Revisions to
ISO 14644 1
ISO 14644-1
2. Eliminate calculation of UCL (Student’s T test)
for sample plans with 2 to 9 sample positions
Forthcoming Revisions to
ISO 14644 1
3 Method of determining minimum number of
ISO 14644-1
3. Method of determining minimum number of
sample positions
– Replace with stated number of minimum sampleReplace with stated number of minimum sample positions as a look-up chart
– Based on 95% confidence levels not on SQRT of area
– May mean a small increase in the number of sample points
A.4.1.1
Derive the minimum number of sampling locations NL from table 3. Table A.1 shows the number of sample locations related to the area of each cleanroom or clean area of each cleanroom or clean zone to be classified and provides at least 95% confidence that at l t 90 % f th t t l d
least 90 % of the total area does not exceed the class limit.
Different levels of confidence and Different levels of confidence and
verification can be specified and agreed upon by the customer and
li supplier.
Forthcoming Revisions to
ISO 14644 1
4 Remove possibility to classify at 5 micron
ISO 14644-1
4. Remove possibility to classify at 5 micron
only for ISO Class 5
Limit number of 29 removed
Replace number with “Note (e)” :
Sample collection limitations for both particles in low concentrations and sizes greater than 1 um make classification inappropriate, due to potential particle pp p , p p
Forthcoming Revisions to
ISO 14644 1
5 Indicate that single digit limits for
ISO 14644-1
5. Indicate that single digit limits for
ISO Class 1 and 2 create challenges
to timely execution
y
R l
b
l
ith
“N t (b)”
Replace number values with “Note (b)” :
These concentrations will lead to large air sample These concentrations will lead to large air sample
volumes for classification. Sequential sampling procedure may be applied; see Annex D.
Forthcoming Revisions to
ISO 14644 1 2
6 New label for classification level
ISO 14644-1, -2
6. New label for classification level
ACP = Air Cleanliness (by) Particles ACC = Air cleanliness (by) Chemicals
SCP = Surface Cleanliness (by) Particles SCC = Surface Cleanliness (by) Chemicals SCC = Surface Cleanliness (by) Chemicals
Cleanroom Designations
Air Surfaces ParticlesACP
SCP
ISO 14644 – 9 ISO 14644 - 1, - 2 Draft #2 of Revision ChemicalsACC
SCC
ISO 14644 10 ISO 14644 8 ISO 14644 – 10 ISO 14644 – 8Classification Limits:
ISO 14644-1:1999
Number of Particles per Cubic Meter by Micrometer Sizep y
0.1 m 0.2 m 0.3 m 0.5 m 1 m 5 m ISO Class 1 10 2 ISO Class 2 100 24 10 4 ISO Class 3 1,000 237 102 35 8 ISO Class 4 10,000 2,370 1,020 352 83 ISO Class 5 100,000 23,700 10,200 3,520 832 29 ISO Class 6 1,000,000 237,000 102,000 35,200 8,320 293 FS 209E Class 100 , , , , , , ISO Class 7 352,000 83,200 2,930 ISO Class 8 3,520,000 832,000 29,300 ISO Class 9 35,200,000 8,320,000 293,000
Proposed new limits and labels:
p
Classification Limits: ISO 14644-1
(2014)
ISO-ACP1 10
0.1 m 0.2 m 0.3 m
Number of Particles per Cubic Meter by Micrometer Size 0.5 m 1 m 5 m 3 1,000 237 102 35 2 100 24 10 1 10 6 1,000,000 237,000 102,000 35,200 8,320 293 5 100,000 23,700 10,200 3,520 832 4 10,000 2,370 1,020 352 83 FS 209E Class 100 293,000 9 35,200,000 8,320,000 2,930 8 3,520,000 832,000 29,300 7 352,000 83,200 , , , , ,
Forthcoming Revisions to
ISO 14644 1 2
7. Calibration of Instruments to ISO 21501-4
ISO 14644-1, -2
7.1 Part of Appendix A (Normative):
A 2 2 I lib i
A.2.2 Instrument calibration
The instrument shall have a valid calibration certificate;
the frequency and method of calibration should be q y
based on current accepted practice as specified in
ISO 21501-4:2007 .
7.2 Also will be added to Bibliography:
ISO 21501-4:2007 Determination of particle size distribution – Single
particle light interaction methods-Part 4: Light scattering airborne particle particle light interaction methods Part 4: Light scattering airborne particle counter for clean spaces.
Probable Revisions to
ISO 14644-2
New Title for Section:
ISO 14644 2
Cleanrooms and associated controlled
environments Part 2: Monitoring to
environments — Part 2: Monitoring to provide evidence of performance by ACP
Focus of 14644-2 will be on methods covering
it i f ti l d k ti t
Probable Revisions to ISO 14644-2
I t
d
ti
Introduction
"Thi
i i
f ISO 14644 2
fl
"This revision of ISO 14644-2 reflects a
philosophical shift that emphasizes
it i
l
ifi ti
"
Probable Revisions to ISO 14644-2
I t
d
ti
"Thi
i i
f ISO 14644 2
fl t
Introduction
"This revision of ISO 14644-2 reflects a
philosophical
shift
that
emphasizes
monitoring over classification
monitoring over classification.
The monitoring process provides a continuing
The monitoring process provides a continuing
flow of data over time, offering a greater
assurance
of
the
performance
of
the
assurance
of
the
performance
of
the
installation."
Probable Revisions to ISO 14644-2
I t
d
ti
"Potential benefits gained from more substantial
Introduction
monitoring are:
Faster response to adverse conditions – Faster response to adverse conditions – Ability to Trend data over time
– Additional parameters in conjunction with airborne counts – Integration of data from multiple instruments
– Enhanced knowledge of installation and process allows more effective risk analysis
Probable Revisions to ISO 14644-2
I t
d
ti
"ISO 14644-2 specifies requirements of a monitoring plan,
Introduction
based on a risk assessment of the intended use. The data obtained provides evidence over time of continuing cleanroom or clean zone performance related to airborne cleanroom or clean zone performance related to airborne cleanliness by particles (ACP).
In some circumstances, relevant regulatory agencies may impose supplementary policies or restrictions. In such situations appropriate adaptations of the monitoring situations, appropriate adaptations of the monitoring procedures may be required."
Probable Revisions to ISO 14644-2
I t
d
ti
"The level of airborne particles measured under a
Introduction
monitoring plan may typically be higher than the level observed during the at-rest classification process. The observed values may fluctuate considerably due to
factors such as, but not limited to, the number of
personnel present, the rate of air flow or exchange, the operation of instruments or machinery, and activities in adjacent spaces "
Probable Revisions to ISO 14644-2
I t
d
ti
"In processes that inherently produce particles as part of
Introduction
the process and where these particles are not a threat to the process or product, it may be appropriate to rely on periodic at-rest classification rather than monitoring of periodic at rest classification rather than monitoring of airborne particles in operation. Other performance and cleanliness attributes may still be required to be
monitored monitored.
After a monitoring plan is initially established, it may be necessary to revise the plan when significant changes y p g g are made to the installation or process requirements. It is also prudent to conduct periodic reviews of a monitoring plan based on data obtained "
Probable Revisions to ISO 14644-2
T bl
f
t
t
IntroductionTable of contents
1 Scope 2 Normative references3 Terms and definitions
4 Creating, executing and maintaining a monitoring plan
4.1 Principle
4 2 Risk Analysis
4.2 Risk Analysis
4.3 Monitoring plan
4.4 Review and approval
4.5 Analyzing Data
Probable Revisions to ISO 14644-2
T bl
f
t
t
Annex A (informative)
Table of contents
Matters to consider when developing a monitoring plan A.1 General Considerations
A.2 Pressure differential monitoring
A.3 Airborne particle monitoring system A.4 Airflow velocity and volume monitoring
Annex B (informative)
Matters to consider when setting Warning levels B.1 Setting levels
B.2 Considerations in setting Warning levels for Pressure Differential B.3 Consideration in setting Warning levels for Airborne Particle
Monitoring
- Environmental - Process
Environmental
Monitoring
Monitoring
Understanding stability of particulate levels through trends
Samples taken daily, weekly or monthly
No requirement to monitor as many points as needed for classification
No requirement to sample 1 cubic meter
Consider using viable sample points as guidance in choosing positions and number of sample points in g p p p an area
Process
Monitoring
Monitoring
In operation or dynamic only
Relates to process and product quality Data may be needed for product release Data may be needed for product release
Often included with Batch Record
Average only count data from same location Average only count data from same location
– Each point must be below limit – But not necessarily each sample
M b OK d l ALERT ACTION il 3 d 5 h – May be OK to delay ALERT or ACTION until 3rd or 5th
consecutive high reading
Cleanroom
Monitoring
Monitoring
No clear guidance from standards or regulation
Number of sample points – Number of sample points – Positions of sample points
It is necessary to consider the purpose of monitoring within the specific manufacturing context.
– Assess the potential for product exposure – Assess the potential for product exposure
– Consider possible contaminants created by process
– Impact to product – Impact to operators – Impact to operators
Cleanroom
Monitoring
However two guidances are offered:
Monitoring
US FDA Guidance for Industry:
Sterile Drug Products Produced by Aseptic Processing
- Sterile Drug Products Produced by Aseptic Processing EU GMP Annex 1U G e
FDA’s Guidance for Industry:
Sterile Drug Products g Produced by Aseptic Processing
Initial release: June 1987
New Revision: September 2004
Restates count values in metric format – Cubic foot cubic meterCubic foot cubic meter
– 1 cubic meter = 35.31 cubic feet Refers to ISO 14644-1
Refers to ISO 14644-1
– Class 100 > ISO Class 5 – Limit restated
3520 counts/m³ for counts 0 5 μm and larger 3520 counts/m for counts 0.5 μm and larger
FDA’s Guidance for Industry:
Sterile Drug Products
F t ti l d t
g
Produced by Aseptic Processing
Focus on potential product exposure Defines two zones
Defines two zones
– Critical (similar to Grade A)
– Controlled (similar to Grade C or D support areas)
Monitor at 0.5 microns Sample point location
– “Not more than 1 foot away from the work site, within the airflow, and during filling/closing operations”
Guidances for
Aseptic Processing
Aseptic Processing
Pharmaceutical Industry Compliance Guidance Compliance Guidance– FDA cGMP Guidance for Industry
• Sterile Drug Products Produced by Aseptic Processing – EU GMP Annex I (EMEA) • Manufacture of Sterile Medicinal Products Medicinal Products
EU GMP Annex 1 is more specific than FDA cGMP Guidance
Guidances for
Aseptic Processing
Aseptic Processing
Guidances for
Aseptic Processing
Aseptic Processing
PIC/S Guidance
7. ENVIRONMENTAL AND PERSONNEL MONITORING Annex I of the EU/PIC/S Guide to GMP provides the basis
for environmental and personnel monitoring requirements and recommendations
and recommendations.
Some specific additional guidance is given below on air borne microbial and non-viable particle monitoring, intervention monitoring and staff training.
PIC/S Guidance
7.1 Air Borne Microbial and Non-Viable Particle Monitoring
7.1.1 It is important to state that the monitoring activity itself should not compromise the product quality Worst case should not compromise the product quality. Worst case scenarios of simulations tests should also include
PIC/S Guidance
7.2 Non-viable monitoring
7.2.1 The location chosen for monitoring should be checked to ensure that the positions reflect the worst case. For
room monitoring the counts should be performed in room monitoring, the counts should be performed in locations where there is most operator activity. For the filling environment the counts should be performed
dj h filli d h
adjacent to the filling zone and where components are exposed in such way as to detect operator activity within these areas.
PIC/S Guidance
7.2 Non-viable monitoring
7.2.1 (continued) Monitoring with sampling probes located in such a way that they monitor the air from the HEPA filter rather than the air immediately surrounding the filter rather than the air immediately surrounding the
critical zones should be avoided. However the location of the sample device should not compromise the laminarity
f h i fl i h i i l I i i l lid i h ld
of the air flow in the critical zone. Initial validation should be checked to confirm that worst case positions have
been adequately identified. These may be reconfirmed q y y
EU Annex 1:
L t t
i i
(2009)
Latest revision (2009)
Classification
– Sections 4 through 7
g
Monitoring
Monitoring
– Sections 8 through 17
EU Annex 1 Summary:
Cl
ifi ti
Classification
Classification – Sections 4 through 7
Classification – Sections 4 through 7
Section 4:
“Classification should be clearly differentiated from operational process environmental monitoring.” Section 5:
“ For classification purposes in Grade A zones, a minimum p p
sample volume of 1 m3 should be taken per sample
EU Grade Definitions
Grade Activity 0.5 m 5 m 0.5 m
A High Risk - filling, open vials, stopper bowls 3 520 20 3 520 20
B Aseptic preparations 3 520 29 352 000 2 000
at rest in operation maximum permitted number of particles/m3 equal to or above
5 m
• Zone grades according to risk of product contamination
C Clean area of less critical operations 352 000 2 000 3 520 000 20 000 D Clean area of less critical operations 3 520 000 20 000 not defined not defined
• Zone grades according to risk of product contamination • Particle count measurements at 0.5 m and 5 m
EU Annex 1:
L
i i
(2009)
Limits at 5 microns for Grade A
Latest revision (2009)
At RestAt Rest In OperationIn Operation
1 per cubic meter 20 per cubic meter
Grade
At Rest In Operation Maximum permitted number of particles/m3
equal to or greater than the tabulated size
Grade
At Rest In Operation
Maximum permitted number of particles/m3 equal to or greater than the tabulated size
0.5 µm 5 µm 0.5 µm 5 µm A 3 500 1 3 500 1 0.5 µm 5 µm 0.5 µm 5 µm A 3 520 20 3 520 20 B 3 500 1 350 000 2 000 C 350 000 2 000 3 500 000 20 000 D 3 5000 000 20 000 t d fi d t d fi d B 3 520 29 352 000 2 900 C 352 000 2 900 3 520 000 29 000
D 3 520 000 29 000 not defined not defined
D 3 5000 000 20 000 not defined not defined
EU Annex 1 Summary:
Classification
Section 5: Section 5:
“For classification purposes EN/ISO 14644-1 methodology p p gy
defines both the minimum number of sample locations and the [minimum] sample size based on the class limit of the largest considered particle size and the method of of the largest considered particle size and the method of evaluation of the data collected.”
EU Annex 1 Summary:
Classification
Section 5 (continued)
“F l ifi ti EN/ISO 14644 1 th d l
“For classification purposes EN/ISO 14644-1 methodology defines both the minimum number of sample locations and the [minimum] sample size based on the class limit of the largest considered particle size and the method of evaluation of the data collected.”
Number of locations
• Currently based on SQRT of Area (M^2)
Sample Volume (B,C,D)
EU Annex 1 Summary:
Monitoring
Monitoring
Monitoring: Sections 8 through 17
Section 8:
“Clean rooms and clean air devices should be routinely y monitored in operation and the monitoring locations based on
a formal risk analysis study – a formal risk analysis study and
– the results obtained during the classification of rooms and/or l d i ”
EU Annex 1 Summary:
Monitoring
Section 9
Monitoring
• “The Grade A zone should be monitored at such a frequency and with suitable sample size that all interventions transient events and any system interventions, transient events and any system
deterioration would be captured and alarms triggered if alert limits are exceeded.
EU Annex 1 Summary:
Monitoring
Section 12:
Monitoring
• “The sample sizes taken for monitoring purposes using automated systems will usually be a function of the
sampling rate of the system used It is not necessary for sampling rate of the system used. It is not necessary for the sample volume to be the same as that used for formal classification of clean rooms and clean air devices.”
• It is not necessary to sample 1m3 during verification
or monitoring
• Particle counters used for monitoring may have the same or different flow rate from those used for
classification classification.
Monitoring Positions:
Risk-based Approach
Lyo 1 Vial Sterilizing Tunnel 4 5 1 Tunnel 3 7 2 Lyo 2 6 • Monitoring must follow the workflow, covering areaswhere product is exposed – Annex 1 (2009)
− Where open vials exit de-pyrogenation – human interaction (1)
Lyo 3 − Where vials are filled (2,3)
− Surrounding Grade B background (4) ─ Where the vials are partially stoppered (5)
─ Loading area in front of lyophilizers must be Grade A if g y p product is not fully stoppered (6,7)
Monitoring Positions:
Risk-based Approach
4 Vial Washing System 1 2 3In a filling operation for which the final product remains liquid, some points established for a lyophilized points established for a lyophilized product would not be needed.
ISO 14644 ISO 21501
d
ISO 14644, ISO 21501 and
EU GMP Annex 1
Sampling of Airborne Particle Counts
I A ti M f t i P
Air Particle Counter Calibration
ISO 21501 and ISO 14644 the link to GMP
ISO 21501 and ISO 14644 - the link to GMP
• Regulators inspect to EU GMP, which calls up ISO 14644
calls up ISO 14644
• Next revision ISO 14644 will refer to ISO 21501-4
• ISO 21501-4 states – “Instruments that
conform to this part of ISO 21501 are used for the classification of air
cleanliness in cleanrooms and
associated controlled environments in accordance with ISO 14644-1”
Proposed wording in -1 -2
Proposed wording in 1, 2
A.2.2 Airborne particle counter calibration
The airborne particle counter shall have a valid calibration certificate; the frequency and method of calibration should be based on current accepted practice as specified in ISO be based on current accepted practice as specified in ISO 21501-4:2007 .
NOTE: Some airborne particle counters cannot be calibrated to all of the required tests in ISO 21501-4:2007.
ISO 21501-4: Additional Tests
Before ISO 21501-4
Si lib ti
ISO 21501-4
Si lib ti
• Size calibration • Size calibration
• Verification of size setting • Counting efficiency
• False count rate
• Counting efficiency • Size resolution
• False count rate • Sampling Flow Rate
• Concentration limit • Sampling flow rate
• Sampling Time • Sampling time
ISO 21501-4 Calibration Standard Delivers:
• Improved compliance
– Removes ambiguity by providing a single internationally recognized standard method for calibration
recognized standard method for calibration
– Harmonization between ISO and GMP guidance
I d it t it d ibilit
• Improved unit-to-unit reproducibility • Improved counting accuracy
• Improved counting accuracy
All current Met One particle counters from Hach may be calibrated using ISO 21501-4 at your facility
ISO 21501-4
ISO 21501 4
ISO 21501-4
ISO 21501-4
ISO 21501-4
ISO 21501 4
What to look for on the calibration certificate
“Hach certifies that the calibration performed
complies with the complies with the
requirements of ISO 21501 . . .
Online and
Portable Particle Counters
Portable Particle Counters
First particle counters designed for ISO 21501 compliance
for ISO 21501 compliance
MET ONE 3400 Cleanroom classification to ISO14644 Cleanroom classification to ISO14644 Portable sampling for environmental validation MET ONE 6015P
Grade A & B automated monitoring Internal vacuum pump
MET ONE 6000 MET ONE 7000
MET ONE 6000
Grade A & B automated monitoring Small and compact
MET ONE 7000 Grade A & B automated monitoring
ADDITIONAL INFORMATION
Road Map
to ISO 21501-4 compliance
Audit your Methods and Equipment ud you e ods a d qu p e
Training
Standard Operating Procedures (SOPs) Validation (IQ/OQ)
Calibration Policies
Equipment Upgrade as needed
Audit your Calibration Service
Factory-trained and authorized Correct equipment and standards
Thank you !
Thank you !
[email protected]
+1 541 210 0194
Presented by
Grants Pass, Oregon USA
Joe GecseyHACH