• No results found

ANALYTICAL METHOD DEVELOPMENT AND VALIDATION OF CLOXACILLIN AND CEFIXIME SIMULTANIOUS IN TABLET DOSAGE FORM BY RP HPLC

N/A
N/A
Protected

Academic year: 2020

Share "ANALYTICAL METHOD DEVELOPMENT AND VALIDATION OF CLOXACILLIN AND CEFIXIME SIMULTANIOUS IN TABLET DOSAGE FORM BY RP HPLC"

Copied!
16
0
0

Loading.... (view fulltext now)

Full text

(1)

ANALYTICAL METHOD DEVELOPMENT AND VALIDATION OF

CLOXACILLIN AND CEFIXIME SIMULTANIOUS IN TABLET

DOSAGE FORM BY RP-HPLC

1

Dr. M. Prasadarao and 2S. Lakshmi Sumanoja*

1

Department of Pharmaceutical Sciences, Acharya Nagarjuna University, Nagarjuna Nagar,

Guntur, Andhra Pradesh- 522 510.

2

M.A.M. College of Pharmacy, Kesanupalli, Narsaraopet, Guntur, A.P.

ABSTRACT

A simple reverse phase high perfomence liquid chromatographic

method has been developed and subsequently validatedfor

simultaneous determination of cefixime and cloxacillin in combined

dosage form. The separation was carried out using a mobile phase

consisting of methanol and mono basic potassium phosphate buffer in

the ratio of 60:40 v/v. The column used was Inertsil –ODS C18 (250 ×

4.6 mm, 5μ) with flow rate of1.0ml/min using PDA detection at

272nm. The described method was linear over a concentration range of

20ppm to 80ppm for the assay of cefixime and cloxacillin respectively.

The retention times of cefixime and cloxacillin were found to be

2.955min and 3.532min respectively. Results of analysis were validated statistically and by

recovery studies. The limit of quantification (LOQ) for cefixime and cloxacillin were found

to be 0.77mg/ml and 1.05mg/ml respectively. Then the limit of detection (LOD) for cefixime

and cloxacillin were found to be 0.25 mg/ml and 20.34 mg/ml respectively. The results of the

study showed that the proposed RP-HPLC method is simple, rapid, precise and accurate

which is useful for the routine determination of cefixime and scloxacillin bulk drug and in its

pharmaceutical dosage form.

KEYWORDS: cefixime, cloxacillin, methanol.

INTRODUCTION

Cefixime and cloxacillin is a combintion drug isa, often consisting of 200mg cefixime with

Volume 4, Issue 8, 2549-2564. Research Article ISSN 2277– 7105

Article Received on 17 June 2015,

Revised on 08 July 2015, Accepted on 29 July 2015

*Correspondence for Author

S. Lakshmi Sumanoja

M.A.M. College of

Pharmacy, Kesanupalli,

(2)

third-generation cephalosporin like ceftriaxone and cefotaxime. Cefixime is highly stable in

the presence of beta-lactamase enzymes. As a result, many organisms resistant to penicillins

and some cephalosporins due to the presence of beta-lactamases, may be susceptible to

cefixime. The antibacterial effect of cefixime results from inhibition of mucopeptide

synthesis in the bacterial cell wall.[2] Cloxacillin is a semisynthetic antibiotic in the same class

as penicillin. Cloxacillin is for use against staphylococci that produce beta-lactamase.[3]

Cefixime Cloxacillin

MATERIALS AND METHOD

Chemicals and solvents: Cefixime, Cloxacillin as gift samples from Mylan Laboratories Limited, Hyderabad, India. The commercial Pharmaceutical tablet formulation of Ceftas

containing 200mg cefixime, 500mg Cloxacillin (manufactured by Intas laboratories) were

procured from local pharmacy. Potassium dihydrogen phosphate – AR grade (SD.Fine chem.

Ltd, mumdai), Acetonitrile-HPLC grade (Merck India), Methanol – HPLC grade (Merck

India).

INSTRUMENTATION

The chromatographic separations were performed using HPLC-Waters alliance

(Model-2690/5) consisting of an in-built auto sampler, a column oven and Waters 996 PDA detector.

The data was acquired through Empower-2-software. The column used was Inertsil ODS

(250×4.6 mm, 5μ). Meltronics sonicator was used for enhancing dissolution of the

compounds. Elico pH meter was used for adjusting the pH of buffer solution. All weighing

was done on Sartorious balance (model AE-160).

(3)

60:40 v/v to the column in the flow rate of 1.0 ml/min whereas run time set was 10 min. The

separation was performed on Inertsil ODS-3V 250mm x 4.6mm, 5μm column and the column

was maintained the temperature ambient and the volume of each injection was 20µl. Prior to

injection, the column was equilibrated for at least 30 min with mobile phase flowing through

the system. The eluents were monitored at 272 nm.

OPTIMISED METHOD

Mobile Phase: Degassed Methanol and Buffer in the ratio of 60:40 V/V.

Preparation of(KH2PO4 0.1M) buffer:Weight 3.8954g of di-sodium hydrogen phosphate and 3.4023 of potassium dihydrogen phosphate in to a beaker containing 1000ml of distilled

water and dissolve completely. Then ph is adjusted with orthophosphoric acid and then

filtered through 0.45µm membrane filter.

Preparation of stock solution: Reference solution: The solution was prepared by dissolving 20.0 mg of accurately weighed Cefixime and 25.0 mg Cloxacillin in Mobile phase, in two

100.0 mL volumetric flasks separately and sonicate for 20min. From the above solutions take

10.0 mL from each solution into a 50.0 mL volumetric flask and then makeup with mobile

phase and sonicate for 10min.

Preparation of working standard solution: The stock solutions equivalent to 20ppm to 80ppm with respect to both drugs were prepared in combination of Cefixime and Cloxacillin

above, sonicated and filtered through 0.45µ membrane.

Optimized chromatographic conditions

Parameters Method

Stationary phase (column) Inertsil -ODS C18(250 x 4.6 mm, 5 µ)

Mobile Phase Methanol : Buffer (60:40)

Flow rate (ml/min) 1.0 ml/min

Run time (minutes) 10 min

Column temperature (°C) Ambient

Volume of injection loop (l) 20

Detection wavelength (nm) 272nm

(4)

C

e

f

ix

im

e

-2

.

9

5

5

C

lo

x

a

c

il

li

n

-3

.

5

3

8

AU

0.00 0.02 0.04 0.06 0.08 0.10 0.12 0.14 0.16 0.18

Minutes

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00

Fig 1: Chromatogram of standard

Inference: Got chromatogram at RT’s of 2.955min to Cefixime and 3.522min to Cloxacillin.

METHOD VALIDATION

The developed method was validated as per the ICH (International Conference on

Harmonization) guidelines with respect to System suitability, Specificity,

Linearity,presission, Accuracy, Limit of detection and Limit of quantification.

1. SYSTEM SUITABILITY

A Standard solution was prepared by using Cefixime and Cloxacillin working standards as

per test method and was injected Five times into the HPLC system.The system suitability

parameters were evaluated from standard chromatograms by calculating the % RSD from five

replicate injections for Cefixime and Cloxacillin , retention times and peak areas.

ACCEPTANCE CRITERIA

1. The % RSD for the retention times of principal peak from 5 replicate injections of each

Standard solution should be not more than 2.0 %

2. The % RSD for the peak area responses of principal peak from 5 replicate injections of

each standard Solution should be not more than 2.0%.

3. The number of theoretical plates (N) for the Cefixime and Cloxacillin peaks is NLT 3000.

4. The Tailing factor (T) for the Cefixime and Cloxacillin

[image:4.595.136.460.74.288.2]
(5)

Table- 1: Data of System Suitability for cefixime and cloxacillin

Cefixime Cloxacillin

Injection RT Peak area USP plate count

USP

tailling RT

Peak area USP platecount USP trailling

1 2.955 1139272 5890.964069 1.238915 3.522 797564 8676.113795 1.099100

2 2.955 1140892 5915.423628 1.230637 3.532 795138 8803.641669 1.103929

3 2.954 1136301 5934.796986 1.240858 3.529 795685 8616.937115 1.111477

4 2.953 1141067 5976.253744 1.238995 3.526 800569 8820.182543 1.117660

5 2.954 1136024 5953.814152 1.241073 3.525 797049 8735.115629 1.119004

MEAN 2.954218 1138711 5934.251 1.236496 3.527241 797201 8730.398 1.110234

SD 0.000837 57540.015 --- --- 0.002683 2124.413 --- ---

%RSD 0.028363 0.213538 --- --- 0.064022 0.266484 --- ---

C e f ix im e -2 . 9 5 5 C lo x a c il li n -3 . 5 3 8 AU 0.00 0.02 0.04 0.06 0.08 0.10 0.12 0.14 0.16 0.18 Minutes

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00

Inference: System suitability Chromatogram for standard – 1

C e f ix im e -2 . 9 5 5 C lo x a c il li n -3 . 5 3 2 AU 0.00 0.02 0.04 0.06 0.08 0.10 0.12 0.14 0.16 0.18 Minutes

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00

(6)

C e f ix im e -2 . 9 5 4 C lo x a c il li n -3 . 5 2 9 AU 0.00 0.02 0.04 0.06 0.08 0.10 0.12 0.14 0.16 0.18 Minutes

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00

Inference: System suitability Chromatogram for standard – 3

C e f ix im e -2 . 9 5 3 C lo x a c il li n -3 . 5 2 6 AU 0.00 0.02 0.04 0.06 0.08 0.10 0.12 0.14 0.16 0.18 Minutes

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00

Inference: System suitability Chromatogram for standard - 4

C e fi x im e - 2 .9 5 4 C lo x a c il li n 3 .5 2 5 AU 0.00 0.02 0.04 0.06 0.08 0.10 0.12 0.14 0.16 0.18 Minutes

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00

(7)

OBSERVATION

The %RSD for retention times and peak areas were found to be within the limit. Refer table:

1 As shown in fig 1 – 5.

2. SPECIFICITY

Cefixime and Cloxacillin

Solutions of standard and sample were prepared as per the test method are injected into

chromatographic system.

ACCEPTENCE CRITERIA

Chromatograms of standard and sample should be identical with near Retention time.

C

e

f

ix

im

e

-2

.

9

5

3

C

lo

x

a

c

il

li

n

-3

.

5

2

3

AU

0.00 0.02 0.04 0.06 0.08 0.10 0.12 0.14 0.16 0.18

Minutes

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00

Fig 6

Inference: Got a peak for sample at an Rt of 2.953min for Cefixime and 3.523min for Cloxacillin.

OBSERVATION: The chromatograms of Standard and Sample were same identical with same retention time. As shown in fig: 6.

3. LINEARITY OF TEST METHOD

A Series of solutions are prepared using Cefixime and Cloxacillin working standards at

concentration levels from 20ppm to 80 ppm of target concentration .Measure the peak area

[image:7.595.144.452.298.528.2]
(8)

ACCEPTANCE CRITERIA

Correlation Coefficient should be not less than 0.9990.

% of y- Intercept should be ±2.0.

% of RSD for level 1 and Level 6 should be not more than 2.0%.

Table 2: Data of Linearity cefixime and cloxacillin Concentration

(ppm)

Average Area

Statistical Analysis of cefixime

Average Area

Statistical analysis of miconazoe

0 0 Slope 31282 0 Slope 20193

20 523467 y-Intercept 11218 412977 y-Intercept 1902

30 829544 Correlation

Coefficient 0.999 605369

Correlation

Coefficient 0.999

40 1139272 807564

50 1448018 1007428

60 1728926 1210925

70 2089505 1409560

80 2407574 1627087

[image:8.595.32.557.181.614.2]
(9)

Fig: 7(b) Linearity Plot (Concentration Vs Response)ofCloxacillin

OBSERVATION

The linear fit of the system was illustrated graphically. The results are presented in table2.

4. PRECISION 4.1Repeatability

a. precision: Standard solution prepared as per test method and injected five times.

b. Method precision: Prepared six sample preparations individually using single as per test

method and injected each solution.

ACCEPTANCE CRITERIA: The % relative standard deviation of individual Cefixime and Cloxacillin, from the six units should be not more than 2.0%.

The individual assays of Cefixime and Cloxacillin should be not less than 98% and not more

than 102.0%.

OBSERVATION

Test results are showing that the test method is precise. Refer tables 3and 4 for system

(10)
[image:10.595.42.556.81.423.2]

Table 3: Data of Repeatability (System precision) for cefixime and cloxacillin

Concentration 40ppm

injection Peak area of

cefixime %Assay

Peak area of

cloxacillin %Assay

1 1146923 99.65 801690 98.84

2 1143596 99.08 797631 99.69

3 1158293 99.98 805783 100.05

4 1147283 100.04 801496 101.11

5 1152490 100.16 806432 100.96

Statistical

Analysis

Mean 1149717 99.78 802606.4 100.13

SD 5754.015 0.435569 3590.034 0.937203

b) Method precision

Concentration 40ppm

Injection Peaks areas

of cefixime %Assay

Peaks areas

of cloxacillin %Assay

1 1152293 99.55 805783 99.85

2 1146923 99.88 801690 99.96

3 1147283 99.40 801496 100.53

4 1152490 99.56 806432 100.30

5 1139272 99.85 797564 100.08

6 1147283 99.40 801496 100.53

Statistical Analysis

Mean 1147591 99.67 801593 100.20

SD 4815.615 0.250093 3262.714 0.290477

%RSD 0.419628 0.250913 0.406614 0.289873

4.2 Intermediate precision (analyst to analyst variability)

A study was conducted by two analysts as per test method

ACCEPTENCE CRITERIA

The individual assays of Cefixime and Cloxacillin should be not less than 98% and not more

than 102% and %RSD of assays should be NMT2.0% by both analysts.

Table 5: Data of Intermediate precision (Analyst 2) for cefixime and cloxacillins

Concentration 40ppm

Injection Peak Areas of

cefixime %Assay

Peak areas of

cloxacillin % Assay

1 1139272 98.80 797564 99.85

2 1140892 99.54 795138 99.68

3 1136601 99.98 795685 100.08

4 1141067 100.02 800569 100.01

5 1136024 101.08 797049 99.52

6 1140892 99.54 795685 100.08

Statistical Analysis

Mean 1139125 99.82 796948.3 100.37

SD 2281.417 0.755001 1998.386 0.337086

[image:10.595.49.552.574.738.2]
(11)

OBSERVATION

Individual %assays and %RSD of Assay are within limit and passes the intermediate

precision, Refer table: 5

5. ACCURACY (RECOVERY)

A study of Accuracy was conducted. Drug Assay was performed in triplicate as per test

method with equivalent amount of Cefixime and Cloxacillin into each volumetric flask for

each spike level to get the concentration of Cefixime and Cloxacillin equivalent to 50%,

100%, and 150% of the labeled amount as per the test method. The average % recovery of

Cefixime and Cloxacillin were calculated.

ACCEPTANCE CRITERIA

The mean % recovery of the Cefixime and Cloxacillin at each spike level should be not less

than 98.0% and not more than 102.0% for both the drugs separately.

OBSERVATION

Amount found

%Recovery = --- × 100 Amount added

The recovery results indicating that the test method has anacceptable level of accuracy. Refer

table: 5

Table-6: (i) Data of Accuracy for cefixime Concentration

% of spiked level

Amount added (ppm)

Amount found (ppm)

% Recovery

Statistical Analysis of % Recovery

50% Injection 1 20 19.85 99.25 MEAN 99.88

50% Injection 2 20 19.96 99.80

50% Injection 3 20 20.12 100.6 %RSD 0.67

100 % Injection 1 40 39.74 99.35 MEAN 99.81

100 % Injection 2 40 40.08 100.2

100% Injection 3 40 40.24 100.6 %RSD 0.399

150%

Injection 1 60 59.04 98.40 MEAN 99.19

150%

Injection 2 60 59.62 99.36

150%

(12)

(ii)Data of Accuracy for Cloxacillin Concentration

% of spiked level

Amount added (ppm)

Amount found (ppm)

% Recovery

Statistical Analysis of % Recovery

50% Injection 1 20 19.86 99.30 MEAN 99.46

50% Injection 2 20 19.98 99.90

50% Injection 3 20 19.84 99.20 %RSD 0.38

100 % Injection 1 40 39.54 98.85 MEAN 99.76

100 % Injection 2 40 39.82 99.55

100% Injection 3 40 39.96 99.9 %RSD 0.189

150% Injection 1 60 59.92 99.86 MEAN 100.0067

150% Injection 2 60 60.08 100.13

150%Injection3 60 60.02 100.03 %RSD 0.136

6. RUGGEDNESS OF TEST METHOD a) System to system variability

System to system variability study was conducted on different HPLC systems, under similar

conditions at different times. Six samples were prepared and each was analyzed as per test

method.

Comparison of both the results obtained on two different HPLC systems, shows that the assay

test method are rugged for System to system variability.

ACCEPTANCE CRITERIA

The % relative standard deviation of Cefixime and Cloxacillin from the six sample

preparations should be not more than 2.0%

The % assay of Cefixime and Cloxacillin should be between 98.0%-102.0%.

OBSERVATION: The % RSD was found within the limit. Ref tables: 3 &7.

b) column to column variability: Column to column variability study was conducted by using different columns. Six samples were prepared and each was analysed as per test

method.

ACCEPTANCE CRITERIA

The %RSD of Cefixime and Cloxacillin tablets should be NMT2.0%. The %assay of

Cefixime and Cloxacillin should be between 98.0% and 102.0% for individual drugs.

(13)
[image:13.595.101.494.92.242.2]

Table 7: Data of system to system variability (cefixime and cloxacillin) System-2

OBSERVATION: The results obtained by comparing with both two types were within limit.

7. ROBUSTNESS

a) Effect of variation of flow rate: A study was conducted to determine the effect of variation in flow rate. Standard solution prepared as per the test method was injected into the

HPLC system using flow rates, 1.0ml/min and 1.2ml/min. The system suitability parameters

were evaluated and found to be within the limits for 1.0ml/min and 1.2ml/min flow. Cefixime

and cloxacillin and was resolved from all other peaks and the retention times were

comparable with those obtained for mobile phase having flow rates 1.0ml/min.

ACCEPTANCE CRITERIA

The Tailing Factor of Cefixime and cloxacillin standards should be NMT 2.0 for Variation

in Flow.

Table: 8(i) Data for Effect of variation in flow rate (cefixime)

Flow 0.8 ml

Std Area Tailing factor

Flow 1.0 ml

Std Area Tailing factor

Flow 1.2 ml

Std Area Tailing factor

1139272 1.238915 1146923 1.251658 1152293 1.262464

1140892 1.230637 1143596 1.245435 1146923 1.251658

1136301 1.240858 1158293 1.262464 1147283 1..237018

1141067 1.238995 1147283 1.237018 1152490 1.239010

1136024 1.241073 1152490 1.239010 1139272 1.238915

Avg 1138711 1.236496 Avg 1149717 1.247117 Avg 1148852 1.245813

SD 2431.578 0.005254 SD 5754.015 0.010328 SD 7076.841 0.010984

%RSD 0.213538 0.424907 %RSD 0.500472 0.008282 %RSD 0.615992 0.00881712

S.NO Peak area Assay % of

Cefixime Peak area

Assay % of Cloxacillin

1 634360 98.65 1203625 99.98

2 634098 98.63 1202225 99.30

3 635696 98.86 1202840 98.60

4 633289 98.52 1204283 99.30

5 634147 98.63 1202735 98.55

6 633495 98.55 1203110 98.73

Mean 634180.8 98.64 1203136.3 99.07667

[image:13.595.39.565.539.688.2]
(14)

Table: 8(ii) Data for Effect of variation in flow rate (Cloxacillin)

Flow 0.8 ml

Std Area

Tailing factor

Flow 1.0 ml

Std Area

Tailing factor

Flow 1.2 ml

Std Area

Tailing factor

797564 1.099100 801690 1.122813 805783 1.121321

795138 1.103929 797631 1.112181 801690 1.122813

795685 1.111477 805783 1.121321 801496 1.124805

800569 1.117660 801496 1.124805 806432 1.123373

797049 1.119004 806432 1.123373 797564 1.099100

Avg 797201 1.110234 Avg 802606.4 1.120899 Avg 801593 1.118282

SD 2124.413 0.008622 SD 3590.034 0.00503 SD 3613.298 0.047969

%RSD 0.500472 0.77655 %RSD 0.447297 0.004488 %RSD 0.450203 0.965376

OBSERVATION: The tailing factor for Cefixime and cloxacillin was found to be within the limits. As shown in table 8.

8. LIMIT OF DETECTION AND QUANTITATION (LOD and LOQ)

Cefixime: From the linearity plot the LOD and LOQ are calculated LOD = 3.3 σ

S

3.3×2431.578

= --- = 0.25 31282

LOQ = 10 σ S

10×2431.578

= --- = 0.77 31282

Cloxacillin

LOD = 3.3 σ S

3.3×2124.413

= --- = 0.34 20193

LOQ = 10 σ S

10×2124.413

(15)

SOMMARY AND CONCLUSSION

The analytical method was developed by studying different parameters. First of all, maximum

absorbance was found to be at 288nm Cefixime for and 225nm for Cloxacillin.Common

wavelength will be 225nm and the peaks purity was excellent. Injection volume was selected

to be 20µl which gave a good peak area. The column used for study was Inertsil C18, ODS

chosen good peak shape. Ambient temperature was found to be suitable for the nature of drug

solution. The flow rate was fixed at 1.0ml/min because of good peak area, satisfactory

retention time and good resolution. Different ratios of mobile phase were studied, mobile

phase with ratio of 60:40 Methanol: Buffer was fixed due to good symmetrical peaks and for

good resolution. So this mobile phase was used for the proposed study.

The present recovery was found to be 98.0-101.50 was linear and precise over the same

range. Both system and method precision was found to be accurate and well within range.

Detection limit was found to be 0.25 Cefixime and 0.34 for Cloxacillin. Linearity study was,

correlation coefficient and curve fitting was found to be. The analytical method was found

linearity over the range of 20-80ppm of the target concentration for both the drugs. The

analytical passed both robustness and ruggedness tests. On both cases, relative standard

deviation was well satisfactory.

REFERENCE

1. www. Drugbank.ceftas.brands.html

2. http://www.drugs.com/cdi/cefixime-tablet.html

3. http://www.drugs.com/cdi/cloxacillin-tablet.html

4. .Pranaykumar Deekonda and Malladi Srinivas Reddy* ISSN 0975-413XCoden (usa):

pchhax der pharma chemica, 2014; 6 (2):31-37

5. Mehul Patel*, Jagdish Kakadiya, Nehal Shah development and validation of first order

derivative spectrophotometric method for simultaneous estimation of cefixime trihydrate

and moxifloxacin hydrochloride in combined tablet dosage form Vol 3|Issue 1|

2013|19-24.

6. Ajit R. Wankhede*1, Prashant Y. Mali2, Vikram Karne3, Anubha R. Khale1, C. S.

Magdum4Estimation of Cefixime and Cloxacillin in Tablet Dosage Form IJPBA, July –

(16)

7. Jigar a goswami1*, nehal j shah review of drugs and itsanalytical methods to treat allergic

rhinitis in combination with other drugs in different dosage form J Int J Pharm Pharm Sci,

5(3): 966-975

8. Bharat G. Chaudhari 1, Heena J. Patel1*Development and Validation of Dual

Wavelength UVSpectrophotometric Method for Simultaneous Estimation of Cefixime

and Dicloxacillin Sodium in Pharmaceutical Dosage Form 2(3):148-154

9. Pranaykumar Deekonda and Malladi Srinivas Reddy*Method development and validation

for the quantitative estimation ofcefixime and ofloxacin in Pharmaceutical preparation by

RP- HPLCDer Pharma Chemica, 2014; 6(2) :31-37

10.Dhoka, Madhura V.; Gawande, Vandana T.; Joshi, Pranav P. Validated HPTLC Method

for Determination of Cefixime Trihydrate and Erdosteine in Bulk and combined

Pharmaceutical Dosage FormEurasian Journal of Analytical Chemistry; 2013; 8(2) 99.

11.Tushar .G. Barot* , Kanhaiyalal Patidar, Nayan Kshartri And Niranjan Vyas

Development and Validation of LC Method for the Determination of Ampicillin and

Dicloxacillin in Pharmaceutical Formulation Using an Experimental Design E-Journal of

Chemistry 2009; 6(4): 955-964

12.Chigozie C. Ezeanokete[a], Kenneth Gerald Ngwoke[a]*, Festus Basden C. Okoye[a],

Patience O. Osadebe[b] Spectrophotometric Determination Of Ampicillin AndCloxacillin

In Pure And Fixed Dosage Forms Through Charge Transfer Complexation Eur. Chem.

Bull., 2013; 2(12): 1009-1012

13.Ajit R Wankhededevelopment And Validation Of Rp-Hplc Method For Simultaneous

Estimation Of Cefixime And Cloxacillin In Tablet Dosage Form Vol1,No3,2010

14.Angshuman Biswas* 1 , A Rajendiran2 , Inder Singh Rawat3 , Asit Kumar De1 , Ayan

Kumar Dey1Validated estimation procedure of Amoxycillin Trihydrate, Cloxacillin

sodium in Pharmaceutical Formulation by RP-HPLC MethodOctober – December 2010

RJPBCS, 2010; 1(4): 344

15.Ch. Ravikumar*1 , V. Valli Kumari2 , G. Tuljarani3method Development And Validation

For The Simultaneous Estimation Of Cefixime And Ofloxacin In A Pharmaceutical

Formulation By Rp-Hplc Method, 2014; 2(1): 1 - 14.

16.Tuani YT* 1 , Kissi FA 2 , Sackey J1,3, Gordon AJ 4 , Akanji O5Development and

Validation of RP-HPLC Method for the Simultaneous Determination of Ampicillin and

Cloxacillin in Oral Suspension Dosage FormInt. J. Pharm. Sci. Rev. Res., September –

Figure

Fig 1:  Chromatogram of standard
Fig 6   Got a peak for sample at an Rt of 2.953min for Cefixime and  3.523min for
Fig: 7(a) Linearity Plot (Concentration Vs Response) of Cefixime
Table 3: Data of Repeatability (System precision) for cefixime and cloxacillin
+2

References

Related documents

The ruggedness of test method was demonstrated by carrying out precision study in six preparations of sample on a single batch sample by different analysts. The results of

The precision of the analytical method was studied by analysis of multiple sampling of homogeneous sample. The precision expressed as standard deviation or

Standard stock solution was prepared by dissolving accurately weighed 100 mg pure drug in.. 100 ml of

For the determination of accuracy of the analytical procedures three identical dilutions of sample were spiked with different concentrations of standard solution

The final volume of the solution was made up to 100 ml with diluent, and the solution was filtered through whatmann filter paper no.41.The sample was analyzed

Carried out the method precision studies for Pioglitazone hydrochloride at 100% concentration of the test solution with a minimum of six determinations of individual

A Standard solution was prepared by using Simvastatin and Ezetimibe working standards as per test method and was injected Five times into the HPLC system.. The

A standard solution was prepared by using IPRA and Sitagliptin working standards as per test method and was injected six times into the HPLC system.. The system